Between October 2001 and September 2009 we lengthened 242 lower-limb segments in 180 patients using the Intramedullary Skeletal Kinetic Distractor (ISKD). Mechanical failure was defined either as breakage of the ISKD or failure of the internal mechanism to activate. Retrieved nails which failed mechanically were examined by the manufacturer for defects. In all, 15 ISKDs in 12 patients (13 limbs) failed mechanically representing an overall failure rate of 6.2%, with fracture of the device occurring in ten of the 15 failures. Two nails in one patient failed to lengthen and had to be replaced. The manufacturer detected an
Patients undergoing femoral lengthening by external fixation
tolerate treatment less well when compared to tibial lengthening.
Lengthening of the femur with an intramedullary device may have
advantages. We reviewed all cases of simple femoral lengthening performed
at our unit from 2009 to 2014. Cases of nonunions, concurrent deformities,
congenital limb deficiencies and lengthening with an unstable hip
were excluded, leaving 33 cases (in 22 patients; 11 patients had
bilateral procedures) for review. Healing index, implant tolerance
and complications were compared.Aims
Patients and Methods
This study describes the Osseointegration Group of Australia’s
Accelerated Protocol two-stage strategy (OGAAP-1) for the osseointegrated
reconstruction of amputated limbs. We report clinical outcomes in 50 unilateral trans-femoral amputees
with a mean age of 49.4 years (24 to 73), with a minimum one-year
follow-up. Outcome measures included the Questionnaire for persons
with a Trans-Femoral Amputation, the health assessment questionnaire
Short-Form-36 Health Survey, the Amputation Mobility Predictor scores
presented as K-levels, 6 Minute Walk Test and timed up and go tests.
Adverse events included soft-tissue problems, infection, fractures
and failure of the implant.Aims
Patients and Methods
Deep vein thrombosis is a common complication
of immobilising the lower limb after surgery. We hypothesised that
intermittent pneumatic compression (IPC) therapy in outpatients
who had undergone surgical repair of acute ruptures of the Achilles
tendon could reduce the incidence of this problem. A total of 150 patients who had undergone surgical repair of
the Achilles tendon were randomised to either treatment with IPC
for six hours per day (n = 74) under an orthosis or treatment as
usual (n = 74) in a plaster cast without IPC. At two weeks post-operatively,
the incidence of deep vein thrombosis was assessed using blinded, double-reported
compression duplex ultrasound. At this point, IPC was discontinued
and all patients were immobilised in an orthosis for a further four
weeks. At six weeks post-operatively, a second compression duplex ultrasound
scan was performed. At two weeks, the incidence of deep vein thrombosis was 21% in
the treated group and 37% in the control group (p = 0.042). Age
over 39 years was found to be a strong risk factor for deep vein
thrombosis (odds ratio (OR) = 4.84, 95% confidence interval (CI)
2.14 to 10.96). Treatment with IPC, corrected for age differences
between groups, reduced the risk of deep vein thrombosis at the
two-week point (OR = 2.60; 95% CI 1.15 to 5.91; p =0.022). At six weeks,
the incidence of deep vein thrombosis was 52% in the treated group
and 48% in the control group (OR 0.94, 95% CI 0.49 to 1.83). IPC
appears to be an effective method of reducing the risk of deep vein
thrombosis in the early stages of post-operative immobilisation
of outpatients. Further research is necessary to elucidate whether
it can confer similar benefits over longer periods of immobilisation
and in a more heterogeneous group of patients. Cite this article:
We compared lower limb coronal alignment measurements
obtained pre- and post-operatively with long-leg radiographs and
computer navigation in patients undergoing primary total knee replacement
(TKR). A series of 185 patients had their pre- and post-implant
radiological and computer-navigation system measurements of coronal alignment
compared using the Bland-Altman method. The study included 81 men
and 104 women with a mean age of 68.5 years (32 to 87) and a mean
body mass index of 31.7 kg/m2 (19 to 49). Pre-implant
Bland–Altman limits of agreement were -9.4° to 8.6° with a repeatability
coefficient of 9.0°. The Bland–Altman plot showed a tendency for the
radiological measurement to indicate a higher level of pre-operative
deformity than the corresponding navigation measurement. Post-implant
limits of agreement were -5.0° to 5.4° with a repeatability coefficient
of 5.2°. The tendency for valgus knees to have greater deformity
on the radiograph was still seen, but was weaker for varus knees. The alignment seen or measured intra-operatively during TKR is
not necessarily the same as the deformity seen on a standing long-leg
radiograph either pre- or post-operatively. Further investigation
into the effect of weight-bearing and surgical exposure of the joint
on the mechanical femorotibial angle is required to enable the most appropriate
intra-operative alignment to be selected.
We carried out a systematic review of the literature
to evaluate the evidence regarding the clinical results of the Ilizarov
method in the treatment of long bone defects of the lower limbs. Only 37 reports (three non-randomised comparative studies, one
prospective study and 33 case-series) met our inclusion criteria.
Although several studies were unsatisfactory in terms of statistical
heterogeneity, our analysis appears to show that the Ilizarov method
of distraction osteogenesis significantly reduced the risk of deep
infection in infected osseous lesions (risk ratio 0.14 (95% confidence
interval (CI) 0.10 to 0.20), p <
0.001). However, there was a
rate of re-fracture of 5% (95% CI 3 to 7), with a rate of neurovascular
complications of 2.2% (95% CI 0.3 to 4) and an amputation rate of
2.9% (95% CI 1.4 to 4.4).The data was generally not statistically
heterogeneous. Where tibial defects were >
8 cm, the risk of re-fracture
increased (odds ratio 3.7 (95% CI 1.1 to 12.5), p = 0.036). The technique is demanding for patients, illustrated by the voluntary
amputation rate of 1.6% (95% CI 0 to 3.1), which underlines the
need for careful patient selection. Cite this article:
Internal lengthening devices in the femur lengthen
along the anatomical axis, potentially creating lateral shift of
the mechanical axis. We aimed to determine whether femoral lengthening
along the anatomical axis has an inadvertent effect on lower limb
alignment. Isolated femoral lengthening using the Intramedullary
Skeletal Kinetic Distractor was performed in 27 femora in 24 patients
(mean age 32 years (16 to 57)). Patients who underwent simultaneous realignment
procedures or concurrent tibial lengthening, or who developed mal-
or nonunion, were excluded. Pre-operative and six-month post-operative
radiographs were used to measure lower limb alignment. The mean lengthening
achieved was 4.4 cm (1.5 to 8.0). In 26 of 27 limbs, the mechanical
axis shifted laterally by a mean of 1.0 mm/cm of lengthening (0
to 3.5). In one femur that was initially in varus, a 3 mm medial
shift occurred during a lengthening of 2.2 cm. In a normally aligned limb, intramedullary lengthening along
the anatomical axis of the femur results in a lateral shift of the
mechanical axis by approximately 1 mm for each 1 cm of lengthening.
We report the results of intramedullary leg lengthening conducted between 2002 and 2009 using the Intramedullary Skeletal Kinetic Distractor in 69 unilateral lengthenings involving 58 femora and 11 tibiae. We identified difficulties that occurred during the treatment and assessed whether they were specifically due to the implant or independent of it. Paley’s classification for evaluating problems, obstacles and complications with external fixators was adopted, and implant-specific difficulties were continuously noted. There were seven failures requiring premature removal of the device, in four due to nail breakage and three for other reasons, and five unsuccessful outcomes after completion of the lengthening. In all, 116 difficulties were noted in 45 patients, with only 24 having problem-free courses. In addition to the difficulties arising from the use of external fixators, there were almost the same number again of implant-specific difficulties. Nevertheless, successful femoral lengthening was achieved in 52 of the 58 patients (90%). However, successful tibial lengthening was only achieved in five of 11 patients (45%).
Patients with diabetes mellitus may develop plantar flexion contractures (equinus) which may increase forefoot pressure during walking. In order to determine the relationship between equinus and forefoot pressure, we measured forefoot pressure during walking in 27 adult diabetics with a mean age of 66.3 years ( Simple linear regression showed that the relationship between equinus and peak forefoot pressure was significant (p <
0.0471), but that only a small portion of the variance was accounted for (R2 = 0.149). This indicates that equinus has only a limited role in causing high forefoot pressure. Our findings suggest caution in undertaking of tendon-lengthening procedures to reduce peak forefoot plantar pressures in diabetic subjects until clearer indications are established.
In a prospective randomised study 31 patients were allocated to either arthrodesis or Mayo resection of the first metatarsophalangeal joint as part of a total reconstruction of the rheumatoid forefoot. Of these, 29 were re-examined after a mean of 72 months (57 to 80), the Foot Function Index was scored and any deformity measured. Load distribution was analysed using a Fscan mat in 14 cases, and time and distance were measured in 12 of these patients using a 3D Motion system. We found excellent patient satisfaction and a significant, lasting reduction of the Foot Function Index, with no statistically significant differences between the groups. There were no significant differences in recurrence of the deformity, the need for special shoes, gait velocity, step length, plantar moment, mean pressure or the position of the centre of force under the forefoot. The cadence was higher and the stance phase shorter in the fusion group. These results suggest that a Mayo resection may be an equally good option for managing the first metatarsophalangeal joint in reconstruction of the rheumatoid forefoot.
We performed two independent, randomised, controlled trials in order to assess the potential benefits of immediate weight-bearing mobilisation after rupture of the tendo Achillis. The first trial, on operatively-treated patients showed an improved functional outcome for patients mobilised fully weight-bearing after surgical repair. Two cases of re-rupture in the treatment group suggested that careful patient selection is required as patients need to follow a structured rehabilitation regimen. The second trial, on conservatively-treated patients, provided no evidence of a functional benefit from immediate weight-bearing mobilisation. However, the practical advantages of immediate weight-bearing did not predispose the patients to a higher complication rate. In particular, there was no evidence of tendon lengthening or a higher re-rupture rate. We would advocate immediate weight-bearing mobilisation for the rehabilitation of all patients with rupture of the tendo Achillis.