Advertisement for orthosearch.org.uk
Results 1 - 50 of 122
Results per page:
The Bone & Joint Journal
Vol. 100-B, Issue 11 | Pages 1499 - 1454
1 Nov 2018
Green CM Buckley SC Hamer AJ Kerry RM Harrison TP

Aims. The management of acetabular defects at the time of revision hip arthroplasty surgery is a challenge. This study presents the results of a long-term follow-up study of the use of irradiated allograft bone in acetabular reconstruction. Patients and Methods. Between 1990 and 2000, 123 hips in 110 patients underwent acetabular reconstruction for aseptic loosening, using impaction bone grafting with frozen, irradiated, and morsellized femoral heads and a cemented acetabular component. A total of 55 men and 55 women with a mean age of 64.3 years (26 to 97) at the time of revision surgery are included in this study. Results. At a mean follow-up of 16.9 years, there had been 23 revisions (18.7%), including ten for infection, eight for aseptic loosening, and three for dislocation. Of the 66 surviving hips (58 patients) that could be reassessed, 50 hips (42 patients; 75.6%) were still functioning satisfactorily. Union of the graft had occurred in all hips with a surviving implant. Survival analysis for all indications was 80.6% at 15 years (55 patients at risk, 95% confidence interval (CI) 71.1 to 87.2) and 73.7% at 20 years (eight patients at risk, 95% CI 61.6 to 82.5). Conclusion. Acetabular reconstruction using frozen, irradiated, and morsellized allograft bone and a cemented acetabular component is an effective method of treatment. It gives satisfactory long-term results and is comparable to other types of reconstruction. Cite this article: Bone Joint J 2018;100-B:1449–54


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 11 | Pages 1419 - 1423
1 Nov 2009
Emms NW Buckley SC Stockley I Hamer AJ Kerry RM

Between 1990 and 2000, 123 hips in 110 patients were reconstructed for aseptic loosening using impaction bone grafting with frozen, irradiated, morsellised femoral heads and cemented acetabular components. This series was reported previously at a mean follow-up of five years. We have extended this follow-up and now describe the outcome of 86 hips in 74 patients at a mean of ten years. There have been 19 revisions, comprising nine for infection, seven for aseptic loosening and three for dislocation. In surviving acetabular reconstructions, union of the graft had occurred in 64 of 67 hips (95.5%). Survival analysis for all indications at ten years was 83.3% (95% confidence interval (CI) 68 to 89) and 71.3% (95% CI 58 to 84) at 15 years. Acetabular reconstruction using irradiated allograft and a cemented acetabular component is an effective method of reconstruction, providing results in the medium- to long-term comparable with those of reported series where non-irradiated freshly-frozen bone was used


The Journal of Bone & Joint Surgery British Volume
Vol. 80-B, Issue 3 | Pages 391 - 395
1 May 1998
Schreurs BW Slooff TJJH Buma P Gardeniers JWM Huiskes R

We report a long-term review of 60 acetabular components revised using impacted, morsellised bone allografts and a cemented polyethylene cup. The acetabular defects were cavitary (37) or combined (23). Follow-up was for a mean 11.8 years (10 to 15). Further revision was needed in five hips, two for septic and three for aseptic loosening. The overall survival rate at 11.8 years was 90%; excluding the septic cases it was 94%. Acetabular reconstruction with impacted morsellised cancellous grafts and cement gives satisfactory long-term results


The Journal of Bone & Joint Surgery British Volume
Vol. 76-B, Issue 5 | Pages 740 - 744
1 Sep 1994
Azuma T Yasuda H Okagaki K Sakai K

We report the results of 24 acetabular reconstructions in which cemented polyethylene cups and tamped corticocancellous allografts were used for severe acetabular bone deficiency. Eleven hips had type-II (cavitary) bone deficiency and 13 had type-III (combined) defects. At a mean follow-up of 5.8 years, two components had migrated more than 5 mm and had accompanying radiolucent zones of more than 2 mm width. A radiolucency 5 mm wide was also seen in zone III of an acetabular implant which had not migrated. None of the patients had required revision because of loosening or infection


The Journal of Bone & Joint Surgery British Volume
Vol. 71-B, Issue 3 | Pages 460 - 464
1 May 1989
Sommelet J Finlayson D Lesur E

We report the results of acetabular reconstruction using a bipolar prosthesis bearing on cancellous bone graft in 37 patients after a minimum of five years. There was a satisfactory clinical outcome in 58% when assessed by pain, range of movement and stability, with greater improvement in the pain score than of the other parameters. Radiological migration of the prosthesis was a frequent finding although this did not always correlate with symptoms. Better results were obtained in cases of primary or secondary protrusio acetabuli than after the revision of previous total arthroplasties


The Bone & Joint Journal
Vol. 103-B, Issue 2 | Pages 299 - 304
1 Feb 2021
Goto E Umeda H Otsubo M Teranishi T

Aims

Various surgical techniques have been described for total hip arthroplasty (THA) in patients with Crowe type III dislocated hips, who have a large acetabular bone defect. The aim of this study was to evaluate the long-term clinical results of patients in whom anatomical reconstruction of the acetabulum was performed using a cemented acetabular component and autologous bone graft from the femoral neck.

Methods

A total of 22 patients with Crowe type III dislocated hips underwent 28 THAs using bone graft from the femoral neck between 1979 and 2000. A Charnley cemented acetabular component was placed at the level of the true acetabulum after preparation with bone grafting. All patients were female with a mean age at the time of surgery of 54 years (35 to 68). A total of 18 patients (21 THAs) were followed for a mean of 27.2 years (20 to 33) after the operation.


The Journal of Bone & Joint Surgery British Volume
Vol. 88-B, Issue 7 | Pages 865 - 869
1 Jul 2006
Comba F Buttaro M Pusso R Piccaluga F

We reviewed the clinical and radiological results of 131 patients who underwent acetabular revision for aseptic loosening with impacted bone allograft and a cemented acetabular component. The mean follow-up was 51.7 months (24 to 156).

The mean post-operative Merle D’Aubigné and Postel scores were 5.7 points (4 to 6) for pain, 5.2 (3 to 6) for gait and 4.5 (2 to 6) for mobility. Radiological evaluation revealed migration greater than 5 mm in four acetabular components. Radiological failure matched clinical failure. Asymptomatic radiolucent lines were observed in 31 of 426 areas assessed (7%). Further revision was required in six patients (4.5%), this was due to infection in three and mechanical failure in three. The survival rate for the reconstruction was 95.8% (95% confidence interval 92.3 to 99.1) overall, and 98%, excluding revision due to sepsis.

Our study, from an independent centre, has reproduced the results of the originators of the method.


The Bone & Joint Journal
Vol. 95-B, Issue 7 | Pages 887 - 892
1 Jul 2013
Karachalios T Roidis N Lampropoulou-Adamidou K Hartofilakidis G

We report the results at a mean of 24.3 years (20 to 32) of 61 previously reported consecutive total hip replacements carried out on 44 patients with severe congenital hip disease, performed with reconstruction of the acetabulum with an impaction grafting technique known as cotyloplasty. The mean age of the patients at operation was 46.7 years (23 to 68) and all were women. The patients were followed post-operatively for a mean of 24.3 years (20 to 32), using the Merle d’Aubigné and Postel scoring system as modified by Charnley, and with serial radiographs. At the time of the latest follow-up, 28 acetabular components had been revised because of aseptic loosening at a mean of 15.9 years (6 to 26), and one at 40 days after surgery because of repeated dislocations. The overall survival rate for aseptic failure of the acetabular component at ten years was 93.1% (95% confidence interval (CI) 86.5 to 96.7) when 53 hips were at risk, and at 23 years was 56.1% (95% CI 49.4 to 62.8), when 22 hips remained at risk. These long-term results are considered satisfactory for the reconstruction of an acetabulum presenting with inadequate bone stock and circumferential segmental defects.

Cite this article: Bone Joint J 2013;95-B:887–92.


The Journal of Bone & Joint Surgery British Volume
Vol. 78-B, Issue 2 | Pages 270 - 275
1 Mar 1996
Hooten JP Engh CA Heekin RD Vinh TN

Two acetabula which contained large bone allografts introduced at revision arthroplasty were obtained at post-mortem. The allografts had been placed in superior defects to support cementless acetabular components, and both hips were functioning well at the time of death. Clinical radiographs demonstrated apparent healing of graft to host bone, no graft collapse and stability of the acetabular components. Microscopic examination of sections through these specimens showed that the bulk allografts were encapsulated in fibrous tissue. Vascularity was increased at the host-graft interface, but there was limited evidence of bone union between the graft and the host. In the few areas where union had occurred, revascularisation extended no more than 2 mm beyond the graft-host interface.

Within the body of the graft, the acellular matrix of trabecular bone maintained structural integrity up to 48 months after surgery. In areas where the allograft was adjacent to an implant, there was fibrous tissue orientated parallel to the implant surface. The acetabulum which contained a porous-coated component showed evidence of bone growth into the porous surface where it was in contact with viable host bone. No ingrowth occurred in areas where the porous coating was in contact with the graft. Although the grafts were functioning well, allograft revascularisation and remodelling were minimal, and the radiological appearance of healing did not correlate with histological findings.


The Bone & Joint Journal
Vol. 96-B, Issue 1 | Pages 48 - 53
1 Jan 2014
Solomon LB Hofstaetter JG Bolt MJ Howie DW

We investigated the detailed anatomy of the gluteus maximus, gluteus medius and gluteus minimus and their neurovascular supply in 22 hips in 11 embalmed adult Caucasian human cadavers. This led to the development of a surgical technique for an extended posterior approach to the hip and pelvis that exposes the supra-acetabular ilium and preserves the glutei during revision hip surgery. Proximal to distal mobilisation of the gluteus medius from the posterior gluteal line permits exposure and mobilisation of the superior gluteal neurovascular bundle between the sciatic notch and the entrance to the gluteus medius, enabling a wider exposure of the supra-acetabular ilium. This technique was subsequently used in nine patients undergoing revision total hip replacement involving the reconstruction of nine Paprosky 3B acetabular defects, five of which had pelvic discontinuity. Intra-operative electromyography showed that the innervation of the gluteal muscles was not affected by surgery. Clinical follow-up demonstrated good hip abduction function in all patients. These results were compared with those of a matched cohort treated through a Kocher–Langenbeck approach. Our modified approach maximises the exposure of the ilium above the sciatic notch while protecting the gluteal muscles and their neurovascular bundle.

Cite this article: Bone Joint J 2014;96-B:48–53.


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 6 | Pages 724 - 730
1 Jun 2008
Hartofilakidis G Georgiades G Babis GC Yiannakopoulos CK

We have evaluated the results of total hip replacement in patients with congenital hip disease using 46 cemented all-polyethylene Charnley acetabular components implanted with the cotyloplasty technique in 34 patients (group A), and compared them with 47 metal-backed cementless acetabular components implanted without bone grafting in 33 patients (group B). Patients in group A were treated between 1988 and 1993 and those in group B between 1990 and 1995. The mean follow-up for group A was 16.6 years (12 to 18) and the mean follow-up for group B was 13.4 years (10 to 16).

Revision for aseptic loosening was undertaken in 15 hips (32.6%) in group A and in four hips (8.5%) in group B. When liner exchange was included, a total of 13 hips were revised in group B (27.7%). The mean polyethylene wear was 0.11 mm/yr (0.002 to 0.43) and 0.107 mm/yr (0 to 0.62) for groups A and B, respectively. Polyethylene wear in group A was associated with linear osteolysis, and in group B with expansile osteolysis.

In patients with congenital hip disease, when 80% cover of the implant can be obtained, a cementless acetabular component appears to be acceptable and provides durable fixation. However, because of the type of osteolysis arising with these devices, early exchange of a worn liner is recommended before extensive bone loss makes revision surgery more complicated.


The Bone & Joint Journal
Vol. 106-B, Issue 5 Supple B | Pages 82 - 88
1 May 2024
Villa JM Rajschmir K Hosseinzadeh S Manrique-Succar J Grieco P Higuera-Rueda CA Riesgo AM

Aims. Large bone defects resulting from osteolysis, fractures, osteomyelitis, or metastases pose significant challenges in acetabular reconstruction for total hip arthroplasty. This study aimed to evaluate the survival and radiological outcomes of an acetabular reconstruction technique in patients at high risk of reconstruction failure (i.e. periprosthetic joint infection (PJI), poor bone stock, immunosuppressed patients), referred to as Hip Reconstruction In Situ with Screws and Cement (HiRISC). This involves a polyethylene liner embedded in cement-filled bone defects reinforced with screws and/or plates for enhanced fixation. Methods. A retrospective chart review of 59 consecutive acetabular reconstructions was performed by four surgeons in a single institution from 18 October 2018 to 5 January 2023. Cases were classified based on the Paprosky classification, excluding type 1 cases (n = 26) and including types 2 or 3 for analysis (n = 33). Radiological loosening was evaluated by an orthopaedic surgeon who was not the operating surgeon, by comparing the immediate postoperative radiographs with the ones at latest follow-up. Mean follow-up was 557 days (SD 441; 31 to 1,707). Results. Out of the 33 cases analyzed, six (18.2%) constructs required revision, with four revisions due to uncontrolled infection, one for dislocation, and one for aseptic loosening. Among the 27 non-revised constructs, only one showed wider radiolucencies compared to immediate postoperative radiographs, indicating potential loosening. Patients who underwent revision (n = 6) were significantly younger and had a higher BMI compared to those with non-revised constructs (p = 0.016 and p = 0.026, respectively). Sex, race, ethnicity, American Society of Anesthesiologists grade, infection status (patients with postoperative PJI diagnosis (septic) vs patients without such diagnosis (aseptic)), and mean follow-up did not significantly differ between revised and non-revised groups. Conclusion. The HiRISC technique may serve as a feasible short-term (about one to two years) alternative in patients with large acetabular defects, particularly in cases of PJI. Longer follow-up is necessary to establish the long-term survival of this technique. Cite this article: Bone Joint J 2024;106-B(5 Supple B):82–88


The Bone & Joint Journal
Vol. 103-B, Issue 2 | Pages 391 - 397
1 Feb 2021
Houdek MT Wunder JS Abdel MP Griffin AM Hevesi M Rose PS Ferguson PC Lewallen DG

Aims. Hip reconstruction after resection of a periacetabular chondrosarcoma is complex and associated with a high rate of complications. Previous reports have compared no reconstruction with historical techniques that are no longer used. The aim of this study was to compare the results of tantalum acetabular reconstruction to both historical techniques and no reconstruction. Methods. We reviewed 66 patients (45 males and 21 females) with a mean age of 53 years (24 to 81) who had undergone acetabular resection for chondrosarcoma. A total of 36 patients (54%) underwent acetabular reconstruction, most commonly with a saddle prosthesis (n = 13; 36%) or a tantalum total hip arthroplasty (THA) (n = 10; 28%). Mean follow-up was nine years (SD 4). Results. There was no difference in the mean age (p = 0.63), sex (p = 0.110), tumour volume (p = 0.646), or type of resection carried out (p > 0.05) between patients with and without reconstruction. Of the original 66 patients, 61 (92%) were ambulant at final follow-up. There was no difference in the proportion of patients who could walk in the reconstruction and 'no reconstruction' groups (p = 0.649). There was no difference in the mean Musculoskeletal Tumor Society (MSTS) score between patients who were reconstructed and those who were not (61% vs 56%; p = 0.378). Patients with a tantalum THA had a significantly (p = 0.015) higher mean MSTS score (78%) than those who were reconstructed with a saddle prosthesis (47%) or who had not been reconstructed (56%). Patients who had undergone reconstruction were more likely to have complications (81% vs 53%; p = 0.033). Conclusion. Reconstruction after resection of the acetabulum is technically demanding. In selected cases, reconstruction is of benefit, especially when reconstruction is by tantalum THA; however, the follow-up for these patients remains mid-term. When not feasible, patients with no reconstruction have an acceptable functional outcome. Level of Evidence: Level III Therapeutic. Cite this article: Bone Joint J 2021;103-B(2):391–397


The Bone & Joint Journal
Vol. 104-B, Issue 4 | Pages 504 - 509
1 Apr 2022
Kennedy JW Farhan-Alanie OM Young D Kelly MP Young PS

Aims. The aim of this study was to assess the clinical and radiological outcomes of an antiprotrusio acetabular cage (APC) when used in the surgical treatment of periacetabular bone metastases. Methods. This retrospective cohort study using a prospectively collected database involved 56 patients who underwent acetabular reconstruction for periacetabular bone metastases or haematological malignancy using a single APC between January 2009 and 2020. The mean follow-up was 20 months (1 to 143). The primary outcome measure was implant survival. Postoperative radiographs were analyzed for loosening and failure. Patient and implant survival were assessed using a competing risk analysis. Secondary parameters included primary malignancy, oncological treatment, surgical factors, length of stay in hospital, and postoperative complications. Results. A total of 33 patients (59%) died during the study period at a mean of 15 months postoperatively (1 to 63). No patient had radiological evidence of loosening or failure. Acetabular component survival was 100%. Three patients (5.4%) had further surgery; one (1.8%) underwent revision of the femoral component for dislocation, one required debridement with implant retention for periprosthetic joint infection, and one required closed reduction for dislocation. Using death as a competing risk, at 100 months, the probability of revision was 0.036 and the risk of death was 0.84. Conclusion. With appropriate patient selection, the antiprotrusio cage offers good implant survival, with a reasonable perioperative complication rate in this high-risk group of patients when managing metastatic disease or haematological malignancy around the acetabulum. Cite this article: Bone Joint J 2022;104-B(4):504–509


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 5 | Pages 617 - 623
1 May 2010
McNamara I Deshpande S Porteous M

The clinical and radiological results of 50 consecutive acetabular reconstructions in 48 patients using impaction grafting have been retrospectively reviewed. A 1:1 mixture of frozen, ground irradiated bone graft and Apapore 60, a synthetic bone graft substitute, was used in all cases. There were 13 complex primary and 37 revision procedures with a mean follow-up of five years (3.4 to 7.6). The clinical survival rate was 100%, with improvements in the mean Harris Hip Scores for pain and function. Radiologically, 30 acetabular grafts showed evidence of incorporation, ten had radiolucent lines and two acetabular components migrated initially before stabilising. Acetabular reconstruction in both primary and revision surgery using a 1:1 mixture of frozen, ground, irriadiated bone and Apapore 60 appears to be a reliable method of managing acetabular defects. Longer follow-up will be required to establish whether this technique is as effective as using fresh-frozen allograft


The Bone & Joint Journal
Vol. 100-B, Issue 7 | Pages 903 - 908
1 Jul 2018
Eachempati KK Malhotra R Pichai S Reddy AVG Podhili Subramani AK Gautam D Bollavaram VR Sheth NP

Aims. The advent of trabecular metal (TM) augments has revolutionized the management of severe bone defects during acetabular reconstruction. The purpose of this study was to evaluate patients undergoing revision total hip arthroplasty (THA) with the use of TM augments for reconstruction of Paprosky IIIA and IIIB defects. Patients and Methods. A retrospective study was conducted at four centres between August 2008 and January 2015. Patients treated with TM augments and TM shell for a Paprosky grade IIIA or IIIB defect, in the absence of pelvic discontinuity, and who underwent revision hip arthroplasty with the use of TM augments were included in the study. A total of 41 patients with minimum follow-up of two years were included and evaluated using intention-to-treat analysis. Results. There were 36 (87.8%) patients with a Paprosky IIIA defect and five (12.2%) patients with a Paprosky IIIB defect. The mean age was 56.7 years (28 to 94). There were 21 (51.2%) women and 20 (48.8%) men. The mean follow-up was 39.4 months (12 to 96). One (2%) patient died after eight years. No failures were noted in the series. The mean survivorship was 100% at the time of latest follow-up. Conclusion. The results of this multicentre study showed encouraging short- and mid-term results for the use of TM augments in the management of Paprosky grade IIIA and IIIB defects. Cite this article: Bone Joint J 2018;100-B:903–8


The Bone & Joint Journal
Vol. 98-B, Issue 6 | Pages 767 - 771
1 Jun 2016
Konan S Duncan CP Masri BA Garbuz DS

Aims. Reconstruction of the acetabulum after failed total hip arthroplasty (THA) can be a surgical challenge in the presence of severe bone loss. We report the long-term survival of a porous tantalum revision acetabular component, its radiological appearance and quality of life outcomes. Patients and Methods. We reviewed the results of 46 patients who had undergone revision of a failed acetabular component with a Paprosky II or III bone defect and reconstruction with a hemispherical, tantalum acetabular component, supplementary screws and a cemented polyethylene liner. Results. After a minimum follow-up of ten years (ten to 12), the survivorship of the porous tantalum acetabular component was 96%, with further revision of the acetabular component as the end point. The ten-year survivorship, with hip revision for any reason as the end point, was 92%. We noted excellent pain relief (mean Western Ontario and McMaster Universities Arthritis Index (WOMAC) score pain 92.6, (40 to 100)) and good functional outcomes (mean WOMAC function 90.3 (30.9 to 100), mean University of California Los Angeles activity scale 5 (2 to 10)) and generic quality of life measures (mean Short Form-12 (SF-12) physical component 48.3 (18.1 to 56.8), mean SF-12 mental component 56.7 (32.9 to 70.3)). Patient satisfaction with pain relief, function and return to recreational activities were excellent. Take home message: Uncemented acetabular reconstruction using a tantalum acetabular component gives excellent clinical and quality of life outcomes at a minimum follow-up of ten years. Cite this article: Bone Joint J 2016;98-B:767–71


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 9 | Pages 1194 - 1200
1 Sep 2011
Akiyama H Yamamoto K Tsukanaka M Kawanabe K Otsuka H So K Goto K Nakamura T

We retrospectively reviewed 40 hips in 36 patients who had undergone acetabular reconstruction using a titanium Kerboull-type acetabular reinforcement device with bone allografts between May 2001 and April 2006. Impacted bone allografts were used for the management of American Academy of Orthopaedic Surgeons Type II defects in 17 hips, and bulk bone allografts together with impacted allografts were used for the management of Type III defects in 23 hips. A total of five hips showed radiological failure at a mean follow-up of 6.7 years (4.5 to 9.3), two of which were infected. The mean pre-operative Merle d’Aubigné score was 10 (5 to 15) vs 13.6 (9 to 18) at the latest follow-up. The Kaplan-Meier survival rate at ten years, calculated using radiological failure or revision of the acetabular component for any reason as the endpoint, was 87% (95% confidence interval 76.3 to 97.7). A separate experimental analysis of the mechanical properties of the device and the load-displacement properties of bone grafts showed that a structurally hard allograft resected from femoral heads of patients with osteoarthritis should be preferentially used in any type of defect. If impacted bone allografts were used, a bone graft thickness of < 25 mm was acceptable in Type II defects. This clinical study indicates that revision total hip replacement using the Kerboull-type acetabular reinforcement device with bone allografts yielded satisfactory mid-term results


The Bone & Joint Journal
Vol. 98-B, Issue 1_Supple_A | Pages 68 - 72
1 Jan 2016
Goodman GP Engh Jr CA

The custom triflange is a patient-specific implant for the treatment of severe bone loss in revision total hip arthroplasty (THA). Through a process of three-dimensional modelling and prototyping, a hydroxyapatite-coated component is created for acetabular reconstruction. There are seven level IV studies describing the clinical results of triflange components. The most common complications include dislocation and infection, although the rates of implant removal are low. Clinical results are promising given the challenging problem. We describe the design, manufacture and implantation process and review the clinical results, contrasting them to other methods of acetabular reconstruction in revision THA. Cite this article: Bone Joint J 2016;98-B(1 Suppl A):68–72


The Bone & Joint Journal
Vol. 106-B, Issue 5 Supple B | Pages 54 - 58
1 May 2024
Wassilew GI Zimmerer A Fischer M Nonnenmacher L O'Hara L Hube R

Aims

The use of a porous metal shell supported by two augments with the ‘footing’ technique is one solution to manage Paprosky IIIB acetabular defects in revision total hip arthroplasty. The aim of this study was to assess the medium-term implant survival and radiological and clinical outcomes of this technique.

Methods

We undertook a retrospective, two-centre series of 39 hips in 39 patients (15 male, 24 female) treated with the ‘footing’ technique for Paprosky IIIB acetabular defects between 2007 and 2020. The median age at the time of surgery was 64.4 years (interquartile range (IQR) 54.4 to 71.0). The median follow-up was 3.9 years (IQR 3.1 to 7.0).


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 12 | Pages 1555 - 1560
1 Dec 2009
Lingaraj K Teo YH Bergman N

We investigated the early results of modular porous metal components used in 23 acetabular reconstructions associated with major bone loss. The series included seven men and 15 women with a mean age of 67 years (38 to 81), who had undergone a mean of two previous revisions (1 to 7). Based on Paprosky’s classification, there were 17 type 3A and six type 3B defects. Pelvic discontinuity was noted in one case. Augments were used in 21 hips to support the shell and an acetabular component-cage construct was implanted in one case. At a mean follow-up of 41 months (24 to 62), 22 components remained well fixed. Two patients required rerevision of the liners for prosthetic joint instability. Clinically, the mean Harris Hip Score improved from 43.0 pre-operatively (14 to 86) to 75.7 post-operatively (53 to 100). The mean pre-operative Merle d’Aubigné score was 8.2 (3 to 15) and improved to a mean of 13.7 (11 to 18) post-operatively. These short-term results suggest that modular porous metal components are a viable option in the reconstruction of Paprosky type 3 acetabular defects. More data are needed to determine whether the system yields greater long-term success than more traditional methods, such as reconstruction cages and structural allografts


The Bone & Joint Journal
Vol. 106-B, Issue 5 Supple B | Pages 47 - 53
1 May 2024
Jones SA Parker J Horner M

Aims

The aims of this study were to determine the success of a reconstruction algorithm used in major acetabular bone loss, and to further define the indications for custom-made implants in major acetabular bone loss.

Methods

We reviewed a consecutive series of Paprosky type III acetabular defects treated according to a reconstruction algorithm. IIIA defects were planned to use a superior augment and hemispherical acetabular component. IIIB defects were planned to receive either a hemispherical acetabular component plus augments, a cup-cage reconstruction, or a custom-made implant. We used national digital health records and registry reports to identify any reoperation or re-revision procedure and Oxford Hip Score (OHS) for patient-reported outcomes. Implant survival was determined via Kaplan-Meier analysis.


The Bone & Joint Journal
Vol. 106-B, Issue 5 Supple B | Pages 74 - 81
1 May 2024
Callary SA Broekhuis D Barends J Ramasamy B Nelissen RGHH Solomon LB Kaptein BL

Aims

The aim of this study was to compare the biomechanical models of two frequently used techniques for reconstructing severe acetabular defects with pelvic discontinuity in revision total hip arthroplasty (THA) – the Trabecular Metal Acetabular Revision System (TMARS) and custom triflange acetabular components (CTACs) – using virtual modelling.

Methods

Pre- and postoperative CT scans from ten patients who underwent revision with the TMARS for a Paprosky IIIB acetabular defect with pelvic discontinuity were retrospectively collated. Computer models of a CTAC implant were designed from the preoperative CT scans of these patients. Computer models of the TMARS reconstruction were segmented from postoperative CT scans using a semi-automated method. The amount of bone removed, the implant-bone apposition that was achieved, and the restoration of the centre of rotation of the hip were compared between all the actual TMARS and the virtual CTAC implants.


The Bone & Joint Journal
Vol. 106-B, Issue 4 | Pages 312 - 318
1 Apr 2024
Sheth NP Jones SA Sanghavi SA Manktelow A

The advent of modular porous metal augments has ushered in a new form of treatment for acetabular bone loss. The function of an augment can be seen as reducing the size of a defect or reconstituting the anterosuperior/posteroinferior columns and/or allowing supplementary fixation. Depending on the function of the augment, the surgeon can decide on the sequence of introduction of the hemispherical shell, before or after the augment. Augments should always, however, be used with cement to form a unit with the acetabular component. Given their versatility, augments also allow the use of a hemispherical shell in a position that restores the centre of rotation and biomechanics of the hip. Progressive shedding or the appearance of metal debris is a particular finding with augments and, with other radiological signs of failure, should be recognized on serial radiographs. Mid- to long-term outcomes in studies reporting the use of augments with hemispherical shells in revision total hip arthroplasty have shown rates of survival of > 90%. However, a higher risk of failure has been reported when augments have been used for patients with chronic pelvic discontinuity.

Cite this article: Bone Joint J 2024;106-B(4):312–318.


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 9 | Pages 1148 - 1153
1 Sep 2009
Schreurs BW Keurentjes JC Gardeniers JWM Verdonschot N Slooff TJJH Veth RPH

We present an update of the clinical and radiological results of 62 consecutive acetabular revisions using impacted morsellised cancellous bone grafts and a cemented acetabular component in 58 patients, at a mean follow-up of 22.2 years (20 to 25). The Kaplan-Meier survivorship for the acetabular component with revision for any reason as the endpoint was 75% at 20 years (95% confidence interval (CI) 62 to 88) when 16 hips were at risk. Excluding two revisions for septic loosening at three and six years, the survivorship at 20 years was 79% (95% CI 67 to 93). With further exclusions of one revision of a well-fixed acetabular component after 12 years during a femoral revision and two after 17 years for wear of the acetabular component, the survivorship for aseptic loosening was 87% at 20 years (95% CI 76 to 97). At the final review 14 of the 16 surviving hips had radiographs available. There was one additional case of radiological loosening and four acetabular reconstructions showed progressive radiolucent lines in one or two zones. Acetabular revision using impacted large morsellised bone chips (0.5 cm to 1 cm in diameter) and a cemented acetabular component remains a reliable technique for reconstruction, even when assessed at more than 20 years after surgery


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 9 | Pages 1149 - 1154
1 Sep 2007
Lian Y Yoo M Pei F Cho Y Cheng J Chun S

We performed 52 total hip replacements in 52 patients using a cementless acetabular component combined with a circumferential osteotomy of the medial acetabular wall for the late sequelae of childhood septic arthritis of the hip. The mean age of the patients at operation was 44.5 years (22 to 66) and the mean follow-up was 7.8 years (5 to 11.8). The mean improvement in the Harris Hip Score was 29.6 points (19 to 51) at final follow-up. The mean cover of the acetabular component was 98.5% (87.8% to 100%). The medial acetabular wall was preserved with a mean thickness of 8.3 mm (1.7 to 17.4) and the mean length of abductor lever arm increased from 43.4 mm (19.1 to 62) to 54.2 mm (36.5 to 68.6). One acetabular component was revised for loosening and osteolysis 4.5 years postoperatively, and one had radiolucent lines in all acetabular zones at final review. Kaplan-Meier survival was 94.2% (95% confidence interval 85.8% to 100%) at 7.3 years, with revision or radiological loosening as an end-point when two hips were at risk. A cementless acetabular component combined with circumferential medial acetabular wall osteotomy provides favourable results for acetabular reconstruction in patients who present with late sequelae of childhood septic hip arthritis


The Bone & Joint Journal
Vol. 106-B, Issue 4 | Pages 359 - 364
1 Apr 2024
Özdemir E de Lange B Buckens CFM Rijnen WHC Visser J

Aims

To investigate the extent of bone development around the scaffold of custom triflange acetabular components (CTACs) over time.

Methods

We performed a single-centre historical prospective cohort study, including all patients with revision THA using the aMace CTAC between January 2017 and March 2021. A total of 18 patients (18 CTACs) were included. Models of the hemipelvis and the scaffold component of the CTACs were created by segmentation of CT scans. The CT scans were performed immediately postoperatively and at least one year after surgery. The amount of bone in contact with the scaffold was analyzed at both times, and the difference was calculated.


The Bone & Joint Journal
Vol. 106-B, Issue 5 Supple B | Pages 59 - 65
1 May 2024
Liu WKT Cheung A Fu H Chan PK Chiu KY

Aims

Isolated acetabular liner exchange with a highly crosslinked polyethylene (HXLPE) component is an option to address polyethylene wear and osteolysis following total hip arthroplasty (THA) in the presence of a well-fixed acetabular shell. The liner can be fixed either with the original locking mechanism or by being cemented within the acetabular component. Whether the method used for fixation of the HXLPE liner has any bearing on the long-term outcomes is still unclear.

Methods

Data were retrieved for all patients who underwent isolated acetabular component liner exchange surgery with a HXLPE component in our institute between August 2000 and January 2015. Patients were classified according to the fixation method used (original locking mechanism (n = 36) or cemented (n = 50)). Survival and revision rates were compared. A total of 86 revisions were performed and the mean duration of follow-up was 13 years.


The Bone & Joint Journal
Vol. 106-B, Issue 5 Supple B | Pages 66 - 73
1 May 2024
Chaudhry F Daud A Greenberg A Braunstein D Safir OA Gross AE Kuzyk PR

Aims

Pelvic discontinuity is a challenging acetabular defect without a consensus on surgical management. Cup-cage reconstruction is an increasingly used treatment strategy. The present study evaluated implant survival, clinical and radiological outcomes, and complications associated with the cup-cage construct.

Methods

We included 53 cup-cage construct (51 patients) implants used for hip revision procedures for pelvic discontinuity between January 2003 and January 2022 in this retrospective review. Mean age at surgery was 71.8 years (50.0 to 92.0; SD 10.3), 43/53 (81.1%) were female, and mean follow-up was 6.4 years (0.02 to 20.0; SD 4.6). Patients were implanted with a Trabecular Metal Revision Shell with either a ZCA cage (n = 12) or a TMARS cage (n = 40, all Zimmer Biomet). Pelvic discontinuity was diagnosed on preoperative radiographs and/or intraoperatively. Kaplan-Meier survival analysis was performed, with failure defined as revision of the cup-cage reconstruction.


The Bone & Joint Journal
Vol. 95-B, Issue 2 | Pages 230 - 238
1 Feb 2013
Giannoudis PV Kanakaris NK Delli Sante E Morell DJ Stengel D Prevezas N

Over a five-year period, adult patients with marginal impaction of acetabular fractures were identified from a registry of patients who underwent acetabular reconstruction in two tertiary referral centres. Fractures were classified according to the system of Judet and Letournel. A topographic classification to describe the extent of articular impaction was used, dividing the joint surface into superior, middle and inferior thirds. Demographic information, hospitalisation and surgery-related complications, functional (EuroQol 5-D) and radiological outcome according to Matta’s criteria were recorded and analysed. In all, 60 patients (57 men, three women) with a mean age of 41 years (18 to 72) were available at a mean follow-up of 48 months (24 to 206). The quality of the reduction was ‘anatomical’ in 44 hips (73.3%) and ‘imperfect’ in 16 (26.7%). The originally achieved anatomical reduction was lost in12 patients (25.8%). Radiologically, 33 hips (55%) were graded as ‘excellent’, 11 (18.3%) as ‘good’, one (1.7%) as ‘fair’ and 15 (25%) as ‘poor’. A total of 11 further operations were required in 11 cases, of which six were total hip replacements. Univariate linear regression analysis of the functional outcome showed that factors associated with worse pain were increasing age and an inferior location of the impaction. Elevation of the articular impaction leads to joint preservation with satisfactory overall medium-term functional results, but secondary collapse is likely to occur in some patients. Cite this article: Bone Joint J 2013;95-B:230–8


The Bone & Joint Journal
Vol. 104-B, Issue 10 | Pages 1180 - 1188
1 Oct 2022
Qu H Mou H Wang K Tao H Huang X Yan X Lin N Ye Z

Aims

Dislocation of the hip remains a major complication after periacetabular tumour resection and endoprosthetic reconstruction. The position of the acetabular component is an important modifiable factor for surgeons in determining the risk of postoperative dislocation. We investigated the significance of horizontal, vertical, and sagittal displacement of the hip centre of rotation (COR) on postoperative dislocation using a CT-based 3D model, as well as other potential risk factors for dislocation.

Methods

A total of 122 patients who underwent reconstruction following resection of periacetabular tumour between January 2011 and January 2020 were studied. The risk factors for dislocation were investigated with univariate and multivariate logistic regression analysis on patient-specific, resection-specific, and reconstruction-specific variables.


The Bone & Joint Journal
Vol. 103-B, Issue 7 | Pages 1247 - 1253
1 Jul 2021
Slullitel PA Oñativia JI Zanotti G Comba F Piccaluga F Buttaro MA

Aims

There is a paucity of long-term studies analyzing risk factors for failure after single-stage revision for periprosthetic joint infection (PJI) following total hip arthroplasty (THA). We report the mid- to long-term septic and non-septic failure rate of single-stage revision for PJI after THA.

Methods

We retrospectively reviewed 88 cases which met the Musculoskeletal Infection Society (MSIS) criteria for PJI. Mean follow-up was seven years (1 to 14). Septic failure was diagnosed with a Delphi-based consensus definition. Any reoperation for mechanical causes in the absence of evidence of infection was considered as non-septic failure. A competing risk regression model was used to evaluate factors associated with septic and non-septic failures. A Kaplan-Meier estimate was used to analyze mortality.


The Bone & Joint Journal
Vol. 103-B, Issue 4 | Pages 795 - 803
1 Apr 2021
Fujiwara T Medellin Rincon MR Sambri A Tsuda Y Clark R Stevenson J Parry MC Grimer RJ Jeys L

Aims

Limb salvage for pelvic sarcomas involving the acetabulum is a major surgical challenge. There remains no consensus about what is the optimum type of reconstruction after resection of the tumour. The aim of this study was to evaluate the surgical outcomes in these patients according to the methods of periacetabular reconstruction.

Methods

The study involved a consecutive series of 122 patients with a periacetabular bone sarcoma who underwent limb-salvage surgery involving a custom-made prosthesis in 65 (53%), an ice-cream cone prosthesis in 21 (17%), an extracorporeal irradiated autograft in 18 (15%), and nonskeletal reconstruction in 18 (15%).


The Bone & Joint Journal
Vol. 96-B, Issue 2 | Pages 188 - 194
1 Feb 2014
Gilbody J Taylor C Bartlett GE Whitehouse SL Hubble MJW Timperley AJ Howell JR Wilson MJ

Impaction bone grafting for the reconstitution of bone stock in revision hip surgery has been used for nearly 30 years. Between 1995 and 2001 we used this technique in acetabular reconstruction, in combination with a cemented component, in 304 hips in 292 patients revised for aseptic loosening. The only additional supports used were stainless steel meshes placed against the medial wall or laterally around the acetabular rim to contain the graft. All Paprosky grades of defect were included. Clinical and radiographic outcomes were collected in surviving patients at a minimum of ten years after the index operation. Mean follow-up was 12.4 years (. sd. 1.5) (10.0 to 16.0). Kaplan–Meier survival with revision for aseptic loosening as the endpoint was 85.9% (95% CI 81.0 to 90.8) at 13.5 years. Clinical scores for pain relief remained satisfactory, and there was no difference in clinical scores between cups that appeared stable and those that appeared radiologically loose. . Cite this article: Bone Joint J 2014;96-B:188–94


The Bone & Joint Journal
Vol. 103-B, Issue 1 | Pages 79 - 86
1 Jan 2021
Slullitel PA Oñativia JI Cima I Zanotti G Comba F Piccaluga F Buttaro MA

Aims

We aimed to report the mid- to long-term rates of septic and aseptic failure after two-stage revision surgery for periprosthetic joint infection (PJI) following total hip arthroplasty (THA).

Methods

We retrospectively reviewed 96 cases which met the Musculoskeletal Infection Society criteria for PJI. The mean follow-up was 90 months (SD 32). Septic failure was assessed using a Delphi-based consensus definition. Any further surgery undertaken for aseptic mechanical causes was considered as aseptic failure. The cumulative incidence with competing risk analysis was used to predict the risk of septic failure. A regression model was used to evaluate factors associated with septic failure. The cumulative incidence of aseptic failure was also analyzed.


The Journal of Bone & Joint Surgery British Volume
Vol. 88-B, Issue 2 | Pages 173 - 178
1 Feb 2006
de Jong PT Haverkamp D van der Vis HM Marti RK

We evaluated the long-term results of 116 total hip replacements with a superolateral shelfplasty in 102 patients with osteoarthritis secondary to developmental dysplasia of the hip. After a mean follow-up of 19.5 years (11.5 to 26.0), 14 acetabular components (12%) had been revised. The cumulative survival at 20 years was 78%, with revision for loosening of the acetabular component as the end-point. All grafts were well integrated and showed remodelling. In six grafts some resorption had occurred under the heads of the screws where the graft was not supporting the socket. Apart from these 14 revisions, seven acetabular components had possible radiological signs of loosening at a mean follow-up of 14.5 years, one had signs of probable loosening, and five had signs of definite loosening. These results indicate that this technique of bone grafting for acetabular reconstruction in hip dysplasia is a durable solution for cemented acetabular components


The Journal of Bone & Joint Surgery British Volume
Vol. 86-B, Issue 4 | Pages 492 - 497
1 May 2004
Schreurs BW Bolder SBT Gardeniers JWM Verdonschot N Slooff TJJH Veth RPH

This study presents the clinical and radiological results of 62 consecutive acetabular revisions in 58 patients, at a mean of 16.5 years follow-up (15 to 20). The Kaplan-Meier survivorship for the cup with end-point revisions for any reason, was 79% at 15 years (95% confidence interval (CI); 67 to 91). Excluding two revisions for septic loosening at three and six years, and one revision of a well-fixed cup after 12 years in the course of a femoral revision, the survivorship was 84% at 15 years (95% CI; 73 to 95). At review there were no additional cases of loosening, although seven acetabular reconstructions showed radiolucent lines in one or two zones. Acetabular revision using impacted large morsellised bone chips (0.7 cm to 1.0 cm) and a cemented cup, is a reliable technique of reconstruction, when assessed at more than 15 years


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 1 | Pages 26 - 31
1 Jan 2007
Kawanabe K Akiyama H Onishi E Nakamura T

We retrospectively evaluated 42 hips which had undergone acetabular reconstruction using the Kerboull acetabular reinforcement device between September 1994 and December 1998. We used autogenous bone chips from the ilium and ceramic particle morsellised grafts, even in large acetabular bone defects, in the early stages of the study. Thereafter, femoral head allograft was used as bulk graft in patients with large acetabular defects. Ceramic blocks and the patients’ contralateral femoral head were also used as bulk graft. The mean follow-up period was 8.7 years (4.3 to 12). Survivorship analysis was performed using radiological failure of the acetabular component, irrespective of whether it was revised, or not, as the end-point. The survival rate of the morsellised graft group (25 hips) and the bulk graft group (17 hips) at ten years was 53% (95% confidence interval (CI) 42.5% to 63.5%) and 82% (95% CI 72.4% to 91.6%), respectively. The mid-term results of revision total hip replacement with the Kerboull device were better when bulk graft was used in any size of bone defect


The Journal of Bone & Joint Surgery British Volume
Vol. 75-B, Issue 5 | Pages 755 - 759
1 Sep 1993
Tasnavites A Murray D Benson M

After late reduction of the hip in children with developmental dysplasia the acetabular contour slowly improves and it is difficult to know if and when acetabular reconstruction is required. We studied the radiographs of 19 patients with unilateral dislocated or subluxated hips which had been reduced between the ages of one and two years. Preoperatively, all the affected hips showed acetabular dysplasia. After reduction they steadily improved for three years by which time none was dysplastic as measured by the acetabular index. After the age of ten years, when assessed by the more sensitive centre-edge angle, two were found to be dysplastic. It was not possible to predict these from early radiographs. We conclude that the decision to reconstruct an acetabulum should not be taken until three years after reduction, and that a few hips which appear to be developing satisfactorily at that time will, nevertheless, become dysplastic


The Journal of Bone & Joint Surgery British Volume
Vol. 88-B, Issue 6 | Pages 721 - 726
1 Jun 2006
Girard J Lavigne M Vendittoli P Roy AG

We have compared the biomechanical nature of the reconstruction of the hip in conventional total hip arthroplasty (THA) and surface replacement arthroplasty (SRA) in a randomised study involving 120 patients undergoing unilateral primary hip replacement. The contralateral hip was used as a control. Post-operatively, the femoral offset was significantly increased with THA (mean 5.1 mm; −2.8 to 11.6) and decreased with SRA (mean −3.3 mm; −8.9 to 8.2). Femoral offset was restored within . sd. 4 mm in 14 (25%) of those with THA and in 28 (57%) of the patients receiving SRA (p < 0.001). In the THA group, the leg was lengthened by a mean of 2.6 mm (−6.04 to +12.9), whereas it was shortened by a mean of 1.9 mm (−7.1 to +2.05) in the SRA group, compared with the contralateral side. Leg-length inequality was restored within . sd. 4 mm in 42 (86%) of the SRA and 33 (60%) of the THA patients. The radiological parameters of acetabular reconstruction were similar in both groups. Restoration of the normal proximal femoral anatomy was more precise with SRA. The enhanced stability afforded by the use of a large-diameter femoral head avoided over-lengthening of the limb or increased offset to improve soft-tissue tension as occurs sometimes in THA. In a subgroup of patients with significant pre-operative deformity, restoration of the normal hip anatomy with lower pre-operative femoral offset or significant shortening of the leg was still possible with SRA


The Journal of Bone & Joint Surgery British Volume
Vol. 83-B, Issue 2 | Pages 177 - 182
1 Mar 2001
Taylor AH Shannon M Whitehouse SL Lee MB Learmonth ID

We describe the results of 76 total arthroplasties of the hip for stage-III or stage-IV avascular necrosis of the femoral head. Harris Galante Porous cups were used in 63 patients between 1986 and 1994 and followed prospectively. We reviewed 70 hips with a follow-up of more than five years (mean 7.6). At the latest review the mean Harris Hip Score had improved from a preoperative value of 29 ± 14.7 to 94 ± 6.8. Radiologically, there was no evidence of acetabular migration. The rate of revision for the femoral component was 8.6%, three being undertaken for loosening and three to allow downsizing of the femoral head. The rate of revision for the acetabular component was 7.1% (five cups). At the time of revision none of the cups was clinically loose, and only required the liner to be changed. The rate of complications was low with no case of deep infection or dislocation, but nine of the 76 hips (11.8%) showed grade-III heterotopic ossification. Previous studies of patients undergoing cemented total hip arthroplasty for the treatment of advanced avascular necrosis have indicated a high incidence of loosening of the acetabular component. Our findings show good medium-term results using the Harris Galante Porous cup for acetabular reconstruction, together with a variety of cemented femoral components, for the treatment of this difficult problem


The Bone & Joint Journal
Vol. 103-B, Issue 1 | Pages 184 - 191
1 Jan 2021
Perrin DL Visgauss JD Wilson DA Griffin AM Abdul Razak AR Ferguson PC Wunder JS

Aims

Local recurrence remains a challenging and common problem following curettage and joint-sparing surgery for giant cell tumour of bone (GCTB). We previously reported a 15% local recurrence rate at a median follow-up of 30 months in 20 patients with high-risk GCTB treated with neoadjuvant Denosumab. The aim of this study was to determine if this initial favourable outcome following the use of Denosumab was maintained with longer follow-up.

Methods

Patients with GCTB of the limb considered high-risk for unsuccessful joint salvage, due to minimal periarticular and subchondral bone, large soft tissue mass, or pathological fracture, were treated with Denosumab followed by extended intralesional curettage with the goal of preserving the joint surface. Patients were followed for local recurrence, metastasis, and secondary sarcoma.


The Journal of Bone & Joint Surgery British Volume
Vol. 82-B, Issue 6 | Pages 830 - 836
1 Aug 2000
Haddad FS Garbuz DS Masri BA Duncan CP

There are few medium- and long-term data on the outcome of the use of proximal femoral structural allografts in revision hip arthroplasty. This is a study of a consecutive series of 40 proximal femoral allografts performed for failed total hip replacements using the same technique with a minimum follow-up of five years (mean 8.8 years; range 5 to 11.5 years). In all cases the stem was cemented into both the allograft and the host femur. The proximal femur of the host was resected in 37 cases. There were four early revisions (10%), two for infection, one for nonunion of the allograft-host junction, and one for allograft resorption noted at the time of revision of a failed acetabular reconstruction. Junctional nonunion was seen in three patients (8%), two of whom were managed successfully by bone grafting, and bone grafting and plating respectively. Instability was observed in four (10%). Trochanteric nonunion was seen in 18 patients (46%) and trochanteric escape in ten of these (27%). The mean Harris hip score improved from 39 to 79. Severe resorption involving the full thickness of the allograft was seen in seven patients (17.5%). This progressed rapidly and silently, but has yet to cause failure of any of the reconstructions. Profound resorption of the allograft may be related to a combination of factors, including a slow form of immune rejection, stress shielding and resorption due to mechanical disuse with solid cemented distal fixation, and the absence of any masking or protective effect which may be provided by the retention of the bivalved host bone as a vascularised onlay autograft. Although continued surveillance is warranted, the very good medium-term clinical results justify the continued use of structural allografts for failed total hip replacements with severe loss of proximal femoral bone


The Bone & Joint Journal
Vol. 102-B, Issue 10 | Pages 1311 - 1318
3 Oct 2020
Huang Y Gao Y Li Y Ding L Liu J Qi X

Aims

Morphological abnormalities are present in patients with developmental dysplasia of the hip (DDH). We studied and compared the pelvic anatomy and morphology between the affected hemipelvis with the unaffected side in patients with unilateral Crowe type IV DDH using 3D imaging and analysis.

Methods

A total of 20 patients with unilateral Crowe-IV DDH were included in the study. The contralateral side was considered normal in all patients. A coordinate system based on the sacral base (SB) in a reconstructed pelvic model was established. The pelvic orientations (tilt, rotation, and obliquity) of the affected side were assessed by establishing a virtual anterior pelvic plane (APP). The bilateral coordinates of the anterior superior iliac spine (ASIS) and the centres of hip rotation were established, and parameters concerning size and volume were compared for both sides of the pelvis.


The Journal of Bone & Joint Surgery British Volume
Vol. 42-B, Issue 4 | Pages 689 - 705
1 Nov 1960
MacKenzie IG Seddon HJ Trevor D

1. The results of treatment of 134 patients with congenital dislocation of 167 hips are reviewed. 2. Late diagnosis is still a major problem. 3. Subluxations rarely give rise to poor results, but in dislocations first treated over the age of five years there is a one-in-three risk of failure. 4. Manipulative reduction is successful less often than reduction on a frame and carries a higher risk of avascular necrosis. 5. Closed reduction on a frame was satisfactory in 58 per cent of patients under the age of three years, and can succeed up to the age of five. 6. Open reduction was required in 20 per cent of cases under the age of three, and can be used successfully up to the age of six. 7. Seven anatomical barriers to closed reduction have been recognised and two or more are commonly found in one hip when open reduction is performed. 8. The acetabular roof may fail to develop after reduction, especially if this is delayed. A C.E. angle of under 20 degrees does not necessarily forebode this, unless measured on an arthrograph. Sclerosis of a sloping acetabular roof is an indication for operation. Acetabuloplasty is the proper operation for a sloping acetabulum and can be done successfully up to the age of twelve. Over this age, a shelf operation should be performed; this is appropriate also in younger patients in whom the curvature of the acetabulum is normal but does not extend far enough laterally. These operations were required in 38 per cent of hips treated in patients under the age of three, and in 64 per cent over this age. There is a one-in-three risk of avascular necrosis when acetabular reconstruction is done in patients under three. 9. Anteversion, if excessive, should be corrected by subtrochanteric osteotomy, and any valgus of the femoral neck should be corrected simultaneously. 10. Unilateral dislocations in patients over the age of six are best treated by Colonna's operation. In our few bilateral cases over this age our results have been disappointing. 11. Avascular necrosis is less common but more serious when it occurs over the age of three. Manipulative reduction and the use of frog-leg plasters are two avoidable factors which appear to increase its incidence. The more serious degrees are accompanied by stiffness of the hip, and when this sign is present weight bearing should be avoided. 12. Prolonged, though rarely permanent, limitation of movement occurs in some 10 per cent of cases. In a few, operative correction was required


The Bone & Joint Journal
Vol. 101-B, Issue 5 | Pages 522 - 528
1 May 2019
Medellin MR Fujiwara T Clark R Stevenson JD Parry M Jeys L

Aims

The aim of this study was to evaluate the prosthesis characteristics and associated conditions that may modify the survival of total femoral endoprosthetic replacements (TFEPR).

Patients and Methods

In all, 81 patients treated with TFEPR from 1976 to 2017 were retrospectively evaluated and failures were categorized according to the Henderson classification. There were 38 female patients (47%) and 43 male patients (53%) with a mean age at diagnosis of 43 years (12 to 86). The mean follow-up time was 10.3 years (0 to 31.7). A survival analysis was performed followed by univariate and multivariate Cox regression to identify independent implant survival factors.


The Bone & Joint Journal
Vol. 101-B, Issue 9 | Pages 1050 - 1057
1 Sep 2019
Lampropoulou-Adamidou K Hartofilakidis G

Aims

To our knowledge, no study has compared the long-term results of cemented and hybrid total hip arthroplasty (THA) in patients with osteoarthritis (OA) secondary to congenital hip disease (CHD). This is a demanding procedure that may require special techniques and implants. Our aim was to compare the long-term outcome of cemented low-friction arthroplasty (LFA) and hybrid THA performed by one surgeon.

Patients and Methods

Between January 1989 and December 1997, 58 hips (44 patients; one man, 43 woman; mean age 56.6 years (25 to 77)) with OA secondary to CHD were treated with a cemented Charnley LFA (group A), and 55 hips (39 patients; two men, 37 women; mean age 49.1 years (27 to 70)) were treated with a hybrid THA (group B), by the senior author (GH). The clinical outcome and survivorship were compared.


The Bone & Joint Journal
Vol. 101-B, Issue 4 | Pages 378 - 385
1 Apr 2019
García-Rey E Carbonell-Escobar R Cordero-Ampuero J García-Cimbrelo E

Aims

We previously reported the long-term results of the cementless Duraloc-Profile total hip arthroplasty (THA) system in a 12- to 15-year follow-up study. In this paper, we provide an update on the clinical and radiological results of a previously reported cohort of patients at 23 to 26 years´ follow-up.

Patients and Methods

Of the 99 original patients (111 hips), 73 patients (82 hips) with a mean age of 56.8 years (21 to 70) were available for clinical and radiological study at a minimum follow-up of 23 years. There were 40 female patients (44 hips) and 33 male patients (38 hips).


The Bone & Joint Journal
Vol. 101-B, Issue 3 | Pages 311 - 316
1 Mar 2019
Löchel J Janz V Hipfl C Perka C Wassilew GI

Aims

The use of trabecular metal (TM) shells supported by augments has provided good mid-term results after revision total hip arthroplasty (THA) in patients with a bony defect of the acetabulum. The aim of this study was to assess the long-term implant survivorship and radiological and clinical outcomes after acetabular revision using this technique.

Patients and Methods

Between 2006 and 2010, 60 patients (62 hips) underwent acetabular revision using a combination of a TM shell and augment. A total of 51 patients (53 hips) had complete follow-up at a minimum of seven years and were included in the study. Of these patients, 15 were men (29.4%) and 36 were women (70.6%). Their mean age at the time of revision THA was 64.6 years (28 to 85). Three patients (5.2%) had a Paprosky IIA defect, 13 (24.5%) had a type IIB defect, six (11.3%) had a type IIC defect, 22 (41.5%) had a type IIIA defect, and nine (17%) had a type IIIB defect. Five patients (9.4%) also had pelvic discontinuity.


The Bone & Joint Journal
Vol. 101-B, Issue 1_Supple_A | Pages 41 - 45
1 Jan 2019
Jones CW De Martino I D’Apolito R Nocon AA Sculco PK Sculco TP

Aims

Instability continues to be a troublesome complication after total hip arthroplasty (THA). Patient-related risk factors associated with a higher dislocation risk include the preoperative diagnosis, an age of 75 years or older, high body mass index (BMI), a history of alcohol abuse, and neurodegenerative diseases. The goal of this study was to assess the dislocation rate, radiographic outcomes, and complications of patients stratified as high-risk for dislocation who received a dual mobility (DM) bearing in a primary THA at a minimum follow-up of two years.

Materials and Methods

We performed a retrospective review of a consecutive series of DM THA performed between 2010 and 2014 at our institution (Hospital for Special Surgery, New York, New York) by a single, high-volume orthopaedic surgeon employing a single prosthesis design (Anatomic Dual Mobility (ADM) Stryker, Mahwah, New Jersey). Patient medical records and radiographs were reviewed to confirm the type of implant used, to identify any preoperative risk factors for dislocation, and any complications. Radiographic analysis was performed to assess for signs of osteolysis or remodelling of the acetabulum.