We prospectively reviewed 1000 consecutive patients who underwent a cementless, hydroxyapatite-coated, stemless, total knee replacement over a period of nine years. Regular post-operative clinical follow-up was performed using the Knee Society score. The mean pre-operative score was 96, improving to 182 and 180 at five and ten years, respectively. To date, there have been seven (0.5%) cases which required revision, primarily for septic loosening (four cases), with low rates of other post-operative complications. The cumulative survival at ten years with revision as the end-point, was 99.14% (95% confidence interval 92.5 to 99.8). These results support the use of hydroxyapatite in a cementless total knee replacement since it can give reliable fixation with an excellent clinical and functional outcome.
We carried out a prospective study of 118 hydroxyapatite-coated, cementless total knee replacements in patients who were ≤ 55 years of age and who had primary (92; 78%) or post-traumatic (26; 22%) osteoarthritis. The mean period of follow-up was 7.9 years (5 to 12.5). The Knee Society clinical scores improved from a pre-operative mean of 98 (0 to 137) to a mean of 185 (135 to 200) at five years, and 173 (137 to 200) at ten years. There were two revisions of the tibial component because of aseptic loosening, and one case of polyethylene wear requiring further surgery. There was no osteolysis or progressive radiological loosening of any other component. At 12 years, the overall rate of implant survival was 97.5% (excluding exchange of spacer) and 92.1% (including exchange of spacer). Cementless total knee replacement can achieve excellent long-term results in young, active patients with osteoarthritis. In contrast to total hip replacement, polyethylene wear, osteolysis and loosening of the prosthesis were not major problems for these patients, although it is possible that this observation could change with longer periods of follow-up.