Advertisement for orthosearch.org.uk
Results 1 - 20 of 32
Results per page:
The Bone & Joint Journal
Vol. 99-B, Issue 6 | Pages 793 - 798
1 Jun 2017
Anderson FL Koch CN Elpers ME Wright TM Haas SB Heyse TJ

Aims

We sought to establish whether an oxidised zirconium (OxZr) femoral component causes less loss of polyethylene volume than a cobalt alloy (CoCr) femoral component in total knee arthroplasty.

Materials and Methods

A total of 20 retrieved tibial inserts that had articulated with OxZr components were matched with 20 inserts from CoCr articulations for patient age, body mass index, length of implantation, and revision diagnosis. Changes in dimensions of the articular surfaces were compared with those of pristine inserts using laser scanning. The differences in volume between the retrieved and pristine surfaces of the two groups were calculated and compared.


The Bone & Joint Journal
Vol. 106-B, Issue 9 | Pages 916 - 923
1 Sep 2024
Fricka KB Wilson EJ Strait AV Ho H Hopper, Jr RH Hamilton WG Sershon RA

Aims. The optimal bearing surface design for medial unicompartmental knee arthroplasty (UKA) remains controversial. The aim of this study was to compare outcomes of fixed-bearing (FB) and mobile-bearing (MB) UKAs from a single high-volume institution. Methods. Prospectively collected data were reviewed for all primary cemented medial UKAs performed by seven surgeons from January 2006 to December 2022. A total of 2,999 UKAs were identified, including 2,315 FB and 684 MB cases. The primary outcome measure was implant survival. Secondary outcomes included 90-day and cumulative complications, reoperations, component revisions, conversion arthroplasties, range of motion, and patient-reported outcome measures. Overall mean age at surgery was 65.7 years (32.9 to 94.3), 53.1% (1,593/2,999) of UKAs were implanted in female patients, and demographics between groups were similar (p > 0.05). The mean follow-up for all UKAs was 3.7 years (0.0 to 15.6). Results. Using revision for any reason as an endpoint, five-year survival for FB UKAs was 97.2% (95% CI 96.4 to 98.1) compared to 96.0% for MB (95% CI 94.1 to 97.9; p = 0.008). The FB group experienced fewer component revisions (14/2,315, 0.6% vs 12/684, 1.8%; p < 0.001) and conversion arthroplasties (38/2315, 1.6% vs 24/684, 3.5%; p < 0.001). A greater number of MB UKAs underwent revision due to osteoarthritis progression (FB = 21/2,315, 0.9% vs MB = 16/684, 2.3%; p = 0.003). In the MB group, 12 (1.8%) subjects experienced bearing dislocations which required revision surgery. There were 15 early periprosthetic tibia fractures (0.6%) in the FB group compared to 0 for MB (p = 0.035). Conclusion. In similar patient populations, FB UKAs demonstrated slightly higher survival than a commonly used MB design. Adverse event profiles differed by bearing type, with an increased risk of bearing dislocation and OA progression with MB designs, and early periprosthetic tibia fractures for FB designs. Cite this article: Bone Joint J 2024;106-B(9):916–923


The Bone & Joint Journal
Vol. 103-B, Issue 12 | Pages 1791 - 1801
1 Dec 2021
Bhalekar RM Nargol ME Shyam N Nargol AVF Wells SR Collier R Pabbruwe M Joyce TJ Langton DJ

Aims

The aim of this study was to investigate whether wear and backside deformation of polyethylene (PE) tibial inserts may influence the cement cover of tibial trays of explanted total knee arthroplasties (TKAs).

Methods

At our retrieval centre, we measured changes in the wear and deformation of PE inserts using coordinate measuring machines and light microscopy. The amount of cement cover on the backside of tibial trays was quantified as a percentage of the total surface. The study involved data from the explanted fixed-bearing components of four widely used contemporary designs of TKA (Attune, NexGen, Press Fit Condylar (PFC), and Triathlon), revised for any indication, and we compared them with components that used previous generations of PE. Regression modelling was used to identify variables related to the amount of cement cover on the retrieved trays.


The Bone & Joint Journal
Vol. 104-B, Issue 2 | Pages 235 - 241
1 Feb 2022
Stone B Nugent M Young SW Frampton C Hooper GJ

Aims

The success of total knee arthroplasty (TKA) is usually measured using functional outcome scores and revision-free survivorship. However, reporting the lifetime risk of revision may be more meaningful to patients when gauging risks, especially in younger patients. We aimed to assess the lifetime risk of revision for patients in different age categories at the time of undergoing primary TKA.

Methods

The New Zealand Joint Registry database was used to obtain revision rates, mortality, and the indications for revision for all primary TKAs performed during an 18-year period between January 1999 and December 2016. Patients were stratified into age groups at the time of the initial TKA, and the lifetime risk of revision was calculated according to age, sex, and the American Society of Anesthesiologists (ASA) grade. The most common indications for revision were also analyzed for each age group.


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 4 | Pages 470 - 475
1 Apr 2011
Kendrick BJL Simpson DJ Kaptein BL Valstar ER Gill HS Murray DW Price AJ

The Oxford unicompartmental knee replacement (UKR) was designed to minimise wear utilising a fully-congruent, mobile, polyethylene bearing. Wear of polyethylene is a significant cause of revision surgery in UKR in the first decade, and the incidence increases in the second decade. Our study used model-based radiostereometric analysis to measure the combined wear of the upper and lower bearing surfaces in 13 medial-compartment Oxford UKRs at a mean of 20.9 years (17.2 to 25.9) post-operatively. The mean linear penetration of the polyethylene bearing was 1.04 mm (0.307 to 2.15), with a mean annual wear rate of 0.045 mm/year (0.016 to 0.099). The annual wear rate of the phase-2 bearings (mean 0.022 mm/year) was significantly less (p = 0.01) than that of phase-1 bearings (mean 0.07 mm/year). The linear wear rate of the Oxford UKR remains very low into the third decade. We believe that phase-2 bearings had lower wear rates than phase-1 implants because of the improved bearing design and surgical technique which decreased the incidence of impingement. We conclude that the design of the Oxford UKR gives low rates of wear in the long term


The Bone & Joint Journal
Vol. 103-B, Issue 6 Supple A | Pages 94 - 101
1 Jun 2021
Roy ME Whiteside LA Ly KK Gauvain MJ

Aims

The aims of this study were to evaluate wear on the surface of cobalt-chromium (CoCr) femoral components used in total knee arthroplasty (TKA) and compare the wear of these components with that of ceramic femoral components.

Methods

Optical profilometry was used to evaluate surface roughness and to examine the features created by the wear process in a knee wear simulator. We developed a method of measuring surface changes on five CoCr femoral components and quantifying the loss of material from the articular surface during the wear process. We also examined the articular surface of three ceramic femoral components from a previous test for evidence of surface damage, and compared it with that of CoCr components.


The Journal of Bone & Joint Surgery British Volume
Vol. 85-B, Issue 3 | Pages 393 - 398
1 Apr 2003
Siddique MS Rao MC Deehan DJ Pinder IM

We carried out 60 revision procedures for failed porous coated anatomic total knee replacements in 54 patients, which were divided into two groups. The 14 knees in group I had a well-fixed femoral component at surgery which was retained, and in the 46 knees in group II both tibial and femoral components were loose and were revised using a variety of implants. Our review comprised clinical and radiological assessment. A total of 13 knees required a second revision. Six (42%) in group I failed very early (mean 2.1 years) when compared with seven (15%) in group II (mean 6.8 years). Failure was due to wear of the polyethylene insert by the abraded, retained femoral component (crude odds ratio 4.07; 95% CI 1.07 to 15.5). We recommend a complete change of primary bearing surfaces at the time of revision of an uncemented total knee replacement in order to prevent early wear of polyethylene


The Bone & Joint Journal
Vol. 103-B, Issue 6 Supple A | Pages 145 - 149
1 Jun 2021
Crawford DA Passias BJ Adams JB Berend KR Lombardi AV

Aims

A limited number of investigations with conflicting results have described perivascular lymphocytic infiltration (PVLI) in the setting of total knee arthroplasty (TKA). The purpose of this study was to determine if PVLI found in TKAs at the time of aseptic revision surgery was associated with worse clinical outcomes and survivorship.

Methods

A retrospective review was conducted on 617 patients who underwent aseptic TKA revision who had histological analysis for PVLI at the time of surgery. Clinical and radiological data were obtained pre- and postoperatively, six weeks postoperatively, and then every year thereafter.


The Bone & Joint Journal
Vol. 103-B, Issue 7 | Pages 1261 - 1269
1 Jul 2021
Burger JA Zuiderbaan HA Sierevelt IN van Steenbergen L Nolte PA Pearle AD Kerkhoffs GMMJ

Aims

Uncemented mobile bearing designs in medial unicompartmental knee arthroplasty (UKA) have seen an increase over the last decade. However, there are a lack of large-scale studies comparing survivorship of these specific designs to commonly used cemented mobile and fixed bearing designs. The aim of this study was to evaluate the survivorship of these designs.

Methods

A total of 21,610 medial UKAs from 2007 to 2018 were selected from the Dutch Arthroplasty Register. Multivariate Cox regression analyses were used to compare uncemented mobile bearings with cemented mobile and fixed bearings. Adjustments were made for patient and surgical factors, with their interactions being considered. Reasons and type of revision in the first two years after surgery were assessed.


The Bone & Joint Journal
Vol. 103-B, Issue 6 Supple A | Pages 59 - 66
1 Jun 2021
Abhari S Hsing TM Malkani MM Smith AF Smith LS Mont MA Malkani AL

Aims

Alternative alignment concepts, including kinematic and restricted kinematic, have been introduced to help improve clinical outcomes following total knee arthroplasty (TKA). The purpose of this study was to evaluate the clinical results, along with patient satisfaction, following TKA using the concept of restricted kinematic alignment.

Methods

A total of 121 consecutive TKAs performed between 11 February 2018 to 11 June 2019 with preoperative varus deformity were reviewed at minimum one-year follow-up. Three knees were excluded due to severe preoperative varus deformity greater than 15°, and a further three due to requiring revision surgery, leaving 109 patients and 115 knees to undergo primary TKA using the concept of restricted kinematic alignment with advanced technology. Patients were stratified into three groups based on the preoperative limb varus deformity: Group A with 1° to 5° varus (43 knees); Group B between 6° and 10° varus (56 knees); and Group C with varus greater than 10° (16 knees). This study group was compared with a matched cohort of 115 TKAs and 115 patients using a neutral mechanical alignment target with manual instruments performed from 24 October 2016 to 14 January 2019.


The Bone & Joint Journal
Vol. 103-B, Issue 1 | Pages 113 - 122
1 Jan 2021
Kayani B Tahmassebi J Ayuob A Konan S Oussedik S Haddad FS

Aims

The primary aim of this study was to compare the postoperative systemic inflammatory response in conventional jig-based total knee arthroplasty (conventional TKA) versus robotic-arm assisted total knee arthroplasty (robotic TKA). Secondary aims were to compare the macroscopic soft tissue injury, femoral and tibial bone trauma, localized thermal response, and the accuracy of component positioning between the two treatment groups.

Methods

This prospective randomized controlled trial included 30 patients with osteoarthritis of the knee undergoing conventional TKA versus robotic TKA. Predefined serum markers of inflammation and localized knee temperature were collected preoperatively and postoperatively at six hours, day 1, day 2, day 7, and day 28 following TKA. Blinded observers used the Macroscopic Soft Tissue Injury (MASTI) classification system to grade intraoperative periarticular soft tissue injury and bone trauma. Plain radiographs were used to assess the accuracy of achieving the planned postioning of the components in both groups.


The Bone & Joint Journal
Vol. 102-B, Issue 10 | Pages 1331 - 1340
3 Oct 2020
Attard V Li CY Self A Mann DA Borthwick LA O’Connor P Deehan DJ Kalson NS

Aims

Stiffness is a common complication after total knee arthroplasty (TKA). Pathogenesis is not understood, treatment options are limited, and diagnosis is challenging. The aim of this study was to investigate if MRI can be used to visualize intra-articular scarring in patients with stiff, painful knee arthroplasties.

Methods

Well-functioning primary TKAs (n = 11), failed non-fibrotic TKAs (n = 5), and patients with a clinical diagnosis of fibrosis1 (n = 8) underwent an MRI scan with advanced metal suppression (Slice Encoding for Metal Artefact Correction, SEMAC) with gadolinium contrast. Fibrotic tissue (low intensity on T1 and T2, low-moderate post-contrast enhancement) was quantified (presence and tissue thickness) in six compartments: supra/infrapatella, medial/lateral gutters, and posterior medial/lateral.


The Bone & Joint Journal
Vol. 98-B, Issue 1 | Pages 58 - 64
1 Jan 2016
Ahmed I Salmon LJ Waller A Watanabe H Roe JP Pinczewski LA

Aims

Oxidised zirconium was introduced as a material for femoral components in total knee arthroplasty (TKA) as an attempt to reduce polyethylene wear. However, the long-term survival of this component is not known.

Methods

We performed a retrospective review of a prospectively collected database to assess the ten year survival and clinical and radiological outcomes of an oxidised zirconium total knee arthroplasty with the Genesis II prosthesis.

The Western Ontario and McMaster Universities Osteoarthritis Index (WOMAC), Knee Injury and Osteoarthritis Outcome Score (KOOS) and a patient satisfaction scale were used to assess outcome.


The Bone & Joint Journal
Vol. 98-B, Issue 5 | Pages 616 - 621
1 May 2016
Teeter MG Thoren J Yuan X McCalden RW MacDonald SJ Lanting BA Naudie DDR

Aims

The purpose of the present study was to examine the long-term fixation of a cemented fixed-bearing polished titanium tibial baseplate (Genesis ll).

Patients and Methods

Patients enrolled in a previous two-year prospective trial (n = 35) were recalled at ten years. Available patients (n = 15) underwent radiostereometric analysis (RSA) imaging in a supine position using a conventional RSA protocol. Migration of the tibial component in all planes was compared between initial and ten-year follow-up. Outcome scores including the Knee Society Score, Western Ontario and McMaster Universities Arthritis Index, 12-item Short Form Health Survey, Forgotten Joint Score, and University of California, Los Angeles Activity Score were recorded.


The Bone & Joint Journal
Vol. 98-B, Issue 1_Supple_A | Pages 105 - 112
1 Jan 2016
Khan M Osman K Green G Haddad FS

Total knee arthroplasty (TKA) is a cost effective and extremely successful operation. As longevity increases, the demand for primary TKA will continue to rise. The success and survivorship of TKAs are dependent on the demographics of the patient, surgical technique and implant-related factors.

Currently the risk of failure of a TKA requiring revision surgery ten years post-operatively is 5%.

The most common indications for revision include aseptic loosening (29.8%), infection (14.8%), and pain (9.5%). Revision surgery poses considerable clinical burdens on patients and financial burdens on healthcare systems.

We present a current concepts review on the epidemiology of failed TKAs using data from worldwide National Joint Registries.

Cite this article: Bone Joint J 2016;98-B(1 Suppl A):105–12.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 10 | Pages 1366 - 1371
1 Oct 2012
Pijls BG Valstar ER Kaptein BL Nelissen RGHH

This single-blinded randomised controlled trial investigated whether one design of mobile-bearing (MB) total knee replacement (TKR) has any advantage over a fixed-bearing (FB) design on long-term fixation as measured by radiostereometry. The amount of wear underneath the mobile bearing was also evaluated. A series of 42 knees was randomised to MB or FB tibial components with appropriate polyethylene inserts and followed for between ten and 12 years, or until the death of the patient. The polyethylene in the MB group was superior in that it was gamma-irradiated in inert gas and was calcium-stearate free; the polyethylene in the FB group was gamma-irradiated in air and contained calcium stearate. In theory this should be advantageous to the wear rate of the MB group. At final follow-up the overall mean migration was 0.75 mm (sd 0.76) in the MB group and 0.66 mm (sd 0.4) in the FB group, with the FB group demonstrating more posterior tilt and the MB group more internal rotation. In the FB group there was one revision for aseptic loosening, but none in the MB group. There were no significant differences in clinical or radiological scores.

For the MB group, the mean linear wear rate on the under-surface was 0.026 mm/year (sd 0.014). This was significantly smaller than the wear rate of 0.11 mm/year (sd 0.06) in the MB between femur and polyethylene (p < 0.001). Nevertheless, even in a best-case setting the mobile bearings of this TKR design had no apparent advantage in terms of fixation over the FB knee prosthesis at ten to 12 years. The wear underneath the mobile bearing was small and is unlikely to be clinically relevant.


The Bone & Joint Journal
Vol. 96-B, Issue 2 | Pages 217 - 223
1 Feb 2014
Namba RS Inacio MCS Cafri G

The outcome of total knee replacement (TKR) using components designed to increase the range of flexion is not fully understood. The short- to mid-term risk of aseptic revision in high flexion TKR was evaluated. The endpoint of the study was aseptic revision and the following variables were investigated: implant design (high flexion vs non-high flexion), the thickness of the tibial insert (≤ 14 mm vs > 14 mm), cruciate ligament (posterior stabilised (PS) vs cruciate retaining), mobility (fixed vs rotating), and the manufacturer (Zimmer, Smith & Nephew and DePuy). Covariates included patient, implant, surgeon and hospital factors. Marginal Cox proportional hazard models were used.

In a cohort of 64 000 TKRs, high flexion components were used in 8035 (12.5%). The high flexion knees with tibial liners of thickness > 14 mm had a density of revision of 1.45/100 years of observation, compared with 0.37/100 in non-high flexion TKR with liners ≤ 14 mm thick. Relative to a standard fixed PS TKR, the NexGen (Zimmer, Warsaw, Indiana) Gender Specific Female high flexion fixed PS TKR had an increased risk of revision (hazard ratio (HR) 2.27 (95% confidence interval (CI) 1.48 to 3.50)), an effect that was magnified when a thicker tibial insert was used (HR 8.10 (95% CI 4.41 to 14.89)).

Surgeons should be cautious when choosing high flexion TKRs, particularly when thicker tibial liners might be required.

Cite this article: Bone Joint J 2014;96-B:217–23.


The Bone & Joint Journal
Vol. 95-B, Issue 3 | Pages 301 - 304
1 Mar 2013
Brennan SA Devitt BM O’Neill CJ Nicholson P

Focal femoral inlay resurfacing has been developed for the treatment of full-thickness chondral defects of the knee. This technique involves implanting a defect-sized metallic or ceramic cap that is anchored to the subchondral bone through a screw or pin. The use of these experimental caps has been advocated in middle-aged patients who have failed non-operative methods or biological repair techniques and are deemed unsuitable for conventional arthroplasty because of their age. This paper outlines the implant design, surgical technique and biomechanical principles underlying their use. Outcomes following implantation in both animal and human studies are also reviewed.

Cite this article: Bone Joint J 2013;95-B:301–4.


The Bone & Joint Journal
Vol. 96-B, Issue 10 | Pages 1333 - 1338
1 Oct 2014
Gustke KA Golladay GJ Roche MW Jerry GJ Elson LC Anderson CR

The aim of this prospective multicentre study was to report the patient satisfaction after total knee replacement (TKR), undertaken with the aid of intra-operative sensors, and to compare these results with previous studies. A total of 135 patients undergoing TKR were included in the study. The soft-tissue balance of each TKR was quantified intra-operatively by the sensor, and 18 (13%) were found to be unbalanced. A total of 113 patients (96.7%) in the balanced group and 15 (82.1%) in the unbalanced group were satisfied or very satisfied one year post-operatively (p = 0.043).

A review of the literature identified no previous study with a mean level of satisfaction that was greater than the reported level of satisfaction of the balanced TKR group in this study. Ensuring soft-tissue balance by using intra-operative sensors during TKR may improve satisfaction.

Cite this article: Bone Joint J 2014;96-B:1333–8


The Bone & Joint Journal
Vol. 96-B, Issue 12 | Pages 1657 - 1662
1 Dec 2014
Stambough JB Clohisy JC Barrack RL Nunley RM Keeney JA

The aims of this retrospective study were to compare the mid-term outcomes following revision total knee replacement (TKR) in 76 patients (81 knees) < 55 years of age with those of a matched group of primary TKRs based on age, BMI, gender and comorbid conditions. We report the activity levels, functional scores, rates of revision and complications. Compared with patients undergoing primary TKR, those undergoing revision TKR had less improvement in the mean Knee Society function scores (8.14 (–55 to +60) vs 23.3 points (–40 to +80), p < 0.001), a similar improvement in UCLA activity level (p = 0.52), and similar minor complication rates (16% vs 13%, p = 0.83) at a mean follow-up of 4.6 years (2 to 13.4). Further revision surgery was more common among revised TKRs (17% vs 5%, p = 0.02), with deep infection and instability being the most common reasons for failure. As many as one-third of patients aged < 55 years in the revision group had a complication or failure requiring further surgery.

Young patients undergoing revision TKR should be counselled that they can expect somewhat less improvement and a higher risk of complications than occur after primary TKR.

Cite this article: Bone Joint J 2014; 96-B:1657–62