Advertisement for orthosearch.org.uk
Results 1 - 15 of 15
Results per page:
The Journal of Bone & Joint Surgery British Volume
Vol. 86-B, Issue 6 | Pages 856 - 860
1 Aug 2004
Rahme H Mattsson P Larsson S

We studied the stability of cemented all-polyethylene keeled glenoid components by radiostereometric analysis (RSA) in 16 shoulders which had received a total shoulder replacement. There were 14 women (one bilateral) and one man with a mean age of 64 years. The diagnosis was osteoarthritis in eight and rheumatoid arthritis in seven. Two of the shoulders were excluded from the RSA study because of loosening of the tantalum markers. Three tantalum markers were inserted in the glenoid socket, two in the coracoid process and two in the acromion. The polyethylene keeled glenoid component was marked with three to five tantalum markers. Conventional radiological and RSA examinations were carried out at five to seven days, at four months and at one and two years after operation. Radiolucent lines were found in all except three shoulders. Migration was most pronounced in the distal direction and exceeded 1 mm in four shoulders. In ten shoulders rotation exceeded 2° in one or more axes with retroversion/anteversion being most common. No correlation was found between migration and the presence of radiolucencies on conventional radiographs


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 10 | Pages 1410 - 1415
1 Oct 2010
Castagna A Randelli M Garofalo R Maradei L Giardella A Borroni M

Total shoulder replacement is a successful procedure for degenerative or some inflammatory diseases of the shoulder. However, fixation of the glenoid seems to be the main weakness with a high rate of loosening. The results using all-polyethylene components have been better than those using metal-backed components. We describe our experience with 35 consecutive total shoulder replacements using a new metal-backed glenoid component with a mean follow-up of 75.4 months (48 to 154). Our implant differs from others because of its mechanism of fixation. It has a convex metal-backed bone interface and the main stabilising factor is a large hollow central peg. The patients were evaulated with standard radiographs and with the Constant Score, the Simple Shoulder Test and a visual analogue scale. All the scores improved and there was no loosening, no polyethylene-glenoid disassembly and no other implant-related complications. We conclude that a metal-backed glenoid component is a good option in total shoulder replacement with no worse results than of those using a cemented all-polyethylene prosthesis


The Journal of Bone & Joint Surgery British Volume
Vol. 88-B, Issue 3 | Pages 341 - 344
1 Mar 2006
Tanaka N Sakahashi H Ishii S Kudo H

The purpose of this study was to assess the long-term results (more than ten years) of two types of cemented ulnar component with type-5 Kudo total elbow arthroplasty in a consecutive series of 56 patients (60 elbows) with rheumatoid arthritis, and to compare the results in elbows above and below a Larsen grade IV. There was no radiolucency around the humeral component. Patients in whom a metal-backed ulnar component and a porous-coated stem were used had better clinical results and significantly less progression of radiolucent line formation around the ulnar component. They also had a significantly better long-term survival than patients with an all-polyethylene ulnar component. The clinical results of arthroplasty using all-polyethylene ulnar components were inferior, regardless of the degree of joint destruction. We conclude that the type-5 Kudo total elbow arthroplasty with cementless fixation of the porous-coated humeral component and cemented fixation of a metal-backed ulnar component is acceptable and well-tolerated by rheumatoid patients


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 5 | Pages 656 - 660
1 May 2010
Ikävalko M Tiihonen R Skyttä ET Belt EA

Between 1982 and 1997, 403 consecutive patients (522 elbows) with rheumatoid arthritis underwent Souter-Strathclyde total elbow replacement. By the end of 2007, there had been 66 revisions for aseptic loosening in 60 patients. The mean time of follow-up was 10.6 years (0 to 25) The survival rates at five-, ten, 15 and 19 years were 96% (95%, confidence interval (CI) 95 to 98), 89% (95% CI 86 to 92), 83% (95% CI 78 to 87), and 77% (95% CI 69 to 85), respectively. The small and medium-sized short-stemmed primary humeral components had a 5.6-fold and 3.6-fold risk of revision for aseptic loosening respectively, compared to the medium-sized long-stemmed component. The small and medium-sized all-polyethylene ulnar components had respectively a 28.2-fold and 8.4-fold risk of revision for aseptic loosening, compared to the metal-backed ulnar components. The use of retentive ulnar components was not associated with an increased risk of aseptic loosening compared to non-retentive implants


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 2 | Pages 210 - 216
1 Feb 2011
Young A Walch G Boileau P Favard L Gohlke F Loew M Molé D

We report the long-term clinical and radiological outcomes of the Aequalis total shoulder replacement with a cemented all-polyethylene flat-back keeled glenoid component implanted for primary osteoarthritis between 1991 and 2003 in nine European centres. A total of 226 shoulders in 210 patients were retrospectively reviewed at a mean of 122.7 months (61 to 219) or at revision. Clinical outcome was assessed using the Constant score, patient satisfaction score and range of movement. Kaplan-Meier survivorship analysis was performed with glenoid revision for loosening and radiological glenoid loosening (. sd. ) as endpoints. The Constant score was found to improve from a mean of 26.8 (. sd. 10.3) pre-operatively to 57.6 (. sd. 20.0) post-operatively (p < 0.001). Active forward flexion improved from a mean of 85.3° (. sd. 27.4) pre-operatively to 125° (. sd. 37.3) postoperatively (p < 0.001). External rotation improved from a mean of 7° (. sd. 6.5) pre-operatively to 30.3° (. sd. 21.8°) post-operatively (p < 0.001). Survivorship with revision of the glenoid component as the endpoint was 99.1% at five years, 94.5% at ten years and 79.4% at 15 years. Survivorship with radiological loosening as the endpoint was 99.1% at five years, 80.3% at ten years and 33.6% at 15 years. Younger patient age and the curettage technique for glenoid preparation correlated with loosening. The rate of glenoid revision and radiological loosening increased with duration of follow-up, but not until a follow-up of five years. Therefore, we recommend that future studies reporting radiological outcomes of new glenoid designs should report follow-up of at least five to ten years


The Journal of Bone & Joint Surgery British Volume
Vol. 87-B, Issue 4 | Pages 496 - 500
1 Apr 2005
Mileti J Sperling JW Cofield RH Harrington JR Hoskin TL

There are theoretical and practical advantages to modular rather than monoblock designs of prostheses for shoulder arthroplasty, but there are no reported studies which specifically compare the clinical and radiological results of their use. We have compared the results of unconstrained total shoulder arthroplasty for osteoarthritis using both types of implant. The monoblock design was used between 1992 and 1995 and the modular design after 1995. Both had cemented all-polyethylene glenoids, the monoblock with matched and the modular with mismatched radii of curvature. There were 34 consecutive shoulders in each group with a mean follow-up of 6.1 years in the first and 5.2 years in the second. There were no significant differences in improvement of pain scores, active elevation, external rotation, internal rotation, patient satisfaction, or the Neer ratings between the two groups. Two of 28 glenoid components in the first group and six of 30 in the second met the criteria for being radiologically at risk for loosening (p = 0.25). There were no significant differences in clinical outcome or radiological changes between the first- and second-generation designs of implant for shoulder arthroplasty


The Journal of Bone & Joint Surgery British Volume
Vol. 84-B, Issue 1 | Pages 77 - 82
1 Jan 2002
Ikävalko M Lehto MUK Repo A Kautiainen H Hämäläinen M

We present the results of 525 primary Souter elbow arthroplasties undertaken in 406 patients between 1982 and 1997. There were 372 women and 34 men with a mean age of 57 years; 119 patients had a bilateral procedure. The elbows were affected by chronic inflammatory disease, usually rheumatoid arthritis, which had been present for a mean of 24.7 years (2 to 70). In about 30% the joints were grossly destroyed with significant loss of bone. In 179 elbows the ulnar components were metal-backed and retentive; in the remaining 346, with better bone stock, non-retentive, all-polyethylene prostheses were used. Because of complications, 108 further operations were required in 82 patients. During the early years the incidence of complications was higher. Dislocation was the indication for 30 further procedures in 26 patients. Thirty patients underwent 33 revision procedures for aseptic loosening, 12 had 29 operations because of deep infection, two for superficial infection, and 14 further operations were done for other reasons. The cumulative rate of success, without aseptic loosening, five and ten years after surgery, was 96% and 85%, respectively


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 3 | Pages 387 - 392
1 Mar 2010
Kasten P Pape G Raiss P Bruckner T Rickert M Zeifang F Loew M

We have investigated the mid-term outcome of total shoulder replacement using a keeled cemented glenoid component and a modern cementing technique with regard to the causes of failure and loosening of the components.

Between 1997 and 2003 we performed 96 total shoulder replacements on 88 patients, 24 men and 64 women with a mean age of 69.7 years (31 to 82). The minimum follow-up was five years and at the time of review 87 shoulders (77 patients) were examined at a mean follow-up of 89.1 months (60 to 127). Cumulative survival curves were generated with re-operations (accomplished and planned), survivorship of the proshesis, loosening of the glenoid (defined as tilt > 5° or subsidence > 5 mm), the presence of radiolucent lines and a Constant score of < 30 as the endpoints.

There were two re-operations not involving revision of the implants and the survival rate of the prosthesis was 100.0% for the follow-up period, with an absolute Constant score of > 30 as the endpoint the survival rate was 98%. Radiological glenoid loosening was 9% after five years, and 33% after nine years. There was an incidence of 8% of radiolucent lines in more than three of six zones in the immediate post-operative period, of 37.0% after the first year which increased to 87.0% after nine years. There was no correlation between the score of Boileau and the total Constant score at the latest follow-up, but there was correlation between glenoid loosening and pain (p = 0.001).

We found that total shoulder replacement had an excellent mid-term survivorship and clinical outcome. The surgical and cementing techniques were related to the decrease in radiolucent lines around the glenoid compared with earlier studies. One concern, however, was the fact that radiolucent lines increased over time and there was a rate of glenoid loosening of 9% after five years and 33% after nine years. This suggests that the design of the glenoid component, and the implantation and cementing techniques may need further improvement.


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 10 | Pages 1416 - 1421
1 Oct 2010
Qureshi F Draviaraj KP Stanley D

Between September 1993 and September 1996, we performed 34 Kudo 5 total elbow replacements in 31 rheumatoid patients. All 22 surviving patients were reviewed at a mean of 11.9 years (10 to 14). Their mean age was 56 years (37 to 78) at the time of operation. All had Larsen grade IV or V rheumatoid changes on X-ray. Nine (three bilateral replacements and six unilateral) had died from unrelated causes. One who had died before ten years underwent revision for dislocation.

Of the 22 total elbow replacements reviewed six had required revision, four for aseptic loosening (one humeral and three ulnar) and two for infection. Post-operatively, one patient had neuropraxia of the ulnar nerve and one of the radial nerve. Two patients had valgus tilting of the ulnar component.

With revision as the endpoint, the mean survival time for the prosthesis was 11.3 years (95% confidence interval (10 to 13) and the estimated survival of the prosthesis at 12 years according to Kaplan-Meier survival analysis was 74% (95% confidence interval 0.53 to 0.91).

Of the 16 surviving implants, ten were free from pain, four had mild pain and two moderate. The mean arc of flexion/extension of the elbow was 106° (65° to 130°) with pronation/supination of 90° (30° to 150°) with the joint at 90° of flexion. The mean Mayo elbow performance score was 82 (60 to 100) with five excellent, ten good and one fair result.

Good long-term results can be expected using the Kudo 5 total elbow replacement in patients with rheumatoid disease, with a low incidence of loosening of the components.


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 7 | Pages 970 - 974
1 Jul 2010
Foruria AM Sperling JW Ankem HK Oh LS Cofield RH

This study was performed to review the safety and outcome of total shoulder replacements in patients who are ≥ 80 years of age. A total of 50 total shoulder replacements in 44 patients at a mean age of 82 years (80 to 89) were studied. Their health and shoulder status, the operation and post-operative course were analysed, including pain, movement, patient satisfaction, medical and surgical complications, radiographs, the need for revision surgery, and implant and patient survival. A total of 27 patients had an ASA classification of III or IV and medical abnormalities were common. Of the 13 shoulders with bony deficiency of the glenoid, nine required grafting. The duration of hospital stay was prolonged and blood transfusions were common. There were no peri-operative deaths. The mean follow-up was for 5.5 years (2 to 12). Pain was significantly reduced (p < 0.001) and movement improved in active elevation and both external and internal rotation (p < 0.001). Using the Neer scale for assessing outcome, 40 (80%) shoulders had an excellent or satisfactory result. There were medical or surgical complications in 17 cases. Four shoulders developed radiological evidence of loosened glenoid components, and three of these had a poor outcome. Three other shoulders required revision, two for instability. By the time of this review 39 of the patients had died from unrelated causes at a mean of 7.5 years (0.8 to 16.4) after surgery.

Total shoulder replacement is a relatively effective treatment in this elderly group of patients. However, there is a requirement for more intense patient care in the peri-operative period, and non-fatal medical or surgical complications are common. Most of these elderly patients will have a comfortable functional shoulder for the rest of their lives.


The Journal of Bone & Joint Surgery British Volume
Vol. 88-B, Issue 11 | Pages 1460 - 1463
1 Nov 2006
Landor I Vavrik P Jahoda D Guttler K Sosna A

We assessed the long-term results of 58 Souter-Strathclyde total elbow replacements in 49 patients with rheumatoid arthritis. The mean length of follow-up was 9.5 years (0.7 to 16.7). The mean pre-operative Mayo Elbow Performance Score was 30 (15 to 80) and at final follow-up was 82 (60 to 95). A total of 13 elbows (22.4%) were revised, ten (17.2%) for aseptic loosening, one (1.7%) for instability, one (1.7%) for secondary loosening after fracture, and one elbow (1.7%) was removed because of deep infection. The Kaplan-Meier survival rate was 70% and 53% at ten and 16 years, respectively. Failure of the ulnar component was found to be the main problem in relation to the loosening. Anterior transposition of the ulnar nerve had no influence on ulnar nerve paresthaesiae in these patients.


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 12 | Pages 1594 - 1600
1 Dec 2009
Khan A Bunker TD Kitson JB

There are no long-term published results on the survival of a third-generation cemented total shoulder replacement. We describe a clinical and radiological study of the Aequalis total shoulder replacement for a minimum of ten years. Between September 1996 and May 1998, 39 consecutive patients underwent a primary cemented total shoulder replacement using this prosthesis. Data were collected prospectively on all patients each year, for a minimum of ten years, or until death or failure of the prosthesis. At a follow-up of at least ten years, 12 patients had died with the prosthesis intact and two had emigrated, leaving 25 available for clinical review. Of these, 13 had rheumatoid arthritis and 12 osteoarthritis. One refused radiological review leaving 24 with fresh radiographs.

Survivorship at ten years was 100% for the humeral component and 92% for the glenoid component. The incidence of lucent lines was low. No humeral component was thought to be at risk and only two glenoid components. The osteoarthritic group gained a mean 65° in forward flexion and their Constant score improved by a mean 41.4 points (13 to 55). The rheumatoid group gained a mean of 24° in flexion and their Constant score improved by 29.4 points. This difference may have been due to failure of the rotator cuff in 75% of the patients with rheumatoid arthritis.

Thus a third-generation total shoulder replacement gives an excellent result in patients with osteoarthritis and an intact rotator cuff. Patients with rheumatoid arthritis have a 75% risk of failure of the rotator cuff at ten years.


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 6 | Pages 764 - 769
1 Jun 2008
Raiss P Aldinger PR Kasten P Rickert M Loew M

Our aim in this prospective study was to evaluate the outcome of total shoulder replacement in the treatment of young and middle-aged active patients with primary glenohumeral osteoarthritis. We reviewed 21 patients (21 shoulders) with a mean age of 55 years (37 to 60). The mean follow-up was seven years (5 to 9). The same anatomical, third-generation, cemented implant had been used in all patients. All the patients were evaluated radiologically and clinically using the Constant and Murley score.

No patients required revision. In one a tear of the supraspinatus tendon occurred. Overall, 20 patients (95%) were either very satisfied (n = 18) or satisfied (n = 2) with the outcome. Significant differences (p < 0.0001) were found for all categories of the Constant and Murley score pre- and post-operatively. The mean Constant and Murley score increased from 24.1 points (10 to 45) to 64.5 points (39 to 93), and the relative score from 30.4% (11% to 50%) to 83% (54% to 116%). No clinical or radiological signs of loosening of the implant were seen.

For young and middle-aged patients with osteoarthritis, third-generation total shoulder replacement is a viable method of treatment with a low rate of complications and excellent results in the mid-term.


The Journal of Bone & Joint Surgery British Volume
Vol. 88-B, Issue 7 | Pages 920 - 924
1 Jul 2006
Mori T Kudo H Iwano K Juji T

We studied 11 patients (14 elbows) with gross rheumatoid deformity of the elbow, treated by total arthroplasty using the Kudo type-5 unlinked prosthesis, and who were evaluated between five and 11 years after operation. Massive bone defects were augmented by autogenous bone grafts. There were no major complications such as infection, subluxation or loosening. In most elbows relief from pain and stability were achieved. The results, according to the Mayo Elbow Performance Score, were excellent in eight, good in five and fair in one. In most elbows there was minimal or no resorption of the grafted bone. There were no radiolucent lines around the stems of the cementless components. This study shows that even highly unstable rheumatoid elbows can be replaced successfully using an unlinked prosthesis, with augmentation by grafting for major defects of bone.


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 3 | Pages 336 - 342
1 Mar 2008
Cuff DJ Virani NA Levy J Frankle MA Derasari A Hines B Pupello DR Cancio M Mighell M

We retrospectively reviewed 21 patients (22 shoulders) who presented with deep infection after surgery to the shoulder, 17 having previously undergone hemiarthroplasty and five open repair of the rotator cuff. Nine shoulders had undergone previous surgical attempts to eradicate their infection. The diagnosis of infection was based on a combination of clinical suspicion (16 shoulders), positive frozen sections (> 5 polymorphonuclear leukocytes per high-power field) at the time of revision (15 shoulders), positive intra-operative cultures (18 shoulders) or the pre-operative radiological appearances. The patients were treated by an extensive debridement, intravenous antibiotics, and conversion to a reverse shoulder prosthesis in either a single- (10 shoulders) or a two-stage (12 shoulders) procedure.

At a mean follow-up of 43 months (25 to 66) there was no evidence of recurrent infection. All outcome measures showed statistically significant improvements. Mean abduction improved from 36.1° (sd 27.8) pre-operatively to 75.7° (sd 36.0) (p < 0.0001), the mean forward flexion from 43.1° (sd 33.5) to 79.5° (sd 43.2) (p = 0.0003), and mean external rotation from 10.2° (sd 18.7) to 25.4° (sd 23.5) (p = 0.0037). There was no statistically significant difference in any outcome between the single-stage and the two-stage group.