Aims. Distraction osteogenesis with intramedullary lengthening devices has undergone rapid development in the past decade with implant enhancement. In this first single-centre matched-pair analysis we focus on the comparison of treatment with the PRECICE and STRYDE intramedullary lengthening devices and aim to clarify any clinical and radiological differences. Methods. A single-centre 2:1 matched-pair retrospective analysis of 42 patients treated with the STRYDE and 82 patients treated with the PRECICE nail between May 2013 and November 2020 was conducted. Clinical and lengthening parameters were compared while focusing radiological assessment on osseous alterations related to the nail’s telescopic junction and locking bolts at four different stages. Results. Osteolysis next to the telescopic junction was observed in 31/48 segments (65%) lengthened with the STRYDE nail before implant removal compared to 1/91 segment (1%) in the PRECICE cohort. In the STRYDE cohort, osteolysis initially increased, but decreased or resolved in almost all lengthened segments (86%) after implant removal. Implant failure was observed in 9/48 STRYDE (19%) and in 8/92 PRECICE nails (9%). Breakage of the distal locking bolts was found in 5/48 STRYDE nails (10%) compared to none in the PRECICE cohort. Treatment-associated pain was generally recorded as mild and found in 30/48 patients (63%) and 39/92 (42%) in the STRYDE and PRECICE cohorts, respectively. Temporary range of motion (ROM) limitations under distraction were registered in 17/48 (35%) segments treated with the STRYDE and 35/92 segments (38%) treated with the PRECICE nail. Conclusion. Osteolysis and periosteal reaction,
Periprosthetic joint infection (PJI) is a challenging complication of any arthroplasty procedure. We reviewed our use of static antibiotic-loaded cement spacers (ABLCSs) for staged management of PJI where segmental bone loss, ligamentous instability, or soft-tissue defects necessitate a static construct. We reviewed factors contributing to their failure and techniques to avoid these complications when using ABLCSs in this context. A retrospective analysis was conducted of 94 patients undergoing first-stage revision of an infected knee prosthesis between September 2007 and January 2020 at a single institution. Radiographs and clinical records were used to assess and classify the incidence and causes of static spacer failure. Of the 94 cases, there were 19 primary total knee arthroplasties (TKAs), ten revision TKAs (varus-valgus constraint), 20 hinged TKAs, one arthrodesis (nail), one failed spacer (performed elsewhere), 21 distal femoral endoprosthetic arthroplasties, and 22 proximal tibial arthroplasties.Aims
Methods
The Motec cementless modular metal-on-metal ball-and-socket
wrist arthroplasty was implanted in 16 wrists with scaphoid nonunion
advanced collapse (SNAC; grades 3 or 4) and 14 wrists with scapholunate
advanced collapse (SLAC) in 30 patients (20 men) with severe (grades
3 or 4) post-traumatic osteoarthritis of the wrist. The mean age of
the patients was 52 years (31 to 71). All prostheses integrated
well radiologically. At a mean follow-up of 3.2 years (1.1 to 6.1)
no luxation or
The aim of this study was to explore the relationship between reason for revision total hip arthroplasty (rTHA) and outcomes in terms of patient-reported outcome measures (PROMs). We reviewed a prospective cohort of 647 patients undergoing full or partial rTHA at a single high-volume centre with a minimum of two years’ follow-up. The reasons for revision were classified as: infection; aseptic loosening; dislocation; structural failure; and painful THA for other reasons. PROMs (modified Oxford Hip Score (mOHS), EuroQol five-dimension three-level health questionnaire (EQ-5D-3L) score, and visual analogue scales for pain during rest and activity), complication rates, and failure rates were compared among the groups.Aims
Methods
This study aimed to describe the use of revision knee arthroplasty in Australia and examine changes in lifetime risk over a decade. De-identified individual-level data on all revision knee arthroplasties performed in Australia from 2007 to 2017 were obtained from the Australian Orthopaedic Association National Joint Replacement Registry. Population data and life tables were obtained from the Australian Bureau of Statistics. The lifetime risk of revision surgery was calculated for each year using a standardized formula. Separate calculations were undertaken for males and females.Aims
Methods
We aimed to compare the implant survival, complications, readmissions, and mortality of Vancouver B2 periprosthetic femoral fractures (PFFs) treated with internal fixation with that of B1 PFFs treated with internal fixation and B2 fractures treated with revision arthroplasty. We retrospectively reviewed the data of 112 PFFs, of which 47 (42%) B1 and 27 (24%) B2 PFFs were treated with internal fixation, whereas 38 (34%) B2 fractures underwent revision arthroplasty. Decision to perform internal fixation for B2 PFFs was based on specific radiological (polished femoral components, intact bone-cement interface) and clinical criteria (low-demand patient). Median follow-up was 36.4 months (24 to 60). Implant survival and mortality over time were estimated with the Kaplan-Meier method. Adverse events (measured with a modified Dindo-Clavien classification) and 90-day readmissions were additionally compared between groups.Aims
Methods
Proximal femoral endoprosthetic replacements (PFEPRs) are the most common reconstruction option for osseous defects following primary and metastatic tumour resection. This study aimed to compare the rate of implant failure between PFEPRs with monopolar and bipolar hemiarthroplasties and acetabular arthroplasties, and determine the optimum articulation for revision PFEPRs. This is a retrospective review of 233 patients who underwent PFEPR. The mean age was 54.7 years (SD 18.2), and 99 (42.5%) were male. There were 90 patients with primary bone tumours (38.6%), 122 with metastatic bone disease (52.4%), and 21 with haematological malignancy (9.0%). A total of 128 patients had monopolar (54.9%), 74 had bipolar hemiarthroplasty heads (31.8%), and 31 underwent acetabular arthroplasty (13.3%).Aims
Methods
Total hip arthroplasty (THA) patients undergoing or having a prior lumbar spine fusion (LSF) have an increased risk of mechanical complications. The aim of this registry-based, retrospective comparative cohort study is to assess the longer term survival of THA in patients who have undergone a LSF during a 17-year period (2000 to 2017). A registry-based population study was conducted on 679 patients who underwent both THA and LSF surgeries. Patients were identified from the regional arthroplasty data base and cross linked to patients with LSF from the regional hospital discharge database between 2000 and 2017. Demographic data, diagnosis leading to primary THA, primary implant survival, perioperative complications, number and causes of failure, and patients requiring revision arthroplasty were collated and compared. For comparison, data from 67,919 primary THAs performed during the same time time period were also retrieved and analyzed.Aims
Methods
Temporary epiphysiodesis (ED) is commonly applied in children and adolescents to treat leg length discrepancies (LLDs) and tall stature. Traditional Blount staples or modern two-hole plates are used in clinical practice. However, they require accurate planning, precise surgical techniques, and attentive follow-up to achieve the desired outcome without complications. This study reports the results of ED using a novel rigid staple (RigidTack) incorporating safety, as well as technical and procedural success according to the idea, development, evaluation, assessment, long-term (IDEAL) study framework. A cohort of 56 patients, including 45 unilateral EDs for LLD and 11 bilateral EDs for tall stature, were prospectively analyzed. ED was performed with 222 rigid staples with a mean follow-up of 24.4 months (8 to 49). Patients with a predicted LLD of ≥ 2 cm at skeletal maturity were included. Mean age at surgery was 12.1 years (8 to 14). Correction and complication rates including implant-associated problems, and secondary deformities as well as perioperative parameters, were recorded (IDEAL stage 2a). These results were compared to historical cohorts treated for correction of LLD with two-hole plates or Blount staples.Aims
Methods
We analyzed the long-term outcomes of patients observed over ten years after resection en bloc and reconstruction with extracorporeal irradiated autografts This retrospective study included 27 patients who underwent resection en bloc and reimplantation of an extracorporeal irradiated autograft. The mean patient age and follow-up period were 31.7 years (9 to 59) and 16.6 years (10.3 to 24.3), respectively. The most common diagnosis was osteosarcoma (n = 10), followed by chondrosarcoma (n = 6). The femur (n = 13) was the most frequently involved site, followed by the tibia (n = 7). There were inlay grafts in five patients, intercalary grafts in 15 patients, and osteoarticular grafts in seven patients. Functional outcome was evaluated with the Musculoskeletal Tumor Society (MSTS) scoring system.Aims
Patients and Methods
A lack of supporting clinical studies have been published to determine the ideal length of intramedullary nail in fixation of trochanteric fractures of the hip. Nevertheless, there has been a trend to use shorter intramedullary nails for the internal fixation of trochanteric hip fractures. Our aim was to determine if the length of nail affected the outcome. We randomized 229 patients with a trochanteric hip fracture between two implants: a ‘standard’ nail of 220 mm and a shorter nail of 175 mm, which had decreased proximal angulation (4° vs 7°) and a reduced diameter at the level of the lesser trochanter. Patients were followed up for one year by a nurse blinded to the type of implant used to determine if there were differences in mobility and pain with two nail designs. Pain was assessed on a scale of 1 (none) to 8 (severe and constant) and mobility on a scale of 1 (full mobility) to 9 (immobile).Aims
Methods
Limb salvage for pelvic sarcomas involving the acetabulum is a major surgical challenge. There remains no consensus about what is the optimum type of reconstruction after resection of the tumour. The aim of this study was to evaluate the surgical outcomes in these patients according to the methods of periacetabular reconstruction. The study involved a consecutive series of 122 patients with a periacetabular bone sarcoma who underwent limb-salvage surgery involving a custom-made prosthesis in 65 (53%), an ice-cream cone prosthesis in 21 (17%), an extracorporeal irradiated autograft in 18 (15%), and nonskeletal reconstruction in 18 (15%).Aims
Methods
There has been a significant reduction in unicompartmental knee arthroplasty (UKA) procedures recorded in Australia. This follows several national joint registry studies documenting high UKA revision rates when compared to total knee arthroplasty (TKA). With the recent introduction of robotically assisted UKA procedures, it is hoped that outcomes improve. This study examines the cumulative revision rate of UKA procedures implanted with a newly introduced robotic system and compares the results to one of the best performing non-robotically assisted UKA prostheses, as well as all other non-robotically assisted UKA procedures. Data from the Australian Orthopaedic Association National Joint Arthroplasty Registry (AOANJRR) for all UKA procedures performed for osteoarthritis (OA) between 2015 and 2018 were analyzed. Procedures using the Restoris MCK UKA prosthesis implanted using the Mako Robotic-Arm Assisted System were compared to non-robotically assisted Zimmer Unicompartmental High Flex Knee System (ZUK) UKA, a commonly used UKA with previously reported good outcomes and to all other non-robotically assisted UKA procedures using Cox proportional hazard ratios (HRs) and Kaplan-Meier estimates of survivorship.Aim
Methods
The aim of this study was to report the results of custom-made endoprostheses with extracortical plates plus or minus a short, intramedullary stem aimed at preserving the physis after resection of bone sarcomas in children. Between 2007 and 2017, 18 children aged less than 16 years old who underwent resection of bone sarcomas, leaving ≤ 5 cm of bone from the physis, and reconstruction with a custom-made endoprosthesis were reviewed. Median follow-up was 67 months (interquartile range 45 to 91). The tumours were located in the femur in 11 patients, proximal humerus in six, and proximal tibia in one.Aims
Patients and Methods
Femoral stems with exchangeable (modular) necks were introduced
to offer surgeons an increased choice when determining the version,
offset and length of the femoral neck during total hip arthroplasty
(THA). It was hoped that this would improve outcomes and reduce
complications, particularly dislocation. In 2010, the Australian Orthopaedic
Association National Joint Replacement Registry (AOANJRR) first
reported an increased rate of revision after primary THA using femoral
stems with an exchangeable neck. The aim of this study was to provide
a more comprehensive up-to-date analysis of primary THA using femoral
stems with exchangeable and fixed necks. The data included all primary THA procedures performed for osteoarthritis
(OA), reported to the AOANJRR between 01 September 1999 and 31 December
2014. There were 9289 femoral stems with an exchangeable neck and
253 165 femoral stems with a fixed neck. The characteristics of
the patients and prostheses including the bearing surface and stem/neck
metal combinations were examined using Cox proportional hazard ratios
(HRs) and Kaplan-Meier estimates of survivorship. Aims
Materials and Methods
The modified Dunn procedure has the potential to restore the anatomy in hips with severe slipped capital femoral epiphyses (SCFE). However, there is a risk of developing avascular necrosis of the femoral head (AVN). In this paper, we report on clinical outcome, radiological outcome, AVN rate and complications, and the cumulative survivorship at long-term follow-up in patients undergoing the modified Dunn procedure for severe SCFE. We performed a retrospective analysis involving 46 hips in 46 patients treated with a modified Dunn procedure for severe SCFE (slip angle > 60°) between 1999 and 2016. At nine-year-follow-up, 40 hips were available for clinical and radiological examination. Mean preoperative age was 13 years, and 14 hips (30%) presented with unstable slips. Mean preoperative slip angle was 64°. Kaplan–Meier survivorship was calculated.Aims
Patients and Methods
Extendible endoprostheses have been available for more than 30
years and have become more sophisticated with time. The latest generation
is ‘non-invasive’ and can be lengthened with an external magnetic
force. Early results have shown a worryingly high rate of complications
such as infection. This study investigates the incidence of complications
and the need for further surgery in a cohort of patients with a
non-invasive growing endoprosthesis. Between 2003 and June 2014, 50 children (51 prostheses) had a
non-invasive growing prosthesis implanted for a primary bone sarcoma.
The minimum follow-up was 24 months for those who survived. Their
mean age was 10.4 years (6 to 14). The incidence of complications
and further surgery was documented.Aims
Patients and Methods
Compared with primary total hip arthroplasty (THA), revision
surgery can be challenging. The cement-in-cement femoral revision
technique involves removing a femoral component from a well-fixed
femoral cement mantle and cementing a new stem into the original
mantle. This technique is widely used and when carried out for the
correct indications, is fast, relatively inexpensive and carries
a reduced short-term risk for the patient compared with the alternative
of removing well-fixed cement. We report the outcomes of this procedure
when two commonly used femoral stems are used. We identified 1179 cement-in-cement stem revisions involving
an Exeter or a Lubinus stem reported to the Swedish Hip Arthroplasty
Register (SHAR) between January 1999 and December 2015. Kaplan-Meier
survival analysis was performed.Aims
Patients and Methods
The use of a noninvasive growing endoprosthesis in the management
of primary bone tumours in children is well established. However,
the efficacy of such a prosthesis in those requiring a revision
procedure has yet to be established. The aim of this series was
to present our results using extendable prostheses for the revision
of previous endoprostheses. All patients who had a noninvasive growing endoprosthesis inserted
at the time of a revision procedure were identified from our database.
A total of 21 patients (seven female patients, 14 male) with a mean
age of 20.4 years (10 to 41) at the time of revision were included.
The indications for revision were mechanical failure, trauma or infection
with a residual leg-length discrepancy. The mean follow-up was 70
months (17 to 128). The mean shortening prior to revision was 44 mm
(10 to 100). Lengthening was performed in all but one patient with
a mean lengthening of 51 mm (5 to 140).Aims
Patients and Methods
Previous studies of failure mechanisms leading
to revision total knee replacement (TKR) performed between 1986 and
2000 determined that many failed early, with a disproportionate
amount accounted for by infection and implant-associated factors
including wear, loosening and instability. Since then, efforts have
been made to improve implant performance and instruct surgeons in
best practice. Recently our centre participated in a multi-centre evaluation
of 844 revision TKRs from 2010 to 2011. The purpose was to report
a detailed analysis of failure mechanisms over time and to see if
failure modes have changed over the past 10 to 15 years. Aseptic
loosening was the predominant mechanism of failure (31.2%), followed
by instability (18.7%), infection (16.2%), polyethylene wear (10.0%),
arthrofibrosis (6.9%) and malalignment (6.6%). The mean time to
failure was 5.9 years (ten days to 31 years), 35.3% of all revisions
occurred at less than two years, and 60.2% in the first five years.
With improvements in implant and polyethylene manufacture, polyethylene
wear is no longer a leading cause of failure. Early mechanisms of
failure are primarily technical errors. In addition to improving
implant longevity, industry and surgeons must work together to decrease
these technical errors. All reports on failure of TKR contain patients
with unexplained pain who not infrequently have unmet expectations.
Surgeons must work to achieve realistic patient expectations pre-operatively,