Aims. The aim of this study was to investigate whether wear and backside deformation of polyethylene (PE) tibial inserts may influence the cement cover of tibial trays of explanted total knee arthroplasties (TKAs). Methods. At our retrieval centre, we measured changes in the wear and deformation of PE inserts using coordinate measuring machines and light microscopy. The amount of cement cover on the backside of tibial trays was quantified as a percentage of the total surface. The study involved data from the explanted fixed-bearing components of four widely used contemporary designs of TKA (Attune, NexGen, Press Fit Condylar (PFC), and Triathlon), revised for any indication, and we compared them with components that used previous generations of PE. Regression modelling was used to identify variables related to the amount of cement cover on the retrieved trays. Results. A total of 114 explanted fixed-bearing TKAs were examined. This included 76 used with contemporary PE inserts which were compared with 15 used with older generation PEs. The Attune and NexGen (central locking) trays were found to have significantly less cement cover than Triathlon and PFC trays (peripheral locking group) (p = 0.001). The median planicity values of the PE inserts used with central locking trays were significantly greater than of those with peripheral locking inserts (205 vs 85 microns; p < 0.001). Attune and NexGen inserts had a characteristic pattern of backside deformation, with the outer edges of the PE deviating inferiorly, leaving the PE margins as the primary areas of articulation. Conclusion. Explanted TKAs with central locking mechanisms were significantly more likely to debond from the cement mantle. The PE inserts of these designs showed characteristic patterns of deformation, which appeared to relate to the manufacturing process and may be exacerbated in vivo. This pattern of deformation was associated with PE wear occurring at the outer edges of the articulation, potentially increasing the
The aim of this study was to evaluate the cost-effectiveness of arthroscopic partial meniscectomy versus physical therapy plus optional delayed arthroscopic partial meniscectomy in young patients aged under 45 years with traumatic meniscal tears. We conducted a multicentre, open-labelled, randomized controlled trial in patients aged 18 to 45 years, with a recent onset, traumatic, MRI-verified, isolated meniscal tear without knee osteoarthritis. Patients were randomized to arthroscopic partial meniscectomy or standardized physical therapy with an optional delayed arthroscopic partial meniscectomy after three months of follow-up. We performed a cost-utility analysis on the randomization groups to compare both treatments over a 24-month follow-up period. Cost utility was calculated as incremental costs per quality-adjusted life year (QALY) gained of arthroscopic partial meniscectomy compared to physical therapy. Calculations were performed from a healthcare system perspective and a societal perspective.Aims
Methods
Loosening of components after total knee arthroplasty (TKA) can be associated with the development of radiolucent lines (RLLs). The aim of this study was to assess the rate of formation of RLLs in the cemented original design of the ATTUNE TKA and their relationship to loosening. A systematic search was undertaken using the Cochrane methodology in three online databases: MEDLINE, Embase, and CINAHL. Studies were screened against predetermined criteria, and data were extracted. Available National Joint Registries in the Network of Orthopaedic Registries of Europe were also screened. A random effects model meta-analysis was undertaken.Aims
Methods
Our objective is to describe our early and mid-term results with the use of a new simple primary knee prosthesis as an articulating spacer in planned two-stage management for infected knee arthroplasty. As a second objective, we compared outcomes between the group with a retained first stage and those with a complete two-stage revision. We included 47 patients (48 knees) with positive criteria for infection, with a minimum two-year follow-up, in which a two-stage approach with an articulating spacer with new implants was used. Patients with infection control, and a stable and functional knee were allowed to retain the initial first-stage components. Outcomes recorded included: infection control rate, reoperations, final range of motion (ROM), and quality of life assessment (QoL) including Western Ontario and McMaster Universities osteoarthritis index, Knee Injury and Osteoarthritis Outcome Score, Oxford Knee Score, 12-Item Short-Form Health Survey questionnaire, and University of California Los Angeles (UCLA) activity score and satisfaction score. These outcomes were evaluated and compared to additional cohorts of patients with retained first-stage interventions and those with a complete two-stage revision. Mean follow-up was 3.7 years (2.0 to 6.5).Aims
Methods
Total knee arthroplasty is an established treatment for knee osteoarthritis with excellent long-term results, but there remains controversy about the role of uncemented prostheses. We present the long-term results of a randomized trial comparing an uncemented tantalum metal tibial component with a conventional cemented component of the same implant design. Patients under the age of 70 years with symptomatic osteoarthritis of the knee were randomized to receive either an uncemented tantalum metal tibial monoblock component or a standard cemented modular component. The mean age at time of recruitment to the study was 63 years (50 to 70), 46 (51.1%) knees were in male patients, and the mean body mass index was 30.4 kg/m2 (21 to 36). The same cruciate retaining total knee system was used in both groups. All patients received an uncemented femoral component and no patients had their patella resurfaced. Patient outcomes were assessed preoperatively and postoperatively using the modified Oxford Knee Score, Knee Society Score, and 12-Item Short-Form Health Survey questionnaire (SF-12) score. Radiographs were analyzed using the American Knee Society Radiograph Evaluation score. Operative complications, reoperations, or revision surgery were recorded. A total of 90 knees were randomized and at last review 77 knees were assessed. In all, 11 patients had died and two were lost to follow-up.Aims
Methods
The purpose of this study was to report the experience of dynamic
intraligamentary stabilisation (DIS) using the Ligamys device for
the treatment of acute ruptures of the anterior cruciate ligament
(ACL). Between March 2011 and April 2012, 50 patients (34 men and 16
women) with an acute rupture of the ACL underwent primary repair
using this device. The mean age of the patients was 30 years (18
to 50). Patients were evaluated for laxity, stability, range of
movement (ROM), Tegner, Lysholm, International Knee Documentation Committee
(IKDC) and visual analogue scale (VAS) scores over a follow-up period
of two years.Aims
Patients and Methods
As the number of younger and more active patients
treated with total knee arthroplasty (TKA) continues to increase,
consideration of better fixation as a means of improving implant
longevity is required. Cemented TKA remains the reference standard
with the largest body of evidence and the longest follow-up to support
its use. However, cementless TKA, may offer the opportunity of a
more bone-sparing procedure with long lasting biological fixation
to the bone. We undertook a review of the literature examining advances
of cementless TKA and the reported results. Cite this article:
Focal femoral inlay resurfacing has been developed
for the treatment of full-thickness chondral defects of the knee. This
technique involves implanting a defect-sized metallic or ceramic
cap that is anchored to the subchondral bone through a screw or
pin. The use of these experimental caps has been advocated in middle-aged
patients who have failed non-operative methods or biological repair
techniques and are deemed unsuitable for conventional arthroplasty
because of their age. This paper outlines the implant design, surgical
technique and biomechanical principles underlying their use. Outcomes
following implantation in both animal and human studies are also reviewed. Cite this article:
The anatomy and microstructure of the menisci
allow the effective distribution of load across the knee. Meniscectomy
alters the biomechanical environment and is a potent risk factor
for osteoarthritis. Despite a trend towards meniscus-preserving
surgery, many tears are irreparable, and many repairs fail. Meniscal allograft transplantation has principally been carried
out for pain in patients who have had a meniscectomy. Numerous case
series have reported a significant improvement in patient-reported
outcomes after surgery, but randomised controlled trials have not
been undertaken. It is scientifically plausible that meniscal allograft transplantation
is protective of cartilage, but this has not been established clinically
to date. Cite this article:
Between April 2004 and July 2007, we performed 241 primary total knee replacements in 204 patients using the e.motion posterior cruciate-retaining, multidirectional mobile-bearing prosthesis. Of these, 100 were carried out using an image-free navigation system, and the remaining 141 with the conventional technique. We conducted a retrospective study from the prospectively collected data of these patients to assess the early results of this new mobile-bearing design. At a mean follow-up of 49 months (32 to 71), 18 knees (7.5%) had mechanical complications of which 13 required revision. Three of these had a peri-prosthetic fracture, and were removed from the study. The indication for revision in the remaining ten was loosening of the femoral component in two, tibiofemoral dislocation in three, disassociation of the polyethylene liner in four, and a broken polyethyene liner in one. There were eight further mechanically unstable knees which presented with recurrent disassociation of the polyethylene liner. There was no significant difference in the incidence of mechanical instability between the navigation-assisted procedures (8 of 99, 8.1%) and the conventionally implanted knees (10 of 139, 7.2%). In our view, the relatively high rate of mechanical complications and revision within 30 months precludes the further use of new design of knee replacement.
We report the results of revision total knee
replacement (TKR) in 26 patients with major metaphyseal osteolytic defects
using 29 trabecular metal cones in conjunction with a rotating hinged
total knee prosthesis. The osteolytic defects were types II and
III (A or B) according to the Anderson Orthopaedic Research Institute
(AORI) classification. The mean age of the patients was 72 years
(62 to 84) and there were 15 men and 11 women. In this series patients had
undergone a mean of 2.34 previous total knee arthroplasties. The
main objective was to restore anatomy along with stability and function
of the knee joint to allow immediate full weight-bearing and active
knee movement. Outcomes were measured using Knee Society scores,
Oxford knee scores, range of movement of the knee and serial radiographs.
Patients were followed for a mean of 36 months (24 to 49). The mean
Oxford knee clinical scores improved from 12.83 (10 to 15) to 35.20
(32 to 38) (p <
0.001) and mean American Knee Society scores
improved from 33.24 (13 to 36) to 81.12 (78 to 86) (p <
0.001).
No radiolucent lines suggestive of loosening were seen around the trabecular
metal cones, and by one year all the radiographs showed good osteo-integration.
There was no evidence of any collapse or implant migration. Our
early results confirm the findings of others that trabecular metal
cones offer a useful way of managing severe bone loss in revision
TKR. Cite this article:
Loss of proprioception following an anterior
cruciate ligament (ACL) injury has been well documented. We evaluated
proprioception in both the injured and the uninjured limb in 25 patients
with ACL injury and in 25 healthy controls, as assessed by joint
position sense (JPS), the threshold for the detection of passive
movement (TDPM) and postural sway during single-limb stance on a
force plate. There were significant proprioceptive deficits in both ACL-deficient
and uninjured knees compared with control knees, as assessed by
the angle reproduction test (on JPS) and postural sway on single
limb stance. The degree of loss of proprioception in the ACL-deficient
knee and the unaffected contralateral knee joint in the same patient
was similar. The TDPM in the injured knee was significantly higher
than that of controls at 30° and 70° of flexion. The TDPM of the
contralateral knee joint was not significantly different from that
in controls. Based on these findings, the effect of proprioceptive training
of the contralateral uninjured knee should be explored. Cite this article:
Antibiotic impregnated articulating spacers are
used in two-stage revision total knee arthroplasty to deliver local antibiotic
therapy while preserving function. We have observed infection control
in greater than 95% of cases with functional outcomes approaching
those seen in revision for aseptic loosening. Higher failure has
been observed with methicillin resistant organisms.
This single-blinded randomised controlled trial
investigated whether one design of mobile-bearing (MB) total knee replacement
(TKR) has any advantage over a fixed-bearing (FB) design on long-term
fixation as measured by radiostereometry. The amount of wear underneath
the mobile bearing was also evaluated. A series of 42 knees was randomised
to MB or FB tibial components with appropriate polyethylene inserts
and followed for between ten and 12 years, or until the death of
the patient. The polyethylene in the MB group was superior in that
it was gamma-irradiated in inert gas and was calcium-stearate free;
the polyethylene in the FB group was gamma-irradiated in air and
contained calcium stearate. In theory this should be advantageous
to the wear rate of the MB group. At final follow-up the overall
mean migration was 0.75 mm ( For the MB group, the mean linear wear rate on the under-surface
was 0.026 mm/year (
We compared the performance of uncemented trabecular metal tibial components in total knee replacement with that of cemented tibial components in patients younger than 60 years over two years using radiostereophotogrammetric analysis (RSA). A total of 22 consecutive patients (mean age 53 years, 33 to 59, 26 knees) received an uncemented NexGen trabecular metal cruciate-retaining monobloc tibial component and 19 (mean 53 years, 44 to 59, 21 knees) a cemented NexGen Option cruciate-retaining modular tibial component. All the trabecular metal components migrated during the initial three months and then stabilised. The exception was external rotation, which did not stabilise until 12 months. Unlike conventional metal-backed implants which displayed a tilting migration comprising subsidence and lift-off from the tibial tray, most of the trabecular metal components showed subsidence only, probably due to the elasticity of the implant. This pattern of subsidence is regarded as being beneficial for uncemented fixation.
We report a retrospective analysis of the results of combined arthroscopically-assisted posterior cruciate ligament reconstruction and open reconstruction of the posterolateral corner in 19 patients with chronic (three or more months) symptomatic instability and pain in the knee. All the operations were performed between 1996 and 2003 and all the patients were assessed pre- and post-operatively by physical examination and by applying three different ligament rating scores. All also had weight-bearing radiographs, MR scans and an examination under anaesthesia and arthroscopy pre-operatively. The posterior cruciate ligament reconstruction was performed using an arthroscopically-assisted single anterolateral bundle technique and the posterolateral corner structures were reconstructed using an open Larson type of tenodesis. The mean follow up was 66.8 months (24 to 110). Pre-operatively, all the patients had a grade III posterior sag according to Clancy and demonstrated more than 20° of external rotation compared with the opposite normal knee on the Dial test. Post-operatively, seven patients (37%) had no residual posterior sag, 11 (58%) had a grade I posterior sag and one (5%) had a grade II posterior sag. In five patients (26%) there was persistent minimal posterolateral laxity. The Lysholm score improved from a mean of 41.2 (28 to 53) to 76.5 (57 to 100) (p = 0.0001) and the Tegner score from a mean of 2.6 (1 to 4) to 6.4 (4 to 9) (p = 0.0001). We conclude that while a combined reconstruction of chronic posterior cruciate ligament and posterolateral corner instability improves the function of the knee, it does not restore complete stability.
Mobile-bearing posterior-stabilised knee replacements have been developed as an alternative to the standard fixed- and mobile-bearing designs. However, little is known about the We conclude that mobile-bearing posterior-stabilised knee replacements reproduce internal rotation of the tibia more closely during flexion than fixed-bearing posterior-stabilised designs. Furthermore, mobile-bearing posterior-stabilised knee replacements demonstrate a unidirectional movement which occurs at the upper and lower sides of the mobile insert. The femur moves in an anteroposterior direction on the upper surface of the insert, whereas the movement at the lower surface is pure rotation. Such unidirectional movement may lead to less wear when compared with the multidirectional movement seen in fixed-bearing posterior-stabilised knee replacements, and should be associated with more evenly applied cam-post stresses.
We carried out a prospective study of 71 patients who had undergone reconstruction of the anterior cruciate ligament with the ABC scaffold. Their mean age was 28 years (18 to 50). All had either sub-acute or chronic traumatic deficiency of the ligament. The mean period of follow-up was five years (four to seven). Assessment included the use of the International Knee Documentation Committee score, the modified Lysholm score, the Tegner Activity score, the Knee Injury and Osteoarthritis Outcome score and measurement with the KT-1000 arthrometer. Two patients had mild recurrent synovitis. There were no infections and no failures of the ligament. During the period of study, two patients sustained a traumatic fracture of a femoral condyle. The implants retained their integrity in both cases. All patients returned to their previous or enhanced levels of daily activity by three months after operation and 56 (79%) achieved their pre-injury level of sporting activity by six months. The patients who were competing in National level sports returned to play at one level less after operation than before. The Lysholm score showed that 58% of the patients (41) were excellent, 34% (24) good, and 8% (6) fair, with a mean post-operative score of 93. According to the International Knee Documentation Committee score, 35% of knees (25) were ‘normal’, 52% (37) ‘nearly normal’ and 13% (9) ‘abnormal’. Complete satisfaction was noted in 90% of patients (64). The development of osteoarthritis and the management of anterior cruciate deficiency associated with laxity of the medial collateral ligament remains uncertain. Our results indicate that in the medium-term, the ABC ligament scaffold is suitable and effective when early and safe return to unrestricted activities is demanded. We acknowledge the current general hostility towards reconstruction of the anterior cruciate ligament with artificial materials following reports of early failure and chronic synovitis associatiated with the production of particulate debris. We did not encounter these problems.