Aims.
Aims. The purpose of the present study was to compare
Accurate placement of the acetabular component during total hip
arthroplasty (THA) is an important factor in the success of the
procedure. However, the reported accuracy varies greatly and is
dependent upon whether free hand or navigated techniques are used.
The aim of this study was to assess the accuracy of an instrument
system that incorporates 3D printed, patient-specific guides designed
to optimise the placement of the acetabular component. A total of 100 consecutive patients were prospectively enrolled
and the accuracy of placement of the acetabular component was measured
using post-operative CT scans.Aims
Patients and Methods
We conducted a meta-analysis, including randomised
controlled trials (RCTs) and cohort studies, to examine the effect
of patient-specific instruments (PSI) on radiological outcomes after
total knee replacement (TKR) including: mechanical axis alignment
and malalignment of the femoral and tibial components in the coronal,
sagittal and axial planes, at a threshold of >
3º from neutral.
Relative risks (RR) for malalignment were determined for all studies
and for RCTs and cohort studies separately. Of 325 studies initially identified, 16 met the eligibility criteria,
including eight RCTs and eight cohort studies. There was no significant
difference in the likelihood of mechanical axis malalignment with
PSI We conclude that PSI does not improve the accuracy of alignment
of the components in TKR compared with conventional instrumentation. Cite this article:
Advanced 3D imaging and CT-based navigation have emerged as valuable tools to use in total knee arthroplasty (TKA), for both preoperative planning and the intraoperative execution of different philosophies of alignment. Preoperative planning using CT-based 3D imaging enables more accurate prediction of the size of components, enhancing surgical workflow and optimizing the precision of the positioning of components. Surgeons can assess alignment, osteophytes, and arthritic changes better. These scans provide improved insights into the patellofemoral joint and facilitate tibial sizing and the evaluation of implant-bone contact area in cementless TKA. Preoperative CT imaging is also required for the development of
The goal of the current systematic review was to assess the impact of implant placement accuracy on outcomes following total knee arthroplasty (TKA). A systematic review was performed in accordance with the Preferred Reporting Items for Systematic Reviews and Meta-Analyses (PRISMA) guidelines using the Ovid Medline, Embase, Cochrane Central, and Web of Science databases in order to assess the impact of the patient-reported outcomes measures (PROMs) and implant placement accuracy on outcomes following TKA. Studies assessing the impact of implant alignment, rotation, size, overhang, or condylar offset were included. Study quality was assessed, evidence was graded (one-star: no evidence, two-star: limited evidence, three-star: moderate evidence, four-star: strong evidence), and recommendations were made based on the available evidence.Aims
Methods
In this study we randomised 140 patients who
were due to undergo primary total knee arthroplasty (TKA) to have the
procedure performed using either patient-specific cutting guides
(PSCG) or conventional instrumentation (CI). The primary outcome measure was the mechanical axis, as measured
at three months on a standing long-leg radiograph by the hip–knee–ankle
(HKA) angle. This was undertaken by an independent observer who
was blinded to the instrumentation. Secondary outcome measures were
component positioning, operating time, Knee Society and Oxford knee
scores, blood loss and length of hospital stay. A total of 126 patients (67 in the CI group and 59 in the PSCG
group) had complete clinical and radiological data. There were 88
females and 52 males with a mean age of 69.3 years (47 to 84) and
a mean BMI of 28.6 kg/m2 (20.2 to 40.8). The mean HKA
angle was 178.9° (172.5 to 183.4) in the CI group and 178.2° (172.4
to 183.4) in the PSCG group (p = 0.34). Outliers were identified
in 22 of 67 knees (32.8%) in the CI group and 19 of 59 knees (32.2%)
in the PSCG group (p = 0.99). There was no significant difference
in the clinical results (p = 0.95 and 0.59, respectively). Operating time,
blood loss and length of hospital stay were not significantly reduced
(p = 0.09, 0.58 and 0.50, respectively) when using PSCG. The use of PSCG in primary TKA did not reduce the proportion
of outliers as measured by post-operative coronal alignment. Cite this article:
Aims. This is a multicentre, non-inventor, prospective observational study of 503 INFINITY fixed bearing total ankle arthroplasties (TAAs). We report our early experience, complications, and radiological and functional outcomes. Methods. Patients were recruited from 11 specialist centres between June 2016 and November 2019. Demographic, radiological, and functional outcome data (Ankle Osteoarthritis Scale, Manchester Oxford Questionnaire, and EuroQol five-dimension five-level score) were collected preoperatively, at six months, one year, and two years. The Canadian Orthopaedic Foot and Ankle Society (COFAS) grading system was used to stratify deformity. Early and late complications and reoperations were recorded as adverse events. Radiographs were assessed for lucencies, cysts, and/or subsidence. Results. In all, 500 patients reached six-month follow-up, 420 reached one-year follow-up, and 188 reached two-year follow-up. The mean age was 67.8 years (23.9 to 88.5). A total of 38 patients (7.5%) presented with inflammatory arthritis. A total of 101 (20.0%) of implantations used
Aims.
Substantial healthcare resources have been devoted
to computer navigation and
Improvements in the surgical technique of total
knee replacement (TKR) are continually being sought. There has recently
been interest in three-dimensional (3D) pre-operative planning using
magnetic resonance imaging (MRI) and CT. The 3D images are increasingly
used for the production of patient-specific models, surgical guides
and custom-made implants for TKR. The users of
L. Spencer-Gardner, J. Pierrepont, M. Topham,
J. Baré, S. McMahon, A. J. Shimmin
Total hip and knee arthroplasty (THA, TKA) are largely successful procedures; however, both have variable outcomes, resulting in some patients being dissatisfied with the outcome. Surgeons are turning to technologies such as robotic-assisted surgery in an attempt to improve outcomes. Robust studies are needed to find out if these innovations are really benefitting patients. The Robotic Arthroplasty Clinical and Cost Effectiveness Randomised Controlled Trials (RACER) trials are multicentre, patient-blinded randomized controlled trials. The patients have primary osteoarthritis of the hip or knee. The operation is Mako-assisted THA or TKA and the control groups have operations using conventional instruments. The primary clinical outcome is the Forgotten Joint Score at 12 months, and there is a built-in analysis of cost-effectiveness. Secondary outcomes include early pain, the alignment of the components, and medium- to long-term outcomes. This annotation outlines the need to assess these technologies and discusses the design and challenges when conducting such trials, including surgical workflows, isolating the effect of the operation, blinding, and assessing the learning curve. Finally, the future of robotic surgery is discussed, including the need to contemporaneously introduce and evaluate such technologies. Cite this article:
Reverse total shoulder arthroplasty (rTSA) can be used in complex cases when the glenoid requires reconstruction. In this study, a baseplate with composite bone autograft and a central trabecular titanium peg was implanted, and its migration was assessed for two years postoperatively using radiostereometric analysis (RSA). A total of 14 patients who underwent a rTSA with an autograft consented to participate. Of these, 11 had a primary rTSA using humeral head autograft and three had a revision rTSA with autograft harvested from the iliac crest. The mean age of the patients was 66 years (39 to 81). Tantalum beads were implanted in the scapula around the glenoid. RSA imaging (stereographic radiographs) was undertaken immediately postoperatively and at three, six, 12, and 24 months. Analysis was completed using model-based RSA software. Outcomes were collected preoperatively and at two years postoperatively, including the Oxford Shoulder Score, the American Shoulder and Elbow Score, and a visual analogue score for pain. A Constant score was also obtained for the assessment of strength and range of motion.Aims
Methods
Intraoperative pressure sensors allow surgeons to quantify soft-tissue balance during total knee arthroplasty (TKA). The aim of this study was to determine whether using sensors to achieve soft-tissue balance was more effective than manual balancing in improving outcomes in TKA. A multicentre randomized trial compared the outcomes of sensor balancing (SB) with manual balancing (MB) in 250 patients (285 TKAs). The primary outcome measure was the mean difference in the four Knee injury and Osteoarthritis Outcome Score subscales (ΔKOOS4) in the two groups, comparing the preoperative and two-year scores. Secondary outcomes included intraoperative balance data, additional patient-reported outcome measures (PROMs), and functional measures.Aims
Methods
Controversy exists as to what should be considered a safe resection margin to minimize local recurrence in high-grade pelvic chondrosarcomas (CS). The aim of this study is to quantify what is a safe margin of resection for high-grade CS of the pelvis. We retrospectively identified 105 non-metastatic patients with high-grade pelvic CS of bone who underwent surgery (limb salvage/amputations) between 2000 and 2018. There were 82 (78%) male and 23 (22%) female patients with a mean age of 55 years (26 to 84). The majority of the patients underwent limb salvage surgery (n = 82; 78%) compared to 23 (22%) who had amputation. In total, 66 (64%) patients were grade 2 CS compared to 38 (36%) grade 3 CS. All patients were assessed for stage, pelvic anatomical classification, type of resection and reconstruction, margin status, local recurrence, distant recurrence, and overall survival. Surgical margins were stratified into millimetres: < 1 mm; > 1 mm but < 2 mm; and > 2 mm.Aims
Methods
This aim of this study was to assess the feasibility of designing and introducing generic 3D-printed instrumentation for routine use in total knee arthroplasty. Instruments were designed to take advantage of 3D-printing technology, particularly ensuring that all parts were pre-assembled, to theoretically reduce the time and skill required during surgery. Concerning functionality, ranges of resection angle and distance were restricted within a safe zone, while accommodating either mechanical or anatomical alignment goals. To identify the most suitable biocompatible materials, typical instrument shapes and mating parts, such as dovetails and screws, were designed and produced.Aims
Materials and Methods
There is little literature about total knee arthroplasty (TKA) after distal femoral osteotomy (DFO). Consequently, the purpose of this study was to analyze the outcomes of TKA after DFO, with particular emphasis on: survivorship free from aseptic loosening, revision, or any re-operation; complications; radiological results; and clinical outcome. We retrospectively reviewed 29 patients (17 women, 12 men) from our total joint registry who had undergone 31 cemented TKAs after a DFO between 2000 and 2012. Their mean age at TKA was 51 years (22 to 76) and their mean body mass index 32 kg/m2 (20 to 45). The mean time between DFO and TKA was ten years (2 to 20). The mean follow-up from TKA was ten years (2 to 16). The prostheses were posterior-stabilized in 77%, varus-valgus constraint (VVC) in 13%, and cruciate-retaining in 10%. While no patient had metaphyseal fixation (e.g. cones or sleeves), 16% needed a femoral stem.Aims
Patients and Methods
Aims
Patients and Methods
Accurate placement of the acetabular component is essential in
total hip arthroplasty (THA). The purpose of this study was to determine
if the ability to achieve inclination of the acetabular component
within the ‘safe-zone’ of 30° to 50° could be improved with the
use of an inclinometer. We reviewed 167 primary THAs performed by a single surgeon over
a period of 14 months. Procedures were performed at two institutions:
an inpatient hospital, where an inclinometer was used (inclinometer
group); and an ambulatory centre, where an inclinometer was not
used as it could not be adequately sterilized (control group). We excluded
47 patients with a body mass index (BMI) of > 40 kg/m2,
age of > 68 years, or a surgical indication other than osteoarthritis
whose treatment could not be undertaken in the ambulatory centre.
There were thus 120 patients in the study, 68 in the inclinometer
group and 52 in the control group. The inclination angles of the acetabular
component were measured from de-identified plain radiographs by
two blinded investigators who were not involved in the surgery.
The effect of the use of the inclinometer on the inclination angle
was determined using multivariate regression analysis.Aims
Patients and Methods
Our aim was to compare kinematic with mechanical alignment in
total knee arthroplasty (TKA). We performed a prospective blinded randomised controlled trial
to compare the functional outcome of patients undergoing TKA in
mechanical alignment (MA) with those in kinematic alignment (KA).
A total of 71 patients undergoing TKA were randomised to either
kinematic (n = 36) or mechanical alignment (n = 35). Pre- and post-operative
hip-knee-ankle radiographs were analysed. The knee injury and osteoarthritis
outcome score (KOOS), American Knee Society Score, Short Form-36,
Euro-Qol (EQ-5D), range of movement (ROM), two minute walk, and timed
up and go tests were assessed pre-operatively and at six weeks,
three and six months and one year post-operatively.Aims
Patients and Methods
We evaluated the accuracy with which a custom-made
acetabular component could be positioned at revision arthroplasty
of the hip in patients with a Paprosky type 3 acetabular defect. A total of 16 patients with a Paprosky type 3 defect underwent
revision surgery using a custom-made trabecular titanium implant.
There were four men and 12 women with a median age of 67 years (48
to 79). The planned inclination (INCL), anteversion (AV), rotation
and centre of rotation (COR) of the implant were compared with the post-operative
position using CT scans. A total of seven implants were malpositioned in one or more parameters:
one with respect to INCL, three with respect to AV, four with respect
to rotation and five with respect to the COR. To the best of our knowledge, this is the first study in which
CT data acquired for the pre-operative planning of a custom-made
revision acetabular implant have been compared with CT data on the
post-operative position. The results are encouraging. Cite this article:
Primary total knee arthroplasty (TKA) is a reliable
procedure with reproducible long-term results. Nevertheless, there
are conditions related to the type of patient or local conditions
of the knee that can make it a difficult procedure. The most common
scenarios that make it difficult are discussed in this review. These
include patients with many previous operations and incisions, and
those with severe coronal deformities, genu recurvatum, a stiff knee,
extra-articular deformities and those who have previously undergone
osteotomy around the knee and those with chronic dislocation of
the patella. Each condition is analysed according to the characteristics of
the patient, the pre-operative planning and the reported outcomes. When approaching the difficult primary TKA surgeons should use
a systematic approach, which begins with the review of the existing
literature for each specific clinical situation. Cite this article: