We compared the complications and outcome of tibial lengthening using the Ilizarov method with and without the use of a supplementary intramedullary nail. In a retrospective case-matched series assembled from 176 patients with tibial lengthening, we matched 52 patients (26 pairs, group A with nail and group B without) according to the following criteria in order of importance: 1) difference in amount of lengthening (± 2 cm); 2) percentage difference in lengthening (± 5%); 3) difference in patient’s age (± seven years); 4) aetiology of the shortening, and 5) level of difficulty in obtaining the correction. The outcome was evaluated using the external fixator index, the healing index and an outcome score according to the criteria of Paley. It was found that some complications were specific to group A or B respectively, but others were common to both groups. The outcome was generally better in lengthenings with a nail, although there was a higher incidence of rectifiable equinus deformity in these patients.
Between October 2001 and September 2009 we lengthened 242 lower-limb segments in 180 patients using the Intramedullary Skeletal Kinetic Distractor (ISKD). Mechanical failure was defined either as breakage of the ISKD or failure of the internal mechanism to activate. Retrieved nails which failed mechanically were examined by the manufacturer for defects. In all, 15 ISKDs in 12 patients (13 limbs) failed mechanically representing an overall failure rate of 6.2%, with fracture of the device occurring in ten of the 15 failures. Two nails in one patient failed to lengthen and had to be replaced. The manufacturer detected an error in the assembly of the nail, which prompted a wide recall. One nail jammed after being forcefully inserted, and two nails failed to lengthen fully. Lengthening was achieved in all 12 patients, although three required a second operation to exchange a defective nail for a new, functioning device. The ISKD is a complex mechanical device which lengthens by the oscillation of two telescopic sections connected by a threaded rod. The junction between these sections is surrounded by a keyring collar. This keyring collar is the weakest part of the device.