The Journal of Bone & Joint Surgery British Volume
Vol. 88-B, Issue 5 | Pages 606 - 613
1 May 2006
Abu-Rajab RB Watson WS Walker B Roberts J Gallacher SJ Meek RMD

We compared peri-prosthetic bone mineral density between identical cemented and cementless LCS rotating platform total knee arthroplasties. Two matched cohorts had dual energy x-ray absorptiometry scans two years post-operatively using a modified validated densitometric analysis protocol, to assess peri-prosthetic bone mineral density. The knee that was not operated on was also scanned to enable the calculation of a relative bone mineral density difference. Oxford Knee and American Knee Society scores were comparable in the two cohorts. Statistical analysis revealed no significant difference in absolute, or relative peri-prosthetic bone mineral density with respect to the method of fixation. However, the femoral peri-prosthetic bone mineral density and relative bone mineral density difference were significantly decreased, irrespective of the method of fixation, particularly in the anterior distal portion of the femur, with a mean reduction in relative bone mineral density difference of 27%. There was no difference in clinical outcome between the cemented and cementless LCS total knee arthroplasty. However, both produce stress-shielding around the femoral implants. This leads us to question the use of more expensive cementless total knee components


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 5 | Pages 656 - 659
1 May 2012
Hwang B Yoon J Nam C Jung K Lee S Han C Moon S

We retrospectively reviewed 30 two-stage revision procedures in 28 patients performed for fungal peri-prosthetic joint infection (PJI) after a primary total knee replacement. Patients were followed for at least two years or until the infection recurred. The mean follow-up for patients who remained free of infection was 4.3 years (2.3 to 6.1). Overall, 17 patients were assessed as American Society of Anesthesiologists grade 3 or 4. The surgical protocol included removal of the infected implant, vigorous debridement and insertion of an articulating cement spacer. This was followed by at least six weeks of antimicrobial treatment and delayed reimplantation in all patients. The mean interval between removal of the prosthesis and reimplantation was 9.5 weeks (6 to 24). After reimplantation, patients took antifungal agents orally for a maximum of six months. Two knees became reinfected at one and two months post-operatively, respectively: one of these subsequently required arthrodesis because of uncontrolled infection. Fungal PJIs can be treated successfully by removal of all infected material, appropriate antimicrobial treatment and delayed reimplantation


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 7 | Pages 874 - 878
1 Jul 2008
Fink B Makowiak C Fuerst M Berger I Schäfer P Frommelt L

We analysed the serum C-reactive protein level, synovial fluid obtained by joint aspiration and five synovial biopsies from 145 knee replacements prior to revision to assess the value of these parameters in diagnosing late peri-prosthetic infection. Five further synovial biopsies were used for histological analysis. Samples were also obtained during the revision and incubated and analysed in an identical manner for 14 days. A total of 40 total knee replacements were found to be infected (prevalence 27.6%). The aspiration technique had a sensitivity of 72.5% (95% confidence interval (CI) 58.7 to 86.3), a specificity of 95.2% (95% CI 91.2 to 99.2), a positive predictive value of 85.3% (95% CI 73.4 to 100), a negative predictive value of 90.1% (95% CI 84.5 to 95.7) and an accuracy of 89%. The biopsy technique had a sensitivity of 100%, a specificity of 98.1% (95% CI 95.5 to 100), a positive predictive value of 95.2% (95% CI 88.8 to 100), a negative predictive value of 100% and an accuracy of 98.6%. C-reactive protein with a cut-off-point of 13.5 mg/l had a sensitivity of 72.5% (95% CI 58.7 to 86.3), a specificity of 80.9% (95% CI 73.4 to 88.4), a positive predictive value of 59.2% (95% CI 45.4 to 73.0), a negative predictive value of 88.5% (95% 81.0 to 96.0 CI) and an accuracy of 78.1%. We found that biopsy was superior to joint aspiration and C-reactive protein in the diagnosis of late peri-prosthetic infection of total knee replacements


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 7 | Pages 908 - 913
1 Jul 2012
Seo JG Moon YW Park SH Lee JH Kang HM Kim SM

Peri-prosthetic patellar fracture following resurfacing as part of total knee replacement (TKR) is an infrequent yet challenging complication. This case-control study was performed to identify clinical, radiological and surgical factors that increase the risk of developing a spontaneous patellar fracture after TKR. Patellar fractures were identified in 74 patients (88 knees) from a series of 7866 consecutive TKRs conducted between 1998 and 2009. After excluding those with a previous history of extensor mechanism realignment or a clear traumatic event, a metal-backed patella, any uncemented component or subsequent infection, the remaining 64 fractures were compared with a matched group of TKRs with an excellent outcome defined by the Knee Society score. The mean age of patients with a fracture was 70 years (51 to 81) at the time of TKR. Patellar fractures were detected at a mean of 13.4 months (2 to 84) after surgery. The incidence of patellar fracture was found to be strongly associated with the number of previous knee operations, greater pre-operative mechanical malalignment, smaller post-operative patellar tendon length, thinner post-resection patellar thickness, and a lower post-operative Insall-Salvati ratio. An understanding of the risk factors associated with spontaneous patellar fracture following TKR provides a valuable insight into prevention of this challenging complication


The Bone & Joint Journal
Vol. 95-B, Issue 7 | Pages 911 - 916
1 Jul 2013
Small SR Ritter MA Merchun JG Davis KE Rogge RD

Stress shielding resulting in diminished bone density following total knee replacement (TKR) may increase the risk of migration and loosening of the prosthesis. This retrospective study was designed to quantify the effects of the method of fixation on peri-prosthetic tibial bone density beneath cemented and uncemented tibial components of similar design and with similar long-term survival rates. Standard radiographs taken between two months and 15 years post-operatively were digitised from a matched group of TKRs using cemented (n = 67) and uncemented (n = 67) AGC tibial prostheses. Digital radiograph densitometry was used to quantify changes in bone density over time. Age, length of follow-up, gender, body mass index and alignment each significantly influenced the long-term pattern of peri-prosthetic bone density. Similar long-term changes in density irrespective of the method of fixation correlated well with the high rate of survival of this TKR at 20 years, and suggest that cemented and uncemented fixation are both equally viable. Cite this article: Bone Joint J 2013;95-B:911–16


The Bone & Joint Journal
Vol. 96-B, Issue 1 | Pages 65 - 69
1 Jan 2014
Gutowski CJ Zmistowski BM Clyde CT Parvizi J

The rate of peri-prosthetic infection following total joint replacement continues to rise, and attempts to curb this trend have included the use of antibiotic-loaded bone cement at the time of primary surgery. We have investigated the clinical- and cost-effectiveness of the use of antibiotic-loaded cement for primary total knee replacement (TKR) by comparing the rate of infection in 3048 TKRs performed without loaded cement over a three-year period versus the incidence of infection after 4830 TKRs performed with tobramycin-loaded cement over a later period of time of a similar duration. In order to adjust for confounding factors, the rate of infection in 3347 and 4702 uncemented total hip replacements (THR) performed during the same time periods, respectively, was also examined. There were no significant differences in the characteristics of the patients in the different cohorts. The absolute rate of infection increased when antibiotic-loaded cement was used in TKR. However, this rate of increase was less than the rate of increase in infection following uncemented THR during the same period. If the rise in the rate of infection observed in THR were extrapolated to the TKR cohort, 18 additional cases of infection would have been expected to occur in the cohort receiving antibiotic-loaded cement, compared with the number observed. Depending on the type of antibiotic-loaded cement that is used, its cost in all primary TKRs ranges between USD $2112.72 and USD $112 606.67 per case of infection that is prevented. Cite this article: Bone Joint J 2014;96-B:65–9


The Bone & Joint Journal
Vol. 95-B, Issue 11 | Pages 1480 - 1483
1 Nov 2013
Hooper GJ Gilchrist N Maxwell R March R Heard A Frampton C

We studied the bone mineral density (BMD) and the bone mineral content (BMC) of the proximal tibia in patients with a well-functioning uncemented Oxford medial compartment arthroplasty using the Lunar iDXA bone densitometer. Our hypothesis was that there would be decreased BMD and BMC adjacent to the tibial base plate and increased BMD and BMC at the tip of the keel. There were 79 consecutive patients (33 men, 46 women) with a mean age of 65 years (44 to 84) with a minimum two-year follow-up (mean 2.6 years (2.0 to 5.0)) after unilateral arthroplasty, who were scanned using a validated standard protocol where seven regions of interest (ROI) were examined and compared with the contralateral normal knee. All had well-functioning knees with a mean Oxford knee score of 43 (14 to 48) and mean Knee Society function score of 90 (20 to 100), showing a correlation with the increasing scores and higher BMC and BMD values in ROI 2 in the non-implanted knee relative to the implanted knee (p = 0.013 and p = 0.015, respectively). The absolute and percentage changes in BMD and BMC were decreased in all ROIs in the implanted knee compared with the non-implanted knee, but this did not reach statistical significance. Bone loss was markedly less than reported losses with total knee replacement. There was no significant association with side, although there was a tendency for the BMC to decrease with age in men. The BMC was less in the implanted side relative to the non-implanted side in men compared with women in ROI 2 (p = 0.027), ROI 3 (p = 0.049) and ROI 4 (p = 0.029). The uncemented Oxford medial compartment arthroplasty appears to allow relative preservation of the BMC and BMD of the proximal tibia, suggesting that the implant acts more physiologically than total knee replacement. Peri-prosthetic bone loss is an important factor in assessing long-term implant stability and survival, and the results of this study are encouraging for the long-term outcome of this arthroplasty. Cite this article: Bone Joint J 2013;95-B:1480–3


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 4 | Pages 479 - 483
1 Apr 2011
Chang C Lai K Yang C Lan S

Between April 2004 and July 2007, we performed 241 primary total knee replacements in 204 patients using the e.motion posterior cruciate-retaining, multidirectional mobile-bearing prosthesis. Of these, 100 were carried out using an image-free navigation system, and the remaining 141 with the conventional technique. We conducted a retrospective study from the prospectively collected data of these patients to assess the early results of this new mobile-bearing design. At a mean follow-up of 49 months (32 to 71), 18 knees (7.5%) had mechanical complications of which 13 required revision. Three of these had a peri-prosthetic fracture, and were removed from the study. The indication for revision in the remaining ten was loosening of the femoral component in two, tibiofemoral dislocation in three, disassociation of the polyethylene liner in four, and a broken polyethyene liner in one. There were eight further mechanically unstable knees which presented with recurrent disassociation of the polyethylene liner. There was no significant difference in the incidence of mechanical instability between the navigation-assisted procedures (8 of 99, 8.1%) and the conventionally implanted knees (10 of 139, 7.2%). In our view, the relatively high rate of mechanical complications and revision within 30 months precludes the further use of new design of knee replacement


The Bone & Joint Journal
Vol. 96-B, Issue 6 | Pages 759 - 764
1 Jun 2014
Tibrewal S Malagelada F Jeyaseelan L Posch F Scott G

Peri-prosthetic infection is amongst the most common causes of failure following total knee replacement (TKR). In the presence of established infection, thorough joint debridement and removal of all components is necessary following which new components may be implanted. This can be performed in one or two stages; two-stage revision with placement of an interim antibiotic-loaded spacer is regarded by many to be the standard procedure for eradication of peri-prosthetic joint infection. . We present our experience of a consecutive series of 50 single-stage revision TKRs for established deep infection performed between 1979 and 2010. There were 33 women and 17 men with a mean age at revision of 66.8 years (42 to 84) and a mean follow-up of 10.5 years (2 to 24). The mean time between the primary TKR and the revision procedure was 2.05 years (1 to 8). Only one patient required a further revision for recurrent infection, representing a success rate of 98%. Nine patients required further revision for aseptic loosening, according to microbiological testing of biopsies taken at the subsequent surgery. Three other patients developed a further septic episode but none required another revision. These results suggest that a single-stage revision can produce comparable results to a two-stage revision. Single-stage revision offers a reduction in costs as well as less morbidity and inconvenience for patients. Cite this article: Bone Joint J 2014;96-B:759–64


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 3 | Pages 339 - 343
1 Mar 2012
Sewell MD Hanna SA Al-Khateeb H Miles J Pollock RC Carrington RWJ Skinner JA Cannon SR Briggs TWR

Patients with skeletal dysplasia are prone to developing advanced osteoarthritis of the knee requiring total knee replacement (TKR) at a younger age than the general population. TKR in this unique group of patients is a technically demanding procedure owing to the deformity, flexion contracture, generalised hypotonia and ligamentous laxity. We retrospectively reviewed the outcome of 11 TKRs performed in eight patients with skeletal dysplasia at our institution using the Stanmore Modular Individualised Lower Extremity System (SMILES) custom-made rotating-hinge TKR. There were three men and five women with mean age of 57 years (41 to 79). Patients were followed clinically and radiologically for a mean of seven years (3 to 11.5). The mean Knee Society clinical and function scores improved from 24 (14 to 36) and 20 (5 to 40) pre-operatively, respectively, to 68 (28 to 80) and 50 (22 to 74), respectively, at final follow-up. Four complications were recorded, including a patellar fracture following a fall, a tibial peri-prosthetic fracture, persistent anterior knee pain, and aseptic loosening of a femoral component requiring revision. Our results demonstrate that custom primary rotating-hinge TKR in patients with skeletal dysplasia is effective at relieving pain, with a satisfactory range of movement and improved function. It compensates for bony deformity and ligament deficiency and reduces the likelihood of corrective osteotomy. Patellofemoral joint complications are frequent and functional outcome is worse than with primary TKR in the general population


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 7 | Pages 879 - 884
1 Jul 2008
Porteous AJ Hassaballa MA Newman JH

We identified 148 patients who had undergone a revision total knee replacement using a single implant system between 1990 and 2000. Of these 18 patients had died, six had developed a peri-prosthetic fracture and ten had incomplete records or radiographs. This left 114 with prospectively-collected radiographs and Bristol knee scores available for study. The height of the joint line before and after revision total knee replacement was measured and classified as either restored to within 5 mm of the pre-operative height or elevated if it was positioned more than 5 mm above the pre-operative height. The joint line was elevated in 41 knees (36%) and restored in 73 (64%). Revision surgery significantly improved the mean Bristol knee score from 41.1 (. sd. 15.9) pre-operatively to 80.5 (. sd. 15) post-operatively (p < 0.001). At one year post-operatively both the total Bristol knee score and its functional component were significantly better in the restored group than in the elevated group (p < 0.01). Overall, revision from a unicondylar knee replacement required less use of bone graft, fewer component augments, restored the joint line more often and gave a significantly better total Bristol knee score (p < 0.02) and functional score (p < 0.01) than revision from total knee replacement. Our findings show that restoration of the joint line at revision total knee replacement gives a significantly better result than leaving it unrestored by more than 5 mm. We recommend the greater use of distal femoral augments to help to achieve this goal


The Bone & Joint Journal
Vol. 96-B, Issue 12 | Pages 1649 - 1656
1 Dec 2014
Lindberg-Larsen M Jørgensen CC Bæk Hansen T Solgaard S Odgaard A Kehlet H

We present detailed information about early morbidity after aseptic revision knee replacement from a nationwide study. All aseptic revision knee replacements undertaken between 1st October 2009 and 30th September 2011 were analysed using the Danish National Patient Registry with additional information from the Danish Knee Arthroplasty Registry. The 1218 revisions involving 1165 patients were subdivided into total revisions, large partial revisions, partial revisions and revisions of unicondylar replacements (UKR revisions). The mean age was 65.0 years (27 to 94) and the median length of hospital stay was four days (interquartile range: 3 to 5), with a 90 days re-admission rate of 9.9%, re-operation rate of 3.5% and mortality rate of 0.2%. The age ranges of 51 to 55 years (p = 0.018), 76 to 80 years (p < 0.001) and ≥ 81 years (p < 0.001) were related to an increased risk of re-admission. The age ranges of 76 to 80 years (p = 0.018) and the large partial revision subgroup (p = 0.073) were related to an increased risk of re-operation. The ages from 76 to 80 years (p < 0.001), age ≥ 81 years (p < 0.001) and surgical time > 120 min (p <  0.001) were related to increased length of hospital stay, whereas the use of a tourniquet (p = 0.008) and surgery in a low volume centre (p = 0.013) were related to shorter length of stay.

In conclusion, we found a similar incidence of early post-operative morbidity after aseptic knee revisions as has been reported after primary procedures. This suggests that a length of hospital stay ≤ four days and discharge home at that time is safe following aseptic knee revision surgery in Denmark.

Cite this article: Bone Joint J 2014;96-B:1649–56.


The Bone & Joint Journal
Vol. 95-B, Issue 2 | Pages 181 - 187
1 Feb 2013
Liddle AD Pandit H O’Brien S Doran E Penny ID Hooper GJ Burn PJ Dodd CAF Beverland DE Maxwell AR Murray DW

The Cementless Oxford Unicompartmental Knee Replacement (OUKR) was developed to address problems related to cementation, and has been demonstrated in a randomised study to have similar clinical outcomes with fewer radiolucencies than observed with the cemented device. However, before its widespread use it is necessary to clarify contraindications and assess the complications. This requires a larger study than any previously published.

We present a prospective multicentre series of 1000 cementless OUKRs in 881 patients at a minimum follow-up of one year. All patients had radiological assessment aligned to the bone–implant interfaces and clinical scores. Analysis was performed at a mean of 38.2 months (19 to 88) following surgery. A total of 17 patients died (comprising 19 knees (1.9%)), none as a result of surgery; there were no tibial or femoral loosenings. A total of 19 knees (1.9%) had significant implant-related complications or required revision. Implant survival at six years was 97.2%, and there was a partial radiolucency at the bone–implant interface in 72 knees (8.9%), with no complete radiolucencies. There was no significant increase in complication rate compared with cemented fixation (p = 0.87), and no specific contraindications to cementless fixation were identified.

Cementless OUKR appears to be safe and reproducible in patients with end-stage anteromedial osteoarthritis of the knee, with radiological evidence of improved fixation compared with previous reports using cemented fixation.

Cite this article: Bone Joint J 2013;95-B:181–7.


The Bone & Joint Journal
Vol. 95-B, Issue 8 | Pages 1069 - 1074
1 Aug 2013
Rao BM Kamal TT Vafaye J Moss M

We report the results of revision total knee replacement (TKR) in 26 patients with major metaphyseal osteolytic defects using 29 trabecular metal cones in conjunction with a rotating hinged total knee prosthesis. The osteolytic defects were types II and III (A or B) according to the Anderson Orthopaedic Research Institute (AORI) classification. The mean age of the patients was 72 years (62 to 84) and there were 15 men and 11 women. In this series patients had undergone a mean of 2.34 previous total knee arthroplasties. The main objective was to restore anatomy along with stability and function of the knee joint to allow immediate full weight-bearing and active knee movement. Outcomes were measured using Knee Society scores, Oxford knee scores, range of movement of the knee and serial radiographs. Patients were followed for a mean of 36 months (24 to 49). The mean Oxford knee clinical scores improved from 12.83 (10 to 15) to 35.20 (32 to 38) (p < 0.001) and mean American Knee Society scores improved from 33.24 (13 to 36) to 81.12 (78 to 86) (p < 0.001). No radiolucent lines suggestive of loosening were seen around the trabecular metal cones, and by one year all the radiographs showed good osteo-integration. There was no evidence of any collapse or implant migration. Our early results confirm the findings of others that trabecular metal cones offer a useful way of managing severe bone loss in revision TKR.

Cite this article: Bone Joint J 2013;95-B:1069–74.


The Bone & Joint Journal
Vol. 98-B, Issue 3 | Pages 334 - 340
1 Mar 2016
Tayton ER Frampton C Hooper GJ Young SW

Aims

The aim of this study was to identify risk factors for prosthetic joint infection (PJI) following total knee arthroplasty (TKA).

Patients and Methods

The New Zealand Joint Registry database was analysed, using revision surgery for PJI at six and 12 months after surgery as primary outcome measures. Statistical associations between revision for infection, with common and definable surgical and patient factors were tested.


The Bone & Joint Journal
Vol. 97-B, Issue 11 | Pages 1493 - 1499
1 Nov 2015
Pandit H Hamilton TW Jenkins C Mellon SJ Dodd CAF Murray DW

This prospective study reports the 15-year survival and ten-year functional outcome of a consecutive series of 1000 minimally invasive Phase 3 Oxford medial UKAs (818 patients, 393 men, 48%, 425 women, 52%, mean age 66 years; 32 to 88). These were implanted by two surgeons involved with the design of the prosthesis to treat anteromedial osteoarthritis and spontaneous osteonecrosis of the knee, which are recommended indications. Patients were prospectively identified and followed up independently for a mean of 10.3 years (5.3 to 16.6).

At ten years, the mean Oxford Knee Score was 40 (standard deviation (sd) 9; 2 to 48): 79% of knees (349) had an excellent or good outcome. There were 52 implant-related re-operations at a mean of 5.5 years (0.2 to 14.7). The most common reasons for re-operation were arthritis in the lateral compartment (2.5%, 25 knees), bearing dislocation (0.7%, seven knees) and unexplained pain (0.7%, seven knees). When all implant-related re-operations were considered as failures, the ten-year rate of survival was 94% (95% confidence interval (CI) 92 to 96) and the 15-year survival rate 91% (CI 83 to 98). When failure of the implant was the endpoint the 15-year survival was 99% (CI 96 to 100).

This is the only large series of minimally invasive UKAs with 15-year survival data. The results support the continued use of minimally invasive UKA for the recommended indications.

Cite this article: Bone Joint J 2015;97-B:1493–99.


The Bone & Joint Journal
Vol. 95-B, Issue 10 | Pages 1348 - 1353
1 Oct 2013
Valenzuela GA Jacobson NA Buzas D Korecki TD Valenzuela RG Teitge RA

The outcome of high tibial osteotomy (HTO) deteriorates with time, and additional procedures may be required. The aim of this study was to compare the clinical and radiological outcomes between unicompartmental knee replacement (UKR) and total knee replacement (TKR) after HTO as well as after primary UKR. A total of 63 patients (63 knees) were studied retrospectively and divided into three groups: UKR after HTO (group A; n = 22), TKR after HTO (group B; n = 18) and primary UKR (group C; n = 22). The Oxford knee score (OKS), Knee Society score (KSS), hip–knee–ankle angles, mechanical axis and patellar height were evaluated pre- and post-operatively. At a mean of 64 months (19 to 180) post-operatively the mean OKS was 43.8 (33 to 49), 43.3 (30 to 48) and 42.5 (29 to 48) for groups A, B and C, respectively (p = 0.73). The mean KSS knee score was 88.8 (54 to 100), 88.11 (51 to 100) and 85.3 (45 to 100) for groups A, B and C, respectively (p = 0.65), and the mean KSS function score was 85.0 (50 to 100) in group A, 85.8 (20 to 100) in group B and 79.3 (50 to 100) in group C (p = 0.48). Radiologically the results were comparable for all groups except for patellar height, with a higher incidence of patella infra following a previous HTO (p = 0.02).

Cite this article: Bone Joint J 2013;95-B:1348–53.


The Bone & Joint Journal
Vol. 96-B, Issue 11_Supple_A | Pages 93 - 95
1 Nov 2014
Gehrke T Kendoff D Haasper C

The use of hinged implants in primary total knee replacement (TKR) should be restricted to selected indications and mainly for elderly patients. Potential indications for a rotating hinge or pure hinge implant in primary TKR include: collateral ligament insufficiency, severe varus or valgus deformity (> 20°) with necessary relevant soft-tissue release, relevant bone loss including insertions of collateral ligaments, gross flexion-extension gap imbalance, ankylosis, or hyperlaxity. Although data reported in the literature are inconsistent, clinical results depend on implant design, proper technical use, and adequate indications. We present our experience with a specific implant type that we have used for over 30 years and which has given our elderly patients good mid-term results. Because revision of implants with long cemented stems can be very challenging, an effort should be made in the future to use shorter stems in modular versions of hinged implants.

Cite this article: Bone Joint J 2014;96-B(11 Suppl A):93–5.


The Bone & Joint Journal
Vol. 96-B, Issue 11_Supple_A | Pages 101 - 104
1 Nov 2014
Lombardi Jr AV Berend KR Adams JB

Previous studies of failure mechanisms leading to revision total knee replacement (TKR) performed between 1986 and 2000 determined that many failed early, with a disproportionate amount accounted for by infection and implant-associated factors including wear, loosening and instability. Since then, efforts have been made to improve implant performance and instruct surgeons in best practice. Recently our centre participated in a multi-centre evaluation of 844 revision TKRs from 2010 to 2011. The purpose was to report a detailed analysis of failure mechanisms over time and to see if failure modes have changed over the past 10 to 15 years. Aseptic loosening was the predominant mechanism of failure (31.2%), followed by instability (18.7%), infection (16.2%), polyethylene wear (10.0%), arthrofibrosis (6.9%) and malalignment (6.6%). The mean time to failure was 5.9 years (ten days to 31 years), 35.3% of all revisions occurred at less than two years, and 60.2% in the first five years. With improvements in implant and polyethylene manufacture, polyethylene wear is no longer a leading cause of failure. Early mechanisms of failure are primarily technical errors. In addition to improving implant longevity, industry and surgeons must work together to decrease these technical errors. All reports on failure of TKR contain patients with unexplained pain who not infrequently have unmet expectations. Surgeons must work to achieve realistic patient expectations pre-operatively, and therefore, improve patient satisfaction post-operatively.

Cite this article: Bone Joint J 2014;96-B(11 Suppl A):101–4.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 7 | Pages 875 - 881
1 Jul 2012
Vanhegan IS Morgan-Jones R Barrett DS Haddad FS

This review summarises the opinions and conclusions reached from a symposium on infected total knee replacement (TKR) held at the British Association of Surgery of the Knee (BASK) annual meeting in 2011. The National Joint Registry for England and Wales reported 5082 revision TKRs in 2010, of which 1157 (23%) were caused by infection. The diagnosis of infection beyond the acute post-operative stage relies on the identification of the causative organism by aspiration and analysis of material obtained at arthroscopy. Ideal treatment then involves a two-stage surgical procedure with extensive debridement and washout, followed by antibiotics. An articulating or non-articulating drug-eluting cement spacer is used prior to implantation of the revision prosthesis, guided by the serum level of inflammatory markers. The use of a single-stage revision is gaining popularity and we would advocate its use in certain patients where the causative organism is known, no sinuses are present, the patient is not immunocompromised, and there is no radiological evidence of component loosening or osteitis.

It is our opinion that single-stage revision produces high-quality reproducible results and will soon achieve the same widespread acceptance as it does in infected hip arthroplasty.