The Bone & Joint Journal
Vol. 99-B, Issue 6 | Pages 708 - 713
1 Jun 2017
Rushton PRP Siddique I Crawford R Birch N Gibson MJ Hutton MJ

The MAGnetic Expansion Control (MAGEC) system is used increasingly in the management of early-onset scoliosis. Good results have been published, but there have been recent reports identifying implant failures that may be associated with significant metallosis surrounding the implants. This article aims to present the current knowledge regarding the performance of this implant, and the potential implications and strategies that may be employed to identify and limit any problems. We urge surgeons to apply caution to patient and construct selection; engage in prospective patient registration using a spine registry; ensure close clinical monitoring until growth has ceased; and send all explanted MAGEC rods for independent analysis. The MAGEC system may be a good instrumentation system for the treatment of early-onset scoliosis. However, it is innovative and like all new technology, especially when deployed in a paediatric population, robust systems to assess long-term outcome are required to ensure that patient safety is maintained. Cite this article: Bone Joint J 2017;99-B:708–13


The Bone & Joint Journal
Vol. 98-B, Issue 9 | Pages 1240 - 1247
1 Sep 2016
Thompson W Thakar C Rolton DJ Wilson-MacDonald J Nnadi C

Aims

We undertook a prospective non-randomised radiological study to evaluate the preliminary results of using magnetically-controlled growing rods (MAGEC System, Ellipse technology) to treat children with early-onset scoliosis.

Patients and Methods

Between January 2011 and January 2015, 19 children were treated with magnetically-controlled growing rods (MCGRs) and underwent distraction at three-monthly intervals. The mean age of our cohort was 9.1 years (4 to 14) and the mean follow-up 22.4 months (5.1 to 35.2). Of the 19 children, eight underwent conversion from traditional growing rods. Whole spine radiographs were carried out pre- and post-operatively: image intensification was used during each lengthening in the outpatient department. The measurements evaluated were Cobb angle, thoracic kyphosis, proximal junctional kyphosis and spinal growth from T1 to S1.


The Bone & Joint Journal
Vol. 98-B, Issue 12 | Pages 1662 - 1667
1 Dec 2016
Teoh KH von Ruhland C Evans SL James SH Jones A Howes J Davies PR Ahuja S

Aims

We present a case series of five patients who had revision surgery following magnetic controlled growing rods (MGCR) for early onset scoliosis. Metallosis was found during revision in four out of five patients and we postulated a mechanism for rod failure based on retrieval analysis.

Patients and Methods

Retrieval analysis was performed on the seven explanted rods. The mean duration of MCGR from implantation to revision was 35 months (17 to 46). The mean age at revision was 12 years (7 to 15; four boys, one girl).