As many as 25% to 40% of unicompartmental knee
replacement (UKR) revisions are performed for pain, a possible cause
of which is proximal tibial strain. The aim of this study was to
examine the effect of UKR implant design and material on cortical
and cancellous proximal tibial strain in a synthetic bone model.
Composite Sawbone tibiae were implanted with cemented UKR components
of different designs, either all-polyethylene or metal-backed. The tibiae
were subsequently loaded in 500 N increments to 2500 N, unloading
between increments. Cortical surface strain was measured using a
digital image correlation technique. Cancellous damage was measured
using acoustic emission, an engineering technique that detects sonic
waves (‘hits’) produced when damage occurs in material. Anteromedial cortical surface strain showed significant differences
between implants at 1500 N and 2500 N in the proximal 10 mm only
(p <
0.001), with relative strain shielding in metal-backed implants.
Acoustic emission showed significant differences in cancellous bone
damage between implants at all loads (p = 0.001). All-polyethylene implants
displayed 16.6 times the total number of cumulative acoustic emission
hits as controls. All-polyethylene implants also displayed more
hits than controls at all loads (p <
0.001), more than metal-backed
implants at loads ≥ 1500 N (p <
0.001), and greater acoustic
emission activity on unloading than controls (p = 0.01), reflecting
a lack of implant stiffness. All-polyethylene implants were associated
with a significant increase in damage at the microscopic level compared
with metal-backed implants, even at low loads. All-polyethylene
implants should be used with caution in patients who are likely
to impose large loads across their knee joint. Cite this article:
Radiological assessment of total and unicompartmental
knee replacement remains an essential part of routine care and follow-up.
Appreciation of the various measurements that can be identified
radiologically is important. It is likely that routine plain radiographs
will continue to be used, although there has been a trend towards
using newer technologies such as CT, especially in a failing knee,
where it provides more detailed information, albeit with a higher
radiation exposure. The purpose of this paper is to outline the radiological parameters
used to evaluate knee replacements, describe how these are measured
or classified, and review the current literature to determine their
efficacy where possible.