We compared lower limb coronal alignment measurements
obtained pre- and post-operatively with long-leg radiographs and
computer navigation in patients undergoing primary total knee replacement
(TKR). A series of 185 patients had their pre- and post-implant
radiological and computer-navigation system measurements of coronal alignment
compared using the Bland-Altman method. The study included 81 men
and 104 women with a mean age of 68.5 years (32 to 87) and a mean
body mass index of 31.7 kg/m2 (19 to 49). Pre-implant
Bland–Altman limits of agreement were -9.4° to 8.6° with a repeatability
coefficient of 9.0°. The Bland–Altman plot showed a tendency for the
radiological measurement to indicate a higher level of pre-operative
deformity than the corresponding navigation measurement. Post-implant
limits of agreement were -5.0° to 5.4° with a repeatability coefficient
of 5.2°. The tendency for valgus knees to have greater deformity
on the radiograph was still seen, but was weaker for varus knees. The alignment seen or measured intra-operatively during TKR is
not necessarily the same as the deformity seen on a standing long-leg
radiograph either pre- or post-operatively. Further investigation
into the effect of weight-bearing and surgical exposure of the joint
on the mechanical femorotibial angle is required to enable the most appropriate
intra-operative alignment to be selected.
We analysed 42 weight-bearing lateral radiographs of the ankle, 20 of which were from patients with a clinical and plain
In this retrospective study we have assessed the results of low tibial valgus osteotomy for varus-type osteoarthritis of the ankle and its indications. We performed an opening wedge osteotomy in 25 women (26 ankles). The mean follow-up was for eight years and three months (2 years 3 months to 17 years 11 months). Of the 26 ankles, 19 showed excellent or good clinical results. Their mean scores for pain, walking, and activities of daily living were significantly improved but there was no change in the range of movement. In the ankles which were classified
We describe a syndrome combining abnormalities of the pelvis, knee and foot in three related patients with a familial history of small dislocated patellae. The clinical and
We treated 35 brachymetatarsal rays of 18 feet in 12 patients by one-stage lengthening with interpositional bone grafts or by gradual lengthening with callotasis combined with shortening of the adjacent metatarsals and phalanges. Definition of the two parabolas which connect the metatarsal heads and the tips of the toes, and recognition of three patterns of metatarsal length, were helpful guides in treatment. In total, 36 excisions of the phalanges and/or the metatarsals were undertaken. The mean shortening was 8 mm. The
Of 111 primary ankle ligament arthroplasties (modified Evans procedure) performed between 1983 and 1994, we were able to identify 89 patients (94 ankles) for follow-up. All were under 50 years of age. Two had died and one refused to co-operate; 86 patients (91 ankles) were therefore reviewed, 25 by telephone and the remainder by clinical examination with all but three also undergoing
This retrospective study analyses the results of subtalar bone block distraction arthrodesis used in the treatment of late complications of calcaneal fractures, acute severely comminuted fractures, nonunion (and malunion) of attempted subtalar arthrodeses, avascular necrosis of the talus, and club-foot deformity. Of 39 patients (41 feet) who had this procedure, 35 (37 feet) returned for follow-up after a mean of 70 months (26 to 140). There were 24 men (25 feet) and 11 women (12 feet) with a mean age of 41 years (16 to 63). Each completed a standardised questionnaire, based on the hindfoot-scoring system of the American Orthopaedic Foot and Ankle Society and were reviewed both clinically and
Patients undergoing femoral lengthening by external fixation
tolerate treatment less well when compared to tibial lengthening.
Lengthening of the femur with an intramedullary device may have
advantages. We reviewed all cases of simple femoral lengthening performed
at our unit from 2009 to 2014. Cases of nonunions, concurrent deformities,
congenital limb deficiencies and lengthening with an unstable hip
were excluded, leaving 33 cases (in 22 patients; 11 patients had
bilateral procedures) for review. Healing index, implant tolerance
and complications were compared.Aims
Patients and Methods
This study describes the Osseointegration Group of Australia’s
Accelerated Protocol two-stage strategy (OGAAP-1) for the osseointegrated
reconstruction of amputated limbs. We report clinical outcomes in 50 unilateral trans-femoral amputees
with a mean age of 49.4 years (24 to 73), with a minimum one-year
follow-up. Outcome measures included the Questionnaire for persons
with a Trans-Femoral Amputation, the health assessment questionnaire
Short-Form-36 Health Survey, the Amputation Mobility Predictor scores
presented as K-levels, 6 Minute Walk Test and timed up and go tests.
Adverse events included soft-tissue problems, infection, fractures
and failure of the implant.Aims
Patients and Methods
Intramedullary infection in long bones represents
a complex clinical challenge, with an increasing incidence due to the
increasing use of intramedullary fixation. We report a prospective
case series using an intramedullary reaming device, the Reamer–Irrigator–Aspirator
(RIA) system, in association with antibiotic cement rods for the
treatment of lower limb long bone infections. A total of 24 such
patients, 16 men and eight women, with a mean age of 44.5 years
(17 to 75), 14 with femoral and 10 with tibial infection, were treated
in a staged manner over a period of 2.5 years in a single referral
centre. Of these, 21 patients had had previous surgery, usually
for fixation of a fracture (seven had sustained an open fracture
originally and one had undergone fasciotomies). According to the
Cierny–Mader classification system, 18 patients were classified
as type 1A, four as 3A (discharging sinus tract), one as type 4A
and one as type 1B. Cite this article:
Although gradual bone transport may permit the
restoration of large-diameter bones, complications are common owing
to the long duration of external fixation. In order to reduce such
complications, a new technique of bone transport involving the use
of an external fixator and a locking plate was devised for segmental
tibial bone defects. A total of ten patients (nine men, one woman) with a mean age
at operation of 40.4 years (16 to 64) underwent distraction osteogenesis
with a locking plate to treat previously infected post-traumatic
segmental tibial defects. The locking plate was fixed percutaneously
to bridge proximal and distal segments, and was followed by external fixation.
After docking, percutaneous screws were fixed at the transported
segment through plate holes. At the same time, bone grafting was
performed at the docking site with the external fixator removed. The mean defect size was 5.9 cm (3.8 to 9.3) and mean external
fixation index was
13.4 days/cm (11.8 to 19.5). In all cases, primary union of the
docking site and distraction callus was achieved, with an excellent
bony result. There was no recurrence of deep infection or osteomyelitis,
and with the exception of one patient with a pre-existing peroneal
nerve injury, all achieved an excellent or good functional result. With short external fixation times and low complication rates,
bone transport with a locking plate could be recommended for patients
with segmental tibial defects. Cite this article:
Between October 2001 and September 2009 we lengthened 242 lower-limb segments in 180 patients using the Intramedullary Skeletal Kinetic Distractor (ISKD). Mechanical failure was defined either as breakage of the ISKD or failure of the internal mechanism to activate. Retrieved nails which failed mechanically were examined by the manufacturer for defects. In all, 15 ISKDs in 12 patients (13 limbs) failed mechanically representing an overall failure rate of 6.2%, with fracture of the device occurring in ten of the 15 failures. Two nails in one patient failed to lengthen and had to be replaced. The manufacturer detected an error in the assembly of the nail, which prompted a wide recall. One nail jammed after being forcefully inserted, and two nails failed to lengthen fully. Lengthening was achieved in all 12 patients, although three required a second operation to exchange a defective nail for a new, functioning device. The ISKD is a complex mechanical device which lengthens by the oscillation of two telescopic sections connected by a threaded rod. The junction between these sections is surrounded by a keyring collar. This keyring collar is the weakest part of the device.
Internal lengthening devices in the femur lengthen
along the anatomical axis, potentially creating lateral shift of
the mechanical axis. We aimed to determine whether femoral lengthening
along the anatomical axis has an inadvertent effect on lower limb
alignment. Isolated femoral lengthening using the Intramedullary
Skeletal Kinetic Distractor was performed in 27 femora in 24 patients
(mean age 32 years (16 to 57)). Patients who underwent simultaneous realignment
procedures or concurrent tibial lengthening, or who developed mal-
or nonunion, were excluded. Pre-operative and six-month post-operative
radiographs were used to measure lower limb alignment. The mean lengthening
achieved was 4.4 cm (1.5 to 8.0). In 26 of 27 limbs, the mechanical
axis shifted laterally by a mean of 1.0 mm/cm of lengthening (0
to 3.5). In one femur that was initially in varus, a 3 mm medial
shift occurred during a lengthening of 2.2 cm. In a normally aligned limb, intramedullary lengthening along
the anatomical axis of the femur results in a lateral shift of the
mechanical axis by approximately 1 mm for each 1 cm of lengthening.
We report the results of intramedullary leg lengthening conducted between 2002 and 2009 using the Intramedullary Skeletal Kinetic Distractor in 69 unilateral lengthenings involving 58 femora and 11 tibiae. We identified difficulties that occurred during the treatment and assessed whether they were specifically due to the implant or independent of it. Paley’s classification for evaluating problems, obstacles and complications with external fixators was adopted, and implant-specific difficulties were continuously noted. There were seven failures requiring premature removal of the device, in four due to nail breakage and three for other reasons, and five unsuccessful outcomes after completion of the lengthening. In all, 116 difficulties were noted in 45 patients, with only 24 having problem-free courses. In addition to the difficulties arising from the use of external fixators, there were almost the same number again of implant-specific difficulties. Nevertheless, successful femoral lengthening was achieved in 52 of the 58 patients (90%). However, successful tibial lengthening was only achieved in five of 11 patients (45%).
The use of autograft bone is the best option
when undertaking a procedure that requires bone graft because it
is osteogenic, osteoconductive and osseo-inductive. Pain, morbidity
and complications associated with harvesting iliac or non-iliac
sites occur in between 6% and 30% of cases. An alternative source
of graft with possibly a lower morbidity is the intramedullary canal.
In this study, 28 patients undergoing 30 arthrodesis procedures
on the hindfoot had a mean of 48 cm3 (43 to 50) of bone
harvested locally from the hindfoot or the tibial shaft by antegrade or
retrograde reaming. No patient sustained a fracture of the calcaneum,
talus or tibia. There was no morbidity except for one complication
when the reamer breached the medial tibial cortex. This healed uneventfully. This method of using the reamer–irrigator–aspirator system is
an extension of the standard technique of intramedullary reaming
of the lower limb: it produces good-quality bone graft with viable
growth factors consistent with that of the iliac crest, and donor
site morbidity is low. This is an efficient method of obtaining
autologous bone for use in arthrodesis of the ankle or hindfoot.
We investigated the fracture-free survival of long bones stabilised by a telescopic intramedullary rod (TIMR) in patients with osteogenesis imperfecta with respect to the remodelling status of fracture or osteotomy sites and TIMR regions, in order to identify risk factors for fracture. A total of 44 femora and 28 tibiae in 25 patients with a mean age of 5.0 years (1.9 to 10.5) at presentation were studied. There were six patients with Sillence type I, five with type III, 13 with type IV and one with type V osteogenesis imperfecta. All received bisphosphonate treatment at the same stage during the mean follow-up of 7.3 years (0.5 to 18.1). The fracture-free survival was estimated at 6.2 years (95% confidence interval 5.1 to 7.3) by Kaplan-Meier analysis. More than half the fracture or osteotomy sites remained in a less-remodelled state at the latest follow-up or time of fracture. Of the 33 fractures, 29 (87.9%) occurred in long bones containing a less-remodelled site, and these fractures were located at this site. The relative fracture risk at the rod tip was significantly greater than in any other TIMR region (p <
0.001), and this was higher in bone segments having a less-remodelled site. This study shows a persistent fracture risk in TIMR-stabilised long bones, especially at less-remodelled fracture or osteotomy sites and at the rod tip.
A silver-containing hydroxyapatite (Ag-HA) coating has been developed using thermal spraying technology. We evaluated the osteoconductivity of this coating on titanium (Ti) implants in rat tibiae in relation to bacterial infection in joint replacement. At 12 weeks, the mean affinity indices of bone formation of a Ti, an HA, a 3%Ag-HA and a 50%Ag-HA coating were 97.3%, 84.9%, 81.0% and 40.5%, respectively. The mean affinity indices of bone contact of these four coatings were 18.8%, 83.7%, 77.2% and 40.5%, respectively. The indices of bone formation and bone contact around the implant of the 3%Ag-HA coating were similar to those of the HA coating, and no significant differences were found between them (bone formation, p = 0.99; bone contact, p = 0.957). However, inhibition of bone formation was observed with the 50%Ag-HA coating. These results indicate that the 3%Ag-HA coating has low toxicity and good osteoconductivity, and that the effect of silver toxicity on osteoconductivity depends on the dose.
We compared the complications and outcome of tibial lengthening using the Ilizarov method with and without the use of a supplementary intramedullary nail. In a retrospective case-matched series assembled from 176 patients with tibial lengthening, we matched 52 patients (26 pairs, group A with nail and group B without) according to the following criteria in order of importance: 1) difference in amount of lengthening (± 2 cm); 2) percentage difference in lengthening (± 5%); 3) difference in patient’s age (± seven years); 4) aetiology of the shortening, and 5) level of difficulty in obtaining the correction. The outcome was evaluated using the external fixator index, the healing index and an outcome score according to the criteria of Paley. It was found that some complications were specific to group A or B respectively, but others were common to both groups. The outcome was generally better in lengthenings with a nail, although there was a higher incidence of rectifiable equinus deformity in these patients.
We present the results of the surgical correction of lower-limb deformities caused by metabolic bone disease. Our series consisted of 17 patients with a diagnosis of hypophosphataemic rickets and two with renal osteodystrophy; their mean age was 25.6 years (14 to 57). In all, 43 lower-limb segments (27 femora and 16 tibiae) were osteotomised and the deformity corrected using a monolateral external fixator. The segment was then stabilised with locked intramedullary nailing. In addition, six femora in three patients were subsequently lengthened by distraction osteogenesis. The mean follow-up was 60 months (18 to 120). The frontal alignment parameters (the mechanical axis deviation, the lateral distal femoral angle and the medial proximal tibial angle) and the sagittal alignment parameters (the posterior distal femoral angle and the posterior proximal tibial angle) improved post-operatively. The external fixator was removed either at the end of surgery or at the end of the lengthening period, allowing for early mobilisation and weight-bearing. We encountered five problems and four obstacles in the programme of treatment. The use of intramedullary nails prevented recurrence of deformity and refracture.
Advances in the design of the components for total ankle replacement have led to a resurgence of interest in this procedure. Between January 1999 and December 2004, 16 patients with a failed total ankle replacement were referred to our unit. In the presence of infection, a two-stage salvage procedure was planned. The first involved the removal of the components and the insertion of a cement spacer. Definitive treatment options included hindfoot fusion with a circular frame or amputation. When there was no infection, a one-stage salvage procedure was planned. Options included hindfoot fusion with an intramedullary nail or revision total ankle replacement. When there was suspicion of infection, a percutaneous biopsy was performed. The patients were followed up for a minimum of 12 months. Of the 16 patients, 14 had aseptic loosening, five of whom underwent a revision total ankle replacement and nine a hindfoot fusion. Of the two with infection, one underwent fusion and the other a below-knee amputation. There were no cases of wound breakdown, nonunion or malunion. Management of the failed total ankle replacement should be performed by experienced surgeons and ideally in units where multidisciplinary support is available. Currently, a hindfoot fusion appears to be preferable to a revision total ankle replacement.