Advertisement for orthosearch.org.uk
Results 1 - 50 of 191
Results per page:
The Bone & Joint Journal
Vol. 103-B, Issue 1 | Pages 98 - 104
1 Jan 2021
van Ooij B Sierevelt IN van der Vis HM Hoornenborg D Haverkamp D

Aims. For many designs of total knee arthroplasty (TKA) it remains unclear whether cemented or uncemented fixation provides optimal long-term survival. The main limitation in most studies is a retrospective or non-comparative study design. The same is true for comparative trials looking only at the survival rate as extensive sample sizes are needed to detect true differences in fixation and durability. Studies using radiostereometric analysis (RSA) techniques have shown to be highly predictive in detecting late occurring aseptic loosening at an early stage. To investigate the difference in predicted long-term survival between cemented, uncemented, and hybrid fixation of TKA, we performed a randomized controlled trial using RSA. Methods. A total of 105 patients were randomized into three groups (cemented, uncemented, and hybrid fixation of the ACS Mobile Bearing (ACS MB) knee system, implantcast). RSA examinations were performed on the first day after surgery and at scheduled follow-up visits at three months, six months, one year, and two years postoperatively. Patient-reported outcome measures (PROMs) were obtained preoperatively and after two years follow-up. Patients and follow-up investigators were blinded for the result of randomization. Results. RSA secondary stabilization did not show a significant difference between the three types of fixation. A maximum total point motion of less than 0.2 mm in the second postoperative year was shown in each group, which suggests stabilization of the implant. At 24 months after surgery, PROMs significantly improved compared to baseline in all treatment groups. No significant difference was observed between the three groups. Conclusion. Secondary stabilization measurements in this study demonstrated no significant difference between the groups. In all groups migration stabilized after initial settling of the implant. For this implant the long-term outcome is not expected to be influenced by the type of fixation to the bone. Cite this article: Bone Joint J 2021;103-B(1):98–104


The Bone & Joint Journal
Vol. 103-B, Issue 5 | Pages 855 - 863
1 May 2021
Koster LA Meinardi JE Kaptein BL Van der Linden - Van der Zwaag E Nelissen RGHH

Aims. The objective of this study was to compare the two-year migration pattern and clinical outcomes of a total knee arthroplasty (TKA) with an asymmetrical tibial design (Persona PS) and a well-proven TKA with a symmetrical tibial design (NexGen LPS). Methods. A randomized controlled radiostereometric analysis (RSA) trial was conducted including 75 cemented posterior-stabilized TKAs. Implant migration was measured with RSA. Maximum total point motion (MTPM), translations, rotations, clinical outcomes, and patient-reported outcome measures (PROMs) were assessed at one week postoperatively and at three, six, 12, and 24 months postoperatively. Results. A linear mixed-effect model using RSA data of 31 asymmetrical and 38 symmetrical TKAs did not show a difference in mean MTPM migration pattern of the tibial or femoral components. Mean tibial component MTPM at two years postoperative of the asymmetrical TKA design was 0.93 mm and 1.00 mm for the symmetrical design. For the femoral component these values were 1.04 mm and 1.14 mm, respectively. No significant differences were observed in other migration parameters or in clinical and PROM measurements. Conclusion. The TKA design with an asymmetrical tibial component has comparable component migration with the proven TKA with a symmetrical tibial component. This suggests the risk of long-term aseptic loosening of the two designs is comparable. Cite this article: Bone Joint J 2021;103-B(5):855–863


The Bone & Joint Journal
Vol. 106-B, Issue 11 | Pages 1293 - 1300
1 Nov 2024
O’Malley O Craven J Davies A Sabharwal S Reilly P

Aims. Reverse shoulder arthroplasty (RSA) has become the most common type of shoulder arthroplasty used in the UK, and a better understanding of the outcomes after revision of a failed RSA is needed. The aim of this study was to review the current evidence systematically to determine patient-reported outcome measures and the rates of re-revision and complications for patients undergoing revision of a RSA. Methods. MEDLINE, Embase, CENTRAL, and the Cochrane Database of Systematic Reviews were searched. Studies involving adult patients who underwent revision of a primary RSA for any indication were included. Those who underwent a RSA for failure of a total shoulder arthroplasty or hemiarthroplasty were excluded. Pre- and postoperative shoulder scores were evaluated in a random effects meta-analysis to determine the mean difference. The rates of re-revision and complications were also calculated. Results. The initial search elicited 3,166 results and, following removal of duplicates and screening, 13 studies with a total of 1,042 RSAs were identified. An increase in shoulder scores pre- to postoperatively was reported in all the studies. Following revision of a RSA to a further RSA, there was a significant increase in the American Shoulder and Elbow Surgeons Score (mean difference 20.78 (95% CI 8.16 to 33.40); p = 0.001). A re-revision rate at final follow-up ranging from 9% to 32%, a one-year re-revision rate of 14%, and a five-year re-revision rate of 23% were reported. The complication rate in all the studies was between 18.5% and 36%, with a total incidence of 29%. Conclusion. This is the largest systematic review of the outcomes following revision of a RSA. We found an improvement in functional outcomes after revision surgery, but the rates of re-revision and complications are high and warrant consideration when planning a revision procedure. Cite this article: Bone Joint J 2024;106-B(11):1293–1300


The Bone & Joint Journal
Vol. 105-B, Issue 11 | Pages 1168 - 1176
1 Nov 2023
Yüksel Y Koster LA Kaptein BL Nelissen RGHH den Hollander P

Aims. Conflicting clinical results are reported for the ATTUNE Total Knee Arthroplasty (TKA). This randomized controlled trial (RCT) evaluated five-year follow-up results comparing cemented ATTUNE and PFC-Sigma cruciate retaining TKAs, analyzing component migration as measured by radiostereometric analysis (RSA), clinical outcomes, patient-reported outcome measures (PROMs), and radiological outcomes. Methods. A total of 74 primary TKAs were included in this single-blind RCT. RSA examinations were performed, and PROMs and clinical outcomes were collected immediate postoperatively, and at three, six, 12, 24, and 60 months’ follow-up. Radiolucent lines (RLLs) were measured in standard anteroposterior radiographs at six weeks, and 12 and 60 months postoperatively. Results. At five-year follow-up, RSA data from 61 patients were available and the mean maximum total point motion (MTPM) of the femoral components were: ATTUNE: 0.96 mm (95% confidence interval (CI) 0.79 to 1.14) and PFC-Sigma 1.37 mm (95% CI 1.18 to 1.59) (p < 0.001). The PFC-Sigma femoral component migrated more in the first postoperative year, but stabilized thereafter. MPTM of the tibial components were comparable at five-year follow-up: ATTUNE 1.12 mm (95% CI 0.95 to 1.31) and PFC-Sigma 1.25 mm (95% CI 1.07 to 1.44) (p = 0.438). RLL at the medial tibial implant-cement interface remained more prevalent for the ATTUNE at five-year follow-up compared to the PFC-Sigma (20% vs 3%). RLL did not progress over time, and varied between patients at different timepoints for both TKA systems. Clinical outcomes and PROMs improved compared with preoperative scores, and were not different between groups. Conclusion. MTPM migration at five-year follow-up of the femoral and tibial component of the ATTUNE were similar and as low as that of the PFC-Sigma. MTPM migration of both knee implants did not significantly change from one year post-surgery, indicating stable fixation. Long-term ATTUNE performance may be expected to be comparable to the clinically well-performing PFC-Sigma. We have not found evidence of increased tibial component migration as measured by RSA to support concerns about cement debonding and a higher risk of aseptic loosening with the ATTUNE TKA. Cite this article: Bone Joint J 2023;105-B(11):1168–1176


The Bone & Joint Journal
Vol. 106-B, Issue 3 | Pages 268 - 276
1 Mar 2024
Park JH Lee JH Kim DY Kim HG Kim JS Lee SM Kim SC Yoo JC

Aims. This study aimed to assess the impact of using the metal-augmented glenoid baseplate (AGB) on improving clinical and radiological outcomes, as well as reducing complications, in patients with superior glenoid wear undergoing reverse shoulder arthroplasty (RSA). Methods. From January 2016 to June 2021, out of 235 patients who underwent primary RSA, 24 received a superior-AGB after off-axis reaming (Group A). Subsequently, we conducted propensity score matching in a 1:3 ratio, considering sex, age, follow-up duration, and glenoid wear (superior-inclination and retroversion), and selected 72 well-balanced matched patients who received a standard glenoid baseplate (STB) after eccentric reaming (Group B). Superior-inclination, retroversion, and lateral humeral offset (LHO) were measured to assess preoperative glenoid wear and postoperative correction, as well as to identify any complications. Clinical outcomes were measured at each outpatient visit before and after surgery. Results. There were no significant differences in demographic data and preoperative characteristics between the two groups. Both groups showed significant improvements in patient-reported outcome measures (visual analogue scale for pain, visual analogue scale for function, American Shoulder and Elbow Surgeons, Constant, and Simple Shoulder Test scores) from preoperative to final assessment (p < 0.001). However, AGB showed no additional benefit. Notably, within range of motion, Group B showed significant postoperative decrease in both external rotation and internal rotation, unlike Group A (p = 0.028 and 0.003, respectively). Both groups demonstrated a significant correction of superior-inclination after surgery, while patients in Group B exhibited a significant decrease in LHO postoperatively (p = 0.001). Regarding complications, Group A experienced more acromial stress fractures (3 cases; 12.5%), whereas Group B had a higher occurrence of scapular notching (24 cases; 33.3%) (p = 0.008). Conclusion. Both eccentric reaming with STB and off-axis reaming with AGB are effective methods for addressing superior glenoid wear in RSA, leading to improved clinical outcomes. However, it is important to be aware of the potential risks associated with eccentric reaming, which include excessive bone loss leading to reduced rotation and scapular notching. Cite this article: Bone Joint J 2024;106-B(3):268–276


The Bone & Joint Journal
Vol. 105-B, Issue 1 | Pages 35 - 46
1 Jan 2023
Mills K Wymenga AB Bénard MR Kaptein BL Defoort KC van Hellemondt GG Heesterbeek PJC

Aims. The aim of this study was to compare a bicruciate-retaining (BCR) total knee arthroplasty (TKA) with a posterior cruciate-retaining (CR) TKA design in terms of kinematics, measured using fluoroscopy and stability as micromotion using radiostereometric analysis (RSA). Methods. A total of 40 patients with end-stage osteoarthritis were included in this randomized controlled trial. All patients performed a step-up and lunge task in front of a monoplane fluoroscope one year postoperatively. Femorotibial contact point (CP) locations were determined at every flexion angle and compared between the groups. RSA images were taken at baseline, six weeks, three, six, 12, and 24 months postoperatively. Clinical and functional outcomes were compared postoperatively for two years. Results. The BCR-TKA demonstrated a kinematic pattern comparable to the natural knee’s screw-home mechanism in the step-up task. In the lunge task, the medial CP of the BCR-TKA was more anterior in the early flexion phase, while laterally the CP was more posterior during the entire movement cycle. The BCR-TKA group showed higher tibial migration. No differences were found for the clinical and functional outcomes. Conclusion. The BCR-TKA shows a different kinematic pattern in early flexion/late extension compared to the CR-TKA. The difference between both implants is mostly visible in the flexion phase in which the anterior cruciate ligament is effective; however, both designs fail to fully replicate the motion of a natural knee. The higher migration of the BCR-TKA was concerning and highlights the importance of longer follow-up. Cite this article: Bone Joint J 2023;105-B(1):35–46


The Bone & Joint Journal
Vol. 105-B, Issue 11 | Pages 1189 - 1195
1 Nov 2023
Kim JS Kim SH Kim SC Park JH Kim HG Lee SM Yoo JC

Aims. The aim of this study was to compare the clinical and radiological outcomes of reverse shoulder arthroplasty (RSA) using small and standard baseplates in Asian patients, and to investigate the impact of a mismatch in the sizes of the glenoid and the baseplate on the outcomes. Methods. This was retrospective analysis of 50 and 33 RSAs using a standard (33.8 mm, ST group) and a small (29.5 mm, SM group) baseplate of the Equinoxe reverse shoulder system, which were undertaken between January 2017 and March 2021. Radiological evaluations included the size of the glenoid, the β-angle, the inclination of the glenoid component, inferior overhang, scapular notching, the location of the central cage in the baseplate within the vault and the mismatch in size between the glenoid and baseplate. Clinical evaluations included the range of motion (ROM) and functional scores. In subgroup analysis, comparisons were performed between those in whom the vault of the glenoid was perforated (VP group) and those in whom it was not perforated (VNP group). Results. Perforation of the vault of the glenoid (p = 0.018) and size mismatch in height (p < 0.001) and width (p = 0.013) were significantly more frequent in the ST group than in the SM group. There was no significant difference in the clinical scores and ROM in the two groups, two years postoperatively (all p > 0.05). In subgroup analysis, the VP group had significantly less inferior overhang (p = 0.009), more scapular notching (p = 0.018), and more size mismatch in height (p < 0.001) and width (p = 0.025) than the VNP group. Conclusion. In Asian patients with a small glenoid, using a 29.5 mm small baseplate at the time of RSA was more effective in reducing size mismatch between the glenoid and the baseplate, decreasing the incidence of perforation of the glenoid vault, and achieving optimal positioning of the baseplate compared with the use of a 33.8 mm standard baseplate. However, longer follow-up is required to assess the impact of these findings on the clinical outcomes. Cite this article: Bone Joint J 2023;105-B(11):1189–1195


The Bone & Joint Journal
Vol. 106-B, Issue 9 | Pages 957 - 963
1 Sep 2024
Baek CH Kim JG Kim BT

Aims. Favourable short-term outcomes have been reported following latissimus dorsi tendon transfer for patients with an irreparable subscapularis (SSC) tendon tear. The aim of this study was to investigate the long-term outcomes of this transfer in these patients. Methods. This was a retrospective study involving 30 patients with an irreparable SSC tear and those with a SSC tear combined with a reparable supraspinatus tear, who underwent a latissimus dorsi tendon transfer. Clinical scores and active range of motion (aROM), SSC-specific physical examination and the rate of return to work were assessed. Radiological assessment included recording the acromiohumeral distance (AHD), the Hamada grade of cuff tear arthropathy and the integrity of the transferred tendon. Statistical analysis compared preoperative, short-term (two years), and final follow-up at a mean of 8.7 years (7 to 10). Results. There were significant improvements in clinical scores, in the range and strength of internal rotation and aROM compared with the preoperative values in the 26 patients (87%) who were available for long-term follow-up. These improvements were maintained between short- and long-term follow-ups. Although there was a decreased mean AHD of 7.3 mm (SD 1.5) and an increased mean Hamada grade of 1.7 (SD 0.5) at final follow-up, the rate of progression of cuff tear arthropathy remained low-grade. Comparison between the isolated SSC and combined SSC and reparable supraspinatus tear groups showed no significant differences. At final follow-up, one patient (3.8%) had undergone revision surgery to a reverse shoulder arthroplasty (RSA). No neurological complications were associated with the procedure. Conclusion. Latissimus dorsi transfer for an irreparable SSC tendon tear resulted in a significant clinical improvement, particularly in pain, range and strength of internal rotation and aROM, which were maintained over a mean of 8.7 years following surgery. Given that this was a long-term outcome study, there was a low-grade progression in the rate of cuff tear arthropathy. Thus, the long-term clinical efficacy of latissimus dorsi tendon transfer in patients with irreparable SSC was confirmed as a joint-preserving procedure for these patients, suggesting it as an effective alternative to RSA in young, active patients without degenerative changes of the glenohumeral joint. Cite this article: Bone Joint J 2024;106-B(9):957–963


The Bone & Joint Journal
Vol. 104-B, Issue 7 | Pages 875 - 883
1 Jul 2022
Mills K Wymenga AB van Hellemondt GG Heesterbeek PJC

Aims. Both the femoral and tibial component are usually cemented at revision total knee arthroplasty (rTKA), while stems can be added with either cemented or press-fit (hybrid) fixation. The aim of this study was to compare the long-term stability of rTKA with cemented and press-fitted stems, using radiostereometric analysis (RSA). Methods. This is a follow-up of a randomized controlled trial, initially involving 32 patients, of whom 19 (nine cemented, ten hybrid) were available for follow-up ten years postoperatively, when further RSA measurements were made. Micromotion of the femoral and tibial components was assessed using model-based RSA software (RSAcore). The clinical outcome was evaluated using the Knee Society Score (KSS), the Knee injury and Osteoarthritis Outcome Score (KOOS), and visual analogue scale (pain and satisfaction). Results. The median total femoral translation and rotation at ten years were 0.39 mm (interquartile range (IQR) 0.20 to 0.54) and 0.59° (IQR 0.46° to 0.73°) for the cemented group and 0.70 mm (IQR 0.15 to 0.77) and 0.78° (IQR 0.47° to 1.43°) for the hybrid group. For the tibial components this was 0.38 mm (IQR 0.33 to 0.85) and 0.98° (IQR 0.38° to 1.34°) for the cemented group and 0.42 mm (IQR 0.30 to 0.52) and 0.72° (IQR 0.62° to 0.82°) for the hybrid group. None of these values were significantly different between the two groups and there were no significant differences between the clinical scores in the two groups at this time. There was only one re-revision, in the hybrid group, for infection and not for aseptic loosening. Conclusion. These results show good long-term fixation with no difference in micromotion and clinical outcome between fully cemented and hybrid fixation in rTKA, which builds on earlier short- to mid-term results. The patients all had type I or II osseous defects, which may in part explain the good results. Cite this article: Bone Joint J 2022;104-B(7):875–883


The Journal of Bone & Joint Surgery British Volume
Vol. 82-B, Issue 8 | Pages 1201 - 1203
1 Nov 2000
Karrholm J Brandsson S Freeman MAR

We studied the knees of 11 volunteers using RSA during a step-up exercise requiring extension while weight-bearing from 50° to 0°. The findings on weight-bearing flexion with and without external rotation of the tibia based on MRI were confirmed


Aims

The aim of this study was to compare the migration of the femoral component, five years postoperatively, between patients with a highly cross-linked polyethylene (HXLPE) insert and those with a conventional polyethylene (PE) insert in an uncemented Triathlon fixed insert cruciate-retaining total knee arthroplasty (TKA). Secondary aims included clinical outcomes and patient-reported outcome measures (PROMs). We have previously reported the migration and outcome of the tibial components in these patients.

Methods

A double-blinded randomized controlled trial was conducted including 96 TKAs. The migration of the femoral component was measured with radiostereometry (RSA) at three and six months and one, two, and five years postoperatively. PROMs were collected preoperatively and at all periods of follow-up.


The Bone & Joint Journal
Vol. 97-B, Issue 1 | Pages 35 - 44
1 Jan 2015
Flivik G Kristiansson I Ryd L

We hypothesised that the removal of the subchondral bone plate (SCBP) for cemented acetabular component fixation in total hip arthroplasty (THA) offers advantages over retention by improving the cement-bone interface, without jeopardising implant stability. We have previously published two-year follow-up data of a randomised controlled trial (RCT), in which 50 patients with primary osteoarthritis were randomised to either retention or removal of the SCBP. The mean age of the retention group (n = 25, 13 males) was 70.0 years (. sd. 6.8). The mean age in the removal group (n = 25, 16 males) was 70.3 years (. sd. 7.9). Now we have followed up the patients at six (retention group, n = 21; removal group, n = 20) and ten years (retention group: n = 17, removal group: n = 18), administering clinical outcome questionnaires and radiostereometric analysis (RSA), and determining the presence of radiolucent lines (RLLs) on conventional radiographs. RSA demonstrated similar translation and rotation patterns up to six years. Between six and ten years, proximal acetabular component migration and changes of inclination were larger in the retention group, although the mean differences did not reach statistical significance. Differences in migration were driven by two patients in the SCBP retention group with extensive migration versus none in the SCBP removal group. The significant difference (p < 0.001) in the development of radiolucent lines in the retention group, previously observed at two years, increased even further during the course of follow-up (p < 0.001). While recognising SCBP removal is a more demanding technique, we conclude that, wherever possible, the SCBP should be removed to improve the cement–bone interface in order to maximise acetabular component stability and longevity. Cite this article: Bone Joint J 2015;97-B:35–44


The Bone & Joint Journal
Vol. 95-B, Issue 5 | Pages 605 - 608
1 May 2013
Murray DW Gulati A Gill HS

The Exeter femoral stem is a double-tapered highly polished collarless cemented implant with good long-term clinical results. In order to determine why the stem functions well we have undertaken a long-term radiostereometric analysis (RSA) study. A total of 20 patients undergoing primary Exeter total hip replacement for osteoarthritis using the Hardinge approach were recruited and followed with RSA for ten years. The stems progressively subsided and internally rotated with posterior head migration. The mean subsidence was 0.7 mm (95% confidence interval (CI) 0.5 to 0.9) at two years and 1.3 mm (95% CI 1.0 to 1.6) at ten years. The mean posterior migration of the head was 0.7 mm (95% CI 0.5 to 0.9) at two years and 1.2 mm (95% CI 1.0 to 1.4) at ten years. There was no significant cement restrictor migration. The Exeter stem continues to subside slowly into the cement mantle in the long term. This appears to compress the cement and the cement bone interface, contributing to secure fixation in the long term. Cite this article: Bone Joint J 2013;95-B:605–8


The Bone & Joint Journal
Vol. 104-B, Issue 1 | Pages 19 - 26
1 Jan 2022
Sevaldsen K Schnell Husby O Lian ØB Farran KM Schnell Husby V

Aims

Highly polished stems with force-closed design have shown satisfactory clinical results despite being related to relatively high early migration. It has been suggested that the minimal thickness of cement mantles surrounding the femoral stem should be 2 mm to 4 mm to avoid aseptic loosening. The line-to-line cementing technique of the femoral stem, designed to achieve stem press-fit, challenges this opinion. We compared the migration of a highly polished stem with force-closed design by standard and line-to-line cementing to investigate whether differences in early migration of the stems occur in a clinical study.

Methods

In this single-blind, randomized controlled, clinical radiostereometric analysis (RSA) study, the migration pattern of the cemented Corail hip stem was compared between line-to-line and standard cementing in 48 arthroplasties. The primary outcome measure was femoral stem migration in terms of rotation and translation around and along with the X-, Y-, and Z- axes measured using model-based RSA at three, 12, and 24 months. A linear mixed-effects model was used for statistical analysis.


The Bone & Joint Journal
Vol. 104-B, Issue 1 | Pages 76 - 82
1 Jan 2022
ten Brinke B Hesseling B Eygendaal D Hoelen MA Mathijssen NMC

Aims

Stemless humeral implants have been developed to overcome stem-related complications in total shoulder arthroplasty (TSA). However, stemless implant designs may hypothetically result in less stable initial fixation, potentially affecting long-term survival. The aim of this study is to investigate early fixation and migration patterns of the stemless humeral component of the Simpliciti Shoulder System and to evaluate clinical outcomes.

Methods

In this prospective cohort study, radiostereometric analysis (RSA) radiographs were obtained in 24 patients at one day, six weeks, six months, one year, and two years postoperatively. Migration was calculated using model-based RSA. Clinical outcomes were evaluated using the visual analogue scale (VAS), the Oxford Shoulder Score (OSS), the Constant-Murley Score (CMS), and the Disabilities of the Arm, Shoulder and Hand (DASH) score.


The Bone & Joint Journal
Vol. 104-B, Issue 1 | Pages 150 - 156
1 Jan 2022
Leino OK Lehtimäki KK Mäkelä K Äärimaa V Ekman E

Aims

Proximal humeral fractures (PHFs) are common. There is increasing evidence that most of these fractures should be treated conservatively. However, recent studies have shown an increase in use of operative treatment. The aim of this study was to identify the trends in the incidence and methods of treatment of PHFs in Finland.

Methods

The study included all Finnish inhabitants aged ≥ 16 years between 1997 and 2019. All records, including diagnostic codes for PHFs and all surgical procedure codes for these fractures, were identified from two national registers. Data exclusion criteria were implemented in order to identify only acute PHFs, and the operations performed to treat them.


The Journal of Bone & Joint Surgery British Volume
Vol. 78-B, Issue 4 | Pages 682 - 682
1 Jul 1996
LOUDON JR


The Journal of Bone & Joint Surgery British Volume
Vol. 78-B, Issue 1 | Pages 1 - 3
1 Jan 1996
Nilsson KG Kärrholm J


The Bone & Joint Journal
Vol. 104-B, Issue 1 | Pages 3 - 5
1 Jan 2022
Rajasekaran RB Ashford R Stevenson JD Pollock R Rankin KS Patton JT Gupta S Cosker TDA


The Journal of Bone & Joint Surgery British Volume
Vol. 88-B, Issue 7 | Pages 976 - 976
1 Jul 2006
DERBYSHIRE B PORTER ML


The Bone & Joint Journal
Vol. 102-B, Issue 10 | Pages 1303 - 1310
3 Oct 2020
Kjærgaard K Ding M Jensen C Bragdon C Malchau H Andreasen CM Ovesen O Hofbauer C Overgaard S

Aims

The most frequent indication for revision surgery in total hip arthroplasty (THA) is aseptic loosening. Aseptic loosening is associated with polyethylene liner wear, and wear may be reduced by using vitamin E-doped liners. The primary objective of this study was to compare proximal femoral head penetration into the liner between a) two cross-linked polyethylene (XLPE) liners (vitamin E-doped (vE-PE)) versus standard XLPE liners, and b) two modular femoral head diameters (32 mm and 36 mm).

Methods

Patients scheduled for a THA were randomized to receive a vE-PE or XLPE liner with a 32 mm or 36 mm metal head (four intervention groups in a 2 × 2 factorial design). Head penetration and acetabular component migration were measured using radiostereometric analysis at baseline, three, 12, 24, and 60 months postoperatively. The Harris Hip Score, University of California, Los Angeles (UCLA) Activity Score, EuroQol five-dimension questionnaire (EQ-5D), and 36-Item Short-Form Health Survey questionnaire (SF-36) were assessed at baseline, three, 12, 36, and 60 months.


The Bone & Joint Journal
Vol. 97-B, Issue 3 | Pages 391 - 397
1 Mar 2015
van Embden D Stollenwerck GANL Koster LA Kaptein BL Nelissen RGHH Schipper IB

The aim of this study was to quantify the stability of fracture-implant complex in fractures after fixation. A total of 15 patients with an undisplaced fracture of the femoral neck, treated with either a dynamic hip screw or three cannulated hip screws, and 16 patients with an AO31-A2 trochanteric fracture treated with a dynamic hip screw or a Gamma Nail, were included. Radiostereometric analysis was used at six weeks, four months and 12 months post-operatively to evaluate shortening and rotation.

Migration could be assessed in ten patients with a fracture of the femoral neck and seven with a trochanteric fracture. By four months post-operatively, a mean shortening of 5.4 mm (-0.04 to 16.1) had occurred in the fracture of the femoral neck group and 5.0 mm (-0.13 to 12.9) in the trochanteric fracture group. A wide range of rotation occurred in both types of fracture. Right-sided trochanteric fractures seem more rotationally stable than left-sided fractures.

This prospective study shows that migration at the fracture site occurs continuously during the first four post-operative months, after which stabilisation occurs. This information may allow the early recognition of patients at risk of failure of fixation.

Cite this article: Bone Joint J 2015;97-B:391–7.


The Bone & Joint Journal
Vol. 97-B, Issue 6 | Pages 755 - 761
1 Jun 2015
Flatøy B Röhrl SM Rydinge J Dahl J Diep LM Nordsletten L

Cemented femoral stems with force closed fixation designs have shown good clinical results despite high early subsidence. A new triple-tapered stem in this category (C-stem AMT) was introduced in 2005. This study compares this new stem with an established stem of similar design (Exeter) in terms of migration (as measured using radiostereometric analysis), peri-prosthetic bone remodelling (measured using dual energy x-ray densitometry, DXA), Oxford Hip Score, and plain radiographs.

A total of 70 patients (70 hips) with a mean age of 66 years (53 to 78) were followed for two years. Owing to missing data of miscellaneous reasons, the final analysis represents data from 51 (RSA) and 65 (DXA) patients. Both stems showed a typical pattern of migration: Subsidence and retroversion that primarily occurred during the first three months. C-stem AMT subsided less during the first three months (p = 0.01), before stabilising at a subsidence rate similar to the Exeter stem from years one to two. The rate of migration into retroversion was slightly higher for C-stem AMT during the second year (p = 0.03). Whilst there were slight differences in movement patterns between the stems, the C-stem AMT exhibits good early clinical outcomes and displays a pattern of migration and bone remodelling that predicts good clinical performance.

Cite this article: Bone Joint J 2015;97-B:755–61.


The Bone & Joint Journal
Vol. 96-B, Issue 2 | Pages 229 - 236
1 Feb 2014
deVos MJ Verdonschot N Luites JWH Anderson PG Eygendaal D

We determined the short-term clinical outcome and migration within the bone of the humeral cementless component of the Instrumented Bone Preserving (IBP) total elbow replacement in a series of 16 patients. There were four men and 12 women with a mean age at operation of 63 years (40 to 81). Migration was calculated using radiostereometric analysis. There were no intra-operative complications and no revisions. At two-year follow-up, all patients showed a significant reduction in pain and functional improvement of the elbow (both p < 0.001). Although ten components (63%) showed movement or micromovement during the first six weeks, 14 (88%) were stable at one year post-operatively. Translation was primarily found in the proximal direction (median 0.3 mm (interquartile range (IQR) -0.09 to 0.8); the major rotational movement was an anterior tilt (median 0.7° (IQR 0.4° to 1.6°)). One malaligned component continued to migrate during the second year, and one component could not be followed beyond three months because migration had caused the markers to break off the prosthesis.

This study shows promising early results for the cementless humeral component of the IBP total elbow replacement. All patients had a good clinical outcome, and most components stabilised within six months of the operation.

Cite this article: Bone Joint J 2014;96-B:229–36.


The Bone & Joint Journal
Vol. 96-B, Issue 5 | Pages 604 - 608
1 May 2014
von Schewelov T Carlsson Å Sanzén L Besjakov J

In 2005, we demonstrated that the polished triple-tapered C-stem at two years had migrated distally and rotated internally. From that series, 33 patients have now been followed radiologically, clinically and by radiostereometric analysis (RSA) for up to ten years. The distal migration within the cement mantle had continued and reached a mean of 2 mm (0.5 to 4.0) at ten years. Internal rotation, also within the cement mantle, was a mean 3.8° (external 1.6° to internal 6.6°) The cement mantle did not show any sign of migration or loosening in relation to the femoral bone. There were no clinical or radiological signs indicating that the migration or rotation within the cement mantle had had any adverse effects for the patients.

Cite this article: Bone Joint J 2014;96-B:604–8.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 10 | Pages 1344 - 1350
1 Oct 2012
Penny JO Ding M Varmarken JE Ovesen O Overgaard S

Radiostereometric analysis (RSA) can detect early micromovement in unstable implant designs which are likely subsequently to have a high failure rate. In 2010, the Articular Surface Replacement (ASR) was withdrawn because of a high failure rate. In 19 ASR femoral components, the mean micromovement over the first two years after implantation was 0.107 mm (sd 0.513) laterally, 0.055 mm (sd 0.204) distally and 0.150 mm (sd 0.413) anteriorly. The mean backward tilt around the x-axis was -0.08° (sd 1.088), mean internal rotation was 0.165° (sd 0.924) and mean varus tilt 0.238° (sd 0.420). The baseline to two-year varus tilt was statistically significant from zero movement, but there was no significant movement from one year onwards.

We conclude that the ASR femoral component achieves initial stability and that early migration is not the mode of failure for this resurfacing arthroplasty.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 10 | Pages 1366 - 1371
1 Oct 2012
Pijls BG Valstar ER Kaptein BL Nelissen RGHH

This single-blinded randomised controlled trial investigated whether one design of mobile-bearing (MB) total knee replacement (TKR) has any advantage over a fixed-bearing (FB) design on long-term fixation as measured by radiostereometry. The amount of wear underneath the mobile bearing was also evaluated. A series of 42 knees was randomised to MB or FB tibial components with appropriate polyethylene inserts and followed for between ten and 12 years, or until the death of the patient. The polyethylene in the MB group was superior in that it was gamma-irradiated in inert gas and was calcium-stearate free; the polyethylene in the FB group was gamma-irradiated in air and contained calcium stearate. In theory this should be advantageous to the wear rate of the MB group. At final follow-up the overall mean migration was 0.75 mm (sd 0.76) in the MB group and 0.66 mm (sd 0.4) in the FB group, with the FB group demonstrating more posterior tilt and the MB group more internal rotation. In the FB group there was one revision for aseptic loosening, but none in the MB group. There were no significant differences in clinical or radiological scores.

For the MB group, the mean linear wear rate on the under-surface was 0.026 mm/year (sd 0.014). This was significantly smaller than the wear rate of 0.11 mm/year (sd 0.06) in the MB between femur and polyethylene (p < 0.001). Nevertheless, even in a best-case setting the mobile bearings of this TKR design had no apparent advantage in terms of fixation over the FB knee prosthesis at ten to 12 years. The wear underneath the mobile bearing was small and is unlikely to be clinically relevant.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 3 | Pages 302 - 307
1 Mar 2012
Kadar T Furnes O Aamodt A Indrekvam K Havelin LI Haugan K Espehaug B Hallan G

In this prospective study we studied the effect of the inclination angle of the acetabular component on polyethylene wear and component migration in cemented acetabular sockets using radiostereometric analysis.

A total of 120 patients received either a cemented Reflection All-Poly ultra-high-molecular-weight polyethylene or a cemented Reflection All-Poly highly cross-linked polyethylene acetabular component, combined with either cobalt–chrome or Oxinium femoral heads. Femoral head penetration and migration of the acetabular component were assessed with repeated radiostereometric analysis for two years. The inclination angle was measured on a standard post-operative anteroposterior pelvic radiograph. Linear regression analysis was used to determine the relationship between the inclination angle and femoral head penetration and migration of the acetabular component.

We found no relationship between the inclination angle and penetration of the femoral head at two years’ follow-up (p = 0.9). Similarly, our data failed to reveal any statistically significant correlation between inclination angle and migration of these cemented acetabular components (p = 0.07 to p = 0.9).


The Bone & Joint Journal
Vol. 95-B, Issue 1 | Pages 23 - 30
1 Jan 2013
Kiernan S Hermann KL Wagner P Ryd L Flivik G

Progressive retroversion of a cemented stem is predictive of early loosening and failure. We assessed the relationship between direct post-operative stem anteversion, measured with CT, and the resulting rotational stability, measured with repeated radiostereometric analysis over ten years. The study comprised 60 cemented total hip replacements using one of two types of matt collared stem with a rounded cross-section. The patients were divided into three groups depending on their measured post-operative anteversion (< 10°, 10° to 25°, >  25°). There was a strong correlation between direct post-operative anteversion and later posterior rotation. At one year the < 10° group showed significantly more progressive retroversion together with distal migration, and this persisted to the ten-year follow-up. In the < 10° group four of ten stems (40%) had been revised at ten years, and an additional two stems (20%) were radiologically loose. In the ‘normal’ (10° to 25°) anteversion group there was one revised (3%) and one loose stem (3%) of a total of 30 stems, and in the > 25° group one stem (5%) was revised and another loose (5%) out of 20 stems. This poor outcome is partly dependent on the design of this prosthesis, but the results strongly suggest that the initial rotational position of cemented stems during surgery affects the subsequent progressive retroversion, subsidence and eventual loosening. The degree of retroversion may be sensitive to prosthetic design and stem size, but < 10° of anteversion appears deleterious to the long-term outcome for cemented hip prosthetic stems.

Cite this article: Bone Joint J 2013;95-B:23–30.


The Bone & Joint Journal
Vol. 105-B, Issue 8 | Pages 912 - 919
1 Aug 2023
Cunningham LJ Walton M Bale S Trail IA

Aims. Reverse total shoulder arthroplasty (rTSA) can be used in complex cases when the glenoid requires reconstruction. In this study, a baseplate with composite bone autograft and a central trabecular titanium peg was implanted, and its migration was assessed for two years postoperatively using radiostereometric analysis (RSA). Methods. A total of 14 patients who underwent a rTSA with an autograft consented to participate. Of these, 11 had a primary rTSA using humeral head autograft and three had a revision rTSA with autograft harvested from the iliac crest. The mean age of the patients was 66 years (39 to 81). Tantalum beads were implanted in the scapula around the glenoid. RSA imaging (stereographic radiographs) was undertaken immediately postoperatively and at three, six, 12, and 24 months. Analysis was completed using model-based RSA software. Outcomes were collected preoperatively and at two years postoperatively, including the Oxford Shoulder Score, the American Shoulder and Elbow Score, and a visual analogue score for pain. A Constant score was also obtained for the assessment of strength and range of motion. Results. RSA analysis showed a small increase in all translation and rotational values up to six months postoperatively, consistent with settling of the implant. The mean values plateaued by 12 months, with no evidence of further migration. In four patients, there was significant variation outside the mean, which corresponded to postoperative complications. There was a significant improvement in the clinical and patient-reported outcomes from the preoperative values to those at two years postoperatively (p < 0.001). Conclusion. These findings show, using RSA, that a glenoid baseplate composite of a trabecular titanium peg with autograft stabilizes within the glenoid about 12 months after surgery, and reinforce findings from a previous study of this implant/graft with CT scans at two years postoperatively, indicating that this type of structural composite results in sound early fixation. Cite this article: Bone Joint J 2023;105-B(8):912–919


The Bone & Joint Journal
Vol. 95-B, Issue 11_Supple_A | Pages 51 - 52
1 Nov 2013
Murray DW

There is a North Atlantic divide, with cementless femoral stems being used more frequently in the USA and cemented stems being used more frequently in many countries in Europe. This is primarily because different cemented stems have been used on different sides of the Atlantic and the results of the cemented stems in the US have often been poor, whereas the results of the stems used in Europe have been good. In the National registers in Europe, cemented stems have tended to achieve better results than cementless.

Cite this article: Bone Joint J 2013;95-B, Supple A:51–2.


The Bone & Joint Journal
Vol. 106-B, Issue 10 | Pages 1133 - 1140
1 Oct 2024
Olsen Kipp J Petersen ET Falstie-Jensen T Frost Teilmann J Zejden A Jellesen Åberg R de Raedt S Thillemann TM Stilling M

Aims. This study aimed to quantify the shoulder kinematics during an apprehension-relocation test in patients with anterior shoulder instability (ASI) and glenoid bone loss using the radiostereometric analysis (RSA) method. Kinematics were compared with the patient’s contralateral healthy shoulder. Methods. A total of 20 patients with ASI and > 10% glenoid bone loss and a healthy contralateral shoulder were included. RSA imaging of the patient’s shoulders was performed during a repeated apprehension-relocation test. Bone volume models were generated from CT scans, marked with anatomical coordinate systems, and aligned with the digitally reconstructed bone projections on the RSA images. The glenohumeral joint (GHJ) kinematics were evaluated in the anteroposterior and superoinferior direction of: the humeral head centre location relative to the glenoid centre; and the humeral head contact point location on the glenoid. Results. During the apprehension test, the centre of the humeral head was 1.0 mm (95% CI 0.0 to 2.0) more inferior on the glenoid for the ASI shoulder compared with the healthy shoulder. Furthermore, the contact point of the ASI shoulder was 1.4 mm (95% CI 0.3 to 2.5) more anterior and 2.0 mm (95% CI 0.8 to 3.1) more inferior on the glenoid compared with the healthy shoulder. The contact point of the ASI shoulder was 1.2 mm (95% CI 0.2 to 2.6) more anterior during the apprehension test compared to the relocation test. Conclusion. The humeral head centre was located more inferior, and the GHJ contact point was located both more anterior and inferior during the apprehension test for the ASI shoulders than the healthy shoulders. Furthermore, the contact point displacement between the apprehension and relocation test revealed increased joint laxity for the ASI shoulder than the healthy shoulders. These results contribute to existing knowledge that ASI shoulders with glenoid bone loss may also suffer from inferior shoulder instability. Cite this article: Bone Joint J 2024;106-B(10):1133–1140


The Bone & Joint Journal
Vol. 105-B, Issue 2 | Pages 148 - 157
1 Feb 2023
Koster LA Rassir R Kaptein BL Sierevelt IN Schager M Nelissen RGHH Nolte PA

Aims. The primary aim of this study was to compare the migration of the femoral and tibial components of the cementless rotating platform Attune and Low Contact Stress (LCS) total knee arthroplasty (TKA) designs, two years postoperatively, using radiostereometric analysis (RSA) in order to assess the risk of the development of aseptic loosening. A secondary aim was to compare clinical and patient-reported outcome measures (PROMs) between the designs. Methods. A total of 61 TKAs were analyzed in this randomized clinical RSA trial. RSA examinations were performed one day and three, six, 12, and 24 months postoperatively. The maximal total point motion (MPTM), translations, and rotations of the components were analyzed. PROMs and clinical data were collected preoperatively and at six weeks and three, six, 12, and 24 months postoperatively. Linear mixed effect modelling was used for statistical analyses. Results. The mean MTPM two years postoperatively (95% confidence interval (CI)) of the Attune femoral component (0.92 mm (0.75 to 1.11)) differed significantly from that of the LCS TKA (1.72 mm (1.47 to 2.00), p < 0.001). The Attune femoral component subsided, tilted (anteroposteriorly), and rotated (internal-external) significantly less. The mean tibial MTPM two years postoperatively did not differ significantly, being 1.11 mm (0.94 to 1.30) and 1.17 mm (0.99 to 1.36, p = 0.447) for the Attune and LCS components, respectively. The rate of migration in the second postoperative year was negligible for the femoral and tibial components of both designs. The mean pain-at-rest (numerical rating scale (NRS)-rest) in the Attune group was significantly less compared with that in the LCS group during the entire follow-up period. At three months postoperatively, the Knee injury and Osteoarthritis Outcome Physical Function Shortform score, the Oxford Knee Score, and the NRS-activity scores were significantly better in the Attune group. Conclusion. The mean MTPM of the femoral components of the cementless rotating platform Attune was significantly less compared with that of the LCS design. This was reflected mainly in significantly less subsidence, posterior tilting, and internal rotation. The mean tibial MTPMs were not significantly different. During the second postoperative year, the components of both designs stabilized and low risks for the development of aseptic loosening are expected. Cite this article: Bone Joint J 2023;105-B(2):148–157


The Bone & Joint Journal
Vol. 101-B, Issue 6 | Pages 627 - 634
1 Jun 2019
King JJ Dalton SS Gulotta LV Wright TW Schoch BS

Aims. Acromial fractures following reverse shoulder arthroplasty (RSA) have a wide range of incidences in reported case series. This study evaluates their incidence following RSA by systematically reviewing the current literature. Materials and Methods. A systematic review using the search terms “reverse shoulder”, “reverse total shoulder”, or “inverted shoulder” was performed using PubMed, Web of Science, and Cochrane databases between 1 January 2010 and 31 March 2018. Preferred Reporting Items for Systematic Reviews and Meta-Analyses (PRISMA) guidelines were used. Studies were included if they reported on RSA outcomes and the incidence rate of acromial and/or scapular spine fractures. The rate of these fractures was evaluated for primary RSA, revision RSA, RSA indications, and RSA implant design. Results. The review included 90 articles out of 686 identified after exclusions. The incidence rate of acromial and/or scapular spine fractures was 2.8% (253/9048 RSAs). The fracture rate was similar for primary and revision RSA (2.8% vs 2.1%; p = 0.4). Acromial fractures were most common after RSA for inflammatory arthritis (10.9%) and massive rotator cuff tears (3.8%). The incidence was lowest in RSA for post-traumatic arthritis (2.1%) and acute proximal humerus fractures (0%). Lateralized glenosphere design had a significantly higher rate of acromial fractures compared with medial glenosphere designs. Conclusion. Based on current English literature, acromial and/or scapular spine fractures occur at a rate of 2.8% after RSA. The incidence is slightly more common after primary compared with revision arthroplasty. Also, higher rates of acromial fractures are reported in RSA performed for inflammatory arthritis and in the lateralized glenoid design. Cite this article: Bone Joint J 2019;101-B:627–634


The Bone & Joint Journal
Vol. 103-B, Issue 2 | Pages 360 - 365
1 Feb 2021
Mahendraraj KA Shields MV Grubhofer F Golenbock SW Jawa A

Aims. Existing literature indicates that inferiorly inclined glenoid baseplates following reverse total shoulder arthroplasty (RSA) produce better outcomes compared to superiorly inclined baseplates. We aim to compare clinical outcomes for RSAs with superiorly and neutrally/inferiorly inclined lateralized glenospheres. Methods. We retrospectively reviewed 154 consecutive patients undergoing RSA between July 2015 and July 2017 by one single-fellowship trained surgeon (AJ). Two raters (KAM and MVS) independently measured glenoid inclination in preoperative and minimum two year follow-up radiographs (anteroposterior/Grashey) using the RSA angle. Inclination was then compared to patient-reported outcomes, range of motion (ROM), and independently assessed degree of scapular notching and staging of heterotopic ossification at two year follow-up. Results. Median postoperative inclination for each group was found to be -3.6° (interquartile range (IQR) -2.1 to -6.9) and 6.0° (3.2° to 10.1°) for the neutrally/inferiorly and superiorly inclined cohorts, respectively. Preoperative inclination was highly associated with postoperative inclination (p = 0.004). When comparing superiorly and neutrally/inferiorly inclined glenospheres, there were no differences in heterotopic ossification (p = 0.606), scapular notching (p = 0.367), American Shoulder and Elbow Surgeons score (p = 0.419), Single Assessment Numeric Evaluation (p = 0.417), Visual Analogue Scale (VAS) pain score (p = 0.290), forward elevation (p = 0.161), external rotation (p = 0.537), or internal rotation (p = 0.656). Conclusion. Compared to neutral and inferior inclination, up to 6° ± 3° of superior glenoid baseplate inclination on a lateralized RSA design produces no differences in postoperative ROM or patient-reported outcomes, and produces similar levels of scapular notching and heterotopic ossification. Additionally, the degree of preoperative inclination represents an important factor in surgical decision-making as it is strongly associated with postoperative inclination. It is important to note that the findings of this study are only reflective of lateralized RSA prostheses. Cite this article: Bone Joint J 2021;103-B(2):360–365


The Bone & Joint Journal
Vol. 101-B, Issue 12 | Pages 1520 - 1525
1 Dec 2019
Clark NJ Samuelsen BT Alentorn-Geli E Assenmacher AT Cofield RH Sperling JW Sánchez-Sotelo J

Aims. Reverse shoulder arthroplasty (RSA) reliably improves shoulder pain and function for a variety of indications. However, the safety and efficacy of RSA in elderly patients is largely unknown. The purpose of this study was to report the mortality, morbidity, complications, reoperations, and outcomes of primary RSA in patients aged > 80 years. Patients and Methods. Between 2004 and 2013, 242 consecutive primary RSAs were performed in patients aged > 80 years (mean 83.3 years (. sd. 3.1)). Of these, 53 were lost to follow-up before two years and ten had died within two years of surgery, leaving 179 for analysis of survivorship, pain, motion, and strength at a minimum of two years or until revision surgery. All 242 patients were considered for the analysis of 90-day, one-year, and overall mortality, medical complications (90-day and overall), surgical complications, and reoperations. The indications for surgery included rotator cuff arthropathy, osteoarthritis, fracture, the sequela of trauma, avascular necrosis, and rheumatoid arthritis. A retrospective review of the medical records was performed to collect all variables. Survivorship free of revision surgery was calculated at two and five years. Results. One patient (0.4%) died within the first 90 days. A total of 45 patients (19%) were known to have died at the time of the final follow-up, with a median time to death of 67.7 months (interquartile range 40.4 to 94.7) postoperatively. Medical complications occurred in six patients (3%) and surgical complications occurred in 21/179 patients (12%). Survivorship free from revision was 98.9% at two years and 98.3% at five years; survivorship free from loosening was 99.5% at final follow-up. The presence of peripheral vascular disease correlated with a higher complication rate. Conclusion. Primary RSA was safe and effective in patients aged > 80 years, with a relatively low rate of medical and surgical complications. Thus, age alone should not be a contraindication to primary RSA in patients aged > 80 years. However, a careful evaluation of comorbidities is required in this age group when considering primary RSA. Cite this article: Bone Joint J 2019;101-B:1520–1525


The Bone & Joint Journal
Vol. 102-B, Issue 11 | Pages 1555 - 1559
1 Nov 2020
Sebastia-Forcada E Lizaur-Utrilla A Mahiques-Segura G Ruiz-Lozano M Lopez-Prats FA Alonso-Montero C

Aims. The purpose of this study was to determine whether there were long-term differences in outcomes of reverse shoulder arthroplasty (RSA) undertaken for acute proximal humeral fracture versus rotator cuff deficiency with a minimum follow-up of five years. Methods. This was a prospective cohort study comparing 67 patients with acute complex proximal humeral fracture and 64 patients with irreparable rotator cuff deficiency who underwent primary RSA. In the fracture group, there were 52 (77.6%) females and 15 (22.4%) males, with a mean age of 73.5 years (51 to 85), while in the arthropathy group, there were 43 (67.1%) females and 21 (32.9%) males, with a mean age of 70.6 years (50 to 84). Patients were assessed by the Constant score, University of California Los Angeles shoulder score (UCLA), short version of the Disability of the Arm Shoulder and Hand score (QuickDASH), and visual analogue scales (VAS) for pain and satisfaction. Radiological evaluation was also performed. Results. Mean follow-up was 8.4 years (5 to 11). There were no significant differences in mean absolute (p = 0.125) or adjusted (p = 0.569) Constant, UCLA (p = 0.088), QuickDASH (p = 0.135), VAS-pain (p = 0.062), or range of movement at the final follow-up. However, patient satisfaction was significantly lower in the fracture group (p = 0.002). The complication rate was 1.5% (one patient) versus 9.3% (six patients), and the revision rate was 1.5% (one patient) versus 7.8% (five patients) in the fracture and arthropathy groups, respectively. The ten-year arthroplasty survival was not significantly different (p = 0.221). Conclusion. RSA may be used not only for patients with irreparable rotator cuff deficiencies, but also for those with acute complex proximal humeral fractures. We found that RSA provided similar functional outcomes and a low revision rate for both indications at long-term. However, satisfaction is lower in patients with an acute fracture. Cite this article: Bone Joint J 2020;102-B(11):1555–1559


The Bone & Joint Journal
Vol. 101-B, Issue 1 | Pages 63 - 67
1 Jan 2019
Erickson BJ Ling D Wong A Eno JJ Dines JS Dines DM Gulotta LV

Aims. The number of rotator cuff repairs that are undertaken is increasing. Reverse shoulder arthroplasty (RSA) is the procedure of choice for patients with rotator cuff arthropathy. We sought to determine whether patients who underwent rotator cuff repair and subsequent RSA had different outcomes compared with a matched control group who underwent RSA without a previous rotator cuff repair. Patients and Methods. All patients with a history of rotator cuff repair who underwent RSA between 2000 and 2015 with a minimum follow-up of two years were eligible for inclusion as the study group. Outcomes, including the American Shoulder and Elbow Surgeons (ASES) scores, were compared with a matched control group of patients who underwent RSA without having previously undergone rotator cuff repair. Results. The study group included 45 patients. Their mean age was 69 years (. sd. 8.6) and 27 patients (60%) were women. The mean ASES score improved from 43.1 to 76.6 two years postoperatively, and to 66.9 five years postoperatively. There was no significant difference between the outcomes at two years in the two groups (all p ≥ 0.05), although there was significantly more improvement in ASES scores in the control group (44.5 vs 33.4; p = 0.01). However, there was no significant difference between ASES scores at two and five years when baseline ASES scores were matched in the two groups (p = 0.42 at two years; p = 0.35 at five years). Conclusion. Significant improvements in ASES scores were seen following RSA in patients who had previously undergone rotator cuff repair. They had higher baseline ASES scores than those who had not previously undergone this surgery. However, there was no significant difference in outcomes between the two groups, two years postoperatively. Previous rotator cuff repair does not appear to affect the early outcome after RSA adversely


The Bone & Joint Journal
Vol. 100-B, Issue 11 | Pages 1493 - 1498
1 Nov 2018
Wagner ER Hevesi M Houdek MT Cofield RH Sperling JW Sanchez-Sotelo J

Aims. Patients with a failed reverse shoulder arthroplasty (RSA) have limited salvage options. The aim of this study was to determine the outcome of revision RSA when used as a salvage procedure for a failed primary RSA. Patients and Methods. We reviewed all revision RSAs performed for a failed primary RSA between 2006 and 2012, excluding patients with a follow-up of less than two years. A total of 27 revision RSAs were included in the study. The mean age of the patients at the time of revision was 70 years (58 to 82). Of the 27 patients, 14 (52% were female). The mean follow-up was 4.4 years (2 to 10). Results. Six patients (22%) developed complications requiring further revision surgery, at a mean of 1.7 years (0.1 to 5.3) postoperatively. The indication for further revision was dislocation in two, glenoid loosening in one, fracture of the humeral component in one, disassociation of the glenosphere in one, and infection in one. The five-year survival free of further revision was 85%. Five additional RSAs developed complications that did not need surgery, including dislocation in three and periprosthetic fracture in two. Overall, patients who did not require further revision had excellent pain relief, and significant improvements in elevation and external rotation of the shoulder (p < 0.01). The mean postoperative American Shoulder and Elbow Surgeons (ASES), and simple shoulder test (SST) scores were 66 and 7, respectively. Radiological results were available in 26 patients (96.3%) at a mean of 4.3 years (1.5 to 9.5). At the most recent follow-up, six patients (23%) had glenoid lucency, which were classified as grade III or higher in three (12%). Smokers had a significantly increased risk of glenoid lucency (p < 0.01). Conclusion. Revision RSA, when used to salvage a failed primary RSA, can be a successful procedure. At intermediate follow-up, survival rates are reasonable, but dislocation and glenoid lucency remain a concern, particularly in smokers. Cite this article: Bone Joint J 2018;100-B:1493–98


The Bone & Joint Journal
Vol. 101-B, Issue 11 | Pages 1348 - 1355
1 Nov 2019
Gascoyne T Parashin S Teeter M Bohm E Laende E Dunbar M Turgeon T

Aims. A retrospective study was conducted to measure short-term in vivo linear and volumetric wear of polyethylene (PE) inserts in 101 total knee arthroplasty (TKA) patients using model-based radiostereometric analysis (MBRSA). Patients and Methods. Nonweightbearing supine RSA exams were performed postoperatively and at six, 12, and 24 months. Weightbearing standing RSA exams were performed on select patients at 12 and 24 months. Wear was measured both linearly (joint space) and volumetrically (digital model overlap) at each available follow-up. Precision of both methods was assessed by comparing double RSA exams. Patient age, sex, body mass index, and Oxford Knee Scores were analyzed for any association with PE wear. Results. Linear wear occurred at 0.015 mm/year (supine) and 0.220 mm/year (standing). Volumetric wear occurred at 10.3 mm. 3. /year (supine) and 39.3 mm. 3. /year (standing). Wear occurred primarily on the medial side of the joint. Weightbearing imaging greatly improved the reliability of measurement. Clinical precision of volumetric wear was 34 mm. 3. No significant associations were found between patient demographics or function scores and measured wear. Conclusion. In vivo volumetric wear of TKAs can be assessed at short-term follow-up using MBRSA. Cite this article: Bone Joint J 2019;101-B:1348–1355


Aims. The aim of this study was to compare the mid-term patient-reported outcome, bone remodelling, and migration of a short stem (Collum Femoris Preserving; CFP) with a conventional uncemented stem (Corail). Methods. Of 81 patients who were initially enrolled, 71 were available at five years’ follow-up. The outcomes at two years have previously been reported. The primary outcome measure was the clinical result assessed using the Oxford Hip Score (OHS). Secondary outcomes were the migration of the stem, measured using radiostereometric analysis (RSA), change of bone mineral density (BMD) around the stem, the development of radiolucent lines, and additional patient-reported outcome measures (PROMs). Results. There were no statistically significant differences between the groups regarding PROMs (median OHS (CFP 45 (interquartile range (IQR) 35 to 48); Corail 45 (IQR 40 to 48); p = 0.568). RSA showed stable stems in both groups, with little or no further subsidence between two and five years. Resorption of the femoral neck was evident in nine patients in the CFP group and in none of the 15 Corail stems with a collar that could be studied. Dual X-ray absorbiometry showed a significantly higher loss of BMD in the proximal Gruen zones in the CFP group (mean changes in BMD: Gruen zone 1, CFP -9.5 (95% confidence interval (CI) -14.8 to -4.2), Corail 1.0 (95% CI 3.4 to 5.4); Gruen zone 7, CFP -23.0 (95% CI -29.4 to -16.6), Corail -7.2 (95% CI -15.9 to 1.4). Two CFP stems were revised before two years’ follow-up due to loosening, and one Corail stem was revised after two years due to chronic infection. Conclusion. The CFP stem has a similar clinical outcome and subsidence pattern when compared with the Corail stem. More pronounced proximal stress-shielding was seen with the CFP stem, suggesting diaphyseal fixation, and questioning its femoral neck-sparing properties in the long term. Cite this article: Bone Joint J 2022;104-B(5):581–588


The Bone & Joint Journal
Vol. 101-B, Issue 10 | Pages 1192 - 1198
1 Oct 2019
Sköldenberg OG Rysinska AD Chammout G Salemyr M Mukka SS Bodén H Eisler T

Aims. Radiostereometric analysis (RSA) studies of vitamin E-doped, highly crosslinked polyethylene (VEPE) liners show low head penetration rates in cementless acetabular components. There is, however, currently no data on cemented VEPE acetabular components in total hip arthroplasty (THA). The aim of this study was to evaluate the safety of a new cemented VEPE component, compared with a conventional polyethylene (PE) component regarding migration, head penetration, and clinical results. Patients and Methods. We enrolled 42 patients (21 male, 21 female) with osteoarthritis and a mean age of 67 years (. sd. 5), in a double-blinded, noninferiority, randomized controlled trial. The subjects were randomized in a 1:1 ratio to receive a reverse hybrid THA with a cemented component of either argon-gas gamma-sterilized PE component (controls) or VEPE, with identical geometry. The primary endpoint was proximal implant migration of the component at two years postoperatively measured with RSA. Secondary endpoints included total migration of the component, penetration of the femoral head into the component, and patient-reported outcome measurements. Results. In total, 19 control implants and 18 implants in the VEPE group were analyzed for the primary endpoint. We found a continuous proximal migration of the component in the VEPE group that was significantly higher with a difference at two years of a mean 0.21 mm (95% confidence interval (CI) 0.05 to 0.37; p = 0.013). The total migration was also significantly higher in the VEPE group, but femoral head penetration was lower. We found no difference in clinical outcomes between the groups. Conclusion. At two years, this cemented VEPE component, although having a low head penetration and excellent clinical results, failed to meet noninferiority compared with the conventional implant by a proximal migration above the proposed safety threshold of RSA. The early proximal migration pattern of the VEPE component is a reason for continued monitoring, although a specific threshold for proximal migration and risk for later failure cannot be defined and needs further study. Cite this article: Bone Joint J 2019;101-B:1192–1198


The Bone & Joint Journal
Vol. 101-B, Issue 4 | Pages 461 - 469
1 Apr 2019
Lädermann A Schwitzguebel AJ Edwards TB Godeneche A Favard L Walch G Sirveaux F Boileau P Gerber C

Aims. The aim of this study was to report the outcomes of different treatment options for glenoid loosening following reverse shoulder arthroplasty (RSA) at a minimum follow-up of two years. Patients and Methods. We retrospectively studied the records of 79 patients (19 men, 60 women; 84 shoulders) aged 70.4 years (21 to 87) treated for aseptic loosening of the glenosphere following RSA. Clinical evaluation included pre- and post-treatment active anterior elevation (AAE), external rotation, and Constant score. Results. From the original cohort, 29 shoulders (35%) were treated conservatively, 27 shoulders (32%) were revised by revision of the glenosphere, and 28 shoulders (33%) were converted to hemiarthroplasty. At last follow-up, conservative treatment and glenoid revision significantly improved AAE, total Constant score, and pain, while hemiarthroplasty did not improve range of movement or clinical scores. Multivariable analysis confirmed that conservative treatment and glenoid revision achieved similar improvements in pain (glenoid revision vs conservative, beta 0.44; p = 0.834) but that outcomes were significantly worse following hemiarthroplasty (beta -5.00; p = 0.029). Conclusion. When possible, glenoid loosening after RSA should first be treated conservatively, then by glenosphere revision if necessary, and last by salvage hemiarthroplasty. Cite this article: Bone Joint J 2019;101-B:461–469


The Bone & Joint Journal
Vol. 101-B, Issue 5 | Pages 610 - 614
1 May 2019
Aibinder WR Bartels DW Sperling JW Sanchez-Sotelo J

Aims. Shoulder arthroplasty using short humeral components is becoming increasingly popular. Some such components have been associated with relatively high rates of adverse radiological findings. The aim of this retrospective review was to evaluate the radiological humeral bone changes and mechanical failure rates with implantation of a short cementless humeral component in anatomical (TSA) and reverse shoulder arthroplasty (RSA). Patients and Methods. A total of 100 shoulder arthroplasties (35 TSA and 65 RSA) were evaluated at a mean of 3.8 years (3 to 8.3). The mean age at the time of surgery was 68 years (31 to 90). The mean body mass index was 32.7 kg/m. 2. (17.3 to 66.4). Results. Greater tuberosity stress shielding was noted in 14 shoulders (two TSA and 12 RSA) and was graded as mild in nine, moderate in two, and severe in three. Medial calcar resorption was noted in 23 shoulders (seven TSA and 16 RSA), and was graded as mild in 21 and moderate in two. No humeral components were revised for loosening or considered to be loose radiologically. Nine shoulders underwent reoperation for infection (n = 3), fracture of the humeral tray (n = 2), aseptic glenoid loosening (n = 1), and instability (n = 3). No periprosthetic fractures occurred. Conclusion. Implantation of this particular short cementless humeral component at the time of TSA or RSA was associated with a low rate of adverse radiological findings on the humeral side at mid-term follow-up. Our data do not raise any concerns regarding the use of a short stem in TSA or RSA. Cite this article: Bone Joint J 2019;101-B:610–614


The Bone & Joint Journal
Vol. 101-B, Issue 7_Supple_C | Pages 55 - 60
1 Jul 2019
Laende EK Richardson CG Dunbar MJ

Aims. Early implant migration measured with radiostereometric analysis (RSA) has been proposed as a useful predictor of long-term fixation of tibial components in total knee arthroplasty. Evaluation of actual long-term fixation is of interest for cemented components, as well as for cementless fixation, which may offer long-term advantages once osseointegration has occurred. The objective of this study was to compare the long-term migration with one- and two-year migration to evaluate the predictive ability of short-term migration data and to compare migration and inducible displacement between cemented and cementless (porous metal monoblock) components at least ten years postoperatively. Patients and Methods. Patients who had participated in RSA migration studies with two-year follow-up were recruited to return for a long-term follow-up, at least ten years from surgery. Two cemented tibial designs from two manufacturers and one porous metal monoblock cementless tibial design were studied. At the long-term follow-up, patients had supine RSA examinations to determine migration and loaded examinations (single leg stance) to determine inducible displacement. In total, 79 patients (54 female) returned, with mean time since surgery of 12 years (10 to 14). There were 58 cemented and 21 cementless tibial components. Results. Migration at one year and two years was significantly correlated with long-term migration (p < 0.001). Median migration at the long-term follow-up was 0.6 mm (maximum total point motion; interquartile range (IQR) 0.4 to 0.9) for the cemented group and 0.6 mm. (IQR 0.3 to 1.1) for the cementless group with no difference between groups (p = 0.99). Inducible displacement was significantly lower for the cementless components (p < 0.001). Conclusion. Long-term migration was strongly correlated with two-year migration. Although long-term migration was not different for cemented or cementless tibial components, inducible displacement at the long-term visit was significantly lower for these cementless components, suggesting superior fixation. These findings support the predictive value of short-term migration in determining long-term fixation. Cite this article: Bone Joint J 2019;101-B(7 Supple C):55–60


The Bone & Joint Journal
Vol. 103-B, Issue 7 | Pages 1206 - 1214
1 Jul 2021
Tsikandylakis G Mortensen KRL Gromov K Mohaddes M Malchau H Troelsen A

Aims. We aimed to investigate if the use of the largest possible cobalt-chromium head articulating with polyethylene acetabular inserts would increase the in vivo wear rate in total hip arthroplasty. Methods. In a single-blinded randomized controlled trial, 96 patients (43 females), at a median age of 63 years (interquartile range (IQR) 57 to 69), were allocated to receive either the largest possible modular femoral head (36 mm to 44 mm) in the thinnest possible insert or a standard 32 mm head. All patients received a vitamin E-doped cross-linked polyethylene insert and a cobalt-chromium head. The primary outcome was proximal head penetration measured with radiostereometric analysis (RSA) at two years. Secondary outcomes were volumetric wear, periacetabular radiolucencies, and patient-reported outcomes. Results. At two years, 44 patients in each group were available for RSA assessment. The median total two-year proximal head penetration was -0.02 mm (IQR -0.09 to 0.07; p = 0.548) for the largest possible head and -0.01 mm (IQR -0.07 to 0.10; p = 0.525) for 32 mm heads. Their difference was not statistically significant (p = 0.323). Neither group demonstrated a bedding-in period. The median steady-state volumetric wear rates were 6.1 mm. 3. /year (IQR -59 to 57) and 3.5 mm. 3. /year (-21 to 34) respectively, and did not differ between the groups (p = 0.848). There were no statistically significant differences in periacetabular radiolucencies or patient-reported outcomes. Conclusion. The use of the largest possible metal head did not increase vitamin E-doped cross-linked polyethylene wear compared with 32 mm heads at two years. Linear wear was negligible and volumetric wear rates were very low in both head size groups. There was a tendency towards higher values of volumetric wear in large heads that warrants longer-term evaluation before any definite conclusions about the association between head size and wear can be drawn. Cite this article: Bone Joint J 2021;103-B(7):1206–1214


The Bone & Joint Journal
Vol. 103-B, Issue 4 | Pages 644 - 649
1 Apr 2021
Alsousou J Oragu E Martin A Strickland L Newman S Kendrick B Taylor A Glyn-Jones S

Aims. The aim of this prospective cohort study was to evaluate the early migration of the TriFit cementless proximally coated tapered femoral stem using radiostereometric analysis (RSA). Methods. A total of 21 patients (eight men and 13 women) undergoing primary total hip arthroplasty (THA) for osteoarthritis of the hip were recruited in this study and followed up for two years. Two patients were lost to follow-up. All patients received a TriFit stem and Trinity Cup with a vitamin E-infused highly cross-linked ultra-high molecular weight polyethylene liner. Radiographs for RSA were taken postoperatively and then at three, 12, and 24 months. Oxford Hip Score (OHS), EuroQol five-dimension questionnaire (EQ-5D), and adverse events were reported. Results. At two years, the mean subsidence of the head and tip for the TriFit stem was 0.38 mm (SD 0.32) and 0.52 mm (SD 0.36), respectively. The total migration of the head and tip was 0.55 mm (SD 0.32) and 0.71 mm (SD 0.38), respectively. There were no statistically significant differences between the three to 12 months' migration (p = 0.105) and 12 to 24 months' migration (p = 0.694). The OHS and EQ-5D showed significant improvements at two years. Conclusion. The results of this study suggest that the TriFit femoral stem achieves initial stability and is likely to be stable in the mid and long term. A long-term outcome study is required to assess late mechanisms of failure and the effects of bone mineral density (BMD) related changes. Cite this article: Bone Joint J 2021;103-B(4):644–649


The Bone & Joint Journal
Vol. 100-B, Issue 12 | Pages 1579 - 1584
1 Dec 2018
Turgeon TR Gascoyne TC Laende EK Dunbar MJ Bohm ER Richardson CG

Aims. The introduction of a novel design of total knee arthroplasty (TKA) must achieve outcomes at least as good as existing designs. A novel design of TKA with a reducing radius of the femoral component and a modified cam-post articulation has been released and requires assessment of the fixation to bone. Radiostereometric analysis (RSA) of the components within the first two postoperative years has been shown to be predictive of medium- to long-term fixation. The aim of this study was to assess the stability of the tibial component of this system during this period of time using RSA. Patients and Methods. A cohort of 30 patients underwent primary, cemented TKA using the novel posterior stabilized fixed-bearing (ATTUNE) design. There was an even distribution of men and women (15:15). The mean age of the patients was 64 years (sd 8) at the time of surgery; their mean body mass index (BMI) was 35.4 kg/m2 (sd 7.9). RSA was used to assess the stability of the tibial component at 6, 12, and 24 months compared with a six-week baseline examination. Patient-reported outcome measures were also assessed. Results. The mean maximum total point motion (MTPM) of the tibial component between 12 and 24 months postoperatively was 0.08 mm (sd 0.08), which is well below the published threshold of 0.2 mm (p < 0.001). Patient-reported outcome measures consistently improved. Conclusion. The tibial component of this novel design of TKA showed stability between assessment 12 and 24 months postoperatively, suggesting an acceptably low risk of medium- to long-term failure due to aseptic loosening


The Bone & Joint Journal
Vol. 100-B, Issue 9 | Pages 1182 - 1186
1 Sep 2018
Werner BS Chaoui J Walch G

Aims. Scapular notching is a frequently observed radiographic phenomenon in reverse shoulder arthroplasty (RSA), signifying impingement of components. The purposes of this study were to evaluate the effect of glenoid component size and glenosphere type on impingement-free range of movement (ROM) for extension and internal and external rotation in a virtual RSA model, and to determine the optimal configuration to reduce the incidence of friction-type scapular notching. Materials and Methods. Preoperative CT scans obtained in 21 patients (three male, 18 female) with primary osteoarthritis were analyzed using modelling software. Two concurrent factors were tested for impingement-free ROM and translation of the centre of rotation: glenosphere diameter (36 mm vs 39 mm) and type (centred, 2 mm inferior eccentric offset, 10° inferior tilt). Results. Glenosphere size was most predictive of increased extension and external rotation, whereas lateralization of the centre of rotation was the most predictive factor for internal rotation. A larger diameter of glenosphere combined with a 10° tilted configuration demonstrated superior values for extension and external rotation, whereas the eccentric component improved internal rotation by a mean 8.9° (standard deviation 2.7°) compared with a standard concentric glenosphere. Conclusion. Glenosphere configuration can be modified to increase range of movement in RSA. Friction-type scapular notching was most effectively reduced by use of a large-diameter glenosphere with 10° inferior tilt. Cite this article: Bone Joint J 2018;100-B:1182–6


The Bone & Joint Journal
Vol. 103-B, Issue 5 | Pages 964 - 970
1 May 2021
Ling DI Schneider B Ode G Lai EY Gulotta LV

Aims. To investigate the impact of the Charlson and Elixhauser comorbidity indices on patient-reported outcomes measures (PROMs) following shoulder arthroplasty. Methods. Patients undergoing total shoulder arthroplasty (TSA), reverse shoulder arthroplasty (RSA), or hemiarthroplasty (HA) from 2016 to 2018 were identified, along with the Charlson and Elixhauser comorbidities listed as their secondary diagnoses in the electronic medical records. Patients were matched to our institution’s registry to obtain their PROMs, including shoulder-specific (American Shoulder and Elbow Society (ASES) and Shoulder Activity Scale (SAS)) and general health scales (12-Item Short Form Survey (SF-12) and Patient-Reported Outcomes Measurement Information System-Pain Interference). Linear regression models adjusting for age and sex were used to evaluate the association between increasing number of comorbidities and PROM scores. A total of 1,817 shoulder arthroplasties were performed: 1,017 (56%) TSA, 726 (40%) RSA, and 74 (4%) HA. The mean age was 67 years (SD 10), and 936 (52%) of the patients were female. Results. The most common comorbidities were obesity (1,256, 69%) and hypertension (990, 55%). Patients with more comorbidities had lower ASES and SAS scores at baseline (p < 0.001). Elixhauser comorbidities continued to negatively impact ASES and SAS scores at one year (p = 0.002) and two-year follow-up (p = 0.002). Patients with more comorbidities reported greater pain interference on PROMIS at baseline (p = 0.007), but not at two years. Higher number of Charlson comorbidities were associated with lower scores on the SF-12 mental component at baseline (p < 0.001) and two years (p = 0.020). Higher number of Elixhauser comorbidities were associated with lower SF-12 physical component scores at baseline (p < 0.001) and two years (p = 0.004). Conclusion. Higher number of comorbidities was associated with lower baseline scores and worse outcomes on both shoulder-specific and general health PROMs. The presence of specific comorbidities may be used during shared decision-making to manage expectations for patients undergoing shoulder arthroplasty. Cite this article: Bone Joint J 2021;103-B(5):964–970