header advert
Results 1 - 20 of 58
Results per page:
The Bone & Joint Journal
Vol. 102-B, Issue 7 Supple B | Pages 27 - 32
1 Jul 2020
Heckmann N Weitzman DS Jaffri H Berry DJ Springer BD Lieberman JR

Aims. Dual mobility (DM) bearings are an attractive treatment option to obtain hip stability during challenging primary and revision total hip arthroplasty (THA) cases. The purpose of this study was to analyze data submitted to the American Joint Replacement Registry (AJRR) to characterize utilization trends of DM bearings in the USA. Methods. All primary and revision THA procedures reported to AJRR from 2012 to 2018 were analyzed. Patients of all ages were included and subdivided into DM and traditional bearing surface cohorts. Patient demographics, geographical region, hospital size, and teaching affiliation were assessed. Associations were determined by chi-squared analysis and logistic regression was performed to assess outcome variables. Results. A total of 406,900 primary and 34,745 revision THAs were identified, of which 35,455 (8.7%) and 8,031 (23.1%) received DM implants respectively. For primary THA, DM usage increased from 6.7% in 2012 to 12.0% in 2018. Among revision THA, DM use increased from 19.5% in 2012 to 30.6% in 2018. Patients < 50 years of age had the highest rates of DM implantation in every year examined. For each year of increase in age, there was a 0.4% decrease in the rate of DM utilization (odds ratio (OR) 0.996 (95% confidence interval (CI) 0.995 to 0.997); p < 0.001). Females were more likely to receive a DM implant compared to males (OR 1.077 (95% CI 1.054 to 1.100); p < 0.001). Major teaching institutions and smaller hospitals were associated with higher rates of utilization. DM articulations were used more commonly for dysplasia compared with osteoarthritis (OR 2.448 (95% CI 2.032 to 2.949); p < 0.001) during primary THA and for instability (OR 3.130 (95% CI 2.751 to 3.562) vs poly-wear; p < 0.001) in the revision setting. Conclusion. DM articulations showed a marked increase in utilization during the period examined. Younger patient age, female sex, and hospital characteristics such as teaching status, smaller size, and geographical location were associated with increased utilization. DM articulations were used more frequently for primary THA in patients with dysplasia and for revision THA in patients being treated for instability. Cite this article: Bone Joint J 2020;102-B(7 Supple B):27–32


The Bone & Joint Journal
Vol. 102-B, Issue 7 Supple B | Pages 20 - 26
1 Jul 2020
Romero J Wach A Silberberg S Chiu Y Westrich G Wright TM Padgett DE

Aims. This combined clinical and in vitro study aimed to determine the incidence of liner malseating in modular dual mobility (MDM) constructs in primary total hip arthroplasties (THAs) from a large volume arthroplasty centre, and determine whether malseating increases the potential for fretting and corrosion at the modular metal interface in malseated MDM constructs using a simulated corrosion chamber. Methods. For the clinical arm of the study, observers independently reviewed postoperative radiographs of 551 primary THAs using MDM constructs from a single manufacturer over a three-year period, to identify the incidence of MDM liner-shell malseating. Multivariable logistic regression analysis was performed to identify risk factors including age, sex, body mass index (BMI), cup design, cup size, and the MDM case volume of the surgeon. For the in vitro arm, six pristine MDM implants with cobalt-chrome liners were tested in a simulated corrosion chamber. Three were well-seated and three were malseated with 6° of canting. The liner-shell couples underwent cyclic loading of increasing magnitudes. Fretting current was measured throughout testing and the onset of fretting load was determined by analyzing the increase in average current. Results. The radiological review identified that 32 of 551 MDM liners (5.8%) were malseated. Malseating was noted in all of the three different cup designs. The incidence of malseating was significantly higher in low-volume MDM surgeons than high-volume MDM surgeons (p < 0.001). Pristine well-seated liners showed significantly lower fretting current values at all peak loads greater than 800 N (p < 0.044). Malseated liner-shell couples had lower fretting onset loads at 2,400 N. Conclusion. MDM malseating remains an issue that can occur in at least one in 20 patients at a high-volume arthroplasty centre. The onset of fretting and increased fretting current throughout loading cycles suggests susceptibility to corrosion when this occurs. These results support the hypothesis that malseated liners may be at risk for fretting corrosion. Clinicians should be aware of this phenomenon. Cite this article: Bone Joint J 2020;102-B(7 Supple B):20–26


The Bone & Joint Journal
Vol. 102-B, Issue 4 | Pages 423 - 425
1 Apr 2020
Hoggett L Cross C Helm A

Aims. Dislocation remains a significant complication after total hip arthroplasty (THA), being the third leading indication for revision. We present a series of acetabular revision using a dual mobility cup (DMC) and compare this with our previous series using the posterior lip augmentation device (PLAD). Methods. A retrospective review of patients treated with either a DMC or PLAD for dislocation in patients with a Charnley THA was performed. They were identified using electronic patient records (EPR). EPR data and radiographs were evaluated to determine operating time, length of stay, and the incidence of complications and recurrent dislocation postoperatively. Results. A total of 28 patients underwent revision using a DMC for dislocation following Charnley THA between 2013 and 2017. The rate of recurrent dislocation and overall complications were compared with those of a previous series of 54 patients who underwent revision for dislocation using a PLAD, between 2007 and 2013. There was no statistically significant difference in the mean distribution of sex or age between the groups. The mean operating time was 71 mins (45 to 113) for DMCs and 43 mins (21 to 84) for PLADs (p = 0.001). There were no redislocations or revisions in the DMC group at a mean follow-up of 55 months (21 to 76), compared with our previous series of PLAD which had a redislocation rate of 16% (n = 9) and an overall revision rate of 25% (n = 14, p = 0.001) at a mean follow-up of 86 months (45 to 128). Conclusion. These results indicate that DMC outperforms PLAD as a treatment for dislocation in patients with a Charnley THA. This should therefore be the preferred form of treatment for these patients despite a slightly longer operating time. Work is currently ongoing to review outcomes of DMC over a longer follow-up period. PLAD should be used with caution in this patient group with preference given to acetabular revision to DMC. Cite this article: Bone Joint J 2020;102-B(4):423–425


The Bone & Joint Journal
Vol. 99-B, Issue 1_Supple_A | Pages 18 - 24
1 Jan 2017
De Martino I D’Apolito R Soranoglou VG Poultsides LA Sculco PK Sculco TP

Aims. The aim of this systematic review was to report the rate of dislocation following the use of dual mobility (DM) acetabular components in primary and revision total hip arthroplasty (THA). Materials and Methods. A systematic review of the literature according to the Preferred Reporting Items for Systematic Reviews and Meta-analyses guidelines was performed. A comprehensive search of Pubmed/Medline, Cochrane Library and Embase (Scopus) was conducted for English articles between January 1974 and March 2016 using various combinations of the keywords “dual mobility”, “dual-mobility”, “tripolar”, “double-mobility”, “double mobility”, “hip”, “cup”, “socket”. The following data were extracted by two investigators independently: demographics, whether the operation was a primary or revision THA, length of follow-up, the design of the components, diameter of the femoral head, and type of fixation of the acetabular component. Results. In all, 59 articles met our inclusion criteria. These included a total of 17 908 THAs which were divided into two groups: studies dealing with DM components in primary THA and those dealing with these components in revision THA. The mean rate of dislocation was 0.9% in the primary THA group, and 3.0% in the revision THA group. The mean rate of intraprosthetic dislocation was 0.7% in primary and 1.3% in revision THAs. Conclusion. Based on the current data, the use of DM acetabular components are effective in minimising the risk of instability after both primary and revision THA. This benefit must be balanced against continuing concerns about the additional modularity, and the new mode of failure of intraprosthetic dislocation. Longer term studies are needed to assess the function of these newer materials compared with previous generations. . Cite this article: Bone Joint J 2017;99-B(1 Supple A):18–24


The Bone & Joint Journal
Vol. 101-B, Issue 4 | Pages 365 - 371
1 Apr 2019
Nam D Salih R Nahhas CR Barrack RL Nunley RM

Aims. Modular dual mobility (DM) prostheses in which a cobalt-chromium liner is inserted into a titanium acetabular shell (vs a monoblock acetabular component) have the advantage of allowing supplementary screw fixation, but the potential for corrosion between the liner and acetabulum has raised concerns. While DM prostheses have shown improved stability in patients deemed ‘high-risk’ for dislocation undergoing total hip arthroplasty (THA), their performance in young, active patients has not been reported. This study’s purpose was to assess clinical outcomes, metal ion levels, and periprosthetic femoral bone mineral density (BMD) in young, active patients receiving a modular DM acetabulum and recently introduced titanium, proximally coated, tapered femoral stem design. Patients and Methods. This was a prospective study of patients between 18 and 65 years of age, with a body mass index (BMI) < 35 kg/m. 2. and University of California at Los Angeles (UCLA) activity score > 6, who received a modular cobalt-chromium acetabular liner, highly crosslinked polyethylene mobile bearing, and cementless titanium femoral stem for their primary THA. Patients with a history of renal disease and metal hardware elsewhere in the body were excluded. A total of 43 patients (30 male, 13 female; mean age 52.6 years (. sd. 6.5)) were enrolled. All patients had a minimum of two years’ clinical follow-up. Patient-reported outcome measures, whole blood metal ion levels (ug/l), and periprosthetic femoral BMD were measured at baseline, as well as at one and two years postoperatively. Power analysis indicated 40 patients necessary to demonstrate a five-fold increase in cobalt levels from baseline (alpha = 0.05, beta = 0.80). A mixed model with repeated measures was used for statistical analysis. Results. Mean Harris Hip Scores improved from 54.1 (. sd. 20.5) to 91.2 (. sd. 10.8) at two years postoperatively (p < 0.001). All patients had radiologically well-fixed components, no patients experienced any instability, and no patients required any further intervention. Mean cobalt levels increased from 0.065 ug/l (. sd. 0.03) preoperatively to 0.30 ug/l (. sd. 0.51) at one year postoperatively (p = 0.01) but decreased at two years postoperatively to 0.16 ug/l (. sd. 0.23; p = 0.2). Four patients (9.3%) had a cobalt level outside the reference range (0.03 ug/l to 0.29 ug/l) at two years postoperatively, with values from 0.32 ug/l to 0.94 ug/l. The mean femoral BMD ratio was maintained in Gruen zones 2 to 7 at both one and two years postoperatively using this stem design. At two years postoperatively, mean BMD in the medial calcar was 101.5% of the baseline value. Conclusion. Use of a modular DM prosthesis and cementless, tapered femoral stem has shown encouraging results in young, active patients undergoing primary THA. Elevation in mean cobalt levels and the presence of four patients outside the reference range at two years postoperatively demonstrates the necessity of continued surveillance in this cohort. Cite this article: Bone Joint J 2019;101-B:365–371


The Bone & Joint Journal
Vol. 98-B, Issue 1_Supple_A | Pages 60 - 63
1 Jan 2016
Ko LM Hozack WJ

Dual mobility cups have two points of articulation, one between the shell and the polyethylene (external bearing) and one between the polyethylene and the femoral head (internal bearing). Movement occurs at the inner bearing; the outer bearing only moves at extremes of movement. Dislocation after total hip arthroplasty (THA) is a cause of much morbidity and its treatment has significant cost implications. Dual mobility cups provide an increased range of movement and a may reduce the risk of dislocation. . This paper reviews the use of these cups in THA, particularly where stability is an issue. Dual mobility cups may be of benefit in primary THA in patients at a high risk of dislocation, such as those who are older with increased comorbidities and a higher American Association of Anesthesiology grade and those with a neuromuscular disease. They may be used at revision surgery where the risk of dislocation is high, such as in patients with many prior dislocations, or those with abductor deficiency. They may also be used in THA for displaced fractures of the femoral neck, which has a notoriously high rate of dislocation. Cite this article: Bone Joint J 2016;98-B(1 Suppl A):60–3


The Bone & Joint Journal
Vol. 103-B, Issue 7 | Pages 1238 - 1246
1 Jul 2021
Hemmerling KJ Weitzler L Bauer TW Padgett DE Wright TM

Aims. Dual mobility implants in total hip arthroplasty are designed to increase the functional head size, thus decreasing the potential for dislocation. Modular dual mobility (MDM) implants incorporate a metal liner (e.g. cobalt-chromium alloy) in a metal shell (e.g. titanium alloy), raising concern for mechanically assisted crevice corrosion at the modular liner-shell connection. We sought to examine fretting and corrosion on MDM liners, to analyze the corrosion products, and to examine histologically the periprosthetic tissues. Methods. A total of 60 retrieved liners were subjectively scored for fretting and corrosion. The corrosion products from the three most severely corroded implants were removed from the implant surface, imaged using scanning electron microscopy, and analyzed using Fourier-transform infrared spectroscopy. Results. Fretting was present on 88% (53/60) of the retrieved liners, and corrosion was present on 97% (58/60). Fretting was most often found on the lip of the taper at the transition between the lip and the dome regions. Macrophages and particles reflecting an innate inflammatory reaction to corrosion debris were noted in six of the 48 cases for which periprosthetic tissues were examined, and all were associated with retrieved components that had high corrosion scores. Conclusion. Our results show that corrosion occurs at the interface between MDM liners and shells and that it can be associated with reactions in the local tissues, suggesting continued concern that this problem may become clinically important with longer-term use of these implants. Cite this article: Bone Joint J 2021;103-B(7):1238–1246


The Bone & Joint Journal
Vol. 103-B, Issue 7 Supple B | Pages 66 - 72
1 Jul 2021
Hernandez NM Hinton ZW Wu CJ Lachiewicz PF Ryan SP Wellman SS

Aims. Modular dual mobility (MDM) acetabular components are often used with the aim of reducing the risk of dislocation in revision total hip arthroplasty (THA). There is, however, little information in the literature about its use in this context. The aim of this study, therefore, was to evaluate the outcomes in a cohort of patients in whom MDM components were used at revision THA, with a mean follow-up of more than five years. Methods. Using the database of a single academic centre, 126 revision THAs in 117 patients using a single design of an MDM acetabular component were retrospectively reviewed. A total of 94 revision THAs in 88 patients with a mean follow-up of 5.5 years were included in the study. Survivorship was analyzed with the endpoints of dislocation, reoperation for dislocation, acetabular revision for aseptic loosening, and acetabular revision for any reason. The secondary endpoints were surgical complications and the radiological outcome. Results. The overall rate of dislocation was 11%, with a six-year survival of 91%. Reoperation for dislocation was performed in seven patients (7%), with a six-year survival of 94%. The dislocations were early (at a mean of 33 days) in six patients, and late (at a mean of 4.3 years) in four patients. There were three intraprosthetic dissociations. An outer head diameter of ≥ 48 mm was associated with a lower risk of dislocation (p = 0.013). Lumbrosacral fusion was associated with increased dislocation (p = 0.004). Four revision THAs (4%) were further revised for aseptic acetabular loosening, and severe bone loss (Paprosky III) at the time of the initial revision was significantly associated with further revision for aseptic acetabular loosening (p = 0.008). Fourteen acetabular components (15%) were re-revised for infection, and a pre-revision diagnosis of reimplantation after periprosthetic joint infection (PJI) was associated with subsequent PJI (p < 0.001). Two THAs had visible metallic changes on the backside of the cobalt chromium liner. Conclusion. When using this MDM component in revision THA, at a mean follow-up of 5.5 years, there was a higher rate of dislocation (11%) than previously reported. The size of the outer bearing was related to the risk of dislocation. There was a low rate of aseptic acetabular loosening. Longer follow-up of this MDM component and evaluation of other designs are warranted. Cite this article: Bone Joint J 2021;103-B(7 Supple B):66–72


The Bone & Joint Journal
Vol. 100-B, Issue 10 | Pages 1297 - 1302
1 Oct 2018
Elbuluk AM Slover J Anoushiravani AA Schwarzkopf R Eftekhary N Vigdorchik JM

Aims

The routine use of dual-mobility (DM) acetabular components in total hip arthroplasty (THA) may not be cost-effective, but an increasing number of patients undergoing THA have a coexisting spinal disorder, which increases the risk of postoperative instability, and these patients may benefit from DM articulations. This study seeks to examine the cost-effectiveness of DM components as an alternative to standard articulations in these patients.

Patients and Methods

A decision analysis model was used to evaluate the cost-effectiveness of using DM components in patients who would be at high risk for dislocation within one year of THA. Direct and indirect costs of dislocation, incremental costs of using DM components, quality-adjusted life-year (QALY) values, and the probabilities of dislocation were derived from published data. The incremental cost-effectiveness ratio (ICER) was established with a willingness-to-pay threshold of $100 000/QALY. Sensitivity analysis was used to examine the impact of variation.


The Bone & Joint Journal
Vol. 101-B, Issue 1_Supple_A | Pages 41 - 45
1 Jan 2019
Jones CW De Martino I D’Apolito R Nocon AA Sculco PK Sculco TP

Aims. Instability continues to be a troublesome complication after total hip arthroplasty (THA). Patient-related risk factors associated with a higher dislocation risk include the preoperative diagnosis, an age of 75 years or older, high body mass index (BMI), a history of alcohol abuse, and neurodegenerative diseases. The goal of this study was to assess the dislocation rate, radiographic outcomes, and complications of patients stratified as high-risk for dislocation who received a dual mobility (DM) bearing in a primary THA at a minimum follow-up of two years. Materials and Methods. We performed a retrospective review of a consecutive series of DM THA performed between 2010 and 2014 at our institution (Hospital for Special Surgery, New York, New York) by a single, high-volume orthopaedic surgeon employing a single prosthesis design (Anatomic Dual Mobility (ADM) Stryker, Mahwah, New Jersey). Patient medical records and radiographs were reviewed to confirm the type of implant used, to identify any preoperative risk factors for dislocation, and any complications. Radiographic analysis was performed to assess for signs of osteolysis or remodelling of the acetabulum. Results. There were 151 patients who met the classification of high-risk according to the inclusion criteria and received DM THA during the study period. Mean age was 82 years old (73 to 95) and 114 patients (77.5%) were female. Mean follow-up was 3.6 years (1.9 to 6.1), with five patients lost to follow-up and one patient who died (for a reason unrelated to the index procedure). One patient (0.66%) sustained an intraprosthetic dislocation; there were no other dislocations. Conclusion. At mid-term follow-up, the use of a DM bearing for primary THA in patients at high risk of dislocation provided a stable reconstruction option with excellent radiographic results. Longer follow-up is needed to confirm the durability of these reconstructions


The Bone & Joint Journal
Vol. 103-B, Issue 7 Supple B | Pages 17 - 24
1 Jul 2021
Vigdorchik JM Sharma AK Buckland AJ Elbuluk AM Eftekhary N Mayman DJ Carroll KM Jerabek SA

Aims. Patients with spinal pathology who undergo total hip arthroplasty (THA) have an increased risk of dislocation and revision. The aim of this study was to determine if the use of the Hip-Spine Classification system in these patients would result in a decreased rate of postoperative dislocation in patients with spinal pathology. Methods. This prospective, multicentre study evaluated 3,777 consecutive patients undergoing THA by three surgeons, between January 2014 and December 2019. They were categorized using The Hip-Spine Classification system: group 1 with normal spinal alignment; group 2 with a flatback deformity, group 2A with normal spinal mobility, and group 2B with a stiff spine. Flatback deformity was defined by a pelvic incidence minus lumbar lordosis of > 10°, and spinal stiffness was defined by < 10° change in sacral slope from standing to seated. Each category determined a patient-specific component positioning. Survivorship free of dislocation was recorded and spinopelvic measurements were compared for reliability using intraclass correlation coefficient. Results. A total of 2,081 patients met the inclusion criteria. There were 987 group 1A, 232 group 1B, 715 group 2A, and 147 group 2B patients. A total of 70 patients had a lumbar fusion, most had L4-5 (16; 23%) or L4-S1 (12; 17%) fusions; 51 patients (73%) had one or two levels fused, and 19 (27%) had > three levels fused. Dual mobility (DM) components were used in 166 patients (8%), including all of those in group 2B and with > three level fusions. Survivorship free of dislocation at five years was 99.2% with a 0.8% dislocation rate. The correlation coefficient was 0.83 (95% confidence interval 0.89 to 0.91). Conclusion. This is the largest series in the literature evaluating the relationship between hip-spine pathology and dislocation after THA, and guiding appropriate treatment. The Hip-Spine Classification system allows surgeons to make appropriate evaluations preoperatively, and it guides the use of DM components in patients with spinopelvic pathology in order to reduce the risk of dislocation in these high-risk patients. Cite this article: Bone Joint J 2021;103-B(7 Supple B):17–24


The Bone & Joint Journal
Vol. 103-B, Issue 7 Supple B | Pages 73 - 77
1 Jul 2021
Lawrie CM Barrack RL Nunley RM

Aims. Dual mobility (DM) implants have been shown to reduce the dislocation rate after total hip arthroplasty (THA), but there remain concerns about the use of cobalt chrome liners inserted into titanium shells. The aim of this study was to assess the clinical outcomes, metal ion levels, and periprosthetic femoral bone mineral density (BMD) at mid-term follow-up in young, active patients receiving a modular DM THA. Methods. This was a prospective study involving patients aged < 65 years, with a BMI of < 35 kg/m. 2. , and University of California, Los Angeles activity score of > 6 who underwent primary THA with a modular cobalt chrome acetabular liner, highly cross-linked polyethylene mobile bearing, and a cementless titanium femoral stem. Patient-reported outcome measures, whole blood metal ion levels (μg/l), and periprosthetic femoral BMD were measured at baseline and at one, two, and five years postoperatively. The results two years postoperatively for this cohort have been previously reported. Results. A total of 43 patients were enrolled. At minimum follow-up of five years, 23 (53.4%) returned for clinical and radiological review, 25 (58.1%) had metal ion analysis performed, 19 (44.2%) underwent dual energy x-ray absorptiometry scans, and 25 (58%) completed a pain-drawing questionnaire. The mean modified Harris Hip Scores improved significantly from 54.8 (SD 19) preoperatively to 93.08 (SD 10.5) five years postoperatively (p < 0.001). One patient was revised for aseptic acetabular loosening. The mean cobalt levels increased from 0.065 μg/l (SD 0.03) to 0.08 (SD 0.05) and the mean titanium levels increased from 0.35 (SD 0.13) to 0.78 (SD 0.29). The femoral BMD ratio decreased in Gruen Zone 1 (91.9%) at five years postoperatively compared with the baseline scores at six weeks potoperatively. The femoral BMD ratio was maintained in Gruen zones 2 to 7. Conclusion. The use of a modular DM component and a cementless, tapered femoral stem shows excellent mid-term survivorship with minimal concerns for corrosion and metal ion release in a cohort of young, active patients undergoing primary THA. Cite this article: Bone Joint J 2021;103-B(7 Supple B):73–77


The Bone & Joint Journal
Vol. 102-B, Issue 9 | Pages 1128 - 1135
14 Sep 2020
Khoshbin A Haddad FS Ward S O hEireamhoin S Wu J Nherera L Atrey A

Aims. The rate of dislocation when traditional single bearing implants are used in revision total hip arthroplasty (THA) has been reported to be between 8% and 10%. The use of dual mobility bearings can reduce this risk to between 0.5% and 2%. Dual mobility bearings are more expensive, and it is not clear if the additional clinical benefits constitute value for money for the payers. We aimed to estimate the cost-effectiveness of dual mobility compared with single bearings for patients undergoing revision THA. Methods. We developed a Markov model to estimate the expected cost and benefits of dual mobility compared with single bearing implants in patients undergoing revision THA. The rates of revision and further revision were calculated from the National Joint Registry of England and Wales, while rates of transition from one health state to another were estimated from the literature, and the data were stratified by sex and age. Implant and healthcare costs were estimated from local procurement prices and national tariffs. Quality-adjusted life-years (QALYs) were calculated using published utility estimates for patients undergoing THA. Results. At a minimum five-year follow-up, the use of dual mobility was cost-effective with an estimated incremental cost-effectiveness ratio (ICER) of between £3,006 and £18,745/QALY for patients aged < 55 years and between 64 and 75 years, respectively. For those aged > 75 years dual mobility was only cost-effective if the timeline was beyond seven years. The use of dual mobility bearings was cost-saving for patients aged < 75 years and cost-effective for those aged > 75 years if the time horizon was beyond ten years. Conclusion. The use of dual mobility bearings is cost-effective compared with single bearings in patients undergoing revision THA. The younger the patient is, the more likely it is that a dual mobility bearing can be more cost-effective and even cost-saving. The results are affected by the time horizon and cost of bearings for those aged > 75 years. For patients aged > 75 years, the surgeon must decide whether the use of a dual mobility bearing is a viable economic and clinical option. Cite this article: Bone Joint J 2020;102-B(9):1128–1135


The Bone & Joint Journal
Vol. 104-B, Issue 5 | Pages 598 - 603
1 May 2022
Siljander MP Gausden EB Wooster BM Karczewski D Sierra RJ Trousdale RT Abdel MP

Aims

The aim of this study was to evaluate the incidence of liner malseating in two commonly used dual-mobility (DM) designs. Secondary aims included determining the risk of dislocation, survival, and clinical outcomes.

Methods

We retrospectively identified 256 primary total hip arthroplasties (THAs) that included a DM component (144 Stryker MDM and 112 Zimmer-Biomet G7) in 233 patients, performed between January 2012 and December 2019. Postoperative radiographs were reviewed independently for malseating of the liner by five reviewers. The mean age of the patients at the time of THA was 66 years (18 to 93), 166 (65%) were female, and the mean BMI was 30 kg/m2 (17 to 57). The mean follow-up was 3.5 years (2.0 to 9.2).


The Bone & Joint Journal
Vol. 103-B, Issue 11 | Pages 1678 - 1685
1 Nov 2021
Abdelaziz H Schröder M Shum Tien C Ibrahim K Gehrke T Salber J Citak M

Aims. One-stage revision hip arthroplasty for periprosthetic joint infection (PJI) has several advantages; however, resection of the proximal femur might be necessary to achieve higher success rates. We investigated the risk factors for resection and re-revisions, and assessed complications and subsequent re-revisions. Methods. In this single-centre, case-control study, 57 patients who underwent one-stage revision arthroplasty for PJI of the hip and required resection of the proximal femur between 2009 and 2018 were identified. The control group consisted of 57 patients undergoing one-stage revision without bony resection. Logistic regression analysis was performed to identify any correlation with resection and the risk factors for re-revisions. Rates of all-causes re-revision, reinfection, and instability were compared between groups. Results. Patients who required resection of the proximal femur were found to have a higher all-cause re-revision rate (29.8% vs 10.5%; p = 0.018), largely due to reinfection (15.8% vs 0%; p = 0.003), and dislocation (8.8% vs 10.5%; p = 0.762), and showed higher rate of in-hospital wound haematoma requiring aspiration or evacuation (p = 0.013), and wound revision (p = 0.008). The use of of dual mobility components/constrained liner in the resection group was higher than that of controls (94.7% vs 36.8%; p < 0.001). The presence and removal of additional metal hardware (odds ratio (OR) = 7.2), a sinus tract (OR 4), ten years’ time interval between primary implantation and index infection (OR 3.3), and previous hip revision (OR 1.4) increased the risk of proximal femoral resection. A sinus tract (OR 9.2) and postoperative dislocation (OR 281.4) were associated with increased risk of subsequent re-revisions. Conclusion. Proximal femoral resection during one-stage revision hip arthroplasty for PJI may be required to reduce the risk of of recurrent or further infection. Patients with additional metalware needing removal or transcortical sinus tracts and chronic osteomyelitis are particularly at higher risk of needing proximal femoral excision. However, radical resection is associated with higher surgical complications and increased re-revision rates. The use of constrained acetabular liners and dual mobility components maintained an acceptable dislocation rate. These results, including identified risk factors, may aid in preoperative planning, patient consultation and consent, and intraoperative decision-making. Cite this article: Bone Joint J 2021;103-B(11):1678–1685


The Bone & Joint Journal
Vol. 100-B, Issue 6 | Pages 720 - 724
1 Jun 2018
Waterson HB Whitehouse MR Greidanus NV Garbuz DS Masri BA Duncan CP

Aims. Fretting and corrosion at the modular head/neck junction, known as trunnionosis, in total hip arthroplasty (THA) is a cause of adverse reaction to metal debris (ARMD). We describe the outcome of revision of metal-on-polyethylene (MoP) THA for ARMD due to trunnionosis with emphasis on the risk of major complications. Patients and Methods. A total of 36 patients with a MoP THA who underwent revision for ARMD due to trunnionosis were identified. Three were excluded as their revision had been to another metal head. The remaining 33 were revised to a ceramic head with a titanium sleeve. We describe the presentation, revision findings, and risk of complications in these patients. Results. The patients presented with pain, swelling, stiffness, or instability and an inflammatory mass was confirmed radiologically. Macroscopic material deposition on the trunnion was seen in all patients, associated with ARMD. Following revision, six (18.2%) dislocated, requiring further revision in four. Three (9.1%) developed a deep infection and six (18.2%) had significant persistent pain without an obvious cause. One developed a femoral artery thrombosis after excision of an iliofemoral pseudotumor, requiring a thrombectomy. Conclusion. The risk of serious complications following revision MoP THA for ARMD associated with trunnionosis is high. In the presence of extensive tissue damage, a constrained liner or dual mobility construct is recommended in these patients. Cite this article: Bone Joint J 2018;100-B:720–4


The Bone & Joint Journal
Vol. 100-B, Issue 7 | Pages 898 - 902
1 Jul 2018
Lachiewicz PF O’Dell JA

Aims. To report our experience with trunnion corrosion following metal-on-polyethylene total hip arthroplasty, in particular to report the spectrum of presentation and determine the mean time to presentation. Patients and Methods. We report the presenting symptoms and signs, intraoperative findings, and early results and complications of operative treatment in nine patients with a mean age of 74 years (60 to 86). The onset of symptoms was at a mean of seven years (3 to 18) after index surgery. Results. Patients presented with a variety of symptoms including pain, limp and rash. The preoperative mean serum cobalt level was 7.1 ppb (2.2 to 12.8) and mean serum chromium level was 2.2 ppb (0.5 to 5.2). Metal artifact reduction sequence (MARS) MRI showed fluid collection and possible pseudotumour formation in five hips, fluid collection in two hips, and synovitis/debris in one hip, with no MRI in one patient. Acetabular revision was performed in three patients, six patients underwent liner and head exchange only. The postoperative metal levels decreased in all patients: mean cobalt 0.5 ppb (0 to 1.8) and mean chromium 0.9 ppb (0 to 2.6) at a mean of five months (3 to 8) postoperatively. Seven patients had good pain relief and no complications at one year. There were two major complications requiring reoperation: acute infection at six weeks, for which the patient required two-stage reimplantation; and recurrent dislocation, for which the patient was revised to a dual mobility component. Conclusion. Trunnion corrosion in metal-on-polyethylene THA has a range of presenting symptoms, and may present later than previously described. A high index of suspicion is warranted, and serum cobalt and chromium levels are recommended for diagnosis. Patients should be counselled about possible postoperative complications. Cite this article: Bone Joint J 2018;100-B:898–902


The Bone & Joint Journal
Vol. 103-B, Issue 7 Supple B | Pages 38 - 45
1 Jul 2021
Horberg JV Coobs BR Jiwanlal AK Betzle CJ Capps SG Moskal JT

Aims

Use of the direct anterior approach (DAA) for total hip arthroplasty (THA) has increased in recent years due to proposed benefits, including a lower risk of dislocation and improved early functional recovery. This study investigates the dislocation rate in a non-selective, consecutive cohort undergoing THA via the DAA without any exclusion or bias in patient selection based on habitus, deformity, age, sex, or fixation method.

Methods

We retrospectively reviewed all patients undergoing THA via the DAA between 2011 and 2017 at our institution. Primary outcome was dislocation at minimum two-year follow-up. Patients were stratified by demographic details and risk factors for dislocation, and an in-depth analysis of dislocations was performed.


The Bone & Joint Journal
Vol. 103-B, Issue 7 Supple B | Pages 129 - 134
1 Jul 2021
Ayekoloye CI Abu Qa'oud M Radi M Leon SA Kuzyk P Safir O Gross AE

Aims

Improvements in functional results and long-term survival are variable following conversion of hip fusion to total hip arthroplasty (THA) and complications are high. The aim of the study was to analyze the clinical and functional results in patients who underwent conversion of hip fusion to THA using a consistent technique and uncemented implants.

Methods

A total of 39 hip fusion conversions to THA were undertaken in 38 patients by a single surgeon employing a consistent surgical technique and uncemented implants. Parameters assessed included Harris Hip Score (HHS) for function, range of motion (ROM), leg length discrepancy (LLD), satisfaction, and use of walking aid. Radiographs were reviewed for loosening, subsidence, and heterotopic ossification (HO). Postoperative complications and implant survival were assessed.


The Bone & Joint Journal
Vol. 103-B, Issue 12 | Pages 1766 - 1773
1 Dec 2021
Sculco PK Windsor EN Jerabek SA Mayman DJ Elbuluk A Buckland AJ Vigdorchik JM

Aims

Spinopelvic mobility plays an important role in functional acetabular component position following total hip arthroplasty (THA). The primary aim of this study was to determine if spinopelvic hypermobility persists or resolves following THA. Our second aim was to identify patient demographic or radiological factors associated with hypermobility and resolution of hypermobility after THA.

Methods

This study investigated patients with preoperative posterior hypermobility, defined as a change in sacral slope (SS) from standing to sitting (ΔSSstand-sit) ≥ 30°. Radiological spinopelvic parameters, including SS, pelvic incidence (PI), lumbar lordosis (LL), PI-LL mismatch, anterior pelvic plane tilt (APPt), and spinopelvic tilt (SPT), were measured on preoperative imaging, and at six weeks and a minimum of one year postoperatively. The severity of bilateral hip osteoarthritis (OA) was graded using Kellgren-Lawrence criteria.