Aims. The aim of this study was to evaluate the effects of using a
portable, accelerometer-based surgical navigation system (KneeAlign2)
in total knee arthroplasty (TKA) on the alignment of the femoral
component, and blood loss. Patients and Methods. A total of 241 consecutive patients with primary osteoarthritis
of the knee were enrolled in this prospective, randomised controlled
study. There were 207 women and 34 men. The mean age of the patients
was 74.0 years (57 to 89). The KneeAlign2 system was used for distal
femoral resection in 121 patients (KA2 group) and a conventional intramedullary
femoral guide was used in 120 patients (IM group). Results. One patient (0.8%) in the KA2 group and 19 in the IM group had
an alignment which was
>
3° away from the neutral mechanical axis (p <
0.01). The mean
deviation from neutral alignment was 1.01° (standard deviation (. sd). 1.0°)
in the KA2 group and 1.93° (. sd. 1.7°) in the IM group (p
<
0.01). Blood loss was significantly less in the KA2 group compared
with the IM group (784 ml (. sd. 357) versus 1071
ml (. sd. 310), p <
0.001). Conclusion. The KneeAlign2 system provides a technically straightforward
method for identifying the femoral head and performing an accurate
distal femoral resection at TKA with significantly less blood loss
compared with a conventional
Aims. A significant percentage of patients remain dissatisfied after total knee arthroplasty (TKA). The aim of this study was to determine whether the sequential addition of accelerometer-based navigation for femoral component preparation and sensor-guided ligament balancing improved complication rates, radiological alignment, or patient-reported outcomes (PROMs) compared with a historical control group using conventional instrumentation. Methods. This retrospective cohort study included 371 TKAs performed by a single surgeon sequentially. A historical control group, with the use of
We undertook a prospective, randomised study of 135 total knee arthroplasties to determine the most accurate and reliable technique for alignment of the tibial prosthesis. Tibial resection was guided by either intramedullary or extramedullary alignment jigs. Of the 135 knees, standardised postoperative radiographs suitable for assessment were available in 100. Correct tibial alignment was found in 85% of the intramedullary group compared with 65% of the extramedullary group (p = 0.019). We conclude that
The aim of this study was to investigate the distribution of phenotypes in Asian patients with end-stage osteoarthritis (OA) and assess whether the phenotype affected the clinical outcome and survival of mechanically aligned total knee arthroplasty (TKA). We also compared the survival of the group in which the phenotype unintentionally remained unchanged with those in which it was corrected to neutral. The study involved 945 TKAs, which were performed in 641 patients with primary OA, between January 2000 and January 2009. These were classified into 12 phenotypes based on the combined assessment of four categories of the arithmetic hip-knee-ankle angle and three categories of actual joint line obliquity. The rates of survival were analyzed using Kaplan-Meier methods and the log-rank test. The Hospital for Special Surgery score and survival of each phenotype were compared with those of the reference phenotype with neutral alignment and a parallel joint line. We also compared long-term survival between the unchanged phenotype group and the corrected to neutral alignment-parallel joint line group in patients with Type IV-b (mild to moderate varus alignment-parallel joint line) phenotype.Aims
Methods
Maquet's line passes from the centre of the femoral head to the centre of the body of the talus. The distance of this line from the centre of the knee on a long-leg radiograph provides the most accurate measure of coronal alignment. Malalignment causes abnormal forces which may lead to loosening after knee replacement. We report a series of 115 Denham knee replacements performed between 1976 and 1981 using the earliest design of components, inserted with
This study compared patient-reported outcomes of three total knee arthroplasty (TKA) designs from one manufacturer: one cruciate-retaining (CR) design, and two cruciate-sacrificing designs, anterior-stabilized (AS) and posterior-stabilized (PS). Patients scheduled for primary TKA were included in a single-centre, prospective, three-armed, blinded randomized trial (n = 216; 72 per group). After intraoperative confirmation of posterior cruciate ligament (PCL) integrity, patients were randomly allocated to receive a CR, AS, or PS design from the same TKA system. Insertion of an AS or PS design required PCL resection. The primary outcome was the mean score of all five subscales of the Knee injury and Osteoarthritis Outcome Score (KOOS) at two-year follow-up. Secondary outcomes included all KOOS subscales, Oxford Knee Score, EuroQol five-dimension health questionnaire, EuroQol visual analogue scale, range of motion (ROM), and willingness to undergo the operation again. Patient satisfaction was also assessed.Aims
Methods
We undertook a prospective, randomised study using a non-invasive transcranial Doppler device to evaluate cranial embolisation in computer-assisted navigated total knee arthroplasty (n = 14) and compared this with a standard conventional surgical technique using
Intraoperative pressure sensors allow surgeons to quantify soft-tissue balance during total knee arthroplasty (TKA). The aim of this study was to determine whether using sensors to achieve soft-tissue balance was more effective than manual balancing in improving outcomes in TKA. A multicentre randomized trial compared the outcomes of sensor balancing (SB) with manual balancing (MB) in 250 patients (285 TKAs). The primary outcome measure was the mean difference in the four Knee injury and Osteoarthritis Outcome Score subscales (ΔKOOS4) in the two groups, comparing the preoperative and two-year scores. Secondary outcomes included intraoperative balance data, additional patient-reported outcome measures (PROMs), and functional measures.Aims
Methods
Cementless total knee arthroplasty (TKA) offers the potential for strong biological fixation compared with cemented TKA where fixation is achieved by the mechanical integration of the cement. Few mid-term results are available for newer cementless TKA designs, which have used additive manufacturing (3D printing). The aim of this study was to present mid-term clinical outcomes and implant survivorship of the cementless Stryker Triathlon Tritanium TKA. This was a single institution registry review of prospectively gathered data from 341 cementless Triathlon Tritanium TKAs at four to 6.8 years follow-up. Outcomes were determined by comparing pre- and postoperative Knee Injury and Osteoarthritis Outcome Score for Joint Replacement (KOOS JR) scores, and pre- and postoperative 12-item Veterans RAND/Short Form Health Survey (VR/SF-12) scores. Aseptic loosening and revision for any reason were the endpoints which were used to determine survivorship at five years.Aims
Methods
Alternative alignment concepts, including kinematic and restricted kinematic, have been introduced to help improve clinical outcomes following total knee arthroplasty (TKA). The purpose of this study was to evaluate the clinical results, along with patient satisfaction, following TKA using the concept of restricted kinematic alignment. A total of 121 consecutive TKAs performed between 11 February 2018 to 11 June 2019 with preoperative varus deformity were reviewed at minimum one-year follow-up. Three knees were excluded due to severe preoperative varus deformity greater than 15°, and a further three due to requiring revision surgery, leaving 109 patients and 115 knees to undergo primary TKA using the concept of restricted kinematic alignment with advanced technology. Patients were stratified into three groups based on the preoperative limb varus deformity: Group A with 1° to 5° varus (43 knees); Group B between 6° and 10° varus (56 knees); and Group C with varus greater than 10° (16 knees). This study group was compared with a matched cohort of 115 TKAs and 115 patients using a neutral mechanical alignment target with manual instruments performed from 24 October 2016 to 14 January 2019.Aims
Methods
The purpose of this study was to report bone adaptive changes after anatomical total shoulder arthroplasty (TSA) using a standard-length hydroxyapatite (HA)-coated humeral component, and to report on a computer-based analysis of radiographs to determine changes in peri-implant bone density objectively. A total of 44 TSAs, performed between 2011 and 2014 using a cementless standard-length humeral component proximally coated with HA, were included. There were 23 males and 21 females with a mean age of 65 years (17 to 65). All shoulders had good quality radiographs at six weeks and five years postoperatively. Three observers graded bone adaptive changes. All radiographs were uploaded into a commercially available photographic software program. The grey value density of humeral radiological areas was corrected to the grey value density of the humeral component and compared over time.Aims
Methods
This study aimed to identify the tibial component and femoral component coronal angles (TCCAs and FCCAs), which concomitantly are associated with the best outcomes and survivorship in a cohort of fixed-bearing, cemented, medial unicompartmental knee arthroplasties (UKAs). We also investigated the potential two-way interactions between the TCCA and FCCA. Prospectively collected registry data involving 264 UKAs from a single institution were analyzed. The TCCAs and FCCAs were measured on postoperative radiographs and absolute angles were analyzed. Clinical assessment at six months, two years, and ten years was undertaken using the Knee Society Knee score (KSKS) and Knee Society Function score (KSFS), the Oxford Knee Score (OKS), the 36-Item Short-Form Health Survey questionnaire (SF-36), and range of motion (ROM). Fulfilment of expectations and satisfaction was also recorded. Implant survivorship was reviewed at a mean follow-up of 14 years (12 to 16). Multivariate regression models included covariates, TCCA, FCCA, and two-way interactions between them. Partial residual graphs were generated to identify angles associated with the best outcomes. Kaplan-Meier analysis was used to compare implant survivorship between groups.Aims
Methods
Metaphyseal fixation during revision total knee arthroplasty (TKA) is important, but potentially difficult when using historical designs of cone. Material and manufacturing innovations have improved the size and shape of the cones which are available, and simplified the required bone preparation. In a large series, we assessed the implant survivorship, radiological results, and clinical outcomes of new porous 3D-printed titanium metaphyseal cones featuring a reamer-based system. We reviewed 142 revision TKAs in 139 patients using 202 cones (134 tibial, 68 femoral) which were undertaken between 2015 and 2016. A total of 60 involved tibial and femoral cones. Most cones (149 of 202; 74%) were used for Type 2B or 3 bone loss. The mean age of the patients was 66 years (44 to 88), and 76 (55 %) were female. The mean body mass index (BMI) was 34 kg/m2 (18 to 60). The patients had a mean of 2.4 (1 to 8) previous operations on the knee, and 68 (48%) had a history of prosthetic infection. The mean follow-up was 2.4 years (2 to 3.6).Aims
Methods
Robotic-assisted total knee arthroplasty (RA-TKA) is theoretically more accurate for component positioning than TKA performed with mechanical instruments (M-TKA). Furthermore, the ability to incorporate soft-tissue laxity data into the plan prior to bone resection should reduce variability between the planned polyethylene thickness and the final implanted polyethylene. The purpose of this study was to compare accuracy to plan for component positioning and precision, as demonstrated by deviation from plan for polyethylene insert thickness in measured-resection RA-TKA versus M-TKA. A total of 220 consecutive primary TKAs between May 2016 and November 2018, performed by a single surgeon, were reviewed. Planned coronal plane component alignment and overall limb alignment were all 0° to the mechanical axis; tibial posterior slope was 2°; and polyethylene thickness was 9 mm. For RA-TKA, individual component position was adjusted to assist gap-balancing but planned coronal plane alignment for the femoral and tibial components and overall limb alignment remained 0 ± 3°; planned tibial posterior slope was 1.5°. Mean deviations from plan for each parameter were compared between groups for positioning and size and outliers were assessed.Aims
Methods
This study aimed to evaluate the association between the sagittal alignment of the femoral component in total knee arthroplasty (TKA) and new Knee Society Score (2011KSS), under the hypothesis that outliers such as the excessive extended or flexed femoral component were related to worse clinical outcomes. A group of 156 knees (134 F:22 M) in 133 patients with a mean age 75.8 years (SD 6.4) who underwent TKA with the cruciate-substituting Bi-Surface Knee prosthesis were retrospectively enrolled. On lateral radiographs, γ angle (the angle between the distal femoral axis and the line perpendicular to the distal rear surface of the femoral component) was measured, and the patients were divided into four groups according to the γ angle. The 2011KSSs among groups were compared using the Kruskal-Wallis test. A secondary regression analysis was used to investigate the association between the 2011KSS and γ angle.Aims
Methods
Aims
Patients and Methods
Shoulder arthroplasty using short humeral components is becoming increasingly popular. Some such components have been associated with relatively high rates of adverse radiological findings. The aim of this retrospective review was to evaluate the radiological humeral bone changes and mechanical failure rates with implantation of a short cementless humeral component in anatomical (TSA) and reverse shoulder arthroplasty (RSA). A total of 100 shoulder arthroplasties (35 TSA and 65 RSA) were evaluated at a mean of 3.8 years (3 to 8.3). The mean age at the time of surgery was 68 years (31 to 90). The mean body mass index was 32.7 kg/m2 (17.3 to 66.4).Aims
Patients and Methods
Hydroxyapatite coatings for uncemented fixation in total knee
arthroplasty can theoretically provide a long-lasting biological
interface with the host bone. The objective of this study was to
test this hypothesis with propriety hydroxyapatite, peri-apatite,
coated tibial components using component migration measured with radiostereometric
analysis over two years as an indicator of long-term fixation. A total of 29 patients at two centres received uncemented PA-coated
tibial components and were followed for two years with radiostereometric
analysis exams to quantify the migration of the component.Aims
Patients and Methods
This study compares the PFC total knee arthroplasty (TKA) system in a prospective randomized control trial (RCT) of the mobile-bearing rotating-platform (RP) TKA against the fixed-bearing (FB) TKA. This is the largest RCT with the longest follow-up where cruciate-retaining PFC total knee arthroplasties are compared in a non-bilateral TKA study. A total of 167 patients (190 knees with 23 bilateral cases), were recruited prospectively and randomly assigned, with 91 knees receiving the RP and 99 knees receiving FB. The mean age was 65.5 years (48 to 82), the mean body mass index (BMI) was 29.7 kg/m2 (20 to 52) and 73 patients were female. The Knee Society Score (KSS), Knee Society Functional Score (KSFS), Oxford Knee Score (OKS), Western Ontario and McMaster Universities Arthritis Index (WOMAC), and 12-Item Short-Form Health Survey Physical and Mental Component Scores (SF-12 PCS, SF-12 MCS) were gathered and recorded preoperatively, at five-years’ follow-up, and at ten years’ follow-up. Additionally, Knee Injury and Osteoarthritis Outcome Scores (KOOS) were collected at five- and ten-year follow-ups. The prevalence of radiolucent lines (RL) on radiographs and implant survival were recorded at five- and ten-year follow-ups.Aims
Patients and Methods
We conducted a randomised controlled trial to assess the accuracy
of positioning and alignment of the components in total knee arthroplasty
(TKA), comparing those undertaken using standard intramedullary
cutting jigs and those with patient-specific instruments (PSI). There were 64 TKAs in the standard group and 69 in the PSI group. The post-operative hip-knee-ankle (HKA) angle and positioning
was investigated using CT scans. Deviation of >
3° from the planned
position was regarded as an outlier. The operating time, Oxford
Knee Scores (OKS) and Short Form-12 (SF-12) scores were recorded.Aims
Patients and Methods
Ensuring correct rotation of the femoral component
is a challenging aspect of patellofemoral replacement surgery. Rotation
equal to the epicondylar axis or marginally more external rotation
is acceptable. Internal rotation is associated with poor outcomes.
This paper comprises two studies evaluating the use of the medial
malleolus as a landmark to guide rotation. We used 100 lower-leg anteroposterior radiographs to evaluate
the reliability of the medial malleolus as a landmark. Assessment
was made of the angle between the tibial shaft and a line from the
intramedullary rod entry site to the medial malleolus. The femoral
cut was made in ten cadaver knees using the inferior tip of the
medial malleolus as a landmark for rotation. Rotation of the cut
relative to the anatomical epicondylar axis was assessed using CT.
The study of radiographs found the position of the medial malleolus
relative to the tibial axis is consistent. Using the inferior tip
of the medial malleolus in the cadaver study produced a mean external
rotation of 1.6° (0.1° to 3.7°) from the anatomical epicondylar
axis. Using the inferior tip of the medial malleolus to guide the
femoral cutting jig avoids internal rotation and introduces an acceptable
amount of external rotation of the femoral component.
Primary total knee arthroplasty (TKA) is a reliable
procedure with reproducible long-term results. Nevertheless, there
are conditions related to the type of patient or local conditions
of the knee that can make it a difficult procedure. The most common
scenarios that make it difficult are discussed in this review. These
include patients with many previous operations and incisions, and
those with severe coronal deformities, genu recurvatum, a stiff knee,
extra-articular deformities and those who have previously undergone
osteotomy around the knee and those with chronic dislocation of
the patella. Each condition is analysed according to the characteristics of
the patient, the pre-operative planning and the reported outcomes. When approaching the difficult primary TKA surgeons should use
a systematic approach, which begins with the review of the existing
literature for each specific clinical situation. Cite this article:
Patient-specific cutting guides (PSCGs) are designed
to improve the accuracy of alignment of total knee replacement (TKR).
We compared the accuracy of limb alignment and component positioning
after TKR performed using PSCGs or conventional instrumentation.
A total of 80 patients were randomised to undergo TKR with either of
the different forms of instrumentation, and radiological outcomes
and peri-operative factors such as operating time were assessed.
No significant difference was observed between the groups in terms
of tibiofemoral angle or femoral component alignment. Although the
tibial component in the PSCGs group was measurably closer to neutral
alignment than in the conventional group, the size of the difference
was very small (89.8° ( Cite this article:
We compared the alignment of 39 total knee replacements implanted using the conventional alignment guide system with 37 implanted using a CT-based navigation system, performed by a single surgeon. The knees were evaluated using full-length weight-bearing anteroposterior radiographs, lateral radiographs and CT scans. The mean hip-knee-ankle angle, coronal femoral component angle and coronal tibial component angle were 181.8° (174.2° to 188.3°), 88.5° (84.0° to 91.8°) and 89.7° (86.3° to 95.1°), respectively for the conventional group and 180.8° (178.2° to 185.1°), 89.3° (85.8° to 92.0°) and 89.9° (88.0° to 93.0°), respectively for the navigated group. The mean sagittal femoral component angle was 85.5° (80.6° to 92.8°) for the conventional group and 89.6° (85.5° to 94.0°) for the navigated group. The mean rotational femoral and tibial component angles were −0.7° (−8.8° to 9.8°) and −3.3° (−16.8° to 5.8°) for the conventional group and −0.6° (−3.5° to 3.0°) and 0.3° (−5.3° to 7.7°) for the navigated group. The ideal angles of all alignments in the navigated group were obtained at significantly higher rates than in the conventional group. Our results demonstrated significant improvements in component positioning with a CT-based navigation system, especially with respect to rotational alignment.
Total knee arthroplasty (TKA) is known to lead
to a reduction in periprosthetic bone mineral density (BMD). In theory,
this may lead to migration, instability and aseptic loosening of
the prosthetic components. Bisphosphonates inhibit bone resorption
and may reduce this loss in BMD. We hypothesised that treatment
with bisphosphonates and calcium would lead to improved BMD and
clinical outcomes compared with treatment with calcium supplementation
alone following TKA. A total of 26 patients, (nine male and 17 female,
mean age 67 years) were prospectively randomised into two study
groups: alendronate and calcium (bisphosphonate group, n = 14) or calcium
only (control group, n = 12). Dual energy X-ray absorptiometry (DEXA)
measurements were performed post-operatively, and at three months,
six months, one, two, four, and seven years post-operatively. Mean femoral metaphyseal BMD was significantly higher in the
bisphosphonate group compared with controls, up to four years following
surgery in some areas of the femur (p = 0.045). BMD was observed
to increase in the lateral tibial metaphysis in the bisphosphonate
group until seven years (p = 0.002), and was significantly higher than
that observed in the control group throughout (p = 0.024). There
were no significant differences between the groups in the central
femoral metaphyseal, tibial medial metaphyseal or diaphyseal regions
of interest (ROI) of either the femur or tibia. Bisphosphonate treatment after TKA may be of benefit for patients
with poor bone quality. However, further studies with a larger number
of patients are necessary to assess whether this is clinically beneficial. Cite this article:
Intra-articular resection of bone with soft-tissue balancing and total knee replacement (TKR) has been described for the treatment of patients with severe osteoarthritis of the knee associated with an ipsilateral malunited femoral fracture. However, the extent to which deformity in the sagittal plane can be corrected has not been addressed. We treated 12 patients with severe arthritis of the knee and an extra-articular malunion of the femur by TKR with intra-articular resection of bone and soft-tissue balancing. The femora had a mean varus deformity of 16° (8° to 23°) in the coronal plane. There were seven recurvatum deformities with a mean angulation of 11° (6° to 15°) and five antecurvatum deformities with a mean angulation of 12° (6° to 15°). The mean follow-up was 93 months (30 to 155). The median Knee Society knee and function scores improved from 18.7 (0 to 49) and 24.5 (10 to 50) points pre-operatively to 93 (83 to 100) and 90 (70 to 100) points at the time of the last follow-up, respectively. The mean mechanical axis of the knee improved from 22.6° of varus (15° to 27° pre-operatively to 1.5° of varus (3° of varus to 2° of valgus) at the last follow-up. The recurvatum deformities improved from a mean of 11° (6° to 15°) pre-operatively to 3° (0° to 6°) at the last follow-up. The antecurvatum deformities in the sagittal plane improved from a mean of 12° (6° to 16°) pre-operatively to 4.4° (0° to 8°) at the last follow-up. Apart from varus deformities, TKR with intra-articular bone resection effectively corrected the extra-articular deformity of the femur in the presence of antecurvatum of up to 16° and recurvatum of up to 15°.
Achieving deep flexion after total knee replacement remains a challenge. In this study we compared the soft-tissue tension and tibiofemoral force in a mobile-bearing posterior cruciate ligament-sacrificing total knee replacement, using equal flexion and extension gaps, and with the gaps increased by 2 mm each. The tests were conducted during passive movement in five cadaver knees, and measurements of strain were made simultaneously in the collateral ligaments. The tibiofemoral force was measured using a customised mini-force plate in the tibial tray. Measurements of collateral ligament strain were not very sensitive to changes in the gap ratio, but tibiofemoral force measurements were. Tibiofemoral force was decreased by a mean of 40% (
Systemic emboli released during total knee replacement have been implicated as a cause of peri-operative morbidity and neurological dysfunction. We undertook a prospective, double-blind, randomised study to compare the cardiac embolic load sustained during computer-assisted and conventional, intramedullary-aligned, total knee replacement, as measured by transoesophageal echocardiography. There were 26 consecutive procedures performed by a single surgeon at a single hospital. The embolic load was scored using the modified Mayo grading system for echogenic emboli. Fourteen patients undergoing computer-assisted total knee replacement had a mean embolic score of 4.89 (3 to 7) and 12 undergoing conventional total knee replacement had a mean embolic score of 6.15 (4 to 8) on release of the tourniquet. Comparison of the groups using a two-tailed This study demonstrates that computer-assisted knee replacement results in the release of significantly fewer systemic emboli than the conventional procedure using intramedullary alignment.
We carried out a prospective study to assess the clinical outcome, complications and survival of the NexGen Legacy posterior-stabilised-Flex total knee replacement (TKR) in a consecutive series of 278 knees between May 2003 and February 2005. Mean follow-up for 259 TKRs (98.2%) was 3.8 years (3.0 to 4.8). Annual follow-up showed improvement in the Knee Society scores (paired This relatively large study indicates that the legacy posterior stabilised-Flex design provides excellent short-term outcome but warrants ongoing evaluation to confirm the long-term durability and functioning of the implant.
The aim of this study was to compare the results in patients having a quadriceps sparing total knee replacement (TKR) with those undergoing a standard TKR at a minimum follow-up of two years. All patients who had a TKR with a high-flex posterior-stabilised prosthesis prior to December 2002 were reviewed retrospectively. There were 57 patients available for follow-up. Those with a quadriceps sparing TKR had less pain peri-operatively with a greater degree of flexion at all the post-operative visits and at the final follow-up, but their operations took longer, with less accurate radiological alignment. There was no difference in the complications and in the Knee Society scores between the two groups at the final follow-up. Total knee replacement through a quadriceps sparing approach has some peri-operative advantages over the standard incision. At a minimum follow-up of two years the clinical results were similar to those with a standard incision, but the radiological outcomes of the quadriceps sparing group were inferior.
A 64-year-old man presented with a severe deformity of the tibia caused by Paget’s disease and osteoarthritis of the ipsilateral knee. Total knee replacement required preliminary correction of the tibial deformity. This was successfully achieved by tibial osteotomy followed by distraction osteogenesis using the Taylor spatial frame. The subsequent knee replacement was successful, with no recurrence of deformity.
We report the results of using a combination of fixator-assisted nailing with lengthening over an intramedullary nail in patients with tibial deformity and shortening. Between 1997 and 2007, 13 tibiae in nine patients with a mean age of 25.4 years (17 to 34) were treated with a unilateral external fixator for acute correction of deformity, followed by lengthening over an intramedullary nail with a circular external fixator applied at the same operating session. At the end of the distraction period locking screws were inserted through the intramedullary nail and the external fixator was removed. The mean amount of lengthening was 5.9 cm (2 to 8). The mean time of external fixation was 90 days (38 to 265). The mean external fixation index was 15.8 days/cm (8.9 to 33.1) and the mean bone healing index was 38 days/cm (30 to 60). One patient developed an equinus deformity which responded to stretching and bracing. Another developed a drop foot due to a compartment syndrome, which was treated by fasciotomy. It recovered in three months. Two patients required bone grafting for poor callus formation. We conclude that the combination of fixator-assisted nailing with lengthening over an intramedullary nail can reduce the overall external fixation time and prevent fractures and deformity of the regenerated bone.
Compartment syndrome is a rare complication of total knee arthroplasty that requires early recognition and prompt decompression in order to prevent long-term disability. We have found only one previous case report in the literature. We present a series of seven cases from four hospitals and five surgeons. Six of the cases resulted in the loss of at least one compartment, and one resulted in amputation. Four of the cases resulted in legal action. We suggest that important risk factors contributing to the development of this condition include complex surgery, soft-tissue compromise, previous surgery, and possibly vascular disease. Delay in the diagnosis and hence delay in decompression was common in our series, and in five cases appeared to be related to the use of a postoperative epidural infusion for pain relief. The presence of associated neurological compromise may have also been a significant factor in the delay to diagnosis in two cases.
We performed a randomised, controlled trial involving 150 patients with a pre-operative level of haemoglobin of 13.0 g/dl or less, to compare the effect of either topical fibrin spray or intravenous tranexamic acid on blood loss after total knee replacement. A total of 50 patients in the topical fibrin spray group had 10 ml of the reconstituted product applied intra-operatively to the operation site. The 50 patients in the tranexamic acid group received 500 mg of tranexamic acid intravenously five minutes before deflation of the tourniquet and a repeat dose three hours later, and a control group of 50 patients received no pharmacological intervention. There was a significant reduction in the total calculated blood loss for those in the topical fibrin spray group (p = 0.016) and tranexamic acid group (p = 0.041) compared with the control group, with mean losses of 1190 ml (708 to 2067), 1225 ml (580 to 2027), and 1415 ml (801 to 2319), respectively. The reduction in blood loss in the topical fibrin spray group was not significantly different from that achieved in the tranexamic acid group (p = 0.72).
This study aimed to determine the correlation between the amount of maximum flexion and the clinical outcome in 207 Koreans (333 knees) undergoing total knee replacement. The association of maximum flexion with clinical outcome was evaluated one year postoperatively using three scoring systems; the American Knee Society score, Western Ontario McMaster Universities Osteoarthritis index and the Short Form-36. The mean maximum flexion decreased post-operatively at 12 months from 140.1° (60° to 160°) to 133.0° (105° to 150°). Only the social function score of the Short Form-36 correlated significantly with maximum flexion (correlation coefficient = 0.180, p = 0.039). In comparative analyses of subgroups divided by a maximum flexion of 120°, we found no significant differences in any parameters except the social function score of the Short Form-36 (41.9
Bilateral sequential total knee replacement was carried out under one anaesthetic in 100 patients. One knee was replaced using a CT-free computer-assisted navigation system and the other conventionally without navigation. The two methods were compared for accuracy of orientation and alignment of the components. There were 85 women and 15 men with a mean age of 67.6 years (54 to 83). Radiological and CT imaging was carried out to determine the alignment of the components. The mean follow-up was 2.3 years (2 to 3). The operating and tourniquet times were significantly longer in the navigation group (p <
0.001). There were no significant pre- or post-operative differences between the knee scores of the two groups (p = 0.288 and p = 0.429, respectively). The results of imaging and the number of outliers for all radiological parameters were not statistically different (p = 0.109 to p = 0.920). In this series computer-assisted navigated total knee replacement did not result in more accurate orientation and alignment of the components than that achieved by conventional total knee replacement.
The appearance of the ‘grand-piano sign’ on the anterior resected surface of the femur has been considered to be a marker for correct femoral rotational alignment during total knee replacement. Our study was undertaken to assess quantitatively the morphological patterns on the resected surface after anterior femoral resection with various angles of external rotation, using a computer-simulation technique. A total of 50 right distal femora with varus osteoarthritis in 50 Korean patients were scanned using computerised tomography. Computer image software was used to simulate the anterior femoral cut, which was applied at an external rotation of 0°, 3° and 6° relative to the posterior condylar axis, and parallel to the surgical and clinical epicondylar axes in each case. The morphological patterns on the resected surface were quantified and classified as the ‘grand-piano sign’, ‘the boot sign’ and the ‘butterfly sign’. The surgeon can use the analogy of these quantified sign patterns to ensure that a correct rotational alignment has been obtained intra-operatively.
We analysed at a mean follow-up of 7.25 years the clinical and radiological outcome of 117 patients (125 knees) who had undergone a primary, cemented, modular Freeman-Samuelson total knee replacement. While the tibial and femoral components were cemented, the patellar component was uncemented. A surface-cementing technique was used to secure the tibial components. A total of 82 knees was available for radiological assessment. Radiolucent lines were seen in 41 knees (50%) and osteolytic lesions were seen in 13 knees (16%). Asymptomatic, rotational loosening of the patellar implant was seen in four patients and osteolysis was more common in patients with a patellar resurfacing. Functional outcome scores were available for 41 patients (41 knees, 35%) and the mean Western Ontario McMasters Universities score was 77.5 (
We compared the performance of uncemented trabecular metal tibial components in total knee replacement with that of cemented tibial components in patients younger than 60 years over two years using radiostereophotogrammetric analysis (RSA). A total of 22 consecutive patients (mean age 53 years, 33 to 59, 26 knees) received an uncemented NexGen trabecular metal cruciate-retaining monobloc tibial component and 19 (mean 53 years, 44 to 59, 21 knees) a cemented NexGen Option cruciate-retaining modular tibial component. All the trabecular metal components migrated during the initial three months and then stabilised. The exception was external rotation, which did not stabilise until 12 months. Unlike conventional metal-backed implants which displayed a tilting migration comprising subsidence and lift-off from the tibial tray, most of the trabecular metal components showed subsidence only, probably due to the elasticity of the implant. This pattern of subsidence is regarded as being beneficial for uncemented fixation.
We conducted a retrospective study to investigate the effect of femoral bowing on the placement of components in total knee replacement (TKR), with regard to its effect on reestablishing the correct mechanical axis, as we hypothesised that computer-assisted total knee replacement (CAS-TKR) would produce more accurate alignment than conventional TKR. Between January 2006 and December 2009, 212 patients (306 knees) underwent TKR. The conventional TKR was compared with CAS-TKR for accuracy of placement of the components and post-operative alignment, as determined by five radiological measurements. There were significant differences in the reconstructed mechanical axes between the bowed and the non-bowed group after conventional TKR (176.2° ( For patients with significant femoral bowing, the reconstructed mechanical axes were significantly closer to normal in the CAS group than in the conventional group (179.2° (
We carried out a prospective randomised study to evaluate the blood loss in 60 patients having a total knee arthroplasty and divided randomly into two equal groups, one having a computer-assisted procedure and the other a standard operation. The surgery was carried out by a single surgeon at one institution using a uniform approach. The only variable in the groups was the use of intramedullary femoral and tibial alignment jigs in the standard group and single tracker pins of the imageless navigation system in the tibia and femur in the navigated group. The mean drainage of blood was 1351 ml (715 to 2890; 95% confidence interval (CI) 1183 to 1518) in the computer-aided group and 1747 ml (1100 to 3030; CI 1581 to 1912) in the conventional group. This difference was statistically significant (p = 0.001). The mean calculated loss of haemoglobin was 36 g/dl in the navigated group There was a highly significant reduction in blood drainage and the calculated Hb loss between the computer-assisted and the conventional techniques. This allows the ordering of less blood before the operation, reduces risks at transfusion and gives financial saving. Computer-assisted surgery may also be useful for patients in whom blood products are not acceptable.
We have undertaken a prospective clinical and radiological analysis of 124 shoulder arthroplasties (113 patients) carried out for osteoarthritis. The clinical results showed improvement in the absolute Constant score and the American Shoulder and Elbow Surgeons score of 22 and 43, respectively. Both were statistically significant (p <
0.001). There was no significant difference in the scores after hemiarthroplasty and total arthroplasty in those patients with an intact rotator cuff. When revision was used as the end-point for survival at ten years, survival of 86%, or 90% if glenoid components made of Hylamer sterilised in air were omitted, was obtained in primary osteoarthritis. The most common cause for revision in the hemiarthroplasty group was glenoid pain at a mean of 1.5 years; in the total arthroplasty group it was loosening of the glenoid at a mean of 4.5 years. Analysis of pre-operative factors showed that the risk of gross loosening of the glenoid increased threefold when there was evidence of erosion of the glenoid at operation. Shoulder arthroplasty should not be delayed once symptomatic osteoarthritis has been established and should be undertaken before failure of the cuff or erosion of the glenoid are present.