We investigated a new intramedullary locking
nail that allows the distal interlocking screws to be locked to
the nail. We compared fixation using this new implant with fixation
using either a conventional nail or a locking plate in a laboratory
simulation of an osteoporotic fracture of the distal femur. A total
of 15 human cadaver femora were used to simulate an AO 33-A3 fracture
pattern. Paired specimens compared fixation using either a locking
or non-locking retrograde nail, and using either a locking retrograde
nail or a locking plate. The constructs underwent cyclical loading
to simulate single-leg stance up to 125 000 cycles. Axial and torsional
stiffness and displacement, cycles to failure and modes of failure
were recorded for each specimen. When compared with locking plate
constructs, locking nail constructs had significantly longer mean
fatigue life (75 800 cycles ( The new locking retrograde femoral nail showed better stiffness
and fatigue life than locking plates, and superior fatigue life
to non-locking nails, which may be advantageous in elderly patients. Cite this article:
Filling the empty holes in peri-articular locking
plates may improve the
Management of bisphosphonate-associated subtrochanteric
fractures remains opinion- or consensus-based. There are limited
data regarding the outcomes of this fracture. We retrospectively reviewed 33 consecutive female patients with
a mean age of 67.5 years (47 to 91) who were treated surgically
between May 2004 and October 2009. The mean follow-up was 21.7 months
(0 to 53). Medical records and radiographs were reviewed to determine
the post-operative ambulatory status, time to clinical and radiological
union and post-fixation complications such as implant failure and
need for second surgery. The predominant fixation method was with an extramedullary device
in 23 patients. 25 (75%) patients were placed on wheelchair mobilisation
or no weight-bearing initially. The mean time to full weight-bearing
was 7.1 months (2.2 to 29.7). The mean time for fracture site pain
to cease was 6.2 months (1.2 to 17.1). The mean time to radiological
union was 10.0 months (2.2 to 27.5). Implant failure was seen in
seven patients (23%, 95 confidence interval (CI) 11.8 to 40.9).
Revision surgery was required in ten patients (33%, 95 CI 19.2 to
51.2). A large proportion of the patients required revision surgery
and suffered implant failure. This fracture is associated with slow
healing and prolonged post-operative immobility. Cite this article: