Advertisement for orthosearch.org.uk
Results 1 - 20 of 474
Results per page:
The Bone & Joint Journal
Vol. 103-B, Issue 6 | Pages 1070 - 1077
1 Jun 2021
Hipfl C Mooij W Perka C Hardt S Wassilew GI

Aims. The purpose of this study was to evaluate unexpected positive cultures in total hip arthroplasty (THA) revisions for presumed aseptic loosening, to assess the prevalence of low-grade infection using two definition criteria, and to analyze its impact on implant survival after revision. Methods. A total of 274 THA revisions performed for presumed aseptic loosening from 2012 to 2016 were reviewed. In addition to obtaining intraoperative tissue cultures from all patients, synovial and sonication fluid samples of the removed implant were obtained in 215 cases (79%) and 101 cases (37%), respectively. Histopathological analysis was performed in 250 cases (91%). Patients were classified as having low-grade infections according to institutional criteria and Musculoskeletal Infection Society (MSIS) International Consensus Meeting (ICM) 2013 criteria. Low-grade infections according to institutional criteria were treated with targeted antibiotics for six weeks postoperatively. Implant failure was defined as the need for re-revision resulting from periprosthetic joint infection (PJI) and aseptic reasons. The mean follow-up was 68 months (26 to 95). Results. Unexpected positive intraoperative samples were found in 77 revisions (28%). Low-grade infection was diagnosed in 36 cases (13%) using institutional criteria and in nine cases (3%) using MSIS ICM 2013 criteria. In all, 41 patients (15%) had single specimen growth of a low-virulent pathogen and were deemed contaminated. Coagulase-negative Staphylococcus and anaerobes were the most commonly isolated bacteria. Implant failure for PJI was higher in revisions with presumed contaminants (5/41, 12%) compared to those with low-grade infections (2/36, 6%) and those with negative samples (5/197, 3%) (p = 0.021). The rate of all-cause re-revision was similar in patients diagnosed with low-grade infections (5/36, 14%) and those with presumed contaminants (6/41, 15%) and negative samples (21/197, 11%) (p = 0.699). Conclusion. Our findings suggest that the presumption of culture contamination in aseptic revision hip arthroplasty may increase the detection of PJI. In this cohort, the presence of low-grade infection did not increase the risk of re-revision. Further studies are needed to assess the relevance of single specimen growth and the benefits of specific postoperative antibiotic regimens. Cite this article: Bone Joint J 2021;103-B(6):1070–1077


The Journal of Bone & Joint Surgery British Volume
Vol. 83-B, Issue 6 | Pages 838 - 842
1 Aug 2001
Chambers IR Fender D McCaskie AW Reeves BC Gregg PJ

The radiological features of the cement mantle around total hip replacements (THRs) have been used to assess aseptic loosening. In this case-control study we investigated the risk of failure of THR as predictable by a range of such features using data from patients recruited to the Trent Regional Arthroplasty Study (TRAS). An independent radiological assessment was undertaken on Charnley THRs with aseptic loosening within five years of surgery and on a control group from the TRAS database. Chi-squared tests were used to test the probability of obtaining the observed data by chance, and odds ratios were calculated to estimate the strength of association for different features. Several features were associated with a clinically important increase (> twofold) in the risk of loosening, which was statistically significant for four features (p < 0.01). Inadequate cementation (Barrack C and D grades) was the most significant feature, with an estimated odds ratio of 9.5 (95% confidence interval 3.2 to 28.4, p < 0.0001) for failure


The Journal of Bone & Joint Surgery British Volume
Vol. 87-B, Issue 6 | Pages 781 - 785
1 Jun 2005
Temmerman OPP Raijmakers PGHM Berkhof J Hoekstra OS Teule GJJ Heyligers IC

In this meta-analysis we included 32 English-language articles published between January 1975 and June 2004 on the diagnostic performance of plain radiography, subtraction arthrography, nuclear arthrography and bone scintigraphy in detecting aseptic loosening of the femoral component, using criteria based on the Cochrane systematic review of screening and diagnostic tests. The mean sensitivity and specificity were, respectively, 82% (95% confidence interval (CI) 76 to 87) and 81% (95% CI 73 to 87) for plain radiography and 85% (95% CI 75 to 91) and 83% (95% CI 75 to 89) for nuclear arthrography. Pooled sensitivity and specificity were, respectively, 86% (95% CI 74 to 93) and 85% (95% CI 77 to 91) for subtraction arthrography and 85% (95% CI 79 to 89) and 72% (95% CI 64 to 79) for bone scintigraphy. Although the diagnostic performance of the imaging techniques was not significantly different, plain radiography and bone scintigraphy are preferred for the assessment of a femoral component because of their efficacy and lower risk of patient morbidity


The Journal of Bone & Joint Surgery British Volume
Vol. 85-B, Issue 1 | Pages 31 - 36
1 Jan 2003
Müller U Gautier E Roeder C Busato A

We assessed differences in the incidence and appearance of the radiological signs of loosening of the cup for various types of design. This was an observational study based on hip registry data of 15 340 patients with 17 951 total hip arthroplasties collected over a period of 33 years in 49 hospitals in Central Europe. The threaded and the press-fit titanium cups showed significantly less aseptic loosening than the other systems. The direction of migration and the frequency of the radiological signs of loosening differed between the cup systems and were time-dependent. Our findings indicate the superiority of uncemented threaded cups and press-fit titanium cups over other designs of cup


The Bone & Joint Journal
Vol. 104-B, Issue 9 | Pages 1039 - 1046
1 Sep 2022
Özdemir E Kuijpers MFL Visser J Schreurs BW Rijnen WHC

Aims. The aim of this study is to report the long-term outcomes of instrumented femoral revisions with impaction allograft bone grafting (IBG) using the X-change femoral revision system at 30 years after introduction of the technique. Methods. We updated the outcomes of our previous study, based on 208 consecutive revisions using IBG and the X-change femoral revision system in combination with a cemented polished stem, performed in our tertiary care institute between 1991 and 2007. Kaplan-Meier survival analyses were used to determine the survival rate of the revisions with endpoint revision for any reason and aseptic loosening. Secondary outcomes were radiological loosening and patient-reported outcome measures. Results. Mean age at revision total hip arthroplasty (THA) was 64.9 years (30 to 86). The most prevalent diagnosis for the femoral revision was aseptic loosening. At review in May 2021, 81 patients (85 hips) were still alive and 118 patients (120 hips; 58%) had died. Three patients (3 hips; 1%) were lost to follow-up at 11, 15, and 16 years after surgery, respectively. Data of all deceased and lost patients were included until final follow-up. The mean follow-up was 13.4 years (0 to 28). During the follow-up, 22 re-revisions were performed. The most common reason for re-revision was infection (n = 12; 54%). The survival with endpoint re-revision for any reason was 86% (95% confidence interval (CI) 79 to 91) at 20 years and 74% (95% CI 43 to 89) at 25 years after surgery. The survival for endpoint re-revision for aseptic loosening was 97% (95% CI 91 to 99) after both 20 and 25 years. Conclusion. We conclude that femoral IBG is a valuable technique that can reconstitute femoral bone loss in the long term. After 25 years of follow-up, few re-revisions for aseptic loosening were required. Also, the overall revision rate is very acceptable at a long follow-up. This technique is especially attractive for younger patients facing femoral revisions with extensive bone loss. Cite this article: Bone Joint J 2022;104-B(9):1039–1046


The Bone & Joint Journal
Vol. 103-B, Issue 9 | Pages 1479 - 1487
1 Sep 2021
Davis ET Pagkalos J Kopjar B

Aims. The aim of our study was to investigate the effect of asymmetric crosslinked polyethylene liner use on the risk of revision of cementless and hybrid total hip arthroplasties (THAs). Methods. We undertook a registry study combining the National Joint Registry dataset with polyethylene manufacturing characteristics as supplied by the manufacturers. The primary endpoint was revision for any reason. We performed further analyses on other reasons including instability, aseptic loosening, wear, and liner dissociation. The primary analytic approach was Cox proportional hazard regression. Results. A total of 213,146 THAs were included in the analysis. Overall, 2,997 revisions were recorded, 1,569 in THAs with a flat liner and 1,428 in THAs using an asymmetric liner. Flat liner THAs had a higher risk of revision for any reason than asymmetric liner THAs when implanted through a Hardinge/anterolateral approach (hazard ratio (HR) 1.169, 95% confidence interval (CI) 1.022 to 1.337) and through a posterior approach (HR 1.122, 95% CI 1.108 to 1.346). There was no increased risk of revision for aseptic loosening when asymmetric liners were used for any surgical approach. A separate analysis of the three most frequently used crosslinked polyethylene liners was in agreement with this finding. When analyzing THAs with flat liners only, THAs implanted through a Hardinge/anterolateral approach were associated with a reduced risk of revision for instability compared to posterior approach THAs (HR 0.561 (95% CI 0.446 to 0.706)). When analyzing THAs with an asymmetric liner, there was no significant difference in the risk of revision for instability between the two approaches (HR 0.838 (95% CI 0.633 to 1.110)). Conclusion. For THAs implanted through the posterior approach, the use of asymmetric liners reduces the risk of revision for instability and revision for any reason. In THAs implanted through a Hardinge/anterolateral approach, the use of an asymmetric liner was associated with a reduced risk of revision. The effect on revision for instability was less pronounced than in the posterior approach. Cite this article: Bone Joint J 2021;103-B(9):1479–1487


The Bone & Joint Journal
Vol. 106-B, Issue 4 | Pages 352 - 358
1 Apr 2024
Wilson JM Trousdale RT Bedard NA Lewallen DG Berry DJ Abdel MP

Aims. Dislocation remains a leading cause of failure following revision total hip arthroplasty (THA). While dual-mobility (DM) bearings have been shown to mitigate this risk, options are limited when retaining or implanting an uncemented shell without modular DM options. In these circumstances, a monoblock DM cup, designed for cementing, can be cemented into an uncemented acetabular shell. The goal of this study was to describe the implant survival, complications, and radiological outcomes of this construct. Methods. We identified 64 patients (65 hips) who had a single-design cemented DM cup cemented into an uncemented acetabular shell during revision THA between 2018 and 2020 at our institution. Cups were cemented into either uncemented cups designed for liner cementing (n = 48; 74%) or retained (n = 17; 26%) acetabular components. Median outer head diameter was 42 mm. Mean age was 69 years (SD 11), mean BMI was 32 kg/m. 2. (SD 8), and 52% (n = 34) were female. Survival was assessed using Kaplan-Meier methods. Mean follow-up was two years (SD 0.97). Results. There were nine cemented DM cup revisions: three for periprosthetic joint infection, three for acetabular aseptic loosening from bone, two for dislocation, and one for a broken cup-cage construct. The two-year survivals free of aseptic DM revision and dislocation were both 92%. There were five postoperative dislocations, all in patients with prior dislocation or abductor deficiency. On radiological review, the DM cup remained well-fixed at the cemented interface in all but one case. Conclusion. While dislocation was not eliminated in this series of complex revision THAs, this technique allowed for maximization of femoral head diameter and optimization of effective acetabular component position during cementing. Of note, there was only one failure at the cemented interface. Cite this article: Bone Joint J 2024;106-B(4):352–358


The Bone & Joint Journal
Vol. 106-B, Issue 6 | Pages 540 - 547
1 Jun 2024
Nandra RS Elnahal WA Mayne A Brash L McBryde CW Treacy RBC

Aims. The Birmingham Hip Resurfacing (BHR) was introduced in 1997 to address the needs of young active patients using a historically proven large-diameter metal-on-metal (MoM) bearing. A single designer surgeon’s consecutive series of 130 patients (144 hips) was previously reported at five and ten years, reporting three and ten failures, respectively. The aim of this study was to extend the follow-up of this original cohort at 25 years. Methods. The study extends the reporting on the first consecutive 144 resurfacing procedures in 130 patients for all indications. All operations were undertaken between August 1997 and May 1998. The mean age at operation was 52.1 years (SD 9.93; 17 to 76), and included 37 female patients (28.5%). Failure was defined as revision of either component for any reason. Kaplan-Meier survival analysis was performed. Routine follow-up with serum metal ion levels, radiographs, and Oxford Hip Scores (OHSs) was undertaken. Results. Overall implant survival was 83.50% (95% confidence interval (CI) 0.79 to 0.90) at 25 years, and the number at risk was 79. Survival in male patients at 25 years was 89.5% (95% CI 0.83 to 0.96) compared to 66.9% for female patients (95% CI 0.51 to 0.83). Ten additional failures occurred in the period of ten to 25 years. These involved an adverse reaction to metal debris in four patients, a periprosthetic femoral neck fracture affecting five patients, and aseptic loosening in one patient. The median chromium levels were 49.50 nmol/l (interquartile range (IQR) 34 to 70), and the median cobalt serum levels were 42 nmol/l (IQR 24.50 to 71.25). The median OHS at last follow-up was 35 (IQR 10 to 48). During the 25-year study period, 29 patients died. Patient survival at 25 years was 75.10% (95% CI 0.67 to 0.83). Conclusion. This study demonstrates that MoM hip resurfacing using the BHR provides a durable alternative to total hip arthroplasty (THA), particularly in younger male patients with osteoarthritis wishing to maintain a high level of function. These results compare favourably to the best results for THAs. Cite this article: Bone Joint J 2024;106-B(6):540–547


The Bone & Joint Journal
Vol. 104-B, Issue 1 | Pages 19 - 26
1 Jan 2022
Sevaldsen K Schnell Husby O Lian ØB Farran KM Schnell Husby V

Aims. Highly polished stems with force-closed design have shown satisfactory clinical results despite being related to relatively high early migration. It has been suggested that the minimal thickness of cement mantles surrounding the femoral stem should be 2 mm to 4 mm to avoid aseptic loosening. The line-to-line cementing technique of the femoral stem, designed to achieve stem press-fit, challenges this opinion. We compared the migration of a highly polished stem with force-closed design by standard and line-to-line cementing to investigate whether differences in early migration of the stems occur in a clinical study. Methods. In this single-blind, randomized controlled, clinical radiostereometric analysis (RSA) study, the migration pattern of the cemented Corail hip stem was compared between line-to-line and standard cementing in 48 arthroplasties. The primary outcome measure was femoral stem migration in terms of rotation and translation around and along with the X-, Y-, and Z- axes measured using model-based RSA at three, 12, and 24 months. A linear mixed-effects model was used for statistical analysis. Results. Results from mixed model analyses revealed a lower mean retroversion for line-to-line (0.72° (95% confidence interval (CI) 0.38° to 1.07°; p < 0.001), but no significant differences in subsidence between the techniques (-0.15 mm (95% CI -0.53 to 0.227; p = 0.429) at 24 months. Radiolucent lines measuring < 2 mm wide were found in three and five arthroplasties cemented by the standard and line-to-line method, respectively. Conclusion. The cemented Corail stem with a force-closed design seems to settle earlier and better with the line-to-line cementing method, although for subsidence the difference was not significant. However, the lower rate of migration into retroversion may reduce the wear and cement deformation, contributing to good long-term fixation and implant survival. Cite this article: Bone Joint J 2022;104-B(1):19–26


The Bone & Joint Journal
Vol. 104-B, Issue 7 | Pages 859 - 866
1 Jul 2022
Innocenti M Smulders K Willems JH Goosen JHM van Hellemondt G

Aims. The aim of this study was to explore the relationship between reason for revision total hip arthroplasty (rTHA) and outcomes in terms of patient-reported outcome measures (PROMs). Methods. We reviewed a prospective cohort of 647 patients undergoing full or partial rTHA at a single high-volume centre with a minimum of two years’ follow-up. The reasons for revision were classified as: infection; aseptic loosening; dislocation; structural failure; and painful THA for other reasons. PROMs (modified Oxford Hip Score (mOHS), EuroQol five-dimension three-level health questionnaire (EQ-5D-3L) score, and visual analogue scales for pain during rest and activity), complication rates, and failure rates were compared among the groups. Results. The indication for revision influenced PROMs improvement over time. This finding mainly reflected preoperative differences between the groups, but diminished between the first and second postoperative years. Preoperatively, patients revised due to infection and aseptic loosening had a lower mOHS than patients with other indications for revision. Pain scores at baseline were highest in patients being revised for dislocation. Infection and aseptic loosening groups showed marked changes over time in both mOHS and EQ-5D-3L. Overall complications and re-revision rates were 35.4% and 9.7% respectively, with no differences between the groups (p = 0.351 and p = 0.470, respectively). Conclusion. Good outcomes were generally obtained regardless of the reason for revision, with patients having the poorest preoperative scores exhibiting the greatest improvement in PROMs. Furthermore, overall complication and reoperation rates were in line with previous reports and did not differ between different indications for rTHA. Cite this article: Bone Joint J 2022;104-B(7):859–866


The Bone & Joint Journal
Vol. 103-B, Issue 7 Supple B | Pages 135 - 144
1 Jul 2021
Kuyl E Shu F Sosa BR Lopez JD Qin D Pannellini T Ivashkiv LB Greenblatt MB Bostrom MPG Yang X

Aims. Aseptic loosening is a leading cause of uncemented arthroplasty failure, often accompanied by fibrotic tissue at the bone-implant interface. A biological target, neutrophil extracellular traps (NETs), was investigated as a crucial connection between the innate immune system’s response to injury, fibrotic tissue development, and proper bone healing. Prevalence of NETs in peri-implant fibrotic tissue from aseptic loosening patients was assessed. A murine model of osseointegration failure was used to test the hypothesis that inhibition (through Pad4-/- mice that display defects in peptidyl arginine deiminase 4 (PAD4), an essential protein required for NETs) or resolution (via DNase 1 treatment, an enzyme that degrades the cytotoxic DNA matrix) of NETs can prevent osseointegration failure and formation of peri-implant fibrotic tissue. Methods. Patient peri-implant fibrotic tissue was analyzed for NETs biomarkers. To enhance osseointegration in loose implant conditions, an innate immune system pathway (NETs) was either inhibited (Pad4-/- mice) or resolved with a pharmacological agent (DNase 1) in a murine model of osseointegration failure. Results. NETs biomarkers were identified in peri-implant fibrotic tissue collected from aseptic loosening patients and at the bone-implant interface in a murine model of osseointegration failure. Inhibition (Pad4-/-) or resolution (DNase 1) of NETs improved osseointegration and reduced fibrotic tissue despite loose implant conditions in mice. Conclusion. This study identifies a biological target (NETs) for potential noninvasive treatments of aseptic loosening by discovering a novel connection between the innate immune system and post-injury bone remodelling caused by implant loosening. By inhibiting or resolving NETs in an osseointegration failure murine model, fibrotic tissue encapsulation around an implant is reduced and osseointegration is enhanced, despite loose implant conditions. Cite this article: Bone Joint J 2021;103-B(7 Supple B):135–144


The Bone & Joint Journal
Vol. 102-B, Issue 1 | Pages 90 - 101
1 Jan 2020
Davis ET Pagkalos J Kopjar B

Aims. The aim of this study was to identify the effect of the manufacturing characteristics of polyethylene acetabular liners on the survival of cementless and hybrid total hip arthroplasty (THA). Methods. Prospective cohort study using linked National Joint Registry (NJR) and manufacturer data. The primary endpoint was revision for aseptic loosening. Cox proportional hazard regression was the primary analytical approach. Manufacturing variables included resin type, crosslinking radiation dose, terminal sterilization method, terminal sterilization radiation dose, stabilization treatment, total radiation dose, packaging, and face asymmetry. Total radiation dose was further divided into G1 (no radiation), G2 (> 0 Mrad to < 5 Mrad), G3 (≥ 5 Mrad to < 10 Mrad), and G4 (≥ 10 Mrad). Results. A total of 5,329 THAs were revised, 1,290 of which were due to aseptic loosening. Total radiation dose, face asymmetry, and stabilization treatments were found to significantly affect implant survival. G1 had the highest revision risk for any reason and for aseptic loosening and G3 and G4 the lowest. Compared with G1, the adjusted hazard ratio for G2 was 0.74 (95% confidence interval (CI) 0.64 to 0.86), G3 was 0.36 (95% CI 0.30 to 0.43), and G4 was 0.38 (95% CI 0.31 to 0.47). The cumulative incidence of revision for aseptic loosening at 12 years was 0.52 and 0.54 per 100 THAs for G3 and G4, respectively, compared with 1.95 per 100 THAs in G1. Asymmetrical liners had a lower revision risk due to aseptic loosening and reasons other than aseptic loosening compared with symmetric (flat) liners. In G3 and G4, stabilization with vitamin E and heating above melting point performed best. Conclusion. Polyethylene liners with a total radiation dose of ≥ 5 Mrad, an asymmetrical liner face, and stabilization with heating above the melting point demonstrate best survival. Cite this article: Bone Joint J 2020;102-B(1):90–101


The Bone & Joint Journal
Vol. 104-B, Issue 7 | Pages 833 - 843
1 Jul 2022
Kayani B Baawa-Ameyaw J Fontalis A Tahmassebi J Wardle N Middleton R Stephen A Hutchinson J Haddad FS

Aims. This study reports the ten-year wear rates, incidence of osteolysis, clinical outcomes, and complications of a multicentre randomized controlled trial comparing oxidized zirconium (OxZr) versus cobalt-chrome (CoCr) femoral heads with ultra-high molecular weight polyethylene (UHMWPE) and highly cross-linked polyethylene (XLPE) liners in total hip arthroplasty (THA). Methods. Patients undergoing primary THA were recruited from four institutions and prospectively allocated to the following treatment groups: Group A, CoCr femoral head with XLPE liner; Group B, OxZr femoral head with XLPE liner; and Group C, OxZr femoral head with UHMWPE liner. All study patients and assessors recording outcomes were blinded to the treatment groups. The outcomes of 262 study patients were analyzed at ten years’ follow-up. Results. Patients in Group C were associated with increased mean liner wear rates compared to patients in Group A (0.133 mm/yr (SD 0.21) vs 0.031 mm/yr (SD 0.07), respectively; p < 0.001) and Group B (0.133 mm/yr (SD 0.21) vs 0.022 mm/yr (SD 0.05), respectively; p < 0.001) at ten years’ follow-up. Patients in Group C were also associated with increased risk of osteolysis and aseptic loosening requiring revision surgery, compared with patients in Group A (7/133 vs 0/133, respectively; p = 0.007) and Group B (7/133 vs 0/135, respectively; p = 0.007). There was a non-statistically significant trend towards increased mean liner wear rates in Group A compared with Group B (0.031 mm/yr (SD 0.07) vs 0.022 mm/yr (SD 0.05), respectively; p = 0.128). All three groups were statistically comparable preoperatively and at ten years’ follow-up when measuring normalized Western Ontario and McMaster Universities Osteoarthritis Index (p = 0.410), 36-Item Short Form Health Survey (p = 0.465 mental, p = 0.713 physical), and pain scale scores (p = 0.451). Conclusion. The use of UHMWPE was associated with progressively increased annual liner wear rates after THA compared to XLPE. At ten years’ follow-up, the group receiving UHMWPE demonstrated an increased incidence of osteolysis and aseptic loosening requiring revision surgery compared to XLPE. Femoral heads composed of OxZr were associated with trend towards reduced wear rates compared to CoCr, but this did not reach statistical significance and did not translate to any differences in osteolysis, functional outcomes, or revision surgery between the two femoral head components. Cite this article: Bone Joint J 2022;104-B(7):833–843


The Bone & Joint Journal
Vol. 103-B, Issue 7 Supple B | Pages 66 - 72
1 Jul 2021
Hernandez NM Hinton ZW Wu CJ Lachiewicz PF Ryan SP Wellman SS

Aims. Modular dual mobility (MDM) acetabular components are often used with the aim of reducing the risk of dislocation in revision total hip arthroplasty (THA). There is, however, little information in the literature about its use in this context. The aim of this study, therefore, was to evaluate the outcomes in a cohort of patients in whom MDM components were used at revision THA, with a mean follow-up of more than five years. Methods. Using the database of a single academic centre, 126 revision THAs in 117 patients using a single design of an MDM acetabular component were retrospectively reviewed. A total of 94 revision THAs in 88 patients with a mean follow-up of 5.5 years were included in the study. Survivorship was analyzed with the endpoints of dislocation, reoperation for dislocation, acetabular revision for aseptic loosening, and acetabular revision for any reason. The secondary endpoints were surgical complications and the radiological outcome. Results. The overall rate of dislocation was 11%, with a six-year survival of 91%. Reoperation for dislocation was performed in seven patients (7%), with a six-year survival of 94%. The dislocations were early (at a mean of 33 days) in six patients, and late (at a mean of 4.3 years) in four patients. There were three intraprosthetic dissociations. An outer head diameter of ≥ 48 mm was associated with a lower risk of dislocation (p = 0.013). Lumbrosacral fusion was associated with increased dislocation (p = 0.004). Four revision THAs (4%) were further revised for aseptic acetabular loosening, and severe bone loss (Paprosky III) at the time of the initial revision was significantly associated with further revision for aseptic acetabular loosening (p = 0.008). Fourteen acetabular components (15%) were re-revised for infection, and a pre-revision diagnosis of reimplantation after periprosthetic joint infection (PJI) was associated with subsequent PJI (p < 0.001). Two THAs had visible metallic changes on the backside of the cobalt chromium liner. Conclusion. When using this MDM component in revision THA, at a mean follow-up of 5.5 years, there was a higher rate of dislocation (11%) than previously reported. The size of the outer bearing was related to the risk of dislocation. There was a low rate of aseptic acetabular loosening. Longer follow-up of this MDM component and evaluation of other designs are warranted. Cite this article: Bone Joint J 2021;103-B(7 Supple B):66–72


The Bone & Joint Journal
Vol. 103-B, Issue 7 Supple B | Pages 129 - 134
1 Jul 2021
Ayekoloye CI Abu Qa'oud M Radi M Leon SA Kuzyk P Safir O Gross AE

Aims. Improvements in functional results and long-term survival are variable following conversion of hip fusion to total hip arthroplasty (THA) and complications are high. The aim of the study was to analyze the clinical and functional results in patients who underwent conversion of hip fusion to THA using a consistent technique and uncemented implants. Methods. A total of 39 hip fusion conversions to THA were undertaken in 38 patients by a single surgeon employing a consistent surgical technique and uncemented implants. Parameters assessed included Harris Hip Score (HHS) for function, range of motion (ROM), leg length discrepancy (LLD), satisfaction, and use of walking aid. Radiographs were reviewed for loosening, subsidence, and heterotopic ossification (HO). Postoperative complications and implant survival were assessed. Results. At mean 12.2 years (2 to 24) follow-up, HHS improved from mean 34.2 (20.8 to 60.5) to 75 (53.6 to 94.0; p < 0.001). Mean postoperative ROM was flexion 77° (50° to 95°), abduction 30° (10° to 40°), adduction 20° (5° to 25°), internal rotation 18° (2° to 30°), and external rotation 17° (5° to 30°). LLD improved from mean -3.36 cm (0 to 8) to postoperative mean -1.14 cm (0 to 4; p < 0.001). Postoperatively, 26 patients (68.4%) required the use of a walking aid. Complications included one (2.5%) dislocation, two (5.1%) partial sciatic nerve injuries, one (2.5%) deep periprosthetic joint infection, two instances of (5.1%) acetabular component aseptic loosening, two (5.1%) periprosthetic fractures, and ten instances of HO (40%), of which three (7.7%) were functionally limiting and required excision. Kaplan-Meier Survival was 97.1% (95% confidence interval (CI) 91.4% to 100%) at ten years and 88.2% (95% CI 70.96 to 100) at 15 years with implant revision for aseptic loosening as endpoint and 81.7% (95% CI 70.9% to 98.0%) at ten years and 74.2% (95% CI 55.6 to 92.8) at 15 years follow-up with implant revision for all cause failure as endpoint. Conclusion. The use of an optimal and consistent surgical technique and cementless implants can result in significant functional improvement, low complication rates, long-term implant survival, and high patient satisfaction following conversion of hip fusion to THA. The possibility of requiring a walking aid should be discussed with the patient before surgery. Cite this article: Bone Joint J 2021;103-B(7 Supple B):129–134


The Bone & Joint Journal
Vol. 103-B, Issue 3 | Pages 492 - 499
1 Mar 2021
Garcia-Rey E Saldaña L Garcia-Cimbrelo E

Aims. Bone stock restoration of acetabular bone defects using impaction bone grafting (IBG) in total hip arthroplasty may facilitate future re-revision in the event of failure of the reconstruction. We hypothesized that the acetabular bone defect during re-revision surgery after IBG was smaller than during the previous revision surgery. The clinical and radiological results of re-revisions with repeated use of IBG were also analyzed. Methods. In a series of 382 acetabular revisions using IBG and a cemented component, 45 hips (45 patients) that had failed due to aseptic loosening were re-revised between 1992 and 2016. Acetabular bone defects graded according to Paprosky during the first and the re-revision surgery were compared. Clinical and radiological findings were analyzed over time. Survival analysis was performed using a competing risk analysis. Results. Intraoperative bone defect during the initial revision included 19 Paprosky type IIIA and 29 Paprosky type IIIB hips; at re-revision, seven hips were Paprosky type II, 27 type IIIA and 11 were type IIIB (p = 0.020). The mean preoperative Harris Hip Score was 45.4 (SD 6.4), becoming 80.7 (SD 12.7) at the final follow-up. In all, 12 hips showed radiological migration of the acetabular component, and three required further revision surgery. The nine-year cumulative failure incidence (nine patients at risk) of the acetabular component for further revision surgery was 9.6% (95% confidence interval (CI) 2.9 to 21.0) for any cause, and 7.5% (95% CI 1.9 to 18.5) for aseptic loosening. Hips with a greater hip height had a higher risk for radiological migration (odds ratio 1.09, 95% CI 1.02 to 1.17; p = 0.008). Conclusion. Bone stock restoration can be obtained using IBG in revision hip surgery. This technique is also useful in re-revision surgery; however, a better surgical technique including a closer distance to hip rotation centre could decrease the risk of radiological migration of the acetabular component. A longer follow-up is required to assess potential fixation deterioration. Cite this article: Bone Joint J 2021;103-B(3):492–499


The Bone & Joint Journal
Vol. 105-B, Issue 7 | Pages 768 - 774
1 Jul 2023
Wooster BM Kennedy NI Dugdale EM Sierra RJ Perry KI Berry DJ Abdel MP

Aims. Contemporary outcomes of primary total hip arthroplasties (THAs) with highly cross-linked polyethylene (HXLPE) liners in patients with inflammatory arthritis have not been well studied. This study examined the implant survivorship, complications, radiological results, and clinical outcomes of THA in patients with inflammatory arthritis. Methods. We identified 418 hips (350 patients) with a primary diagnosis of inflammatory arthritis who underwent primary THA with HXLPE liners from January 2000 to December 2017. Of these hips, 68% had rheumatoid arthritis (n = 286), 13% ankylosing spondylitis (n = 53), 7% juvenile rheumatoid arthritis (n = 29), 6% psoriatic arthritis (n = 24), 5% systemic lupus erythematosus (n = 23), and 1% scleroderma (n = 3). Mean age was 58 years (SD 14.8), 66.3% were female (n = 277), and mean BMI was 29 kg/m. 2. (SD 7). Uncemented femoral components were used in 77% of cases (n = 320). Uncemented acetabular components were used in all patients. Competing risk analysis was used accounting for death. Mean follow-up was 4.5 years (2 to 18). Results. The ten-year cumulative incidence of any revision was 3%, and was highest in psoriatic arthritis patients (16%). The most common indications for the 15 revisions were dislocations (n = 8) and periprosthetic joint infections (PJI; n = 4, all on disease-modifying antirheumatic drugs (DMARDs)). The ten-year cumulative incidence of reoperation was 6.1%, with the most common indications being wound infections (six cases, four on DMARDs) and postoperative periprosthetic femur fractures (two cases, both uncemented femoral components). The ten-year cumulative incidence of complications not requiring reoperation was 13.1%, with the most common being intraoperative periprosthetic femur fracture (15 cases, 14 uncemented femoral components; p = 0.13). Radiological evidence of early femoral component subsidence was observed in six cases (all uncemented). Only one femoral component ultimately developed aseptic loosening. Harris Hip Scores substantially improved (p < 0.001). Conclusion. Contemporary primary THAs with HXLPE in patients with inflammatory arthritis had excellent survivorship and good functional outcomes regardless of fixation method. Dislocation, PJI, and periprosthetic fracture were the most common complications in this cohort with inflammatory arthritis. Cite this article: Bone Joint J 2023;105-B(7):768–774


The Bone & Joint Journal
Vol. 106-B, Issue 2 | Pages 136 - 143
1 Feb 2024
van der Lelij TJN Marang-van de Mheen PJ Kaptein BL Koster LA Ljung P Nelissen RGHH Toksvig-Larsen S

Aims. The objective of this study was to compare the two-year migration and clinical outcomes of a new cementless hydroxyapatite (HA)-coated titanium acetabular shell with its previous version, which shared the same geometrical design but a different manufacturing process for applying the titanium surface. Methods. Overall, 87 patients undergoing total hip arthroplasty (THA) were randomized to either a Trident II HA or Trident HA shell, each cementless with clusterholes and HA-coating. All components were used in combination with a cemented Exeter V40 femoral stem. Implant migration was measured using radiostereometric analysis (RSA), with radiographs taken within two days of surgery (baseline), and at three, 12, and 24 months postoperatively. Proximal acetabular component migration was the primary outcome measure. Clinical scores and patient-reported outcome measures (PROMs) were collected at each follow-up. Results. Mean proximal migrations at three, 12, and 24 months were 0.08 mm (95% confidence interval (CI) 0.03 to 0.14), 0.11 mm (95% CI 0.06 to 0.16), and 0.14 mm (95% CI 0.09 to 0.20), respectively, in the Trident II HA group, versus 0.11 mm (95% CI 0.06 to 0.16), 0.12 mm (95% CI 0.07 to 0.17), and 0.14 mm (95% CI 0.09 to 0.19) in the Trident HA group (p = 0.875). No significant differences in translations or rotations between the two designs were found in any other direction. Clinical scores and PROMs were comparable between groups, except for an initially greater postoperative improvement in Hip disability and Osteoarthritis Outcome Symptoms score in the Trident HA group (p = 0.033). Conclusion. The Trident II clusterhole HA shell has comparable migration with its predecessor, the Trident hemispherical HA cluster shell, suggesting a similar risk of long-term aseptic loosening. Cite this article: Bone Joint J 2024;106-B(2):136–143


The Bone & Joint Journal
Vol. 102-B, Issue 12 | Pages 1670 - 1674
5 Dec 2020
Khan T Middleton R Alvand A Manktelow ARJ Scammell BE Ollivere BJ

Aims. To determine mortality risk after first revision total hip arthroplasty (THA) for periprosthetic femoral fracture (PFF), and to compare this to mortality risk after primary and first revision THA for other common indications. Methods. The study cohort consisted of THAs recorded in the National Joint Registry between 2003 and 2015, linked to national mortality data. First revision THAs for PFF, infection, dislocation, and aseptic loosening were identified. We used a flexible parametric model to estimate the cumulative incidence function of death at 90 days, one year, and five years following first revision THA and primary THA, in the presence of further revision as a competing risk. Analysis covariates were age, sex, and American Society of Anesthesiologists (ASA) grade. Results. A total of 675,078 primary and 74,223 first revision THAs were included (of which 6,131 were performed for PFF). Following revision for PFF, mortality ranged from 9% at 90 days, 21% at one year, and 60% at five years in the highest risk group (males, ≥ 75 years, ASA ≥ 3) to 0.6%, 1.4%, and 5.5%, respectively, for the lowest risk group (females, < 75 years, ASA ≤ 2). Mortality was greater in all groups following first revision THA for PFF than for primary THA. Compared to mortality risk after first revision THA for infection, dislocation, or aseptic loosening, revision for PFF was associated with higher five-year mortality in all groups except males < 75 years with an ASA ≤ 2. Conclusion. Mortality risk after revision THA for PFF is high, reaching 60% at five years in the highest risk patient group. In comparison to other common indications for revision, PFF demonstrated the highest overall risk of mortality at five years. These estimates can be used in the surgical decision-making process and when counselling patients and carers regarding surgical risk. Cite this article: Bone Joint J 2020;102-B(12):1670–1674


The Bone & Joint Journal
Vol. 103-B, Issue 7 Supple B | Pages 122 - 128
1 Jul 2021
Tibbo ME Limberg AK Gausden EB Huang P Perry KI Yuan BJ Berry DJ Abdel MP

Aims. The prevalence of ipsilateral total hip arthroplasty (THA) and total knee arthroplasty (TKA) is rising in concert with life expectancy, putting more patients at risk for interprosthetic femur fractures (IPFFs). Our study aimed to assess treatment methodologies, implant survivorship, and IPFF clinical outcomes. Methods. A total of 76 patients treated for an IPFF from February 1985 to April 2018 were reviewed. Prior to fracture, at the hip/knee sites respectively, 46 femora had primary/primary, 21 had revision/primary, three had primary/revision, and six had revision/revision components. Mean age and BMI were 74 years (33 to 99) and 30 kg/m. 2. (21 to 46), respectively. Mean follow-up after fracture treatment was seven years (2 to 24). Results. Overall, 59 fractures were classified as Vancouver C (Unified Classification System (UCS) D), 17 were Vancouver B (UCS B). In total, 57 patients (75%) were treated with open reduction and internal fixation (ORIF); three developed nonunion, three developed periprosthetic joint infection, and two developed aseptic loosening. In all, 18 patients (24%) underwent revision arthroplasty including 13 revision THAs, four distal femoral arthroplasties (DFAs), and one revision TKA: of these, one patient developed aseptic loosening and two developed nonunion. Survivorship free from any reoperation was 82% (95% confidence interval (CI) 66.9% to 90.6%) and 77% (95% CI 49.4% to 90.7%) in the ORIF and revision groups at two years, respectively. ORIF patients who went on to union tended to have stemmed knee components and greater mean interprosthetic distance (IPD = 189 mm (SD 73.6) vs 163 mm (SD 36.7); p = 0.546) than nonunited fractures. Patients who went on to nonunion in the revision arthroplasty group had higher medullary diameter: cortical width ratio (2.5 (SD 1.7) vs 1.3 (SD 0.3); p = 0.008) and lower IPD (36 mm (SD 30.6) vs 214 mm (SD 32.1); p < 0.001). At latest follow-up, 95% of patients (n = 72) were ambulatory. Conclusion. Interprosthetic femur fractures are technically and biologically challenging cases. Individualized approaches to internal fixation versus revision arthroplasty led to an 81% (95% CI 68.3% to 88.6%) survivorship free from reoperation at two years with 95% of patients ambulatory. Continued improvements in management are warranted. Cite this article: Bone Joint J 2021;103-B(7 Supple B):122–128