Advertisement for orthosearch.org.uk
Results 1 - 50 of 137
Results per page:
The Journal of Bone & Joint Surgery British Volume
Vol. 74-B, Issue 3 | Pages 372 - 376
1 May 1992
Curtis M Jinnah R Wilson V Hungerford D

The osseo-integration of an uncemented acetabular component depends on its initial stability. This is usually provided by under-reaming of the acetabulum. We have assessed the fixation of 52 mm porous-coated hemispherical prostheses inserted into cadaveric acetabula under-reamed by 1, 2, 3 and 4 mm. We tested the torsional stability of fixation, after preloading with 686 N in compression, by measuring the torque required to produce 1 degree and 2 degrees of rotation. Under-reaming by 2 mm and 3 mm gave significantly better fixation than 1 mm (p less than 0.01, p less than 0.02). Insertion after under-reaming of 4 mm caused some fractures. To obtain maximum interference fit and optimal implant stability, we recommend the use of an implant 2 mm or 3 mm larger than the last reamer


The Journal of Bone & Joint Surgery British Volume
Vol. 76-B, Issue 4 | Pages 575 - 578
1 Jul 1994
Berry D Barnes C Scott R Cabanela M Poss R

Ten cases are described of catastrophic failure of the polyethylene liner of three different designs of uncemented acetabular component. Failure occurred as a result of either 'wearthrough' to the metal backing, liner fracture or a combination of both, at a mean of 4.6 years after implantation (2 to 7.6). At revision there was metallosis in all hips and osteolysis of the femur or the pelvis in six. Catastrophic failure was seen only in cups with a minimum polyethylene thickness of less than 5 mm


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 5 | Pages 591 - 594
1 May 2007
Lie SA Hallan G Furnes O Havelin LI Engesæter LB

We analysed the results of different strategies in the revision of primary uncemented acetabular components reported to the Norwegian Arthroplasty Register. The aim was to compare the risk of further acetabular revision after isolated liner exchange and complete component revision. The results of exchanging well-fixed components were also compared with those of exchanging loose acetabular components. The period studied was between September 1987 and April 2005. The following groups were compared: group 1, exchange of liner only in 318 hips; group 2, exchange of well-fixed components in 398; and group 3, exchange of loose components in 933. We found that the risk of a further cup revision was lower after revision of well-fixed components (relative risk from a Cox model (RR) = 0.56, 95% confidence interval 0.37% to 0.87%) and loose components (RR = 0.56, 95% confidence interval 0.39% to 0.80%), compared with exchange of the liner in isolation. The most frequent reason for a further acetabular revision was dislocation, accounting for 61 (28%) of the re-revisions. Other reasons for further revision included pain in 27 (12%), loosening in 24 (11%) and infection in 20 (9%). Re-revisions because of pain were less frequent when complete component (fixed or loose) revision was undertaken compared with isolated exchange of the liner (RR = 0.20 (95% confidence interval 0.06% to 0.65%) and RR = 0.10 (95% confidence interval 0.03% to 0.30%), respectively). The risk of further acetabular revision for infection, however, did not differ between the groups. In this study, exchange of the liner only had a higher risk of further cup revision than revision of the complete acetabular component. Our results suggest that the threshold for revising well-fixed components in the case of liner wear and osteolysis should be lowered


The Journal of Bone & Joint Surgery British Volume
Vol. 76-B, Issue 2 | Pages 263 - 266
1 Mar 1994
Hernandez Keating E Faris P Meding J Ritter M

We measured polyethylene wear in 231 porous-coated uncemented acetabular cups. We divided the hips into two groups according to the fixation of the femoral component, by cementing (n = 97) or press-fit (n = 134). Follow-up was from three to five years. The patients in two sub-groups were matched for weight, diagnosis, sex, age and length of follow-up. The linear wear rate of cups articulated with uncemented femoral components (0.22 mm/year) was significantly higher than the wear rate (0.15 mm/year) of cups articulated within cemented femoral components (p < 0.05). These results can be compared with previously reported wear rates of 0.08 mm/year for cemented all-polyethylene cups and 0.11 mm/year for cemented metal-backed cups. The higher wear rates of uncemented arthroplasties could jeopardize the long-term results of this type of hip replacement.


The Journal of Bone & Joint Surgery British Volume
Vol. 87-B, Issue 11 | Pages 1475 - 1479
1 Nov 2005
Goosen JHM Verheyen CCPM Tulp NJA

We investigated the rate of polyethylene wear of a cementless acetabular component at different periods of follow-up in order to test the hypothesis than an irrecoverable deformation process (creep) was followed by an initially low, but gradually increasing wear rate. We studied prospectively 93 uncemented total hip arthroplasties in 83 patients (mean age 50 years (22 to 63)) with a mean follow-up of 8.2 years (3 to 12). We measured the penetration of the femoral head from radiographs taken immediately after surgery at three, six and nine years, or at the latest follow-up.

The median wear rate was 0.17 mm per year in the first three years, a finding which we considered to be caused by creep. Thereafter, the rate of wear declined to 0.07 mm per year (four- to six-year period) and then increased to 0.17 mm per year (seven to nine years) and 0.27 mm per year (more than nine years), which we considered to be a reflection of genuine polyethylene wear. After the nine-year follow-up the wear rates were higher in patients with marked osteolysis. We found no relationship between the inclination angle of the acetabular component or femoral head orientation and the rate of wear. No acetabular component required revision.


The Bone & Joint Journal
Vol. 95-B, Issue 9 | Pages 1178 - 1183
1 Sep 2013
Li H Mao Y Oni JK Dai K Zhu Z

In developmental dysplasia of the hip (DDH), a bone defect is often observed superior to the acetabulum after the reconstruction at the level of the true acetabulum during total hip replacement (THR). However, the essential amount of uncemented acetabular component coverage required for a satisfactory outcome remains controversial. The purpose of this study was to assess the stability and function of acetabular components with a lack of coverage > 30% (31% to 50%). A total of 760 DDH patients underwent THR with acetabular reconstruction at the level of the true floor. Lack of coverage above the acetabular component of > 30% occurred in 56 patients. Intra-operatively, autogenous morcellised bone grafts were used to fill the uncovered portion. Other than two screws inserted through the acetabular shell, no additional structural supports were used in these hips. In all, four patients were lost to follow-up. Therefore, 52 patients (52 hips, 41 women and 11 men) with a mean age of 60.1 years (42 to 78) were available for this study at a mean of 4.8 years (3 to 7). There were no instances of prosthesis revision or marked loosening during the follow-up. The Harris hip score improved from a mean of 40.7 points (sd 12.2) pre-operatively to 91.1 (sd 5.0) at the last follow-up. Radiological analysis with medical imaging software allowed us to calculate the extent of the uncoverage in terms of the uncovered arc of the implant as viewed on the anteroposterior pelvic radiograph. From this we propose that up to 17 mm of lateral undercoverage in the presence of a stable initial implantation in the presence of bone autografting, with an inclination angle of the acetabular component between 40° and 55°, is acceptable. This represents undercoverage of ≤ 50%.

Cite this article: Bone Joint J 2013;95-B:1178–83.


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 8 | Pages 1037 - 1043
1 Aug 2009
Krieg AH Speth BM Ochsner PE

Polyethylene wear of acetabular components is a key factor in the development of periprosthetic osteolysis and wear at the articular surface has been well documented and quantified, but fewer data are available about changes which occur at the backside of the liner.

At revision surgery for loosening of the femoral component we retrieved 35 conventional modular acetabular liners of the same design. Linear and volumetric articular wear, backside volumetric change and the volume of the screw-head indentations were quantified. These volumes, clinical data and the results from radiological Ein Bild Röntgen Analyse migration analysis were used to identify potential factors influencing the volumetric articular wear and backside volumetric change.

The rate of backside volumetric change was found to be 2.8% of the rate of volumetric articular wear and decreased with increasing liner size. Migrated acetabular components showed significantly higher rates of backside volumetric change plus screw-head indentations than those without migration.

The backside volumetric change was at least ten times larger than finite-element simulation had suggested. In a stable acetabular component with well-anchored screws, the amount of backside wear should not cause clinical problems. Impingement of the screw-heads could produce more wear particles than those generated at the liner-shell interface. Because the rate of backside volumetric change is only 2.8% of the rate of volumetric articular wear and since creep is likely to contribute a significant portion to this, the debris generated by wear at the backside of the liner may not be sufficient to create a strong osteolytic response.


The Bone & Joint Journal
Vol. 106-B, Issue 4 | Pages 352 - 358
1 Apr 2024
Wilson JM Trousdale RT Bedard NA Lewallen DG Berry DJ Abdel MP

Aims

Dislocation remains a leading cause of failure following revision total hip arthroplasty (THA). While dual-mobility (DM) bearings have been shown to mitigate this risk, options are limited when retaining or implanting an uncemented shell without modular DM options. In these circumstances, a monoblock DM cup, designed for cementing, can be cemented into an uncemented acetabular shell. The goal of this study was to describe the implant survival, complications, and radiological outcomes of this construct.

Methods

We identified 64 patients (65 hips) who had a single-design cemented DM cup cemented into an uncemented acetabular shell during revision THA between 2018 and 2020 at our institution. Cups were cemented into either uncemented cups designed for liner cementing (n = 48; 74%) or retained (n = 17; 26%) acetabular components. Median outer head diameter was 42 mm. Mean age was 69 years (SD 11), mean BMI was 32 kg/m2 (SD 8), and 52% (n = 34) were female. Survival was assessed using Kaplan-Meier methods. Mean follow-up was two years (SD 0.97).


The Journal of Bone & Joint Surgery British Volume
Vol. 82-B, Issue 2 | Pages 185 - 187
1 Mar 2000
Journeaux SF Morgan DAF Donnelly WJ

We reviewed 120 consecutive primary total hip replacements in 109 patients in whom a Freeman uncemented metal-backed component had been used. Acetabular components were used with a Freeman neck-retaining stem in one of three configurations: cemented smooth stem, uncemented smooth stem or uncemented ridged stem. After a mean follow up of 72 months (62 to 113) there were nine cases of aseptic loosening; a survivorship at eight years of 83%. In addition, more than one-third of the remaining surviving cups showed loosening radiologically. The Freeman acetabular component demonstrated a characteristic pattern of loosening on radiographs. The high incidence of aseptic loosening can be explained by poor design and material failure.


The Bone & Joint Journal
Vol. 106-B, Issue 3 Supple A | Pages 130 - 136
1 Mar 2024
Morlock M Perka C Melsheimer O Kirschbaum SM

Aims. Despite higher rates of revision after total hip arthroplasty (THA) being reported for uncemented stems in patients aged > 75 years, they are frequently used in this age group. Increased mortality after cemented fixation is often used as a justification, but recent data do not confirm this association. The aim of this study was to investigate the influence of the design of the stem and the type of fixation on the rate of revision and immediate postoperative mortality, focusing on the age and sex of the patients. Methods. A total of 333,144 patients with primary osteoarthritis (OA) of the hip who underwent elective THA between November 2012 and September 2022, using uncemented acetabular components without reconstruction shells, from the German arthroplasty registry were included in the study. The revision rates three years postoperatively for four types of stem (uncemented, uncemented with collar, uncemented short, and cemented) were compared within four age groups: < 60 years (Young), between 61 and 70 years (Mid-I), between 71 and 80 years (Mid-II), and aged > 80 years (Old). A noninferiority analysis was performed on the most frequently used designs of stem. Results. The design of the stem was found to have no significant influence on the rate of revision for either sex in the Young group. Uncemented collared stems had a significantly lower rate of revision compared with the other types of stem for females in the Mid-I group. There was a significantly higher rate of revision for uncemented stems in females in the Mid-II group compared with all other types of stem, while in males the rate for uncemented stems was only significantly higher than the rate for cemented stems. Cemented stems had a significantly lower revision rate compared with uncemented and short stems for both sexes in the Old cohort, as did females with collared stems. The rate of immediate postoperative mortality was similar for all types of stem in the Old age group, as were the American Society of Anesthesiologists grades. Conclusion. In patients aged > 80 years, uncemented and short stems had significantly higher revision rates compared with cemented and collared stems, especially in females. The design of the stem and type of fixation have to be analyzed in more detail than only considering cemented and uncemented fixation, in order to further improve the success of THA. Cite this article: Bone Joint J 2024;106-B(3 Supple A):130–136


The Bone & Joint Journal
Vol. 103-B, Issue 12 | Pages 1774 - 1782
1 Dec 2021
Divecha HM O'Neill TW Lunt M Board TN

Aims. The aim of this study was to determine if uncemented acetabular polyethylene (PE) liner geometry, and lip size, influenced the risk of revision for instability or loosening. Methods. A total of 202,511 primary total hip arthroplasties (THAs) with uncemented acetabular components were identified from the National Joint Registry (NJR) dataset between 2003 and 2017. The effect of liner geometry on the risk of revision for instability or loosening was investigated using competing risk regression analyses adjusting for age, sex, American Society of Anesthesiologists grade, indication, side, institution type, surgeon grade, surgical approach, head size, and polyethylene crosslinking. Stratified analyses by surgical approach were performed, including pairwise comparisons of liner geometries. Results. The distribution of liner geometries were neutral (39.4%; 79,822), 10° (34.5%; 69,894), 15° (21.6%; 43,722), offset reorientating (2.8%; 5705), offset neutral (0.9%; 1,767), and 20° (0.8%; 1,601). There were 690 (0.34%) revisions for instability. Compared to neutral liners, the adjusted subhazard ratios of revision for instability were: 10°, 0.64 (p < 0.001); 15°, 0.48 (p < 0.001); and offset reorientating, 1.6 (p = 0.010). No association was found with other geometries. 10° and 15° liners had a time-dependent lower risk of revision for instability within the first 1.2 years. In posterior approaches, 10° and 15° liners had a lower risk of revision for instability, with no significant difference between them. The protective effect of lipped over neutral liners was not observed in laterally approached THAs. There were 604 (0.3%) revisions for loosening, but no association between liner geometry and revision for loosening was found. Conclusion. This registry-based study confirms a lower risk of revision for instability in posterior approach THAs with 10° or 15° lipped liners compared to neutral liners, but no significant difference between these lip sizes. A higher revision risk is seen with offset reorientating liners. The benefit of lipped geometries against revision for instability was not seen in laterally approached THAs. Liner geometry does not seem to influence the risk of revision for loosening. Cite this article: Bone Joint J 2021;103-B(12):1774–1782


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 2 | Pages 196 - 201
1 Feb 2010
Hallan G Dybvik E Furnes O Havelin LI

The Norwegian Arthroplasty Register has shown that several designs of uncemented femoral stems give good or excellent survivorship. The overall findings for uncemented total hip replacement however, have been disappointing because of poor results with the use of metal-backed acetabular components. In this study, we exclusively investigated the medium-to long-term performance of primary uncemented metal-backed acetabular components. A total of 9113 primary uncemented acetabular components were implanted in 7937 patients between 1987 and 2007. These were included in a prospective, population-based observational study. All the implants were modular and metal-backed with ultra-high-molecular-weight polyethylene liners. The femoral heads were made of stainless steel, cobalt-chrome (CoCr) alloy or alumina ceramic. In all, seven different designs of acetabular component were evaluated by the Kaplan-Meier survivorship method and Cox regression analysis. Most acetabular components performed well up to seven years. When the endpoint was revision of the acetabular component because of aseptic loosening, the survival ranged between 87% and 100% at ten years. However, when the endpoint was revision for any reason, the survival estimates were 81% to 92% for the same implants at ten years. Aseptic loosening, wear, osteolysis and dislocation were the main reasons for the relatively poor overall performance of the acetabular components. Prostheses with alumina heads performed slightly better than those with stainless steel or CoCr alloy in subgroups. Whereas most acetabular components performed well at seven years, the survivorship declined with longer follow-up. Fixation was generally good. None of the metal-backed uncemented acetabular components with ultra-high-molecular-weight polyethylene liners in our study had satisfactory long-term results because of high rates of wear, osteolysis, aseptic loosening and dislocation


The Bone & Joint Journal
Vol. 102-B, Issue 4 | Pages 414 - 422
1 Apr 2020
Gwynne-Jones DP Gray AR

Aims. To compare long-term survival of all-cemented and hybrid total hip arthroplasty (THA) using the Exeter Universal stem. Methods. Details of 1,086 THAs performed between 1999 and 2005 using the Exeter stem and either a cemented (632) or uncemented acetabular component (454) were collected from local records and the New Zealand Joint Registry. A competing risks regression survival analysis was performed with death as the competing risk with adjustments made for age, sex, approach, and bearing. Results. There were 61 revisions (9.7%; 0.82 revisions/100 observed component years, (OCYs)) in the all-cemented group and 18 (4.0%; 0.30/100 OCYs) in the hybrid group. The cumulative incidence of revision at 18 years was 12.1% for cemented and 5.2% for hybrids. There was a significantly greater risk of revision for all-cemented compared with hybrids (unadjusted sub-hazard ratio (SHR) 2.44; p = 0.001), and of revision for loosening, wear, or osteolysis (unadjusted SHR 3.77; p < 0.001). After adjustment, the increased risk of all-cause revision did not reach significance at age 70 years and above. The advantage for revision for loosening, wear, and osteolysis remained at all ages. Conclusion. This study supports the use of uncemented acetabular fixation when used in combination with the Exeter stem with improved survivorship for revision for aseptic loosening, wear, and osteolysis at all ages and for all-cause revision in patients less than 70 years. Cite this article: Bone Joint J 2020;102-B(4):414–422


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 11 | Pages 1422 - 1427
1 Nov 2008
Utting MR Raghuvanshi M Amirfeyz R Blom AW Learmonth ID Bannister GC

We have reviewed 70 Harris-Galante uncemented acetabular components implanted as hybrid hip replacements with cemented stems between 1991 and 1995 in 53 patients whose mean age was 40 years (19 to 49). The mean follow-up was for 13.6 years (12 to 16) with no loss to follow-up. We assessed the patients both clinically and radiologically. The mean Oxford hip score was 20 (12 to 46) and the mean Harris hip score 81 (37 to 100) at the final review. Radiologically, 27 hips (39%) had femoral osteolysis, 13 (19%) acetabular osteolysis, and 31 (44%) radiolucent lines around the acetabular component. Kaplan-Meier survival curves were constructed for the outcomes of revision of the acetabular component, revision of the component and polyethylene liner, and impending revision for progressive osteolysis. The cumulative survival for revision of the acetabular component was 94% (95% confidence interval 88.4 to 99.7), for the component and liner 84% (95% confidence interval 74.5 to 93.5) and for impending revision 55.3% (95% confidence interval 40.6 to 70) at 16 years. Uncemented acetabular components with polyethylene liners undergo silent lysis and merit regular long-term radiological review


The Bone & Joint Journal
Vol. 105-B, Issue 7 | Pages 768 - 774
1 Jul 2023
Wooster BM Kennedy NI Dugdale EM Sierra RJ Perry KI Berry DJ Abdel MP

Aims. Contemporary outcomes of primary total hip arthroplasties (THAs) with highly cross-linked polyethylene (HXLPE) liners in patients with inflammatory arthritis have not been well studied. This study examined the implant survivorship, complications, radiological results, and clinical outcomes of THA in patients with inflammatory arthritis. Methods. We identified 418 hips (350 patients) with a primary diagnosis of inflammatory arthritis who underwent primary THA with HXLPE liners from January 2000 to December 2017. Of these hips, 68% had rheumatoid arthritis (n = 286), 13% ankylosing spondylitis (n = 53), 7% juvenile rheumatoid arthritis (n = 29), 6% psoriatic arthritis (n = 24), 5% systemic lupus erythematosus (n = 23), and 1% scleroderma (n = 3). Mean age was 58 years (SD 14.8), 66.3% were female (n = 277), and mean BMI was 29 kg/m. 2. (SD 7). Uncemented femoral components were used in 77% of cases (n = 320). Uncemented acetabular components were used in all patients. Competing risk analysis was used accounting for death. Mean follow-up was 4.5 years (2 to 18). Results. The ten-year cumulative incidence of any revision was 3%, and was highest in psoriatic arthritis patients (16%). The most common indications for the 15 revisions were dislocations (n = 8) and periprosthetic joint infections (PJI; n = 4, all on disease-modifying antirheumatic drugs (DMARDs)). The ten-year cumulative incidence of reoperation was 6.1%, with the most common indications being wound infections (six cases, four on DMARDs) and postoperative periprosthetic femur fractures (two cases, both uncemented femoral components). The ten-year cumulative incidence of complications not requiring reoperation was 13.1%, with the most common being intraoperative periprosthetic femur fracture (15 cases, 14 uncemented femoral components; p = 0.13). Radiological evidence of early femoral component subsidence was observed in six cases (all uncemented). Only one femoral component ultimately developed aseptic loosening. Harris Hip Scores substantially improved (p < 0.001). Conclusion. Contemporary primary THAs with HXLPE in patients with inflammatory arthritis had excellent survivorship and good functional outcomes regardless of fixation method. Dislocation, PJI, and periprosthetic fracture were the most common complications in this cohort with inflammatory arthritis. Cite this article: Bone Joint J 2023;105-B(7):768–774


The Bone & Joint Journal
Vol. 103-B, Issue 5 | Pages 872 - 880
1 May 2021
Young PS Macarico DT Silverwood RK Farhan-Alanie OM Mohammed A Periasamy K Nicol A Meek RMD

Aims. Uncemented metal acetabular components show good osseointegration, but material stiffness causes stress shielding and retroacetabular bone loss. Cemented monoblock polyethylene components load more physiologically; however, the cement bone interface can suffer fibrous encapsulation and loosening. It was hypothesized that an uncemented titanium-sintered monoblock polyethylene component may offer the optimum combination of osseointegration and anatomical loading. Methods. A total of 38 patients were prospectively enrolled and received an uncemented monoblock polyethylene acetabular (pressfit) component. This single cohort was then retrospectively compared with previously reported randomized cohorts of cemented monoblock (cemented) and trabecular metal (trabecular) acetabular implants. The primary outcome measure was periprosthetic bone density using dual-energy x-ray absorptiometry over two years. Secondary outcomes included radiological and clinical analysis. Results. Although there were differences in the number of males and females in each group, no significant sex bias was noted (p = 0.080). Furthermore, there was no significant difference in age (p = 0.910) or baseline lumbar bone mineral density (BMD) (p = 0.998) found between any of the groups (pressfit, cemented, or trabecular). The pressfit implant initially behaved like the trabecular component with an immediate fall in BMD in the inferior and medial regions, with preserved BMD laterally, suggesting lateral rim loading. However, the pressfit component subsequently showed a reversal in BMD medially with recovery back towards baseline, and a continued rise in lateral BMD. This would suggest that the pressfit component begins to reload the medial bone over time, more akin to the cemented component. Analysis of postoperative radiographs revealed no pressfit component subsidence or movement up to two years postoperatively (100% interobserver reliability). Medial defects seen immediately postoperatively in five cases had completely resolved by two years in four patients. Conclusion. Initially, the uncemented monoblock component behaved similarly to the rigid trabecular metal component with lateral rim loading; however, over two years this changed to more closely resemble the loading pattern of a cemented polyethylene component with increasing medial pelvic loading. This indicates that the uncemented monoblock acetabular component may result in optimized fixation and preservation of retroacetabular bone stock. Cite this article: Bone Joint J 2021;103-B(5):872–880


The Bone & Joint Journal
Vol. 104-B, Issue 7 | Pages 852 - 858
1 Jul 2022
Grothe T Günther K Hartmann A Blum S Haselhoff R Goronzy J

Aims. Head-taper corrosion is a cause of failure in total hip arthroplasty (THA). Recent reports have described an increasing number of V40 taper failures with adverse local tissue reaction (ALTR). However, the real incidence of V40 taper damage and its cause remain unknown. The aim of this study was to evaluate the long-term incidence of ALTR in a consecutive series of THAs using a V40 taper and identify potentially related factors. Methods. Between January 2006 and June 2007, a total of 121 patients underwent THA using either an uncemented (Accolade I, made of Ti. 12. Mo. 6. Zr. 2. Fe; Stryker, USA) or a cemented (ABG II, made of cobalt-chrome-molybdenum (CoCrMo); Stryker) femoral component, both with a V40 taper (Stryker). Uncemented acetabular components (Trident; Stryker) with crosslinked polyethylene liners and CoCr femoral heads of 36 mm diameter were used in all patients. At a mean folllow-up of 10.8 years (SD 1.1), 94 patients (79%) were eligible for follow-up (six patients had already undergone a revision, 15 had died, and six were lost to follow-up). A total of 85 THAs in 80 patients (mean age 61 years (24 to 75); 47 (56%) were female) underwent clinical and radiological evaluation, including the measurement of whole blood levels of cobalt and chrome. Metal artifact reduction sequence MRI scans of the hip were performed in 71 patients. Results. A total of 20 ALTRs were identified on MRI, with an incidence of 26%. Patients with ALTR had significantly higher median Co levels compared with those without ALTR (2.96 μg/l (interquartile range (IQR) 1.35 to 4.98) vs 1.44 μg/l (IQR 0.79 to 2.5); p = 0.019). Radiological evidence of osteolysis was also significantly associated with ALTR (p = 0.014). Median Cr levels were not significantly higher in those with ALTR compared with those without one (0.97 μg/l (IQR 0.72 to 1.9) v 0.67 μg/l (IQR 0.5 to 1.19; p = 0.080). BMI, sex, age, type of femoral component, head length, the inclination of the acetabular component, and heterotopic ossification formation showed no significant relationship with ALTR. Conclusion. Due to the high incidence of local ALTR in our cohort after more than ten years postoperatively, we recommend regular follow-up investigation even in asymptomatic patients with V40 taper and metal heads. As cobalt levels correlate with ALTR occurrence, routine metal ion screening and consecutive MRI investigation upon elevation could be discussed. Cite this article: Bone Joint J 2022;104-B(7):852–858


The Bone & Joint Journal
Vol. 96-B, Issue 7 | Pages 884 - 888
1 Jul 2014
Insull PJ Cobbett H Frampton CM Munro JT

We compared the rate of revision for instability after total hip replacement (THR) when lipped and non-lipped acetabular liners were used. We hypothesised that the use of a lipped liner in a modular uncemented acetabular component reduces the risk of revision for instability after primary THR. Using data from the New Zealand Joint Registry, we found that the use of a lipped liner was associated with a significantly decreased rate of revision for instability and for all other indications. Adjusting for the size of the femoral head, the surgical approach and the age and gender of the patient, this difference remained strongly significant (p < 0.001). We conclude that evidence from the New Zealand registry suggests that the use of lipped liners with modular uncemented acetabular components is associated with a decreased rate of revision for instability after primary THR. Cite this article: Bone Joint J 2014;96-B:884–8


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 3 | Pages 319 - 323
1 Mar 2008
Moonot P Singh PJ Cronin MD Kalairajah YE Kavanagh TG Field RE

Hip resurfacing is a bone-conserving procedure with respect to proximal femoral resection, but there is debate in the literature as to whether the same holds true for the acetabulum. We have investigated whether the Birmingham hip resurfacing conserves acetabular bone. Between 1998 and 2005, 500 Birmingham hip resurfacings were performed by two surgeons. Between 1996 and 2005 they undertook 700 primary hip replacements, with an uncemented acetabular component. These patients formed the clinical material to compare acetabular component sizing. The Birmingham hip resurfacing group comprised 350 hips in men and 150 hips in women. The uncemented total hip replacement group comprised 236 hips in men and 464 hips in women. Age- and gender-matched analysis of a cohort of patients for the sizes of the acetabular components required for the two types of replacement was also undertaken. Additionally, an analysis of the sizes of the components used by each surgeon was performed. For age-matched women, the mean outside diameter of the Birmingham hip resurfacing acetabular components was 2.03 mm less than that of the acetabular components in the uncemented total hip replacements (p < 0.0001). In similarly matched men there was no significant difference (p = 0.77). A significant difference was also found between the size of acetabular components used by the two surgeons for Birmingham hip resurfacing for both men (p = 0.0015) and women (p = 0.001). In contrast, no significant difference was found between the size of acetabular components used by the two surgeons for uncemented total hip replacement in either men or women (p = 0.06 and p = 0.14, respectively). This suggests that variations in acetabular preparation also influence acetabular component size in hip resurfacing


The Journal of Bone & Joint Surgery British Volume
Vol. 88-B, Issue 1 | Pages 31 - 34
1 Jan 2006
Loughead JM Starks I Chesney D Matthews JNS McCaskie AW Holland JP

Resurfacing arthroplasty of the hip is being performed more frequently in the United Kingdom. The majority of these patients are younger than 55 years of age, and in this group the key benefits include conservation of femoral bone stock and the potential reduction in the rate of dislocation afforded by the larger resurfacing head. Early aseptic loosening is well recognised in patients younger than 55 years of age, and proponents of resurfacing believe that the improved wear characteristics of the metal-on-metal bearing may improve the long-term survival of this implant. There has been some concern, however, that resurfacing may not be conservative of acetabular bone. We compared a series of 33 consecutive patients who had a hybrid total hip arthroplasty with an uncemented acetabular component and a cemented femoral implant, with 35 patients undergoing a Birmingham hip resurfacing arthroplasty. We compared the diameter of the implanted acetabulum in both groups and, because they were not directly comparable, we corrected for patient size by measuring the diameter of the contralateral femoral head. The data were analysed using unpaired t-tests and analysis of covariance. There was a significantly larger acetabulum in the Birmingham arthroplasty group (mean diameter 56.6 mm vs 52.0 mm; p < 0.001). However, this group had a significantly larger femoral head diameter on the contralateral side (p = 0.03). Analysis of covariance revealed a significant difference between the mean size of the acetabular component implanted in the two operations. The greatest difference in the size of acetabulum was in those patients with a larger diameter of the femoral head. This study shows that more bone is removed from the acetabulum in hip resurfacing than during hybrid total hip arthroplasty, a difference which is most marked in larger patients


The Bone & Joint Journal
Vol. 98-B, Issue 1_Supple_A | Pages 54 - 59
1 Jan 2016
González Della Valle A Sharrock N Barlow M Caceres L Go G Salvati EA

We describe our technique and rationale using hybrid fixation for primary total hip arthroplasty (THA) at the Hospital for Special Surgery. Modern uncemented acetabular components have few screw holes, or no holes, polished inner surfaces, improved locking mechanisms, and maximised thickness and shell-liner conformity. Uncemented sockets can be combined with highly cross-linked polyethylene liners, which have demonstrated very low wear and osteolysis rates after ten to 15 years of implantation. The results of cement fixation with a smooth or polished surface finished stem have been excellent, virtually eliminating complications seen with cementless fixation like peri-operative femoral fractures and thigh pain. Although mid-term results of modern cementless stems are encouraging, the long-term data do not show reduced revision rates for cementless stems compared with cemented smooth stems. In this paper we review the conduct of a hybrid THA, with emphasis on pre-operative planning, surgical technique, hypotensive epidural anaesthesia, and intra-operative physiology. Cite this article: Bone Joint J 2016;98-B(1 Suppl A):54–9


The Bone & Joint Journal
Vol. 105-B, Issue 10 | Pages 1052 - 1059
1 Oct 2023
El-Sahoury JAN Kjærgaard K Ovesen O Hofbauer C Overgaard S Ding M

Aims

The primary outcome was investigating differences in wear, as measured by femoral head penetration, between cross-linked vitamin E-diffused polyethylene (vE-PE) and cross-linked polyethylene (XLPE) acetabular component liners and between 32 and 36 mm head sizes at the ten-year follow-up. Secondary outcomes included acetabular component migration and patient-reported outcome measures (PROMs) such as the EuroQol five-dimension questionnaire, 36-Item Short-Form Health Survey, Harris Hip Score, and University of California, Los Angeles Activity Scale (UCLA).

Methods

A single-blinded, multi-arm, 2 × 2 factorial randomized controlled trial was undertaken. Patients were recruited between May 2009 and April 2011. Radiostereometric analyses (RSAs) were performed from baseline to ten years. Of the 220 eligible patients, 116 underwent randomization, and 82 remained at the ten-year follow-up. Eligible patients were randomized into one of four interventions: vE-PE acetabular liner with either 32 or 36 mm femoral head, and XLPE acetabular liner with either 32 or 36 mm femoral head. Parameters were otherwise identical except for acetabular liner material and femoral head size.


The Journal of Bone & Joint Surgery British Volume
Vol. 86-B, Issue 2 | Pages 177 - 184
1 Mar 2004
Daniel J Pynsent PB McMinn DJW

The results of conventional hip replacement in young patients with osteoarthritis have not been encouraging even with improvements in the techniques of fixation and in the bearing surfaces. Modern metal-on-metal hip resurfacing was introduced as a less invasive method of joint reconstruction for this particular group. This is a series of 446 hip resurfacings (384 patients) performed by one of the authors (DJWM) using cemented femoral components and hydroxyapatite-coated uncemented acetabular components with a maximum follow-up of 8.2 years (mean 3.3). Their survival rate, Oxford hip scores and activity levels are reviewed. Six patients died due to unrelated causes. There was one revision (0.02%) out of 440 hips. The mean Oxford score of the surviving 439 hips is 13.5. None of the patients were told to change their activities at work or leisure; 31% of the men with unilateral resurfacings and 28% with bilateral resurfacings were involved in jobs that they considered heavy or moderately heavy; 92% of men with unilateral hip resurfacings and 87% of the whole group participate in leisure-time sporting activity. The extremely low rate of failure in spite of the resumption of high level occupational and leisure activities provides early evidence of the suitability of this procedure for young and active patients with arthritis


The Journal of Bone & Joint Surgery British Volume
Vol. 76-B, Issue 4 | Pages 568 - 574
1 Jul 1994
Huk O Bansal M Betts F Rimnac C Lieberman Huo M Salvati E

We report a prospective study of the liner-metal interfaces of modular uncemented acetabular components as sources of debris. We collected the pseudomembrane from the screw-cup junction and the empty screw holes of the metal backing of 19 acetabula after an average implantation of 22 months. Associated osteolytic lesions were separately collected in two cases. The back surfaces of the liners and the screws were examined for damage, and some liners were scanned by electron microscopy. The tissues were studied histologically and by atomic absorption spectrophotometry to measure titanium content. The pseudomembrane from the screw-cup junction contained polyethylene debris in seven specimens and metal debris in ten. The material from empty screw holes was necrotic tissue or dense fibroconnective tissue with a proliferative histiocytic infiltrate and foreign-body giant-cell reaction. It contained polyethylene debris in 14 cases and metal in five. The two acetabular osteolytic lesions also showed a foreign-body giant-cell reaction to particulate debris. The average titanium levels in pseudomembranes from the screw-cup junction and the empty screw holes were 959 micrograms/g (48 to 11,900) and 74 micrograms/g (0.72 to 331) respectively. The tissue from the two lytic lesions showed average titanium levels of 139 and 147 micrograms/g respectively. The back surfaces of the PE liners showed surface deformation, burnishing, and embedded metal debris. All 30 retrieved screws demonstrated fretting at the base of the head and on the proximal shaft. Non-articular modular junctions create new interfaces for the generation of particulate debris, which may cause granulomatous reaction


The Bone & Joint Journal
Vol. 106-B, Issue 5 Supple B | Pages 54 - 58
1 May 2024
Wassilew GI Zimmerer A Fischer M Nonnenmacher L O'Hara L Hube R

Aims

The use of a porous metal shell supported by two augments with the ‘footing’ technique is one solution to manage Paprosky IIIB acetabular defects in revision total hip arthroplasty. The aim of this study was to assess the medium-term implant survival and radiological and clinical outcomes of this technique.

Methods

We undertook a retrospective, two-centre series of 39 hips in 39 patients (15 male, 24 female) treated with the ‘footing’ technique for Paprosky IIIB acetabular defects between 2007 and 2020. The median age at the time of surgery was 64.4 years (interquartile range (IQR) 54.4 to 71.0). The median follow-up was 3.9 years (IQR 3.1 to 7.0).


The Journal of Bone & Joint Surgery British Volume
Vol. 88-B, Issue 8 | Pages 997 - 1002
1 Aug 2006
Vendittoli P Lavigne M Girard J Roy AG

We have undertaken a prospective, randomised study to compare conservation of acetabular bone after total hip replacement and resurfacing arthroplasty of the hip. We randomly assigned 210 hips to one of the two treatment groups. Uncemented, press-fit acetabular components were used for both. No significant difference was found in the mean diameter of acetabular implant inserted in the groups (54.74 mm for total hip replacement and 54.90 mm for resurfacing arthroplasty). In seven resurfacing procedures (6.8%), the surgeon used a larger size of component in order to match the corresponding diameter of the femoral component. With resurfacing arthroplasty, conservation of bone is clearly advantageous on the femoral side. Our study has shown that, with a specific design of acetabular implant and by following a careful surgical technique, removal of bone on the acetabular side is comparable with that of total hip replacement


The Bone & Joint Journal
Vol. 106-B, Issue 1 | Pages 11 - 15
1 Jan 2024
Jain S Lamb JN Pandit H

Polished taper-slip (PTS) cemented stems have an excellent clinical track record and are the most common stem type used in primary total hip arthroplasty (THA) in the UK. Due to low rates of aseptic loosening, they have largely replaced more traditional composite beam (CB) cemented stems. However, there is now emerging evidence from multiple joint registries that PTS stems are associated with higher rates of postoperative periprosthetic femoral fracture (PFF) compared to their CB stem counterparts. The risk of both intraoperative and postoperative PFF remains greater with uncemented stems compared to either of these cemented stem subtypes. PFF continues to be a devastating complication following primary THA and is associated with high complication and mortality rates. Recent efforts have focused on identifying implant-related risk factors for PFF in order to guide preventative strategies, and therefore the purpose of this article is to present the current evidence on the effect of cemented femoral stem design on the risk of PFF.

Cite this article: Bone Joint J 2024;106-B(1):11–15.


The Bone & Joint Journal
Vol. 106-B, Issue 4 | Pages 312 - 318
1 Apr 2024
Sheth NP Jones SA Sanghavi SA Manktelow A

The advent of modular porous metal augments has ushered in a new form of treatment for acetabular bone loss. The function of an augment can be seen as reducing the size of a defect or reconstituting the anterosuperior/posteroinferior columns and/or allowing supplementary fixation. Depending on the function of the augment, the surgeon can decide on the sequence of introduction of the hemispherical shell, before or after the augment. Augments should always, however, be used with cement to form a unit with the acetabular component. Given their versatility, augments also allow the use of a hemispherical shell in a position that restores the centre of rotation and biomechanics of the hip. Progressive shedding or the appearance of metal debris is a particular finding with augments and, with other radiological signs of failure, should be recognized on serial radiographs. Mid- to long-term outcomes in studies reporting the use of augments with hemispherical shells in revision total hip arthroplasty have shown rates of survival of > 90%. However, a higher risk of failure has been reported when augments have been used for patients with chronic pelvic discontinuity.

Cite this article: Bone Joint J 2024;106-B(4):312–318.


The Bone & Joint Journal
Vol. 106-B, Issue 3 Supple A | Pages 97 - 103
1 Mar 2024
Baujard A Martinot P Demondion X Dartus J Faure PA Girard J Migaud H

Aims

Mechanical impingement of the iliopsoas (IP) tendon accounts for 2% to 6% of persistent postoperative pain after total hip arthroplasty (THA). The most common initiator is anterior acetabular component protrusion, where the anterior margin is not covered by anterior acetabular wall. A CT scan can be used to identify and measure this overhang; however, no threshold exists for determining symptomatic anterior IP impingement due to overhang. A case-control study was conducted in which CT scan measurements were used to define a threshold that differentiates patients with IP impingement from asymptomatic patients after THA.

Methods

We analyzed the CT scans of 622 patients (758 THAs) between May 2011 and May 2020. From this population, we identified 136 patients with symptoms suggestive of IP impingement. Among them, six were subsequently excluded: three because the diagnosis was refuted intraoperatively, and three because they had another obvious cause of impingement, leaving 130 hips (130 patients) in the study (impingement) group. They were matched to a control group of 138 asymptomatic hips (138 patients) after THA. The anterior acetabular component overhang was measured on an axial CT slice based on anatomical landmarks (orthogonal to the pelvic axis).


The Bone & Joint Journal
Vol. 106-B, Issue 5 Supple B | Pages 125 - 132
1 May 2024
Carender CN Perry KI Sierra RJ Trousdale RT Berry DJ Abdel MP

Aims

Uncemented implants are now commonly used at reimplantation of a two-stage revision total hip arthoplasty (THA) following periprosthetic joint infection (PJI). However, there is a paucity of data on the performance of the most commonly used uncemented femoral implants – modular fluted tapered (MFT) femoral components – in this setting. This study evaluated implant survival, radiological results, and clinical outcomes in a large cohort of reimplantation THAs using MFT components.

Methods

We identified 236 reimplantation THAs from a single tertiary care academic institution from September 2000 to September 2020. Two designs of MFT femoral components were used as part of an established two-stage exchange protocol for the treatment of PJI. Mean age at reimplantation was 65 years (SD 11), mean BMI was 32 kg/m2 (SD 7), and 46% (n = 109) were female. Mean follow-up was seven years (SD 4). A competing risk model accounting for death was used.


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 10 | Pages 1278 - 1283
1 Oct 2008
Eswaramoorthy V Moonot P Kalairajah Y Biant LC Field RE

We report the clinical and radiological outcome at ten years of 104 primary total hip replacements (100 patients) using the Metasul metal-on-metal bearing. Of these, 52 had a cemented Stuehmer-Weber polyethylene acetabular component with a Metasul bearing and 52 had an uncemented Allofit acetabular component with a Metasul liner. A total of 15 patients (16 hips) died before their follow-up at ten years and three were lost to follow-up. The study group therefore comprised 82 patients (85 hips). The mean Oxford score at ten years was 20.7 (12 to 42). Six of 85 hips required revision surgery. One was performed because of infection, one for aseptic loosening of the acetabular component and four because of unexplained pain. Histological examination showed an aseptic lymphocytic vasculitis associated lesion-type tissue response in two of these. Continued follow-up is advocated in order to monitor the long-term performance of the Metasul bearing and tissue responses to metal debris


The Bone & Joint Journal
Vol. 105-B, Issue 3 | Pages 261 - 268
1 Mar 2023
Ruhr M Huber G Niki Y Lohner L Ondruschka B Morlock MM

Aims

The aim of the study was to investigate whether the primary stability of press-fit acetabular components can be improved by altering the impaction procedure.

Methods

Three impaction procedures were used to implant acetabular components into human cadaveric acetabula using a powered impaction device. An impaction frequency of 1 Hz until complete component seating served as reference. Overimpaction was simulated by adding ten strokes after complete component seating. High-frequency implantation was performed at 6 Hz. The lever-out moment of the acetabular components was used as measure for primary stability. Permanent bone deformation was assessed by comparison of double micro-CT (µCT) measurements before and after impaction. Acetabular component deformation and impaction forces were recorded, and the extent of bone-implant contact was determined from 3D laser scans.


The Bone & Joint Journal
Vol. 106-B, Issue 3 Supple A | Pages 81 - 88
1 Mar 2024
Lustig S Cotte M Foissey C Asirvatham RD Servien E Batailler C

Aims

The benefit of a dual-mobility acetabular component (DMC) for primary total hip arthroplasties (THAs) is controversial. This study aimed to compare the dislocation and complication rates when using a DMC compared to single-mobility (SM) acetabular component in primary elective THA using data collected at a single centre, and compare the revision rates and survival outcomes in these two groups.

Methods

Between 2010 and 2019, 2,075 primary THAs using either a cementless DM or SM acetabular component were included. Indications for DMC were patients aged older than 70 years or with high risk of dislocation. All other patients received a SM acetabular component. Exclusion criteria were cemented implants, patients treated for femoral neck fracture, and follow-up of less than one year. In total, 1,940 THAs were analyzed: 1,149 DMC (59.2%) and 791 SM (40.8%). The mean age was 73 years (SD 9.2) in the DMC group and 57 years (SD 12) in the SM group. Complications and revisions have been analyzed retrospectively.


The Bone & Joint Journal
Vol. 104-B, Issue 7 | Pages 801 - 810
1 Jul 2022
Krull P Steinbrück A Grimberg AW Melsheimer O Morlock M Perka C

Aims

Registry studies on modified acetabular polyethylene (PE) liner designs are limited. We investigated the influence of standard and modified PE acetabular liner designs on the revision rate for mechanical complications in primary cementless total hip arthroplasty (THA).

Methods

We analyzed 151,096 primary cementless THAs from the German Arthroplasty Registry (EPRD) between November 2012 and November 2020. Cumulative incidence of revision for mechanical complications for standard and four modified PE liners (lipped, offset, angulated/offset, and angulated) was determined using competing risk analysis at one and seven years. Confounders were investigated with a Cox proportional-hazards model.


The Bone & Joint Journal
Vol. 103-B, Issue 9 | Pages 1479 - 1487
1 Sep 2021
Davis ET Pagkalos J Kopjar B

Aims

The aim of our study was to investigate the effect of asymmetric crosslinked polyethylene liner use on the risk of revision of cementless and hybrid total hip arthroplasties (THAs).

Methods

We undertook a registry study combining the National Joint Registry dataset with polyethylene manufacturing characteristics as supplied by the manufacturers. The primary endpoint was revision for any reason. We performed further analyses on other reasons including instability, aseptic loosening, wear, and liner dissociation. The primary analytic approach was Cox proportional hazard regression.


Aims

The aim of this study was to compare the mid-term patient-reported outcome, bone remodelling, and migration of a short stem (Collum Femoris Preserving; CFP) with a conventional uncemented stem (Corail).

Methods

Of 81 patients who were initially enrolled, 71 were available at five years’ follow-up. The outcomes at two years have previously been reported. The primary outcome measure was the clinical result assessed using the Oxford Hip Score (OHS). Secondary outcomes were the migration of the stem, measured using radiostereometric analysis (RSA), change of bone mineral density (BMD) around the stem, the development of radiolucent lines, and additional patient-reported outcome measures (PROMs).


The Bone & Joint Journal
Vol. 103-B, Issue 11 | Pages 1678 - 1685
1 Nov 2021
Abdelaziz H Schröder M Shum Tien C Ibrahim K Gehrke T Salber J Citak M

Aims

One-stage revision hip arthroplasty for periprosthetic joint infection (PJI) has several advantages; however, resection of the proximal femur might be necessary to achieve higher success rates. We investigated the risk factors for resection and re-revisions, and assessed complications and subsequent re-revisions.

Methods

In this single-centre, case-control study, 57 patients who underwent one-stage revision arthroplasty for PJI of the hip and required resection of the proximal femur between 2009 and 2018 were identified. The control group consisted of 57 patients undergoing one-stage revision without bony resection. Logistic regression analysis was performed to identify any correlation with resection and the risk factors for re-revisions. Rates of all-causes re-revision, reinfection, and instability were compared between groups.


The Bone & Joint Journal
Vol. 103-B, Issue 3 | Pages 492 - 499
1 Mar 2021
Garcia-Rey E Saldaña L Garcia-Cimbrelo E

Aims

Bone stock restoration of acetabular bone defects using impaction bone grafting (IBG) in total hip arthroplasty may facilitate future re-revision in the event of failure of the reconstruction. We hypothesized that the acetabular bone defect during re-revision surgery after IBG was smaller than during the previous revision surgery. The clinical and radiological results of re-revisions with repeated use of IBG were also analyzed.

Methods

In a series of 382 acetabular revisions using IBG and a cemented component, 45 hips (45 patients) that had failed due to aseptic loosening were re-revised between 1992 and 2016. Acetabular bone defects graded according to Paprosky during the first and the re-revision surgery were compared. Clinical and radiological findings were analyzed over time. Survival analysis was performed using a competing risk analysis.


The Bone & Joint Journal
Vol. 103-B, Issue 7 Supple B | Pages 129 - 134
1 Jul 2021
Ayekoloye CI Abu Qa'oud M Radi M Leon SA Kuzyk P Safir O Gross AE

Aims

Improvements in functional results and long-term survival are variable following conversion of hip fusion to total hip arthroplasty (THA) and complications are high. The aim of the study was to analyze the clinical and functional results in patients who underwent conversion of hip fusion to THA using a consistent technique and uncemented implants.

Methods

A total of 39 hip fusion conversions to THA were undertaken in 38 patients by a single surgeon employing a consistent surgical technique and uncemented implants. Parameters assessed included Harris Hip Score (HHS) for function, range of motion (ROM), leg length discrepancy (LLD), satisfaction, and use of walking aid. Radiographs were reviewed for loosening, subsidence, and heterotopic ossification (HO). Postoperative complications and implant survival were assessed.


The Bone & Joint Journal
Vol. 103-B, Issue 7 | Pages 1206 - 1214
1 Jul 2021
Tsikandylakis G Mortensen KRL Gromov K Mohaddes M Malchau H Troelsen A

Aims

We aimed to investigate if the use of the largest possible cobalt-chromium head articulating with polyethylene acetabular inserts would increase the in vivo wear rate in total hip arthroplasty.

Methods

In a single-blinded randomized controlled trial, 96 patients (43 females), at a median age of 63 years (interquartile range (IQR) 57 to 69), were allocated to receive either the largest possible modular femoral head (36 mm to 44 mm) in the thinnest possible insert or a standard 32 mm head. All patients received a vitamin E-doped cross-linked polyethylene insert and a cobalt-chromium head. The primary outcome was proximal head penetration measured with radiostereometric analysis (RSA) at two years. Secondary outcomes were volumetric wear, periacetabular radiolucencies, and patient-reported outcomes.


The Bone & Joint Journal
Vol. 103-B, Issue 2 | Pages 299 - 304
1 Feb 2021
Goto E Umeda H Otsubo M Teranishi T

Aims

Various surgical techniques have been described for total hip arthroplasty (THA) in patients with Crowe type III dislocated hips, who have a large acetabular bone defect. The aim of this study was to evaluate the long-term clinical results of patients in whom anatomical reconstruction of the acetabulum was performed using a cemented acetabular component and autologous bone graft from the femoral neck.

Methods

A total of 22 patients with Crowe type III dislocated hips underwent 28 THAs using bone graft from the femoral neck between 1979 and 2000. A Charnley cemented acetabular component was placed at the level of the true acetabulum after preparation with bone grafting. All patients were female with a mean age at the time of surgery of 54 years (35 to 68). A total of 18 patients (21 THAs) were followed for a mean of 27.2 years (20 to 33) after the operation.


Aims

Vitamin E-infused highly crosslinked polyethylene (VEPE) has been introduced into total hip arthroplasty (THA) with the aim of further improving the wear characteristics of moderately and highly crosslinked polyethylenes (ModXLPE and HXLPE). There are few studies analyzing the outcomes of vitamin E-infused components in cemented arthroplasty, though early acetabular component migration has been reported. The aim of this study was to measure five-year polyethylene wear and acetabular component stability of a cemented VEPE acetabular component compared with a ModXLPE cemented acetabular component.

Methods

In a prospective randomized controlled trial (RCT), we assessed polyethylene wear and acetabular component stability (primary outcome) with radiostereometric analysis (RSA) in 68 patients with reverse hybrid THA at five years follow-up. Patients were randomized to either a VEPE or a ModXLPE cemented acetabular component.


The Bone & Joint Journal
Vol. 102-B, Issue 2 | Pages 198 - 204
1 Feb 2020
Perlbach R Palm L Mohaddes M Ivarsson I Schilcher J

Aims. This single-centre observational study aimed to describe the results of extensive bone impaction grafting of the whole acetabular cavity in combination with an uncemented component in acetabular revisions performed in a standardized manner since 1993. Methods. Between 1993 and 2013, 370 patients with a median age of 72 years (interquartile range (IQR) 63 to 79 years) underwent acetabular revision surgery. Of these, 229 were more than ten years following surgery and 137 were more than 15 years. All revisions were performed with extensive use of morcellized allograft firmly impacted into the entire acetabular cavity, followed by insertion of an uncemented component with supplementary screw fixation. All types of reoperation were captured using review of radiographs and medical charts, combined with data from the local surgical register and the Swedish Hip Arthroplasty Register. Results. Among patients with possible follow-up of ten and 15 years, 152 and 72 patients remained alive without revision of the acetabular component. The number of deaths was 61 and 50, respectively. Of those who died, six patients in each group had a reoperation performed before death. The number of patients with a reoperation was 22 for those with ten-year follow-up and 21 for those with 15 years of follow-up. The Kaplan-Meier implant survival rate for aseptic loosening among all 370 patients in the cohort was 96.3% (95% confidence interval (CI) 94.1 to 98.5) after ten years and 92.8% (95% CI 89.2 to 96.6) after 15 years. Conclusion. Extensive bone impaction grafting combined with uncemented revision components appears to be a reliable method with favourable long-term survival. This technique offers the advantage of bone stock restoration and disputes the long-standing perception that uncemented components require > 50% of host bone contact for successful implant survival. Cite this article: Bone Joint J 2020;102-B(2):198–204


The Bone & Joint Journal
Vol. 101-B, Issue 8 | Pages 951 - 959
1 Aug 2019
Preston N McHugh GA Hensor EMA Grainger AJ O’Connor PJ Conaghan PG Stone MH Kingsbury SR

Aims

This study aimed to develop a virtual clinic for the purpose of reducing face-to-face orthopaedic consultations.

Patients and Methods

Anonymized experts (hip and knee arthroplasty patients, surgeons, physiotherapists, radiologists, and arthroplasty practitioners) gave feedback via a Delphi Consensus Technique. This consisted of an iterative sequence of online surveys, during which virtual documents, made up of a patient-reported questionnaire, standardized radiology report, and decision-guiding algorithm, were modified until consensus was achieved. We tested the patient-reported questionnaire on seven patients in orthopaedic clinics using a ‘think-aloud’ process to capture difficulties with its completion.


The Bone & Joint Journal
Vol. 100-B, Issue 11 | Pages 1499 - 1454
1 Nov 2018
Green CM Buckley SC Hamer AJ Kerry RM Harrison TP

Aims

The management of acetabular defects at the time of revision hip arthroplasty surgery is a challenge. This study presents the results of a long-term follow-up study of the use of irradiated allograft bone in acetabular reconstruction.

Patients and Methods

Between 1990 and 2000, 123 hips in 110 patients underwent acetabular reconstruction for aseptic loosening, using impaction bone grafting with frozen, irradiated, and morsellized femoral heads and a cemented acetabular component. A total of 55 men and 55 women with a mean age of 64.3 years (26 to 97) at the time of revision surgery are included in this study.


The Bone & Joint Journal
Vol. 101-B, Issue 11 | Pages 1423 - 1430
1 Nov 2019
Wiik AV Lambkin R Cobb JP

Aims

The aim of this study was to assess the functional gain achieved following hip resurfacing arthroplasty (HRA).

Patients and Methods

A total of 28 patients (23 male, five female; mean age, 56 years (25 to 73)) awaiting Birmingham HRA volunteered for this prospective gait study, with an age-matched control group of 26 healthy adults (16 male, ten female; mean age, 56 years (33 to 84)). The Oxford Hip Score (OHS) and gait analysis using an instrumented treadmill were used preoperatively and more than two years postoperatively to measure the functional change attributable to the intervention.


The Bone & Joint Journal
Vol. 101-B, Issue 3 | Pages 311 - 316
1 Mar 2019
Löchel J Janz V Hipfl C Perka C Wassilew GI

Aims

The use of trabecular metal (TM) shells supported by augments has provided good mid-term results after revision total hip arthroplasty (THA) in patients with a bony defect of the acetabulum. The aim of this study was to assess the long-term implant survivorship and radiological and clinical outcomes after acetabular revision using this technique.

Patients and Methods

Between 2006 and 2010, 60 patients (62 hips) underwent acetabular revision using a combination of a TM shell and augment. A total of 51 patients (53 hips) had complete follow-up at a minimum of seven years and were included in the study. Of these patients, 15 were men (29.4%) and 36 were women (70.6%). Their mean age at the time of revision THA was 64.6 years (28 to 85). Three patients (5.2%) had a Paprosky IIA defect, 13 (24.5%) had a type IIB defect, six (11.3%) had a type IIC defect, 22 (41.5%) had a type IIIA defect, and nine (17%) had a type IIIB defect. Five patients (9.4%) also had pelvic discontinuity.


The Bone & Joint Journal
Vol. 98-B, Issue 6 | Pages 767 - 771
1 Jun 2016
Konan S Duncan CP Masri BA Garbuz DS

Aims

Reconstruction of the acetabulum after failed total hip arthroplasty (THA) can be a surgical challenge in the presence of severe bone loss. We report the long-term survival of a porous tantalum revision acetabular component, its radiological appearance and quality of life outcomes.

Patients and Methods

We reviewed the results of 46 patients who had undergone revision of a failed acetabular component with a Paprosky II or III bone defect and reconstruction with a hemispherical, tantalum acetabular component, supplementary screws and a cemented polyethylene liner.


The Bone & Joint Journal
Vol. 100-B, Issue 1_Supple_A | Pages 17 - 21
1 Jan 2018
Konan S Duncan CP

Patients with neuromuscular imbalance who require total hip arthroplasty (THA) present particular technical problems due to altered anatomy, abnormal bone stock, muscular imbalance and problems of rehabilitation.

In this systematic review, we studied articles dealing with THA in patients with neuromuscular imbalance, published before April 2017. We recorded the demographics of the patients and the type of neuromuscular pathology, the indication for surgery, surgical approach, concomitant soft-tissue releases, the type of implant and bearing, pain and functional outcome as well as complications and survival.

Recent advances in THA technology allow for successful outcomes in these patients. Our review suggests excellent benefits for pain relief and good functional outcome might be expected with a modest risk of complication.

Cite this article: Bone Joint J 2018;100-B(1 Supple A):17–21.


The Bone & Joint Journal
Vol. 99-B, Issue 4_Supple_B | Pages 27 - 32
1 Apr 2017
Cnudde PHJ Kärrholm J Rolfson O Timperley AJ Mohaddes M

Aims

Compared with primary total hip arthroplasty (THA), revision surgery can be challenging. The cement-in-cement femoral revision technique involves removing a femoral component from a well-fixed femoral cement mantle and cementing a new stem into the original mantle. This technique is widely used and when carried out for the correct indications, is fast, relatively inexpensive and carries a reduced short-term risk for the patient compared with the alternative of removing well-fixed cement. We report the outcomes of this procedure when two commonly used femoral stems are used.

Patients and Methods

We identified 1179 cement-in-cement stem revisions involving an Exeter or a Lubinus stem reported to the Swedish Hip Arthroplasty Register (SHAR) between January 1999 and December 2015. Kaplan-Meier survival analysis was performed.