The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 10 | Pages 1410 - 1415
1 Oct 2010
Castagna A Randelli M Garofalo R Maradei L Giardella A Borroni M

Total shoulder replacement is a successful procedure for degenerative or some inflammatory diseases of the shoulder. However, fixation of the glenoid seems to be the main weakness with a high rate of loosening. The results using all-polyethylene components have been better than those using metal-backed components. We describe our experience with 35 consecutive total shoulder replacements using a new metal-backed glenoid component with a mean follow-up of 75.4 months (48 to 154). Our implant differs from others because of its mechanism of fixation. It has a convex metal-backed bone interface and the main stabilising factor is a large hollow central peg. The patients were evaulated with standard radiographs and with the Constant Score, the Simple Shoulder Test and a visual analogue scale. All the scores improved and there was no loosening, no polyethylene-glenoid disassembly and no other implant-related complications. We conclude that a metal-backed glenoid component is a good option in total shoulder replacement with no worse results than of those using a cemented all-polyethylene prosthesis


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 2 | Pages 235 - 241
1 Feb 2010
van der Lugt JCT Valstar ER Witvoet-Braam SW Nelissen RGHH

Mechanical loosening which begins with early-onset migration of the prosthesis is the major reason for failure of the Souter-Strathclyde elbow replacement. In a prospective study of 18 Souter-Strathclyde replacements we evaluated the patterns of migration using roentgen stereophotogrammetric analysis. We had previously reported the short-term results after a follow-up of two years which we have now extended to a mean follow-up of 8.2 years (1 to 11.3). Migration was assessed along the co-ordinal axes and overall micromovement was expressed as the maximum total point movement. The alignment of the prosthesis and the presence of radiolucent lines were examined on conventional standardised radiographs. All the humeral components showed increased and variable patterns of migration at the extended follow-up and four humeral components were revised. The maximum total point movement at two years in the revised prostheses was 1.8 mm (. sd. 1.0) and in the non-revised 0.7 mm (. sd. 0.5, p = 0.01). Most humeral components migrated into external rotation resulting in an anterior and varus tilt. The ulnar components remained stable


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 1 | Pages 75 - 81
1 Jan 2009
Cil A Veillette CJH Sanchez-Sotelo J Sperling JW Schleck C Cofield RH

Between 1976 and 2004, 38 revision arthroplasties (35 patients) were performed for aseptic loosening of the humeral component. The mean interval from primary arthroplasty to revision was 7.1 years (0.4 to 16.6). A total of 35 shoulders (32 patients) were available for review at a mean follow-up of seven years (2 to 19.3). Pre-operatively, 34 patients (97%) had moderate or severe pain; at final follow-up, 29 (83%) had no or only mild pain (p < 0.0001). The mean active abduction improved from 88° to 107° (p < 0.01); and the mean external rotation from 37° to 46° (p = 0.27). Excellent or satisfactory results were achieved in 25 patients (71%) according to the modified Neer rating system. Humeral components were cemented in 29, with ingrowth implants used in nine cases. There were 19 of standard length and 17 were longer (two were custom replacements and are not included). Bone grafting was required for defects in 11 humeri. Only two glenoid components were left unrevised. Intra-operative complications included cement extrusion in eight cases, fracture of the shaft of the humerus is two and of the tuberosity in four. There were four re-operations, one for recurrent humeral loosening, with 89% survival free of re-operations at ten years. Revision surgery for aseptic loosening of the humeral component provides reliable pain relief and modest improvement of movement, although there is a substantial risk of intra-operative complications. Revision to a total shoulder replacement gives better results than to a hemiarthroplasty


The Journal of Bone & Joint Surgery British Volume
Vol. 84-B, Issue 1 | Pages 83 - 87
1 Jan 2002
Nagels J Valstar ER Stokdijk M Rozing PM

The incidence of loosening of a cemented glenoid componentin total shoulder arthroplasty, detected by means of radiolucent lines or positional shift of the component on true anteroposterior radiographs, has been reported to be between 0% and 44%. Radiolucent lines are, however, difficult to detect and to interpret because of the mobility of the shoulder girdle and the obliquity of the glenoid which hinder standardisation of radiographs. We examined radiolucencies around cemented glenoid components in 48 patients, with a mean follow-up of 5.3 years, and found progressive changes to be present predominantly at the inferior pole of the component. This may hold a clue for the mechanism of loosening of this implant. In five patients we performed an additional analysis of loosening of the glenoid component using digital roentgen stereophotogrammetric analysis (RSA). After three years, three of the five implants had loosened (migration 1.2 to 5.5 mm). In only one, with gross loosening, were the radiological signs consistent with the RSA findings. When traditional radiographs are used for assessment, the rate of early loosening is underestimated. We recommend that RSA be used for this


The Journal of Bone & Joint Surgery British Volume
Vol. 88-B, Issue 3 | Pages 341 - 344
1 Mar 2006
Tanaka N Sakahashi H Ishii S Kudo H

The purpose of this study was to assess the long-term results (more than ten years) of two types of cemented ulnar component with type-5 Kudo total elbow arthroplasty in a consecutive series of 56 patients (60 elbows) with rheumatoid arthritis, and to compare the results in elbows above and below a Larsen grade IV. There was no radiolucency around the humeral component. Patients in whom a metal-backed ulnar component and a porous-coated stem were used had better clinical results and significantly less progression of radiolucent line formation around the ulnar component. They also had a significantly better long-term survival than patients with an all-polyethylene ulnar component. The clinical results of arthroplasty using all-polyethylene ulnar components were inferior, regardless of the degree of joint destruction. We conclude that the type-5 Kudo total elbow arthroplasty with cementless fixation of the porous-coated humeral component and cemented fixation of a metal-backed ulnar component is acceptable and well-tolerated by rheumatoid patients


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 6 | Pages 757 - 761
1 Jun 2009
Nuttall D Haines JF Trail IA

In a prospective study between 2000 and 2005, 22 patients with primary osteoarthritis of the shoulder had a total shoulder arthroplasty with a standard five-pegged glenoid component, 12 with non-offset humeral head and ten with offset humeral head components. Over a period of 24 months the relative movement of the glenoid component with respect to the scapula was measured using radiostereometric analysis. Nine glenoids needed reaming for erosion. There was a significant increase in rotation about all three axes with time (p < 0.001), the largest occurring about the longitudinal axis (anteversion-retroversion), with mean values of 3.8° and 1.9° for the non-offset and offset humeral head eroded subgroups, respectively. There was also a significant difference in rotation about the anteversion-retroversion axis (p = 0.01) and the varus-valgus (p < 0.001) z-axis between the two groups. The offset humeral head group reached a plateau at early follow-up with rotation about the z-axis, whereas the mean of the non-offset humeral head group at 24 months was three times greater than that of the offset group accounting for the highly significant difference between them


The Journal of Bone & Joint Surgery British Volume
Vol. 86-B, Issue 6 | Pages 856 - 860
1 Aug 2004
Rahme H Mattsson P Larsson S

We studied the stability of cemented all-polyethylene keeled glenoid components by radiostereometric analysis (RSA) in 16 shoulders which had received a total shoulder replacement. There were 14 women (one bilateral) and one man with a mean age of 64 years. The diagnosis was osteoarthritis in eight and rheumatoid arthritis in seven. Two of the shoulders were excluded from the RSA study because of loosening of the tantalum markers. Three tantalum markers were inserted in the glenoid socket, two in the coracoid process and two in the acromion. The polyethylene keeled glenoid component was marked with three to five tantalum markers. Conventional radiological and RSA examinations were carried out at five to seven days, at four months and at one and two years after operation. Radiolucent lines were found in all except three shoulders. Migration was most pronounced in the distal direction and exceeded 1 mm in four shoulders. In ten shoulders rotation exceeded 2° in one or more axes with retroversion/anteversion being most common. No correlation was found between migration and the presence of radiolucencies on conventional radiographs


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 2 | Pages 210 - 216
1 Feb 2011
Young A Walch G Boileau P Favard L Gohlke F Loew M Molé D

We report the long-term clinical and radiological outcomes of the Aequalis total shoulder replacement with a cemented all-polyethylene flat-back keeled glenoid component implanted for primary osteoarthritis between 1991 and 2003 in nine European centres. A total of 226 shoulders in 210 patients were retrospectively reviewed at a mean of 122.7 months (61 to 219) or at revision. Clinical outcome was assessed using the Constant score, patient satisfaction score and range of movement. Kaplan-Meier survivorship analysis was performed with glenoid revision for loosening and radiological glenoid loosening (. sd. ) as endpoints. The Constant score was found to improve from a mean of 26.8 (. sd. 10.3) pre-operatively to 57.6 (. sd. 20.0) post-operatively (p < 0.001). Active forward flexion improved from a mean of 85.3° (. sd. 27.4) pre-operatively to 125° (. sd. 37.3) postoperatively (p < 0.001). External rotation improved from a mean of 7° (. sd. 6.5) pre-operatively to 30.3° (. sd. 21.8°) post-operatively (p < 0.001). Survivorship with revision of the glenoid component as the endpoint was 99.1% at five years, 94.5% at ten years and 79.4% at 15 years. Survivorship with radiological loosening as the endpoint was 99.1% at five years, 80.3% at ten years and 33.6% at 15 years. Younger patient age and the curettage technique for glenoid preparation correlated with loosening. The rate of glenoid revision and radiological loosening increased with duration of follow-up, but not until a follow-up of five years. Therefore, we recommend that future studies reporting radiological outcomes of new glenoid designs should report follow-up of at least five to ten years


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 10 | Pages 1382 - 1389
1 Oct 2012
Sewell MD Kang SN Al-Hadithy N Higgs DS Bayley I Falworth M Lambert SM

There is little information about the management of peri-prosthetic fracture of the humerus after total shoulder replacement (TSR). This is a retrospective review of 22 patients who underwent a revision of their original shoulder replacement for peri-prosthetic fracture of the humerus with bone loss and/or loose components. There were 20 women and two men with a mean age of 75 years (61 to 90) and a mean follow-up 42 months (12 to 91): 16 of these had undergone a previous revision TSR. Of the 22 patients, 12 were treated with a long-stemmed humeral component that bypassed the fracture. All their fractures united after a mean of 27 weeks (13 to 94). Eight patients underwent resection of the proximal humerus with endoprosthetic replacement to the level of the fracture. Two patients were managed with a clam-shell prosthesis that retained the original components. The mean Oxford shoulder score (OSS) of the original TSRs before peri-prosthetic fracture was 33 (14 to 48). The mean OSS after revision for fracture was 25 (9 to 31). Kaplan-Meier survival using re-intervention for any reason as the endpoint was 91% (95% confidence interval (CI) 68 to 98) and 60% (95% CI 30 to 80) at one and five years, respectively. There were two revisions for dislocation of the humeral head, one open reduction for modular humeral component dissociation, one internal fixation for nonunion, one trimming of a prominent screw and one re-cementation for aseptic loosening complicated by infection, ultimately requiring excision arthroplasty. Two patients sustained nerve palsies. Revision TSR after a peri-prosthetic humeral fracture associated with bone loss and/or loose components is a salvage procedure that can provide a stable platform for elbow and hand function. Good rates of union can be achieved using a stem that bypasses the fracture. There is a high rate of complications and function is not as good as with the original replacement


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 5 | Pages 656 - 660
1 May 2010
Ikävalko M Tiihonen R Skyttä ET Belt EA

Between 1982 and 1997, 403 consecutive patients (522 elbows) with rheumatoid arthritis underwent Souter-Strathclyde total elbow replacement. By the end of 2007, there had been 66 revisions for aseptic loosening in 60 patients. The mean time of follow-up was 10.6 years (0 to 25) The survival rates at five-, ten, 15 and 19 years were 96% (95%, confidence interval (CI) 95 to 98), 89% (95% CI 86 to 92), 83% (95% CI 78 to 87), and 77% (95% CI 69 to 85), respectively. The small and medium-sized short-stemmed primary humeral components had a 5.6-fold and 3.6-fold risk of revision for aseptic loosening respectively, compared to the medium-sized long-stemmed component. The small and medium-sized all-polyethylene ulnar components had respectively a 28.2-fold and 8.4-fold risk of revision for aseptic loosening, compared to the metal-backed ulnar components. The use of retentive ulnar components was not associated with an increased risk of aseptic loosening compared to non-retentive implants


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 9 | Pages 1240 - 1246
1 Sep 2011
Melis B DeFranco M Lädermann A Molé D Favard L Nérot C Maynou C Walch G

Radiological changes and differences between cemented and uncemented components of Grammont reverse shoulder arthroplasties (DePuy) were analysed at a mean follow-up of 9.6 years (8 to 12). Of 122 reverse shoulder arthroplasties implanted in five shoulder centres between 1993 and 2000, a total of 68 (65 patients) were available for study. The indications for reversed shoulder arthroplasty were cuff tear arthropathy in 48 shoulders, revision of shoulder prostheses of various types in 11 and massive cuff tear in nine. The development of scapular notching, bony scapular spur formation, heterotopic ossification, glenoid and humeral radiolucencies, stem subsidence, radiological signs of stress shielding and resorption of the tuberosities were assessed on standardised true anteroposterior and axillary radiographs. A scapular notch was observed in 60 shoulders (88%) and was associated with the superolateral approach (p = 0.009). Glenoid radiolucency was present in 11 (16%), bony scapular spur and/or ossifications in 51 (75%), and subsidence of the stem and humeral radiolucency in more than three zones were present in three (8.8%) and in four (11.8%) of 34 cemented components, respectively, and in one (2.9%) and two (5.9%) of 34 uncemented components, respectively. Radiological signs of stress shielding were significantly more frequent with uncemented components (p < 0.001), as was resorption of the greater (p < 0.001) and lesser tuberosities (p = 0.009)


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 12 | Pages 1594 - 1600
1 Dec 2009
Khan A Bunker TD Kitson JB

There are no long-term published results on the survival of a third-generation cemented total shoulder replacement. We describe a clinical and radiological study of the Aequalis total shoulder replacement for a minimum of ten years. Between September 1996 and May 1998, 39 consecutive patients underwent a primary cemented total shoulder replacement using this prosthesis. Data were collected prospectively on all patients each year, for a minimum of ten years, or until death or failure of the prosthesis. At a follow-up of at least ten years, 12 patients had died with the prosthesis intact and two had emigrated, leaving 25 available for clinical review. Of these, 13 had rheumatoid arthritis and 12 osteoarthritis. One refused radiological review leaving 24 with fresh radiographs. Survivorship at ten years was 100% for the humeral component and 92% for the glenoid component. The incidence of lucent lines was low. No humeral component was thought to be at risk and only two glenoid components. The osteoarthritic group gained a mean 65° in forward flexion and their Constant score improved by a mean 41.4 points (13 to 55). The rheumatoid group gained a mean of 24° in flexion and their Constant score improved by 29.4 points. This difference may have been due to failure of the rotator cuff in 75% of the patients with rheumatoid arthritis. Thus a third-generation total shoulder replacement gives an excellent result in patients with osteoarthritis and an intact rotator cuff. Patients with rheumatoid arthritis have a 75% risk of failure of the rotator cuff at ten years


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 3 | Pages 387 - 392
1 Mar 2010
Kasten P Pape G Raiss P Bruckner T Rickert M Zeifang F Loew M

We have investigated the mid-term outcome of total shoulder replacement using a keeled cemented glenoid component and a modern cementing technique with regard to the causes of failure and loosening of the components. Between 1997 and 2003 we performed 96 total shoulder replacements on 88 patients, 24 men and 64 women with a mean age of 69.7 years (31 to 82). The minimum follow-up was five years and at the time of review 87 shoulders (77 patients) were examined at a mean follow-up of 89.1 months (60 to 127). Cumulative survival curves were generated with re-operations (accomplished and planned), survivorship of the proshesis, loosening of the glenoid (defined as tilt > 5° or subsidence > 5 mm), the presence of radiolucent lines and a Constant score of < 30 as the endpoints. There were two re-operations not involving revision of the implants and the survival rate of the prosthesis was 100.0% for the follow-up period, with an absolute Constant score of > 30 as the endpoint the survival rate was 98%. Radiological glenoid loosening was 9% after five years, and 33% after nine years. There was an incidence of 8% of radiolucent lines in more than three of six zones in the immediate post-operative period, of 37.0% after the first year which increased to 87.0% after nine years. There was no correlation between the score of Boileau and the total Constant score at the latest follow-up, but there was correlation between glenoid loosening and pain (p = 0.001). We found that total shoulder replacement had an excellent mid-term survivorship and clinical outcome. The surgical and cementing techniques were related to the decrease in radiolucent lines around the glenoid compared with earlier studies. One concern, however, was the fact that radiolucent lines increased over time and there was a rate of glenoid loosening of 9% after five years and 33% after nine years. This suggests that the design of the glenoid component, and the implantation and cementing techniques may need further improvement


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 5 | Pages 671 - 677
1 May 2012
Raiss P Sowa B Bruckner T Eck S Woerz S Rohr K Rickert M Kasten P

The aim of this study was to compare a third-generation cementing procedure for glenoid components with a new technique for cement pressurisation. In 20 pairs of scapulae, 20 keeled and 20 pegged glenoid components were implanted using either a third-generation cementing technique (group 1) or a new pressuriser (group 2). Cement penetration was measured by three-dimensional (3D) analysis of micro-CT scans. The mean 3D depth of penetration of the cement was significantly greater in group 2 (p < 0.001). The mean thickness of the cement mantle for keeled glenoids was 2.50 mm (2.0 to 3.3) in group 1 and 5.18 mm (4.4 to 6.1) in group 2, and for pegged glenoids it was 1.72 mm (0.9 to 2.3) in group 1 and 5.63 mm (3.6 to 6.4) in group 2. A cement mantle < 2 mm was detected less frequently in group 2 (p < 0.001). Using the cement pressuriser the proportion of cement mantles < 2 mm was significantly reduced compared with the third-generation cementing technique


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 9 | Pages 1197 - 1200
1 Sep 2009
Betts HM Abu-Rajab R Nunn T Brooksbank AJ

We describe the longer term clinical and radiological findings in a prospectively followed series of 49 rheumatoid patients (58 shoulders) who had undergone Neer II total shoulder replacement. The early and intermediate results have been published previously. At a mean follow-up of 19.8 years (16.5 to 23.8) 14 shoulders survived. Proximal migration of the humeral component was associated with progressive loosening of the glenoid and humeral components, but was independent of the state of the rotator cuff at the time of operation. Despite these changes the range of movement was preserved. Most patients had little or no pain in the shoulder, could sleep undisturbed and could attend to personal hygiene and grooming


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 8 | Pages 1112 - 1117
1 Aug 2010
Clement ND Hallett A MacDonald D Howie C McBirnie J

We compared the outcome of arthroscopic repair of the rotator cuff in 32 diabetic patients with the outcome in 32 non-diabetic patients matched for age, gender, size of tear and comorbidities. The Constant-Murley score improved from a mean of 49.2 (24 to 80) pre-operatively to 60.8 (34 to 95) post-operatively (p = 0.0006) in the diabetic patients, and from 46.4 (23 to 90) pre-operatively to 65.2 (25 to 100) post-operatively (p = 0.0003) in the non-diabetic patients at six months. This was significantly greater (p = 0.0002) in non-diabetic patients (18.8) than in diabetics (11.6). There was no significant change in the mean mental component of the Short-Form 12, but the mean physical component increased from 35 to 41 in non-diabetics (p = 0.0001), and from 37 to 39 (p = 0.15) in diabetics. These trends were observed at one year. Patients with diabetes showed improvement of pain and function following arthroscopic rotator cuff repair in the short term, but less than their non-diabetic counterparts


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 10 | Pages 1416 - 1421
1 Oct 2010
Qureshi F Draviaraj KP Stanley D

Between September 1993 and September 1996, we performed 34 Kudo 5 total elbow replacements in 31 rheumatoid patients. All 22 surviving patients were reviewed at a mean of 11.9 years (10 to 14). Their mean age was 56 years (37 to 78) at the time of operation. All had Larsen grade IV or V rheumatoid changes on X-ray. Nine (three bilateral replacements and six unilateral) had died from unrelated causes. One who had died before ten years underwent revision for dislocation. Of the 22 total elbow replacements reviewed six had required revision, four for aseptic loosening (one humeral and three ulnar) and two for infection. Post-operatively, one patient had neuropraxia of the ulnar nerve and one of the radial nerve. Two patients had valgus tilting of the ulnar component. With revision as the endpoint, the mean survival time for the prosthesis was 11.3 years (95% confidence interval (10 to 13) and the estimated survival of the prosthesis at 12 years according to Kaplan-Meier survival analysis was 74% (95% confidence interval 0.53 to 0.91). Of the 16 surviving implants, ten were free from pain, four had mild pain and two moderate. The mean arc of flexion/extension of the elbow was 106° (65° to 130°) with pronation/supination of 90° (30° to 150°) with the joint at 90° of flexion. The mean Mayo elbow performance score was 82 (60 to 100) with five excellent, ten good and one fair result. Good long-term results can be expected using the Kudo 5 total elbow replacement in patients with rheumatoid disease, with a low incidence of loosening of the components


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 3 | Pages 330 - 334
1 Mar 2007
Cesar M Roussanne Y Bonnel F Canovas F

Between 1993 and 2002, 58 GSB III total elbow replacements were implanted in 45 patients with rheumatoid arthritis by the same surgeon. At the most recent follow-up, five patients had died (five elbows) and six (nine elbows) had been lost to follow-up, leaving 44 total elbow replacements in 34 patients available for clinical and radiological review at a mean follow-up of 74 months (25 to 143). There were 26 women and eight men with a mean age at operation of 55.7 years (24 to 77). At the latest follow-up, 31 excellent (70%), six good (14%), three fair (7%) and four poor (9%) results were noted according to the Mayo elbow performance score. Five humeral (11%) and one ulnar (2%) component were loose according to radiological criteria (type III or type IV). Of the 44 prostheses, two (5%) had been revised, one for type-IV humeral loosening after follow-up for ten years and one for fracture of the ulnar component. Seven elbows had post-operative dysfunction of the ulnar nerve, which was transient in five and permanent in two. Despite an increased incidence of loosening with time, the GSB III prosthesis has given favourable mid-term results in patients with rheumatoid arthritis


The Bone & Joint Journal
Vol. 95-B, Issue 5 | Pages 668 - 672
1 May 2013
Abdel MP Hattrup SJ Sperling JW Cofield RH Kreofsky CR Sanchez-Sotelo J

Instability after arthroplasty of the shoulder is difficult to correct surgically. Soft-tissue procedures and revision surgery using unconstrained anatomical components are associated with a high rate of failure. The purpose of this study was to determine the results of revision of an unstable anatomical shoulder arthroplasty to a reverse design prosthesis. Between 2004 and 2007, 33 unstable anatomical shoulder arthroplasties were revised to a reverse design. The mean age of the patients was 71 years (53 to 86) and their mean follow-up was 42 months (25 to 71). The mean time to revision was 26 months (4 to 164). Pain scores improved significantly (pre-operative visual analogue scale (VAS) of 7.2 (. sd. 1.6); most recent VAS 2.2 (. sd. 1.9); p = 0.001). There was a statistically significant increase in mean active forward elevation from 40.2° (. sd. 27.3) to 97.0° (. sd. 36.2) (p = 0.001). There was no significant difference in internal (p = 0.93) or external rotation (p = 0.40). Radiological findings included notching in five shoulders (15%) and heterotopic ossification of the inferior capsular region in three (9%). At the last follow-up 31 shoulders (94%) were stable. The remaining two shoulders dislocated at 2.5 weeks and three months post-operatively, respectively. According to the Neer rating system, there were 13 excellent (40%), ten satisfactory (30%) and ten unsatisfactory results (30%). Revision of hemiarthroplasty or anatomical total shoulder replacement for instability using a reverse design prosthesis gives good short-term results. Cite this article: Bone Joint J 2013;95-B:668–72


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 1 | Pages 80 - 85
1 Jan 2012
Malone AA Sanchez JS Adams R Morrey B

We report the effectiveness of revision of total elbow replacement by re-cementing. Between 1982 and 2004, 53 elbows in 52 patients were treated with re-cementing of a total elbow replacement into part or all of the existing cement mantle or into the debrided host-bone interface, without the use of structural bone augmentation or a custom prosthesis. The original implant revision was still in situ and functional in 42 of 53 elbows (79%) at a mean of 94.5 months (26 to 266) after surgery. In 31 of these 42 elbows (74%) the Mayo Elbow Performance Score was good or excellent. Overall, of the 53 elbows, 18 (34%) required re-operation, ten (19%) for loosening. A classification system was developed to identify those not suitable for revision by this technique, and using this we have showed that successful re-implantation is statistically correlated to properly addressing the bone deficiency for both the humeral (p = 0.005) and the ulnar (p = 0.039) components