The Gamma nail is frequently used in unstable
peri-trochanteric hip fractures. We hypothesised that mechanical failure
of the Gamma nail was associated with inadequate proximal three-point
fixation. We identified a consecutive series of 299 Gamma nails
implanted in 299 patients over a five-year period, 223 of whom fulfilled
our inclusion criteria for investigation. The series included 61
men and 162 women with a mean age of 81 years (20 to 101). Their fractures
were classified according to the Modified Evans’ classification
and the quality of fracture reduction was graded. The technical
adequacy of three points of
We investigated a new intramedullary locking
nail that allows the distal interlocking screws to be locked to
the nail. We compared fixation using this new implant with fixation
using either a conventional nail or a locking plate in a laboratory
simulation of an osteoporotic fracture of the distal femur. A total
of 15 human cadaver femora were used to simulate an AO 33-A3 fracture
pattern. Paired specimens compared fixation using either a locking
or non-locking retrograde nail, and using either a locking retrograde
nail or a locking plate. The constructs underwent cyclical loading
to simulate single-leg stance up to 125 000 cycles. Axial and torsional
stiffness and displacement, cycles to failure and modes of failure
were recorded for each specimen. When compared with locking plate
constructs, locking nail constructs had significantly longer mean
fatigue life (75 800 cycles ( The new locking retrograde femoral nail showed better stiffness
and fatigue life than locking plates, and superior fatigue life
to non-locking nails, which may be advantageous in elderly patients. Cite this article:
In a randomised trial involving 598 patients
with 600 trochanteric fractures of the hip, the fractures were treated with
either a sliding hip screw (n = 300) or a Targon PF intramedullary
nail (n = 300). The mean age of the patients was 82 years (26 to
104). All surviving patients were reviewed at one year with functional
outcome assessed by a research nurse blinded to the treatment used.
The intramedullary nail was found to have a slightly increased mean
operative time (46 minutes ( In summary, both implants produced comparable results but there
was a tendency to better return of mobility for those treated with
the intramedullary nail.
Fractures and nonunions of the proximal humerus are increasingly treated by open reduction and internal fixation. The extended deltopectoral approach remains the most widely used for this purpose. However, it provides only limited exposure of the lateral and posterior aspects of the proximal humerus. We have previously described the alternative extended deltoid-splitting approach. In this paper we outline variations and extensions of this technique that we have developed in the management of further patients with these fractures.