The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 7 | Pages 855 - 858
1 Jul 2009
Wroblewski BM Siney PD Fleming PA

The design of the Charnley total hip replacement follows the principle of low frictional torque. It is based on the largest possible difference between the radius of the femoral head and that of the outer aspect of the acetabular component. The aim is to protect the bone-cement interface by movement taking place at the smaller radius, the articulation. This is achieved in clinical practice by a 22.225 mm diameter head articulating with a 40 mm or 43 mm diameter acetabular component of ultra-high molecular weight polyethylene. We compared the incidence of aseptic loosening of acetabular components with an outer diameter of 40 mm and 43 mm at comparable depths of penetration with a mean follow-up of 17 years (1 to 40). In cases with no measurable wear none of the acetabular components were loose. With increasing acetabular penetration there was an increased incidence of aseptic loosening which reflected the difference in the external radii, with 1.5% at 1 mm, 8.8% at 2 mm, 9.7% at 3 mm and 9.6% at 4 mm of penetration in favour of the larger 43 mm acetabular component. Our findings support the Charnley principle of low frictional torque. The level of the benefit is in keeping with the predicted values


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 8 | Pages 1171 - 1175
1 Aug 2010
Hajipour L Gulihar A Dias J

We carried out lacerations of 50%, followed by trimming, in ten turkey flexor tendons in vitro and measured the coefficient of friction at the tendon-pulley interface with loads of 200 g and 400 g and in 10°, 30°, 50° and 70° of flexion. Laceration increased the coefficient of friction from 0.12 for the intact tendon to 0.3 at both the test loads. Trimming the laceration reduced the coefficient of friction to 0.2. An exponential increase in the gliding resistance was found at 50° and 70° of flexion (p = 0.02 and p = 0.003, respectively) following trimming compared to that of the intact tendon. We concluded that trimming partially lacerated flexor tendons will reduce the gliding resistance at the tendon-pulley interface, but will lead to fragmentation and triggering of the tendon at higher degrees of flexion and loading. We recommend that higher degrees of flexion be avoided during early post-operative rehabilitation following trimming of a flexor tendon


The Bone & Joint Journal
Vol. 97-B, Issue 4 | Pages 463 - 472
1 Apr 2015
Panagiotidou A Meswania J Osman K Bolland B Latham J Skinner J Haddad FS Hart A Blunn G

The aim of this study was to assess the effect of frictional torque and bending moment on fretting corrosion at the taper interface of a modular femoral component and to investigate whether different combinations of material also had an effect. The combinations we examined were 1) cobalt–chromium (CoCr) heads on CoCr stems 2) CoCr heads on titanium alloy (Ti) stems and 3) ceramic heads on CoCr stems. In test 1 increasing torque was imposed by offsetting the stem in the anteroposterior plane in increments of 0 mm, 4 mm, 6 mm and 8 mm when the torque generated was equivalent to 0 Nm, 9 Nm, 14 Nm and 18 Nm. In test 2 we investigated the effect of increasing the bending moment by offsetting the application of axial load from the midline in the mediolateral plane. Increments of offset equivalent to head + 0 mm, head + 7 mm and head + 14 mm were used. Significantly higher currents and amplitudes were seen with increasing torque for all combinations of material. However, Ti stems showed the highest corrosion currents. Increased bending moments associated with using larger offset heads produced more corrosion: Ti stems generally performed worse than CoCr stems. Using ceramic heads did not prevent corrosion, but reduced it significantly in all loading configurations. Cite this article: Bone Joint J 2015;97-B:463–72


The Journal of Bone & Joint Surgery British Volume
Vol. 59-B, Issue 4 | Pages 385 - 392
1 Nov 1977
Hardinge K Williams D Etienne A MacKenzie D Charnley J

Fifty-four hips converted to low friction arthroplasty between 1965 and 1975 have been reviewed one to eleven years after operation. In many cases malposition had led to degenerative changes in the opposite hip, the lumbar spine or the knee, often with severe loss of function due to pain. It was found that total replacement could give useful relief of pain and improved function, though the range of movement obtained was not as good as in primary replacement. An outstanding feature was the correction of inequality of leg length. In general, the results were much better in cases of ankylosis acquired in adult life than in cases of spontaneous fusion after sepsis in childhood. The most important complication was a single case of sciatic palsy


The Journal of Bone & Joint Surgery British Volume
Vol. 70-B, Issue 3 | Pages 373 - 376
1 May 1988
Laforgia R Murphy J Redfern T

We report the results of low friction arthroplasty in 72 old tuberculous hips and 42 hips with old quiescent septic arthritis. Follow-up was for a minimum of two years. Recrudescence of tuberculosis was seen in only one patient, but deep infection occurred in four of the hips with previous septic arthritis. The reasons for this are discussed. It is recommended that patients with old tuberculous hips should have a course of anti-tuberculous drugs before operation, that patients with previous septic infection should have prophylactic antibiotics and antibiotic-loaded cement and that the operation should be performed by an experienced surgeon


The Journal of Bone & Joint Surgery British Volume
Vol. 70-B, Issue 2 | Pages 182 - 186
1 Mar 1988
Hartofilakidis G Stamos K Ioannidis T

We report the replacement of 42 hips in 34 adults with untreated congenital dislocation. We used Charnley low friction implants, cementing the cup at the level of the true acetabulum after deepening and enlarging it by our own technique of cotyloplasty. Results were evaluated in 38 hips after a mean of 5.5 years. All the patients showed marked improvement, with no infection and, as yet, no late revision. The technical difficulties of the operation and the complications are discussed


The Journal of Bone & Joint Surgery British Volume
Vol. 70-B, Issue 5 | Pages 756 - 760
1 Nov 1988
Kim Y Ko C Ahn J Yoon Y Kwak B

We report 60 patients with tuberculosis of the hip treated by Charnley low friction arthroplasty and followed for eight to 13 years. Eight of them had active tuberculosis of the hip at the time of operation, and all were covered by relatively short courses of antituberculous drugs. Our study suggests that arthroplasty can be recommended for these patients provided that adequate chemotherapy is given both before and after operation


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 6 | Pages 735 - 745
1 Jun 2012
Jeffers JRT Walter WL

This systematic review of the literature summarises the clinical experience with ceramic-on-ceramic hip bearings over the past 40 years and discusses the concerns that exist in relation to the bearing combination. Loosening, fracture, liner chipping on insertion, liner canting and dissociation, edge-loading and squeaking have all been reported, and the relationship between these issues and implant design and surgical technique is investigated. New design concepts are introduced and analysed with respect to previous clinical experience.


The Journal of Bone & Joint Surgery British Volume
Vol. 86-B, Issue 1 | Pages 149 - 149
1 Jan 2004
FRÖLKE JPM REELING BROUWER GC


The Journal of Bone & Joint Surgery British Volume
Vol. 84-B, Issue 4 | Pages 492 - 496
1 May 2002
Giannoudis PV Snowden S Matthews SJ Smye SW Smith RM

We have carried out a prospective, randomised trial to measure the rise of temperature during reaming of the tibia before intramedullary nailing. We studied 34 patients with a mean age of 35.1 years (18 to 63) and mean injury severity score of 10 (9 to 13). The patients were randomised into two groups: group 1 included 18 patients whose procedure was undertaken without a tourniquet and group 2, 16 patients in whom a tourniquet was used. The temperature in the bone was measured directly by two thermocouples inserted into the cortical bone near the isthmus of the tibial diaphysis. Reaming was carried out to at least 1.5 mm above the required diameter of the nail. Blood loss was assessed by recording the preoperative and postoperative haemoglobin (Hb) level.

The minimum clinical follow-up was six months. In group 1 (no tourniquet), the mean Hb dropped 2.8 g/dl from 14.3 ± 1.02 g/dl to 11.5 ± 1.04 g/dl (p = 0.0001), whereas with the tourniquet, the mean decrease was 1.3 g/dl from 14 ± 1 g/dl to 12.7 ± 1.3 g/dl (p = 0.007). This difference was not statistically significant. The mean initial tibial temperature was 35.6°C (sd 0.6) and rose with reaming to levels between 36.3°C and 51.6°C. The highest temperatures were obtained with the largest reamers (11 and 12 mm, p = 0.0001) and the most rapid rise with the smallest diameters of medullary canal (8 or 9 mm). The rise of temperature was transient (20s).

We were unable to identify any effect of the use of a tourniquet on the temperature achieved. Reamed intramedullary tibial nailing induces a transient elevation of temperature which is directly related to the amount of reaming.


The Journal of Bone & Joint Surgery British Volume
Vol. 71-B, Issue 3 | Pages 523 - 525
1 May 1989
Hodgkinson J Shelley P Wroblewski B

In a prospective study, a double crossover wire with a compression spring was used to re-attach 52 un-united trochanters at revision operations on total hip arthroplasties. Bony union was achieved in 42 (81%) and was not influenced by the duration of the nonunion or the separation gap. The new method compared favourably with earlier revisions at which other methods of trochanteric re-attachment had been used.


The Bone & Joint Journal
Vol. 103-B, Issue 12 | Pages 1791 - 1801
1 Dec 2021
Bhalekar RM Nargol ME Shyam N Nargol AVF Wells SR Collier R Pabbruwe M Joyce TJ Langton DJ

Aims. The aim of this study was to investigate whether wear and backside deformation of polyethylene (PE) tibial inserts may influence the cement cover of tibial trays of explanted total knee arthroplasties (TKAs). Methods. At our retrieval centre, we measured changes in the wear and deformation of PE inserts using coordinate measuring machines and light microscopy. The amount of cement cover on the backside of tibial trays was quantified as a percentage of the total surface. The study involved data from the explanted fixed-bearing components of four widely used contemporary designs of TKA (Attune, NexGen, Press Fit Condylar (PFC), and Triathlon), revised for any indication, and we compared them with components that used previous generations of PE. Regression modelling was used to identify variables related to the amount of cement cover on the retrieved trays. Results. A total of 114 explanted fixed-bearing TKAs were examined. This included 76 used with contemporary PE inserts which were compared with 15 used with older generation PEs. The Attune and NexGen (central locking) trays were found to have significantly less cement cover than Triathlon and PFC trays (peripheral locking group) (p = 0.001). The median planicity values of the PE inserts used with central locking trays were significantly greater than of those with peripheral locking inserts (205 vs 85 microns; p < 0.001). Attune and NexGen inserts had a characteristic pattern of backside deformation, with the outer edges of the PE deviating inferiorly, leaving the PE margins as the primary areas of articulation. Conclusion. Explanted TKAs with central locking mechanisms were significantly more likely to debond from the cement mantle. The PE inserts of these designs showed characteristic patterns of deformation, which appeared to relate to the manufacturing process and may be exacerbated in vivo. This pattern of deformation was associated with PE wear occurring at the outer edges of the articulation, potentially increasing the frictional torque generated at this interface. Cite this article: Bone Joint J 2021;103-B(12):1791–1801


The Bone & Joint Journal
Vol. 99-B, Issue 4 | Pages 554 - 560
1 Apr 2017
Tamai K Suzuki A Takahashi S Akhgar J Rahmani MS Hayashi K Ohyama S Nakamura H

Aims. We aimed to evaluate the temperature around the nerve root during drilling of the lamina and to determine whether irrigation during drilling can reduce the chance of nerve root injury. Materials and Methods. Lumbar nerve roots were exposed to frictional heat by high-speed drilling of the lamina in a live rabbit model, with saline (room temperature (RT) or chilled saline) or without saline (control) irrigation. We measured temperatures surrounding the nerve root and made histological evaluations. Results. In the control group, the mean temperature around the nerve root was 52.0°C (38.0°C to 75.5°C) after 60 seconds of drilling, and nerve root injuries were found in one out of 13 (7.7%) immediately, three out of 14 (21.4%) at three days, and 11 out of 25 (44.0%) at seven days post-operatively. While the RT group showed a significantly lower temperature around the nerve root compared with the control group (mean 46.5°C; 34.5°C to 66.9°C, p < 0.001), RT saline failed to significantly reduce the incidence of nerve root injury (ten out of 26; 38.5%; odds ratio (OR) 0.96; 95% confidence interval (CI) 0.516 to 1.785; p = 0.563). However, chilled saline irrigation resulted in a significantly lower temperature than the control group (mean 39.0°C; 35.3°C to 52.3°C; p < 0.001) and a lower rate of nerve root injury (two out of 21; 9.5%, OR 0.13; 95% CI 0.02 to 0.703, p = 0.010). Conclusion. Frictional heat caused by a high-speed drill can cause histological nerve root injury. Chilled saline irrigation had a more prominent effect than RT in reducing the incidence of the thermal injury during extended drilling. Cite this article: Bone Joint J 2017;99-B:554–60


The Bone & Joint Journal
Vol. 106-B, Issue 4 | Pages 312 - 318
1 Apr 2024
Sheth NP Jones SA Sanghavi SA Manktelow A

The advent of modular porous metal augments has ushered in a new form of treatment for acetabular bone loss. The function of an augment can be seen as reducing the size of a defect or reconstituting the anterosuperior/posteroinferior columns and/or allowing supplementary fixation. Depending on the function of the augment, the surgeon can decide on the sequence of introduction of the hemispherical shell, before or after the augment. Augments should always, however, be used with cement to form a unit with the acetabular component. Given their versatility, augments also allow the use of a hemispherical shell in a position that restores the centre of rotation and biomechanics of the hip. Progressive shedding or the appearance of metal debris is a particular finding with augments and, with other radiological signs of failure, should be recognized on serial radiographs. Mid- to long-term outcomes in studies reporting the use of augments with hemispherical shells in revision total hip arthroplasty have shown rates of survival of > 90%. However, a higher risk of failure has been reported when augments have been used for patients with chronic pelvic discontinuity.

Cite this article: Bone Joint J 2024;106-B(4):312–318.


The Bone & Joint Journal
Vol. 106-B, Issue 5 Supple B | Pages 74 - 81
1 May 2024
Callary SA Broekhuis D Barends J Ramasamy B Nelissen RGHH Solomon LB Kaptein BL

Aims

The aim of this study was to compare the biomechanical models of two frequently used techniques for reconstructing severe acetabular defects with pelvic discontinuity in revision total hip arthroplasty (THA) – the Trabecular Metal Acetabular Revision System (TMARS) and custom triflange acetabular components (CTACs) – using virtual modelling.

Methods

Pre- and postoperative CT scans from ten patients who underwent revision with the TMARS for a Paprosky IIIB acetabular defect with pelvic discontinuity were retrospectively collated. Computer models of a CTAC implant were designed from the preoperative CT scans of these patients. Computer models of the TMARS reconstruction were segmented from postoperative CT scans using a semi-automated method. The amount of bone removed, the implant-bone apposition that was achieved, and the restoration of the centre of rotation of the hip were compared between all the actual TMARS and the virtual CTAC implants.


The Bone & Joint Journal
Vol. 106-B, Issue 11 | Pages 1273 - 1283
1 Nov 2024
Mahmud H Wang D Topan-Rat A Bull AMJ Heinrichs CH Reilly P Emery R Amis AA Hansen UN

Aims

The survival of humeral hemiarthroplasties in patients with relatively intact glenoid cartilage could theoretically be extended by minimizing the associated postoperative glenoid erosion. Ceramic has gained attention as an alternative to metal as a material for hemiarthroplasties because of its superior tribological properties. The aim of this study was to assess the in vitro wear performance of ceramic and metal humeral hemiarthroplasties on natural glenoids.

Methods

Intact right cadaveric shoulders from donors aged between 50 and 65 years were assigned to a ceramic group (n = 8, four male cadavers) and a metal group (n = 9, four male cadavers). A dedicated shoulder wear simulator was used to simulate daily activity by replicating the relevant joint motion and loading profiles. During testing, the joint was kept lubricated with diluted calf serum at room temperature. Each test of wear was performed for 500,000 cycles at 1.2 Hz. At intervals of 125,000 cycles, micro-CT scans of each glenoid were taken to characterize and quantify glenoid wear by calculating the change in the thickness of its articular cartilage.


The Journal of Bone & Joint Surgery British Volume
Vol. 44-B, Issue 3 | Pages 662 - 674
1 Aug 1962
Barnett CH Cobbold AF

1. By the use of a device that allows movement of a human finger joint to take place uninfluenced by muscle activity, the coefficient of friction has been determined between living articular surfaces. 2. The force of friction rises as the range of movement is increased, because of the tension then developing in the ligaments and the soft tissues surrounding the joint. 3. Measurements have also been made of the forces of friction within the ankle joint of the dog and within four types of reciprocating bearings (plastic, oil-lubricated, "floating" and hydrostatic). 4. By altering the load borne by the joints and bearings it has been shown that joints resemble in their behaviour those bearings in which a film of fluid is maintained between the fixed and moving surfaces, rather than bearings depending wholly or in part on boundary lubrication. 5. It is suggested that joints normally owe their great freedom of movement to a special type of fluid film lubrication that has been termed "weeping lubrication," supplemented by "floating lubrication," though on occasions boundary conditions may prevail


The Bone & Joint Journal
Vol. 105-B, Issue 8 | Pages 912 - 919
1 Aug 2023
Cunningham LJ Walton M Bale S Trail IA

Aims

Reverse total shoulder arthroplasty (rTSA) can be used in complex cases when the glenoid requires reconstruction. In this study, a baseplate with composite bone autograft and a central trabecular titanium peg was implanted, and its migration was assessed for two years postoperatively using radiostereometric analysis (RSA).

Methods

A total of 14 patients who underwent a rTSA with an autograft consented to participate. Of these, 11 had a primary rTSA using humeral head autograft and three had a revision rTSA with autograft harvested from the iliac crest. The mean age of the patients was 66 years (39 to 81). Tantalum beads were implanted in the scapula around the glenoid. RSA imaging (stereographic radiographs) was undertaken immediately postoperatively and at three, six, 12, and 24 months. Analysis was completed using model-based RSA software. Outcomes were collected preoperatively and at two years postoperatively, including the Oxford Shoulder Score, the American Shoulder and Elbow Score, and a visual analogue score for pain. A Constant score was also obtained for the assessment of strength and range of motion.


The Bone & Joint Journal
Vol. 105-B, Issue 11 | Pages 1177 - 1183
1 Nov 2023
van der Graaff SJA Reijman M Meuffels DE Koopmanschap MA

Aims

The aim of this study was to evaluate the cost-effectiveness of arthroscopic partial meniscectomy versus physical therapy plus optional delayed arthroscopic partial meniscectomy in young patients aged under 45 years with traumatic meniscal tears.

Methods

We conducted a multicentre, open-labelled, randomized controlled trial in patients aged 18 to 45 years, with a recent onset, traumatic, MRI-verified, isolated meniscal tear without knee osteoarthritis. Patients were randomized to arthroscopic partial meniscectomy or standardized physical therapy with an optional delayed arthroscopic partial meniscectomy after three months of follow-up. We performed a cost-utility analysis on the randomization groups to compare both treatments over a 24-month follow-up period. Cost utility was calculated as incremental costs per quality-adjusted life year (QALY) gained of arthroscopic partial meniscectomy compared to physical therapy. Calculations were performed from a healthcare system perspective and a societal perspective.


The Bone & Joint Journal
Vol. 106-B, Issue 3 Supple A | Pages 81 - 88
1 Mar 2024
Lustig S Cotte M Foissey C Asirvatham RD Servien E Batailler C

Aims

The benefit of a dual-mobility acetabular component (DMC) for primary total hip arthroplasties (THAs) is controversial. This study aimed to compare the dislocation and complication rates when using a DMC compared to single-mobility (SM) acetabular component in primary elective THA using data collected at a single centre, and compare the revision rates and survival outcomes in these two groups.

Methods

Between 2010 and 2019, 2,075 primary THAs using either a cementless DM or SM acetabular component were included. Indications for DMC were patients aged older than 70 years or with high risk of dislocation. All other patients received a SM acetabular component. Exclusion criteria were cemented implants, patients treated for femoral neck fracture, and follow-up of less than one year. In total, 1,940 THAs were analyzed: 1,149 DMC (59.2%) and 791 SM (40.8%). The mean age was 73 years (SD 9.2) in the DMC group and 57 years (SD 12) in the SM group. Complications and revisions have been analyzed retrospectively.


The Bone & Joint Journal
Vol. 105-B, Issue 6 | Pages 610 - 621
1 Jun 2023
Prodromidis AD Chloros GD Thivaios GC Sutton PM Pandit H Giannoudis PV Charalambous CP

Aims

Loosening of components after total knee arthroplasty (TKA) can be associated with the development of radiolucent lines (RLLs). The aim of this study was to assess the rate of formation of RLLs in the cemented original design of the ATTUNE TKA and their relationship to loosening.

Methods

A systematic search was undertaken using the Cochrane methodology in three online databases: MEDLINE, Embase, and CINAHL. Studies were screened against predetermined criteria, and data were extracted. Available National Joint Registries in the Network of Orthopaedic Registries of Europe were also screened. A random effects model meta-analysis was undertaken.


The Bone & Joint Journal
Vol. 98-B, Issue 1 | Pages 117 - 124
1 Jan 2016
Takenaka S Hosono N Mukai Y Tateishi K Fuji T

Aims. The aim of this study was to determine whether chilled irrigation saline decreases the incidence of clinical upper limb palsy (ULP; a reduction of one grade or more on manual muscle testing; MMT), based on the idea that ULP results from thermal damage to the nerve roots by heat generated by friction during bone drilling. Methods. Irrigation saline for drilling was used at room temperature (RT, 25.6°C) in open-door laminoplasty in 400 patients (RT group) and chilled to a mean temperature of 12.1°C during operations for 400 patients (low-temperature (LT) group). We assessed deltoid, biceps, and triceps brachii muscle strength by MMT. ULP occurring within two days post-operatively was categorised as early-onset palsy. Results. The incidence of ULP (4.0% vs 9.5%, p = 0.003), especially early-onset palsy (1.0% vs 5.5%, p < 0.001), was significantly lower for the LT group than for the RT group. Multivariate analysis indicated that RT irrigation saline use, concomitant foraminotomy, and opened side were significant predictors for ULP. Discussion. Using chilled irrigation saline during bone drilling significantly decreased the ULP incidence, particularly the early-onset type, and shortened the recovery period for ULP. Chilled irrigation saline can thus be recommended as a simple method for preventing ULP. Take home message: Chilled irrigation during laminoplasty reduces C5 palsy. Cite this article: Bone Joint J 2016;98-B:117–24


The Bone & Joint Journal
Vol. 104-B, Issue 9 | Pages 1032 - 1038
1 Sep 2022
García-Rey E Cruz-Pardos A Saldaña L

Aims

A significant reduction in wear at five and ten years was previously reported when comparing Durasul highly cross-linked polyethylene with nitrogen-sterilized Sulene polyethylene in total hip arthroplasty (THA). We investigated whether the improvement observed at the earlier follow-up continued, resulting in decreased osteolysis and revision surgery rates over the second decade.

Methods

Between January 1999 and December 2001, 90 patients underwent surgery using the same acetabular and femoral components with a 28 mm metallic femoral head and either a Durasul or Sulene liner. A total of 66 hips of this prospective randomized study were available for a minimum follow-up of 20 years. The linear femoral head penetration rate was measured at six weeks, one year, and annually thereafter, using the Dorr method on digitized radiographs with a software package.


The Journal of Bone & Joint Surgery British Volume
Vol. 82-B, Issue 8 | Pages 1182 - 1188
1 Nov 2000
Barker DS Wang AW Yeo MF Nawana NS Brumby SA Pearcy MJ Howie DW

We studied the effect of the surface finish of the stem on the transfer of load in the proximal femur in a sheep model of cemented hip arthroplasty. Strain-gauge analysis and corresponding finite-element (FE) analysis were performed to assess the effect of friction and creep at the cement-stem interface. No difference was seen between the matt and polished stems. FE analysis showed that the effects of cement creep and friction at the stem-cement interface on femoral strain were small compared with the effect of inserting a cemented stem


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 4 | Pages 549 - 556
1 Apr 2007
Udofia I Liu F Jin Z Roberts P Grigoris P

Finite element analysis was used to examine the initial stability after hip resurfacing and the effect of the procedure on the contact mechanics at the articulating surfaces. Models were created with the components positioned anatomically and loaded physiologically through major muscle forces. Total micromovement of less than 10 μm was predicted for the press-fit acetabular components models, much below the 50 μm limit required to encourage osseointegration. Relatively high compressive acetabular and contact stresses were observed in these models. The press-fit procedure showed a moderate influence on the contact mechanics at the bearing surfaces, but produced marked deformation of the acetabular components. No edge contact was predicted for the acetabular components studied. It is concluded that the frictional compressive stresses generated by the 1 mm to 2 mm interference-fit acetabular components, together with the minimal micromovement, would provide adequate stability for the implant, at least in the immediate post-operative situation


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 2 | Pages 158 - 162
1 Feb 2012
Sternheim A Backstein D Kuzyk PRT Goshua G Berkovich Y Safir O Gross AE

We report the use of porous metal acetabular revision shells in the treatment of contained bone loss. The outcomes of 53 patients with ≤ 50% acetabular bleeding host bone contact were compared with a control group of 49 patients with > 50% to 85% bleeding host bone contact. All patients were treated with the same type of trabecular metal acetabular revision shell. The mean age at revision was 62.4 years (42 to 80) and the mean follow-up of both groups was 72.4 months (60 to 102). Clinical, radiological and functional outcomes were assessed. There were four (7.5%) mechanical failures in the ≤ 50% host bone contact group and no failures in the > 50% host bone contact group (p = 0.068). Out of both groups combined there were four infections (3.9%) and five recurrent dislocations (4.9%) with a stable acetabular component construct that were revised to a constrained liner. Given the complexity of the reconstructive challenge, porous metal revision acetabular shells show acceptable failure rates at five to ten years’ follow-up in the setting of significant contained bone defects. This favourable outcome might be due to the improved initial stability achieved by a high coefficient of friction between the acetabular implant and the host bone, and the high porosity, which affords good bone ingrowth


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 4 | Pages 486 - 488
1 Apr 2010
Wroblewski BM Purbach B Siney PD Fleming PA

Increasing follow-up identifies the outcome in younger patients who have undergone total hip replacement (THR) and reveals the true potential for survival of the prosthesis. We identified 28 patients (39 THRs) who had undergone cemented Charnley low friction arthroplasty between 1969 and 2001. Their mean age at operation was 17.9 years (12 to 19) and the maximum follow-up was 34 years. Two patients (4 THRs) were lost to follow-up, 13 (16 THRs) were revised at a mean period of 19.1 years (8 to 34) and 13 (19 THRs) continue to attend regular follow-up at a mean of 12.6 years (2.3 to 29). In this surviving group one acetabular component was radiologically loose and all femoral components were secure. In all the patients the diameter of the femoral head was 22.225 mm with Charnley femoral components used in 29 hips and C-stem femoral components in ten. In young patients who require THR the acetabular bone stock is generally a limiting factor for the size of the component. Excellent long-term results can be obtained with a cemented polyethylene acetabular component and a femoral head of small diameter


The Journal of Bone & Joint Surgery British Volume
Vol. 54-B, Issue 4 | Pages 590 - 599
1 Nov 1972
Andersson GBJ Freeman MAR Swanson SAV

1. Prosthetic acetabular cups of the Charnley and McKee-Farrar designs were cemented into cadaveric pelves using different procedures for preparing the acetabulum. 2. The torsional moments needed to loosen these cups were measured. 3. The torsional moments so measured were found to be from about four to more than twenty times higher than the frictional moments measured in independent tests on the two designs of prosthesis. 4. It is argued that late looseness of the acetabular component after total hip replacement, in the absence of infection, seems most likely to be due to thermal damage to the bone occurring at the time of polymerisation of the cement, and to subsequent bone resorption. 5. Surgical preparation of the acetabulum should include removal of all the articular cartilage and cleaning of the acetabular fossa, but the drilling of additional holes in the floor of the acetabulum seems unimportant. 6. The possibility of fatigue fracture in bone as a factor contributing to late loosening is an argument in favour of metal-on-polyethylene prostheses with their lower frictional moments, although the importance of this factor cannot be estimated


The Bone & Joint Journal
Vol. 104-B, Issue 7 | Pages 833 - 843
1 Jul 2022
Kayani B Baawa-Ameyaw J Fontalis A Tahmassebi J Wardle N Middleton R Stephen A Hutchinson J Haddad FS

Aims

This study reports the ten-year wear rates, incidence of osteolysis, clinical outcomes, and complications of a multicentre randomized controlled trial comparing oxidized zirconium (OxZr) versus cobalt-chrome (CoCr) femoral heads with ultra-high molecular weight polyethylene (UHMWPE) and highly cross-linked polyethylene (XLPE) liners in total hip arthroplasty (THA).

Methods

Patients undergoing primary THA were recruited from four institutions and prospectively allocated to the following treatment groups: Group A, CoCr femoral head with XLPE liner; Group B, OxZr femoral head with XLPE liner; and Group C, OxZr femoral head with UHMWPE liner. All study patients and assessors recording outcomes were blinded to the treatment groups. The outcomes of 262 study patients were analyzed at ten years’ follow-up.


The Journal of Bone & Joint Surgery British Volume
Vol. 55-B, Issue 4 | Pages 759 - 773
1 Nov 1973
Swanson SAV Freeman MAR Heath JC

1. Currently available total replacement hip and knee prostheses were tested in a machine enabling flexion-extension movements to be applied whilst the prostheses were surrounded with Ringer's solution or other liquid and loaded within the physiological range. 2. Prostheses of which both components were made in cobalt-chromium-molybdenum alloy produced visible quantities of alloy particles, whose sizes ranged down to about 0·1 microns, and cobalt and molybdenum ions in solution. 3. No metallic or plastic particles were detected during tests on a hip prosthesis made of stainless steel and high density polyethylene. 4. The frictional moments in cobalt-chromium-molybdenum hip prostheses were higher than in stainless steel-polyethylene hip prostheses, by a factor of at least 2 to 1. 5. It is accepted that the conditions of these tests were probably more severe than in life, but the difference is held to be one of degree and not one of kind. 6. The particulate alloy debris, when injected in massive doses into the muscles of rats, gave an incidence of malignant tumours which was comparable to that already established for pure cobalt powder, whereas particles of several other metals, tested in the same way, gave no tumours. 7. It is argued that the particles which are known to be produced in at least some patients using cobalt-chromium-molybdenum total replacement joint prostheses constitute a risk of tumour formation which is certainly small, possibly negligible, but not accurately calculable at present. 8. The results of these tests, particularly the differences in frictional moment and in the production of particulate debris, suggest a preference for high density polyethylene as one component of a total joint replacement prosthesis


The Bone & Joint Journal
Vol. 103-B, Issue 7 | Pages 1206 - 1214
1 Jul 2021
Tsikandylakis G Mortensen KRL Gromov K Mohaddes M Malchau H Troelsen A

Aims

We aimed to investigate if the use of the largest possible cobalt-chromium head articulating with polyethylene acetabular inserts would increase the in vivo wear rate in total hip arthroplasty.

Methods

In a single-blinded randomized controlled trial, 96 patients (43 females), at a median age of 63 years (interquartile range (IQR) 57 to 69), were allocated to receive either the largest possible modular femoral head (36 mm to 44 mm) in the thinnest possible insert or a standard 32 mm head. All patients received a vitamin E-doped cross-linked polyethylene insert and a cobalt-chromium head. The primary outcome was proximal head penetration measured with radiostereometric analysis (RSA) at two years. Secondary outcomes were volumetric wear, periacetabular radiolucencies, and patient-reported outcomes.


The Journal of Bone & Joint Surgery British Volume
Vol. 46-B, Issue 3 | Pages 544 - 545
1 Aug 1964
Smithuis T

1. Exostosis bursata in a patient with hereditary multiple exostoses is described. 2. Fracture of the tip of the exostosis had occurred causing acute pain and haemorrhage into the bursa. 3. Although it appeared that it might be a pseudo-joint cavity the histological appearance indicated that it was a bursa caused by friction between the exostosis and the surrounding soft tissue


The Journal of Bone & Joint Surgery British Volume
Vol. 70-B, Issue 1 | Pages 49 - 51
1 Jan 1988
Echeverri A Shelley P Wroblewski B

We report a retrospective review of 127 low friction arthroplasties carried out for the failure of a previous hip operation. After an average follow-up of 10.4 years, 20% of cases required further revision, over half of them for deep infection. We estimated from the radiographic appearances that eventual failure by loosening was probable in 58% on the femoral side and 56% on the acetabular side of the arthroplasties


The Journal of Bone & Joint Surgery British Volume
Vol. 67-B, Issue 5 | Pages 736 - 740
1 Nov 1985
Wroblewski B Shelley P

A method of performing a biplanar intracapsular trochanteric osteotomy with a Gigli saw was designed and tested prospectively in 431 cases of Charnley low friction arthroplasty. Three methods of trochanteric reattachment were tested, and a double cross-over wire with a compression spring was best; this method was successful in 222 out of 226 patients (98.2%), of which half were revision operations. Adduction seemed to be the main movement leading to trochanteric detachment


The Journal of Bone & Joint Surgery British Volume
Vol. 71-B, Issue 4 | Pages 624 - 628
1 Aug 1989
Loudon Older M

We compared the clinical outcome with femoral subsidence and radiographic changes in 102 patients at 9 to 13 years after low friction arthroplasty. In 92 cases with a satisfactory outcome there was an average of 2.3 radiological signs and mean subsidence of less than 5 mm. An unsatisfactory outcome was associated with 3.4 signs per film and with subsidence of more than 5 mm. The radiological signs we describe are often the hallmark of successful load transmission, but if they increase with time or are associated with subsidence of over 5 mm, then clinical failure is likely. Fracture of the cement tip is associated with increased subsidence and adversely affects the long-term clinical outcome


The Journal of Bone & Joint Surgery British Volume
Vol. 70-B, Issue 4 | Pages 596 - 599
1 Aug 1988
Pacheco V Shelley P Wroblewski B

A retrospective review of 72 cases of Charnley low friction arthroplasty revised for stem loosening, has identified a number of "at risk" factors. These were: previous hip surgery and, in radiographs taken at one year, demarcation of the distal cement and fracture of the cement near the tip of the stem. Separation of the back of the stem from the cement, as an isolated feature, was not considered significant. Endosteal cavitation of the femoral shaft, rare in the first year, indicated loosening of some duration. Patients whose radiographs show the "at risk" changes, should be followed-up indefinitely in order to plan timely revision and avoid gross loss of the femoral bone stock


The Journal of Bone & Joint Surgery British Volume
Vol. 70-B, Issue 2 | Pages 277 - 279
1 Mar 1988
Letts M Kaylor D Gouw G

The use of prolonged halo stabilisation in a child is increasingly indicated for trauma and congenital instability of the cervical spine, but complications of pin fixation in this age group are frequent. We have analysed four aspects of the mechanics of the halo pin: the forces applied by each of six surgeons was shown to vary widely, penetration of the inner table occurred relatively easily, friction at the pin-halo interface influenced forces, and the skull thickness measured by CT scan varied from 1.1 mm to 4.3 mm in children under six years of age. We recommend CT scanning of the skull before elective halo application in young children to ascertain the safest pin sites


The Journal of Bone & Joint Surgery British Volume
Vol. 68-B, Issue 3 | Pages 483 - 488
1 May 1986
Sakurai M Miyasaka Y

Thickening of the fibrous element of a peripheral nerve may be caused by repeated friction, traction, constriction, ischaemia or partial rupture. The sequel may be a conduction disorder and a clinical condition such as an entrapment neuropathy or a tardy nerve palsy. Neural fibrosis is typically associated with a pseudoneuroma in continuity which has resulted from scarring and adhesions around the nerve as well as proliferation of the fibrous element within the nerve; the fibrosis may be classified as extraneural, intraneural or dispersive. We report 17 cases treated by external neurolysis, with 14 satisfactory results, and 42 patients treated by internal neurolysis with success in 37. Seven of the eight failures were in cases of dispersive fibrosis. A technique of internal neurolysis is described


The Journal of Bone & Joint Surgery British Volume
Vol. 50-B, Issue 3 | Pages 570 - 577
1 Aug 1968
Berkin CR

1. Twelve patients with suspected pseudarthroses in previous spine grafts were subjected to operation. 2. Correction of the curves was produced by Harrington's instrumentation at the sites of pseudarthroses, osteotomies and fractures. 3. A significant increase in stature was produced in all the patients. 4. Harrington's hooks were safely inserted into cavities cut in intact portions of previous grafts. 5. Most patients were able to leave hospital in less than four weeks. 6. Deliberate bending of the Harrington rods has not led to subsequent fracture of the rod. 7. A sliding scoliosis fusion frame is described. It reduces friction during correction and it reduces haemorrhage during surgery


The Journal of Bone & Joint Surgery British Volume
Vol. 82-B, Issue 2 | Pages 297 - 303
1 Mar 2000
Ramaniraka NA Rakotomanana LR Leyvraz P

After cemented total hip arthroplasty (THA) there may be failure at either the cement-stem or the cement-bone interface. This results from the occurrence of abnormally high shear and compressive stresses within the cement and excessive relative micromovement. We therefore evaluated micromovement and stress at the cement-bone and cement-stem interfaces for a titanium and a chromium-cobalt stem. The behaviour of both implants was similar and no substantial differences were found in the size and distribution of micromovement on either interface with respect to the stiffness of the stem. Micromovement was minimal with a cement mantle 3 to 4 mm thick but then increased with greater thickness of the cement. Abnormally high micromovement occurred when the cement was thinner than 2 mm and the stem was made of titanium. The relative decrease in surface roughness augmented slipping but decreased debonding at the cement-bone interface. Shear stress at this site did not vary significantly for the different coefficients of cement-bone friction while compressive and hoop stresses within the cement increased slightly


The Journal of Bone & Joint Surgery British Volume
Vol. 42-B, Issue 2 | Pages 360 - 366
1 May 1960
Nisbet NW

1. Certain macroscopical and microscopical features of the tendo calcaneus of the rabbit are described and illustrated, and the vascularisation as revealed by Spalteholz clearing is presented. 2. The vessels of the epitenon are chiefly derived from proximal and distal sources. 3. The vessels of the paratenon are derived from the main arteries of the leg. 4. The two vascular systems are largely independent of each other except along one edge of the tendon by way of a mesotenon. 5. The paratenon, epitenon and mesotenon and the related vessels are comparable to those found in tendons with synovial sheaths. By inference and from evidence obtained by dissection on the living human subject it is suggested that the arrangements are similar in the human tendo calcaneus. 6. Considerable friction develops on movement between the surfaces of the paratenon and epitenon. This might be significant in pathological states of the human tendon


The Bone & Joint Journal
Vol. 102-B, Issue 7 | Pages 832 - 837
1 Jul 2020
Dover C Kuiper JH Craig P Shaylor P

Aims

We have previously demonstrated raised cobalt and chromium levels in patients with larger diameter femoral heads, following metal-on-polyethylene uncemented total hip arthroplasty. Further data have been collected, to see whether these associations have altered with time and to determine the long-term implications for these patients and our practice.

Methods

Patients from our previous study who underwent Trident-Accolade primary total hip arthroplasties using a metal-on-polyethylene bearing in 2009 were reviewed. Patients were invited to have their cobalt and chromium levels retested, and were provided an Oxford Hip Score. Serum ion levels were then compared between groups (28 mm, 36 mm, and 40 mm heads) and over time.


The Journal of Bone & Joint Surgery British Volume
Vol. 86-B, Issue 3 | Pages 406 - 412
1 Apr 2004
Deshmukh SC Kumar D Mathur K Thomas B

We reviewed 13 patients with a complex fracture-dislocation of the proximal interphalangeal joint of a finger and one patient with a complex fracture-dislocation of the interphalangeal joint of thumb. We had treated these injuries using a pins and rubbers traction system which had been modified to avoid friction of the pins against the bone during mobilisation of the joint in order to minimise the risk of osteolysis. A Michigan hand outcome questionnaire was used for subjective assessment. The active range of movement (AROM) of the proximal and distal interphalangeal joints and the grip strength were used for objective assessment. The mean follow-up was 34 months (12 to 49). The mean normalised Michigan hand outcome score was 84. The mean AROM of the proximal interphalangeal joint was 85° and that of the distal interphalangeal joint 48°. The mean grip strength was 92% of the uninvolved hand. Twelve patients have returned to their original occupations. There has been no radiological osteolysis or clinical osteomyelitis. This modified traction system has given acceptable results with a low rate of complications. It is light, cheap, effective and easy to apply


The Bone & Joint Journal
Vol. 102-B, Issue 7 | Pages 852 - 860
1 Jul 2020
Zamora T Garbuz DS Greidanus NV Masri BA

Aims

Our objective is to describe our early and mid-term results with the use of a new simple primary knee prosthesis as an articulating spacer in planned two-stage management for infected knee arthroplasty. As a second objective, we compared outcomes between the group with a retained first stage and those with a complete two-stage revision.

Methods

We included 47 patients (48 knees) with positive criteria for infection, with a minimum two-year follow-up, in which a two-stage approach with an articulating spacer with new implants was used. Patients with infection control, and a stable and functional knee were allowed to retain the initial first-stage components. Outcomes recorded included: infection control rate, reoperations, final range of motion (ROM), and quality of life assessment (QoL) including Western Ontario and McMaster Universities osteoarthritis index, Knee Injury and Osteoarthritis Outcome Score, Oxford Knee Score, 12-Item Short-Form Health Survey questionnaire, and University of California Los Angeles (UCLA) activity score and satisfaction score. These outcomes were evaluated and compared to additional cohorts of patients with retained first-stage interventions and those with a complete two-stage revision. Mean follow-up was 3.7 years (2.0 to 6.5).


The Journal of Bone & Joint Surgery British Volume
Vol. 46-B, Issue 2 | Pages 179 - 190
1 May 1964
Outerbridge RE

1 . Current theories of the etiology of chondromalacia patellae do not explain satisfactorily either its great frequency or its common site of origin on the medial patellar facet. 2. The etiology can be more logically explained by the presence of a ridge on the upper anterior border of the cartilage of the medial femoral condyle, in most knees. This ridge, consisting of cartilage, or cartilage and bone, varies considerably in height and, in normal knee joint movement, causes considerable friction on the medial patellar facet. 3. The degenerative changes were found to be greater in the presence of the larger ridges, and–because of longer wear and tear–in the older patients. 4. This study indicates that chondromalacia was more severe in women than in men, and in patients overweight. Although the activity of the individual and the power of the quadriceps mechanism must play an extremely important part in this condition, it was not possible to assess this. 5. Two factors previously considered to be important in the etiology of this condition, namely, the length of the patellar tendon and Wiberg's Type III patellar shape, have not been confirmed in this study. 6. Resulting from the present investigation certain precautions are suggested in rehabilitation after operations on the knee, and a surgical method for discouraging the progress of this common, and sometimes disabling, condition has been devised


The Journal of Bone & Joint Surgery British Volume
Vol. 77-B, Issue 4 | Pages 528 - 533
1 Jul 1995
Gillespie W Pekarsky B O'Connell D

Cost is a factor in the choice of prosthetic components in joint replacement. For a given performance, the least expensive components are the most cost-effective. When evaluating a new prosthesis with an unknown outcome, the use of an economic model allows estimation of potential cost-effectiveness. We used published data for the survival of cemented total hip replacements from Sweden, and cost and demographic information from New South Wales, Australia, in such a model. In young active total hip recipients a new prosthetic design which offered a 90% improvement in survivorship over 15 years and a 15% reduction in the cost of revision surgery, could be sold at a price of 2 to 2.5 times that of conventional cemented components such as the Charnley Low Friction Arthroplasty and still be cost-effective. Using more likely estimates of the improved performance of new technology, however, the upper limit of cost-effectiveness is an increase of 1.5 to 1. Only a very small increase in the cost of a prosthesis could ever be justified for older patients of either sex. Most of the potential benefits of a better level of survivorship appear towards the end of the 15-year period. The results of modelling may be incorporated in clinical trial design. Given the known performance of some well-established and relatively inexpensive designs of prostheses, very large randomised studies would be required to prove an improvement in performance


The Bone & Joint Journal
Vol. 102-B, Issue 10 | Pages 1405 - 1411
3 Oct 2020
Martynov I Klink T Slowik V Stich R Zimmermann P Engel C Lacher M Boehm R

Aims

This exploratory randomized controlled trial (RCT) aimed to determine the splint-related outcomes when using the novel biodegradable wood-composite splint (Woodcast) compared to standard synthetic fibreglass (Dynacast) for the immobilization of undisplaced upper limb fractures in children.

Methods

An exploratory RCT was performed at a tertiary paediatric referral hospital between 1 June 2018 and 30 September 2019. The intention-to-treat population consisted of 170 patients (mean age 8.42 years (SD 3.42); Woodcast (WCG), n = 84, 57 male (67.9%); Dynacast (DNG), n = 86, 58 male (67.4%)). Patients with undisplaced upper limb fractures were randomly assigned to WCG or DNG treatment groups. Primary outcome was the stress stability of the splint material, defined as absence of any deformations or fractures within the splint during study period. Secondary outcomes included patient satisfaction and medical staff opinion. Additionally, biomechanical and chemical analysis of the splint samples was carried out.


The Bone & Joint Journal
Vol. 102-B, Issue 8 | Pages 1025 - 1032
1 Aug 2020
Hampton M Mansoor J Getty J Sutton PM

Aims

Total knee arthroplasty is an established treatment for knee osteoarthritis with excellent long-term results, but there remains controversy about the role of uncemented prostheses. We present the long-term results of a randomized trial comparing an uncemented tantalum metal tibial component with a conventional cemented component of the same implant design.

Methods

Patients under the age of 70 years with symptomatic osteoarthritis of the knee were randomized to receive either an uncemented tantalum metal tibial monoblock component or a standard cemented modular component. The mean age at time of recruitment to the study was 63 years (50 to 70), 46 (51.1%) knees were in male patients, and the mean body mass index was 30.4 kg/m2 (21 to 36). The same cruciate retaining total knee system was used in both groups. All patients received an uncemented femoral component and no patients had their patella resurfaced. Patient outcomes were assessed preoperatively and postoperatively using the modified Oxford Knee Score, Knee Society Score, and 12-Item Short-Form Health Survey questionnaire (SF-12) score. Radiographs were analyzed using the American Knee Society Radiograph Evaluation score. Operative complications, reoperations, or revision surgery were recorded. A total of 90 knees were randomized and at last review 77 knees were assessed. In all, 11 patients had died and two were lost to follow-up.


The Bone & Joint Journal
Vol. 102-B, Issue 7 Supple B | Pages 33 - 40
1 Jul 2020
Gustafson JA Pourzal R Levine BR Jacobs JJ Lundberg HJ

Aims

The aim of this study was to develop a novel computational model for estimating head/stem taper mechanics during different simulated assembly conditions.

Methods

Finite element models of generic cobalt-chromium (CoCr) heads on a titanium stem taper were developed and driven using dynamic assembly loads collected from clinicians. To verify contact mechanics at the taper interface, comparisons of deformed microgroove characteristics (height and width of microgrooves) were made between model estimates with those measured from five retrieved implants. Additionally, these models were used to assess the role of assembly technique—one-hit versus three-hits—on the taper interlock mechanical behaviour.


The Bone & Joint Journal
Vol. 102-B, Issue 5 | Pages 573 - 579
1 May 2020
Krueger DR Guenther K Deml MC Perka C

Aims

We evaluated a large database with mechanical failure of a single uncemented modular femoral component, used in revision hip arthroplasty, as the end point and compared them to a control group treated with the same implant. Patient- and implant-specific risk factors for implant failure were analyzed.

Methods

All cases of a fractured uncemented modular revision femoral component from one manufacturer until April 2017 were identified and the total number of implants sold until April 2017 was used to calculate the fracture rate. The manufacturer provided data on patient demographics, time to failure, and implant details for all notified fractured devices. Patient- and implant-specific risk factors were evaluated using a logistic regression model with multiple imputations and compared to data from a previously published reference group, where no fractures had been observed. The results of a retrieval analysis of the fractured implants, performed by the manufacturer, were available for evaluation.