The risk of mechanical failure of modular revision hip stems is frequently mentioned in the literature, but little is currently known about the actual clinical failure rates of this type of prosthesis. The current retrospective long-term analysis examines the distal and modular failure patterns of the Prevision hip stem from 18 years of clinical use. A design improvement of the modular taper was introduced in 2008, and the data could also be used to compare the original and the current design of the modular connection. We performed an analysis of the Prevision modular hip stem using the manufacturer’s vigilance database and investigated different mechanical failure patterns of the hip stem from January 2004 to December 2022.Aims
Methods
Adverse spinal motion or balance (spine mobility) and adverse pelvic mobility, in combination, are often referred to as adverse spinopelvic mobility (SPM). A stiff lumbar spine, large posterior standing pelvic tilt, and severe sagittal spinal deformity have been identified as risk factors for increased hip instability. Adverse SPM can create functional malposition of the acetabular components and hence is an instability risk. Adverse pelvic mobility is often, but not always, associated with abnormal spinal motion parameters. Dislocation rates for dual-mobility articulations (DMAs) have been reported to be between 0% and 1.1%. The aim of this study was to determine the early survivorship from the Australian Orthopaedic Association National Joint Replacement Registry (AOANJRR) of patients with adverse SPM who received a DMA. A multicentre study was performed using data from 227 patients undergoing primary total hip arthroplasty (THA), enrolled consecutively. All the patients who had one or more adverse spine or pelvic mobility parameter had a DMA inserted at the time of their surgery. The mean age was 76 years (22 to 93) and 63% were female (n = 145). At a mean of 14 months (5 to 31) postoperatively, the AOANJRR was analyzed for follow-up information. Reasons for revision and types of revision were identified.Aims
Methods
Registry studies on modified acetabular polyethylene (PE) liner designs are limited. We investigated the influence of standard and modified PE acetabular liner designs on the revision rate for mechanical complications in primary cementless total hip arthroplasty (THA). We analyzed 151,096 primary cementless THAs from the German Arthroplasty Registry (EPRD) between November 2012 and November 2020. Cumulative incidence of revision for mechanical complications for standard and four modified PE liners (lipped, offset, angulated/offset, and angulated) was determined using competing risk analysis at one and seven years. Confounders were investigated with a Cox proportional-hazards model.Aims
Methods
The purpose of this study is to examine six types of bearing surfaces implanted at a single institution over three decades to determine whether the reasons for revision vary among the groups and how long it takes to identify differences in survival. We considered six cohorts that included a total of 1,707 primary hips done between 1982 and 2010. These included 223 conventional polyethylene sterilized with γ irradiation in air (CPE-GA), 114 conventional polyethylene sterilized with gas plasma (CPE-GP), 116 crosslinked polyethylene (XLPE), 1,083 metal-on-metal (MOM), 90 ceramic-on-ceramic (COC), and 81 surface arthroplasties (SAs). With the exception of the COC, all other groups used cobalt-chromium (CoCr) femoral heads. The mean follow-up was 10 (0.008 to 35) years. Descriptive statistics with revisions per 100 component years (re/100 yr) and survival analysis with revision for any reason as the endpoint were used to compare bearing surfaces.Aims
Methods
We retrospectively reviewed 44 consecutive patients
(50 hips) who underwent acetabular re-revision after a failed previous
revision that had been performed using structural or morcellised
allograft bone, with a cage or ring for uncontained defects. Of
the 50 previous revisions, 41 cages and nine rings were used with
allografts for 14 minor-column and 36 major-column defects. We routinely
assessed the size of the acetabular bone defect at the time of revision
and re-revision surgery. This allowed us to assess whether host
bone stock was restored. We also assessed the outcome of re-revision
surgery in these circumstances by means of radiological characteristics,
rates of
Aims. Joint-preserving surgery of the hip (JPSH) has evolved considerably
and now includes a number of procedures, including arthroscopy,
surgical dislocation, and redirectional osteotomies of the femur
and acetabulum. There are a number of different factors which lead
to failure of JPSH. Consequently, it is of interest to assess the
various
To present a surgically relevant update of trunnionosis. Systematic review performed April 2017.Aims
Materials and Methods
Hip resurfacing has been proposed as an alternative
to traditional total hip arthroplasty in young, active patients.
Much has been learned following the introduction of metal-on-metal resurfacing
devices in the 1990s. The triad of a well-designed device, implanted
accurately, in the correct patient has never been more critical
than with these implants. Following Food and Drug Administration approval in 2006, we studied
the safety and effectiveness of one hip resurfacing device (Birmingham
Hip Resurfacing) at our hospital in a large, single-surgeon series.
We report our early to mid–term results in 1333 cases followed for
a mean of 4.3 years (2 to 5.7) using a prospective, observational
registry. The mean patient age was 53.1 years (12 to 84); 70% were
male and 91% had osteoarthritis. Complications were few, including
no dislocations, no femoral component loosening, two femoral neck
fractures (0.15%), one socket loosening (0.08%), three deep infections
(0.23%), and three cases of metallosis (0.23%). There were no destructive
pseudotumours. Overall survivorship at up to 5.7 years was 99.2%. Aseptic survivorship
in males under the age of 50 was 100%. We believe this is the largest
United States series of a single surgeon using a single resurfacing
system. Cite this article:
A retrospective study was conducted to investigate
the changes in metal ion levels in a consecutive series of Birmingham
Hip Resurfacings (BHRs) at a minimum ten-year follow-up. We reviewed
250 BHRs implanted in 232 patients between 1998 and 2001. Implant
survival, clinical outcome (Harris hip score), radiographs and serum chromium
(Cr) and cobalt (Co) ion levels were assessed. Of 232 patients, 18 were dead (five bilateral BHRs), 15 lost
to follow-up and ten had been revised. The remaining 202 BHRs in
190 patients (136 men and 54 women; mean age at surgery 50.5 years
(17 to 76)) were evaluated at a minimum follow-up of ten years (mean
10.8 years (10 to 13.6)). The overall implant survival at 13.2 years
was 92.4% (95% confidence interval 90.8 to 94.0). The mean Harris
hip score was 97.7 (median 100; 65 to 100). Median and mean ion
levels were low for unilateral resurfacings (Cr: median 1.3 µg/l,
mean
1.95 µg/l (<
0.5 to 16.2); Co: median 1.0 µg/l, mean 1.62 µg/l
(<
0.5 to 17.3)) and bilateral resurfacings (Cr: median 3.2 µg/l,
mean 3.46 µg/l (<
0.5 to 10.0); Co: median 2.3 µg/l, mean 2.66
µg/l (<
0.5 to 9.5)). In 80 unilateral BHRs with sequential ion
measurements, Cr and Co levels were found to decrease significantly
(p <
0.001) from the initial assessment at a median of six years
(4 to 8) to the last assessment at a median of 11 years (9 to 13),
with a mean reduction of 1.24 µg/l for Cr and 0.88 µg/l for Co.
Three female patients had a >
2.5 µg/l increase of Co ions, associated with
head sizes ≤ 50 mm, clinical symptoms and osteolysis. Overall, there
was no significant difference in change of ion levels between genders
(Cr, p = 0.845; Co, p = 0.310) or component sizes (Cr, p = 0.505;
Co, p = 0.370). Higher acetabular component inclination angles correlated
with greater change in ion levels (Cr, p = 0.013; Co, p = 0.002).
Patients with increased ion levels had lower Harris hip scores (p
= 0.038). In conclusion, in well-functioning BHRs the metal ion levels
decreased significantly at ten years. An increase >
2.5 µg/l was
associated with poor function. Cite this article:
We report a 12- to 15-year implant survival assessment
of a prospective single-surgeon series of Birmingham Hip Resurfacings
(BHRs). The earliest 1000 consecutive BHRs including 288 women (335
hips) and 598 men (665 hips) of all ages and diagnoses with no exclusions
were prospectively followed-up with postal questionnaires, of whom
the first 402 BHRs (350 patients) also had clinical and radiological
review. Mean follow-up was 13.7 years (12.3 to 15.3). In total, 59 patients
(68 hips) died 0.7 to 12.6 years following surgery from unrelated
causes. There were 38 revisions, 0.1 to 13.9 years (median 8.7)
following operation, including 17 femoral failures (1.7%) and seven
each of infections, soft-tissue reactions and other causes. With
revision for any reason as the end-point Kaplan–Meier survival analysis
showed 97.4% (95% confidence interval (CI) 96.9 to 97.9) and 95.8%
(95% CI 95.1 to 96.5) survival at ten and 15 years, respectively.
Radiological assessment showed 11 (3.5%) femoral and 13 (4.1%) acetabular
radiolucencies which were not deemed failures and one radiological
femoral failure (0.3%). Our study shows that the performance of the BHR continues to
be good at 12- to 15-year follow-up. Men have better implant survival
(98.0%; 95% CI 97.4 to 98.6) at 15 years than women (91.5%; 95%
CI 89.8 to 93.2), and women <
60 years (90.5%; 95% CI 88.3 to
92.7) fare worse than others. Hip dysplasia and osteonecrosis are
risk factors for failure. Patients under 50 years with osteoarthritis
fare best (99.4%; 95% CI 98.8 to 100 survival at 15 years), with
no failures in men in this group. Cite this article:
Metal-on-metal hip arthroplasty gained significant
favor in the first decade of the millennium. However, the past several
years have seen increasing reports of failure, pseudotumor and other
adverse reactions. This study presents the results of a single center’s
15-year experience with metal-on-metal total hip arthroplasty as
strong evidence that metal-on-metal is going, going, gone.
Despite the increasing interest and subsequent published literature on hip resurfacing arthroplasty, little is known about the prevalence of its complications and in particular the less common