The purpose of this study was to assess mid-term survivorship following primary total knee arthroplasty (TKA) with Optetrak Logic components and identify the most common revision indications at a single institution. We identified a retrospective cohort of 7,941 Optetrak primary TKAs performed from January 2010 to December 2018. We reviewed the intraoperative findings of 369 TKAs that required revision TKA from January 2010 to December 2021 and the details of the revision implants used. Kaplan-Meier analysis was used to determine survivorship. Cox regression analysis was used to examine the impact of patient variables and year of implantation on survival time.Aims
Methods
A limited number of investigations with conflicting results have described perivascular lymphocytic infiltration (PVLI) in the setting of total knee arthroplasty (TKA). The purpose of this study was to determine if PVLI found in TKAs at the time of aseptic revision surgery was associated with worse clinical outcomes and survivorship. A retrospective review was conducted on 617 patients who underwent aseptic TKA revision who had histological analysis for PVLI at the time of surgery. Clinical and radiological data were obtained pre- and postoperatively, six weeks postoperatively, and then every year thereafter.Aims
Methods
This study aimed to determine outcomes of isolated tibial insert exchange (ITIE) during revision total knee arthroplasty (TKA). From 1985 to 2016, 270 ITIEs were performed at one institution for instability (55%, n = 148), polyethylene wear (39%, n = 105), insert fracture/dissociation (5%, n = 14), or stiffness (1%, n = 3). Patients with component loosening, implant malposition, infection, and extensor mechanism problems were excluded.Aims
Methods
To assess migration of the tibial component we used roentgen stereophotogrammetric analysis in 40 patients who had had a
Since redesign of the Oxford phase III mobile-bearing unicompartmental
knee arthroplasty (UKA) femoral component to a twin-peg design,
there has not been a direct comparison to total knee arthroplasty
(TKA). Thus, we explored differences between the two cohorts. A total of 168 patients (201 knees) underwent medial UKA with
the Oxford Partial Knee Twin-Peg. These patients were compared with
a randomly selected group of 177 patients (189 knees) with primary
Vanguard TKA. Patient demographics, Knee Society (KS) scores and
range of movement (ROM) were compared between the two cohorts. Additionally,
revision, re-operation and manipulation under anaesthesia rates
were analysed.Aims
Patients and Methods
Little is known about employment following total knee arthroplasty
(TKA). This study aims to identify factors which predict return
to work following TKA in patients of working age in the United Kingdom. We prospectively assessed 289 patients (289 TKAs) aged ≤ 65 years
who underwent TKA between 2010 and 2013. There were 148 women. The
following were recorded pre-operatively: age, gender, body mass
index, social deprivation, comorbidities, indication for surgery,
work status and nature of employment, activity level as assessed by
the University of California, Los Angeles (UCLA) activity score
and Oxford Knee Score (OKS). The intention of patients to return
to work or to retire was not assessed pre-operatively. At a mean
of 3.4 years (2 to 4) post-operatively, the return to work status,
OKS, the EuroQol-5 dimensions (EQ-5D) score, UCLA activity score
and Work, Osteoarthritis and joint-Replacement (WORQ) score were
obtained. Univariate and multivariate analyses were performed.Aims
Patients &
Methods
As the number of younger and more active patients
treated with total knee arthroplasty (TKA) continues to increase,
consideration of better fixation as a means of improving implant
longevity is required. Cemented TKA remains the reference standard
with the largest body of evidence and the longest follow-up to support
its use. However, cementless TKA, may offer the opportunity of a
more bone-sparing procedure with long lasting biological fixation
to the bone. We undertook a review of the literature examining advances
of cementless TKA and the reported results. Cite this article:
Previous studies of failure mechanisms leading
to revision total knee replacement (TKR) performed between 1986 and
2000 determined that many failed early, with a disproportionate
amount accounted for by infection and implant-associated factors
including wear, loosening and instability. Since then, efforts have
been made to improve implant performance and instruct surgeons in
best practice. Recently our centre participated in a multi-centre evaluation
of 844 revision TKRs from 2010 to 2011. The purpose was to report
a detailed analysis of failure mechanisms over time and to see if
failure modes have changed over the past 10 to 15 years. Aseptic
loosening was the predominant mechanism of failure (31.2%), followed
by instability (18.7%), infection (16.2%), polyethylene wear (10.0%),
arthrofibrosis (6.9%) and malalignment (6.6%). The mean time to
failure was 5.9 years (ten days to 31 years), 35.3% of all revisions
occurred at less than two years, and 60.2% in the first five years.
With improvements in implant and polyethylene manufacture, polyethylene
wear is no longer a leading cause of failure. Early mechanisms of
failure are primarily technical errors. In addition to improving
implant longevity, industry and surgeons must work together to decrease
these technical errors. All reports on
This study reports on the first 150 consecutive
Oxford cementless unicompartmental knee arthroplasties (UKA) performed
in an independent centre (126 patients). All eligible patients had
functional scores (Oxford knee score and high activity arthroplasty
score) recorded pre-operatively and at two- and five-years of follow-up. Fluoroscopically
aligned radiographs were taken at five years and analysed for any
evidence of radiolucent lines (RLLs), subsidence or loosening. The
mean age of the cohort was 63.6 years (39 to 86) with 81 (53.1%)
males. Excellent functional scores were maintained at five years
and there were no progressive RLLs demonstrated on radiographs.
Two patients underwent revision to a total knee arthroplasty giving
a revision rate of 0.23/100 (95% confidence interval 0.03 to 0.84)
component years with overall component survivorship of 98.7% at
five years. There were a further four patients who underwent further
surgery on the same knee, two underwent bearing exchanges for dislocation
and two underwent lateral UKAs for disease progression. This was
a marked improvement from other UKAs reported in New Zealand Joint
Registry data and supports the designing centre’s early results. Cite this article:
The strain on clinic and surgeon resources resulting
from a rise in demand for total knee replacement (TKR) requires reconsideration
of when and how often patients need to be seen for follow-up. Surgeons
will otherwise require increased paramedical staff or need to limit
the number of TKRs they undertake. We reviewed the outcome data
of 16 414 primary TKRs undertaken at our centre to determine the
time to re-operation for any reason and for specific failure mechanisms.
Peak risk years for failure were determined by comparing the conditional
probability of failure, the number of failures divided by the total
number of TKRs cases, for each year. The median times to failure
for the most common failure mechanisms were 4.9 years (interquartile
range (IQR) 1.7 to 10.7) for femoral and tibial loosening, 1.9 years
(IQR 0.8 to 3.9) for infection, 3.1 years (IQR 1.6 to 5.5) for tibial
collapse and 5.6 years (IQR 3.4 to 9.3) for instability. The median
time to failure for all revisions was 3.3 years (IQR 1.2 to 8.5),
with an overall revision rate of 1.7% (n = 282). Results from our
patient population suggest that patients be seen for follow-up at
six months, one year, three years, eight years, 12 years, and every
five years thereafter. Patients with higher pain in the early post-operative
period or high body mass index (≥ 41 kg/m2) should be
monitored more closely. Cite this article:
Minimally invasive total knee replacement (MIS-TKR)
has been reported to have better early recovery than conventional
TKR. Quadriceps-sparing (QS) TKR is the least invasive MIS procedure,
but it is technically demanding with higher reported rates of complications
and outliers. This study was designed to compare the early clinical
and radiological outcomes of TKR performed by an experienced surgeon
using the QS approach with or without navigational assistance (NA),
or using a mini-medial parapatellar (MP) approach. In all, 100 patients
completed a minimum two-year follow-up: 30 in the NA-QS group, 35
in the QS group, and 35 in the MP group. There were no significant
differences in clinical outcome in terms of ability to perform a
straight-leg raise at 24 hours (p = 0.700), knee score (p = 0.952),
functional score (p = 0.229) and range of movement (p = 0.732) among
the groups. The number of outliers for all three radiological parameters
of mechanical axis, frontal femoral component alignment and frontal
tibial component alignment was significantly lower in the NA-QS
group than in the QS group (p = 0.008), but no outlier was found
in the MP group. In conclusion, even after the surgeon completed a substantial
number of cases before the commencement of this study, the supplementary
intra-operative use of computer-assisted navigation with QS-TKR
still gave inferior radiological results and longer operating time,
with a similar outcome at two years when compared with a MP approach. Cite this article:
We carried out a prospective study of 132 patients (159 knees) who underwent closed-wedge high tibial osteotomy for severe medial compartment osteoarthritis between 1988 and 1997. A total of 94 patients (118 knees) was available for review at a mean of 16.4 years (16 to 20). Seven patients (7.4%) (11 knees) required conversion to total knee replacement. Kaplan-Meier survival was 97.6% (95% confidence interval 95.0 to 100) at ten years and 90.4% (95% confidence interval 84.1 to 96.7) at 15 years. Excellent and good results as assessed by the Hospital for Special Surgery knee score were achieved in 87 knees (73.7%). A pre-operative body mass index >
27.5 kg/m2 and range of movement <
100° were risk factors predicting early failure. Although our long-term results were satisfactory, strict indications for osteotomy are required if long-term survival is required.