In 20 patients undergoing hybrid total hip arthroplasty, the reproducibility and accuracy of templating using
Magnification of anteroposterior radiographs of the pelvis is variable. To improve the accuracy of templating, reliable and radiographer-friendly methods of scaling are necessary. We assessed two methods of scaling
In this randomized study, we aimed to compare quality of regenerate in monolateral Both groups were comparable in demographic and injury characteristics. A phantom (aluminium step wedge of increasing thickness) was designed to compare the density of regenerate on radiographs. A CT scan was performed at three and six months postoperatively to assess regenerate density. A total of 30 patients (29 male, one female; mean age 32.54 years (18 to 60)) with an infected nonunion of a tibial fracture presenting to our tertiary institute between June 2011 and April 2016 were included in the study.Aims
Patients and Methods
Aims. The aims of this study were to identify means to quantify coronal plane displacement associated with distal radius fractures (DRFs), and to understand their relationship to radial inclination (RI). Methods. From posteroanterior
We have investigated the accuracy of the templating of
We aimed to determine the reliability, accuracy and the clinical role of digital templating in the pre-operative work-up for total knee replacement. Initially a sample of ten pre-operative
We have compared four computer-assisted methods to measure penetration of the femoral head into the acetabular component in total hip replacement. These were the Martell Hip Analysis suite 7.14, Rogan HyperOrtho, Rogan View Pro-X and Roman v1.70. The images used for the investigation comprised 24 anteroposterior
The Birmingham Hip Resurfacing (BHR) was introduced in 1997 to address the needs of young active patients using a historically proven large-diameter metal-on-metal (MoM) bearing. A single designer surgeon’s consecutive series of 130 patients (144 hips) was previously reported at five and ten years, reporting three and ten failures, respectively. The aim of this study was to extend the follow-up of this original cohort at 25 years. The study extends the reporting on the first consecutive 144 resurfacing procedures in 130 patients for all indications. All operations were undertaken between August 1997 and May 1998. The mean age at operation was 52.1 years (SD 9.93; 17 to 76), and included 37 female patients (28.5%). Failure was defined as revision of either component for any reason. Kaplan-Meier survival analysis was performed. Routine follow-up with serum metal ion levels, radiographs, and Oxford Hip Scores (OHSs) was undertaken.Aims
Methods
There is no level I evidence dealing with the optimal period of immobilization for patients with a displaced distal radial fracture following closed reduction. A shorter period might lead to a better functional outcome due to less stiffness and pain. The aim of this study was to investigate whether this period could be safely reduced from six to four weeks. This multicentre randomized controlled trial (RCT) included adult patients with a displaced distal radial fracture, who were randomized to be treated with immobilization in a cast for four or six weeks following closed reduction. The primary outcome measure was the Patient-Rated Wrist Evaluation (PRWE) score after follow-up at one year. Secondary outcomes were the abbreviated version of the Disability of Arm, Shoulder and Hand (QuickDASH) score after one year, the functional outcome at six weeks, 12 weeks, and six months, range of motion (ROM), the level of pain after removal of the cast, and complications.Aims
Methods
The aim of this retrospective cohort study was to assess and investigate the safety and efficacy of using a distal tibial osteotomy compared to proximal osteotomy for limb lengthening in children. In this study, there were 59 consecutive tibial lengthening and deformity corrections in 57 children using a circular frame. All were performed or supervised by the senior author between January 2013 and June 2019. A total of 25 who underwent a distal tibial osteotomy were analyzed and compared to a group of 34 who had a standard proximal tibial osteotomy. For each patient, the primary diagnosis, time in frame, complications, and lengthening achieved were recorded. From these data, the frame index was calculated (days/cm) and analyzed.Aims
Methods
The aim of this study was to investigate whether anterior pelvic plane-pelvic tilt (APP-PT) is associated with distinct hip pathomorphologies. We asked: is there a difference in APP-PT between young symptomatic patients being evaluated for joint preservation surgery and an asymptomatic control group? Does APP-PT vary among distinct acetabular and femoral pathomorphologies? And does APP-PT differ in symptomatic hips based on demographic factors? This was an institutional review board-approved, single-centre, retrospective, case-control, comparative study, which included 388 symptomatic hips in 357 patients who presented to our tertiary centre for joint preservation between January 2011 and December 2015. Their mean age was 26 years (SD 2; 23 to 29) and 50% were female. They were allocated to 12 different morphological subgroups. The study group was compared with a control group of 20 asymptomatic hips in 20 patients. APP-PT was assessed in all patients based on supine anteroposterior pelvic radiographs using validated HipRecon software. Values in the two groups were compared using an independent-samples Aims
Methods
Excessive posterior pelvic tilt (PT) may increase the risk of anterior instability after total hip arthroplasty (THA). The aim of this study was to investigate the changes in PT occurring from the preoperative supine to postoperative standing position following THA, and identify factors associated with significant changes in PT. Supine PT was measured on preoperative CT scans and standing PT was measured on preoperative and one-year postoperative standing lateral radiographs in 933 patients who underwent primary THA. Negative values indicate posterior PT. Patients with > 13° of posterior PT from preoperative supine to postoperative standing (ΔPT ≤ -13°) radiographs, which corresponds to approximately a 10° increase in functional anteversion of the acetabular component, were compared with patients with less change (ΔPT > -13°). Logistic regression analysis was used to assess preoperative demographic and spinopelvic parameters predictive of PT changes of ≤ -13°. The area under receiver operating characteristic curve (AUC) determined the diagnostic accuracy of the predictive factors.Aims
Methods
BoneMaster is a thin electrochemically applied hydroxyapatite (HA) coating for orthopaedic implants that is quickly resorbed during osseointegration. Early stabilization is a surrogacy marker of good survival of femoral stems. The hypothesis of this study was that a BoneMaster coating yields a fast early and lasting fixation of stems. A total of 53 patients were randomized to be treated using Bi-Metric cementless femoral stems with either only a porous titanium plasma-sprayed coating (P group) or a porous titanium plasma-sprayed coating with an additional BoneMaster coating (PBM group). The patients were examined with radiostereometry until five years after surgery.Aims
Methods
Developmental dysplasia of the hip (DDH) describes a pathological relationship between the femoral head and acetabulum. Periacetabular osteotomy (PAO) may be used to treat this condition. The aim of this study was to evaluate the results of PAO in adolescents and adults with persistent DDH. Patients were divided into four groups: A, adolescents who had not undergone surgery for DDH in childhood (25 hips); B, adolescents who had undergone surgery for DDH in childhood (20 hips); C, adults with DDH who had not undergone previous surgery (80 hips); and D, a control group of patients with healthy hips (70 hips). The radiological evaluation of digital anteroposterior views of hips included the Wiberg angle (centre-edge angle (CEA)), femoral head cover (FHC), medialization, distalization, and the ilioischial angle. Clinical assessment involved the Harris Hip Score (HHS) and gluteal muscle performance assessment.Aims
Methods
The aim of the study was to compare two methods of calculating pelvic incidence (PI) and pelvic tilt (PT), either by using the femoral heads or acetabular domes to determine the bicoxofemoral axis, in patients with unilateral or bilateral primary hip osteoarthritis (OA). PI and PT were measured on standing lateral radiographs of the spine in two groups: 50 patients with unilateral (Group I) and 50 patients with bilateral hip OA (Group II), using the femoral heads or acetabular domes to define the bicoxofemoral axis. Agreement between the methods was determined by intraclass correlation coefficient (ICC) and the standard error of measurement (SEm). The intraobserver reproducibility and interobserver reliability of the two methods were analyzed on 31 radiographs in both groups to calculate ICC and SEm.Aims
Methods
To benchmark the radiation dose to patients during the course of treatment for a spinal deformity. Our radiation dose database identified 25,745 exposures of 6,017 children (under 18 years of age) and adults treated for a spinal deformity between 1 January 2008 and 31 December 2016. Patients were divided into surgical (974 patients) and non-surgical (5,043 patients) cohorts. We documented the number and doses of ionizing radiation imaging events (radiographs, CT scans, or intraoperative fluoroscopy) for each patient. All the doses for plain radiographs, CT scans, and intraoperative fluoroscopy were combined into a single effective dose by a medical physicist (milliSivert (mSv)).Aims
Methods
Cement-in-cement revision of the femoral component represents a widely practised technique for a variety of indications in revision total hip arthroplasty. In this study, we compare the clinical and radiological outcomes of two polished tapered femoral components. From our prospectively collated database, we identified all patients undergoing cement-in-cement revision from January 2005 to January 2013 who had a minimum of two years' follow-up. All cases were performed by the senior author using either an Exeter short revision stem or the C-Stem AMT high offset No. 1 prosthesis. Patients were followed-up annually with clinical and radiological assessment.Aims
Methods
The migration percentage (MP) is one criterion used for surgery in dislocated or displaced hips in children with cerebral palsy (CP). The MP at which a displaced hip can no longer return to normal is unclear. The aim of this paper was to identify the point of no return of the MP through a large population-based study. All children registered on the Cerebral Palsy Integrated Pathway Scotland surveillance programme undergo regular pelvic radiographs. Any child who had a MP measuring over 35% since the programme’s inception in 2013, in at least one hip and at one timepoint, was identified. The national radiography database was then interrogated to identify all pelvic radiographs for each of these children from birth through to the date of analysis. A minimum of a further two available radiographs following the initial measurement of MP ≥ 35% was required for inclusion.Aims
Methods
We aimed to investigate if the use of the largest possible cobalt-chromium head articulating with polyethylene acetabular inserts would increase the in vivo wear rate in total hip arthroplasty. In a single-blinded randomized controlled trial, 96 patients (43 females), at a median age of 63 years (interquartile range (IQR) 57 to 69), were allocated to receive either the largest possible modular femoral head (36 mm to 44 mm) in the thinnest possible insert or a standard 32 mm head. All patients received a vitamin E-doped cross-linked polyethylene insert and a cobalt-chromium head. The primary outcome was proximal head penetration measured with radiostereometric analysis (RSA) at two years. Secondary outcomes were volumetric wear, periacetabular radiolucencies, and patient-reported outcomes.Aims
Methods
The most frequent indication for revision surgery in total hip arthroplasty (THA) is aseptic loosening. Aseptic loosening is associated with polyethylene liner wear, and wear may be reduced by using vitamin E-doped liners. The primary objective of this study was to compare proximal femoral head penetration into the liner between a) two cross-linked polyethylene (XLPE) liners (vitamin E-doped (vE-PE)) versus standard XLPE liners, and b) two modular femoral head diameters (32 mm and 36 mm). Patients scheduled for a THA were randomized to receive a vE-PE or XLPE liner with a 32 mm or 36 mm metal head (four intervention groups in a 2 × 2 factorial design). Head penetration and acetabular component migration were measured using radiostereometric analysis at baseline, three, 12, 24, and 60 months postoperatively. The Harris Hip Score, University of California, Los Angeles (UCLA) Activity Score, EuroQol five-dimension questionnaire (EQ-5D), and 36-Item Short-Form Health Survey questionnaire (SF-36) were assessed at baseline, three, 12, 36, and 60 months.Aims
Methods
Current treatments of prosthetic joint infection (PJI) are minimally effective against The ability of PlySs2 and vancomycin to kill biofilm and colony-forming units (CFUs) on orthopaedic implants were compared using in vitro models. An in vivo murine PJI model of DAIR was used to assess the efficacy of a combination of PlySs2 and vancomycin on periprosthetic bacterial load.Aims
Methods
The Wrightington classification system of fracture-dislocations of the elbow divides these injuries into six subtypes depending on the involvement of the coronoid and the radial head. The aim of this study was to assess the reliability and reproducibility of this classification system. This was a blinded study using radiographs and CT scans of 48 consecutive patients managed according to the Wrightington classification system between 2010 and 2018. Four trauma and orthopaedic consultants, two post CCT fellows, and one speciality registrar based in the UK classified the injuries. The seven observers reviewed preoperative radiographs and CT scans twice, with a minimum four-week interval. Radiographs and CT scans were reviewed separately. Inter- and intraobserver reliability were calculated using Fleiss and Cohen kappa coefficients. The Landis and Koch criteria were used to interpret the strength of the kappa values. Validity was assessed by calculating the percentage agreement against intraoperative findings.Aims
Methods
The purpose of this study was to evaluate spinopelvic mechanics from standing and sitting positions in subjects with and without femoroacetabular impingement (FAI). We hypothesize that FAI patients will experience less flexion at the lumbar spine and more flexion at the hip whilst changing from standing to sitting positions than subjects without FAI. This increase in hip flexion may contribute to symptomatology in FAI. Male subjects were prospectively enrolled to the study (n = 20). Mean age was 31 years old (22 to 41). All underwent clinical examination, plain radiographs, and dynamic imaging using EOS. Subjects were categorized into three groups: non-FAI (no radiographic or clinical FAI or pain), asymptomatic FAI (radiographic and clinical FAI but no pain), and symptomatic FAI (patients with both pain and radiographic FAI). FAI was defined as internal rotation less than 15° and alpha angle greater than 60°. Subjects underwent standing and sitting radiographs in order to measure spine and femoroacetabular flexion.Aims
Patients and Methods
The primary aim of this study was to determine the morbidity
of a tibial strut autograft and characterize the rate of bony union
following its use. We retrospectively assessed a series of 104 patients from a single
centre who were treated with a tibial strut autograft of > 5 cm
in length. A total of 30 had a segmental reconstruction with continuity
of bone, 27 had a segmental reconstruction without continuity of
bone, 29 had an arthrodesis and 18 had a nonunion. Donor-site morbidity
was defined as any event that required a modification of the postoperative
management. Union was assessed clinically and radiologically at
a median of 36 months (IQR, 14 to 74).Aims
Patients and Methods
To evaluate interobserver reliability of the Orthopaedic Trauma
Association’s open fracture classification system (OTA-OFC). Patients of any age with a first presentation of an open long
bone fracture were included. Standard radiographs, wound photographs,
and a short clinical description were given to eight orthopaedic
surgeons, who independently evaluated the injury using both the
Gustilo and Anderson (GA) and OTA-OFC classifications. The responses
were compared for variability using Cohen’s kappa.Aims
Patients and Methods
This prospective randomised controlled trial was designed to
evaluate the outcome of both the MRI- and CT-based patient-specific
matched guides (PSG) from the same manufacturer. A total of 137 knees in 137 patients (50 men, 87 women) were
included, 67 in the MRI- and 70 in the CT-based PSG group. Their
mean age was 68.4 years (47.0 to 88.9). Outcome was expressed as
the biomechanical limb alignment (centre hip-knee-ankle: HKA-axis)
achieved post-operatively, the position of the individual components
within 3° of the pre-operatively planned alignment, correct planned
implant size and operative data (e.g. operating time and blood loss).Aims
Patients and Methods
The aim of this study was to report the clinical, functional
and radiological outcomes of children and adolescents with tibial
fractures treated using the Ilizarov method. Between 2013 and 2016 a total of 74 children with 75 tibial fractures
underwent treatment at our major trauma centre using an Ilizarov
frame. Demographic and clinical information from a prospective database
was supplemented by routine functional and psychological assessment
and a retrospective review of the notes and radiographs.Aims
Patients and Methods
The purpose of this study was to evaluate whether an innovative
templating technique could predict the need for acetabular augmentation
during primary total hip arthroplasty for patients with dysplastic
hips. We developed a simple templating technique to estimate acetabular
component coverage at total hip arthroplasty, the True Cup: False
Cup (TC:FC) ratio. We reviewed all patients with dysplastic hips
who underwent primary total hip arthroplasty between 2005 and 2012.
Traditional radiological methods of assessing the degree of acetabular dysplasia
(Sharp’s angle, Tönnis angle, centre-edge angle) as well as the
TC:FC ratio were measured from the pre-operative radiographs. A
comparison of augmented and non-augmented hips was undertaken to
determine any difference in pre-operative radiological indices between
the two cohorts. The intra- and inter-observer reliability for all
radiological indices used in the study were also calculated.Aims
Patients and Methods
The aim of this study was to investigate differences in pain,
range of movement function and satisfaction at three months and
one year after total knee arthroplasty (TKA) in patients with an
oblique pattern of kinematic graph of the knee and those with a
varus pattern. A total of 91 patients who underwent TKA were included in this
retrospective study. Patients (59 women and 32 men with mean age
of 68.7 years; 38.6 to 88.4) were grouped according to kinematic
graphs which were generated during navigated TKA and the outcomes
between the groups were compared.Aims
Patients and Methods
We wished to examine the effectiveness of tibial lengthening
using a two ring Ilizarov frame in skeletally immature patients.
This is a potentially biomechanically unstable construct which risks
the loss of axial control. We retrospectively reviewed a consecutive series of 24 boys and
26 girls, with a mean age of 8.6 years (4 to 14), who underwent
52 tibial lengthening procedures with a mean follow-up of 4.3 years
(4.0 to 16.9). Tibial alignment was measured before and after treatment
using joint orientation lines from the knee and a calculation of
the oblique plane axis.Aims
Patients and Methods
Our first aim was to determine whether there are significant
changes in the level of metal ions in the blood at mid-term follow-up,
in patients with an Articular Surface Replacement (ASR) arthroplasty.
Secondly, we sought to identify risk factors for any increases. The study involved 435 patients who underwent unilateral, metal-on-metal
(MoM) hip resurfacing (HRA) or total hip arthroplasty (THA). These
patients all had one measurement of the level of metal ions in the
blood before seven years had passed post-operatively (early evaluation)
and one after seven years had passed post-operatively (mid-term evaluation).
Changes in ion levels were tested using a Wilcoxon signed-rank test.
We identified subgroups at the highest risk of increase using a
multivariable linear logistic regression model.Aims
Patients and Methods
Although success has been achieved with implantation of bone marrow mesenchymal stem cells (bMSCs) in degenerative discs, its full potential may not be achieved if the harsh environment of the degenerative disc remains. Axial distraction has been shown to increase hydration and nutrition. Combining both therapies may have a synergistic effect in reversing degenerative disc disease. In order to evaluate the effect of bMSC implantation, axial distraction and combination therapy in stimulating regeneration and retarding degeneration in degenerative discs, we first induced disc degeneration by axial loading in a rabbit model. The rabbits in the intervention groups performed better with respect to disc height, morphological grading, histological scoring and average dead cell count. The groups with distraction performed better than those without on all criteria except the average dead cell count. Our findings suggest that bMSC implantation and distraction stimulate regenerative changes in degenerative discs in a rabbit model.
The aim of this study was to investigate the long-term clinical
and radiological outcome of patients who suffer recurrent injuries
to the anterior cruciate ligament (ACL) after reconstruction and
require revision surgery. From a consecutive series of 200 patients who underwent primary
reconstruction following rupture of the ACL, we identified 36 who
sustained a further rupture, 29 of whom underwent revision surgery.
Patients were reviewed prospectively at one, two, seven, 15 and
about 20 years after their original surgery. Primary outcome measures
were the number of further ruptures, the posterior tibial slope
(PTS), and functional and radiological outcomes. These were compared
with a gender and age matched cohort of patients who underwent primary
ACL reconstruction only.Aims
Patients and Methods
We report patient-reported outcomes and complications
associated with contemporary periacetabular osteotomy (PAO) surgery
in treating symptomatic acetabular dysplasia and compare these outcomes
with total hip arthroplasty (THA) in patients with similar demographic
details. Two consecutive cohorts included patients between aged
18 to 40 years who had undergone either PAO (100 hips; 24 male,
76 female) or THA (55 hips; 18 male, 37 female). At a mean follow-up
of 5.9 years (2 to 13), there was significant improvement in the
modified Harris hip pain (p <
0.001, PAO and p <
0.001, THA),
function (p <
0.001, PAO and p = 0.001, THA), and total scores
(p <
0.001, PAO and p <
0.001, THA) within each cohort. There
were no significant differences in the clinical outcome scores between
the groups. Complication rates were low and similar in each cohort
(p = 0.68). Similar to THA, contemporary PAO surgery is a clinically
effective procedure that improves function and activity levels,
provides pain relief and is associated with an acceptable complication
rate. Cite this article:
We prospectively assessed the diagnostic accuracy
of the gravity stress test and clinical findings to evaluate the stability
of the ankle mortise in patients with supination–external rotation-type
fractures of the lateral malleolus without widening of the medial
clear space. The cohort included 79 patients with a mean age of
44 years (16 to 82). Two surgeons assessed medial tenderness, swelling
and ecchymosis and performed the external rotation (ER) stress test
(a reference standard). A diagnostic radiographer performed the
gravity stress test. For the gravity stress test, the positive likelihood ratio (LR)
was 5.80 with a 95% confidence interval (CI) of 2.75 to 12.27, and
the negative LR was 0.15 (95% CI 0.07 to 0.35), suggesting a moderate
change from the pre-test probability. Medial tenderness, both alone
and in combination with swelling and/or ecchymosis, indicated a
small change (positive LR, 2.74 to 3.25; negative LR, 0.38 to 0.47),
whereas swelling and ecchymosis indicated only minimal changes (positive
LR, 1.41 to 1.65; negative LR, 0.38 to 0.47). In conclusion, when gravity stress test results are in agreement
with clinical findings, the result is likely to predict stability
of the ankle mortise with an accuracy equivalent to ER stress test
results. When clinical examination suggests a medial-side injury,
however, the gravity stress test may give a false negative result. Cite this article:
Digital radiography is becoming widespread. Accurate pre-operative templating of digital images of the hip traditionally involves positioning a calibration object at its centre. This can be difficult and cause embarrassment. We have devised a method whereby a planar disc placed on the radiographic cassette accounts for the expected magnification. Initial examination of 50 pelvic CT scans showed a mean hip centre distance of 117 mm (79 to 142) above the gluteal skin. Further calculations predicted that a disc of 37.17 mm diameter, placed on the cassette, would appear identical to a 30 mm sphere placed at the level of the centre of the hip as requested by our templating software. We assessed accuracy and reproducibility by ‘reverse calibration’ of 20 radiographs taken three months after hip replacement using simultaneous sphere and disc methods, and a further 20 with a precision disc of accurate size. Even when variations in patient size were ignored, the disc proved more accurate and reliable than the sphere. The technique is reliable, robust, cost effective and acceptable to patients and radiographers. It can easily be used in any radiography department after a few simple calculations and manufacture of appropriately-sized discs.
The aim of this study was to analyse the effect of altered viewing
perspectives on the measurement of the glenopolar angle (GPA) and
the differences between these measurements made on 3D CT reconstructions
and anteroposterior (AP) scapular view radiographs. The influence of the viewing perspective on the GPA was assessed,
as were the differences in the measurements of the GPA between 3D
CT reconstructions and AP scapular view radiographs in 68 cadaveric
scapulae.Aims
Materials and Methods
We evaluated an operative technique, described
by the Exeter Hip Unit, to assist accurate introduction of the femoral
component. We assessed whether it led to a reduction in the rate
of leg-length discrepancy after total hip arthroplasty (THA). A total of 100 patients undergoing THA were studied retrospectively;
50 were undertaken using the test method and 50 using conventional
methods as a control group. The groups were matched with respect
to patient demographics and the grade of surgeon. Three observers
measured the depth of placement of the femoral component on post-operative
radiographs and measured the length of the legs. There was a strong correlation between the depth of insertion
of the femoral component and the templated depth in the test group
(R = 0.92), suggesting accuracy of the technique. The mean leg-length
discrepancy was 5.1 mm (0.6 to 21.4) pre-operatively and 1.3 mm
(0.2 to 9.3) post-operatively. There was no difference between Consultants
and Registrars as primary surgeons. Agreement between the templated
and post-operative depth of insertion was associated with reduced
post-operative leg-length discrepancy. The intra-class coefficient
was R ≥ 0.88 for all measurements, indicating high observer agreement.
The post-operative leg-length discrepancy was significantly lower
in the test group (1.3 mm) compared with the control group (6.3
mm, p <
0.001). The Exeter technique is reproducible and leads to a lower incidence
of leg-length discrepancy after THA. Cite this article:
A high radiographic inclination angle (RI) contributes
to accelerated wear and has been associated with dislocation after
total hip arthroplasty (THA). With freehand positioning of the acetabular
component there is a lack of accuracy, with a trend towards a high
radiographic inclination angle. The aim of this study was to investigate
whether the use of a digital protractor to measure the operative
inclination angle (OI) could improve the positioning of the acetabular
component in relation to a ‘safe zone’. We measured the radiographic inclination angles of 200 consecutive
uncemented primary THAs. In the first 100 the component was introduced
freehand and in the second 100 a digital protractor was used to
measure the operative inclination angle. The mean difference between the operative and the radiographic
inclination angles
(∆RI–OI) in the second cohort was 12.3° (3.8° to 19.8°). There was
a strong correlation between the circumference of the hip and ∆RI–OI.
The number of RI outliers was significantly reduced in the protractor
group (p = 0.002). Adjusting the OI, using a digital protractor and taking into
account the circumference of the patient’s hip, improves the RI
significantly (p <
0.001) and does not require additional operating
time. Cite this article:
We have developed a novel method of calculating the radiological magnification of the hip using two separate radio-opaque markers. We recruited 74 patients undergoing radiological assessment following total hip replacement. Both the new double marker and a conventional single marker were used by the radiographer at the time of x-ray. The predicted magnification according to each marker was calculated, as was the true radiological magnification of the components. The correlation between true and predicted magnification was good using the double marker (r = 0.90, n = 74, p <
0.001), but only moderate for the single marker (r = 0.50, n = 63, p <
0.001). The median error was significantly less for the double marker than for the single (1.1% The double marker method appears to be superior to the single marker method when used in the clinical environment.
We hypothesised that the removal of the subchondral
bone plate (SCBP) for cemented acetabular component fixation in
total hip arthroplasty (THA) offers advantages over retention by
improving the cement-bone interface, without jeopardising implant
stability. We have previously published two-year follow-up data
of a randomised controlled trial (RCT), in which 50 patients with
primary osteoarthritis were randomised to either retention or removal
of the SCBP. The mean age of the retention group (n = 25, 13 males)
was 70.0 years ( Cite this article:
A total of 20 pairs of fresh-frozen cadaver femurs were assigned to four alignment groups consisting of relative varus (10° and 20°) and relative valgus (10° and 20°), 75 composite femurs of two neck geometries were also used. In both the cadaver and the composite femurs, placing the component in 20° of valgus resulted in a significant increase in load to failure. Placing the component in 10° of valgus had no appreciable effect on increasing the load to failure except in the composite femurs with varus native femoral necks. Specimens in 10° of varus were significantly weaker than the neutrally-aligned specimens. The results suggest that retention of the intact proximal femoral strength occurs at an implant angulation of ≥ 142°. However, the benefit of extreme valgus alignment may be outweighed in clinical practice by the risk of superior femoral neck notching, which was avoided in this study.
Metatarsus primus varus deformity correction
is one of the main objectives in hallux valgus surgery. A ‘syndesmosis’
procedure may be used to correct hallux valgus. An osteotomy is
not involved. The aim is to realign the first metatarsal using soft
tissues and a cerclage wire around the necks of the first and second
metatarsals. We have retrospectively assessed 27 patients (54 feet) using
the American Orthopaedic Foot and Ankle Society (AOFAS) score, radiographs
and measurements of the plantar pressures after bilateral syndesmosis
procedures. There were 26 women. The mean age of the patients was
46 years (18 to 70) and the mean follow-up was 26.4 months (24 to
33.4). Matched-pair comparisons of the AOFAS scores, the radiological
parameters and the plantar pressure measurements were conducted
pre- and post-operatively, with the mean of the left and right feet.
The mean AOFAS score improved from 62.8 to 94.4 points (p <
0.001).
Significant differences were found on all radiological parameters
(p <
0.001). The mean hallux valgus and first intermetatarsal
angles were reduced from 33.2° (24.3° to 49.8°) to 19.1° (10.1°
to 45.3°) (p <
0.001) and from 15.0° (10.2° to 18.6°) to 7.2°
(4.2° to 11.4°) (p <
0.001) respectively. The mean medial sesamoid
position changed from 6.3(4.5 to 7) to 3.6 (2 to 7) (p <
0.001)
according to the Hardy’s scale (0 to 7). The mean maximum force
and the force–time integral under the hallux region were significantly
increased by 71.1% (p = 0.001), (20.57 (0.08 to 58.3) to 35.20 (6.63
to 67.48)) and 73.4% (p = 0.014), (4.44 (0.00 to 22.74) to 7.70
(1.28 to 19.23)) respectively. The occurrence of the maximum force
under the hallux region was delayed by 11% (p = 0.02), (87.3% stance
(36.3% to 100%) to 96.8% stance (93.0% to 100%)). The force data
reflected the restoration of the function of the hallux. Three patients
suffered a stress fracture of the neck of the second metatarsal.
The short-term results of this surgical procedure for the treatment
of hallux valgus are satisfactory. Cite this article:
The ideal bearing surface for young patients
undergoing total hip replacement (THR) remains controversial. We report
the five-year results of a randomised controlled trial comparing
the clinical and radiological outcomes of 102 THRs in 91 patients
who were <
65 years of age. These patients were randomised to
receive a cobalt–chrome on ultra-high-molecular-weight polyethylene,
cobalt–chrome on highly cross-linked polyethylene, or a ceramic-on-ceramic
bearing. In all, 97 hip replacements in 87 patients were available
for review at five years. Two hips had been revised, one for infection
and one for peri-prosthetic fracture. At the final follow-up there were no significant differences
between the groups for the mean Western Ontario and McMaster Universities
osteoarthritis index (pain, p = 0.543; function, p = 0.10; stiffness,
p = 0.99), Short Form-12 (physical component, p = 0.878; mental
component, p = 0.818) or Harris hip scores (p = 0.22). Radiological
outcomes revealed no significant wear in the ceramic group. Comparison of standard and highly cross-linked polyethylene,
however, revealed an almost threefold difference in the mean annual
linear wear rates (0.151 mm/year
Digital templating in hip replacement is commonly performed with radiological markers to determine the magnification. The latter can also be determined by measuring the distance from the x-ray focal spot to the object and the distance from the x-ray focal spot to the radiological cassette or image receptor. We used post-operative radiographs of total hip replacements and hemiarthroplasties from 22 patients to calculate the magnification using both methods. The accuracy of each method was ascertained by measuring the size of the head of the implant projected on to the radiographs and comparing the result with the known size recorded in the medical records. The accuracy was found to be similar with a mean absolute measurement error of 2.6% ( We conclude that the distance measuring method is as accurate as the radiological marker method, but may avoid some of the disadvantages such as misplacement of the marker or placement outside the radiological field. It may also be more acceptable to the patient and radiographer.
Noise generation has been reported with ceramic-on-ceramic
articulations in total hip replacement (THR). This study evaluated
208 consecutive Delta Motion THRs at a mean follow-up of 21 months
(12 to 35). There were 141 women and 67 men with a mean age of 59
years (22 to 84). Patients were reviewed clinically and radiologically,
and the incidence of noise was determined using a newly described
assessment method. Noise production was examined against range of
movement, ligamentous laxity, patient-reported outcome scores, activity
level and orientation of the acetabular component. There were 143
silent hips (69%), 22 (11%) with noises other than squeaking, 17
(8%) with unreproducible squeaking and 26 (13%) with reproducible
squeaking. Hips with reproducible squeaking had a greater mean range
of movement (p <
0.001) and mean ligament laxity (p = 0.004), smaller
median head size (p = 0.01) and decreased mean acetabular component
inclination (p = 0.02) and anteversion angle (p = 0.02) compared
with the other groups. There was no relationship between squeaking
and age (p = 0.13), height (p = 0.263), weight (p = 0.333), body
mass index (p = 0.643), gender (p = 0.07) or patient outcome score
(p = 0.422). There were no revisions during follow-up. Despite the
surprisingly high incidence of squeaking, all patients remain satisfied
with their hip replacement. Cite this article:
The Cementless Oxford Unicompartmental Knee Replacement
(OUKR) was developed to address problems related to cementation,
and has been demonstrated in a randomised study to have similar
clinical outcomes with fewer radiolucencies than observed with the
cemented device. However, before its widespread use it is necessary
to clarify contraindications and assess the complications. This
requires a larger study than any previously published. We present a prospective multicentre series of 1000 cementless
OUKRs in 881 patients at a minimum follow-up of one year. All patients
had radiological assessment aligned to the bone–implant interfaces
and clinical scores. Analysis was performed at a mean of 38.2 months
(19 to 88) following surgery. A total of 17 patients died (comprising
19 knees (1.9%)), none as a result of surgery; there were no tibial
or femoral loosenings. A total of 19 knees (1.9%) had significant
implant-related complications or required revision. Implant survival
at six years was 97.2%, and there was a partial radiolucency at
the bone–implant interface in 72 knees (8.9%), with no complete radiolucencies.
There was no significant increase in complication rate compared
with cemented fixation (p = 0.87), and no specific contraindications
to cementless fixation were identified. Cementless OUKR appears to be safe and reproducible in patients
with end-stage anteromedial osteoarthritis of the knee, with radiological
evidence of improved fixation compared with previous reports using
cemented fixation. Cite this article:
We compared lower limb coronal alignment measurements
obtained pre- and post-operatively with long-leg radiographs and
computer navigation in patients undergoing primary total knee replacement
(TKR). A series of 185 patients had their pre- and post-implant
radiological and computer-navigation system measurements of coronal alignment
compared using the Bland-Altman method. The study included 81 men
and 104 women with a mean age of 68.5 years (32 to 87) and a mean
body mass index of 31.7 kg/m2 (19 to 49). Pre-implant
Bland–Altman limits of agreement were -9.4° to 8.6° with a repeatability
coefficient of 9.0°. The Bland–Altman plot showed a tendency for the
radiological measurement to indicate a higher level of pre-operative
deformity than the corresponding navigation measurement. Post-implant
limits of agreement were -5.0° to 5.4° with a repeatability coefficient
of 5.2°. The tendency for valgus knees to have greater deformity
on the radiograph was still seen, but was weaker for varus knees. The alignment seen or measured intra-operatively during TKR is
not necessarily the same as the deformity seen on a standing long-leg
radiograph either pre- or post-operatively. Further investigation
into the effect of weight-bearing and surgical exposure of the joint
on the mechanical femorotibial angle is required to enable the most appropriate
intra-operative alignment to be selected.
The computed neck-shaft angle and the size of the femoral component were recorded in 100 consecutive hip resurfacings using imageless computer-navigation and compared with the angle measured before operation and with actual component implanted. The reliability of the registration was further analysed using ten cadaver femora. The mean absolute difference between the measured and navigated neck-shaft angle was 16.3° (0° to 52°). Navigation underestimated the measured neck-shaft angle in 38 patients and the correct implant size in 11. Registration of the cadaver femora tended to overestimate the correct implant size and provided a low level of repeatability in computing the neck-shaft angle. Prudent pre-operative planning is advisable for use in conjunction with imageless navigation since misleading information may be registered intraoperatively, which could lead to inappropriate sizing and positioning of the femoral component in hip resurfacing.
Radiological assessment of total and unicompartmental
knee replacement remains an essential part of routine care and follow-up.
Appreciation of the various measurements that can be identified
radiologically is important. It is likely that routine plain radiographs
will continue to be used, although there has been a trend towards
using newer technologies such as CT, especially in a failing knee,
where it provides more detailed information, albeit with a higher
radiation exposure. The purpose of this paper is to outline the radiological parameters
used to evaluate knee replacements, describe how these are measured
or classified, and review the current literature to determine their
efficacy where possible.
There has been debate about the role of unicompartmental knee replacement in the presence of radiologically identifiable degenerative changes of the patellofemoral joint. We studied 195 knees in 163 patients in whom an Oxford unicompartmental knee replacement had been performed for medial osteoarthritis between January 2004 and July 2007. The mean age of the patients was 66 years (51 to 93). The degree of degenerative change of the patellofemoral joint was assessed using Jones’ criteria. Functional outcome was assessed at a mean of 3.4 years (2 to 7) post-operatively, using the Oxford knee score and the Short-form 12 score. Degenerative changes of the patellofemoral joint were seen pre-operatively in 125 knees (64%) on the skyline radiographs. There was no statistically significant difference in the Oxford knee or Short-form 12 scores between those patients who had patellofemoral osteoarthritis pre-operatively and those who did not (p = 0.22 and 0.54, respectively). These results support the opinion expressed at the designer’s hospital that degenerative changes of the patellofemoral joint in isolation should not be considered to be a contraindication to medial Oxford unicompartmental knee replacement.
Lengthening of the humerus is now an established
technique. We compared the complications of humeral lengthening
with those of femoral lengthening and investigated whether or not
the callus formation in the humerus proceeds at a higher rate than
that in the femur. A total of 24 humeral and 24 femoral lengthenings
were performed on 12 patients with achondroplasia. We measured the
pixel value ratio (PVR) of the lengthened area on radiographs and
each radiograph was analysed for the shape, type and density of
the callus. The quality of life (QOL) of the patients after humeral
lengthening was compared with that prior to surgery. The complication
rate per segment of humerus and femur was 0.87% and 1.37%, respectively.
In the humerus the PVR was significantly higher than that of the
femur. Lower limbs were associated with an increased incidence of
concave, lateral and central callus shapes. Humeral lengthening
had a lower complication rate than lower-limb lengthening, and QOL
increased significantly after humeral lengthening. Callus formation
in the humerus during the distraction period proceeded at a significantly
higher rate than that in the femur. These findings indicate that humeral lengthening has an important
role in the management of patients with achondroplasia.
Congenital pseudarthrosis of the tibia is an uncommon manifestation of neurofibromatosis type 1 (NF1), but one that remains difficult to treat due to anabolic deficiency and catabolic excess. Bone grafting and more recently recombinant human bone morphogenetic proteins (rhBMPs) have been identified as pro-anabolic stimuli with the potential to improve the outcome after surgery. As an additional pharmaceutical intervention, we describe the combined use of rhBMP-2 and the bisphosphonate zoledronic acid in a mouse model of NF1-deficient fracture repair. Fractures were generated in the distal tibiae of neurofibromatosis type 1-deficient ( When only rhBMP but no zoledronic acid was used to promote repair, 75% of fractures in These data support the concept that preventing bone loss in combination with anabolic stimulation may improve the outcome following surgical treatment for children with congenital pseudarthoris of the tibia and NF1.
This study used CT analysis to determine the rotational alignment of 39 painful and 26 painless fixed-bearing total knee replacements (TKRs) from a cohort of 740 NexGen Legacy posterior-stabilised and cruciate-retaining prostheses implanted between May 1996 and August 2003. The mean rotation of the tibial component was 4.3° of internal rotation (25.4° internal to 13.9° external rotation) in the painful group and 2.2° of external rotation (8.5° internal to 18.2° external rotation) in the painfree group (p = 0.024). In the painful group 17 tibial components were internally rotated more than 9° compared with none in the painfree group (p <
0.001). Additionally, six femoral components in the painful group were internally rotated more than 6° compared with none in the painfree group (p = 0.017). External rotational errors were not found to be associated with pain. Overall, 22 (56.4%) of the painful TKRs had internal rotational errors involving the femoral, the tibial or both components. It is estimated that at least 4.6% of all our TKRs have been implanted with significant internal rotational errors.
Our aim was to determine the effect of the initial pattern of fracture and the displacement of fragments on the outcome of proximal humeral fractures treated conservatively. We followed 93 consecutive patients prospectively for one year. Final movement and strength were compared with those of the contralateral side. The final American Shoulder and Elbow Society score and the Disabilities of Arm, Shoulder and Hand and Short-Form 36 questionnaires were compared with those provided by the patient on the day of the injury. Radiographs and CT scans with three-dimensional reconstruction were obtained in all patients. The pattern of the fracture and the displacement of individual fragments were analysed and correlated with the final outcome. There were two cases of nonunion and six of avascular necrosis. The majority of the fractures (84 patients; 90%) followed one of the following four patterns: posteromedial (varus) impaction in 50 patients (54%), lateral (valgus) impaction in 13 (14%), isolated greater tuberosity in 15 (16%), and anteromedial impaction fracture in six (6%). Head orientation, impaction of the surgical neck and displacement of the tuberosity correlated strongly with the outcome. In fractures with posteromedial impaction, a poor outcome was noted as the articular surface displaced inferiorly increasing its distance from the acromion. A poorer outcome was noted as a fractured greater tuberosity displaced medially overlapping with the posterior articular surface. Lateral impaction fractures had a worse outcome than other patterns of fracture.
Radiographs of 110 patients who had undergone 120 high tibial osteotomies (60 closed-wedge, 60 open-wedge) were assessed for posterior tibial slope before and after operation, and before removal of the hardware. In the closed-wedge group the mean slope was 5.7° ( Posterior tibial slope decreases after closed-wedge high tibial osteotomy and increases after an open-wedge procedure because of the geometry of the proximal tibia. The changes in the slope are stable over time, emphasising the influence of the operative procedure rather than of the implant.
Many radiographic techniques have been described for measuring patellar height. They can be divided into two groups: those that relate the position of the patella to the femur (direct) and those that relate it to the tibia (indirect). This article looks at the methods that have been described, the logic behind their conception and the critical analyses that have been performed to test them.
We have evaluated the effect of vacuum aspiration of the iliac wing on the osseointegration of cement into the acetabulum. We entered a total of 40 patients undergoing primary total hip arthroplasty into two consecutive study groups. Group 1 underwent acetabular cement pressurisation for 60 seconds before insertion of the acetabular component. Group 2 had the same pressurisation with simultaneous vacuum suction of the ilium using an iliac-wing aspirator. Standard post-operative radiographs were reviewed blindly to assess the penetration of cement into the iliac wing. Penetration was significantly greater in the group with aspiration of the iliac wing.
A cavovarus foot deformity was simulated in cadaver specimens by inserting metallic wedges of 15° and 30° dorsally into the first tarsometatarsal joint. Sensors in the ankle joint recorded static tibiotalar pressure distribution at physiological load. The peak pressure increased significantly from neutral alignment to the 30° cavus deformity, and the centre of force migrated medially. The anterior migration of the centre of force was significant for both the 15° (repeated measures analysis of variance (ANOVA), p = 0.021) and the 30° (repeated measures ANOVA, p = 0.007) cavus deformity. Differences in ligament laxity did not influence the peak pressure. These findings support the hypothesis that the cavovarus foot deformity causes an increase in anteromedial ankle joint pressure leading to anteromedial arthrosis in the long term, even in the absence of lateral hindfoot instability.
We compared the alignment of 39 total knee replacements implanted using the conventional alignment guide system with 37 implanted using a CT-based navigation system, performed by a single surgeon. The knees were evaluated using full-length weight-bearing anteroposterior radiographs, lateral radiographs and CT scans. The mean hip-knee-ankle angle, coronal femoral component angle and coronal tibial component angle were 181.8° (174.2° to 188.3°), 88.5° (84.0° to 91.8°) and 89.7° (86.3° to 95.1°), respectively for the conventional group and 180.8° (178.2° to 185.1°), 89.3° (85.8° to 92.0°) and 89.9° (88.0° to 93.0°), respectively for the navigated group. The mean sagittal femoral component angle was 85.5° (80.6° to 92.8°) for the conventional group and 89.6° (85.5° to 94.0°) for the navigated group. The mean rotational femoral and tibial component angles were −0.7° (−8.8° to 9.8°) and −3.3° (−16.8° to 5.8°) for the conventional group and −0.6° (−3.5° to 3.0°) and 0.3° (−5.3° to 7.7°) for the navigated group. The ideal angles of all alignments in the navigated group were obtained at significantly higher rates than in the conventional group. Our results demonstrated significant improvements in component positioning with a CT-based navigation system, especially with respect to rotational alignment.
External fixation of distal tibial fractures is often associated with delayed union. We have investigated whether union can be enhanced by using recombinant bone morphogenetic protein-7 (rhBMP-7). Osteoinduction with rhBMP-7 and bovine collagen was used in 20 patients with distal tibial fractures which had been treated by external fixation (BMP group). Healing of the fracture was compared with that of 20 matched patients in whom treatment was similar except that rhBMP-7 was not used. Significantly more fractures had healed by 16 (p = 0.039) and 20 weeks (p = 0.022) in the BMP group compared with the matched group. The mean time to union (p = 0.002), the duration of absence from work (p = 0.018) and the time for which external fixation was required (p = 0.037) were significantly shorter in the BMP group than in the matched group. Secondary intervention due to delayed healing was required in two patients in the BMP group and seven in the matched group. RhBMP-7 can enhance the union of distal tibial fractures treated by external fixation.
The aim of this study was to assess the accuracy
of placement of pelvic binders and to determine whether circumferential
compression at the level of the greater trochanters is the best
method of reducing a symphyseal diastasis. Patients were identified by a retrospective review of all pelvic
radiographs performed at a military hospital over a period of 30
months. We analysed any pelvic radiograph on which the buckle of
the pelvic binder was clearly visible. The patients were divided
into groups according to the position of the buckle in relation
to the greater trochanters: high, trochanteric or low. Reduction
of the symphyseal diastasis was measured in a subgroup of patients
with an open-book fracture, which consisted of an injury to the
symphysis and disruption of the posterior pelvic arch (AO/OTA 61-B/C). We identified 172 radiographs with a visible pelvic binder. Five
cases were excluded due to inadequate radiographs. In 83 (50%) the
binder was positioned at the level of the greater trochanters. A
high position was the most common site of inaccurate placement,
occurring in 65 (39%). Seventeen patients were identified as a subgroup
to assess the effect of the position of the binder on reduction
of the diastasis. The mean gap was 2.8 times greater (mean difference
22 mm) in the high group compared with the trochanteric group (p
<
0.01). Application of a pelvic binder above the level of the greater
trochanters is common and is an inadequate method of reducing pelvic
fractures and is likely to delay cardiovascular recovery in these
seriously injured patients.
Valgus high tibial osteotomy for osteoarthritis of the medial compartment of the knee can be performed using medial opening- and lateral closing-wedge techniques. The latter have been thought to offer greater initial stability. We measured and compared the stability of opening- and closing-wedge osteotomies fixed by TomoFix plates using radiostereometry in a series of 42 patients in a prospective, randomised clinical trial. There were no differences between the opening- and closing-wedge groups in the time to regain knee function and full weight-bearing. Pain and knee function were significantly improved in both groups without any differences between them. All the osteotomies united within one year. Radiostereometry showed no clinically relevant movement of bone or differences between either group. Medial opening-wedge high tibial osteotomy secured by a TomoFix plate offers equal stability to a lateral closing-wedge technique. Both give excellent initial stability and provide significantly improved knee function and reduction in pain, although the opening-wedge technique was more likely to produce the intended correction.