The aim of this study was to quantify the stability
of fracture-implant complex in fractures after fixation. A total
of 15 patients with an undisplaced fracture of the femoral neck,
treated with either a dynamic hip screw or three cannulated hip
screws, and 16 patients with an AO31-A2 trochanteric fracture treated
with a dynamic hip screw or a Gamma Nail, were included. Radiostereometric
analysis was used at six weeks, four months and 12 months post-operatively
to evaluate shortening and rotation. Migration could be assessed in ten patients with a fracture of
the femoral neck and seven with a trochanteric fracture. By four
months post-operatively, a mean shortening of 5.4 mm (-0.04 to 16.1)
had occurred in the fracture of the femoral neck group and 5.0 mm
(-0.13 to 12.9) in the trochanteric fracture group. A wide range
of rotation occurred in both types of fracture. Right-sided trochanteric
fractures seem more rotationally stable than left-sided fractures. This prospective study shows that migration at the fracture site
occurs continuously during the first four post-operative months,
after which stabilisation occurs. This information may allow the
early recognition of patients at risk of
Aims. The primary aim of this study was to determine if delayed clavicular fixation results in a greater risk of operative complications and revision surgery. Patients and Methods. A retrospective case series was undertaken of all displaced clavicular fractures that underwent plate fixation over a ten-year period (2007 to 2017). Patient demographics, time to surgery, complications, and mode of failure were collected. Logistic regression was used to identify independent risk factors contributing towards operative complications. Receiver operating characteristic (ROC) curve analysis was used to determine if a potential ‘safe window’ exists from injury to delayed surgery. Propensity score matching was used to construct a case control study for comparison of risk. Results. A total of 259 patients were included in the analysis. Postoperative infection occurred in 3.9% of all patients (n = 10); the only variable associated was a greater time interval from injury to fixation (p = 0.001). Failed primary surgery requiring revision fixation was required in 7.7% of the cohort (n = 20), with smoking (p < 0.001), presence of a postoperative infection (p < 0.001), increasing age (p = 0.018), and greater time delay from injury to surgery (p = 0.015) identified as significant independent predictors on regression analysis. ROC analysis revealed that surgery beyond 96 days from injury increased the rate of major complications and revision surgery. Using a matched case cohort of cases before (n = 67) and after (n = 77) the ‘safe window’, the risk of postoperative infection increased (odds ratio (OR) 7.7, 95% confidence interval (CI) 1.9 to 62.9; p = 0.028),
The purpose of this study was to determine the weightbearing practice of operatively managed fragility fractures in the setting of publically funded health services in the UK and Ireland. The Fragility Fracture Postoperative Mobilisation (FFPOM) multicentre audit included all patients aged 60 years and older undergoing surgery for a fragility fracture of the lower limb between 1 January 2019 and 30 June 2019, and 1 February 2021 and 14 March 2021. Fractures arising from high-energy transfer trauma, patients with multiple injuries, and those associated with metastatic deposits or infection were excluded. We analyzed this patient cohort to determine adherence to the British Orthopaedic Association Standard, “all surgery in the frail patient should be performed to allow full weight-bearing for activities required for daily livingAims
Methods
The aim of this study was to describe the management and associated outcomes of patients sustaining a femoral hip periprosthetic fracture (PPF) in the UK population. This was a multicentre retrospective cohort study including adult patients who presented to 27 NHS hospitals with 539 new PPFs between 1 January 2018 and 31 December 2018. Data collected included: management strategy (operative and nonoperative), length of stay, discharge destination, and details of post-treatment outcomes (reoperation, readmission, and 30-day and 12-month mortality). Descriptive analysis by fracture type was performed, and predictors of PPF management and outcomes were assessed using mixed-effects logistic regression.Aims
Methods
This study evaluated variation in the surgical treatment of stable (A1) and unstable (A2) trochanteric hip fractures among an international group of orthopaedic surgeons, and determined the influence of patient, fracture, and surgeon characteristics on choice of implant (intramedullary nailing (IMN) versus sliding hip screw (SHS)). A total of 128 orthopaedic surgeons in the Science of Variation Group evaluated radiographs of 30 patients with Type A1 and A2 trochanteric hip fractures and indicated their preferred treatment: IMN or SHS. The management of Type A3 (reverse obliquity) trochanteric fractures was not evaluated. Agreement between surgeons was calculated using multirater kappa. Multivariate logistic regression models were used to assess whether patient, fracture, and surgeon characteristics were independently associated with choice of implant.Aims
Methods
To compare the functionality of adults with displaced mid-shaft clavicular fractures treated either operatively or nonoperatively and to compare the relative risk of nonunion and reoperation between the two groups. Based on specific eligibility criteria, 120 adults (median age 37.5 years (interquartile range (18 to 61)) and 84% males (n = 101)) diagnosed with an acute displaced mid-shaft fracture were recruited, and randomized to either the operative (n = 60) or nonoperative (n = 60) treatment group. This randomized controlled, partially blinded trial followed patients for 12 months following initial treatment. Functionality was assessed by the Constant score (CS) (assessor blinded to treatment) and Disability of the Arm, Shoulder and Hand (DASH) score. Clinical and radiological evaluation, and review of patient files for complications and reoperations, were added as secondary outcomes.Aims
Methods
A lack of supporting clinical studies have been published to determine the ideal length of intramedullary nail in fixation of trochanteric fractures of the hip. Nevertheless, there has been a trend to use shorter intramedullary nails for the internal fixation of trochanteric hip fractures. Our aim was to determine if the length of nail affected the outcome. We randomized 229 patients with a trochanteric hip fracture between two implants: a ‘standard’ nail of 220 mm and a shorter nail of 175 mm, which had decreased proximal angulation (4° vs 7°) and a reduced diameter at the level of the lesser trochanter. Patients were followed up for one year by a nurse blinded to the type of implant used to determine if there were differences in mobility and pain with two nail designs. Pain was assessed on a scale of 1 (none) to 8 (severe and constant) and mobility on a scale of 1 (full mobility) to 9 (immobile).Aims
Methods
Currently, periprosthetic fractures are excluded from the American Society for Bone and Mineral Research (ASBMR) definition of atypical femoral fracture (AFFs). This study aims to report on a series of periprosthetic femoral fractures (PFFs) that otherwise meet the criteria for AFFs. Secondary aims were to identify predictors of periprosthetic atypical femoral fractures (PAFFs) and quantify the complications of treatment. This was a retrospective case control study of consecutive patients with periprosthetic femoral fractures between 2007 and 2017. Two observers identified 16 PAFF cases (mean age 73.9 years (44 to 88), 14 female patients) and 17 typical periprosthetic fractures in patients on bisphosphonate therapy as controls (mean age 80.7 years (60 to 86, 13 female patients). Univariate and multivariate analysis was performed to identify predictors of PAFF. Management and complications were recorded.Aims
Patients and Methods
Displaced, comminuted acetabular fractures in the elderly are increasingly common, but there is no consensus on whether they should be treated non-surgically, surgically with open reduction and internal fixation (ORIF), or with acute total hip arthroplasty (THA). A combination of ORIF and acute THA, an approach called ’combined hip procedure’ (CHP), has been advocated and our aim was to compare the outcome after CHP or ORIF alone. A total of 27 patients with similar acetabular fractures (severe acetabular impaction with or without concomitant femoral head injury) with a mean age of 72.2 years (50 to 89) were prospectively followed for a minimum of two years. In all, 14 were treated with ORIF alone and 13 were treated with a CHP. Hip joint and patient survival were estimated. Operating times, blood loss, radiological outcomes, and patient-reported outcomes were assessed.Aims
Patients and Methods
To evaluate the effect of a single early high-dose vitamin D
supplement on fracture union in patients with hypovitaminosis D
and a long bone fracture. Between July 2011 and August 2013, 113 adults with a long bone
fracture were enrolled in a prospective randomised double-blind
placebo-controlled trial. Their serum vitamin D levels were measured
and a total of 100 patients were found to be vitamin D deficient
(<
20 ng/ml) or insufficient (<
30 ng/mL). These were then
randomised to receive a single dose of vitamin D3 orally
(100 000 IU) within two weeks of injury (treatment group, n = 50)
or a placebo (control group, n = 50). We recorded patient demographics,
fracture location and treatment, vitamin D level, time to fracture
union and complications, including vitamin D toxicity. Outcomes included union, nonunion or complication requiring an
early, unplanned secondary procedure. Patients without an outcome
at 15 months and no scheduled follow-up were considered lost to
follow-up. The Aims
Patients and Methods
To evaluate the outcomes of cemented total hip arthroplasty (THA)
following a fracture of the acetabulum, with evaluation of risk
factors and comparison with a patient group with no history of fracture. Between 1992 and 2016, 49 patients (33 male) with mean age of
57 years (25 to 87) underwent cemented THA at a mean of 6.5 years
(0.1 to 25) following acetabular fracture. A total of 38 had undergone
surgical fixation and 11 had been treated non-operatively; 13 patients
died at a mean of 10.2 years after THA (0.6 to 19). Patients were
assessed pre-operatively, at one year and at final follow-up (mean
9.1 years, 0.5 to 23) using the Oxford Hip Score (OHS). Implant
survivorship was assessed. An age and gender-matched cohort of THAs
performed for non-traumatic osteoarthritis (OA) or avascular necrosis
(AVN) (n = 98) were used to compare complications and patient-reported outcome
measures (PROMs).Aims
Patients and Methods
The Gamma nail is frequently used in unstable
peri-trochanteric hip fractures. We hypothesised that mechanical failure
of the Gamma nail was associated with inadequate proximal three-point
fixation. We identified a consecutive series of 299 Gamma nails
implanted in 299 patients over a five-year period, 223 of whom fulfilled
our inclusion criteria for investigation. The series included 61
men and 162 women with a mean age of 81 years (20 to 101). Their fractures
were classified according to the Modified Evans’ classification
and the quality of fracture reduction was graded. The technical
adequacy of three points of proximal fixation was recorded from
intra-operative fluoroscopic images, and technical inadequacy for
each point was defined. All patients were followed to final follow-up
and mechanical failures were identified. A multivariate statistical
analysis was performed, adjusting for confounders. A total of 16
failures (7.2%) were identified. The position of the lag screw relative
to the lateral cortex was the most important point of proximal fixation,
and when inadequate the failure rate was 25.8% (eight of 31: odds
ratio 7.5 (95% confidence interval 2.5 to 22.7), p <
0.001). Mechanical failure of the Gamma nail in peri-trochanteric femoral
fractures is rare (<
1%) when three-point proximal fixation is
achieved. However, when proximal fixation is inadequate, failure
rates increase. The strongest predictor of failure is positioning
the lateral end of the lag screw short of the lateral cortex. Adherence
to simple technical points minimises the risk of fixation failure
in this vulnerable patient group. Cite this article:
We compared a new fixation system, the Targon
Femoral Neck (TFN) hip screw, with the current standard treatment of
cannulated screw fixation. This was a single-centre, participant-blinded,
randomised controlled trial. Patients aged 65 years and over with
either a displaced or undisplaced intracapsular fracture of the
hip were eligible. The primary outcome was the risk of revision
surgery within one year of fixation. A total of 174 participants were included in the trial. The absolute
reduction in risk of revision was of 4.7% (95% CI 14.2 to 22.5)
in favour of the TFN hip screw (chi-squared test, p = 0.741), which
was less than the pre-specified level of minimum clinically important
difference. There were no significant differences in any of the
secondary outcome measures. We found no evidence of a clinical difference in the risk of
revision surgery between the TFN hip screw and cannulated screw
fixation for patients with an intracapsular fracture of the hip. Cite this article:
The spiral blade modification of the Dynamic
Hip Screw (DHS) was designed for superior biomechanical fixation
in the osteoporotic femoral head. Our objective was to compare clinical
outcomes and in particular the incidence of loss of fixation. In a series of 197 consecutive patients over the age of 50 years
treated with DHS-blades (blades) and 242 patients treated with conventional
DHS (screw) for AO/OTA 31.A1 or A2 intertrochanteric fractures were
identified from a prospectively compiled database in a level 1 trauma
centre. Using propensity score matching, two groups comprising 177
matched patients were compiled and radiological and clinical outcomes
compared. In each group there were 66 males and 111 females. Mean
age was 83.6 (54 to 100) for the conventional DHS group and 83.8
(52 to 101) for the blade group. Loss of fixation occurred in two blades and 13 DHSs. None of
the blades had observable migration while nine DHSs had gross migration
within the femoral head before the fracture healed. There were two
versus four implant cut-outs respectively and one side plate pull-out
in the DHS group. There was no significant difference in mortality
and eventual walking ability between the groups. Multiple logistic
regression suggested that poor reduction (odds ratio (OR) 11.49,
95% confidence intervals (CI) 1.45 to 90.9, p = 0.021) and fixation
by DHS (OR 15.85, 95%CI 2.50 to 100.3, p = 0.003) were independent
predictors of loss of fixation. The spiral blade design may decrease the risk of implant migration
in the femoral head but does not reduce the incidence of cut-out
and reoperation. Reduction of the fracture is of paramount importance
since poor reduction was an independent predictor for loss of fixation
regardless of the implant being used. Cite this article:
Hip fracture is a global public health problem.
The National Hip Fracture Database provides a framework for service evaluation
in this group of patients in the United Kingdom, but does not collect
patient-reported outcome data and is unable to provide meaningful
data about the recovery of quality of life. We report one-year patient-reported outcomes of a prospective
cohort of patients treated at a single major trauma centre in the
United Kingdom who sustained a hip fracture between January 2012
and March 2014. There was an initial marked decline in quality of life from baseline
measured using the EuroQol 5 Dimensions score (EQ-5D). It was followed
by a significant improvement to 120 days for all patients. Although
their quality of life improved during the year after the fracture,
it was still significantly lower than before injury irrespective
of age group or cognitive impairment (mean reduction EQ-5D 0.22;
95% confidence interval (CI) 0.17 to 0.26). There was strong evidence
that quality of life was lower for patients with cognitive impairment.
There was a mean reduction in EQ-5D of 0.28 (95% CI 0.22 to 0.35)
in patients <
80 years of age. This difference was consistent
(and fixed) throughout follow-up. Quality of life does not improve
significantly during recovery from hip fracture in patients over
80 years of age (p = 0.928). Secondary measures of function showed
similar trends. Hip fracture marks a step down in the quality of life of a patient:
it accounts for approximately 0.22 disability adjusted life years
in the first year after fracture. This is equivalent to serious
neurological conditions for which extensive funding for research
and treatment is made available. Cite this article:
We evaluated the cost and consequences of proximal femoral fractures requiring further surgery because of complications. The data were collected prospectively in a standard manner from all patients with a proximal femoral fracture presenting to the trauma unit at the John Radcliffe Hospital over a five-year period. The total cost of treatment for each patient was calculated by separating it into its various components. The risk factors for the complications that arose, the location of their discharge and the mortality rates for these patients were compared to those of a matched control group. There were 2360 proximal femoral fractures in 2257 patients, of which 144 (6.1%) required further surgery. The mean cost of treatment in patients with complications was £18 709 (£2606.30 to £60.827.10), compared with £8610 (£918.54 to £45 601.30) for uncomplicated cases (p <
0.01), with a mean length of stay of 62.8 (44.5 to 79.3) and 32.7 (23.8 to 35.0) days, respectively. The probability of mortality after one month in these cases was significantly higher than in the control group, with a mean survival of 209 days, compared with 496 days for the controls. Patients with complications were statistically less likely to return to their own home (p <
0.01). Greater awareness and understanding are required to minimise the complications of proximal femoral fractures and consequently their cost.
In a randomised trial involving 598 patients
with 600 trochanteric fractures of the hip, the fractures were treated with
either a sliding hip screw (n = 300) or a Targon PF intramedullary
nail (n = 300). The mean age of the patients was 82 years (26 to
104). All surviving patients were reviewed at one year with functional
outcome assessed by a research nurse blinded to the treatment used.
The intramedullary nail was found to have a slightly increased mean
operative time (46 minutes ( In summary, both implants produced comparable results but there
was a tendency to better return of mobility for those treated with
the intramedullary nail.
We treated 47 patients with a mean age of 57 years (22 to 88) who had a proximal humeral fracture in which there was a severe varus deformity, using a standard operative protocol of anatomical reduction, fixation with a locking plate and supplementation by structural allografts in unstable fractures. The functional and radiological outcomes were reviewed. At two years after operation the median Constant score was 86 points and the median Disabilities of the Arm, Shoulder and Hand score 17 points. Seven of the patients underwent further surgery, two for
In a series of 450 patients over 70 years of age with displaced fractures of the femoral neck sustained between 1995 and 1997 treatment was randomised either to internal fixation or replacement. Depending on age and level of activity the latter was either a total hip replacement or a hemiarthroplasty. Patients who were confused or bed-ridden were excluded, as were those with rheumatoid arthritis. At ten years there were 99 failures (45.6%) after internal fixation compared with 17 (8.8%) after replacement. The rate of mortality was high at 75% at ten years, and was the same in both groups at all times. Patient-reported pain and function were similar in both groups at five and ten years. Those with successfully healed fractures had more hip pain and reduction of mobility at four months compared with patients with an uncomplicated replacement, and they never attained a better outcome than the latter patients regarding pain or function. Primary replacement gave reliable long-term results in patients with a displaced fracture of the femoral neck.
We conducted a randomised controlled trial to compare external fixation of trochanteric fractures of the femur with the more costly option of the sliding hip screw. Patients in both groups were matched for age (mean 67 years, 50 to 100) and gender. We excluded all pathological fractures, patients presenting at more than one week, fractures with subtrochanteric extension or reverse obliquity, multiple fractures or any bone and joint disease interfering with rehabilitation. The interval between injury and operation, the duration of surgery, the amount of blood loss, the length of hospital stay and the cost of treatment were all significantly higher in the sliding hip screw group (p <
0.05). The time to union, range of movement, mean Harris hip scores and Western Ontario and McMaster University knee scores were comparable at six months. The number of patients showing shortening or malrotation was too small to show a significant difference between the groups. Pin-track infection occurred in 18 patients (60%) treated with external fixation, whereas there was a single case of wound infection (3.3%) in the sliding hip screw group.