Advertisement for orthosearch.org.uk
Results 1 - 1 of 1
Results per page:
Applied filters
The Bone & Joint Journal
Dates
Year From

Year To
The Journal of Bone & Joint Surgery British Volume
Vol. 76-B, Issue 4 | Pages 575 - 578
1 Jul 1994
Berry D Barnes C Scott R Cabanela M Poss R

Ten cases are described of catastrophic failure of the polyethylene liner of three different designs of uncemented acetabular component. Failure occurred as a result of either 'wearthrough' to the metal backing, liner fracture or a combination of both, at a mean of 4.6 years after implantation (2 to 7.6). At revision there was metallosis in all hips and osteolysis of the femur or the pelvis in six. Catastrophic failure was seen only in cups with a minimum polyethylene thickness of less than 5 mm.