The osseo-integration of an
Ten cases are described of catastrophic failure of the polyethylene liner of three different designs of
We analysed the results of different strategies in the revision of primary
We measured polyethylene wear in 231 porous-coated uncemented acetabular cups. We divided the hips into two groups according to the fixation of the femoral component, by cementing (n = 97) or press-fit (n = 134). Follow-up was from three to five years. The patients in two sub-groups were matched for weight, diagnosis, sex, age and length of follow-up. The linear wear rate of cups articulated with uncemented femoral components (0.22 mm/year) was significantly higher than the wear rate (0.15 mm/year) of cups articulated within cemented femoral components (p <
0.05). These results can be compared with previously reported wear rates of 0.08 mm/year for cemented all-polyethylene cups and 0.11 mm/year for cemented metal-backed cups. The higher wear rates of uncemented arthroplasties could jeopardize the long-term results of this type of hip replacement.
We investigated the rate of polyethylene wear of a cementless acetabular component at different periods of follow-up in order to test the hypothesis than an irrecoverable deformation process (creep) was followed by an initially low, but gradually increasing wear rate. We studied prospectively 93 uncemented total hip arthroplasties in 83 patients (mean age 50 years (22 to 63)) with a mean follow-up of 8.2 years (3 to 12). We measured the penetration of the femoral head from radiographs taken immediately after surgery at three, six and nine years, or at the latest follow-up. The median wear rate was 0.17 mm per year in the first three years, a finding which we considered to be caused by creep. Thereafter, the rate of wear declined to 0.07 mm per year (four- to six-year period) and then increased to 0.17 mm per year (seven to nine years) and 0.27 mm per year (more than nine years), which we considered to be a reflection of genuine polyethylene wear. After the nine-year follow-up the wear rates were higher in patients with marked osteolysis. We found no relationship between the inclination angle of the acetabular component or femoral head orientation and the rate of wear. No acetabular component required revision.
In developmental dysplasia of the hip (DDH),
a bone defect is often observed superior to the acetabulum after
the reconstruction at the level of the true acetabulum during total
hip replacement (THR). However, the essential amount of uncemented
acetabular component coverage required for a satisfactory outcome
remains controversial. The purpose of this study was to assess the
stability and function of acetabular components with a lack of coverage >
30% (31% to 50%). A total of 760 DDH patients underwent THR with
acetabular reconstruction at the level of the true floor. Lack of
coverage above the acetabular component of >
30% occurred in 56
patients. Intra-operatively, autogenous morcellised bone grafts
were used to fill the uncovered portion. Other than two screws inserted through
the acetabular shell, no additional structural supports were used
in these hips. In all, four patients were lost to follow-up. Therefore,
52 patients (52 hips, 41 women and 11 men) with a mean age of 60.1
years (42 to 78) were available for this study at a mean of 4.8
years (3 to 7). There were no instances of prosthesis revision or
marked loosening during the follow-up. The Harris hip score improved
from a mean of 40.7 points ( Cite this article:
Polyethylene wear of acetabular components is a key factor in the development of periprosthetic osteolysis and wear at the articular surface has been well documented and quantified, but fewer data are available about changes which occur at the backside of the liner. At revision surgery for loosening of the femoral component we retrieved 35 conventional modular acetabular liners of the same design. Linear and volumetric articular wear, backside volumetric change and the volume of the screw-head indentations were quantified. These volumes, clinical data and the results from radiological The rate of backside volumetric change was found to be 2.8% of the rate of volumetric articular wear and decreased with increasing liner size. Migrated acetabular components showed significantly higher rates of backside volumetric change plus screw-head indentations than those without migration. The backside volumetric change was at least ten times larger than finite-element simulation had suggested. In a stable acetabular component with well-anchored screws, the amount of backside wear should not cause clinical problems. Impingement of the screw-heads could produce more wear particles than those generated at the liner-shell interface. Because the rate of backside volumetric change is only 2.8% of the rate of volumetric articular wear and since creep is likely to contribute a significant portion to this, the debris generated by wear at the backside of the liner may not be sufficient to create a strong osteolytic response.
Aims. Despite higher rates of revision after total hip arthroplasty (THA) being reported for uncemented stems in patients aged > 75 years, they are frequently used in this age group. Increased mortality after cemented fixation is often used as a justification, but recent data do not confirm this association. The aim of this study was to investigate the influence of the design of the stem and the type of fixation on the rate of revision and immediate postoperative mortality, focusing on the age and sex of the patients. Methods. A total of 333,144 patients with primary osteoarthritis (OA) of the hip who underwent elective THA between November 2012 and September 2022, using
Dislocation remains a leading cause of failure following revision total hip arthroplasty (THA). While dual-mobility (DM) bearings have been shown to mitigate this risk, options are limited when retaining or implanting an uncemented shell without modular DM options. In these circumstances, a monoblock DM cup, designed for cementing, can be cemented into an uncemented acetabular shell. The goal of this study was to describe the implant survival, complications, and radiological outcomes of this construct. We identified 64 patients (65 hips) who had a single-design cemented DM cup cemented into an uncemented acetabular shell during revision THA between 2018 and 2020 at our institution. Cups were cemented into either uncemented cups designed for liner cementing (n = 48; 74%) or retained (n = 17; 26%) acetabular components. Median outer head diameter was 42 mm. Mean age was 69 years (SD 11), mean BMI was 32 kg/m2 (SD 8), and 52% (n = 34) were female. Survival was assessed using Kaplan-Meier methods. Mean follow-up was two years (SD 0.97).Aims
Methods
We reviewed 120 consecutive primary total hip replacements in 109 patients in whom a Freeman uncemented metal-backed component had been used. Acetabular components were used with a Freeman neck-retaining stem in one of three configurations: cemented smooth stem, uncemented smooth stem or uncemented ridged stem. After a mean follow up of 72 months (62 to 113) there were nine cases of aseptic loosening; a survivorship at eight years of 83%. In addition, more than one-third of the remaining surviving cups showed loosening radiologically. The Freeman acetabular component demonstrated a characteristic pattern of loosening on radiographs. The high incidence of aseptic loosening can be explained by poor design and material failure.
Aims. The aim of this study was to determine if uncemented acetabular polyethylene (PE) liner geometry, and lip size, influenced the risk of revision for instability or loosening. Methods. A total of 202,511 primary total hip arthroplasties (THAs) with
The Norwegian Arthroplasty Register has shown that several designs of uncemented femoral stems give good or excellent survivorship. The overall findings for uncemented total hip replacement however, have been disappointing because of poor results with the use of metal-backed acetabular components. In this study, we exclusively investigated the medium-to long-term performance of primary
Aims. To compare long-term survival of all-cemented and hybrid total hip arthroplasty (THA) using the Exeter Universal stem. Methods. Details of 1,086 THAs performed between 1999 and 2005 using the Exeter stem and either a cemented (632) or
We have reviewed 70 Harris-Galante
Aims. Contemporary outcomes of primary total hip arthroplasties (THAs) with highly cross-linked polyethylene (HXLPE) liners in patients with inflammatory arthritis have not been well studied. This study examined the implant survivorship, complications, radiological results, and clinical outcomes of THA in patients with inflammatory arthritis. Methods. We identified 418 hips (350 patients) with a primary diagnosis of inflammatory arthritis who underwent primary THA with HXLPE liners from January 2000 to December 2017. Of these hips, 68% had rheumatoid arthritis (n = 286), 13% ankylosing spondylitis (n = 53), 7% juvenile rheumatoid arthritis (n = 29), 6% psoriatic arthritis (n = 24), 5% systemic lupus erythematosus (n = 23), and 1% scleroderma (n = 3). Mean age was 58 years (SD 14.8), 66.3% were female (n = 277), and mean BMI was 29 kg/m. 2. (SD 7). Uncemented femoral components were used in 77% of cases (n = 320).
Aims.
Aims. Head-taper corrosion is a cause of failure in total hip arthroplasty (THA). Recent reports have described an increasing number of V40 taper failures with adverse local tissue reaction (ALTR). However, the real incidence of V40 taper damage and its cause remain unknown. The aim of this study was to evaluate the long-term incidence of ALTR in a consecutive series of THAs using a V40 taper and identify potentially related factors. Methods. Between January 2006 and June 2007, a total of 121 patients underwent THA using either an uncemented (Accolade I, made of Ti. 12. Mo. 6. Zr. 2. Fe; Stryker, USA) or a cemented (ABG II, made of cobalt-chrome-molybdenum (CoCrMo); Stryker) femoral component, both with a V40 taper (Stryker).
We compared the rate of revision for instability
after total hip replacement (THR) when lipped and non-lipped acetabular
liners were used. We hypothesised that the use of a lipped liner
in a modular
Hip resurfacing is a bone-conserving procedure with respect to proximal femoral resection, but there is debate in the literature as to whether the same holds true for the acetabulum. We have investigated whether the Birmingham hip resurfacing conserves acetabular bone. Between 1998 and 2005, 500 Birmingham hip resurfacings were performed by two surgeons. Between 1996 and 2005 they undertook 700 primary hip replacements, with an
Resurfacing arthroplasty of the hip is being performed more frequently in the United Kingdom. The majority of these patients are younger than 55 years of age, and in this group the key benefits include conservation of femoral bone stock and the potential reduction in the rate of dislocation afforded by the larger resurfacing head. Early aseptic loosening is well recognised in patients younger than 55 years of age, and proponents of resurfacing believe that the improved wear characteristics of the metal-on-metal bearing may improve the long-term survival of this implant. There has been some concern, however, that resurfacing may not be conservative of acetabular bone. We compared a series of 33 consecutive patients who had a hybrid total hip arthroplasty with an
We describe our technique and rationale using
hybrid fixation for primary total hip arthroplasty (THA) at the Hospital
for Special Surgery. Modern
The primary outcome was investigating differences in wear, as measured by femoral head penetration, between cross-linked vitamin E-diffused polyethylene (vE-PE) and cross-linked polyethylene (XLPE) acetabular component liners and between 32 and 36 mm head sizes at the ten-year follow-up. Secondary outcomes included acetabular component migration and patient-reported outcome measures (PROMs) such as the EuroQol five-dimension questionnaire, 36-Item Short-Form Health Survey, Harris Hip Score, and University of California, Los Angeles Activity Scale (UCLA). A single-blinded, multi-arm, 2 × 2 factorial randomized controlled trial was undertaken. Patients were recruited between May 2009 and April 2011. Radiostereometric analyses (RSAs) were performed from baseline to ten years. Of the 220 eligible patients, 116 underwent randomization, and 82 remained at the ten-year follow-up. Eligible patients were randomized into one of four interventions: vE-PE acetabular liner with either 32 or 36 mm femoral head, and XLPE acetabular liner with either 32 or 36 mm femoral head. Parameters were otherwise identical except for acetabular liner material and femoral head size.Aims
Methods
The results of conventional hip replacement in young patients with osteoarthritis have not been encouraging even with improvements in the techniques of fixation and in the bearing surfaces. Modern metal-on-metal hip resurfacing was introduced as a less invasive method of joint reconstruction for this particular group. This is a series of 446 hip resurfacings (384 patients) performed by one of the authors (DJWM) using cemented femoral components and hydroxyapatite-coated
We report a prospective study of the liner-metal interfaces of modular
The use of a porous metal shell supported by two augments with the ‘footing’ technique is one solution to manage Paprosky IIIB acetabular defects in revision total hip arthroplasty. The aim of this study was to assess the medium-term implant survival and radiological and clinical outcomes of this technique. We undertook a retrospective, two-centre series of 39 hips in 39 patients (15 male, 24 female) treated with the ‘footing’ technique for Paprosky IIIB acetabular defects between 2007 and 2020. The median age at the time of surgery was 64.4 years (interquartile range (IQR) 54.4 to 71.0). The median follow-up was 3.9 years (IQR 3.1 to 7.0).Aims
Methods
Polished taper-slip (PTS) cemented stems have an excellent clinical track record and are the most common stem type used in primary total hip arthroplasty (THA) in the UK. Due to low rates of aseptic loosening, they have largely replaced more traditional composite beam (CB) cemented stems. However, there is now emerging evidence from multiple joint registries that PTS stems are associated with higher rates of postoperative periprosthetic femoral fracture (PFF) compared to their CB stem counterparts. The risk of both intraoperative and postoperative PFF remains greater with uncemented stems compared to either of these cemented stem subtypes. PFF continues to be a devastating complication following primary THA and is associated with high complication and mortality rates. Recent efforts have focused on identifying implant-related risk factors for PFF in order to guide preventative strategies, and therefore the purpose of this article is to present the current evidence on the effect of cemented femoral stem design on the risk of PFF. Cite this article:
We have undertaken a prospective, randomised study to compare conservation of acetabular bone after total hip replacement and resurfacing arthroplasty of the hip. We randomly assigned 210 hips to one of the two treatment groups.
The advent of modular porous metal augments has ushered in a new form of treatment for acetabular bone loss. The function of an augment can be seen as reducing the size of a defect or reconstituting the anterosuperior/posteroinferior columns and/or allowing supplementary fixation. Depending on the function of the augment, the surgeon can decide on the sequence of introduction of the hemispherical shell, before or after the augment. Augments should always, however, be used with cement to form a unit with the acetabular component. Given their versatility, augments also allow the use of a hemispherical shell in a position that restores the centre of rotation and biomechanics of the hip. Progressive shedding or the appearance of metal debris is a particular finding with augments and, with other radiological signs of failure, should be recognized on serial radiographs. Mid- to long-term outcomes in studies reporting the use of augments with hemispherical shells in revision total hip arthroplasty have shown rates of survival of > 90%. However, a higher risk of failure has been reported when augments have been used for patients with chronic pelvic discontinuity. Cite this article:
Mechanical impingement of the iliopsoas (IP) tendon accounts for 2% to 6% of persistent postoperative pain after total hip arthroplasty (THA). The most common initiator is anterior acetabular component protrusion, where the anterior margin is not covered by anterior acetabular wall. A CT scan can be used to identify and measure this overhang; however, no threshold exists for determining symptomatic anterior IP impingement due to overhang. A case-control study was conducted in which CT scan measurements were used to define a threshold that differentiates patients with IP impingement from asymptomatic patients after THA. We analyzed the CT scans of 622 patients (758 THAs) between May 2011 and May 2020. From this population, we identified 136 patients with symptoms suggestive of IP impingement. Among them, six were subsequently excluded: three because the diagnosis was refuted intraoperatively, and three because they had another obvious cause of impingement, leaving 130 hips (130 patients) in the study (impingement) group. They were matched to a control group of 138 asymptomatic hips (138 patients) after THA. The anterior acetabular component overhang was measured on an axial CT slice based on anatomical landmarks (orthogonal to the pelvic axis).Aims
Methods
Uncemented implants are now commonly used at reimplantation of a two-stage revision total hip arthoplasty (THA) following periprosthetic joint infection (PJI). However, there is a paucity of data on the performance of the most commonly used uncemented femoral implants – modular fluted tapered (MFT) femoral components – in this setting. This study evaluated implant survival, radiological results, and clinical outcomes in a large cohort of reimplantation THAs using MFT components. We identified 236 reimplantation THAs from a single tertiary care academic institution from September 2000 to September 2020. Two designs of MFT femoral components were used as part of an established two-stage exchange protocol for the treatment of PJI. Mean age at reimplantation was 65 years (SD 11), mean BMI was 32 kg/m2 (SD 7), and 46% (n = 109) were female. Mean follow-up was seven years (SD 4). A competing risk model accounting for death was used.Aims
Methods
We report the clinical and radiological outcome at ten years of 104 primary total hip replacements (100 patients) using the Metasul metal-on-metal bearing. Of these, 52 had a cemented Stuehmer-Weber polyethylene acetabular component with a Metasul bearing and 52 had an
The aim of the study was to investigate whether the primary stability of press-fit acetabular components can be improved by altering the impaction procedure. Three impaction procedures were used to implant acetabular components into human cadaveric acetabula using a powered impaction device. An impaction frequency of 1 Hz until complete component seating served as reference. Overimpaction was simulated by adding ten strokes after complete component seating. High-frequency implantation was performed at 6 Hz. The lever-out moment of the acetabular components was used as measure for primary stability. Permanent bone deformation was assessed by comparison of double micro-CT (µCT) measurements before and after impaction. Acetabular component deformation and impaction forces were recorded, and the extent of bone-implant contact was determined from 3D laser scans.Aims
Methods
The benefit of a dual-mobility acetabular component (DMC) for primary total hip arthroplasties (THAs) is controversial. This study aimed to compare the dislocation and complication rates when using a DMC compared to single-mobility (SM) acetabular component in primary elective THA using data collected at a single centre, and compare the revision rates and survival outcomes in these two groups. Between 2010 and 2019, 2,075 primary THAs using either a cementless DM or SM acetabular component were included. Indications for DMC were patients aged older than 70 years or with high risk of dislocation. All other patients received a SM acetabular component. Exclusion criteria were cemented implants, patients treated for femoral neck fracture, and follow-up of less than one year. In total, 1,940 THAs were analyzed: 1,149 DMC (59.2%) and 791 SM (40.8%). The mean age was 73 years (SD 9.2) in the DMC group and 57 years (SD 12) in the SM group. Complications and revisions have been analyzed retrospectively.Aims
Methods
Registry studies on modified acetabular polyethylene (PE) liner designs are limited. We investigated the influence of standard and modified PE acetabular liner designs on the revision rate for mechanical complications in primary cementless total hip arthroplasty (THA). We analyzed 151,096 primary cementless THAs from the German Arthroplasty Registry (EPRD) between November 2012 and November 2020. Cumulative incidence of revision for mechanical complications for standard and four modified PE liners (lipped, offset, angulated/offset, and angulated) was determined using competing risk analysis at one and seven years. Confounders were investigated with a Cox proportional-hazards model.Aims
Methods
The aim of our study was to investigate the effect of asymmetric crosslinked polyethylene liner use on the risk of revision of cementless and hybrid total hip arthroplasties (THAs). We undertook a registry study combining the National Joint Registry dataset with polyethylene manufacturing characteristics as supplied by the manufacturers. The primary endpoint was revision for any reason. We performed further analyses on other reasons including instability, aseptic loosening, wear, and liner dissociation. The primary analytic approach was Cox proportional hazard regression.Aims
Methods
The aim of this study was to compare the mid-term patient-reported outcome, bone remodelling, and migration of a short stem (Collum Femoris Preserving; CFP) with a conventional uncemented stem (Corail). Of 81 patients who were initially enrolled, 71 were available at five years’ follow-up. The outcomes at two years have previously been reported. The primary outcome measure was the clinical result assessed using the Oxford Hip Score (OHS). Secondary outcomes were the migration of the stem, measured using radiostereometric analysis (RSA), change of bone mineral density (BMD) around the stem, the development of radiolucent lines, and additional patient-reported outcome measures (PROMs).Aims
Methods
One-stage revision hip arthroplasty for periprosthetic joint infection (PJI) has several advantages; however, resection of the proximal femur might be necessary to achieve higher success rates. We investigated the risk factors for resection and re-revisions, and assessed complications and subsequent re-revisions. In this single-centre, case-control study, 57 patients who underwent one-stage revision arthroplasty for PJI of the hip and required resection of the proximal femur between 2009 and 2018 were identified. The control group consisted of 57 patients undergoing one-stage revision without bony resection. Logistic regression analysis was performed to identify any correlation with resection and the risk factors for re-revisions. Rates of all-causes re-revision, reinfection, and instability were compared between groups.Aims
Methods
Bone stock restoration of acetabular bone defects using impaction bone grafting (IBG) in total hip arthroplasty may facilitate future re-revision in the event of failure of the reconstruction. We hypothesized that the acetabular bone defect during re-revision surgery after IBG was smaller than during the previous revision surgery. The clinical and radiological results of re-revisions with repeated use of IBG were also analyzed. In a series of 382 acetabular revisions using IBG and a cemented component, 45 hips (45 patients) that had failed due to aseptic loosening were re-revised between 1992 and 2016. Acetabular bone defects graded according to Paprosky during the first and the re-revision surgery were compared. Clinical and radiological findings were analyzed over time. Survival analysis was performed using a competing risk analysis.Aims
Methods
Improvements in functional results and long-term survival are variable following conversion of hip fusion to total hip arthroplasty (THA) and complications are high. The aim of the study was to analyze the clinical and functional results in patients who underwent conversion of hip fusion to THA using a consistent technique and uncemented implants. A total of 39 hip fusion conversions to THA were undertaken in 38 patients by a single surgeon employing a consistent surgical technique and uncemented implants. Parameters assessed included Harris Hip Score (HHS) for function, range of motion (ROM), leg length discrepancy (LLD), satisfaction, and use of walking aid. Radiographs were reviewed for loosening, subsidence, and heterotopic ossification (HO). Postoperative complications and implant survival were assessed.Aims
Methods
We aimed to investigate if the use of the largest possible cobalt-chromium head articulating with polyethylene acetabular inserts would increase the in vivo wear rate in total hip arthroplasty. In a single-blinded randomized controlled trial, 96 patients (43 females), at a median age of 63 years (interquartile range (IQR) 57 to 69), were allocated to receive either the largest possible modular femoral head (36 mm to 44 mm) in the thinnest possible insert or a standard 32 mm head. All patients received a vitamin E-doped cross-linked polyethylene insert and a cobalt-chromium head. The primary outcome was proximal head penetration measured with radiostereometric analysis (RSA) at two years. Secondary outcomes were volumetric wear, periacetabular radiolucencies, and patient-reported outcomes.Aims
Methods
Various surgical techniques have been described for total hip arthroplasty (THA) in patients with Crowe type III dislocated hips, who have a large acetabular bone defect. The aim of this study was to evaluate the long-term clinical results of patients in whom anatomical reconstruction of the acetabulum was performed using a cemented acetabular component and autologous bone graft from the femoral neck. A total of 22 patients with Crowe type III dislocated hips underwent 28 THAs using bone graft from the femoral neck between 1979 and 2000. A Charnley cemented acetabular component was placed at the level of the true acetabulum after preparation with bone grafting. All patients were female with a mean age at the time of surgery of 54 years (35 to 68). A total of 18 patients (21 THAs) were followed for a mean of 27.2 years (20 to 33) after the operation.Aims
Methods
Vitamin E-infused highly crosslinked polyethylene (VEPE) has been introduced into total hip arthroplasty (THA) with the aim of further improving the wear characteristics of moderately and highly crosslinked polyethylenes (ModXLPE and HXLPE). There are few studies analyzing the outcomes of vitamin E-infused components in cemented arthroplasty, though early acetabular component migration has been reported. The aim of this study was to measure five-year polyethylene wear and acetabular component stability of a cemented VEPE acetabular component compared with a ModXLPE cemented acetabular component. In a prospective randomized controlled trial (RCT), we assessed polyethylene wear and acetabular component stability (primary outcome) with radiostereometric analysis (RSA) in 68 patients with reverse hybrid THA at five years follow-up. Patients were randomized to either a VEPE or a ModXLPE cemented acetabular component.Aims
Methods
Aims. This single-centre observational study aimed to describe the results of extensive bone impaction grafting of the whole acetabular cavity in combination with an
This study aimed to develop a virtual clinic for the purpose of reducing face-to-face orthopaedic consultations. Anonymized experts (hip and knee arthroplasty patients, surgeons, physiotherapists, radiologists, and arthroplasty practitioners) gave feedback via a Delphi Consensus Technique. This consisted of an iterative sequence of online surveys, during which virtual documents, made up of a patient-reported questionnaire, standardized radiology report, and decision-guiding algorithm, were modified until consensus was achieved. We tested the patient-reported questionnaire on seven patients in orthopaedic clinics using a ‘think-aloud’ process to capture difficulties with its completion.Aims
Patients and Methods
The management of acetabular defects at the time of revision hip arthroplasty surgery is a challenge. This study presents the results of a long-term follow-up study of the use of irradiated allograft bone in acetabular reconstruction. Between 1990 and 2000, 123 hips in 110 patients underwent acetabular reconstruction for aseptic loosening, using impaction bone grafting with frozen, irradiated, and morsellized femoral heads and a cemented acetabular component. A total of 55 men and 55 women with a mean age of 64.3 years (26 to 97) at the time of revision surgery are included in this study.Aims
Patients and Methods
The aim of this study was to assess the functional gain achieved following hip resurfacing arthroplasty (HRA). A total of 28 patients (23 male, five female; mean age, 56 years (25 to 73)) awaiting Birmingham HRA volunteered for this prospective gait study, with an age-matched control group of 26 healthy adults (16 male, ten female; mean age, 56 years (33 to 84)). The Oxford Hip Score (OHS) and gait analysis using an instrumented treadmill were used preoperatively and more than two years postoperatively to measure the functional change attributable to the intervention.Aims
Patients and Methods
The use of trabecular metal (TM) shells supported by augments has provided good mid-term results after revision total hip arthroplasty (THA) in patients with a bony defect of the acetabulum. The aim of this study was to assess the long-term implant survivorship and radiological and clinical outcomes after acetabular revision using this technique. Between 2006 and 2010, 60 patients (62 hips) underwent acetabular revision using a combination of a TM shell and augment. A total of 51 patients (53 hips) had complete follow-up at a minimum of seven years and were included in the study. Of these patients, 15 were men (29.4%) and 36 were women (70.6%). Their mean age at the time of revision THA was 64.6 years (28 to 85). Three patients (5.2%) had a Paprosky IIA defect, 13 (24.5%) had a type IIB defect, six (11.3%) had a type IIC defect, 22 (41.5%) had a type IIIA defect, and nine (17%) had a type IIIB defect. Five patients (9.4%) also had pelvic discontinuity.Aims
Patients and Methods
Reconstruction of the acetabulum after failed total hip arthroplasty
(THA) can be a surgical challenge in the presence of severe bone
loss. We report the long-term survival of a porous tantalum revision
acetabular component, its radiological appearance and quality of
life outcomes. We reviewed the results of 46 patients who had undergone revision
of a failed acetabular component with a Paprosky II or III bone
defect and reconstruction with a hemispherical, tantalum acetabular
component, supplementary screws and a cemented polyethylene liner.Aims
Patients and Methods
Patients with neuromuscular imbalance who require
total hip arthroplasty (THA) present particular technical problems
due to altered anatomy, abnormal bone stock, muscular imbalance
and problems of rehabilitation. In this systematic review, we studied articles dealing with THA
in patients with neuromuscular imbalance, published before April
2017. We recorded the demographics of the patients and the type
of neuromuscular pathology, the indication for surgery, surgical
approach, concomitant soft-tissue releases, the type of implant
and bearing, pain and functional outcome as well as complications
and survival. Recent advances in THA technology allow for successful outcomes
in these patients. Our review suggests excellent benefits for pain
relief and good functional outcome might be expected with a modest
risk of complication. Cite this article:
Compared with primary total hip arthroplasty (THA), revision
surgery can be challenging. The cement-in-cement femoral revision
technique involves removing a femoral component from a well-fixed
femoral cement mantle and cementing a new stem into the original
mantle. This technique is widely used and when carried out for the
correct indications, is fast, relatively inexpensive and carries
a reduced short-term risk for the patient compared with the alternative
of removing well-fixed cement. We report the outcomes of this procedure
when two commonly used femoral stems are used. We identified 1179 cement-in-cement stem revisions involving
an Exeter or a Lubinus stem reported to the Swedish Hip Arthroplasty
Register (SHAR) between January 1999 and December 2015. Kaplan-Meier
survival analysis was performed.Aims
Patients and Methods