Advertisement for orthosearch.org.uk
Results 1 - 4 of 4
Results per page:
The Bone & Joint Journal
Vol. 103-B, Issue 6 | Pages 1168 - 1172
1 Jun 2021
Iliadis AD Wright J Stoddart MT Goodier WD Calder P

Aims

The STRYDE nail is an evolution of the PRECICE Intramedullary Limb Lengthening System, with unique features regarding its composition. It is designed for load bearing throughout treatment in order to improve patient experience and outcomes and allow for simultaneous bilateral lower limb lengthening. The literature published to date is limited regarding outcomes and potential problems. We report on our early experience and raise awareness for the potential of adverse effects from this device.

Methods

This is a retrospective review of prospective data collected on all patients treated in our institution using this implant. We report the demographics, nail accuracy, reliability, consolidation index, and cases where concerning clinical and radiological findings were encountered. There were 14 STRYDE nails implanted in nine patients (three male and six female) between June 2019 and September 2020. Mean age at surgery was 33 years (14 to 65). Five patients underwent bilateral lengthening (two femoral and three tibial) and four patients unilateral femoral lengthening for multiple aetiologies.


The Bone & Joint Journal
Vol. 99-B, Issue 11 | Pages 1537 - 1544
1 Nov 2017
Wahl P Guidi M Benninger E Rönn K Gautier E Buclin T Magnin J Livio F

Aims

Calcium sulphate (CaSO4) is a resorbable material that can be used simultaneously as filler of a dead space and as a carrier for the local application of antibiotics. Our aim was to describe the systemic exposure and the wound fluid concentrations of vancomycin in patients treated with vancomycin-loaded CaSO4 as an adjunct to the routine therapy of bone and joint infections.

Patients and Methods

A total of 680 post-operative blood and 233 wound fluid samples were available for analysis from 94 implantations performed in 87 patients for various infective indications. Up to 6 g of vancomycin were used. Non-compartmental pharmacokinetic analysis was performed on the data from 37 patients treated for an infection of the hip.


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 5 | Pages 662 - 667
1 May 2008
Strauss EJ Egol KA Alaia M Hansen D Bashar M Steiger D

This study was undertaken to evaluate the safety and efficacy of retrievable inferior vena cava filters in high-risk orthopaedic patients. A total of 58 patients had a retrievable inferior vena cava filter placed as an adjunct to chemical and mechanical prophylaxis, most commonly for a history of previous deep-vein thrombosis or pulmonary embolism, polytrauma, or expected prolonged immobilisation. In total 56 patients (96.6%) had an uncomplicated post-operative course. Two patients (3.4%) died in the peri-operative period for unrelated reasons.

Of the 56 surviving patients, 50 (89%) were available for follow-up. A total of 32 filters (64%) were removed without complication at a mean of 37.8 days (4 to 238) after placement. There were four filters (8%) which were retained because of thrombosis at the filter site, and four (8%) were retained because of incorporation of the filter into the wall of the inferior vena cava. In ten cases (20%) the retrievable filter was left in place to continue as primary prophylaxis. No patient had post-removal thromboembolic complications.

A retrievable inferior vena cava filter, as an adjunct to chemical and mechanical prophylaxis, was a safe and effective means of reducing the acute risk of pulmonary embolism in this high-risk group of patients. Although most filters were removed without complications, thereby avoiding the long-term complications that have plagued permanent indwelling filters, a relatively high percentage of filters had to be left in situ.


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 7 | Pages 962 - 967
1 Jul 2009
Aston WJS Calder PR Baker D Hartley J Hill RA

We present a retrospective review of a single-surgeon series of 30 consecutive lengthenings in 27 patients with congenital short femur using the Ilizarov technique performed between 1994 and 2005.

The mean increase in length was 5.8 cm/18.65% (3.3 to 10.4, 9.7% to 48.8%), with a mean time in the frame of 223 days (75 to 363). By changing from a distal to a proximal osteotomy for lengthening, the mean range of knee movement was significantly increased from 98.1° to 124.2° (p = 0.041) and there was a trend towards a reduced requirement for quadricepsplasty, although this was not statistically significant (p = 0.07). The overall incidence of regenerate deformation or fracture requiring open reduction and internal fixation was similar in the distal and proximal osteotomy groups (56.7% and 53.8%, respectively). However, in the proximal osteotomy group, pre-placement of a Rush nail reduced this rate from 100% without a nail to 0% with a nail (p < 0.001). When comparing a distal osteotomy with a proximal one over a Rush nail for lengthening, there was a significant decrease in fracture rate from 58.8% to 0% (p = 0.043).

We recommend that in this group of patients lengthening of the femur with an Ilizarov construct be carried out through a proximal osteotomy over a Rush nail. Lengthening should also be limited to a maximum of 6 cm during one treatment, or 20% of the original length of the femur, in order to reduce the risk of complications.