The practice of overlapping surgery has been increasing in the delivery of orthopaedic surgery, aiming to provide efficient, high-quality care. However, there are concerns about the safety of this practice. The purpose of this study was to examine the safety and efficacy of a model of partially overlapping surgery that we termed ‘swing room’ in the practice of primary total hip (THA) and knee arthroplasty (TKA). A retrospective review of prospectively collected data was carried out on patients who underwent primary THA and TKA between 2006 and 2017 in two academic centres. Cases were stratified as partially overlapping (swing room), in which the surgeon is in one operating room (OR) while the next patient is being prepared in another, or nonoverlapping surgery. The demographic details of the patients which were collected included operating time, length of stay (LOS), postoperative complications within six weeks of the procedure, unplanned hospital readmissions, and unplanned reoperations. Fisher's exact, Wilcoxon rank-sum tests, chi-squared tests, and logistic regression analysis were used for statistical analysis.Aims
Methods
Fretting and corrosion at the modular head/neck junction, known
as trunnionosis, in total hip arthroplasty (THA) is a cause of adverse
reaction to metal debris (ARMD). We describe the outcome of revision
of metal-on-polyethylene (MoP) THA for ARMD due to trunnionosis
with emphasis on the risk of major complications. A total of 36 patients with a MoP THA who underwent revision
for ARMD due to trunnionosis were identified. Three were excluded
as their revision had been to another metal head. The remaining
33 were revised to a ceramic head with a titanium sleeve. We describe
the presentation, revision findings, and risk of complications in
these patients.Aims
Patients and Methods
To present a surgically relevant update of trunnionosis. Systematic review performed April 2017.Aims
Materials and Methods
Patients with neuromuscular imbalance who require
total hip arthroplasty (THA) present particular technical problems
due to altered anatomy, abnormal bone stock, muscular imbalance
and problems of rehabilitation. In this systematic review, we studied articles dealing with THA
in patients with neuromuscular imbalance, published before April
2017. We recorded the demographics of the patients and the type
of neuromuscular pathology, the indication for surgery, surgical
approach, concomitant soft-tissue releases, the type of implant
and bearing, pain and functional outcome as well as complications
and survival. Recent advances in THA technology allow for successful outcomes
in these patients. Our review suggests excellent benefits for pain
relief and good functional outcome might be expected with a modest
risk of complication. Cite this article:
Fractures around total knee arthroplasties pose
a significant surgical challenge. Most can be managed with osteosynthesis
and salvage of the replacement. The techniques of fixation of these
fractures and revision surgery have evolved and so has the assessment
of outcome. This specialty update summarises the current evidence
for the classification, methods of fixation, revision surgery and
outcomes of the management of periprosthetic fractures associated
with total knee arthroplasty. Cite this article:
Reconstruction of the acetabulum after failed total hip arthroplasty
(THA) can be a surgical challenge in the presence of severe bone
loss. We report the long-term survival of a porous tantalum revision
acetabular component, its radiological appearance and quality of
life outcomes. We reviewed the results of 46 patients who had undergone revision
of a failed acetabular component with a Paprosky II or III bone
defect and reconstruction with a hemispherical, tantalum acetabular
component, supplementary screws and a cemented polyethylene liner.Aims
Patients and Methods
We present a case series of ten metal-on-polyethylene total hip
arthroplasties (MoP THAs) with delayed dislocation associated with
unrecognised adverse local tissue reaction due to corrosion at the
trunnion and pseudotumour formation. The diagnosis was not suspected in nine of the ten patients (six
female/four male; mean age 66 years), despite treatment in a specialist
unit (mean time from index surgery to revision was 58 months, 36
to 84). It was identified at revision surgery and subsequently confirmed
by histological examination of resected tissue. Pre-operative assessment
and culture results ruled out infection. A variety of treatment
strategies were used, including resection of the pseudotumour and
efforts to avoid recurrent dislocation. Aims
Methods
Tapered fluted titanium stems are increasingly
used for femoral revision arthroplasty. They are available in modular and
non-modular forms. Modularity has advantages when the bone loss
is severe, the proximal femur is mis shapen or the surgeon is unfamiliar
with the implant, but it introduces the risk of fracture of the
stem at the junction between it and the proximal body segment. For
that reason, and while awaiting intermediate-term results of more recently
introduced designs of this junction, non-modularity has attracted
attention, at least for straightforward revision cases. We review the risks and causes of fracture of tapered titanium
modular revision stems and present an argument in favour of the
more selective use of modular designs. Cite this article:
Periprosthetic joint infection (PJI) complicates
between 0.5% and 1.2% primary total hip arthroplasties (THAs) and
may have devastating consequences. The traditional assessment of
patients suffering from PJI has involved the serological study of
inflammatory markers and microbiological analysis of samples obtained
from the joint space. Treatment has involved debridement and revision
arthroplasty performed in either one or two stages. We present an update on the burden of PJI, strategies for its
diagnosis and treatment, the challenge of resistant organisms and
the need for definitive evidence to guide the treatment of PJI after
THA. Cite this article:
Adverse reaction to wear and corrosion debris
is a cause for concern in total hip arthroplasty (THA). Modular junctions
are a potential source of such wear products and are associated
with secondary pseudotumour formation. We present a consecutive series of 17 patients treated at our
unit for this complication following metal-on-highly cross-linked
polyethylene (MoP) THA. We emphasise the risk of misdiagnosis as
infection, and present the aggregate laboratory results and pathological
findings in this series. The clinical presentation was pain, swelling or instability.
Solid, cystic and mixed soft-tissue lesions were noted on imaging
and confirmed intra-operatively. Corrosion at the head–neck junction
was noted in all cases. No bacteria were isolated on multiple pre-
and intra-operative samples yet the mean erythrocyte sedimentation
rate was 49 (9 to 100) and C-reactive protein 32 (0.6 to 106) and
stromal polymorphonuclear cell counts were noted in nine cases. Adverse soft–tissue reactions can occur in MoP THA owing to corrosion
products released from the head–neck junction. The diagnosis should
be carefully considered when investigating pain after THA. This
may avoid the misdiagnosis of periprosthetic infection with an unidentified
organism and mitigate the unnecessary management of these cases
with complete single- or two-stage exchange. Cite this article:
The Unified Classification System (UCS) was introduced
because of a growing need to have a standardised universal classification
system of periprosthetic fractures. It combines and simplifies many
existing classification systems, and can be applied to any fracture
around any partial or total joint replacement occurring during or
after operation. Our goal was to assess the inter- and intra-observer
reliability of the UCS in association with knee replacement when
classifying fractures affecting one or more of the femur, tibia
or patella. We used an international panel of ten orthopaedic surgeons with
subspecialty fellowship training and expertise in adult hip and
knee reconstruction (‘experts’) and ten residents of orthopaedic
surgery in the last two years of training (‘pre-experts’). They
each received 15 radiographs for evaluation. After six weeks they
evaluated the same radiographs again but in a different order. The reliability was assessed using the Kappa and weighted Kappa
values. The Kappa values for inter-observer reliability for the experts
and the pre-experts were 0.741 (95% confidence interval (CI) 0.707
to 0.774) and 0.765 (95% CI 0.733 to 0.797), respectively. The weighted
Kappa values for intra-observer reliability for the experts and
pre-experts were 0.898 (95% CI 0.846 to 0.950) and 0.878 (95% CI
0.815 to 0.942) respectively. The UCS has substantial inter-observer reliability and ‘near
perfect’ intra-observer reliability when used for periprosthetic
fractures in association with knee replacement in the hands of experienced
and inexperienced users. Cite this article:
Non-modular tapered fluted, titanium stems are
available for use in femoral revision. The combination of taper
and flutes on the stem provides axial and rotational stability,
respectively. The material and surface properties of the stem promotes
bone on-growth. If the surgeon is confident and reasonably experienced
in the surgical use of this sort of design and the case is relatively
straightforward, a non-modular design is effective. It also potentially reduces
implant inventory, and circumvents the potential problems of taper
junction corrosion and fatigue fracture. There are reports of excellent
survival, good clinical and functional results and evidence of subsequent
increase in proximal bone stock. Cite this article: Bo
The Unified Classification System (UCS) emphasises
the key principles in the assessment and management of peri-prosthetic
fractures complicating partial or total joint replacement. We tested the inter- and intra-observer agreement for the UCS
as applied to the pelvis and femur using 20 examples of peri-prosthetic
fracture in 17 patients. Each subtype of the UCS was represented
by at least one case. Specialist orthopaedic surgeons (experts)
and orthopaedic residents (pre-experts) assessed reliability on
two separate occasions. For the pelvis, the UCS showed inter-observer agreement of 0.837
(95% confidence intervals (CI) 0.798 to 0.876) for the experts and
0.728 (95% CI 0.689 to 0.767) for the pre-experts. The intra-observer
agreement for the experts was 0.861 (95% CI 0.760 to 0.963) and
0.803 (95% 0.688 to 0.918) for the pre-experts. For the femur, the
UCS showed an inter-observer kappa value of 0.805 (95% CI 0.765
to 0.845) for the experts and a value of 0.732 (95% CI 0.690 to 0.773)
for the pre-experts. The intra-observer agreement was 0.920 (95%
CI 0.867 to 0.973) for the experts, and 0.772 (95% CI 0.652 to 0.892)
for the pre-experts. This corresponds to a substantial and ‘almost
perfect’ inter- and intra-observer agreement for the UCS for peri-prosthetic
fractures of the pelvis and femur. We hope that unifying the terminology of these injuries will
assist in their assessment, treatment and outcome. Cite this article:
Advances in the treatment of periprosthetic joint
infections of the hip have once more pushed prosthesis preserving techniques
into the limelight. At the same time, the common infecting organisms
are evolving to become more resistant to conventional antimicrobial
agents. Whilst the epidemiology of resistant staphylococci is changing,
a number of recent reports have advocated the use of irrigation
and debridement and one-stage revision for the treatment of periprosthetic
joint infections due to resistant organisms. This review presents
the available evidence for the treatment of periprosthetic joint
infections of the hip, concentrating in particular on methicillin
resistant staphylococci. Cite this article:
Periprosthetic fractures are an increasingly
common complication following joint replacement. The principles
which underpin their evaluation and treatment are common across
the musculoskeletal system. The Unified Classification System proposes
a rational approach to treatment, regardless of the bone that is
broken or the joint involved. Cite this article:
Tapered, fluted, modular, titanium stems are
increasingly popular in the operative management of Vancouver B2 and
selected B3 peri-prosthetic femoral fractures. We have reviewed
the results at our institution looking at stem survival and clinical
outcomes and compared this with reported outcomes in the literature.
Stem survival at a mean of 54Â months was 96% in our series and 97%
for combined published cases. Review of radiology showed maintenance
or improvement of bone stock in 89% of cases with high rates of
femoral union. Favourable clinical outcome scores have reported
by several authors. No difference in survival or clinical scores
was observed between B2 and B3 fractures. Tapered stems are a useful
option in revision for femoral fracture across the spectrum of femoral
bone deficiency. Cite this article:
Hip arthrodesis remains a viable surgical technique
in well selected patients, typically the young manual labourer with
isolated unilateral hip disease. Despite this, its popularity with
patients and surgeons has decreased due to the evolution of hip
replacement, and is seldom chosen by young adult patients today.
The surgeon is more likely to encounter a patient who requests conversion
to total hip replacement (THR). The most common indications are
a painful pseudarthrosis, back pain, ipsilateral knee pain or contralateral
hip pain. Occasionally the patient will request conversion because
of difficulty with activities of daily living, body image and perceived
cosmesis. The technique of conversion and a discussion of the results
are presented. Cite this article:
Tapered, fluted, modular, titanium stems have
a long history in Europe and are increasing in popularity in North America.
We have reviewed the results at our institution looking at stem
survival and clinical outcomes. Radiological outcomes and quality
of life assessments have been performed and compared to cylindrical
non-modular cobalt chromium stems. Survival at five years was 94%.
This fell to 85% at ten years due to stem breakage with older designs.
Review of radiology showed maintenance or improvement of bone stock
in 87% of cases. Outcome scores were superior in tapered stems despite
worse pre-operative femoral deficiency. Tapered stems have proved
to be a useful alternative in revision total hip arthroplasty across
the spectrum of femoral bone deficiency.
Antibiotic impregnated articulating spacers are
used in two-stage revision total knee arthroplasty to deliver local antibiotic
therapy while preserving function. We have observed infection control
in greater than 95% of cases with functional outcomes approaching
those seen in revision for aseptic loosening. Higher failure has
been observed with methicillin resistant organisms.
We report the outcome at ten to 15 years of two-stage revision for hip infection in 99 patients using the Prostalac articulated hip spacer system. All the patients were contacted to determine their current functional and infection status using the Oxford-12, Short form-12, and Western Ontario and McMaster University Osteoarthritis Index questionnaires. A total of 11 of the 99 patients had a further infection, of whom seven responded to repeat surgery with no further sequelae. The mean interval between the stages was five months (1 to 36). We were able to review 48 living patients, with a mean age of 72 years (46 to 86), 34 (71%) of whom provided health-related quality-of-life outcome scores. The mean follow-up was 12 years (10 to 15). The long-term success rate was 89% and with additional surgery this rose to 96%. The mean global Western Ontario and McMaster University Osteoarthritis Index score was 80.6 ( Two-stage revision for hip infection using a Prostalac interim spacer offers a predictable and lasting solution for patients with this difficult problem.
A prospective cohort of 222 patients who underwent revision hip replacement between April 2001 and March 2004 was evaluated to determine predictors of function, pain and activity level between one and two years post-operatively, and to define quality of life outcomes using validated patient reported outcome tools. Predictive models were developed and proportional odds regression analyses were performed to identify factors that predict quality of life outcomes at one and two years post-operatively. The dependent outcome variables were the Western Ontario and McMaster Osteoarthritis Index (WOMAC) function and pain scores, and University of California Los Angeles activity scores. The independent variables included patient demographics, operative factors, and objective quality of life parameters, including pre-operative WOMAC, and the Short Form-12 mental component score. There was a significant improvement ( Predictors of quality of life outcomes after revision hip replacement were established. Although some patient-specific and surgery-specific variables were important, age, gender, Charnley class and pre-operative WOMAC function score had the most robust associations with outcome.
We studied various aspects of graft impaction and penetration of cement in an experimental model. Cancellous bone was removed proximally and local diaphyseal lytic defects were simulated in six human cadaver femora. After impaction grafting the specimens were sectioned and prepared for histomorphometric analysis. The porosity of the graft was lowest in Gruen zone 4 (52%) and highest in Gruen zone 1 (76%). At the levels of Gruen zones 6 and 2 the entire cross-section was almost filled with cement. Cement sometimes reached the endosteal surface in other Gruen zones. The mean peak impaction forces exerted with the impactors were negatively correlated with the porosity of the graft.
Removal of well-fixed, cementless, acetabular components during revision arthroplasty remains a challenging problem. Further damage to host bone may limit options for reconstruction and compromise the long-term result of the revision operation. We report the results of 31 hips with well-fixed, cementless sockets which were removed using a new cup extraction system. In all hips the socket was removed without difficulty and with minimal further bone loss.
There are few medium- and long-term data on the outcome of the use of proximal femoral structural allografts in revision hip arthroplasty. This is a study of a consecutive series of 40 proximal femoral allografts performed for failed total hip replacements using the same technique with a minimum follow-up of five years (mean 8.8 years; range 5 to 11.5 years). In all cases the stem was cemented into both the allograft and the host femur. The proximal femur of the host was resected in 37 cases. There were four early revisions (10%), two for infection, one for nonunion of the allograft-host junction, and one for allograft resorption noted at the time of revision of a failed acetabular reconstruction. Junctional nonunion was seen in three patients (8%), two of whom were managed successfully by bone grafting, and bone grafting and plating respectively. Instability was observed in four (10%). Trochanteric nonunion was seen in 18 patients (46%) and trochanteric escape in ten of these (27%). The mean Harris hip score improved from 39 to 79. Severe resorption involving the full thickness of the allograft was seen in seven patients (17.5%). This progressed rapidly and silently, but has yet to cause failure of any of the reconstructions. Profound resorption of the allograft may be related to a combination of factors, including a slow form of immune rejection, stress shielding and resorption due to mechanical disuse with solid cemented distal fixation, and the absence of any masking or protective effect which may be provided by the retention of the bivalved host bone as a vascularised onlay autograft. Although continued surveillance is warranted, the very good medium-term clinical results justify the continued use of structural allografts for failed total hip replacements with severe loss of proximal femoral bone.
The PROSTALAC functional spacer is made of antibiotic-loaded acrylic cement but has a small metal-on-polythene articular surface. We have used it as an interim spacer in two-stage exchange arthroplasty for infected total knee replacement. PROSTALAC allows continuous rehabilitation between stages as it maintains good alignment and stability of the knee and a reasonable range of movement. It also helps to maintain the soft-tissue planes, which facilitates the second-stage procedure. We reviewed 45 consecutive patients, treated over a period of nine years. The mean follow-up was for 48 months (20 to 112). At final review, there was no evidence of infection in 41 patients (91%); only one had a recurrent infection with the same organism. There was improvement in the Hospital for Special Surgery knee score between stages and at final review. The range of movement was maintained between stages. Complications were primarily related to the extensor mechanism and stability of the knee between stages. Both of these problems decreased with refinement of the design of the implant. The rate of cure of the infection in our patients was similar to that using other methods. Movement of the knee does not appear to hinder control of infection.
We have previously described a simple and reproducible three-dimensional technique of CT for the measurement of the cover of the femoral head in acetabular dysplasia in adults. We now describe the application of this technique in ten patients with symptomatic dysplasia to assess the degree and direction of dysplasia and to measure the cover obtained at acetabular osteotomy. The indices obtained gave a useful indication of the degree and direction of the dysplasia and confirmed which components had been used most efficiently to achieve cover. The information is easily presented in graphical form and gives a clearer indication of the cover obtained than the indices derived from plain radiographs.
We describe a U-shaped approach to the distal femur which, having divided the extensor mechanism and elevated the entire quadriceps muscle, gives excellent exposure and allows a number of reconstructive options. It was used in 14 patients, 13 of whom were followed up for a mean of 3.5 years (1 to 11). There was no case of flap necrosis, and complications related to the reconstruction were acceptable.
An analysis of the cement mantle obtained with the Exeter impaction allografting system at one centre showed that it was either deficient or absent in almost 47% of Gruen zones. We therefore examined the mantle obtained using this system at another hospital and compared the results with those from the CPT and Harris Precoat Systems at other centres. The surgical indications for the procedure and the patient details were broadly similar in all four hospitals. There was some variation in the frequency of use of cortical strut allografts, cerclage wires and wire mesh to supplement the impaction allograft. Analysis of the cement mantles showed that when uncertain Gruen zones were excluded, the incidence of zones with areas of absence or deficiency of the cement was 47% and 50%, respectively, for the two centres using the Exeter system, 21% for the CPT system and 18% for the Harris Precoat system. We measured the difference in size between the proximal allograft impactors and the definitive prosthesis for each system. The Exeter system impactors are shorter than the definitive prosthesis and taper sharply so that the cavity created is inadequate, especially distally. The CPT proximal impactors are considerably longer than the definitive prosthesis and are designed to give a mantle of approximately 2 mm medially and laterally and 1.5 mm anteriorly and posteriorly. The Harris Precoat proximal impactors allow for a mantle with a circumference of 0.75 mm in the smaller sizes and 1 mm in the larger. Many reports link the longevity of a cemented implant to the adequacy of the cement mantle. For this reason, femoral impaction systems require careful design to achieve a cement mantle which is uninterrupted in its length and adequate in its thickness. Our results suggest that some current systems require modification.