Bisphosphonates reduce peri-prosthetic bone loss in the short term after total hip replacement but the mid- and longer term effects are not known. The aims of this randomised trial were to examine the effect of a single dose of 90 mg of pamidronate on the clinical and radiological outcome and peri-prosthetic bone mineral density in 50 patients (56 hips) over a five-year period, following total hip replacement. At five years, 37 patients (42 hips) returned for assessment. The Harris hip scores were similar in the pamidronate and placebo groups throughout the study. Also at five years, four patients, two from each group had osteolytic lesions on plain radiography. These were located around the acetabular component in three patients and in the femoral calcar in one. The femoral and acetabular peri-prosthetic bone mineral density in the pamidronate group and the control group was similar at five years. Pamidronate given as a single post-operative dose does not appear to influence the clinical outcome or prevent the development of osteolytic lesions at five years after total hip replacement.
The aims of this study were to examine the repeatability of measurements of bone mineral density (BMD) around a cemented polyethylene Charnley acetabular component using dual-energy x-ray absorptiometry and to determine the longitudinal pattern of change in BMD during the first 24 months after surgery. The precision of measurements of BMD in 19 subjects ranged from 7.7% to 10.8% between regions, using a four-region-of-interest model. A longitudinal study of 27 patients demonstrated a transient decrease in net pelvic BMD during the first 12 months, which recovered to baseline at 24 months. The BMD in the region medial to the dome of the component reduced by between 7% and 10% during the first three months, but recovered to approximately baseline values by two years. Changes in BMD in the pelvis around cemented acetabular components may be measured using dual-energy x-ray absorptiometry. Bone loss after insertion of a cemented Charnley acetabular component is small, transient and occurs mainly at the medial wall of the acetabulum. After two years, bone mass returns to baseline values, with a pattern suggesting a uniform transmission of load to the acetabulum.
We aimed to evaluate the precision and longitudinal sensitivity of measurement of bone mineral density (BMD) in the pelvis and to determine the effect of bone cement on the measurement of BMD in femoral regions of interest (ROI) after total hip arthroplasty (THA). A series of 29 patients had duplicate dual-energy x-ray absorptiometry (DXA) scans of the hip within 13 months of THA. Pelvic analyses using 3- and 4-ROI models gave a coefficient of variation (CV) of 2.5% to 3.6% and of 2.5% to 4.8%, respectively. Repeat scans in 17 subjects one year later showed a significant change in BMD in three regions using the 4-ROI model, compared with change in only one region with the 3-ROI model (p <
0.05). Manual exclusion of cement from femoral ROIs increased the net CV from 1.6% to 3.6% (p = 0.001), and decreased the measured BMD by 20% (t = 12.1, p <
0.001). Studies of two cement phantoms in vitro showed a small downward drift in bone cement BMD giving a measurement error of less than 0.03 g/cm2/year associated with inclusion of cement in femoral ROIs. Changes in pelvic periprosthetic BMD are best detected using a 4-ROI model. Analysis of femoral ROI is more precise without exclusion of cement although an awareness of its effect on the measurement of the BMD is needed.