Aims. The Birmingham Hip Resurfacing (BHR) arthroplasty has been used as a surgical treatment of coxarthrosis since 1997. We present 20-year results of 234 consecutive BHRs performed in our unit. Methods. Between 1999 and 2001, there were 217 patients: 142 males (65.4%), mean age 52 years (18 to 68) who had 234 implants (17 bilateral). They had patient-reported outcome measures collected, imaging (radiograph and ultrasound), and serum metal ion assessment. Survivorship analysis was performed using Kaplan-Meier estimates. Revision for any cause was considered as an endpoint for the analysis. Results. Mean follow-up was 20.9 years (19.3 to 22.4). Registry data revealed that 19 hips (8.1%) had been revised and 26 patients (12%) had died from causes unrelated to the
Modern metal-on-metal hip resurfacing has been
widely performed in the United Kingdom for over a decade. However,
the literature reports conflicting views of the benefits: excellent
medium- to long-term results with some brands in specific subgroups,
but high failure rates and local soft-tissue reactions in others.
The National Joint Registry for England and Wales (NJR) has collected
data on all hip resurfacings performed since 2003. This retrospective
cohort study recorded survival time to revision from a resurfacing
procedure, exploring risk factors independently associated with
failure. All patients with a primary diagnosis of osteoarthritis
who underwent resurfacing between 2003 and 2010 were included in
the analyses. Cox’s proportional hazard models were used to analyse
the extent to which the risk of revision was related to patient,
surgeon and implant covariates. A total of 27 971 hip resurfacings were performed during the
study period, of which 1003 (3.59%) underwent revision surgery.
In the final adjusted model, we found that women were at greater
risk of revision than men (hazard ratio (HR) = 1.30, p = 0.007),
but the risk of revision was independent of age. Of the implant-specific
predictors, five brands had a significantly greater risk of revision
than the Birmingham Hip Resurfacing (BHR) (ASR: HR = 2.82, p <
0.001,
Conserve: HR = 2.03, p <
0.001, Cormet: HR = 1.43, p = 0.001,
Durom: HR = 1.67, p <
0.001, Recap: HR = 1.58, p = 0.007). Smaller
femoral head components were also significantly more likely to require
revision (≤ 44 mm: HR = 2.14, p <
0.001, 45 to 47 mm: HR = 1.48,
p = 0.001) than medium or large heads, as were operations performed
by low-volume surgeons (HR = 1.36, p <
0.001). Once these influences
had been removed, in 4873 male patients <
60 years old undergoing
resurfacing with a
Recent events have highlighted the importance
of implant design for survival and wear-related complications following
metal-on-metal hip resurfacing arthroplasty. The mid-term survival
of the most widely used implant, the Birmingham Hip Resurfacing
(BHR), has been described by its designers. The aim of this study
was to report the ten-year survival and patient-reported functional
outcome of the
Aims. The Birmingham Hip Resurfacing (BHR) was introduced in 1997 to address the needs of young active patients using a historically proven large-diameter metal-on-metal (MoM) bearing. A single designer surgeon’s consecutive series of 130 patients (144 hips) was previously reported at five and ten years, reporting three and ten failures, respectively. The aim of this study was to extend the follow-up of this original cohort at 25 years. Methods. The study extends the reporting on the first consecutive 144 resurfacing procedures in 130 patients for all indications. All operations were undertaken between August 1997 and May 1998. The mean age at operation was 52.1 years (SD 9.93; 17 to 76), and included 37 female patients (28.5%). Failure was defined as revision of either component for any reason. Kaplan-Meier survival analysis was performed. Routine follow-up with serum metal ion levels, radiographs, and Oxford Hip Scores (OHSs) was undertaken. Results. Overall implant survival was 83.50% (95% confidence interval (CI) 0.79 to 0.90) at 25 years, and the number at risk was 79. Survival in male patients at 25 years was 89.5% (95% CI 0.83 to 0.96) compared to 66.9% for female patients (95% CI 0.51 to 0.83). Ten additional failures occurred in the period of ten to 25 years. These involved an adverse reaction to metal debris in four patients, a periprosthetic femoral neck fracture affecting five patients, and aseptic loosening in one patient. The median chromium levels were 49.50 nmol/l (interquartile range (IQR) 34 to 70), and the median cobalt serum levels were 42 nmol/l (IQR 24.50 to 71.25). The median OHS at last follow-up was 35 (IQR 10 to 48). During the 25-year study period, 29 patients died. Patient survival at 25 years was 75.10% (95% CI 0.67 to 0.83). Conclusion. This study demonstrates that MoM hip resurfacing using the
Aims. The first aim of this study was to evaluate whether preoperative renal function is associated with postoperative changes in whole blood levels of metal ions in patients who have undergone a Birmingham Hip Resurfacing (BHR) arthroplasty with a metal-on-metal bearing. The second aim was to evaluate whether exposure to increased cobalt (Co) and chromium (Cr) levels for ten years adversely affected renal function. Methods. As part of a multicentre, prospective post-approval study, whole blood samples were sent to a single specialized laboratory to determine Co and Cr levels, and the estimated glomerular filtration rate (eGFR). The study included patients with 117 unrevised unilateral BHRs. There were 36 females (31%). The mean age of the patients at the time of surgery of 51.3 years (SD 6.5), and they all had preoperative one-, four-, five-, and ten-year laboratory data. The mean follow-up was 10.1 years (SD 0.2). Results. Median Co levels at one year postoperatively increased significantly compared with the preoperative values, by a factor of 9.7, from 0.13 to 1.26 ppb (p < 0.001), and the median Cr levels increased significantly by a factor of 2.5, from 0.60 to 1.50 ppb (p < 0.001). Lower preoperative eGFRs were associated with significantly larger increases in Co at one year compared with the preoperative levels (ρ = -0.26; p = 0.005), but there was no relationship between preoperative eGFRs and changes in Cr at one year (ρ = -0.13; p = 0.153). Metal levels remained relatively constant with the passage of time, with a median ten-year value of 1.12 ppb for Co and 1.29 ppb for Cr. There was no significant relationship between the Co and Cr levels at ten-year follow-up and the change in eGFR from the preoperative level to that at ten years (ρ = -0.02; p = 0.827 for Co; ρ = -0.008; p = 0.933 for Cr). Conclusion. Although patients with lower preoperative eGFRs tended to have larger increases in Co levels at one year, increased metal levels for patients who underwent unilateral
Aims. The aim of this study was to report the implant survival and patient-reported outcome measures (PROMs) in a consecutive series of patients aged less than 50 years at the time of arthroplasty using the Birmingham Hip Resurfacing system (BHR), with a minimum follow-up of ten years. Patients and Methods. A total of 226 patients with osteoarthritis of the hip, who underwent
This study compared the demographic, clinical and patient-reported outcomes after total hip replacement (THR) and Birmingham Hip Resurfacing (BHR) carried out by a single surgeon. Patients completed a questionnaire that included the WOMAC, SF-36 scores and comorbid medical conditions. Data were collected before operation and one year after. The outcome scores were adjusted for age, gender, comorbid conditions and, at one year, for the pre-operative scores. There were 214 patients with a THR and 132 with a
The Articular Surface Replacement (ASR) hip resurfacing arthroplasty has a failure rate of 12.0% at five years, compared with 4.3% for the Birmingham Hip Resurfacing (BHR). We analysed 66 ASR and 64
There have been no large comparative studies of the blood levels of metal ions after implantation of commercially available hip resurfacing devices which have taken into account the effects of femoral size and inclination and anteversion of the acetabular component. We present the results in 90 patients with unilateral articular surface replacement (ASR) hip resurfacings (mean time to blood sampling 26 months) and 70 patients with unilateral Birmingham Hip Resurfacing (BHR) implants (mean time 47 months). The whole blood and serum chromium (Cr) and cobalt (Co) concentrations were inversely related to the size of the femoral component in both groups (p <
0.05). Cr and Co were more strongly influenced by the position of the acetabular component in the case of the ASR, with an increase in metal ions observed at inclinations >
45° and anteversion angles of <
10° and >
20°. These levels were only increased in the
We compared the medium-term clinical and radiological results of hybrid total hip replacement (THR) with metal-on-metal Birmingham hip resurfacing (BHR) in two groups of 54 young patients matched for age, gender, body mass index and pre-operative levels of activity. The clinical outcome was assessed by the University of California, Los Angeles (UCLA) activity score, the Oxford Hip Score (OHS) and the EuroQol scores. Radiologically, all hips were assessed for migration and osteolysis, the hybrid THRs for polyethylene wear and the BHRs for a pedestal sign. The mean follow-up of the patients with a hybrid THR was ten years and for those with a
The Birmingham hip resurfacing (BHR) arthroplasty is a metal-on-metal prosthesis for which no medium- or long-term results have been published. Despite this, it is increasing in popularity as an alternative to stemmed prostheses for younger patients. Since the fixation of the socket is conventional, the major concern is long-term failure of the femoral component. This can be predicted by the use of roentgen stereophotogrammetric analysis (RSA). We have therefore undertaken such a study of the
We report the long-term survival and functional
outcome of the Birmingham Hip Resurfacing (BHR) in patients aged <
50 years at operation, and explore the factors affecting survival.
Between 1997 and 2006, a total of 447 BHRs were implanted in 393
patients (mean age 41.5 years (14.9 to 49.9)) by one designing surgeon.
The mean follow-up was 10.1 years (5.2 to 14.7), with no loss to
follow-up. In all, 16 hips (3.6%) in 15 patients were revised, giving
an overall cumulative survival of 96.3% (95% confidence interval
(CI) 93.7 to 98.3) at ten years and 94.1% (95% CI 84.9 to 97.3)
at 14 years. Using aseptic revision as the endpoint, the survival
for men with primary osteoarthritis (n = 195) was 100% (95% CI 100
to 100) at both ten years and 14 years, and in women with primary
osteoarthritis (n = 109) it was 96.1% (95% CI 90.1 to 99.9) at ten
years and 91.2% (95% CI 68.6 to 98.7) at 14 years. Female gender
(p = 0.047) and decreasing femoral head size (p = 0.044) were significantly
associated with an increased risk of revision. The median Oxford
hip score (OHS, modified as a percentage with 100% indicating worst
outcome) at last follow-up was 4.2% (46 of 48; interquartile range
(IQR) 0% to 24%) and the median University of California, Los Angeles
(UCLA) score was 6.0 (IQR 5 to 8). Men had significantly better
OHS (p = 0.02) and UCLA scores (p = 0.01) than women. The
We report a 12- to 15-year implant survival assessment
of a prospective single-surgeon series of Birmingham Hip Resurfacings
(BHRs). The earliest 1000 consecutive BHRs including 288 women (335
hips) and 598 men (665 hips) of all ages and diagnoses with no exclusions
were prospectively followed-up with postal questionnaires, of whom
the first 402 BHRs (350 patients) also had clinical and radiological
review. Mean follow-up was 13.7 years (12.3 to 15.3). In total, 59 patients
(68 hips) died 0.7 to 12.6 years following surgery from unrelated
causes. There were 38 revisions, 0.1 to 13.9 years (median 8.7)
following operation, including 17 femoral failures (1.7%) and seven
each of infections, soft-tissue reactions and other causes. With
revision for any reason as the end-point Kaplan–Meier survival analysis
showed 97.4% (95% confidence interval (CI) 96.9 to 97.9) and 95.8%
(95% CI 95.1 to 96.5) survival at ten and 15 years, respectively.
Radiological assessment showed 11 (3.5%) femoral and 13 (4.1%) acetabular
radiolucencies which were not deemed failures and one radiological
femoral failure (0.3%). . Our study shows that the performance of the
Hip resurfacing is being performed more frequently in the United Kingdom. The possible benefits include more accurate restoration of leg length, femoral offset and femoral anteversion than occurs after total hip arthroplasty (THA). We compared anteroposterior radiographs from 26 patients who had undergone hybrid THA (uncemented cup/cemented stem), with 28 who had undergone Birmingham Hip Resurfacing arthroplasty (BHR). We measured the femoral offset, femoral length, acetabular offset and acetabular height with reference to the normal contralateral hip. The data were analysed by paired t-tests. There was a significant reduction in femoral offset (p = 0.0004) and increase in length (p = 0.001) in the
We report the findings of an independent review
of 230 consecutive Birmingham hip resurfacings (BHRs) in 213 patients
(230 hips) at a mean follow-up of 10.4 years (9.6 to 11.7). A total
of 11 hips underwent revision; six patients (six hips) died from
unrelated causes; and 13 patients (16 hips) were lost to follow-up.
The survival rate for the whole cohort was 94.5% (95% confidence
interval (CI) 90.1 to 96.9). The survival rate in women was 89.1%
(95% CI 79.2 to 94.4) and in men was 97.5% (95% CI 92.4 to 99.2).
Women were 1.4 times more likely to suffer failure than men. For
each millimetre increase in component size there was a 19% lower
chance of a failure. The mean Oxford hip score was 45.0 (median
47.0, 28 to 48); mean University of California, Los Angeles activity
score was 7.4 (median 8.0, 3 to 9); mean patient satisfaction score
was 1.4 (median 1.0, 0 to 9). A total of eight hips had lysis in
the femoral neck and two hips had acetabular lysis. One hip had
progressive radiological changes around the peg of the femoral component.
There was no evidence of progressive neck narrowing between five
and ten years. Our results confirm that
We investigated the effect of the Birmingham hip resurfacing (BHR) arthroplasty on the bone mineral density (BMD) of the femur. A comparative study was carried out on 26 hips in 25 patients. Group A consisted of 13 patients (13 hips) who had undergone resurfacing hip arthroplasty with the
To find out if there is an inverse association between estimated glomerular filtration rate (eGFR) and whole blood cobalt (Co) and chromium (Cr) levels in patients with metal-on-metal (MoM) hip arthroplasties and renal insufficiency, suggesting that renal insufficiency could cause accumulation of Co and Cr in blood. Out of 2,520 patients with 3,013 MoM hip arthroplasties, we identified 1,244 patients with whole blood Co, Cr, and creatinine measured within no more than a one-year interval. We analyzed the correlation of blood metal ion levels and eGFR to identify a potential trend of accumulating Co or Cr with decreasing eGFR.Aims
Methods
It is not known whether change in patient-reported outcome measures (PROMs) over time can be predicted by factors present at surgery, or early follow-up. The aim of this study was to identify factors associated with changes in PROM status between two-year evaluation and medium-term follow-up. Patients undergoing Birmingham Hip Resurfacing completed the Veteran’s Rand 36 (VR-36), modified Harris Hip Score (mHHS), Tegner Activity Score, and Western Ontario and McMaster Universities Osteoarthritis Index (WOMAC) at two years and a minimum of three years. A change in score was assessed against minimal clinically important difference (MCID) and patient-acceptable symptom state (PASS) thresholds. Binary logistic regression was used to assess the relationship between patient factors and deterioration in PASS status between follow-ups.Aims
Patients and Methods
The aim of this study was to assess the functional gain achieved following hip resurfacing arthroplasty (HRA). A total of 28 patients (23 male, five female; mean age, 56 years (25 to 73)) awaiting Birmingham HRA volunteered for this prospective gait study, with an age-matched control group of 26 healthy adults (16 male, ten female; mean age, 56 years (33 to 84)). The Oxford Hip Score (OHS) and gait analysis using an instrumented treadmill were used preoperatively and more than two years postoperatively to measure the functional change attributable to the intervention.Aims
Patients and Methods
The purpose of this study is to examine six types of bearing surfaces implanted at a single institution over three decades to determine whether the reasons for revision vary among the groups and how long it takes to identify differences in survival. We considered six cohorts that included a total of 1,707 primary hips done between 1982 and 2010. These included 223 conventional polyethylene sterilized with γ irradiation in air (CPE-GA), 114 conventional polyethylene sterilized with gas plasma (CPE-GP), 116 crosslinked polyethylene (XLPE), 1,083 metal-on-metal (MOM), 90 ceramic-on-ceramic (COC), and 81 surface arthroplasties (SAs). With the exception of the COC, all other groups used cobalt-chromium (CoCr) femoral heads. The mean follow-up was 10 (0.008 to 35) years. Descriptive statistics with revisions per 100 component years (re/100 yr) and survival analysis with revision for any reason as the endpoint were used to compare bearing surfaces.Aims
Methods
We investigated whether blood metal ion levels could effectively
identify patients with bilateral Birmingham Hip Resurfacing (BHR)
implants who have adverse reactions to metal debris (ARMD). Metal ion levels in whole blood were measured in 185 patients
with bilateral BHRs. Patients were divided into those with ARMD
who either had undergone a revision for ARMD or had ARMD on imaging
(n = 30), and those without ARMD (n = 155). Receiver operating characteristic
analysis was used to determine the optimal thresholds of blood metal
ion levels for identifying patients with ARMD.Aims
Patients and Methods
The aim of this study was to evaluate the ten-year
clinical and functional outcome of hip resurfacing and to compare it
with that of cementless hip arthroplasty in patients under the age
of 55 years. Between 1999 and 2002, 80 patients were enrolled into the study:
24 were randomised (11 to hip resurfacing, 13 to total hip arthroplasty),
18 refused hip resurfacing and chose cementless total hip arthroplasty
with a 32 mm bearing, and 38 insisted on resurfacing. The mean follow-up
for all patients was 12.1 years (10 to 14). Patients were assessed clinically and radiologically at one year,
five years and ten years. Outcome measures included EuroQol EQ5D,
Oxford, Harris hip, University of California Los Angeles and University
College Hospital functional scores. No differences were seen between the two groups in the Oxford
or Harris hip scores or in the quality of life scores. Despite a
similar aspiration to activity pre-operatively, a higher proportion
of patients with a hip resurfacing were running and involved in
sport and heavy manual labour after ten years. We found significantly higher function scores in patients who
had undergone hip resurfacing than in those with a cementless hip
arthroplasty at ten years. This suggests a functional advantage
for hip resurfacing. There were no other attendant problems. Cite this article:
We wished to investigate the influence of metal debris exposure
on the subsequent immune response and resulting soft-tissue injury
following metal-on-metal (MoM) hip arthroplasty. Some reports have
suggested that debris generated from the head-neck taper junction
is more destructive than equivalent doses from metal bearing surfaces. We investigated the influence of the source and volume of metal
debris on chromium (Cr) and cobalt (Co) concentrations in corresponding
blood and hip synovial fluid samples and the observed agglomerated
particle sizes in excised tissues using multiple regression analysis
of prospectively collected data. A total of 199 explanted MoM hips
(177 patients; 132 hips female) were analysed to determine rates
of volumetric wear at the bearing surfaces and taper junctions. Aims
Patients and Methods
Hip resurfacing has been proposed as an alternative
to traditional total hip arthroplasty in young, active patients.
Much has been learned following the introduction of metal-on-metal resurfacing
devices in the 1990s. The triad of a well-designed device, implanted
accurately, in the correct patient has never been more critical
than with these implants. Following Food and Drug Administration approval in 2006, we studied
the safety and effectiveness of one hip resurfacing device (Birmingham
Hip Resurfacing) at our hospital in a large, single-surgeon series.
We report our early to mid–term results in 1333 cases followed for
a mean of 4.3 years (2 to 5.7) using a prospective, observational
registry. The mean patient age was 53.1 years (12 to 84); 70% were
male and 91% had osteoarthritis. Complications were few, including
no dislocations, no femoral component loosening, two femoral neck
fractures (0.15%), one socket loosening (0.08%), three deep infections
(0.23%), and three cases of metallosis (0.23%). There were no destructive
pseudotumours. Overall survivorship at up to 5.7 years was 99.2%. Aseptic survivorship
in males under the age of 50 was 100%. We believe this is the largest
United States series of a single surgeon using a single resurfacing
system. Cite this article:
Surgeons have commonly used modular femoral heads and stems from
different manufacturers, although this is not recommended by orthopaedic
companies due to the different manufacturing processes. We compared the rate of corrosion and rate of wear at the trunnion/head
taper junction in two groups of retrieved hips; those with mixed
manufacturers (MM) and those from the same manufacturer (SM). We identified 151 retrieved hips with large-diameter cobalt-chromium
heads; 51 of two designs that had been paired with stems from different
manufacturers (MM) and 100 of seven designs paired with stems from
the same manufacturer (SM). We determined the severity of corrosion
with the Goldberg corrosion score and the volume of material loss
at the head/stem junction. We used multivariable statistical analysis
to determine if there was a significant difference between the two
groups. Aims
Materials and Methods
This study presents the long-term survivorship, risk factors for prosthesis survival, and an assessment of the long-term effects of changes in surgical technique in a large series of patients treated by metal-on-metal (MoM) hip resurfacing arthroplasty (HRA). Between November 1996 and January 2012, 1074 patients (1321 hips) underwent HRA using the Conserve Plus Hip Resurfacing System. There were 787 men (73%) and 287 women (27%) with a mean age of 51 years (14 to 83). The underlying pathology was osteoarthritis (OA) in 1003 (75.9%), developmental dysplasia of the hip (DDH) in 136 (10.3%), avascular necrosis in 98 (7.4%), and other conditions, including inflammatory arthritis, in 84 (6.4%).Aims
Patients and Methods
The aims of this study were to compare the diagnostic test characteristics
of ultrasound alone, metal artefact reduction sequence MRI (MARS-MRI)
alone, and ultrasound combined with MARS-MRI for identifying intra-operative
pseudotumours in metal-on-metal hip resurfacing (MoMHR) patients
undergoing revision surgery. This retrospective diagnostic accuracy study involved 39 patients
(40 MoMHRs). The time between imaging modalities was a mean of 14.6
days (0 to 90), with imaging performed at a mean of 5.3 months (0.06
to 12) before revision. The prevalence of intra-operative pseudotumours
was 82.5% (n = 33).Aims
Methods
A retrospective study was conducted to investigate
the changes in metal ion levels in a consecutive series of Birmingham
Hip Resurfacings (BHRs) at a minimum ten-year follow-up. We reviewed
250 BHRs implanted in 232 patients between 1998 and 2001. Implant
survival, clinical outcome (Harris hip score), radiographs and serum chromium
(Cr) and cobalt (Co) ion levels were assessed. Of 232 patients, 18 were dead (five bilateral BHRs), 15 lost
to follow-up and ten had been revised. The remaining 202 BHRs in
190 patients (136 men and 54 women; mean age at surgery 50.5 years
(17 to 76)) were evaluated at a minimum follow-up of ten years (mean
10.8 years (10 to 13.6)). The overall implant survival at 13.2 years
was 92.4% (95% confidence interval 90.8 to 94.0). The mean Harris
hip score was 97.7 (median 100; 65 to 100). Median and mean ion
levels were low for unilateral resurfacings (Cr: median 1.3 µg/l,
mean
1.95 µg/l (<
0.5 to 16.2); Co: median 1.0 µg/l, mean 1.62 µg/l
(<
0.5 to 17.3)) and bilateral resurfacings (Cr: median 3.2 µg/l,
mean 3.46 µg/l (<
0.5 to 10.0); Co: median 2.3 µg/l, mean 2.66
µg/l (<
0.5 to 9.5)). In 80 unilateral BHRs with sequential ion
measurements, Cr and Co levels were found to decrease significantly
(p <
0.001) from the initial assessment at a median of six years
(4 to 8) to the last assessment at a median of 11 years (9 to 13),
with a mean reduction of 1.24 µg/l for Cr and 0.88 µg/l for Co.
Three female patients had a >
2.5 µg/l increase of Co ions, associated with
head sizes ≤ 50 mm, clinical symptoms and osteolysis. Overall, there
was no significant difference in change of ion levels between genders
(Cr, p = 0.845; Co, p = 0.310) or component sizes (Cr, p = 0.505;
Co, p = 0.370). Higher acetabular component inclination angles correlated
with greater change in ion levels (Cr, p = 0.013; Co, p = 0.002).
Patients with increased ion levels had lower Harris hip scores (p
= 0.038). In conclusion, in well-functioning BHRs the metal ion levels
decreased significantly at ten years. An increase >
2.5 µg/l was
associated with poor function. Cite this article:
We systematically reviewed the peer-reviewed literature to relate the survival of hybrid metal-on-metal hip resurfacing arthroplasty devices to a National Institute of Clinical Excellence (NICE) benchmark for choosing a primary total hip replacement, which is a survival rate of 90% at a follow-up of ten years. A total of 29 articles (10 621 resurfaced hips) met the inclusion criteria. The mean follow-up ranged from 0.6 to 10.5 years and the survival of the implant ranged from 84% to 100%. Of the 10 621 hips, 370 were revised (3.5%), with aseptic loosening as the most frequent mode of failure. None of the hip resurfacing arthroplasty implants used to date met the full ten-year NICE benchmark of survival. A total of 13 studies showed satisfactory survival compared with the three-year NICE benchmark.
We present the clinical results and survivorship
of consecutive 100 Birmingham Hip Resurfacings in 90 patients at
a minimum follow-up of ten years. All procedures were carried out
by an independent surgeon who commenced a prospective study in 1998.
Patients were assessed clinically using the Western Ontario and
McMaster Universities osteoarthritis index, Short-Form 36, Harris
hip score and University of California, Los Angeles activity score. Radiological
analysis was performed by an independent observer and blood metal
ion levels concentrations were measured at ten years post-operatively
in 62 patients. The median acetabular component inclination was
46.2° (34° to 59°) and anteversion 11.0° (0° to 30°). The median
chromium concentration in the unilateral group was 1.74 (0.41 to
15.23) and for the bilateral group was 2.98 (1.57 to 18.01). The
equivalent values for cobalt were 1.67 (0.54 to 20.4) and 1.88 (1.38
to 19.32). In total there were eight failures giving an overall survival
at ten years of 92% (95% confidence interval (CI) 86.7 to 97.3).
BHR in male patients had an improved survivorship of 94.6% (95%
CI 89.4 to 100) compared with females at 84.6% (95% CI 70.7 to 98.5),
but this did not reach statistical significance (p = 0.119). Four
of the nine BHRs with a bearing diameter of 42 mm failed. The overall results were consistent with data produced from other
centres in that the clinical outcome of large male patients was
extremely encouraging, whereas the survival of the smaller joints
was less satisfactory.
We sought to determine whether cobalt-chromium alloy (CoCr) femoral
stem tapers (trunnions) wear more than titanium (Ti) alloy stem
tapers (trunnions) when used in a large diameter (LD) metal-on-metal
(MoM) hip arthroplasty system. We performed explant analysis using validated methodology to
determine the volumetric material loss at the taper surfaces of
explanted LD CoCr MoM hip arthroplasties used with either a Ti alloy
(n = 28) or CoCr femoral stem (n = 21). Only 12/14 taper constructs
with a rough male taper surface and a nominal included angle close
to 5.666° were included. Multiple regression modelling was undertaken
using taper angle, taper roughness, bearing diameter (horizontal
lever arm) as independent variables. Material loss was mapped using
a coordinate measuring machine, profilometry and scanning electron
microscopy.Aims
Patients and Methods
Metal-on-metal hip resurfacing is commonly performed for osteoarthritis in young active patients. We have observed cystic or solid masses, which we have called inflammatory pseudotumours, arising around these devices. They may cause soft-tissue destruction with severe symptoms and a poor outcome after revision surgery. The aim of this study was to determine the incidence of and risk factors for pseudotumours that are serious enough to require revision surgery. Since 1999, 1419 metal-on-metal hip resurfacings have been implanted by our group in 1224 patients; 1.8% of the patients had a revision for pseudotumour. In this series the Kaplan-Meier cumulative revision rate for pseudotumour increased progressively with time. At eight years, in all patients, it was 4% (95% confidence interval (CI) 2.2 to 5.8). Factors significantly associated with an increase in revision rate were female gender (p <
0.001), age under 40 (p = 0.003), small components (p = 0.003), and dysplasia (p = 0.019), whereas implant type was not (p = 0.156). These factors were inter-related, however, and on fitting a Cox proportional hazard model only gender (p = 0.002) and age (p = 0.024) had a significant independent influence on revision rate; size nearly reached significance (p = 0.08). Subdividing the cohort according to significant factors, we found that the revision rate for pseudotumours in men was 0.5% (95% CI 0 to 1.1) at eight years wheras in women over 40 years old it was 6% (95% CI 2.3 to 10.1) at eight years and in women under 40 years it was 13.1% at six years (95% CI 0 to 27) (p <
0.001). We recommend that resurfacings are undertaken with caution in women, particularly those under 40 years of age but they remain a good option in young men. Further work is required to understand the aetiology of pseudotumours so that this complication can be avoided.
Hip resurfacing is a bone-conserving procedure with respect to proximal femoral resection, but there is debate in the literature as to whether the same holds true for the acetabulum. We have investigated whether the Birmingham hip resurfacing conserves acetabular bone. Between 1998 and 2005, 500 Birmingham hip resurfacings were performed by two surgeons. Between 1996 and 2005 they undertook 700 primary hip replacements, with an uncemented acetabular component. These patients formed the clinical material to compare acetabular component sizing. The Birmingham hip resurfacing group comprised 350 hips in men and 150 hips in women. The uncemented total hip replacement group comprised 236 hips in men and 464 hips in women. Age- and gender-matched analysis of a cohort of patients for the sizes of the acetabular components required for the two types of replacement was also undertaken. Additionally, an analysis of the sizes of the components used by each surgeon was performed. For age-matched women, the mean outside diameter of the Birmingham hip resurfacing acetabular components was 2.03 mm less than that of the acetabular components in the uncemented total hip replacements (p <
0.0001). In similarly matched men there was no significant difference (p = 0.77). A significant difference was also found between the size of acetabular components used by the two surgeons for Birmingham hip resurfacing for both men (p = 0.0015) and women (p = 0.001). In contrast, no significant difference was found between the size of acetabular components used by the two surgeons for uncemented total hip replacement in either men or women (p = 0.06 and p = 0.14, respectively). This suggests that variations in acetabular preparation also influence acetabular component size in hip resurfacing.
The effects of the method of fixation and interface conditions on the biomechanics of the femoral component of the Birmingham hip resurfacing arthroplasty were examined using a highly detailed three-dimensional computer model of the hip. Stresses and strains in the proximal femur were compared for the natural femur and for the femur resurfaced with the Birmingham hip resurfacing. A comparison of cemented
We report 17 patients (20 hips) in whom metal-on-metal resurfacing had been performed and who presented with various symptoms and a soft-tissue mass which we termed a pseudotumour. Each patient underwent plain radiography and in some, CT, MRI and ultrasonography were also performed. In addition, histological examination of available samples was undertaken. All the patients were women and their presentation was variable. The most common symptom was discomfort in the region of the hip. Other symptoms included spontaneous dislocation, nerve palsy, a noticeable mass or a rash. The common histological features were extensive necrosis and lymphocytic infiltration. To date, 13 of the 20 hips have required revision to a conventional hip replacement. Two are awaiting revision. We estimate that approximately 1% of patients who have a metal-on-metal resurfacing develop a pseudotumour within five years. The cause is unknown and is probably multifactorial. There may be a toxic reaction to an excess of particulate metal wear debris or a hypersensitivity reaction to a normal amount of metal debris. We are concerned that with time the incidence of these pseudotumours may increase. Further investigation is required to define their cause.
The recent resurgence in the use of metal-on-metal bearings has led to fresh concerns over metal wear and elevated systemic levels of metal ions. In order to establish if bearing diameter influences the release of metal ions, we compared the whole blood levels of cobalt and chromium (at one year) and the urinary cobalt and chromium output (at one to three and four to six years) following either a 50 mm or 54 mm Birmingham hip resurfacing or a 28 mm Metasul total hip replacement. The whole blood concentrations and daily output of cobalt and chromium in these time periods for both bearings were in the same range and without significant difference.
We undertook a retrospective cohort study to
determine clinical outcomes following the revision of metal-on-metal (MoM)
hip replacements for adverse reaction to metal debris (ARMD), and
to identify predictors of time to revision and outcomes following
revision. Between 1998 and 2012 a total of 64 MoM hips (mean age
at revision of 57.8 years; 46 (72%) female; 46 (72%) hip resurfacings
and 18 (28%) total hip replacements) were revised for ARMD at one specialist
centre. At a mean follow-up of 4.5 years (1.0 to 14.6) from revision
for ARMD there were 13 hips (20.3%) with post-operative complications
and eight (12.5%) requiring re-revision. The Kaplan–Meier five-year survival rate for ARMD revision was
87.9% (95% confidence interval 78.9 to 98.0; 19 hips at risk). Excluding
re-revisions, the median absolute Oxford hip score (OHS) following
ARMD revision using the percentage method (0% best outcome and 100%
worst outcome) was 18.8% (interquartile range (IQR) 7.8% to 48.3%),
which is equivalent to 39/48 (IQR 24.8/48 to 44.3/48) when using
the modified OHS. Histopathological response did not affect time
to revision for ARMD (p = 0.334) or the subsequent risk of re-revision
(p = 0.879). Similarly, the presence or absence of a contralateral
MoM hip bearing did not affect time to revision for ARMD (p = 0.066)
or the subsequent risk of re-revision (p = 0.178). Patients revised to MoM bearings had higher rates of re-revision
(five of 16 MoM hips re-revised; p = 0.046), but those not requiring
re-revision had good functional results (median absolute OHS 14.6%
or 41.0/48). Short-term morbidity following revision for ARMD was
comparable with previous reports. Caution should be exercised when choosing
bearing surfaces for ARMD revisions. Cite this article:
We report a retrospective review of the incidence of venous thromboembolism in 463 consecutive patients who underwent primary total hip arthroplasty (487 procedures). Treatment included both total hip replacement and hip resurfacing, and the patients were managed without anticoagulants. The thromboprophylaxis regimen included an antiplatelet agent, generally aspirin, hypotensive epidural anaesthesia, elastic compression stockings and early mobilisation. In 258 of these procedures (244 patients) performed in 2005 (cohort A) mechanical compression devices were not used, whereas in 229 (219 patients) performed during 2006 (cohort B) bilateral intermittent pneumatic calf compression was used. All operations were performed through a posterior mini-incision approach. Patients who required anticoagulation for pre-existing medical problems and those undergoing revision arthroplasty were excluded. Doppler ultrasonographic screening for deep-vein thrombosis was performed in all patients between the fourth and sixth post-operative days. All patients were reviewed at a follow-up clinic six to ten weeks after the operation. In addition, reponse to a questionnaire was obtained at the end of 12 weeks post-operatively. No symptomatic calf or above-knee deep-vein thrombosis or pulmonary embolism occurred. In 25 patients in cohort A (10.2%) and in ten patients in cohort B (4.6%) asymptomatic calf deep-vein thromboses were detected ultrasonographically. This difference was statistically significant (p = 0.03). The regimen followed by cohort B offers the prospect of a low incidence of venous thromboembolism without subjecting patients to the higher risk of bleeding associated with anticoagulant use.
The dysplasia cup, which was devised as an adjunct to the Birmingham Hip Resurfacing system, has a hydroxyapatite-coated porous surface and two supplementary neutralisation screws to provide stable primary fixation, permit early weight-bearing, and allow incorporation of morcellised autograft without the need for structural bone grafting. A total of 110 consecutive dysplasia resurfacing arthroplasties in 103 patients (55 men and 48 women) performed between 1997 and 2000 was reviewed with a minimum follow-up of six years. The mean age at operation was 47.2 years (21 to 62) and 104 hips (94%) were Crowe grade II or III. During the mean follow-up of 7.8 years (6 to 9.6), three hips (2.7%) were converted to a total hip replacement at a mean of 3.9 years (2 months to 8.1 years), giving a cumulative survival of 95.2% at nine years (95% confidence interval 89 to 100). The revisions were due to a fracture of the femoral neck, a collapse of the femoral head and a deep infection. There was no aseptic loosening or osteolysis of the acetabular component associated with either of the revisions performed for failure of the femoral component. No patient is awaiting a revision. The median Oxford hip score in 98 patients with surviving hips at the final review was 13 and the 10th and the 90th percentiles were 12 and 23, respectively.
This study compared component wear rates and pre-revision blood metal ions levels in two groups of failed metal-on-metal hip arthroplasties: hip resurfacing and modular total hip replacement (THR). There was no significant difference in the median rate of linear wear between the groups for both acetabular (p = 0.4633) and femoral (p = 0.0872) components. There was also no significant difference in the median linear wear rates when failed hip resurfacing and modular THR hips of the same type (ASR and Birmingham hip resurfacing (BHR)) were compared. Unlike other studies of well-functioning hips, there was no significant difference in pre-revision blood metal ion levels between hip resurfacing and modular THR. Edge loading was common in both groups, but more common in the resurfacing group (67%) than in the modular group (57%). However, this was not significant (p = 0.3479). We attribute this difference to retention of the neck in resurfacing of the hip, leading to impingement-type edge loading. This was supported by visual evidence of impingement on the femur. These findings show that failed metal-on-metal hip resurfacing and modular THRs have similar component wear rates and are both associated with raised pre-revision blood levels of metal ions.
We examined the relationships between the serum levels of chromium and cobalt ions and the inclination angle of the acetabular component and the level of activity in 214 patients implanted with a metal-on-metal resurfacing hip replacement. Each patient had a single resurfacing and no other metal in their body. All serum measurements were performed at a minimum of one year after operation. The inclination of the acetabular component was considered to be steep if the abduction angle was greater than 55°. There were significantly higher levels of metal ions in patients with steeply-inclined components (p = 0.002 for chromium, p = 0.003 for cobalt), but no correlation was found between the level of activity and the concentration of metal ions. A highly significant (p <
0.001) correlation with the arc of cover was found. Arcs of cover of less than 10 mm were correlated with a greater risk of high concentrations of serum metal ions. The arc of coverage was also related to the design of the component and to size as well as to the abduction angle of the acetabular component. Steeply-inclined acetabular components, with abduction angles greater than 55°, combined with a small size of component are likely to give rise to higher serum levels of cobalt and chromium ions. This is probably due to a greater risk of edge-loading.
Resurfacing arthroplasty of the hip is being performed more frequently in the United Kingdom. The majority of these patients are younger than 55 years of age, and in this group the key benefits include conservation of femoral bone stock and the potential reduction in the rate of dislocation afforded by the larger resurfacing head. Early aseptic loosening is well recognised in patients younger than 55 years of age, and proponents of resurfacing believe that the improved wear characteristics of the metal-on-metal bearing may improve the long-term survival of this implant. There has been some concern, however, that resurfacing may not be conservative of acetabular bone. We compared a series of 33 consecutive patients who had a hybrid total hip arthroplasty with an uncemented acetabular component and a cemented femoral implant, with 35 patients undergoing a Birmingham hip resurfacing arthroplasty. We compared the diameter of the implanted acetabulum in both groups and, because they were not directly comparable, we corrected for patient size by measuring the diameter of the contralateral femoral head. The data were analysed using unpaired There was a significantly larger acetabulum in the Birmingham arthroplasty group (mean diameter 56.6 mm
Early failure associated with adverse reactions to metal debris is an emerging problem after hip resurfacing but the exact mechanism is unclear. We analysed our entire series of 660 metal-on-metal resurfacings (Articular Surface Replacement (ASR) and Birmingham Hip Resurfacing (BHR)) and large-bearing ASR total hip replacements, to establish associations with metal debris-related failures. Clinical and radiological outcomes, metal ion levels, explant studies and lymphocyte transformation tests were performed. A total of 17 patients (3.4%) were identified (all ASR bearings) with adverse reactions to metal debris, for which revision was required. This group had significantly smaller components, significantly higher acetabular component anteversion, and significantly higher whole concentrations of blood and joint chromium and cobalt ions than asymptomatic patients did (all p <
0.001). Post-revision lymphocyte transformation tests on this group showed no reactivity to chromium or cobalt ions. Explants from these revisions had greater surface wear than retrievals for uncomplicated fractures. The absence of adverse reactions to metal debris in patients with well-positioned implants usually implies high component wear. Surgeons must consider implant design, expected component size and acetabular component positioning in order to reduce early failures when performing large-bearing metal-on-metal hip resurfacing and replacement.
Wear of metal-on-metal bearings causes elevated levels of cobalt and chromium in blood and body fluids. Metal-on-metal bearings have two distinct wear phases. In the early phase, the wear rate is high. Later, it decreases and the bearing enters a steady-state phase. It is expected that as the wear rates decline, the level of cobalt detected in plasma will also decrease. We studied the baseline and exercise-related cobalt rise in 21 patients (13 men and eight women) with a mean age of 54 years (38 to 80) who had undergone successful hip resurfacing at a mean of 44 months (10 to 96) earlier. Our results showed that circulating baseline cobalt levels were not significantly correlated with the time since implantation (r = 0.08, p = 0.650). By contrast, the exercise-related cobalt rise was directly correlated with the inclination angle of the acetabular component (r = 0.47, p = 0.032) and inversely correlated with the time since implantation (r = −0.5, p = 0.020). Inclination of the acetabular component should be kept less than 40° to decrease the production of wear debris.
We report an independent prospective review of the first 230 Birmingham hip resurfacings in 212 patients at a mean follow-up of five years (4 to 6). Two patients, one with a loose acetabular component and the other with suspected avascular necrosis of the femoral head, underwent revision. There were two deaths from unrelated causes and one patient was lost to follow-up. The survivorship with the worst-case scenario was 97.8% (95% confidence interval 95.8 to 99.5). The mean Harris hip score improved significantly (paired On radiological review at five years, one patient had a progressive lucent line around the acetabular component and six had progressive lucent lines around the femoral component. A total of 18 femoral components (8%) had migrated into varus and those with lucent lines present migrated a mean of 3.8° (1.02° to 6.54°) more than the rest. Superolateral notching of the femoral neck and reactive sclerosis at the tip of the peg of the femoral component were associated with the presence of lucent lines (chi-squared test, p <
0.05), but not with migration of the femoral component, and are of unknown significance. Our results with the Birmingham hip resurfacing continue to be satisfactory at a mean follow-up of five years.
The rate and mode of early failure in 463 Birmingham hip resurfacings in a two-centre, multisurgeon series were examined. Of the 463 patients two have died and three were lost to follow-up. The mean radiological and clinical follow-up was for 43 months (6 to 90). We have revised 13 resurfacings (2.8%) including seven for pain, three for fracture, two for dislocation and another for sepsis. Of these, nine had macroscopic and histological evidence of metallosis. The survival at five years was 95.8% (95% confidence interval (CI) 94.1 to 96.8) for revision for all causes and 96.9% (95% CI 95.5 to 98.3) for metallosis. The rate of metallosis related revision was 3.1% at five years. Risk factors for metallosis were female gender, a small femoral component, a high abduction angle and obesity. We do not advocate the use of the Birmingham Hip resurfacing procedure in patients with these risk factors.
Despite the increasing interest and subsequent published literature on hip resurfacing arthroplasty, little is known about the prevalence of its complications and in particular the less common modes of failure. The aim of this study was to identify the prevalence of failure of hip resurfacing arthroplasty and to analyse the reasons for it. From a multi-surgeon series (141 surgeons) of 5000 Birmingham hip resurfacings we have analysed the modes, prevalence, gender differences and times to failure of any hip requiring revision. To date 182 hips have been revised (3.6%). The most common cause for revision was a fracture of the neck of the femur (54 hips, prevalence 1.1%), followed by loosening of the acetabular component (32 hips, 0.6%), collapse of the femoral head/avascular necrosis (30 hips, 0.6%), loosening of the femoral component (19 hips, 0.4%), infection (17 hips, 0.3%), pain with aseptic lymphocytic vascular and associated lesions (ALVAL)/metallosis (15 hips, 0.3%), loosening of both components (five hips, 0.1%), dislocation (five hips, 0.1%) and malposition of the acetabular component (three hips, 0.1%). In two cases the cause of failure was unknown. Comparing men with women, we found the prevalence of revision to be significantly higher in women (women = 5.7%; men = 2.6%, p <
0.001). When analysing the individual modes of failure women had significantly more revisions for loosening of the acetabular component, dislocation, infection and pain/ALVAL/metallosis (p <
0.001, p = 0.004, p = 0.008, p = 0.01 respectively). The mean time to failure was 2.9 years (0.003 to 11.0) for all causes, with revision for fracture of the neck of the femur occurring earlier than other causes (mean 1.5 years, 0.02 to 11.0). There was a significantly shorter time to failure in men (mean 2.1 years, 0.4 to 8.7) compared with women (mean 3.6 years, 0.003 to 11.0) (p <
0.001).
The purpose of this study was twofold: first,
to determine whether the five-year results of hip resurfacing arthroplasty
(HRA) in Canada justified the continued use of HRA; and second,
to identify whether greater refinement of patient selection was
warranted. This was a retrospective cohort study that involved a review
of 2773 HRAs performed between January 2001 and December 2008 at
11 Canadian centres. Cox’s proportional hazards models were used
to analyse the predictors of failure of HRA. Kaplan–Meier survival
analysis was performed to predict the cumulative survival rate at
five years. The factors analysed included age, gender, body mass
index, pre-operative hip pathology, surgeon’s experience, surgical
approach, implant sizes and implant types. The most common modes
of failure were also analysed. The 2773 HRAs were undertaken in 2450 patients: 2127 in men and
646 in women. The mean age at operation was 50.5 years ( The failure rates of HRA at five years justify the ongoing use
of this technique in men. Female gender is an independent predictor
of failure, and a higher failure rate at five years in women leads
the authors to recommend this technique only in exceptional circumstances
for women. Cite this article:
We describe the experience with the first consecutive 230 Birmingham hip resurfacings at our centre. At a mean follow-up of three years (25 to 52 months) survivorship was 99.14% with revision in one patient for a loose acetabular component and one death from unrelated causes. One patient developed a fracture of the femoral neck at six weeks which united unremarkably after a period of non-weight-bearing. The Harris hip score improved from a mean of 62.54 (8 to 92) to 97.74 (61 to 100). The mean flexion improved from 91.52° (25 to 140) to 110.41° (80 to 145). Most patients (97%) considered the outcome to be good or excellent. Our preliminary experience with this implant is encouraging and the results are superior to the earlier generation of resurfacings for the same length of follow-up.