Aims. Vancomycin is commonly added to acrylic bone cement during revision
arthroplasty surgery. Proprietary cement preparations containing
vancomycin are available, but are significantly more expensive.
We investigated whether the elution of antibiotic from ‘home-made’
cement containing vancomycin was comparable with more expensive
commercially available vancomycin impregnated cement. Materials and Methods. A total of 18 cement discs containing either proprietary CopalG+V;
or ‘home-made’ CopalR+G with vancomycin added by hand, were made.
Each disc contained the same amount of antibiotic (0.5 g gentamycin,
2 g vancomycin) and was immersed in ammonium acetate buffer in a
sealed container. Fluid from each container was sampled at eight
time points over a two-week period. The concentrations of gentamicin
and vancomycin in the fluid were analysed using high performance
liquid chromatography mass spectrometry. Results. The highest peak concentrations of antibiotic were observed from
the ‘home-made’ cements containing vancomycin, added as in the operating
theatre. The overall elution of antibiotic was, fivefold (vancomycin)
and twofold (gentamicin) greater from the ‘home-made’ mix compared
with the commercially mixed cement. The use of a vacuum during mixing
had no significant effect on antibiotic elution in any of the samples. Conclusion. These findings suggest that the addition of 2 g vancomycin powder
to gentamicin-impregnated
Aims. The aim of this study was to develop a single-layer hybrid organic-inorganic sol-gel coating that is capable of a controlled antibiotic release for cementless hydroxyapatite (HA)-coated titanium orthopaedic prostheses. Methods. Coatings containing gentamicin at a concentration of 1.25% weight/volume (wt/vol), similar to that found in commercially available antibiotic-loaded
It can be extremely challenging to determine whether to perform reimplantation in patients who have contradictory serum inflammatory markers and frozen section results. We investigated whether patients with a positive frozen section at reimplantation were at a higher risk of reinfection despite normal ESR and CRP. We retrospectively reviewed 163 consecutive patients with periprosthetic joint infections (PJIs) who had normal ESR and CRP results pre-reimplantation in our hospital from 2014 to 2018. Of these patients, 26 had positive frozen sections at reimplantation. The minimum follow-up time was two years unless reinfection occurred within this period. Univariable and multivariable logistic regression analyses were performed to identify the association between positive frozen sections and treatment failure.Aims
Methods
The success rates of two-stage revision arthroplasty for infection have evolved since their early description. The implementation of internationally accepted outcome criteria led to the readjustment of such rates. However, patients who do not undergo reimplantation are usually set aside from these calculations. The aim of this study was to investigate the outcomes of two-stage revision arthroplasty when considering those who do not undergo reimplantation, and to investigate the characteristics of this subgroup. A retrospective cohort study was conducted. Patients with chronic hip or knee periprosthetic joint infection (PJI) treated with two-stage revision between January 2010 and October 2018, with a minimum follow-up of one year, were included. Variables including demography, morbidity, microbiology, and outcome were collected. The primary endpoint was the eradication of infection. Patients who did not undergo reimplantation were analyzed in order to characterize this subgroup better.Aims
Methods
Aseptic loosening is seen in a significant proportion of cemented total hip replacements (THR). In a small subgroup of patients who suffer early loosening polyethylene debris is unlikely to be responsible. We recently reported one case of allergic contact dermatitis to N,N-dimethylparatoluidine (DMT), an accelerator used in
Preoperative nasal All primary total hip arthroplasties (THA) and total knee arthroplasties (TKA) performed from January 2006 to April 2018 were retrospectively reviewed for the incidence of early PJI. Demographic parameters, risk factors for PJI (American Society of Anaesthesiologists classification, body mass index, smoking status, and diabetes mellitus) and implant types were collected. A preoperative screening and eradication protocol for nasal colonization of Aims
Methods
Biopsy of the periprosthetic tissue is an important diagnostic tool for prosthetic joint infection (PJI) as it enables the detection of the responsible microorganism with its sensitivity to antibiotics. We aimed to investigate how often the bacteria identified in the tissue analysis differed between samples obtained from preoperative biopsy and intraoperative revision surgery in cases of late PJI; and whether there was a therapeutic consequence. A total of 508 patients who required revision surgery of total hip arthroplasty (THA) (n = 231) or total knee arthroplasty (TKA) (n = 277) because of component loosening underwent biopsy before revision surgery. The tissue samples collected at biopsy and during revision surgery were analyzed according to the criteria of the Musculoskeletal Infection Society (MSIS).Aims
Methods
The aim of this study was to conduct the largest low contact stress (LCS) retrieval study to elucidate the failure mechanisms of the Porocoat and Duofix femoral component. The latter design was voluntarily recalled by the manufacturer. Uncemented LCS explants were divided into three groups: Duofix, Porocoat, and mixed. Demographics, polyethylene wear, tissue ingrowth, and metallurgical analyses were performed.Aims
Materials and Methods
The aim of this study was to give estimates of the incidence of component incompatibility in hip and knee arthroplasty and to test the effect of an online, real-time compatibility check. Intraoperative barcode registration of arthroplasty implants was introduced in Denmark in 2013. We developed a compatibility database and, from May 2017, real-time compatibility checking was implemented and became part of the registration. We defined four classes of component incompatibility: A-I, A-II, B-I, and B-II, depending on an assessment of the level of risk to the patient (A/B), and on whether incompatibility was knowingly accepted (I/II).Aims
Materials and Methods
We describe the development and early clinical application of a ported, proximally-cemented titanium stem for cemented total hip arthroplasty. PMMA
Fungal prosthetic joint infections (PJIs) are rare and account for about 1% of total PJIs. Our aim was to present clinical and microbiological results in treating these patients with a two-stage approach and antifungal spacers. We retrospectively reviewed our institutional database and identified 26 patients with positive fungal cultures and positive Musculoskeletal Infection Society (MSIS) criteria for PJI who were treated between 2009 and 2017. We identified 18 patients with total hip arthroplasty (THA) and eight patients with total knee arthroplasty (TKA). The surgical and antifungal treatment, clinical and demographic patient data, complications, relapses, and survival were recorded and analyzed.Aims
Patients and Methods
The aim of this study was to establish the diagnostic accuracy
of culture of joint aspirate with and without saline injection-reaspiration. This is a retrospective analysis of 580 hip and knee aspirations
in patients who were deemed to have a moderate to high risk of infection,
and who subsequently proceeded to revision arthroplasty over a period
of 12 years. It was carried out at a large quaternary referral centre
where preoperative aspiration is routine.Aims
Patients and Methods
Tissue responses to debris formed by abrasion of polymethylmethacrylate
(PMMA) spacers at two-stage revision arthroplasty for prosthetic
joint infection are not well described. We hypothesised that PMMA
debris induces immunomodulation in periprosthetic tissues. Samples of tissue were taken during 35 two-stage revision arthroplasties
(nine total hip and 26 total knee arthroplasties) in patients whose
mean age was 67 years (44 to 85). Fourier transform infrared microscopy
was used to confirm the presence of PMMA particles. Histomorphometry
was performed using Sudan Red and Haematoxylin-Eosin staining.
CD-68, CD-20, CD-11(c), CD-3 and IL-17 antibodies were used to immunophenotype
the inflammatory cells. All slides were scored semi-quantitatively
using the modified Willert scoring system.Aims
Patients and Methods
We investigated whether the indentation of bone
cement spacers used in revision of infected joint arthroplasty with a
MacDonald dissector increased the elution of antibiotic The fluid sampled at 72 hours from the indented discs containing
0.17 g gentamicin (0.88% w/w) contained a mean of 113 mcg/ml (90.12
to 143.5) compared with 44.5 mcg/ml (44.02 to 44.90) in the fluid
sampled from the plain discs (p = 0.012). In discs containing 0.33
g gentamicin (1.75% w/w), the concentration eluted from the indented discs
at 72 hours was a mean of 316 mcg/ml (223 to 421) compared with
a mean of 118 mcg/ml (100 to 140) from the plain discs (p <
0.001). At two weeks, these significant differences persisted. At nine
weeks the indented discs eluted a greater concentration for all
gentamicin doses, but the difference was only significant for the
discs containing 0.17 g (0.88% w/w, p = 0.006). However if the area
under the curve is taken as a measure of the total antibiotic eluted,
the indented discs eluted more gentamicin than the plain discs for
the 0.17 g (0.88% w/w, p = 0.031), the 0.25 g (1.41% w/w, p <
0.001) and the 0.33 g (1.75% w/w, p <
0.001) discs. When preparing antibiotic spacers for use in staged revision
arthroplasty surgery we recommend indenting the spacer with a MacDonald
dissector to increase the elution of antibiotic. Cite this article:
The purpose of this study was to evaluate whether
the serum level of interleukin 6 (IL-6) could be used to identify the
persistence of infection after the first stage of a two-stage revision
for periprosthetic joint infection. Between 2010 and 2011, we prospectively studied 55 patients (23
men, 32 women; mean age 69.5 years; 36 to 86) with a periprosthetic
joint infection. Bacteria were identified in two intra-operative
tissue samples during re-implantation in 16 patients. These cases
were classified as representing persistent infection. To calculate a precise cut-off value which could be used in everyday
clinical practice, a 3 x 2 contingency table was constructed and
manually defined. We found that a serum IL-6 ≥ 13 pg/mL can be regarded as indicating
infection: its positive-predictive value is 90.9%. A serum IL-6 ≤ 8
pg/mL can be regarded as indicating an absence of infection: its
negative predictive value is 92.1%. The serum IL-6 level seems to be a reasonable marker for identifying
persistent infection after the first stage of a revision joint arthroplasty
and before attempting re-implantation. Cite this article:
Fungal peri-prosthetic infections of the knee
and hip are rare but likely to result in devastating complications.
In this study we evaluated the results of their management using
a single-stage exchange technique. Between 2001 and 2011, 14 patients
(ten hips, four knees) were treated for a peri-prosthetic fungal
infection. One patient was excluded because revision surgery was
not possible owing to a large acetabular defect. One patient developed
a further infection two months post-operatively and was excluded
from the analysis. Two patients died of unrelated causes. After a mean of seven years (3 to 11) a total of ten patients
were available for follow-up. One patient, undergoing revision replacement
of the hip, had a post-operative dislocation. Another patient, undergoing
revision replacement of the knee, developed a wound infection and
required revision 29 months post-operatively following a peri-prosthetic femoral
fracture. The mean Harris hip score increased to 74 points (63 to 84; p
<
0.02) in those undergoing revision replacement of the hip,
and the mean Hospital for Special Surgery knee score increased to
75 points (70 to 80; p <
0.01) in those undergoing revision replacement
of the knee. A single-stage revision following fungal peri-prosthetic infection
is feasible, with an acceptable rate of a satisfactory outcome. Cite this article:
The aim of this study was to define return to
theatre (RTT) rates for elective hip and knee replacement (HR and
KR), to describe the predictors and to show the variations in risk-adjusted
rates by surgical team and hospital using national English hospital
administrative data. We examined information on 260 206 HRs and 315 249 KRs undertaken
between April 2007 and March 2012. The 90-day RTT rates were 2.1%
for HR and 1.8% for KR. Male gender, obesity, diabetes and several
other comorbidities were associated with higher odds for both index
procedures. For HR, hip resurfacing had half the odds of cement fixation
(OR = 0.58, 95% confidence intervals (CI) 0.47 to 0.71). For KR,
unicondylar KR had half the odds of total replacement (OR = 0.49,
95% CI 0.42 to 0.56), and younger ages had higher odds (OR = 2.23,
95% CI 1.65 to 3.01) for ages <
40 years compared with ages 60
to 69 years). There were more funnel plot outliers at three standard deviations
than would be expected if variation occurred on a random basis. Hierarchical modelling showed that three-quarters of the variation
between surgeons for HR and over half the variation between surgeons
for KR are not explained by the hospital they operated at or by
available patient factors. We conclude that 90-day RTT rate may
be a useful quality indicator for orthopaedics. Cite this article:
Antibiotic prophylaxis is routinely administered during joint replacement surgery and may predispose patients to
We have investigated the contaminating bacteria in primary hip arthroplasty and their sensitivity to the prophylactic antibiotics currently in use. Impressions (627) of the gloved hands of the surgical team in 50 total hip arthroplasties were obtained on blood agar. The gloves were changed after draping, at intervals of 20 minutes thereafter, and before using cement. Changes were also undertaken whenever a visible puncture was detected. The culture plates were incubated at 37°C for 48 hours. Isolates were identified and tested for sensitivity to flucloxacillin, which is a recognised indicator of sensitivity to cefuroxime. They were also tested against other agents depending upon their appearance on Gram staining. We found contamination in 57 (9%) impressions and 106 bacterial isolates. Coagulase-negative staphylococci were seen most frequently (68.9%), but we also isolated