We present the clinical results and survivorship
of consecutive 100 Birmingham Hip Resurfacings in 90 patients at
a minimum follow-up of ten years. All procedures were carried out
by an independent surgeon who commenced a prospective study in 1998.
Patients were assessed clinically using the Western Ontario and
McMaster Universities osteoarthritis index, Short-Form 36, Harris
hip score and University of California, Los Angeles activity score. Radiological
analysis was performed by an independent observer and blood metal
ion levels concentrations were measured at ten years post-operatively
in 62 patients. The median acetabular component inclination was
46.2° (34° to 59°) and anteversion 11.0° (0° to 30°). The median
chromium concentration in the unilateral group was 1.74 (0.41 to
15.23) and for the bilateral group was 2.98 (1.57 to 18.01). The
equivalent values for cobalt were 1.67 (0.54 to 20.4) and 1.88 (1.38
to 19.32). In total there were eight failures giving an overall survival
at ten years of 92% (95% confidence interval (CI) 86.7 to 97.3).
BHR in male patients had an improved survivorship of 94.6% (95%
CI 89.4 to 100) compared with females at 84.6% (95% CI 70.7 to 98.5),
but this did not reach statistical significance (p = 0.119). Four
of the nine BHRs with a bearing diameter of 42 mm failed. The overall results were consistent with data produced from other
centres in that the clinical outcome of large male patients was
extremely encouraging, whereas the survival of the smaller joints
was less satisfactory.
Aims. The Birmingham Hip
Aims. The Birmingham Hip
Aims. The first aim of this study was to evaluate whether preoperative renal function is associated with postoperative changes in whole blood levels of metal ions in patients who have undergone a Birmingham Hip
Aims. It is not known whether change in patient-reported outcome measures (PROMs) over time can be predicted by factors present at surgery, or early follow-up. The aim of this study was to identify factors associated with changes in PROM status between two-year evaluation and medium-term follow-up. Patients and Methods. Patients undergoing Birmingham Hip
Elevated blood cobalt levels secondary to metal-on-metal (MoM) hip arthroplasties are a suggested risk factor for developing cardiovascular complications including cardiomyopathy. Clinical studies assessing patients with MoM hips using left ventricular ejection fraction (LVEF) have found conflicting evidence of cobalt-induced cardiomyopathy. Global longitudinal strain (GLS) is an echocardiography measurement known to be more sensitive than LVEF when diagnosing early cardiomyopathies. The extent of cardiovascular injury, as measured by GLS, in patients with elevated blood cobalt levels has not previously been examined. A total of 16 patients with documented blood cobalt ion levels above 13 µg/l (13 ppb, 221 nmol/l) were identified from a regional arthroplasty database. They were matched with eight patients awaiting hip arthroplasty. All patients underwent echocardiography, including GLS, investigating potential signs of cardiomyopathy.Aims
Methods
It is accepted that resurfacing hip replacement
preserves the bone mineral density (BMD) of the femur better than total
hip replacement (THR). However, no studies have investigated any
possible difference on the acetabular side. Between April 2007 and March 2009, 39 patients were randomised
into two groups to receive either a resurfacing or a THR and were
followed for two years. One patient’s resurfacing subsequently failed,
leaving 19 patients in each group. Resurfaced replacements maintained proximal femoral BMD and,
compared with THR, had an increased bone mineral density in Gruen
zones 2, 3, 6, and particularly zone 7, with a gain of 7.5% (95%
confidence interval (CI) 2.6 to 12.5) compared with a loss of 14.6%
(95% CI 7.6 to 21.6).
We review the history and literature of hip resurfacing arthroplasty.
The Articular Surface Replacement (ASR) hip resurfacing arthroplasty has a failure rate of 12.0% at five years, compared with 4.3% for the Birmingham Hip
The early designs of hip resurfacing implants suffered high rates of early failure, making it impossible to obtain valuable mid-term radiostereophotogrammetric (RSA) results. The metal-on-metal Birmingham Hip
A cadaver study using six pairs of lower limbs was conducted to investigate the accuracy of computer navigation and standard instrumentation for the placement of the Birmingham Hip
The dysplasia cup, which was devised as an adjunct to the Birmingham Hip
Few independent studies have reported the outcome of resurfacing arthroplasty of the hip. The aim of this study was to report the five-year clinical outcome and seven-year survival of an independent series. A total of 610 Birmingham Hip
The hip’s capsular ligaments passively restrain extreme range of movement (ROM) by wrapping around the native femoral head/neck. We determined the effect of hip resurfacing arthroplasty (HRA), dual-mobility total hip arthroplasty (DM-THA), conventional THA, and surgical approach on ligament function. Eight paired cadaveric hip joints were skeletonized but retained the hip capsule. Capsular ROM restraint during controlled internal rotation (IR) and external rotation (ER) was measured before and after HRA, DM-THA, and conventional THA, with a posterior (right hips) and anterior capsulotomy (left hips).Aims
Materials and Methods
The aim of this study was to assess the functional gain achieved following hip resurfacing arthroplasty (HRA). A total of 28 patients (23 male, five female; mean age, 56 years (25 to 73)) awaiting Birmingham HRA volunteered for this prospective gait study, with an age-matched control group of 26 healthy adults (16 male, ten female; mean age, 56 years (33 to 84)). The Oxford Hip Score (OHS) and gait analysis using an instrumented treadmill were used preoperatively and more than two years postoperatively to measure the functional change attributable to the intervention.Aims
Patients and Methods
The purpose of this study was twofold: first,
to determine whether the five-year results of hip resurfacing arthroplasty
(HRA) in Canada justified the continued use of HRA; and second,
to identify whether greater refinement of patient selection was
warranted. This was a retrospective cohort study that involved a review
of 2773 HRAs performed between January 2001 and December 2008 at
11 Canadian centres. Cox’s proportional hazards models were used
to analyse the predictors of failure of HRA. Kaplan–Meier survival
analysis was performed to predict the cumulative survival rate at
five years. The factors analysed included age, gender, body mass
index, pre-operative hip pathology, surgeon’s experience, surgical
approach, implant sizes and implant types. The most common modes
of failure were also analysed. The 2773 HRAs were undertaken in 2450 patients: 2127 in men and
646 in women. The mean age at operation was 50.5 years ( The failure rates of HRA at five years justify the ongoing use
of this technique in men. Female gender is an independent predictor
of failure, and a higher failure rate at five years in women leads
the authors to recommend this technique only in exceptional circumstances
for women. Cite this article:
The painful subluxed or dislocated hip in adults
with cerebral palsy presents a challenging problem. Prosthetic dislocation
and heterotopic ossification are particular concerns. We present
the first reported series of 19 such patients (20 hips) treated
with hip resurfacing and proximal femoral osteotomy. The pre-operative
Gross Motor Function Classification System (GMFCS) was level V in
13 (68%) patients, level IV in three (16%), level III in one (5%) and
level II in two (11%). The mean age at operation was 37 years (13
to 57). The mean follow-up was 8.0 years (2.7 to 11.6), and 16 of the
18 (89%) contactable patients or their carers felt that the surgery
had been worthwhile. Pain was relieved in 16 of the 18 surviving
hips (89%) at the last follow-up, and the GMFCS level had improved
in seven (37%) patients. There were two (10%) early dislocations;
three hips (15%) required revision of femoral fixation, and two
hips (10%) required revision, for late traumatic fracture of the
femoral neck and extra-articular impingement, respectively. Hence
there were significant surgical complications in a total of seven
hips (35%). No hips required revision for instability, and there
were no cases of heterotopic ossification. We recommend hip resurfacing with proximal femoral osteotomy
for the treatment of the painful subluxed or dislocated hip in patients
with cerebral palsy.
We report the findings of an independent review
of 230 consecutive Birmingham hip resurfacings (BHRs) in 213 patients
(230 hips) at a mean follow-up of 10.4 years (9.6 to 11.7). A total
of 11 hips underwent revision; six patients (six hips) died from
unrelated causes; and 13 patients (16 hips) were lost to follow-up.
The survival rate for the whole cohort was 94.5% (95% confidence
interval (CI) 90.1 to 96.9). The survival rate in women was 89.1%
(95% CI 79.2 to 94.4) and in men was 97.5% (95% CI 92.4 to 99.2).
Women were 1.4 times more likely to suffer failure than men. For
each millimetre increase in component size there was a 19% lower
chance of a failure. The mean Oxford hip score was 45.0 (median
47.0, 28 to 48); mean University of California, Los Angeles activity
score was 7.4 (median 8.0, 3 to 9); mean patient satisfaction score
was 1.4 (median 1.0, 0 to 9). A total of eight hips had lysis in
the femoral neck and two hips had acetabular lysis. One hip had
progressive radiological changes around the peg of the femoral component.
There was no evidence of progressive neck narrowing between five
and ten years. Our results confirm that BHR provides good functional outcome
and durability for men, at a mean follow-up of ten years. We are
now reluctant to undertake hip resurfacing in women with this implant.
In this prospective study a total of 80 consecutive
Chinese patients with Crowe type I or II developmental dysplasia of
the hip were randomly assigned for hip resurfacing arthroplasty
(HRA) or total hip replacement (THR). Three patients assigned to HRA were converted to THR, and three
HRA patients and two THR patients were lost to follow-up. This left
a total of 34 patients (37 hips) who underwent HRA and 38 (39 hips)
who underwent THR. The mean follow-up was 59.4 months (52 to 70)
in the HRA group and 60.6 months (50 to 72) in the THR group. There was
no failure of the prosthesis in either group. Flexion of the hip
was significantly better after HRA, but there was no difference
in the mean post-operative Harris hip scores between the groups.
The mean size of the acetabular component in the HRA group was significantly
larger than in the THR group (49.5 mm vs 46.1 mm, p = 0.001). There was
no difference in the mean abduction angle of the acetabular component
between the two groups. Although the patients in this series had risk factors for failure
after HRA, such as low body weight, small femoral heads and dysplasia,
the clinical results of resurfacing in those with Crowe type I or
II hip dysplasia were satisfactory. Patients in the HRA group had
a better range of movement, although neck-cup impingement was observed.
However, more acetabular bone was sacrificed in HRA patients, and
it is unclear whether this will have an adverse effect in the long
term.
Patellofemoral complications are common after
total knee replacement (TKR). Leaving the patellar unsurfaced after
TKR may lead to complications such as anterior knee pain, and re-operation
to surface it. Complications after patellar resurfacing include
patellar fracture, aseptic loosening, patellar instability, polyethylene
wear, patellar clunk and osteonecrosis. Historically, patellar complications
account for one of the larger proportions of causes of failure in
TKR, however, with contemporary implant designs, complication rates
have decreased. Most remaining failures relate to patellofemoral
tracking. Understanding the causes of patellofemoral maltracking
is essential to prevent these complications as well as manage them
when they occur. Cite this article:
A 34-year-old woman with a benign form of osteopetrosis developed osteoarthritis of the hip. In order to avoid the difficulties associated with inserting the femoral component of a conventional total hip arthroplasty, a hybrid metal-on-metal resurfacing was performed. There were several technical challenges associated with the procedure, including the sizing of the component, press-fit fixation of the acetabular component and femoral head preparation, as well as trying to avoid a fracture. No surgical complication occurred. After more than a year following surgery, the patient showed excellent clinical function and remained satisfied with the outcome. We conclude that the hybrid metal-on-metal resurfacing arthroplasty represents a valuable option for the treatment of patients with osteopetrosis and secondary hip osteoarthritis.
We performed 96 Birmingham resurfacing arthroplasties of the hip in 71 consecutive patients with avascular necrosis of the femoral head. A modified neck-capsule-preserving approach was used which is described in detail. The University of California, Los Angeles outcome score, the radiological parameters and survival rates were assessed. The mean follow-up was for 5.4 years (4.0 to 8.1). All the patients remained active with a mean University of California, Los Angeles activity score of 6.86 (6 to 9). Three hips failed, giving a cumulative survival rate of 95.4%. With failure of the femoral component as the endpoint, the cumulative survival rate was 98.0%. We also describe the combined abduction-valgus angle of the bearing couple, which is the sum of the inclination angle of the acetabular component and the stem-shaft angle, as an index of the optimum positioning of the components in the coronal plane. Using a modified surgical technique, it is possible to preserve the femoral head in avascular necrosis by performing hip resurfacing in patients with good results.
We have evaluated the clinical effectiveness
of a metal resurfacing inlay implant for osteochondral defects of
the medial talar dome after failed previous surgical treatment.
We prospectively studied 20 consecutive patients with a mean age
of 38 years (20 to 60), for a mean of three years (2 to 5) post-surgery.
There was statistically significant reduction of pain in each of
four situations (i.e., rest, walking, stair climbing and running;
p ≤ 0.01). The median American Orthopaedic Foot and Ankle Society
ankle-hindfoot score improved from 62 (interquartile range (IQR)
46 to 72) pre-operatively to 87 (IQR 75 to 95) at final follow-up
(p <
0.001). The Foot and Ankle Outcome Score improved on all
subscales (p ≤ 0.03). The mean Short-Form 36 physical component
scale improved from 36 (23 to 50) pre-operatively to 45 (29 to 55)
at final follow-up (p = 0.001); the mental component scale did not
change significantly. On radiographs, progressive degenerative changes
of the opposing tibial plafond were observed in two patients. One
patient required additional surgery for the osteochondral defect.
This study shows that a metal implant is a promising treatment for
osteochondral defects of the medial talar dome after failed previous
surgery. Cite this article:
The aim of this study was to establish the natural
course of unrevised asymptomatic pseudotumours after metal-on-metal
(MoM) hip resurfacing during a six- to 12-month follow-up period.
We used repeated metal artefact reduction sequence (MARS)-magnetic
resonance imaging (MRI), serum metal ion analysis and clinical examination to
study 14 unrevised hips (mean patient age 52.7 years, 46 to 68,
5 female, 7 male) with a pseudotumour and 23 hips (mean patient
age 52.8 years, 38 to 69, 7 female, 16 male) without a pseudotumour.
The mean post-operative time to the first MARS-MRI scan was 4.3 years
(2.2 to 8.3), and mean time between the first and second MARS-MRI scan
was eight months (6 to 12). At the second MRI scan, the grade of
severity of the pseudotumour had not changed in 35 hips. One new
pseudotumour (Anderson C2 score, moderate) was observed, and one
pseudotumour was downgraded from C2 (moderate) to C1 (mild). In
general, the characteristics of the pseudotumours hardly changed. Repeated MARS-MRI scans within one year in patients with asymptomatic
pseudotumours after MoM hip resurfacing showed little or no variation.
In 23 patients without pseudotumour, one new asymptomatic pseudotumour
was detected. This is the first longitudinal study on the natural history of
pseudotumours using MARS-MRI scans in hip resurfacing, and mirrors
recent results for 28 mm diameter MoM total hip replacement. Cite this article:
A total of 219 hips in 192 patients aged between
18 and 65 years were randomised to 28-mm metal-on-metal uncemented
total hip replacements (THRs, 107 hips) or hybrid hip resurfacing
(HR, 112 hips). At a mean follow-up of eight years (6.6 to 9.3)
there was no significant difference between the THR and HR groups
regarding rate of revision (4.0% (4 of 99) Cite this article:
Radiostereometric analysis (RSA) can detect early
micromovement in unstable implant designs which are likely subsequently
to have a high failure rate. In 2010, the Articular Surface Replacement
(ASR) was withdrawn because of a high failure rate. In 19 ASR femoral
components, the mean micromovement over the first two years after implantation
was 0.107 mm ( We conclude that the ASR femoral component achieves initial stability
and that early migration is not the mode of failure for this resurfacing
arthroplasty.
This systematic review of the literature summarises
the clinical experience with ceramic-on-ceramic hip bearings over
the past 40 years and discusses the concerns that exist in relation
to the bearing combination. Loosening, fracture, liner chipping
on insertion, liner canting and dissociation, edge-loading and squeaking
have all been reported, and the relationship between these issues
and implant design and surgical technique is investigated. New design
concepts are introduced and analysed with respect to previous clinical
experience.
We studied 40 patients in whom the patella was not severely deformed and who were undergoing primary total knee arthroplasty (TKA) for osteoarthritis by one surgeon using one type of prosthesis. They were randomly allocated either to have the patella retained or resurfaced with a cemented, all-polyethylene component regardless of the state of the patellar articular cartilage. Apart from removal of osteophytes, no surgery was undertaken on the retained patellae. All 38 surviving patients were evaluated at three years using the HSS knee score and a new, specifically designed Patellar score (maximum score of 30). No TKA was revised, but two patients in the resurfacing group had a further unrelated procedure. The mean HSS and Patellar scores at follow-up were 89 and 28 in the patellar retention group and 83 and 26 in the patellar resurfacing group. Statistically significant lower scores for both were recorded in women and in heavier patients. Stair-climbing ability was significantly better in the retention group. Although there were no complications related to patellar resurfacing, in the medium term we did not find any significant benefit from resurfacing the patella during TKA for osteoarthritis if it was not severely deformed.
We describe a patient who sustained a peri-prosthetic fracture of the femoral neck two weeks after undergoing a resurfacing procedure for osteoarthritis of the hip. The fracture was minimally displaced and a satisfactory outcome was obtained following conservative treatment.
Aims. The aim of this study was to analyze the results of two radiocapitellar prostheses in a large case series followed prospectively, with medium-term follow-up. Patients and Methods. A total of 31 patients with a mean age of 54 years (27 to 73) were analyzed; nine had primary osteoarthritis (OA) and 17 had post-traumatic OA, three had capitellar osteonecrosis, and two had a fracture. Overall, 17 Lateral
Aims. The aim of this study was to report the implant survival and patient-reported outcome measures (PROMs) in a consecutive series of patients aged less than 50 years at the time of arthroplasty using the Birmingham Hip
We sought to establish the incidence of joint failure secondary to adverse reaction to metal debris (ARMD) following metal-on-metal hip resurfacing in a large, three surgeon, multicentre study involving 4226 hips with a follow-up of 10 to 142 months. Three implants were studied: the Articular Surface Replacement; the Birmingham Hip
Aims. This study presents the long-term survivorship, risk factors for prosthesis survival, and an assessment of the long-term effects of changes in surgical technique in a large series of patients treated by metal-on-metal (MoM) hip resurfacing arthroplasty (HRA). Patients and Methods. Between November 1996 and January 2012, 1074 patients (1321 hips) underwent HRA using the Conserve Plus Hip
Metal and ceramic humeral head bearing surfaces are available choices in anatomical shoulder arthroplasties. Wear studies have shown superior performance of ceramic heads, however comparison of clinical outcomes according to bearing surface in total shoulder arthroplasty (TSA) and hemiarthroplasty (HA) is limited. This study aimed to compare the rates of revision and reoperation following metal and ceramic humeral head TSA and HA using data from the National Joint Registry (NJR), which collects data from England, Wales, Northern Ireland, Isle of Man and the States of Guernsey. NJR shoulder arthroplasty records were linked to Hospital Episode Statistics and the National Mortality Register. TSA and HA performed for osteoarthritis (OA) in patients with an intact rotator cuff were included. Metal and ceramic humeral head prostheses were matched within separate TSA and HA groups using propensity scores based on 12 and 11 characteristics, respectively. The primary outcome was time to first revision and the secondary outcome was non-revision reoperation.Aims
Methods
Aims. We investigated whether blood metal ion levels could effectively
identify patients with bilateral Birmingham Hip
Modern metal-on-metal hip resurfacing has been
widely performed in the United Kingdom for over a decade. However,
the literature reports conflicting views of the benefits: excellent
medium- to long-term results with some brands in specific subgroups,
but high failure rates and local soft-tissue reactions in others.
The National Joint Registry for England and Wales (NJR) has collected
data on all hip resurfacings performed since 2003. This retrospective
cohort study recorded survival time to revision from a resurfacing
procedure, exploring risk factors independently associated with
failure. All patients with a primary diagnosis of osteoarthritis
who underwent resurfacing between 2003 and 2010 were included in
the analyses. Cox’s proportional hazard models were used to analyse
the extent to which the risk of revision was related to patient,
surgeon and implant covariates. A total of 27 971 hip resurfacings were performed during the
study period, of which 1003 (3.59%) underwent revision surgery.
In the final adjusted model, we found that women were at greater
risk of revision than men (hazard ratio (HR) = 1.30, p = 0.007),
but the risk of revision was independent of age. Of the implant-specific
predictors, five brands had a significantly greater risk of revision
than the Birmingham Hip
In this systematic review, our aim was to explore
whether or not patients are able to return to athletic activity
following lower limb joint replacement. We also investigated any evidence
as to whether participation in athletic activity post-joint replacement
increases complications and reduces implant survival. A PubMed, Embase and Sports Discus search was performed using
the MeSH terms ‘Sport’, ‘Athletic’, ‘Athlete’, ‘Physical’, ‘Activity’,
‘Arthroplasty’, ‘Total Hip Replacement’, ‘Hip
We report the long-term survival and functional
outcome of the Birmingham Hip
Mechanical impingement of the iliopsoas (IP) tendon accounts for 2% to 6% of persistent postoperative pain after total hip arthroplasty (THA). The most common initiator is anterior acetabular component protrusion, where the anterior margin is not covered by anterior acetabular wall. A CT scan can be used to identify and measure this overhang; however, no threshold exists for determining symptomatic anterior IP impingement due to overhang. A case-control study was conducted in which CT scan measurements were used to define a threshold that differentiates patients with IP impingement from asymptomatic patients after THA. We analyzed the CT scans of 622 patients (758 THAs) between May 2011 and May 2020. From this population, we identified 136 patients with symptoms suggestive of IP impingement. Among them, six were subsequently excluded: three because the diagnosis was refuted intraoperatively, and three because they had another obvious cause of impingement, leaving 130 hips (130 patients) in the study (impingement) group. They were matched to a control group of 138 asymptomatic hips (138 patients) after THA. The anterior acetabular component overhang was measured on an axial CT slice based on anatomical landmarks (orthogonal to the pelvic axis).Aims
Methods
Perthes’ disease (PD) is a childhood hip disorder that can affect the quality of life in adulthood due to femoral head deformity and osteoarthritis. There is very little data on how PD patients function as adults, especially from the patients’ perspective. The purpose of this study was to collect treatment history, demographic details, the University of California, Los Angeles activity score (UCLA), the 36-Item Short Form survey (SF-36) score, and the Hip disability and Osteoarthritis Outcome score (HOOS) of adults who had PD using a web-based survey method and to compare their outcomes to the outcomes from an age- and sex-matched normative population. The English REDCap-based survey was made available on a PD study group website. The survey included childhood and adult PD history, UCLA, SF-36, and HOOS. Of the 1,182 participants who completed the survey, the 921 participants who did not have a total hip arthroplasty are the focus of this study. The mean age at survey was 38 years (SD 12) and the mean duration from age at PD onset to survey participation was 30.8 years (SD 12.6).Aims
Methods
Professional dancers represent a unique patient population in the setting of hip arthroplasty, given the high degree of hip strength and mobility required by their profession. We sought to determine the clinical outcomes and ability to return to professional dance after total hip arthroplasty (THA) or hip resurfacing arthroplasty (HRA). Active professional dancers who underwent primary THA or HRA at a single institution with minimum one-year follow-up were included in the study. Primary outcomes included the rate of return to professional dance, three patient-reported outcome measures (PROMs) (modified Harris Hip Score (mHHS), Hip disability and Osteoarthritis Outcome Score for Joint Replacement (HOOS-JR), and Lower Extremity Activity Scale (LEAS)), and postoperative complications.Aims
Methods
Hip resurfacing is being performed more frequently in the United Kingdom. The possible benefits include more accurate restoration of leg length, femoral offset and femoral anteversion than occurs after total hip arthroplasty (THA). We compared anteroposterior radiographs from 26 patients who had undergone hybrid THA (uncemented cup/cemented stem), with 28 who had undergone Birmingham Hip
We report the outcome at a minimum of five years of 110 consecutive metal-on-metal Birmingham Hip
Early failure associated with adverse reactions to metal debris is an emerging problem after hip resurfacing but the exact mechanism is unclear. We analysed our entire series of 660 metal-on-metal resurfacings (Articular Surface Replacement (ASR) and Birmingham Hip
We describe the findings at six years in an ongoing prospective clinicoradiological and metal ion study in a cohort of 26 consecutive male patients with unilateral Birmingham Hip
There have been no large comparative studies of the blood levels of metal ions after implantation of commercially available hip resurfacing devices which have taken into account the effects of femoral size and inclination and anteversion of the acetabular component. We present the results in 90 patients with unilateral articular surface replacement (ASR) hip resurfacings (mean time to blood sampling 26 months) and 70 patients with unilateral Birmingham Hip
This study compared the demographic, clinical and patient-reported outcomes after total hip replacement (THR) and Birmingham Hip
Debridement, antibiotics, and implant retention (DAIR) is a widely accepted form of surgical treatment for patients with an early periprosthetic joint infection (PJI) after primary arthroplasty. The outcome of DAIR after revision arthroplasty, however, has not been reported. The aim of this study was to report the success rate of DAIR after revision arthroplasty with a follow-up of two years. This retrospective study, conducted at the Sint Maartenskliniek, Nijmegen, the Netherlands, included 88 patients who underwent DAIR within 90 days of revision total hip or total knee arthroplasty between 2012 and 2019. Details of the surgical procedures and PJI were collected. Univariate analysis and a subgroup analysis of the culture-positive group were performed. Kaplan-Meier survivorship curves were constructed.Aims
Methods
The main objective of this study was to examine whether the Oxford Shoulder Score (OSS) demonstrated floor or ceiling effects when used to measure outcomes following shoulder arthroplasty in a large national cohort. Secondary objectives were to assess its pain and function subscales, and to identify independent predictors for patients achieving a postoperative ceiling score following shoulder arthroplasty. Secondary database analysis of the National Joint Registry (NJR), which included 48,270 patients undergoing shoulder arthroplasty, was conducted. The primary outcome measure was the OSS. Secondary outcome measures were the OSS-Function Component Subscale and OSS-Pain Component Subscale. Floor and ceiling effects were considered to be present if > 15% of patients scored either the lowest or highest possible score. Logistic regression analysis was used to identify independent predictors for scoring the highest possible OSS score postoperatively.Aims
Methods
Total hip arthroplasties (THAs) are performed by surgeons at various stages in training with varying levels of supervision, but we do not know if this is safe practice with comparable outcomes to consultant-performed THA. Our aim was to examine the association between surgeon grade, the senior supervision of trainees, and the risk of revision following THA. We performed an observational study using National Joint Registry (NJR) data. We included adult patients who underwent primary THA for osteoarthritis, recorded in the NJR between 2003 and 2016. Exposures were operating surgeon grade (consultant or trainee) and whether or not trainees were directly supervised by a scrubbed consultant. Outcomes were all-cause revision and the indication for revision up to ten years. We used methods of survival analysis, adjusted for patient, operation, and healthcare setting factors.Aims
Methods
Knee arthroplasty surgery is a highly effective treatment for arthritis and disorders of the knee. There are a wide variety of implant brands and types of knee arthroplasty available to surgeons. As a result of a number of highly publicized failures, arthroplasty surgery is highly regulated in the UK and many other countries through national registries, introduced to monitor implant performance, surgeons, and hospitals. With time, the options available within many brand portfolios have grown, with alternative tibial or femoral components, tibial insert materials, or shapes and patella resurfacings. In this study we have investigated the effect of the expansion of implant brand portfolios and where there may be a lack of transparency around a brand name. We also aimed to establish the potential numbers of compatible implant construct combinations. Hypothetical implant brand portfolios were proposed, and the number of compatible implant construct combinations was calculated.Aims
Methods
Earlier studies dealing with trends in the management of osteonecrosis of the femoral head (ONFH) identified an increasing rate of total hip arthroplasties (THAs) and a decreasing rate of joint-preserving procedures between 1992 and 2008. In an effort to assess new trends in the management of this condition, this study evaluated the annual trends of joint-preserving versus arthroplasties for patients aged < or > 50 years old, and the incidence of specific operative management techniques. A total of 219,371 patients with ONFH were identified from a nationwide database between 1 January 2009 and 31 December 2015. The mean age was 54 years (18 to 90) and 105,298 (48%) were female. The diagnosis was made using International Classification of Disease, Ninth revision, Clinical Modification (ICD-9-CM) and Tenth Revision, Clinical Modification (ICD-10-CM) procedure codes. The percentage of patients managed using each procedure during each year was calculated and compared between years. The trends in the use of the types of procedure were also evaluated.Aims
Methods
The aim of this study was to compare robotic arm-assisted bi-unicompartmental knee arthroplasty (bi-UKA) with conventional mechanically aligned total knee arthroplasty (TKA) in order to determine the changes in the anatomy of the knee and alignment of the lower limb following surgery. An analysis of 38 patients who underwent TKA and 32 who underwent bi-UKA was performed as a secondary study from a prospective, single-centre, randomized controlled trial. CT imaging was used to measure coronal, sagittal, and axial alignment of the knee preoperatively and at three months postoperatively to determine changes in anatomy that had occurred as a result of the surgery. The hip-knee-ankle angle (HKAA) was also measured to identify any differences between the two groups.Aims
Methods
The purpose of this study is to examine six types of bearing surfaces implanted at a single institution over three decades to determine whether the reasons for revision vary among the groups and how long it takes to identify differences in survival. We considered six cohorts that included a total of 1,707 primary hips done between 1982 and 2010. These included 223 conventional polyethylene sterilized with γ irradiation in air (CPE-GA), 114 conventional polyethylene sterilized with gas plasma (CPE-GP), 116 crosslinked polyethylene (XLPE), 1,083 metal-on-metal (MOM), 90 ceramic-on-ceramic (COC), and 81 surface arthroplasties (SAs). With the exception of the COC, all other groups used cobalt-chromium (CoCr) femoral heads. The mean follow-up was 10 (0.008 to 35) years. Descriptive statistics with revisions per 100 component years (re/100 yr) and survival analysis with revision for any reason as the endpoint were used to compare bearing surfaces.Aims
Methods
The aim of this study was to use national registry database information to estimate cumulative rates and relative risk of revision due to infection after reverse shoulder arthroplasty. We included 17 730 primary shoulder arthroplasties recorded between 2004 and 2013 in The Nordic Arthroplasty Register Association (NARA) data set. With the Kaplan–Meier method, we illustrated the ten-year cumulative rates of revision due to infection and with the Cox regression model, we reported the hazard ratios as a measure of the relative risk of revision due to infection.Aims
Patients and Methods
The aim of this study was to describe temporal trends and survivorship of total hip arthroplasty (THA) in very young patients, aged ≤ 20 years. A descriptive observational study was undertaken using data from the National Joint Registry (NJR) for England, Wales, Northern Ireland and the Isle of Man between April 2003 and March 2017. All patients aged ≤ 20 years at the time of THA were included and the primary outcome was revision surgery. Descriptive statistics were used to summarize the data and Kaplan–Meier estimates calculated for the cumulative implant survival.Aims
Patients and Methods
Surface replacement arthroplasty (SRA), compared with traditional total hip arthroplasty (THA), is more expensive and carries unique concern related to metal ions production and hypersensitivity. Additionally, SRA is a more demanding procedure with a decreased margin for error compared with THA. To justify its use, SRA must demonstrate comparable component survival and some clinical advantages. We therefore performed a systematic literature review to investigate the differences in complication rates, patient-reported outcomes, stress shielding, and hip biomechanics between SRA and THA. A systematic review of the literature was completed using MEDLINE and EMBASE search engines. Inclusion criteria were level I to level III articles that reported clinical outcomes following primary SRA compared with THA. An initial search yielded 2503 potential articles for inclusion. Exclusion criteria included review articles, level IV or level V evidence, less than one year’s follow-up, and previously reported data. In total, 27 articles with 4182 patients were available to analyze.Aims
Materials and Methods
We wished to investigate the influence of metal debris exposure
on the subsequent immune response and resulting soft-tissue injury
following metal-on-metal (MoM) hip arthroplasty. Some reports have
suggested that debris generated from the head-neck taper junction
is more destructive than equivalent doses from metal bearing surfaces. We investigated the influence of the source and volume of metal
debris on chromium (Cr) and cobalt (Co) concentrations in corresponding
blood and hip synovial fluid samples and the observed agglomerated
particle sizes in excised tissues using multiple regression analysis
of prospectively collected data. A total of 199 explanted MoM hips
(177 patients; 132 hips female) were analysed to determine rates
of volumetric wear at the bearing surfaces and taper junctions. Aims
Patients and Methods
The aim of this study was to report the long-term outcome and
implant survival of the lateral resurfacing elbow (LRE) arthroplasty
in the treatment of elbow arthritis. We reviewed a consecutive series of 27 patients (30 elbows) who
underwent LRE arthroplasty between December 2005 and January 2008.
There were 15 women and 12 men, with a mean age of 61 years (25
to 82). The diagnosis was primary hypotrophic osteoarthritis (OA)
in 12 patients (14 elbows), post-traumatic osteoarthritis (PTOA)
in five (five elbows) and rheumatoid arthritis (RA) in ten patients
(11 elbows). The mean clinical outcome scores including the Mayo
Elbow Performance Score (MEPS), the American Shoulder and Elbow
Surgeons elbow score (ASES-e), the mean range of movement and the
radiological outcome were recorded at three, six and 12 months and
at a mean final follow-up of 8.3 years (7.3 to 9.4). A one sample Aim
Patients and Methods
Metal-on-metal resurfacing of the hip (MoMHR)
has enjoyed a resurgence in the last decade, but is now again in question
as a routine option for osteoarthritis of the hip. Proponents of
hip resurfacing suggest that its survival is superior to that of
conventional hip replacement (THR), and that hip resurfacing is
less invasive, is easier to revise than THR, and provides superior
functional outcomes. Our argument serves to illustrate that none
of these proposed advantages have been realised and new and unanticipated
serious complications, such as pseudotumors, have been associated
with the procedure. As such, we feel that the routine use of MoMHR
is not justified. Cite this article:
Surgeons have commonly used modular femoral heads and stems from
different manufacturers, although this is not recommended by orthopaedic
companies due to the different manufacturing processes. We compared the rate of corrosion and rate of wear at the trunnion/head
taper junction in two groups of retrieved hips; those with mixed
manufacturers (MM) and those from the same manufacturer (SM). We identified 151 retrieved hips with large-diameter cobalt-chromium
heads; 51 of two designs that had been paired with stems from different
manufacturers (MM) and 100 of seven designs paired with stems from
the same manufacturer (SM). We determined the severity of corrosion
with the Goldberg corrosion score and the volume of material loss
at the head/stem junction. We used multivariable statistical analysis
to determine if there was a significant difference between the two
groups. Aims
Materials and Methods
Hip resurfacing has been proposed as an alternative
to traditional total hip arthroplasty in young, active patients.
Much has been learned following the introduction of metal-on-metal resurfacing
devices in the 1990s. The triad of a well-designed device, implanted
accurately, in the correct patient has never been more critical
than with these implants. Following Food and Drug Administration approval in 2006, we studied
the safety and effectiveness of one hip resurfacing device (Birmingham
Hip Resurfacing) at our hospital in a large, single-surgeon series.
We report our early to mid–term results in 1333 cases followed for
a mean of 4.3 years (2 to 5.7) using a prospective, observational
registry. The mean patient age was 53.1 years (12 to 84); 70% were
male and 91% had osteoarthritis. Complications were few, including
no dislocations, no femoral component loosening, two femoral neck
fractures (0.15%), one socket loosening (0.08%), three deep infections
(0.23%), and three cases of metallosis (0.23%). There were no destructive
pseudotumours. Overall survivorship at up to 5.7 years was 99.2%. Aseptic survivorship
in males under the age of 50 was 100%. We believe this is the largest
United States series of a single surgeon using a single resurfacing
system. Cite this article:
Currently, there is little information about the need for peri-operative
blood transfusion in patients undergoing shoulder arthroplasty. The purpose of this study was to identify the rate of transfusion
and its predisposing factors, and to establish a blood conservation
strategy. We identified all patients who had undergone shoulder arthroplasty
at our hospital between 1 January 2011 and 31 December 2013. The
rate of transfusion was determined from the patient’s records. While
there were exceptions, patients typically underwent transfusion
if they had a level of haemoglobin of <
7.5 g/dl if asymptomatic,
<
9.0 g/dl if they had a significant cardiac history or symptoms
of dizziness or light headedness. Multivariable regression analysis was undertaken to identify
predictors of transfusion. High- and low-risk cohorts for transfusion
were identified from a receiver operating characteristic (ROC) curve.Aims
Methods
To investigate whether elective joint arthroplasty performed
at the weekend is associated with a different 30-day mortality We examined the 30-day cumulative mortality rate (Kaplan-Meier)
for all elective hip and knee arthroplasties performed in England
and Wales between 1st April 2003 and 31st December 2014, comprising
118 096 episodes undertaken at the weekend and 1 233 882 episodes
performed on a weekday. We used Cox proportional-hazards regression
models to assess for time-dependent variation and adjusted for identified
risk factors for mortality.Aims
Patients and Methods
A contact patch to rim (CPR) distance of <
10 mm has been
associated with edge-loading and excessive wear. However, not all
arthroplasties with a low CPR distance show problems with wear.
Therefore, CPR distance may not be the only variable affecting the
post-operative metal ion concentrations. We used multiple logistic regression to determine what variables
differed between the patients who had high and low cobalt (CoS)
and chromium (CrS) serum ion concentrations within a cohort of patients
with low (<
10 mm) CPR distances. A total of 56 patients treated
with unilateral hip resurfacing arthroplasty (HRA) had CoS and CrS
ion studies performed more than one year after surgery. The mean
age of the patients at the time of surgery was 51.7 years (29 to
70), with 38 women (68%) and 18 men (32%).Aims
Patients and Methods
The aims of this study were to describe the prevalence of previous
lumbar surgery in patients who undergo total hip arthroplasty (THA)
and to investigate their patient-reported outcomes (PROMs) one year
post-operatively. Data from the Swedish Hip Arthroplasty Register and the Swedish
Spine Register gathered from 2002 to 2013 were merged to identify
a group of patients who had undergone lumbar surgery before THA
(n = 997) and a carefully matched one-to-one control group. We investigated
differences in the one-year post-operative PROMs between the groups.
Linear regression analyses were used to explore the associations
between previous lumbar surgery and these PROMs following THA. The
prevalence of prior lumbar surgery was calculated as the ratio of
patients identified with previous lumbar surgery between 2002 and
2012, and divided by the total number of patients who underwent
a THA in 2012.Aims
Patients and Methods
The aims of this study were to compare the diagnostic test characteristics
of ultrasound alone, metal artefact reduction sequence MRI (MARS-MRI)
alone, and ultrasound combined with MARS-MRI for identifying intra-operative
pseudotumours in metal-on-metal hip resurfacing (MoMHR) patients
undergoing revision surgery. This retrospective diagnostic accuracy study involved 39 patients
(40 MoMHRs). The time between imaging modalities was a mean of 14.6
days (0 to 90), with imaging performed at a mean of 5.3 months (0.06
to 12) before revision. The prevalence of intra-operative pseudotumours
was 82.5% (n = 33).Aims
Methods
Recent events have highlighted the importance
of implant design for survival and wear-related complications following
metal-on-metal hip resurfacing arthroplasty. The mid-term survival
of the most widely used implant, the Birmingham Hip Resurfacing
(BHR), has been described by its designers. The aim of this study
was to report the ten-year survival and patient-reported functional
outcome of the BHR from an independent centre. In this cohort of 554 patients (646 BHRs) with a mean age of
51.9 years (16.5 to 81.5) followed for a mean of eight years (1
to 12), the survival and patient-reported functional outcome depended
on gender and the size of the implant. In female hips (n = 267)
the ten-year survival was 74% (95% confidence interval (CI) 83 to
91), the ten-year revision rate for pseudotumour was 7%, the mean
Oxford hip score (OHS) was 43 ( This study supports the ongoing use of resurfacing in young active
men, who are a subgroup of patients who tend to have problems with
conventional THR. In contrast, the results in women have been poor
and we do not recommend metal-on-metal resurfacing in women. Continuous
follow-up is recommended because of the increasing incidence of
pseudotumour with the passage of time.
A retrospective study was conducted to investigate
the changes in metal ion levels in a consecutive series of Birmingham
Hip Resurfacings (BHRs) at a minimum ten-year follow-up. We reviewed
250 BHRs implanted in 232 patients between 1998 and 2001. Implant
survival, clinical outcome (Harris hip score), radiographs and serum chromium
(Cr) and cobalt (Co) ion levels were assessed. Of 232 patients, 18 were dead (five bilateral BHRs), 15 lost
to follow-up and ten had been revised. The remaining 202 BHRs in
190 patients (136 men and 54 women; mean age at surgery 50.5 years
(17 to 76)) were evaluated at a minimum follow-up of ten years (mean
10.8 years (10 to 13.6)). The overall implant survival at 13.2 years
was 92.4% (95% confidence interval 90.8 to 94.0). The mean Harris
hip score was 97.7 (median 100; 65 to 100). Median and mean ion
levels were low for unilateral resurfacings (Cr: median 1.3 µg/l,
mean
1.95 µg/l (<
0.5 to 16.2); Co: median 1.0 µg/l, mean 1.62 µg/l
(<
0.5 to 17.3)) and bilateral resurfacings (Cr: median 3.2 µg/l,
mean 3.46 µg/l (<
0.5 to 10.0); Co: median 2.3 µg/l, mean 2.66
µg/l (<
0.5 to 9.5)). In 80 unilateral BHRs with sequential ion
measurements, Cr and Co levels were found to decrease significantly
(p <
0.001) from the initial assessment at a median of six years
(4 to 8) to the last assessment at a median of 11 years (9 to 13),
with a mean reduction of 1.24 µg/l for Cr and 0.88 µg/l for Co.
Three female patients had a >
2.5 µg/l increase of Co ions, associated with
head sizes ≤ 50 mm, clinical symptoms and osteolysis. Overall, there
was no significant difference in change of ion levels between genders
(Cr, p = 0.845; Co, p = 0.310) or component sizes (Cr, p = 0.505;
Co, p = 0.370). Higher acetabular component inclination angles correlated
with greater change in ion levels (Cr, p = 0.013; Co, p = 0.002).
Patients with increased ion levels had lower Harris hip scores (p
= 0.038). In conclusion, in well-functioning BHRs the metal ion levels
decreased significantly at ten years. An increase >
2.5 µg/l was
associated with poor function. Cite this article:
Arthroplasty registries are important for the
surveillance of joint replacements and the evaluation of outcome. Independent
validation of registry data ensures high quality. The ability for
orthopaedic implant retrieval centres to validate registry data
is not known. We analysed data from the National Joint Registry
for England, Wales and Northern Ireland (NJR) for primary metal-on-metal
hip arthroplasties performed between 2003 and 2013. Records were
linked to the London Implant Retrieval Centre (RC) for validation.
A total of 67 045 procedures on the NJR and 782 revised pairs of
components from the RC were included. We were able to link 476 procedures
(60.9%) recorded with the RC to the NJR successfully. However, 306
procedures (39.1%) could not be linked. The outcome recorded by the
NJR (as either revised, unrevised or death) for a primary procedure
was incorrect in 79 linked cases (16.6%). The rate of registry-retrieval
linkage and correct assignment of outcome code improved over time.
The rates of error for component reference numbers on the NJR were
as follows: femoral head category number 14/229 (5.0%); femoral head
batch number 13/232 (5.3%); acetabular component category number
2/293 (0.7%) and acetabular component batch number 24/347 (6.5%). Registry-retrieval linkage provided a novel means for the validation
of data, particularly for component fields. This study suggests
that NJR reports may underestimate rates of revision for many types
of metal-on-metal hip replacement. This is topical given the increasing
scope for NJR data. We recommend a system for continuous independent
evaluation of the quality and validity of NJR data. Cite this article:
Femoroacetabular impingement (FAI) is commonly
associated with early hip arthritis. We reviewed our series of 1300
hip resurfacing procedures. More than 90% of our male patients,
with an average age of 53 years, had cam impingement lesions. In
this condition, there are anterior femoral neck osteophytes, and
a retroverted femoral head on a normally anteverted neck. It is
postulated that FAI results in collision of the anterior neck of
the femur against the rim of the acetabulum, causing damage to the
acetabular labrum and articular cartilage, resulting in osteoarthritis.
Early treatment of FAI involves arthroscopic or open removal of
bone from the anterior femoral neck, as well as repair or removal
of labral tears. However, once osteoarthritis has developed, hip
replacement or hip resurfacing is indicated. Hip resurfacing can
re-orient the head and re-shape the neck. This helps to restore
normal biomechanics to the hip, eliminate FAI, and improve range
of motion. Since many younger men with hip arthritis have FAI, and
are also considered the best candidates for hip resurfacing, it
is evident that resurfacing has a role in these patients.
Hip resurfacing arthroplasty (HRA) is an alternative to conventional
total hip arthroplasty for patients with osteonecrosis (ON) of the
femoral head. Our aim was to report the long-term outcome of HRA,
which is not currently known. Long-term survivorship, clinical scores and radiographic results
for 82 patients (99 hips) treated with HRA for ON over a period
of 18 years were reviewed retrospectively. The mean age of the 67
men and 15 women at the time of surgery was 40.8 years (14 to 64).
Patients were resurfaced regardless of the size of the osteonecrotic
lesion.Aims
Patients and Methods
Metal-on-metal hip resurfacing was performed for developmental dysplasia in 96 hips in 85 patients, 78 in women and 18 in men, with a mean age at the time of surgery of 43 years (14 to 65). These cases were matched for age, gender, operating surgeon and date of operation with a group of patients with primary osteoarthritis who had been treated by resurfacing, to provide a control group of 96 hips (93 patients). A clinical and radiological follow-up study was performed. The dysplasia group were followed for a mean of 4.4 years (2.0 to 8.5) and the osteoarthritis group for a mean of 4.5 years (2.2 to 9.4). Of the dysplasia cases, 17 (18%) were classified as Crowe grade III or IV. There were five (5.2%) revisions in the dysplasia group and none in the osteoarthritic patients. Four of the failures were due to acetabular loosening and the other sustained a fracture of the neck of femur. There was a significant difference in survival between the two groups (p = 0.02). The five-year survival was 96.7% (95% confidence interval 90.0 to 100) for the dysplasia group and 100% (95% confidence interval 100 to 100) for the osteoarthritic group. There was no significant difference in the median Oxford hip score between the two groups at any time during the study. The medium-term results of metal-on-metal hip resurfacing in all grades of developmental dysplasia are encouraging, although they are significantly worse than in a group of matched patients with osteoarthritis treated in the same manner.
The long-term functional outcome of total hip arthroplasty (THA)
performed by trainees is not known. A multicentre retrospective
study of 879 THAs was undertaken to investigate any differences
in outcome between those performed by trainee surgeons and consultants. A total of 879 patients with a mean age of 69.5 years (37 to
94) were included in the study; 584 THAs (66.4%) were undertaken
by consultants, 138 (15.7%) by junior trainees and 148 (16.8%) by
senior trainees. Patients were scored using the Harris Hip Score
(HHS) pre-operatively and at one, three, five, seven and ten years
post-operatively. Surgical outcome, complications and survival were
compared between groups. The effect of supervision was determined
by comparing supervised and unsupervised trainees. A primary univariate
analysis was used to select variables for inclusion in multivariate
analysis. Aims
Patients and Methods
Total knee arthroplasty (TKA) is a cost effective
and extremely successful operation. As longevity increases, the demand
for primary TKA will continue to rise. The success and survivorship
of TKAs are dependent on the demographics of the patient, surgical
technique and implant-related factors. Currently the risk of failure of a TKA requiring revision surgery
ten years post-operatively is 5%. The most common indications for revision include aseptic loosening
(29.8%), infection (14.8%), and pain (9.5%). Revision surgery poses
considerable clinical burdens on patients and financial burdens
on healthcare systems. We present a current concepts review on the epidemiology of failed
TKAs using data from worldwide National Joint Registries. Cite this article:
Vascular injuries during total hip arthroplasty
(THA) are rare but when they occur, have serious consequences. These have
traditionally been managed with open exploration and repair, but
more recently there has been a trend towards percutaneous endovascular
management. We performed a systematic review of the literature to assess
if this change in trend has led to an improvement in the overall
reported rates of morbidity and mortality during the last 22 years
in comparison with the reviews of the literature published previously. We found a total of 61 articles describing 138 vascular injuries
in 124 patients. Injuries because of a laceration were the most
prevalent (n = 51, 44%) and the most common presenting feature,
when recorded, was bleeding (n = 41, 53.3%). Delay in diagnosis
was associated with the type of vascular lesion (p <
0.001) and
the clinical presentation (p = 0.002). Open exploration and repair was the most common form of management,
however percutaneous endovascular intervention was used in one third
of the injuries and more constantly during the last 13 years. The main overall reported complications included death (n = 9,
7.3%), amputation (n = 2, 1.6%), and persistent ischaemia (n = 9,
7.3%). When compared with previous reviews there was a similar rate
of mortality but lower rates of amputation and permanent disability,
especially in patients managed by endovascular strategies. Cite this article:
In this paper, we will consider the current role
of metal-on-metal bearings by looking at three subtypes of MoM hip
arthroplasty separately: Hip resurfacing, large head (>
36 mm) MoM
THA and MoM THA with traditional femoral head sizes.
Fracture of the neck of the femur after resurfacing arthroplasty usually leads to failure and conversion to a total hip replacement. We describe an intracapsular fracture of the femoral neck sustained after hip resurfacing which was treated by cannulated screws, resulting in union and retention of the resurfacing implant. The result at follow-up three years later was very satisfactory with a Harris hip score of 99.
Wear of metal-on-metal bearings causes elevated levels of cobalt and chromium in blood and body fluids. Metal-on-metal bearings have two distinct wear phases. In the early phase, the wear rate is high. Later, it decreases and the bearing enters a steady-state phase. It is expected that as the wear rates decline, the level of cobalt detected in plasma will also decrease. We studied the baseline and exercise-related cobalt rise in 21 patients (13 men and eight women) with a mean age of 54 years (38 to 80) who had undergone successful hip resurfacing at a mean of 44 months (10 to 96) earlier. Our results showed that circulating baseline cobalt levels were not significantly correlated with the time since implantation (r = 0.08, p = 0.650). By contrast, the exercise-related cobalt rise was directly correlated with the inclination angle of the acetabular component (r = 0.47, p = 0.032) and inversely correlated with the time since implantation (r = −0.5, p = 0.020). Inclination of the acetabular component should be kept less than 40° to decrease the production of wear debris.
The aim of this study was to evaluate the ten-year
clinical and functional outcome of hip resurfacing and to compare it
with that of cementless hip arthroplasty in patients under the age
of 55 years. Between 1999 and 2002, 80 patients were enrolled into the study:
24 were randomised (11 to hip resurfacing, 13 to total hip arthroplasty),
18 refused hip resurfacing and chose cementless total hip arthroplasty
with a 32 mm bearing, and 38 insisted on resurfacing. The mean follow-up
for all patients was 12.1 years (10 to 14). Patients were assessed clinically and radiologically at one year,
five years and ten years. Outcome measures included EuroQol EQ5D,
Oxford, Harris hip, University of California Los Angeles and University
College Hospital functional scores. No differences were seen between the two groups in the Oxford
or Harris hip scores or in the quality of life scores. Despite a
similar aspiration to activity pre-operatively, a higher proportion
of patients with a hip resurfacing were running and involved in
sport and heavy manual labour after ten years. We found significantly higher function scores in patients who
had undergone hip resurfacing than in those with a cementless hip
arthroplasty at ten years. This suggests a functional advantage
for hip resurfacing. There were no other attendant problems. Cite this article:
Fractures of the proximal humerus can lead to malalignment of the humeral head, necrosis and post-traumatic osteoarthritis. In such cases surface replacement might be a promising option. A total of 28 shoulders with glenohumeral arthritis subsequent to a fracture underwent surface replacement arthroplasty of the humeral head in patients with a mean age of 60 years (35 to 83). On the basis of the inclination of the impacted head, post-traumatic arthritis was divided into three types: type 1, an impacted fracture of the head in an anatomical position (seven cases); type 2, a valgus impacted fracture (13 cases); type 3, a varus impacted fracture (eight cases). The outcome was measured by means of the Constant score. According to the Boileau classification of the sequelae of fractures of the proximal humerus, all 28 patients had a final result of intra-capsular category 1. The mean Constant score for the 28 shoulders increased from 23.2 points (2 to 45) pre-operatively to 55.1 points (20 to 89) at a mean of 31 months (24 to 66) post-operatively. Valgus impacted fractures had significantly better results (p <
0.039). Surface replacement arthroplasty can provide good results for patients with post-traumatic osteoarthritis of the shoulder. Their use avoids post-operative complications of the humeral shaft, such as peri-prosthetic fractures. Further surgery can be undertaken more easily as the bone stock is preserved.
The survivorship of contemporary resurfacing arthroplasty of the hip using metal-on-metal bearings is better than that of first generation designs, but short-term failures still occur. The most common reasons for failure are fracture of the femoral neck, loosening of the component, osteonecrosis of the femoral head, reaction to metal debris and malpositioning of the component. In 2008 the Australian National Joint Registry reported an inverse relationship between the size of the head component and the risk of revision in resurfacing hip arthroplasty. Hips with a femoral component size of ≤ 44 mm have a fivefold increased risk of revision than those with femoral components of ≥ 55 mm irrespective of gender. We have reviewed the literature to explore this observation and to identify possible reasons including the design of the implant, loading of the femoral neck, the orientation of the component, the production of wear debris and the effects of metal ions, penetration of cement and vascularity of the femoral head. Our conclusion is that although multifactorial, the most important contributors to failure in resurfacing arthroplasty of the hip are likely to be the design and geometry of the component and the orientation of the acetabular component.
Many different designs of total hip arthroplasty
(THA) with varying performance and cost are available. The identification
of those which are the most cost-effective could allow significant
cost-savings. We used an established Markov model to examine the
cost effectiveness of five frequently used categories of THA which differed
according to bearing surface and mode of fixation, using data from
the National Joint Registry for England and Wales. Kaplan–Meier
analyses of rates of revision for men and women were modelled with
parametric distributions. Costs of devices were provided by the
NHS Supply Chain and associated costs were taken from existing studies.
Lifetime costs, lifetime quality-adjusted-life-years (QALYs) and
the probability of a device being cost effective at a willingness
to pay £20 000/QALY were included in the models. The differences in QALYs between different categories of implant
were extremely small (<
0.0039 QALYs for men or women over the
patient’s lifetime) and differences in cost were also marginal (£2500
to £3000 in the same time period). As a result, the probability
of any particular device being the most cost effective was very
sensitive to small, plausible changes in quality of life estimates
and cost. Our results suggest that available evidence does not support
recommending a particular device on cost effectiveness grounds alone.
We would recommend that the choice of prosthesis should be determined
by the rate of revision, local costs and the preferences of the
surgeon and patient. Cite this article:
The presence of pseudotumours, which are soft-tissue masses relating to the hip, after metal-on-metal hip resurfacing arthroplasty has been associated with elevated levels of metal ions in serum, suggesting that pseudotumours occur when there is increased wear. We aimed to quantify the wear in vivo of implants revised for pseudotumours (eight) and of a control group of implants (22) revised for other reasons of failure. We found that the implant group with pseudotumours had a significantly higher rate of median linear wear of the femoral component at 8.1 μm/year (2.75 to 25.4) than the 1.79 μm/year (0.82 to 4.15; p = 0.002) of the non-pseudotumour group. For the acetabular component a significantly higher rate of median linear wear of 7.36 μm/year (1.61 to 24.9) was observed in the pseudotumour group compared with 1.28 μm/year (0.81 to 3.33, p = 0.001) in the other group. Wear of the acetabular component in the pseudotumour group always involved the edge of the implant, indicating that edge-loading had occurred. Our findings are the first direct evidence that pseudotumour is associated with increased wear at the metal-on-metal articulation. Furthermore, edge-loading with the loss of fluid-film lubrication may be an important mechanism of generation of wear in patients with a pseudotumour.
Abnormal wear of cobalt-containing metal-on-metal
joints is associated with inflammatory pseudotumours. Cobalt ions
activate human toll-like receptor 4 (TLR4), which normally responds
to bacterial lipopolysaccharide (LPS) in sepsis. Activation of TLR4
by LPS increases the expression of chemokines IL-8 and CXCL10, which
recruit leukocytes and activated T-cells, respectively. This study
was designed to determine whether cobalt induces a similar inflammatory
response to LPS by promoting the expression of IL-8 and CXCL10.
A human monocytic cell line, derived from acute monocytic leukaemia,
was treated with cobalt ions and expression of IL-8 and CXCL10 measured at
mRNA and protein levels. Cobalt-treated macrophages showed a 60-fold
increase in IL-8 mRNA, and an eightfold increase in production of
the mature chemokine (both p <
0.001); expression of the CXCL10
gene and protein was also significantly increased by cobalt (both
p <
0.001). Experiments were also performed in the presence of
CLI-095, a TLR4-specific antagonist which abrogated the cobalt-mediated
increase in IL-8 and CXCL10 expression. These findings suggest that cobalt ions induce inflammation similar
to that observed during sepsis by the simultaneous activation of
two TLR4-mediated signalling pathways. These pathways result in
increased production of IL-8 and CXCL10, and may be implicated in
pseudotumour formation following metal-on-metal replacement. Cite this article:
The aim of this study was to evaluate the functional
and oncological outcome of extracorporeally irradiated autografts
used to reconstruct the pelvis after a P1/2 internal hemipelvectomy. The study included 18 patients with a primary malignant bone
tumour of the pelvis. There were 13 males and five females with
a mean age of 24.8 years (8 to 62). Of these, seven had an osteogenic
sarcoma, six a Ewing’s sarcoma, and five a chondrosarcoma. At a
mean follow-up of 51.6 months (4 to 185), nine patients had died
with metastatic disease while nine were free from disease. Local
recurrence occurred in three patients all of whom eventually died of
their disease. Deep infection occurred in three patients and required
removal of their graft in two while the third underwent a hindquarter
amputation for extensive flap necrosis. The mean Musculoskeletal Tumor Society functional score of the
16 patients who could be followed-up for at least 12 months was
77% (50 to 90). Those 15 patients who completed the Toronto Extremity
Salvage Score questionnaire had a mean score of 71% (53 to 85). Extracorporeal irradiation and re-implantation of bone is a valid
method of reconstruction after an internal hemipelvectomy. It has
an acceptable morbidity and a functional outcome that compares favourably
with other available reconstructive techniques. Cite this article:
Blood metal ions have been widely used to investigate
metal-on-metal hip replacements, but their ability to discriminate
between well-functioning and failed hips is not known. The Medicines
and Healthcare products Regulatory Agency (MHRA) has suggested a
cut-off level of 7 parts per billion (ppb). We performed a pair-matched, case-control study to investigate
the sensitivity and specificity of blood metal ion levels for diagnosing
failure in 176 patients with a unilateral metal-on-metal hip replacement.
We recruited 88 cases with a pre-revision, unexplained failed hip
and an equal number of matching controls with a well-functioning
hip. We investigated the 7 ppb cut-off level for the maximum of
cobalt or chromium and determined optimal mathematical cut-off levels
from receiver-operating characteristic curves. The 7 ppb cut-off level for the maximum of cobalt or chromium
had a specificity of 89% and sensitivity 52% for detecting a pre-operative
unexplained failed metal on metal hip replacement. The optimal cut-off
level for the maximum of cobalt or chromium was 4.97 ppb and had
sensitivity 63% and specificity 86%. Blood metal ions had good discriminant ability to separate failed
from well-functioning hip replacements. The MHRA cut-off level of
7 ppb provides a specific test but has poor sensitivity.
We present the early clinical and radiological results of Articular Surface Replacement (ASR) resurfacings in 214 hips (192 patients) with a mean follow-up of 43 months (30 to 57). The mean age of the patients was 56 years (28 to 74) and 85 hips (40%) were in 78 women. The mean Harris hip score improved from 52 (11 to 81) to 95 (27 to 100) at two years and the mean University of California, Los Angeles activity score from 3.9 (1 to 10) to 7.4 (2 to 10) in the same period. Narrowing of the neck (to a maximum of 9%) was noted in 124 of 209 hips (60%). There were 12 revisions (5.6%) involving four (1.9%) early fractures of the femoral neck and two (0.9%) episodes of collapse of the femoral head secondary to avascular necrosis. Six patients (2.8%) had failure related to metal wear debris. The overall survival for our series was 93% (95% confidence interval 80 to 98) and 89% (95% confidence interval 82 to 96) for hips with acetabular components smaller than 56 mm in diameter. The ASR implant has a lower diametrical clearance and a subhemispherical acetabular component when compared with other more frequently implanted metal-on-metal hip resurfacings. These changes may contribute to the higher failure rate than in other series, compared with other designs. Given our poor results with the small components we are no longer implanting the smaller size.
The aim of this study was to define return to
theatre (RTT) rates for elective hip and knee replacement (HR and
KR), to describe the predictors and to show the variations in risk-adjusted
rates by surgical team and hospital using national English hospital
administrative data. We examined information on 260 206 HRs and 315 249 KRs undertaken
between April 2007 and March 2012. The 90-day RTT rates were 2.1%
for HR and 1.8% for KR. Male gender, obesity, diabetes and several
other comorbidities were associated with higher odds for both index
procedures. For HR, hip resurfacing had half the odds of cement fixation
(OR = 0.58, 95% confidence intervals (CI) 0.47 to 0.71). For KR,
unicondylar KR had half the odds of total replacement (OR = 0.49,
95% CI 0.42 to 0.56), and younger ages had higher odds (OR = 2.23,
95% CI 1.65 to 3.01) for ages <
40 years compared with ages 60
to 69 years). There were more funnel plot outliers at three standard deviations
than would be expected if variation occurred on a random basis. Hierarchical modelling showed that three-quarters of the variation
between surgeons for HR and over half the variation between surgeons
for KR are not explained by the hospital they operated at or by
available patient factors. We conclude that 90-day RTT rate may
be a useful quality indicator for orthopaedics. Cite this article:
We report a 12- to 15-year implant survival assessment
of a prospective single-surgeon series of Birmingham Hip Resurfacings
(BHRs). The earliest 1000 consecutive BHRs including 288 women (335
hips) and 598 men (665 hips) of all ages and diagnoses with no exclusions
were prospectively followed-up with postal questionnaires, of whom
the first 402 BHRs (350 patients) also had clinical and radiological
review. Mean follow-up was 13.7 years (12.3 to 15.3). In total, 59 patients
(68 hips) died 0.7 to 12.6 years following surgery from unrelated
causes. There were 38 revisions, 0.1 to 13.9 years (median 8.7)
following operation, including 17 femoral failures (1.7%) and seven
each of infections, soft-tissue reactions and other causes. With
revision for any reason as the end-point Kaplan–Meier survival analysis
showed 97.4% (95% confidence interval (CI) 96.9 to 97.9) and 95.8%
(95% CI 95.1 to 96.5) survival at ten and 15 years, respectively.
Radiological assessment showed 11 (3.5%) femoral and 13 (4.1%) acetabular
radiolucencies which were not deemed failures and one radiological
femoral failure (0.3%). Our study shows that the performance of the BHR continues to
be good at 12- to 15-year follow-up. Men have better implant survival
(98.0%; 95% CI 97.4 to 98.6) at 15 years than women (91.5%; 95%
CI 89.8 to 93.2), and women <
60 years (90.5%; 95% CI 88.3 to
92.7) fare worse than others. Hip dysplasia and osteonecrosis are
risk factors for failure. Patients under 50 years with osteoarthritis
fare best (99.4%; 95% CI 98.8 to 100 survival at 15 years), with
no failures in men in this group. Cite this article:
We retrospectively analysed concentrations of chromium and cobalt ions in samples of synovial fluid and whole blood taken from a group of 92 patients with failed current-generation metal-on-metal hip replacements. We applied acid oxidative digestion to our trace metal analysis protocol, which found significantly higher levels of metal ion concentrations in blood and synovial fluid than a non-digestive method. Patients were subcategorised by mode of failure as either ‘unexplained pain’ or ‘defined causes’. Using this classification, chromium and cobalt ion levels were present over a wider range in synovial fluid and not as strongly correlated with blood ion levels as previously reported. There was no significant difference between metal ion concentrations and manufacturer of the implant, nor femoral head size below or above 50 mm. There was a moderately positive correlation between metal ion levels and acetabular component inclination angle as measured on three-dimensional CT imaging. Our results suggest that acid digestion of samples of synovial fluid samples is necessary to determine metal ion concentrations accurately so that meaningful comparisons can be made between studies.
Lately, concerns have arisen following the use of large metal-on-metal bearings in hip replacements owing to reports of catastrophic soft-tissue reactions resulting in implant failure and associated complications. This review examines the literature and contemporary presentations on current clinical dilemmas in metal-on-metal hip replacement.