Dreyer L Bader C Flörkemeier T Wagner M

Aims. The risk of mechanical failure of modular revision hip stems is frequently mentioned in the literature, but little is currently known about the actual clinical failure rates of this type of prosthesis. The current retrospective long-term analysis examines the distal and modular failure patterns of the Prevision hip stem from 18 years of clinical use. A design improvement of the modular taper was introduced in 2008, and the data could also be used to compare the original and the current design of the modular connection. Methods. We performed an analysis of the Prevision modular hip stem using the manufacturer’s vigilance database and investigated different mechanical failure patterns of the hip stem from January 2004 to December 2022. Results. Two mechanical failure patterns were identified: fractures in the area of the distal fluted profile (distal stem fracture) and failure of the modular taper (modular fracture). A failure rate of 0.07% was observed for distal stem fracture, and modular fracture rates of 1.74% for the original and 0.013% for the current taper design. Conclusion. A low risk of mechanical failure for both fracture types was observed compared to other known complications in revision hip arthroplasty. In addition, the data show that a design change did significantly reduce the risk of a modular fracture. Cite this article: Bone Joint J 2024;106-B(2):151–157


The Bone & Joint Journal
Vol. 105-B, Issue 5 | Pages 504 - 510
1 May 2023
Evans JT Salar O Whitehouse SL Sayers A Whitehouse MR Wilton T Hubble MJW

Aims

The Exeter V40 femoral stem is the most implanted stem in the National Joint Registry (NJR) for primary total hip arthroplasty (THA). In 2004, the 44/00/125 stem was released for use in ‘cement-in-cement’ revision cases. It has, however, been used ‘off-label’ as a primary stem when patient anatomy requires a smaller stem with a 44 mm offset. We aimed to investigate survival of this implant in comparison to others in the range when used in primary THAs recorded in the NJR.

Methods

We analyzed 328,737 primary THAs using the Exeter V40 stem, comprising 34.3% of the 958,869 from the start of the NJR to December 2018. Our exposure was the stem, and the outcome was all-cause construct revision. We stratified analyses into four groups: constructs using the 44/00/125 stem, those using the 44/0/150 stem, those including a 35.5/125 stem, and constructs using any other Exeter V40 stem.


The Bone & Joint Journal
Vol. 104-B, Issue 2 | Pages 212 - 220
1 Feb 2022
Fishley WG Selvaratnam V Whitehouse SL Kassam AM Petheram TG

Aims

Femoral cement-in-cement revision is a well described technique to reduce morbidity and complications in hip revision surgery. Traditional techniques for septic revision of hip arthroplasty necessitate removal of all bone cement from the femur. In our two centres, we have been using a cement-in-cement technique, leaving the distal femoral bone cement in selected patients for septic hip revision surgery, both for single and the first of two-stage revision procedures. A prerequisite for adoption of this technique is that the surgeon considers the cement mantle to be intimately fixed to bone without an intervening membrane between cement and host bone. We aim to report our experience for this technique.

Methods

We have analyzed patients undergoing this cement-in-cement technique for femoral revision in infection, and present a consecutive series of 89 patients. Follow-up was undertaken at a mean of 56.5 months (24.0 to 134.7) for the surviving cases.


The Bone & Joint Journal
Vol. 104-B, Issue 2 | Pages 206 - 211
1 Feb 2022
Bloch BV White JJE Matar HE Berber R Manktelow ARJ

Aims

Total hip arthroplasty (THA) is a very successful and cost-effective operation, yet debate continues about the optimum fixation philosophy in different age groups. The concept of the 'cementless paradox' and the UK 'Getting it Right First Time' initiative encourage increased use of cemented fixation due to purported lower revision rates, especially in elderly patients, and decreased cost.

Methods

In a high-volume, tertiary referral centre, we identified 10,112 THAs from a prospectively collected database, including 1,699 cemented THAs, 5,782 hybrid THAs, and 2,631 cementless THAs. The endpoint was revision for any reason. Secondary analysis included examination of implant survivorship in patients aged over 70 years, over 75 years, and over 80 years at primary THA.


The Bone & Joint Journal
Vol. 103-B, Issue 7 | Pages 1215 - 1221
1 Jul 2021
Kennedy JW Ng NYB Young D Kane N Marsh AG Meek RMD

Aims

Cement-in-cement revision of the femoral component represents a widely practised technique for a variety of indications in revision total hip arthroplasty. In this study, we compare the clinical and radiological outcomes of two polished tapered femoral components.

Methods

From our prospectively collated database, we identified all patients undergoing cement-in-cement revision from January 2005 to January 2013 who had a minimum of two years' follow-up. All cases were performed by the senior author using either an Exeter short revision stem or the C-Stem AMT high offset No. 1 prosthesis. Patients were followed-up annually with clinical and radiological assessment.


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 5 | Pages 577 - 582
1 May 2009
Duncan WW Hubble MJW Howell JR Whitehouse SL Timperley AJ Gie GA

The removal of well-fixed bone cement from the femoral canal during revision of a total hip replacement (THR) can be difficult and risks the loss of excessive bone stock and perforation or fracture of the femoral shaft. Retaining the cement mantle is attractive, yet the technique of cement-in-cement revision is not widely practised. We have used this procedure at our hospital since 1989. The stems were removed to gain a better exposure for acetabular revision, to alter version or leg length, or for component incompatibility. We studied 136 hips in 134 patients and followed them up for a mean of eight years (5 to 15). A further revision was required in 35 hips (25.7%), for acetabular loosening in 26 (19.1%), sepsis in four, instability in three, femoral fracture in one and stem fracture in one. No femoral stem needed to be re-revised for aseptic loosening. A cement-in-cement revision of the femoral stem is a reliable technique in the medium term. It also reduces the risk of perforation or fracture of the femoral shaft


The Bone & Joint Journal
Vol. 103-B, Issue 8 | Pages 1339 - 1344
1 Aug 2021
Jain S Mohrir G Townsend O Lamb JN Palan J Aderinto J Pandit H

Aims

This aim of this study was to assess the reliability and validity of the Unified Classification System (UCS) for postoperative periprosthetic femoral fractures (PFFs) around cemented polished taper-slip (PTS) stems.

Methods

Radiographs of 71 patients with a PFF admitted consecutively at two centres between 25 February 2012 and 19 May 2020 were collated by an independent investigator. Six observers (three hip consultants and three trainees) were familiarized with the UCS. Each PFF was classified on two separate occasions, with a mean time between assessments of 22.7 days (16 to 29). Interobserver reliability for more than two observers was assessed using percentage agreement and Fleiss’ kappa statistic. Intraobserver reliability between two observers was calculated with Cohen kappa statistic. Validity was tested on surgically managed UCS type B PFFs where stem stability was documented in operation notes (n = 50). Validity was assessed using percentage agreement and Cohen kappa statistic between radiological assessment and intraoperative findings. Kappa statistics were interpreted using Landis and Koch criteria. All six observers were blinded to operation notes and postoperative radiographs.


The Bone & Joint Journal
Vol. 103-B, Issue 2 | Pages 309 - 320
1 Feb 2021
Powell-Bowns MFR Oag E Ng N Pandit H Moran M Patton JT Clement ND Scott CEH

Aims

The aim of this study was to determine whether fixation, as opposed to revision arthroplasty, can be safely used to treat reducible Vancouver B type fractures in association with a cemented collarless polished tapered femoral stem (the Exeter).

Methods

This retrospective cohort study assessed 152 operatively managed consecutive unilateral Vancouver B fractures involving Exeter stems; 130 were managed with open reduction and internal fixation (ORIF) and 22 with revision arthroplasty. Mean follow-up was 6.5 years (SD 2.6; 3.2 to 12.1). The primary outcome measure was revision of at least one component. Kaplan–Meier survival analysis was performed. Regression analysis was used to identify risk factors for revision following ORIF. Secondary outcomes included any reoperation, complications, blood transfusion, length of hospital stay, and mortality.


The Bone & Joint Journal
Vol. 103-B, Issue 2 | Pages 299 - 304
1 Feb 2021
Goto E Umeda H Otsubo M Teranishi T

Aims

Various surgical techniques have been described for total hip arthroplasty (THA) in patients with Crowe type III dislocated hips, who have a large acetabular bone defect. The aim of this study was to evaluate the long-term clinical results of patients in whom anatomical reconstruction of the acetabulum was performed using a cemented acetabular component and autologous bone graft from the femoral neck.

Methods

A total of 22 patients with Crowe type III dislocated hips underwent 28 THAs using bone graft from the femoral neck between 1979 and 2000. A Charnley cemented acetabular component was placed at the level of the true acetabulum after preparation with bone grafting. All patients were female with a mean age at the time of surgery of 54 years (35 to 68). A total of 18 patients (21 THAs) were followed for a mean of 27.2 years (20 to 33) after the operation.


The Bone & Joint Journal
Vol. 102-B, Issue 10 | Pages 1289 - 1296
1 Oct 2020
Amstutz HC Le Duff M

Aims

Hip resurfacing arthroplasty (HRA) is typically indicated for young and active patients. Due to the longevity of arthroplasty, these patients are likely to undergo revision surgery during their lifetime. There is a paucity of information on the long-term outcome of revision surgeries performed after failed HRA. The aim of our study was to provide survivorship data as well as clinical scores after HRA revisions.

Methods

A total of 42 patients (43 hips) were revised after HRA at our centre to a variety of devices, including four HRA and 39 total hip arthroplasties (THAs). In addition to perioperative complications, University of California, Los Angeles (UCLA) hip scores and 12-Item Short-Form Health Survey questionnaire (SF-12) quality of life scores were collected at follow-up visits after the primary HRA and after revision surgery.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 2 | Pages 167 - 172
1 Feb 2012
Buttaro MA Costantini J Comba F Piccaluga F

We determined the midterm survival, incidence of peri-prosthetic fracture and the enhancement of the width of the femur when combining struts and impacted bone allografts in 24 patients (25 hips) with severe femoral bone loss who underwent revision hip surgery. The pre-operative diagnosis was aseptic loosening in 16 hips, second-stage reconstruction in seven, peri-prosthetic fracture in one and stem fracture in one hip. A total of 14 hips presented with an Endoklinik grade 4 defect and 11 hips a grade 3 defect. The mean pre-operative Merle D’Aubigné and Postel score was 5.5 points (1 to 8). The survivorship was 96% (95% confidence interval 72 to 98) at a mean of 54.5 months (36 to 109). The mean functional score was 17.3 points (16 to 18). One patient in which the strut did not completely bypass the femoral defect was further revised using a long cemented stem due to peri-prosthetic fracture at six months post-operatively. The mean subsidence of the stem was 1.6 mm (1 to 3). There was no evidence of osteolysis, resorption or radiolucencies during follow-up in any hip. Femoral width was enhanced by a mean of 41% (19% to 82%). A total of 24 hips had partial or complete bridging of the strut allografts. This combined biological method was associated with a favourable survivorship, a low incidence of peri-prosthetic fracture and enhancement of the width of the femur in revision total hip replacement in patients with severe proximal femoral bone loss


The Bone & Joint Journal
Vol. 101-B, Issue 9 | Pages 1050 - 1057
1 Sep 2019
Lampropoulou-Adamidou K Hartofilakidis G

Aims

To our knowledge, no study has compared the long-term results of cemented and hybrid total hip arthroplasty (THA) in patients with osteoarthritis (OA) secondary to congenital hip disease (CHD). This is a demanding procedure that may require special techniques and implants. Our aim was to compare the long-term outcome of cemented low-friction arthroplasty (LFA) and hybrid THA performed by one surgeon.

Patients and Methods

Between January 1989 and December 1997, 58 hips (44 patients; one man, 43 woman; mean age 56.6 years (25 to 77)) with OA secondary to CHD were treated with a cemented Charnley LFA (group A), and 55 hips (39 patients; two men, 37 women; mean age 49.1 years (27 to 70)) were treated with a hybrid THA (group B), by the senior author (GH). The clinical outcome and survivorship were compared.


The Journal of Bone & Joint Surgery British Volume
Vol. 80-B, Issue 2 | Pages 273 - 278
1 Mar 1998
Wroblewski BM Fleming PA Hall RM Siney PD

We report a prospective study of the use of intramedullary bone blocks to improve the fixation of a matt-finish femoral stem in Charnley low-friction arthroplasties. There were 379 patients (441 hips), but at a minimum follow-up of ten years there were 258 arthroplasties in 221 patients including some which had been revised. The mean age at surgery was 41 years (17 to 51) and the mean follow-up was 13.4 years (1 to 20 including the early revisions). Nine stems (3.5%) had been revised for aseptic loosening, but there were no stem fractures. Survivorship of stems was 99.2% at ten years and 94.35% at 15 and 20 years. We found that the patient’s gender, the position of the stem and the experience of the surgeon all influenced the outcome. Our findings suggest that using our method of stem fixation, follow-up of over 11 years was needed to reveal the effects of endosteal cavitation of the femur, and of over 13 years to assess any divergence between the clinical and the radiological outcomes of stem fixation


The Journal of Bone & Joint Surgery British Volume
Vol. 80-B, Issue 3 | Pages 404 - 410
1 May 1998
Matsui M Nakata K Masuhara K Ohzono K Sugano N Ochi T

We implanted 51 Metal-Cancellous Cementless Lübeck (MCCL) prostheses into 45 patients with dysplastic hips and followed 49 hips (96.1%) for five to nine years. One had needed revision for stem fracture and one for infection; the clinical outcome of the other 47 hips was assessed using the Merle d’Aubigné and Postel hip score. All hips were either excellent (63%) or good (37%). Three patients (6%) had mild thigh pain at six months, but this had settled within two years. Serial radiographs showed stable fixation with bone ingrowth in all hips, with increased density of the cancellous bone in contact with the implant and some trabecular ingrowth. There was early varus shift of the stem in one hip, but this stabilised in three months. Osteolysis of the femoral cortex was seen in one hip at seven years after surgery, and mild bone resorption due to stress shielding in 31 (63%). Acetabular bone grafting with autogenous bone from the femoral head gave successful support to the socket in 13 hips. The MCCL prosthesis gave satisfactory mid-term results in patients with osteoarthritis secondary to hip dysplasia


The Bone & Joint Journal
Vol. 101-B, Issue 10 | Pages 1199 - 1208
1 Oct 2019
Lamb JN Matharu GS Redmond A Judge A West RM Pandit HG

Aims

We compared implant and patient survival following intraoperative periprosthetic femoral fractures (IOPFFs) during primary total hip arthroplasty (THA) with matched controls.

Patients and Methods

This retrospective cohort study compared 4831 hips with IOPFF and 48 154 propensity score matched primary THAs without IOPFF implanted between 2004 and 2016, which had been recorded on a national joint registry. Implant and patient survival rates were compared between groups using Cox regression.


The Bone & Joint Journal
Vol. 98-B, Issue 4 | Pages 468 - 474
1 Apr 2016
Abdel MP Houdek MT Watts CD Lewallen DG Berry DJ

Aims. The goals of this study were to define the risk factors, characteristics, and chronology of fractures in 5417 revision total hip arthroplasties (THAs). . Patients and Methods. From our hospital’s prospectively collected database we identified all patients who had undergone a revision THA between 1969 and 2011 which involved the femoral stem. The patients’ medical records and radiographs were examined and the relevant data extracted. Post-operative periprosthetic fractures were classified using the Vancouver system. A total of 5417 revision THAs were identified. Results. There were 668 intra-operative fractures, giving an incidence of 12%. Fractures were three times more common with uncemented stems (19%) than with cemented stems (6%) (p <  0.001). The incidence of intra-operative femoral fracture varied by uncemented stem type: fully-coated (20%); proximally-coated (19%); modular fluted tapered (16%) (p < 0.05). Most fractures occurred during the insertion of the femoral component (35%). One-third involved the diaphysis and 26% were of the calcar: 69% were undisplaced. There were 281 post-operative fractures of the femur (20-year probability = 11%). There was no difference in risk for cemented and uncemented stems. Post-operative fractures were more common in men < 70 years (p = 0.02). Periprosthetic fractures occurred earlier after uncemented revision of the femoral component, but later after a cemented revision. The most common fracture type was a Vancouver B. 1. (31%). Of all post-operative fractures, 24% underwent open reduction and internal fixation and 15% revision arthroplasty. Conclusion. In revision THA, intra-operative fractures occurred three times more often with an uncemented stem. Many were undisplaced diaphyseal fractures treated with cerclage fixation. . While the risk of post-operative fracture is similar between uncemented and cemented components, they occur at notably different times depending on the type of stem fixation. . Take home message: In revision THA, intra-operative periprosthetic femoral fractures occur three times more often with uncemented stems. Many are non-displaced diaphyseal fractures treated with cerclage fixation. While postoperative fracture risks are equivalent between uncemented and cemented components, they occur at notably different time periods based on stem fixation type. Cite this article: Bone Joint J 2016;98-B:468–74


The Bone & Joint Journal
Vol. 101-B, Issue 1 | Pages 96 - 103
1 Jan 2019
Colo E Leenders LAM Rijnen WHC Schreurs BW Hannink G

Aims

The aim of this study was to analyze the effect of a lateral rim mesh on the survival of primary total hip arthroplasty (THA) in young patients, aged 50 years or younger.

Patients and Methods

We compared a study group of 235 patients (257 hips) who received a primary THA with the use of impaction bone grafting (IBG) with an additional lateral rim mesh with a group of 306 patients (343 hips) who received IBG in the absence of a lateral rim mesh during the same period from 1988 to 2015. In the mesh group, there were 74 male and 183 female patients, with a mean age of 35 years (13 to 50). In the no-mesh group, there were 173 male and 170 female patients, with a mean age of 38 years (12.6 to 50). Cox regression analyses were performed to study the effect of a lateral rim mesh on acetabular component survival. Kaplan–Meier analyses with 95% confidence intervals (CIs) were performed to estimate the survival of the acetabular implant.


The Bone & Joint Journal
Vol. 100-B, Issue 11 | Pages 1499 - 1454
1 Nov 2018
Green CM Buckley SC Hamer AJ Kerry RM Harrison TP

Aims

The management of acetabular defects at the time of revision hip arthroplasty surgery is a challenge. This study presents the results of a long-term follow-up study of the use of irradiated allograft bone in acetabular reconstruction.

Patients and Methods

Between 1990 and 2000, 123 hips in 110 patients underwent acetabular reconstruction for aseptic loosening, using impaction bone grafting with frozen, irradiated, and morsellized femoral heads and a cemented acetabular component. A total of 55 men and 55 women with a mean age of 64.3 years (26 to 97) at the time of revision surgery are included in this study.


The Bone & Joint Journal
Vol. 100-B, Issue 12 | Pages 1565 - 1571
1 Dec 2018
Kristensen TB Dybvik E Furnes O Engesæter LB Gjertsen J

Aims

The aim of this large registry-based study was to compare mid-term survival rates of cemented femoral stems of different designs used in hemiarthroplasty for a fracture of the femoral neck.

Patients and Methods

From the Norwegian Hip Fracture Register (NHFR), 20 532 primary cemented bipolar hemiarthroplasties, which were undertaken in patients aged > 70 years with a femoral neck fracture between 2005 and 2016, were included. Polished tapered stems (n = 12 065) (Exeter and CPT), straight stems (n = 5545) (Charnley, Charnley Modular, and Spectron EF), and anatomical stems (n = 2922) (Lubinus SP2) were included. The survival of the implant with any reoperation as the endpoint was calculated using the Kaplan–Meier method and hazard ratios (HRs), and the different indications for reoperation were calculated using Cox regression analysis.


The Bone & Joint Journal
Vol. 100-B, Issue 8 | Pages 1002 - 1009
1 Aug 2018
Westerman RW Whitehouse SL Hubble MJW Timperley AJ Howell JR Wilson MJ

Aims

The aim of this study was to report the initial results of the Exeter V40 stem, which became available in 2000.

Patients and Methods

A total of 540 total hip arthroplasties (THAs) were performed in our unit using this stem between December 2000 and May 2002. Our routine protocol is to review patients postoperatively and at one, five, and ten years following surgery.