In a prospective observational study we compared the two-year outcome of lumbar fusion by a simple technique using translaminar screws (n = 57) with a more extensive method using transforaminal lumbar interbody fusion and pedicular screw
The purpose of this study was to determine whether
it would be feasible to use oblique lumbar interbody
We reviewed seven children with torticollis due to refractory atlanto-axial rotatory
We reviewed the outcome of a retrospective case series of eight patients with atlantoaxial instability who had been treated by percutaneous anterior transarticular screw
We have reviewed 81 patients with fractures of the odontoid process treated between May 1983 and July 1997, by anterior screw
We analysed the complications encountered in 102 consecutive patients who had posterolateral lumbosacral fusion performed with transpedicular screw and rod
The purpose of this study was to determine whether
patients with a burst fracture of the thoracolumbar spine treated
by short segment pedicle screw
We investigated the effects of hydroxyapatite (HA) coating on the purchase of pedicle screws. A total of 23 consecutive patients undergoing lumbar fusion was randomly assigned to one of three treatment groups. The first received uncoated stainless-steel screws, the second screws which were partly coated with HA, and the third screws which were fully coated. The insertion torque was recorded. After 11 to 16 months, 21 screws had been extracted. The extraction torque was recorded. Radiographs were taken to assess fusion and to detect loosening of the screws. At removal, the extraction torques exceeded the upper limit of the torque wrench (600 Ncm) for many HA-coated screws. The calculated mean extraction torque was 29 ± 36 Ncm for the uncoated group, 447 ± 114 Ncm for the partly-coated group and 574 ± 52 Ncm for the fully-coated group. There were significant differences between all three groups (p <
0.001). There were more radiolucent zones surrounding the uncoated screws than the HA-coated screws (p <
0.001). HA coating of pedicle screws resulted in improved
In patients with osteoporosis there is always
a strong possibility that pedicle screws will loosen. This makes
it difficult to select the appropriate osteoporotic patient for
a spinal fusion. The purpose of this study was to determine the
correlation between bone mineral density (BMD) and the magnitude
of torque required to insert a pedicle screw. To accomplish this,
181 patients with degenerative disease of the lumbar spine were
studied prospectively. Each underwent dual-energy x-ray absorptiometry
(DEXA) and intra-operative measurement of the torque required to
insert each pedicle screw. The levels of torque generated in patients
with osteoporosis and osteopenia were significantly lower than those
achieved in normal patients. Positive correlations were observed between
BMD and T-value at the instrumented lumbar vertebrae, mean BMD and
mean T-value of the lumbar vertebrae, and mean BMD and mean T-value
of the proximal femur. The predictive torque (Nm) generated during pedicle
screw insertion was [-0.127 + 1.62 × (BMD at the corresponding lumbar
vertebrae)], as measured by linear regression analysis. The positive
correlation between BMD and the maximum torque required to insert
a pedicle screw suggests that pre-operative assessment of BMD may
be useful in determining the ultimate strength of
Aims. The optimal procedure for the treatment of ossification of the posterior longitudinal ligament (OPLL) remains controversial. The aim of this study was to compare the outcome of anterior cervical ossified posterior longitudinal ligament en bloc resection (ACOE) with posterior laminectomy and fusion with bone graft and internal
Aims. The timing of surgical fixation in spinal fractures is a contentious topic. Existing literature suggests that early stabilization leads to reduced morbidity, improved neurological outcomes, and shorter hospital stay. However, the quality of evidence is low and equivocal with regard to the safety of early
Aims. It is uncertain whether instrumented spinal fixation in nonambulatory children with neuromuscular scoliosis should finish at L5 or be extended to the pelvis. Pelvic
The aim of this study was to reassess the rate of neurological, psoas-related, and abdominal complications associated with L4-L5 lateral lumbar interbody fusion (LLIF) undertaken using a standardized preoperative assessment and surgical technique. This was a multicentre retrospective study involving consecutively enrolled patients who underwent L4-L5 LLIF by seven surgeons at seven institutions in three countries over a five-year period. The demographic details of the patients and the details of the surgery, reoperations and complications, including femoral and non-femoral neuropraxia, thigh pain, weakness of hip flexion, and abdominal complications, were analyzed. Neurological and psoas-related complications attributed to LLIF or posterior instrumentation and persistent symptoms were recorded at one year postoperatively.Aims
Methods
The aim of this study was to investigate the incidence and characteristics of instrumentation failure (IF) after total en bloc spondylectomy (TES), and to analyze risk factors for IF. The medical records from 136 patients (65 male, 71 female) with a mean age of 52.7 years (14 to 80) who underwent TES were retrospectively reviewed. The mean follow-up period was 101 months (36 to 232). Analyzed factors included incidence of IF, age, sex, BMI, history of chemotherapy or radiotherapy, tumour histology (primary or metastasis; benign or malignant), surgical approach (posterior or combined), tumour location (thoracic or lumbar; junctional or non-junctional), number of resected vertebrae (single or multilevel), anterior resection line (disc-to-disc or intravertebra), type of bone graft (autograft or frozen autograft), cage subsidence (CS), and local alignment (LA). A survival analysis of the instrumentation was performed, and relationships between IF and other factors were investigated using the Cox regression model.Aims
Methods
The aim of this study was to compare the clinical and radiological outcomes of patients with early-onset scoliosis (EOS), who had undergone spinal fusion after distraction-based spinal growth modulation using either traditional growing rods (TGRs) or magnetically controlled growing rods (MCGRs). We undertook a retrospective review of skeletally mature patients who had undergone fusion for an EOS, which had been previously treated using either TGRs or MCGRs. Measured outcomes included sequential coronal T1 to S1 height and major curve (Cobb) angle on plain radiographs and any complications requiring unplanned surgery before final fusion.Aims
Methods
This study addressed two questions: first, does surgical correction of an idiopathic scoliosis increase the volume of the rib cage, and second, is it possible to evaluate the change in lung function after corrective surgery for adolescent idiopathic scoliosis (AIS) using biplanar radiographs of the ribcage with 3D reconstruction? A total of 45 patients with a thoracic AIS which needed surgical correction and fusion were included in a prospective study. All patients underwent pulmonary function testing (PFT) and low-dose biplanar radiographs both preoperatively and one year after surgery. The following measurements were recorded: forced vital capacity (FVC), slow vital capacity (SVC), and total lung capacity (TLC). Rib cage volume (RCV), maximum rib hump, main thoracic curve Cobb angle (MCCA), medial-lateral and anteroposterior diameter, and T4-T12 kyphosis were calculated from 3D reconstructions of the biplanar radiographs.Aims
Methods
The outcome following the development of neurological complications after corrective surgery for scoliosis varies from full recovery to a permanent deficit. This study aimed to assess the prognosis and recovery of major neurological deficits in these patients, and to determine the risk factors for non-recovery, at a minimum follow-up of two years. A major neurological deficit was identified in 65 of 8,870 patients who underwent corrective surgery for scoliosis, including eight with complete paraplegia and 57 with incomplete paraplegia. There were 23 male and 42 female patients. Their mean age was 25.0 years (SD 16.3). The aetiology of the scoliosis was idiopathic (n = 6), congenital (n = 23), neuromuscular (n = 11), neurofibromatosis type 1 (n = 6), and others (n = 19). Neurological function was determined by the American Spinal Injury Association (ASIA) impairment scale at a mean follow-up of 45.4 months (SD 17.2). the patients were divided into those with recovery and those with no recovery according to the ASIA scale during follow-up.Aims
Methods
To report the outcome of spinal deformity correction through anterior spinal fusion in wheelchair-bound patients with myelomeningocele. We reviewed 12 consecutive patients (7M:5F; mean age 12.4 years (9.2 to 16.8)) including demographic details, spinopelvic parameters, surgical correction, and perioperative data. We assessed the impact of surgery on patient outcomes using the Spina Bifida Spine Questionnaire and a qualitative questionnaire.Aims
Methods
To evaluate the perioperative complications associated with total en bloc spondylectomy (TES) in patients with spinal tumours, based on the extent and level of tumour resection. In total, 307 patients who underwent TES in a single centre were reviewed retrospectively. There were 164 male and 143 female patients with a mean age at the time of surgery of 52.9 years (SD 13.3). A total of 225 patients were operated on for spinal metastases, 34 for a malignant primary tumour, 41 for an aggressive benign tumour, and seven with a primary of unknown origin. The main lesion was located in the thoracic spine in 213, and in the lumbar spine in 94 patients. There were 97 patients who underwent TES for more than two consecutive vertebrae.Aims
Methods
Aims. The aim of this study was to compare the peak pull-out force
(PPF) of pedicle-lengthening screws (PLS) and traditional pedicle
screws (TPS) using instant and cyclic fatigue testing. Materials and Methods. A total of 60 lumbar vertebrae were divided into six groups:
PLS submitted to instant pull-out and fatigue-resistance testing
(groups A1 and A2, respectively), TPS submitted to instant pull-out
and fatigue-resistance testing (groups B1 and B2, respectively)
and PLS augmented with 2 ml polymethylmethacrylate, submitted to
instant pull-out and fatigue-resistance testing (groups C1 and C2,
respectively). The PPF and normalized PPF (PPFn) for bone mineral density
(BMD) were compared within and between all groups. Results. In all groups, BMD was significantly correlated with PPF (r =
0.83, p < 0.001). The PPFn in A1 was significantly less than
in B1 (p = 0.006) and C1 (p = 0.002). The PPFn of A2 was significantly
less than in B2 (p < 0.001) and C2 (p < 0.001). The PPFn in
A1, B1, and C1 was significantly greater than in A2 (p = 0.002),
B2 (p = 0.027), and C2 (p = 0.003). There were no significant differences
in PPFn between B1 and C1, or between B2 and C2. Conclusion. Pedicle lengthening screws with cement augmentation can provide
the same