Ahmad A Subramanian T Panteliadis P Wilson-Macdonald J Rothenfluh DA Nnadi C

Aims

Magnetically controlled growing rods (MCGRs) allow non-invasive correction of the spinal deformity in the treatment of early-onset scoliosis. Conventional growing rod systems (CGRS) need repeated surgical distractions: these are associated with the effect of the ‘law of diminishing returns’.

The primary aim of this study was to quantify this effect in MCGRs over sequential distractions.

Patients and Methods

A total of 35 patients with a maximum follow-up of 57 months were included in the study. There were 17 boys and 18 girls with a mean age of 7.4 years (2 to 14). True Distraction (TD) was determined by measuring the expansion gap on fluoroscopy. This was compared with Intended Distraction (ID) and expressed as the ‘T/I’ ratio. The T/I ratio and the Cobb angle were calculated at several time points during follow-up.


The Bone & Joint Journal
Vol. 102-B, Issue 10 | Pages 1375 - 1383
3 Oct 2020
Zhang T Sze KY Peng ZW Cheung KMC Lui YF Wong YW Kwan KYH Cheung JPY

Aims. To investigate metallosis in patients with magnetically controlled growing rods (MCGRs) and characterize the metal particle profile of the tissues surrounding the rod. Methods. This was a prospective observational study of patients with early onset scoliosis (EOS) treated with MCGRs and undergoing rod exchange who were consecutively recruited between February 2019 and January 2020. Ten patients were recruited (mean age 12 years (SD 1.3); 2 M:8 F). The configurations of the MCGR were studied to reveal the distraction mechanisms, with crucial rod parts being the distractable piston rod and the magnetically driven rotor inside the barrel of the MCGR. Metal-on-metal contact in the form of ring-like wear marks on the piston was found on the distracted portion of the piston immediately outside the barrel opening (BO) through which the piston rod distracts. Biopsies of paraspinal muscles and control tissue samples were taken over and away from the wear marks, respectively. Spectral analyses of the rod alloy and biopsies were performed to reveal the metal constituents and concentrations. Histological analyses of the biopsies were performed with haematoxylin and eosin staining. Results. Titanium (Ti), vanadium (V), and neodymium (Nd) concentrations in the biopsies taken near the wear marks were found to be significantly higher than those in the control tissue samples. Significantly increased Nd concentrations were also found in the tissues near the barrel of the MCGR. Chronic inflammation was revealed by the histological studies with fibrosis and macrophage infiltration. Black particles were present within the macrophages in the fibrotic tissues. Conclusion. Ti and V were generated mainly at the BO due to metal-on-metal contact, whereas the Nd from the rotor of the MCGR is likely released from the BO during distraction sessions. Phagocytotic immune cells with black particles inside raise concern regarding the long-term implications of metallosis. Cite this article: Bone Joint J 2020;102-B(10):1375–1383


The Bone & Joint Journal
Vol. 104-B, Issue 2 | Pages 257 - 264
1 Feb 2022
Tahir M Mehta D Sandhu C Jones M Gardner A Mehta JS

Aims. The aim of this study was to compare the clinical and radiological outcomes of patients with early-onset scoliosis (EOS), who had undergone spinal fusion after distraction-based spinal growth modulation using either traditional growing rods (TGRs) or magnetically controlled growing rods (MCGRs). Methods. We undertook a retrospective review of skeletally mature patients who had undergone fusion for an EOS, which had been previously treated using either TGRs or MCGRs. Measured outcomes included sequential coronal T1 to S1 height and major curve (Cobb) angle on plain radiographs and any complications requiring unplanned surgery before final fusion. Results. We reviewed 43 patients (63% female) with a mean age of 6.4 years (SD 2.6) at the index procedure, and 12.2 years (SD 2.2) at final fusion. Their mean follow-up was 8.1 years (SD 3.4). A total of 16 patients were treated with MCGRs and 27 with TGRs. The mean number of distractions was 7.5 in the MCGR group and ten in the TGR group (p = 0.471). The mean interval between distractions was 3.4 months in the MCGR group and 8.6 months in the TGR group (p < 0.001). The mean Cobb angle had improved by 25.1° in the MCGR group and 23.2° in TGR group (p = 0.664) at final follow-up. The mean coronal T1 to S1 height had increased by 16% in the MCGR group and 32.9% in TGR group (p = 0.001), although the mean T1 to S1 height achieved at final follow-up was similar in both. Unplanned operations were needed in 43.8% of the MCGR group and 51.2% of TGR group (p = 0.422). Conclusion. In this retrospective, single-centre review, there were no significant differences in major curve correction or gain in spinal height at fusion. Although the number of planned procedures were fewer in patients with MCGRs, the rates of implant-related complications needing unplanned revision surgery were similar in the two groups. Cite this article: Bone Joint J 2022;104-B(2):257–264


The Bone & Joint Journal
Vol. 98-B, Issue 12 | Pages 1662 - 1667
1 Dec 2016
Teoh KH von Ruhland C Evans SL James SH Jones A Howes J Davies PR Ahuja S

Aims. We present a case series of five patients who had revision surgery following magnetic controlled growing rods (MGCR) for early onset scoliosis. Metallosis was found during revision in four out of five patients and we postulated a mechanism for rod failure based on retrieval analysis. Patients and Methods. Retrieval analysis was performed on the seven explanted rods. The mean duration of MCGR from implantation to revision was 35 months (17 to 46). The mean age at revision was 12 years (7 to 15; four boys, one girl). Results. A total of six out of seven rods had tissue metallosis and pseudo-capsule surrounding the actuator. A total of four out of seven rods were pistoning. There were two rods which were broken. All rods had abrasive circumferential markings. A significant amount of metal debris was found when the actuators were carefully cut open. Analytical electron microscopy demonstrated metal fragments of predominantly titanium with a mean particle size of 3.36 microns (1.31 to 6.61). Conclusion. This study highlights concerns with tissue metallosis in MCGR. We recommend careful follow-up of patients who have received this implant. Cite this article: Bone Joint J 2016;98-B:1662–7


The Bone & Joint Journal
Vol. 100-B, Issue 4 | Pages 507 - 515
1 Apr 2018
Nnadi C Thakar C Wilson-MacDonald J Milner P Rao A Mayers D Fairbank J Subramanian T

Aims. The primary aim of this study was to evaluate the performance and safety of magnetically controlled growth rods in the treatment of early onset scoliosis. Secondary aims were to evaluate the clinical outcome, the rate of further surgery, the rate of complications, and the durability of correction. Patients and Methods. We undertook an observational prospective cohort study of children with early onset scoliosis, who were recruited over a one-year period and followed up for a minimum of two years. Magnetically controlled rods were introduced in a standardized manner with distractions performed three-monthly thereafter. Adverse events which were both related and unrelated to the device were recorded. Ten children, for whom relevant key data points (such as demographic information, growth parameters, Cobb angles, and functional outcomes) were available, were recruited and followed up over the period of the study. There were five boys and five girls. Their mean age was 6.2 years (2.5 to 10). Results. The mean coronal Cobb angle improved from 57.6° (40° to 81°) preoperatively, 32.8° (28° to 46°) postoperatively, and 41° (19° to 57°) at two years. Five children had an adverse event, with four requiring return to theatre, but none were related to the device. There were no neurological complications or infections. No devices failed. One child developed a proximal junctional kyphosis. The mean gain in spinal column height from T1 to S1 was 45.4 mm (24 to 81) over the period of the study. Conclusion. Magnetically controlled growth rods provide an alternative solution to traditional growing rods in the surgical management of children with early onset scoliosis, supporting growth of the spine while controlling curve progression. Their use has clear psychosocial and economic benefits, with the reduction of the need for repeat surgery as required with traditional growing rods. Cite this article: Bone Joint J 2018;100-B:507–15


The Bone & Joint Journal
Vol. 95-B, Issue 1 | Pages 75 - 80
1 Jan 2013
Dannawi Z Altaf F Harshavardhana NS El Sebaie H Noordeen H

Conventional growing rods are the most commonly used distraction-based devices in the treatment of progressive early-onset scoliosis. This technique requires repeated lengthenings with the patient anaesthetised in the operating theatre. We describe the outcomes and complications of using a non-invasive magnetically controlled growing rod (MCGR) in children with early-onset scoliosis. Lengthening is performed on an outpatient basis using an external remote control with the patient awake.

Between November 2009 and March 2011, 34 children with a mean age of eight years (5 to 12) underwent treatment. The mean length of follow-up was 15 months (12 to 18). In total, 22 children were treated with dual rod constructs and 12 with a single rod. The mean number of distractions per patient was 4.8 (3 to 6). The mean pre-operative Cobb angle was 69° (46° to 108°); this was corrected to a mean 47° (28° to 91°) post-operatively. The mean Cobb angle at final review was 41° (27° to 86°). The mean pre-operative distance from T1 to S1 was 304 mm (243 to 380) and increased to 335 mm (253 to 400) in the immediate post-operative period. At final review the mean distance from T1 to S1 had increased to 348 mm (260 to 420).

Two patients developed a superficial wound infection and a further two patients in the single rod group developed a loss of distraction. In the dual rod group, one patient had pull-out of a hook and one developed prominent metalwork. Two patients had a rod breakage; one patient in the single rod group and one patient in the dual rod group. Our early results show that the MCGR is safe and effective in the treatment of progressive early-onset scoliosis with the avoidance of repeated surgical lengthenings.

Cite this article: Bone Joint J 2013;95-B:75–80.


The Bone & Joint Journal
Vol. 100-B, Issue 9 | Pages 1187 - 1200
1 Sep 2018
Subramanian T Ahmad A Mardare DM Kieser DC Mayers D Nnadi C

Aims

Magnetically controlled growing rod (MCGR) systems use non-invasive spinal lengthening for the surgical treatment of early-onset scoliosis (EOS). The primary aim of this study was to evaluate the performance of these devices in the prevention of progression of the deformity. A secondary aim was to record the rate of complications.

Patients and Methods

An observational study of 31 consecutive children with EOS, of whom 15 were male, who were treated between December 2011 and October 2017 was undertaken. Their mean age was 7.7 years (2 to 14). The mean follow-up was 47 months (24 to 69). Distractions were completed using the tailgating technique. The primary outcome measure was correction of the radiographic deformity. Secondary outcomes were growth, functional outcomes and complication rates.


The Journal of Bone & Joint Surgery British Volume
Vol. 82-B, Issue 5 | Pages 695 - 701
1 Jul 2000
Kawaguchi Y Kitagawa H Nakamura H Gejo R Kimura T

We recorded compound muscle action potentials (CMAPs) from the diaphragm in 15 normal volunteers, nine patients with lesions of the lower cervical cord (C5 to C8), one completely quadriplegic patient (C6) and seven patients with lesions at a higher cervical level (C1 to C4). Transcranial magnetic stimulation and electrical stimulation of the phrenic nerve were carried out. When the centre of the coil was placed on the interauricular line at a point 3 cm lateral to the vertex on the scalp, the CMAPs from the diaphragm had the largest amplitude and the shortest latency. There was no difference in the mean latency of the CMAPs recorded by transcranial magnetic stimulation in the normal volunteers and in the patients with lesions of the lower cervical cord. In the quadriplegic patient, the latency of the CMAPs was not delayed, but was prolonged in the patients with lesions at a higher level. Those evoked by electrical stimulation of the phrenic nerve were not prolonged in the patients with higher lesions. Our findings suggest that the prolongation of the latency by transcranial magnetic stimulation reflects dysfunction of the higher cervical cord. The combination of transcranial magnetic stimulation and electrical stimulation of the phrenic nerve can detect the precise level of the lesion in the motor tract to the diaphragm


The Bone & Joint Journal
Vol. 106-B, Issue 6 | Pages 596 - 602
1 Jun 2024
Saarinen AJ Sponseller P Thompson GH White KK Emans J Cahill PJ Hwang S Helenius I

Aims

The aim of this study was to compare outcomes after growth-friendly treatment for early-onset scoliosis (EOS) between patients with skeletal dysplasias versus those with other syndromes.

Methods

We retrospectively identified 20 patients with skeletal dysplasias and 292 with other syndromes (control group) who had completed surgical growth-friendly EOS treatment between 1 January 2000 and 31 December 2018. We compared radiological parameters, complications, and health-related quality of life (HRQoL) at mean follow-up of 8.6 years (SD 3.3) in the dysplasia group and 6.6 years (SD 2.6) in the control group.


The Bone & Joint Journal
Vol. 103-B, Issue 3 | Pages 547 - 552
1 Mar 2021
Magampa RS Dunn R

Aims

Spinal deformity surgery carries the risk of neurological injury. Neurophysiological monitoring allows early identification of intraoperative cord injury which enables early intervention resulting in a better prognosis. Although multimodal monitoring is the ideal, resource constraints make surgeon-directed intraoperative transcranial motor evoked potential (TcMEP) monitoring a useful compromise. Our experience using surgeon-directed TcMEP is presented in terms of viability, safety, and efficacy.

Methods

We carried out a retrospective review of a single surgeon’s prospectively maintained database of cases in which TcMEP monitoring had been used between 2010 and 2017. The upper limbs were used as the control. A true alert was recorded when there was a 50% or more loss of amplitude from the lower limbs with maintained upper limb signals. Patients with true alerts were identified and their case history analyzed.


The Bone & Joint Journal
Vol. 102-B, Issue 6 | Pages 671 - 676
1 Jun 2020
Giorgi PD Villa F Gallazzi E Debernardi A Schirò GR Crisà FM Talamonti G D’Aliberti G

Aims

The current pandemic caused by COVID-19 is the biggest challenge for national health systems for a century. While most medical resources are allocated to treat COVID-19 patients, several non-COVID-19 medical emergencies still need to be treated, including vertebral fractures and spinal cord compression. The aim of this paper is to report the early experience and an organizational protocol for emergency spinal surgery currently being used in a large metropolitan area by an integrated team of orthopaedic surgeons and neurosurgeons.

Methods

An organizational model is presented based on case centralization in hub hospitals and early management of surgical cases to reduce hospital stay. Data from all the patients admitted for emergency spinal surgery from the beginning of the outbreak were prospectively collected and compared to data from patients admitted for the same reason in the same time span in the previous year, and treated by the same integrated team.


The Bone & Joint Journal
Vol. 98-B, Issue 9 | Pages 1240 - 1247
1 Sep 2016
Thompson W Thakar C Rolton DJ Wilson-MacDonald J Nnadi C

Aims

We undertook a prospective non-randomised radiological study to evaluate the preliminary results of using magnetically-controlled growing rods (MAGEC System, Ellipse technology) to treat children with early-onset scoliosis.

Patients and Methods

Between January 2011 and January 2015, 19 children were treated with magnetically-controlled growing rods (MCGRs) and underwent distraction at three-monthly intervals. The mean age of our cohort was 9.1 years (4 to 14) and the mean follow-up 22.4 months (5.1 to 35.2). Of the 19 children, eight underwent conversion from traditional growing rods. Whole spine radiographs were carried out pre- and post-operatively: image intensification was used during each lengthening in the outpatient department. The measurements evaluated were Cobb angle, thoracic kyphosis, proximal junctional kyphosis and spinal growth from T1 to S1.


The Bone & Joint Journal
Vol. 97-B, Issue 12 | Pages 1683 - 1692
1 Dec 2015
Patel A James SL Davies AM Botchu R

The widespread use of MRI has revolutionised the diagnostic process for spinal disorders. A typical protocol for spinal MRI includes T1 and T2 weighted sequences in both axial and sagittal planes. While such an imaging protocol is appropriate to detect pathological processes in the vast majority of patients, a number of additional sequences and advanced techniques are emerging. The purpose of the article is to discuss both established techniques that are gaining popularity in the field of spinal imaging and to introduce some of the more novel ‘advanced’ MRI sequences with examples to highlight their potential uses.

Cite this article: Bone Joint J 2015;97-B:1683–92.


The Bone & Joint Journal
Vol. 100-B, Issue 6 | Pages 772 - 779
1 Jun 2018
Helenius IJ Oksanen HM McClung A Pawelek JB Yazici M Sponseller PD Emans JB Sánchez Pérez-Grueso FJ Thompson GH Johnston C Shah SA Akbarnia BA

Aims

The aim of this study was to compare the outcomes of surgery using growing rods in patients with severe versus moderate early-onset scoliosis (EOS).

Patients and Methods

A review of a multicentre EOS database identified 107 children with severe EOS (major curve ≥ 90°) treated with growing rods before the age of ten years with a minimum follow-up of two years and three or more lengthening procedures. From the same database, 107 matched controls with moderate EOS were identified.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 8 | Pages 1024 - 1031
1 Aug 2012
Rajasekaran S Kanna RM Shetty AP

The identification of the extent of neural damage in patients with acute or chronic spinal cord injury is imperative for the accurate prediction of neurological recovery. The changes in signal intensity shown on routine MRI sequences are of limited value for predicting functional outcome. Diffusion tensor imaging (DTI) is a novel radiological imaging technique which has the potential to identify intact nerve fibre tracts, and has been used to image the brain for a variety of conditions. DTI imaging of the spinal cord is currently only a research tool, but preliminary studies have shown that it holds considerable promise in predicting the severity of spinal cord injury.

This paper briefly reviews our current knowledge of this technique.


The Bone & Joint Journal
Vol. 95-B, Issue 9 | Pages 1244 - 1249
1 Sep 2013
Jeon C Park J Chung N Son K Lee Y Kim J

We investigated the spinopelvic morphology and global sagittal balance of patients with a degenerative retrolisthesis or anterolisthesis. A total of 269 consecutive patients with a degenerative spondylolisthesis were included in this study. There were 95 men and 174 women with a mean age of 64.3 years (sd 10.5; 40 to 88). A total of 106 patients had a pure retrolisthesis (R group), 130 had a pure anterolisthesis (A group), and 33 had both (R+A group).

A backward slip was found in the upper lumbar levels (mostly L2 or L3) with an almost equal gender distribution in both the R and R+A groups. The pelvic incidence and sacral slope of the R group were significantly lower than those of the A (both p < 0.001) and R+A groups (both p < 0.001). The lumbar lordosis of the R+A group was significantly greater than that of the R (p = 0.025) and A groups (p = 0.014). The C7 plumb line of the R group was located more posteriorly than that of the A group (p = 0.023), but was no different from than that of the R+A group (p = 0.422). The location of C7 plumb line did not differ between the three groups (p = 0.068). The spinosacral angle of the R group was significantly smaller than that of the A group (p < 0.001) and R+A group (p < 0.001).

Our findings imply that there are two types of degenerative retrolisthesis: one occurs primarily as a result of degeneration in patients with low pelvic incidence, and the other occurs secondarily as a compensatory mechanism in patients with an anterolisthesis and high pelvic incidence.

Cite this article: Bone Joint J 2013;95-B:1244–9.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 5 | Pages 678 - 683
1 May 2012
Matsumoto M Okada E Ichihara D Chiba K Toyama Y Fujiwara H Momoshima S Nishiwaki Y Takahata T

We conducted a prospective follow-up MRI study of originally asymptomatic healthy subjects to clarify the development of Modic changes in the cervical spine over a ten-year period and to identify related factors. Previously, 497 asymptomatic healthy volunteers with no history of cervical trauma or surgery underwent MRI. Of these, 223 underwent a second MRI at a mean follow-up of 11.6 years (10 to 12.7). These 223 subjects comprised 133 men and 100 women with a mean age at second MRI of 50.5 years (23 to 83). Modic changes were classified as not present and types 1 to 3. Changes in Modic types over time and relationships between Modic changes and progression of degeneration of the disc or clinical symptoms were evaluated. A total of 31 subjects (13.9%) showed Modic changes at follow-up: type 1 in nine, type 2 in 18, type 3 in two, and types 1 and 2 in two. Modic changes at follow-up were significantly associated with numbness or pain in the arm, but not with neck pain or shoulder stiffness. Age (≥ 40 years), gender (male), and pre-existing disc degeneration were significantly associated with newly developed Modic changes.

In the cervical spine over a ten-year period, type 2 Modic changes developed most frequently. Newly developed Modic changes were significantly associated with age, gender, and pre-existing disc degeneration.


The Journal of Bone & Joint Surgery British Volume
Vol. 87-B, Issue 8 | Pages 1111 - 1116
1 Aug 2005
Ranson CA Kerslake RW Burnett AF Batt ME Abdi S

Low back injuries account for the greatest loss of playing time for professional fast bowlers in cricket. Previous radiological studies have shown a high prevalence of degeneration of the lumbar discs and stress injuries of the pars interarticularis in elite junior fast bowlers. We have examined MRI appearance of the lumbar spines of 36 asymptomatic professional fast bowlers and 17 active control subjects. The fast bowlers had a relatively high prevalence of multi-level degeneration of the lumbar discs and a unique pattern of stress lesions of the pars interarticularis on the non-dominant side. The systems which have been used to classify the MR appearance of the lumbar discs and pars were found to be reliable. However, the relationship between the radiological findings, pain and dysfunction remains unclear.