Internal lengthening devices in the femur lengthen
along the anatomical axis, potentially creating lateral shift of
the mechanical axis. We aimed to determine whether femoral lengthening
along the anatomical axis has an inadvertent effect on lower limb
alignment. Isolated femoral lengthening using the Intramedullary
Skeletal Kinetic Distractor was performed in 27 femora in 24 patients
(mean age 32 years (16 to 57)). Patients who underwent simultaneous realignment
procedures or concurrent tibial lengthening, or who developed mal-
or nonunion, were excluded. Pre-operative and six-month post-operative
radiographs were used to measure lower limb alignment. The mean lengthening
achieved was 4.4 cm (1.5 to 8.0). In 26 of 27 limbs, the mechanical
axis shifted laterally by a mean of 1.0 mm/cm of lengthening (0
to 3.5). In one femur that was initially in varus, a 3 mm medial
shift occurred during a lengthening of 2.2 cm. In a normally aligned limb, intramedullary lengthening along
the anatomical axis of the femur results in a lateral shift of the
mechanical axis by approximately 1 mm for each 1 cm of lengthening.
We performed nine metatarsal and three proximal phalangeal lengthenings in five patients with congenital brachymetatarsia of the first and one or two other metatarsal bones, by a one-stage combined shortening and lengthening procedure using intercalcary autogenous bone grafts from adjacent shortened metatarsal bones. Instead of the isolated lengthening of the first and the other metatarsal bones, we shortened the adjacent normal metatarsal and used the excised
We report the results of intramedullary leg lengthening conducted between 2002 and 2009 using the Intramedullary Skeletal Kinetic Distractor in 69 unilateral lengthenings involving 58 femora and 11 tibiae. We identified difficulties that occurred during the treatment and assessed whether they were specifically due to the implant or independent of it. Paley’s classification for evaluating problems, obstacles and complications with external fixators was adopted, and implant-specific difficulties were continuously noted. There were seven failures requiring premature removal of the device, in four due to nail breakage and three for other reasons, and five unsuccessful outcomes after completion of the lengthening. In all, 116 difficulties were noted in 45 patients, with only 24 having problem-free courses. In addition to the difficulties arising from the use of external fixators, there were almost the same number again of implant-specific difficulties. Nevertheless, successful femoral lengthening was achieved in 52 of the 58 patients (90%). However, successful tibial lengthening was only achieved in five of 11 patients (45%).