We have developed a
In a prospective randomised clinical study acetabular components were implanted either freehand (n = 30) or using CT-based (n = 30) or imageless navigation (n = 30). The position of the component was determined post-operatively on CT scans of the pelvis. Following conventional freehand placement of the acetabular component, only 14 of the 30 were within the safe zone as defined by Lewinnek et al (40° inclination . sd. 10°; 15° anteversion . sd. 10°). After computer-assisted navigation 25 of 30 acetabular components (CT-based) and 28 of 30 components (imageless) were positioned within this limit (overall p <
0.001). No significant differences were observed between CT-based and imageless navigation (p = 0.23); both showed a significant reduction in variation of the position of the acetabular component compared with conventional freehand arthroplasty (p <
0.001). The duration of the operation was increased by eight minutes with imageless and by 17 minutes with
We analyzed the acetabular morphology of Crowe type IV hips using CT data to identify a landmark for the ideal placement of the centre of the acetabular component, as assessed by morphometric geometrical analysis, and its reliability. A total of 52 Crowe IV hips (42 patients; seven male, 35 female; mean age 68.5 years (32 to 82)) and 50 normal hips (50 patients; eight male, 42 female; mean age 60.7 years (34 to 86)) undergoing total hip arthroplasty were retrospectively identified. In this CT-based simulation study, the acetabular component was positioned at the true acetabulum with a radiological inclination of 40° and anteversion of 20°. Acetabular shape and the position of the centre of the acetabular component were analyzed by morphometric geometrical analysis using the generalized Procrustes analysis.Aims
Patients and Methods
The aim of this study was to evaluate the accuracy of implant placement when using robotic assistance during total hip arthroplasty (THA). A total of 20 patients underwent a planned THA using preoperative CT scans and robotic-assisted software. There were nine men and 11 women (n = 20 hips) with a mean age of 60.8 years (Aims
Patients and Methods
The computed neck-shaft angle and the size of the femoral component were recorded in 100 consecutive hip resurfacings using imageless computer-navigation and compared with the angle measured before operation and with actual component implanted. The reliability of the registration was further analysed using ten cadaver femora. The mean absolute difference between the measured and navigated neck-shaft angle was 16.3° (0° to 52°). Navigation underestimated the measured neck-shaft angle in 38 patients and the correct implant size in 11. Registration of the cadaver femora tended to overestimate the correct implant size and provided a low level of repeatability in computing the neck-shaft angle. Prudent pre-operative planning is advisable for use in conjunction with imageless navigation since misleading information may be registered intraoperatively, which could lead to inappropriate sizing and positioning of the femoral component in hip resurfacing.
We have investigated the accuracy of placement of the femoral component using imageless navigation in 100 consecutive Birmingham Hip Resurfacings. Pre-operative templating determined the native neck-shaft angle and planned stem-shaft angle of the implant. The latter were verified post-operatively using digital anteroposterior unilateral radiographs of the hip. The mean neck-shaft angle determined before operation was 132.7° (118° to 160°). The mean planned stem-shaft angle was a relative valgus alignment of 9.7° ( Navigation in hip resurfacing may afford the surgeon a reliable and accurate method of placement of the femoral component.
The angle of inclination of the acetabular component in total hip replacement is a recognised contributing factor in dislocation and early wear. During non-navigated surgery, insertion of the acetabular component has traditionally been performed at an angle of 45° relative to the sagittal plane as judged by the surgeon’s eye, the operative inclination. Typically, the method used to assess inclination is the measurement made on the postoperative anteroposterior radiograph, the radiological inclination. The aim of this study was to measure the intra-operative angle of inclination of the acetabular component on 60 consecutive patients in the lateral decubitus position when using a posterior approach during total hip replacement. This was achieved by taking intra-operative photographs of the acetabular inserter, representing the acetabular axis, and a horizontal reference. The results were compared with the post-operative radiological inclination. The mean post-operative radiological inclination was 13° greater than the photographed operative inclination, which was unexpectedly high. It appears that in the lateral decubitus position with a posterior approach, the uppermost hemipelvis adducts, thus reducing the apparent operative inclination. Surgeons using the posterior approach in lateral decubitus need to aim for a lower operative inclination than when operating with the patient supine in order to achieve an acceptable radiological inclination.
New brands of joint prosthesis are released for general implantation with limited evidence of their long-term performance in patients. The CUSUM continuous monitoring method is a statistical testing procedure which could be used to provide prospective evaluation of brands as soon as implantation in patients begins and give early warning of poor performance. We describe the CUSUM and illustrate the potential value of this monitoring tool by applying it retrospectively to the 3M Capital Hip experience. The results show that if the clinical data and methodology had been available, the CUSUM would have given an alert to the underperformance of this prosthesis almost four years before the issue of a Hazard Notice by the Medical Devices Agency. This indicates that the CUSUM can be a valuable tool in monitoring joint prostheses, subject to timely and complete collection of data. Regional or national joint registries provide an opportunity for future centralised, continuous monitoring of all hip and knee prostheses using these techniques.