The Fassier Duval (FD) rod is a third-generation telescopic implant for children with osteogenesis imperfecta (OI). Threaded fixation enables proximal insertion without opening the knee or ankle joint. We have reviewed our combined two-centre experience with this implant. In total, 34 children with a mean age of five years (1 to 14) with severe OI have undergone rodding of 72 lower limb long bones (27 tibial, 45 femoral) for recurrent fractures with progressive deformity despite optimized bone health and bisphosphonate therapy. Data were collected prospectively, with 1.5 to 11 years follow-up.Aims
Methods
Percutaneous epiphysiodesis using transphyseal
screws (PETS) has been developed for the treatment of lower limb discrepancies
with the aim of replacing traditional open procedures. The goal
of this study was to evaluate its efficacy and safety at skeletal
maturity. A total of 45 consecutive patients with a mean skeletal
age of 12.7 years (8.5 to 15) were included and followed until maturity.
The mean efficacy of the femoral epiphysiodesis was 35% (14% to 87%)
at six months and 66% (21% to 100%) at maturity. The mean efficacy
of the tibial epiphysiodesis was 46% (18% to 73%) at six months
and 66% (25% to 100%) at maturity. In both groups of patients the
under-correction was significantly reduced between six months post-operatively
and skeletal maturity. The overall rate of revision was 18% (eight
patients), and seven of these revisions (87.5%) involved the tibia.
This series showed that use of the PETS technique in the femur was
safe, but that its use in the tibia was associated with a significant
rate of complications, including a valgus deformity in nine patients
(20%), leading us to abandon it in the tibia. The arrest of growth
was delayed and the final loss of growth at maturity was only 66%
of that predicted pre-operatively. This should be taken into account
in the pre-operative planning.