The Vantage Total Ankle System is a fourth-generation low-profile fixed-bearing implant that has been available since 2016. We aimed to describe our early experience with this implant. This is a single-centre retrospective review of patients who underwent primary total ankle arthroplasty (TAA) with a Vantage implant between November 2017 and February 2020, with a minimum of two years’ follow-up. Four surgeons contributed patients. The primary outcome was reoperation and revision rate of the Vantage implant at two years. Secondary outcomes included radiological alignment, peri-implant complications, and pre- and postoperative patient-reported outcomes.Aims
Methods
Preoperative talar valgus deformity ≥ 15° is considered a contraindication for total ankle arthroplasty (TAA). We compared operative procedures and clinical outcomes of TAA in patients with talar valgus deformity ≥ 15° and < 15°. A matched cohort of patients similar for demographics and components used but differing in preoperative coronal-plane tibiotalar valgus deformity ≥ 15° (valgus, n = 50; 52% male, mean age 65.8 years (SD 10.3), mean body mass index (BMI) 29.4 (SD 5.2)) or < 15° (control, n = 50; 58% male, mean age 65.6 years (SD 9.8), mean BMI 28.7 (SD 4.2)), underwent TAA by one surgeon. Preoperative and postoperative radiographs, Ankle Osteoarthritis Scale (AOS) pain and disability and 36-item Short Form Health Survey (SF-36) version 2 scores were collected prospectively. Ancillary procedures, secondary procedures, and complications were recorded.Aims
Methods
We report the medium-term outcomes of a consecutive series of 118 Zenith total ankle arthroplasties (TAAs) from a single, non-designer centre. Between December 2010 and May 2016, 118 consecutive Zenith prostheses were implanted in 114 patients. Demographic, clinical, and patient-reported outcome measures (PROMs) data were collected. The endpoint of the study was failure of the implant requiring revision of one or all of the components. Kaplan-Meier survival curves were generated with 95% confidence intervals (CIs) and the rate of failure calculated for each year.Aims
Methods
We report the long-term clinical and radiological outcomes of a consecutive series of 200 total ankle arthroplasties (TAAs, 184 patients) at a single centre using the Scandinavian Total Ankle Replacement (STAR) implants. Between November 1993 and February 2000, 200 consecutive STAR prostheses were implanted in 184 patients by a single surgeon. Demographic and clinical data were collected prospectively and the last available status was recorded for further survival analysis. All surviving patients underwent regular clinical and radiological review. Pain and function were assessed using the American Orthopaedic Foot and Ankle Society (AOFAS) hindfoot scoring system. The principal endpoint of the study was failure of the implant requiring revision of one or all of the components. Kaplan–Meier survival curves were generated with 95% confidence intervals and the rate of failure calculated for each year.Aims
Patients and Methods
The purpose of this study was to determine the functional outcome and implant survivorship of mobile-bearing total ankle arthroplasty (TAA) performed by a single surgeon. We reviewed 205 consecutive patients (210 ankles) who had undergone mobile-bearing TAA (205 patients) for osteoarthritis of the ankle between January 2005 and December 2015. Their mean follow-up was 6.4 years (2.0 to 13.4). Functional outcome was assessed using the Ankle Osteoarthritis Scale, American Orthopaedic Foot and Ankle Society (AOFAS) ankle-hindfoot score, 36-Item Short-Form Health Survey (SF-36) score, visual analogue scale, and range of movement. Implant survivorship and complications were also evaluated.Aims
Patients and Methods
The last decade has seen a considerable increase
in the use of in total ankle arthroplasty (TAA) to treat patients
with end-stage arthritis of the ankle. However, the longevity of
the implants is still far from that of total knee and hip arthroplasties. The aim of this review is to outline a diagnostic and treatment
algorithm for the painful TAA to be used when considering revision
surgery. Cite this article: