After obtaining informed consent, 80 patients were randomised to undergo a navigated or conventional total knee replacement. All received a cemented,
Wear of the polyethylene (PE) tibial insert of total knee arthroplasty (TKA) increases the risk of revision surgery with a significant cost burden on the healthcare system. This study quantifies wear performance of tibial inserts in a large and diverse series of retrieved TKAs to evaluate the effect of factors related to the patient, knee design, and bearing material on tibial insert wear performance. An institutional review board-approved retrieval archive was surveyed for modular PE tibial inserts over a range of in vivo duration (mean 58 months (0 to 290)). Five knee designs, totalling 1,585 devices, were studied. Insert wear was estimated from measured thickness change using a previously published method. Linear regression statistical analyses were used to test association of 12 patient and implant design variables with calculated wear rate.Aims
Methods
We reviewed 100 cemented Endo-model rotating-hinge (Waldemar Link GMBH &
Co, Hamburg, Germany) total knee replacements in 80 patients with a mean age of 70 years (56 to 85) at a mean post-operative follow-up of 11 years (7 to 15). Good or excellent results were seen in 91% of knees and survival at 15 years was 96.1%. There were two deep infections, one dislocation and one supracondylar fracture. There were no peri-operative deaths and there was no loosening, malalignment, migration or wear. We consider this prosthesis to be ideally suited for the replacement of the deformed knee when the use of an
The long-term success of total knee replacement is multifactorial, including factors relating to the patient, the operation and the implant. The purpose of this study was to examine the 20-year survival of the cemented Anatomical Graduated Component (AGC) total knee replacement. Between 1983 and 2004, 7760 of these were carried out at our institution. Of these, 6726 knees which received the non-modular metal-backed tibial component with compression-moulded polyethylene and had a minimum two-year follow-up were available for study. In all, 36 knees were followed over 20 years with a survival of the tibial and femoral components together of 97.8% (95% confidence interval (CI) 0.9851 to 0.9677), with no implants being revised for polyethylene wear or osteolysis. Age >
70 was associated with increased survival (99.6%, 95% CI 99.0 to 99.8) (p <
0.0001) but pre-operative valgus alignment reduced survival (95.1%, 95% CI 90.0 to 97.6) (p = 0.0056). Age <
55 (p = 0.129), pre-operative varus alignment (p = 0.707), osteonecrosis (p = 0.06), rheumatoid arthritis (p = 0.247), and gender (p = 0.666) were not statistically associated with failure. We attribute the success of the AGC implant to its relatively
The aim of this study was to compare the clinical outcomes of robotic arm-assisted bi-unicompartmental knee arthroplasty (bi-UKA) with conventional mechanically aligned total knee arthroplasty (TKA) during the first six weeks and at one year postoperatively. A per protocol analysis of 76 patients, 43 of whom underwent TKA and 34 of whom underwent bi-UKA, was performed from a prospective, single-centre, randomized controlled trial. Diaries kept by the patients recorded pain, function, and the use of analgesics daily throughout the first week and weekly between the second and sixth weeks. Patient-reported outcome measures (PROMs) were compared preoperatively, and at three months and one year postoperatively. Data were also compared longitudinally and a subgroup analysis was conducted, stratified by preoperative PROM status.Aims
Methods
Anterior cruciate ligament (ACL) and multiligament knee (MLK) injuries increase the risk of development of knee osteoarthritis and eventual need for total knee arthroplasty (TKA). There are limited data regarding implant use and outcomes in these patients. The aim of this study was to compare the use of constrained implants and outcomes among patients undergoing TKA with a history of prior knee ligament reconstruction (PKLR) Patients with a history of ACL or MLK reconstruction who underwent TKA between 2007 and 2017 were identified in a single-institution registry. There were 223 patients who met inclusion criteria (188 ACL reconstruction patients, 35 MLK reconstruction patients). A matched cohort, also of 223 patients, was identified based on patient age, body mass index (BMI), sex, and year of surgery. There were 144 male patients and 79 female patients in both cohorts. Mean age at the time of TKA was 57.2 years (31 to 88). Mean BMI was 29.7 kg/m2 (19.5 to 55.7).Aims
Patients and Methods
Antibiotic-loaded bone cements (ALBCs) may offer early protection against the formation of bacterial biofilm after joint arthroplasty. Use in hip arthroplasty is widely accepted, but there is a lack of evidence in total knee arthroplasty (TKA). The objective of this study was to evaluate the use of ALBC in a large population of TKA patients. Data from the National Joint Registry (NJR) of England and Wales were obtained for all primary cemented TKAs between March 2003 and July 2016. Patient, implant, and surgical variables were analyzed. Cox proportional hazards models were used to assess the influence of ALBC on risk of revision. Body mass index (BMI) data were available in a subset of patients.Aims
Materials and Methods
Modern total knee arthroplasty (TKA) prostheses are designed to restore near normal kinematics including high flexion. Kneeling is a high flexion, kinematically demanding activity after TKA. The debate about design choice has not yet been informed by six-degrees-of-freedom in vivo kinematics. This prospective randomized clinical trial compared kneeling kinematics in three TKA designs. In total, 68 patients were randomized to either a posterior stabilized (PS-FB), cruciate-retaining (CR-FB), or rotating platform (CR-RP) design. Of these patients, 64 completed a minimum one year follow-up. Patients completed full-flexion kneeling while being imaged using single-plane fluoroscopy. Kinematics were calculated by registering the 3D implant models onto 2D-dynamic fluoroscopic images and exported for analysis.Aims
Methods
The aim of this study was to compare robotic arm-assisted bi-unicompartmental knee arthroplasty (bi-UKA) with conventional mechanically aligned total knee arthroplasty (TKA) in order to determine the changes in the anatomy of the knee and alignment of the lower limb following surgery. An analysis of 38 patients who underwent TKA and 32 who underwent bi-UKA was performed as a secondary study from a prospective, single-centre, randomized controlled trial. CT imaging was used to measure coronal, sagittal, and axial alignment of the knee preoperatively and at three months postoperatively to determine changes in anatomy that had occurred as a result of the surgery. The hip-knee-ankle angle (HKAA) was also measured to identify any differences between the two groups.Aims
Methods
A retrospective longitudinal study was conducted to compare directly volumetric wear of retrieved polyethylene inserts to predicted volumetric wear modelled from individual gait mechanics of total knee arthroplasty (TKA) patients. In total, 11 retrieved polyethylene tibial inserts were matched with gait analysis testing performed on those patients. Volumetric wear on the articular surfaces was measured using a laser coordinate measure machine and autonomous reconstruction. Knee kinematics and kinetics from individual gait trials drove computational models to calculate medial and lateral tibiofemoral contact paths and forces. Sliding distance along the contact path, normal forces and implantation time were used as inputs to Archard’s equation of wear to predict volumetric wear from gait mechanics. Measured and modelled wear were compared for each component.Aims
Methods
The primary aim of this study was to compare the knee-specific functional outcome of patellofemoral arthroplasty with total knee arthroplasty (TKA) in the management of patients with patellofemoral osteoarthritis. A total of 54 consecutive Avon patellofemoral arthroplasties were identified and propensity-score-matched to a group of 54 patients undergoing a TKA with patellar resurfacing for patellofemoral osteoarthritis. The Oxford Knee Score (OKS), the 12-Item Short-Form Health Survey (SF-12), and patient satisfaction were collected at a mean follow up of 9.2 years (8 to 15). Survival was defined by revision or intention to revise.Aims
Patients and Methods
The objectives of this study were to compare postoperative pain, analgesia requirements, inpatient functional rehabilitation, time to hospital discharge, and complications in patients undergoing conventional jig-based unicompartmental knee arthroplasty (UKA) This prospective cohort study included 146 patients with symptomatic medial compartment knee osteoarthritis undergoing primary UKA performed by a single surgeon. This included 73 consecutive patients undergoing conventional jig-based mobile bearing UKA, followed by 73 consecutive patients receiving robotic-arm assisted fixed bearing UKA. All surgical procedures were performed using the standard medial parapatellar approach for UKA, and all patients underwent the same postoperative rehabilitation programme. Postoperative pain scores on the numerical rating scale and opiate analgesia consumption were recorded until discharge. Time to attainment of predefined functional rehabilitation outcomes, hospital discharge, and postoperative complications were recorded by independent observers.Aims
Patients and Methods
The aim of this independent multicentre study was to assess the
mid-term results of mobile bearing unicondylar knee arthroplasty
(UKA) for isolated lateral osteoarthritis of the knee joint. We retrospectively evaluated 363 consecutive, lateral UKAs (346
patients) performed using the Oxford domed lateral prosthesis undertaken
in three high-volume knee arthroplasty centres between 2006 and
2014. Mean age of the patients at surgery was 65 years (36 to 88)
with a mean final follow-up of 37 months (12 to 93)Aims
Patients and Methods
Total knee arthroplasty (TKA) is a cost effective
and extremely successful operation. As longevity increases, the demand
for primary TKA will continue to rise. The success and survivorship
of TKAs are dependent on the demographics of the patient, surgical
technique and implant-related factors. Currently the risk of failure of a TKA requiring revision surgery
ten years post-operatively is 5%. The most common indications for revision include aseptic loosening
(29.8%), infection (14.8%), and pain (9.5%). Revision surgery poses
considerable clinical burdens on patients and financial burdens
on healthcare systems. We present a current concepts review on the epidemiology of failed
TKAs using data from worldwide National Joint Registries. Cite this article:
Our aim was to compare kinematic with mechanical alignment in
total knee arthroplasty (TKA). We performed a prospective blinded randomised controlled trial
to compare the functional outcome of patients undergoing TKA in
mechanical alignment (MA) with those in kinematic alignment (KA).
A total of 71 patients undergoing TKA were randomised to either
kinematic (n = 36) or mechanical alignment (n = 35). Pre- and post-operative
hip-knee-ankle radiographs were analysed. The knee injury and osteoarthritis
outcome score (KOOS), American Knee Society Score, Short Form-36,
Euro-Qol (EQ-5D), range of movement (ROM), two minute walk, and timed
up and go tests were assessed pre-operatively and at six weeks,
three and six months and one year post-operatively.Aims
Patients and Methods
There have been differing descriptions of the
anterolateral structures of the knee, and not all have been named
or described clearly. The aim of this study was to provide a clear
anatomical interpretation of these structures. We dissected 40 fresh-frozen
cadaveric knees to view the relevant anatomy and identified a consistent
structure in 33 knees (83%); we termed this the anterolateral ligament
of the knee. This structure passes antero-distally from an attachment
proximal and posterior to the lateral femoral epicondyle to the
margin of the lateral tibial plateau, approximately midway between
Gerdy’s tubercle and the head of the fibula. The ligament is superficial
to the lateral (fibular) collateral ligament proximally, from which
it is distinct, and separate from the capsule of the knee. In the
eight knees in which it was measured, we observed that the ligament
was isometric from 0° to 60° of flexion of the knee, then slackened
when the knee flexed further to 90° and was lengthened by imposing
tibial internal rotation. Cite this article:
The use of hinged implants in primary total knee
replacement (TKR) should be restricted to selected indications and mainly
for elderly patients. Potential indications for a rotating hinge
or pure hinge implant in primary TKR include: collateral ligament
insufficiency, severe varus or valgus deformity (>
20°) with necessary
relevant soft-tissue release, relevant bone loss including insertions
of collateral ligaments, gross flexion-extension gap imbalance,
ankylosis, or hyperlaxity. Although data reported in the literature
are inconsistent, clinical results depend on implant design, proper
technical use, and adequate indications. We present our experience
with a specific implant type that we have used for over 30 years
and which has given our elderly patients good mid-term results.
Because revision of implants with long cemented stems can be very
challenging, an effort should be made in the future to use shorter stems
in modular versions of hinged implants. Cite this article:
We compared patient-reported outcomes of the Kinemax fixed- and mobile-bearing total knee replacement in a multi-centre randomised controlled trial. Patients were randomised to the fixed- or the mobile-bearing prosthesis via a sealed envelope method after the bone cuts had been made in the operating theatre. Randomisation was stratified by centre and diagnosis. Patients were assessed pre-operatively and at eight to 12 weeks, one year and two years post-operatively. Validated questionnaires were used which included the Western Ontario MacMasters University, Short-Form 12, Mental Health Index-5, Knee Injury and Osteoarthritis Outcome Score for Knee-Related Quality of Life and Function in Sport and Recreation scales and a validated scale of satisfaction post-operatively. In total, 242 patients (250 knees) with a mean age of 68 years (40 to 80) were recruited from four NHS orthopaedic centres. Of these, 132 patients (54.5%) were women. No statistically significant differences could be identified in any of the patient-reported outcome scores between patients who received the fixed-bearing or the mobile-bearing knee up to two-years post-operatively.
We compared the results of 146 patients who received an anatomic modular knee fixed-bearing total knee replacement (TKR) in one knee and a low contact stress rotating platform mobile-bearing TKR in the other. There were 138 women and eight men with a mean age of 69.8 years (42 to 80). The mean follow-up was 13.2 years (11.0 to 14.5). The patients were assessed clinically and radiologically using the rating systems of the Hospital for Special Surgery and the Knee Society at three months, six months, one year, and annually thereafter. The assessment scores of both rating systems pre-operatively and at the final review did not show any statistically significant differences between the two designs of implant. In the anatomic modular knee group, one knee was revised because of aseptic loosening of the tibial component and one because of infection. In addition, three knees were revised because of wear of the polyethylene tibial bearing. In the low contact stress group, two knees were revised because of instability requiring exchange of the polyethylene insert and one because of infection. The radiological analysis found no statistical difference in the incidence of radiolucent lines at the final review (Student’s We found no evidence of the superiority of one design over the other at long-term follow-up.
This prospective study compares the outcome of
157 hydroxyapatite (HA)-coated tibial components with 164 cemented
components in the ROCC Rotating Platform total knee replacement
in 291 patients. The mean follow-up was 7.6 years (5.2 to 11). There
were two revisions for loosening: one for an HA-coated and one for
a cemented tibial component. Radiological evaluation demonstrated
no radiolucent lines with the HA-coated femoral components. A total
of three HA-coated tibial components exhibited radiolucent lines
at three months post-operatively and these disappeared after three
further months of protected weight-bearing. With HA-coated components
the operating time was shorter (p <
0.006) and the radiological
assessment of the tibial interface was more stable (p <
0.01).
Using revision for aseptic loosening of the tibial component as
the end point, the survival rates at nine years was identical for
both groups at 99.1%. Our results suggest that HA-coated components perform at least
as well as the same design with cemented components and compare
favourably with those of series describing cemented or porous-coated
knee replacements, suggesting that fixation of both components with