The aim of this study was to review the provision of total elbow arthroplasties (TEAs) in England, including the incidence, the characteristics of the patients and the service providers, the types of implant, and the outcomes. We analyzed the primary TEAs recorded in the National Joint Registry (NJR) between April 2012 and December 2022, with mortality data from the Civil Registration of Deaths dataset. Linkage with Hospital Episode Statistics-Admitted Patient Care (HES-APC) data provided further information not collected by the NJR. The incidences were calculated using estimations of the populations from the Office for National Statistics. The annual number of TEAs performed by surgeons and hospitals was analyzed on a national and regional basis.Aims
Methods
Arthroplasties of the elbow, including total elbow arthroplasty, radial head arthroplasty, distal humeral hemiarthroplasty, and radiocapitellar arthroplasty, are rarely undertaken. This scoping review aims to outline the current research in this area to inform the development of future research. A scoping review was undertaken adhering to the Joanna Briggs Institute guidelines using Medline, Embase, CENTRAL, and trial registries, limited to studies published between 1 January 1990 and 7 February 2021. Endnote software was used for screening and selection, and included randomized trials, non-randomized controlled trials, prospective and retrospective cohort studies, case-control studies, analytical cross-sectional studies, and case series of ten or more patients reporting the clinical outcomes of elbow arthroplasty. The results are presented as the number of types of studies, sample size, length of follow-up, clinical outcome domains and instruments used, sources of funding, and a narrative review.Aims
Methods
The aim of this study was to investigate the feasibility of application of a 3D-printed megaprosthesis with hemiarthroplasty design for defects of the distal humerus or proximal ulna following tumour resection. From June 2018 to January 2020, 13 patients with aggressive or malignant tumours involving the distal humerus (n = 8) or proximal ulna (n = 5) were treated by en bloc resection and reconstruction with a 3D-printed megaprosthesis with hemiarthroplasty, designed in our centre. In this paper, we summarize the baseline and operative data, oncological outcome, complication profiles, and functional status of these patients.Aims
Methods
Nerve transfer has become a common and often effective reconstructive strategy for proximal and complex peripheral nerve injuries of the upper limb. This case-based discussion explores the principles and potential benefits of nerve transfer surgery and offers in-depth discussion of several established and valuable techniques including: motor transfer for elbow flexion after musculocutaneous nerve injury, deltoid reanimation for axillary nerve palsy, intrinsic re-innervation following proximal ulnar nerve repair, and critical sensory recovery despite non-reconstructable median nerve lesions.Abstract
The aim of this study was to report the mid-term clinical outcome
of cemented unlinked J-alumina ceramic elbow (JACE) arthroplasties
when used in patients with rheumatoid arthritis (RA). We retrospectively reviewed 87 elbows, in 75 patients with RA,
which was replaced using a cemented JACE total elbow arthroplasty
(TEA) between August 2003 and December 2012, with a follow-up of
96%. There were 72 women and three men, with a mean age of 62 years
(35 to 79). The mean follow-up was nine years (2 to 14). The clinical condition
of each elbow before and after surgery was assessed using the Mayo
Elbow Performance Index (MEPI, 0 to 100 points). Radiographic loosening
was defined as a progressive radiolucent line of >1 mm that was
completely circumferential around the prosthesis.Aims
Patients and Methods
Aims. We assessed the long-term (more than ten-year) outcomes of the
Kudo type-5 elbow prosthesis in patients with rheumatoid arthritis
(RA). Materials and Methods. We reviewed 41 elbows (Larsen Grade IV, n = 21; Grade V, n =
20) in 31 patients with RA who had undergone a Kudo type-5 total
elbow arthroplasty (TEA) between 1994 and 2003, and had been followed
up for more than ten years. The humeral component was cementless
and the all-polyethylene
Revision total elbow arthroplasty (TEA) is often challenging.
The aim of this study was to report on the clinical and radiological
results of revision arthroplasty of the elbow with the Latitude
TEA. Between 2006 and 2010 we used the Latitude TEA for revision in
18 consecutive elbows (17 patients); mean age 53 years (28 to 80);
14 women. A Kudo TEA was revised in 15 elbows and a Souter-Strathclyde
TEA in three. Stability, range of movement (ROM), visual analogue score (VAS)
for pain and functional scores, Elbow Functional Assessment Scale
(EFAS), the Functional Rating Index of Broberg and Morrey (FRIBM)
and the Modified Andrews’ Elbow Scoring System (MAESS) were assessed
pre-operatively and at each post-operative follow-up visit (six,
12 months and biennially thereafter). Radiographs were analysed
for loosening, fractures and dislocation. The mean follow-up was
59 months (26 to 89).Aims
Patients and Methods
We describe the use of a protocol of irrigation and debridement
(I&
D) with retention of the implant for the treatment of periprosthetic
infection of a total elbow arthroplasty (TEA). This may be an attractive
alternative to staged re-implantation. Between 1990 and 2010, 23 consecutive patients were treated in
this way. Three were lost to follow-up leaving 20 patients (21 TEAs)
in the study. There were six men and 14 women. Their mean age was
58 years (23 to 76). The protocol involved: component unlinking,
irrigation and debridement (I&
D), and the introduction of antibiotic
laden cement beads; organism-specific intravenous antibiotics; repeat
I&
D and re-linkage of the implant if appropriate; long-term
oral antibiotic therapy. Aims
Patients and Methods
We review our experience of Coonrad-Morrey total elbow arthroplasty
(TEA) for fractures of the distal humerus in non-rheumatoid patients
with a minimum of ten years follow-up. TEA through a triceps splitting approach was peformed in 37 non-rheumatoid
patients for a fracture of the distal humerus between 1996 and 2004.
One patient could not be traced and 17 had died before the tenth
anniversary of their surgery. This left 19 patients with a minimum
follow-up of ten years to form the study group. Of these, 13 patients
were alive at the time of final review. The other six had died,
but after the tenth anniversary of their elbow arthroplasty. Their
clinical and radiological data were included in the study. Aims
Patients and Methods
Determining and accurately restoring the flexion-extension
axis of the elbow is essential for functional recovery after total
elbow arthroplasty (TEA). We evaluated the effect of morphological
features of the elbow on variations of alignment of the components
at TEA. Morphological and positioning variables were compared by
systematic CT scans of 22 elbows in 21 patients after TEA. There were five men and 16 women, and the mean age was 63 years
(38 to 80). The mean follow-up was 22 months (11 to 44). The anterior offset and version of the humeral components were
significantly affected by the anterior angulation of the humerus
(p = 0.052 and p = 0.004, respectively). The anterior offset and
version of the
To prevent insufficiency of the triceps after
total elbow arthroplasty, we have, since 2008, used a triceps-sparing ulnar
approach. This study evaluates the clinical results and post-operative
alignment of the prosthesis using this approach. We reviewed 25 elbows in 23 patients. There were five men and
18 women with a mean age of 69 years (54 to 83). There were 18 elbows
with rheumatoid arthritis, six with a fracture or pseudoarthrosis
and one elbow with osteoarthritis. Post-operative complications included one intra-operative fracture,
one elbow with heterotopic ossification, one transient ulnar nerve
palsy, and one elbow with skin necrosis, but no elbow was affected
by insufficiency of the triceps. Patients were followed for a mean of 42 months (24 to 77). The
mean post-operative Japanese Orthopaedic Association Elbow Score
was 90.8 (51 to 100) and the mean Mayo Elbow Performance score 93.8
(65 to 100). The mean post-operative flexion/extension of the elbow
was 135°/-8°. The Manual Muscle Testing score of the triceps was
5 in 23 elbows and 2 in two elbows (one patient). The mean alignment
of the implants examined by 3D-CT was 2.8° pronation (standard deviation
(. sd). 5.5), 0.3° valgus (. sd. 2.7), and 0.7° extension
(. sd. 3.2) for the humeral component, and 9.3° pronation
(. sd. 9.7), 0.3° valgus (. sd. 4.0), and 8.6° extension
(. sd. 3.1) for the
Medium-term results of the Discovery elbow replacement
are presented. We reviewed 51 consecutive primary Discovery total
elbow replacements (TERs) implanted in 48 patients. The mean age
of the patients was 69.2 years (49 to 92), there were 19 males and
32 females (37%:63%) The mean follow-up was 40.6 months (24 to 69).
A total of six patients were lost to follow-up. Statistically significant
improvements in range movement and Oxford Elbow Score were found
(p <
0.001). Radiolucent lines were much more common in, and
aseptic loosening was exclusive to, the humeral component. Kaplan–Meier
survivorship at five years was 92.2% (95% CI 74.5% to 96.4%) for
aseptic loosening. In four TERs, periprosthetic infection occurred
resulting in failure. A statistically significant association between
infection and increased BMI was found (p = 0.0268). Triceps failure
was more frequent after the Mayo surgical approach and TER performed
after previous trauma surgery. No failures of the implant were noted. Our comparison shows that the Discovery has early clinical results
that are similar to other semi-constrained TERs. We found continued
radiological surveillance with particular focus on humeral lucency
is warranted and has not previously been reported. Despite advances
in the design of total elbow replacement prostheses, rates of complication
remain high. Cite this article:
We determined the short-term clinical outcome
and migration within the bone of the humeral cementless component
of the Instrumented Bone Preserving (IBP) total elbow replacement
in a series of 16 patients. There were four men and 12 women with
a mean age at operation of 63 years (40 to 81). Migration was calculated
using radiostereometric analysis. There were no intra-operative
complications and no revisions. At two-year follow-up, all patients
showed a significant reduction in pain and functional improvement
of the elbow (both p <
0.001). Although ten components (63%)
showed movement or micromovement during the first six weeks, 14
(88%) were stable at one year post-operatively. Translation was
primarily found in the proximal direction (median 0.3 mm (interquartile range
(IQR) -0.09 to 0.8); the major rotational movement was an anterior
tilt (median 0.7° (IQR 0.4° to 1.6°)). One malaligned component
continued to migrate during the second year, and one component could
not be followed beyond three months because migration had caused
the markers to break off the prosthesis. This study shows promising early results for the cementless humeral
component of the IBP total elbow replacement. All patients had a
good clinical outcome, and most components stabilised within six
months of the operation. Cite this article:
We report our experience of staged revision surgery
for the treatment of infected total elbow arthroplasty (TEA). Between
1998 and 2010 a consecutive series of 33 patients (34 TEAs) underwent
a first-stage procedure with the intention to proceed to second-stage
procedure when the infection had been controlled. A single first-stage procedure
with removal of the components and cement was undertaken for 29
TEAs (85%), followed by the insertion of antibiotic-impregnated
cement beads, and five (15%) required two or more first-stage procedures.
The most common organism isolated was coagulase-negative A second-stage procedure was performed for 26 TEAs (76%); seven
patients (seven TEAs, 21%) had a functional resection arthroplasty
with antibiotic beads There were three recurrent infections (11.5%) in those patients
who underwent a second-stage procedure. The infection presented
at a mean of eight months (5 to 10) post-operatively. The mean Mayo
Elbow Performance Score (MEPS) in those who underwent a second stage
revision without recurrent infection was 81.1 (65 to 95). Staged revision surgery is successful in the treatment of patients
with an infected TEA and is associated with a low rate of recurrent
infection. However, when infection does occur, this study would
suggest that it becomes apparent within ten months of the second
stage procedure. Cite this article:
There is little information about the management
of peri-prosthetic fracture of the humerus after total shoulder replacement
(TSR). This is a retrospective review of 22 patients who underwent
a revision of their original shoulder replacement for peri-prosthetic
fracture of the humerus with bone loss and/or loose components.
There were 20 women and two men with a mean age of 75 years (61
to 90) and a mean follow-up 42 months (12 to 91): 16 of these had
undergone a previous revision TSR. Of the 22 patients, 12 were treated
with a long-stemmed humeral component that bypassed the fracture.
All their fractures united after a mean of 27 weeks (13 to 94).
Eight patients underwent resection of the proximal humerus with
endoprosthetic replacement to the level of the fracture. Two patients
were managed with a clam-shell prosthesis that retained the original
components. The mean Oxford shoulder score (OSS) of the original
TSRs before peri-prosthetic fracture was 33 (14 to 48). The mean
OSS after revision for fracture was 25 (9 to 31). Kaplan-Meier survival
using re-intervention for any reason as the endpoint was 91% (95%
confidence interval (CI) 68 to 98) and 60% (95% CI 30 to 80) at
one and five years, respectively. There were two revisions for dislocation of the humeral head,
one open reduction for modular humeral component dissociation, one
internal fixation for nonunion, one trimming of a prominent screw
and one re-cementation for aseptic loosening complicated by infection,
ultimately requiring excision arthroplasty. Two patients sustained
nerve palsies. Revision TSR after a peri-prosthetic humeral fracture associated
with bone loss and/or loose components is a salvage procedure that
can provide a stable platform for elbow and hand function. Good
rates of union can be achieved using a stem that bypasses the fracture.
There is a high rate of complications and function is not as good as
with the original replacement.
Rarely, the extent of a malignant bone tumour
may necessitate resection of the complete humerus to achieve adequate
oncological clearance. We present our experience with reconstruction
in such cases using a total humeral endoprosthesis (THER) in 20
patients (12 male and eight female) with a mean age of 22 years
(6 to 59). We assessed the complications, the oncological and functional
outcomes and implant survival. Surgery was performed between June
2001 and October 2009. The diagnosis included osteosarcoma in nine,
Ewing’s sarcoma in eight and chondrosarcoma in three. One patient
was lost to follow-up. The mean follow-up was 41 months (10 to 120)
for all patients and 56 months (25 to 120) in survivors. There were
five local recurrences (26.3%) and 11 patients were alive at time
of last follow-up, with overall survival for all patients being
52% (95% confidence interval (CI) 23.8 to 74) at five years. The
mean Musculoskeletal Tumor Society score for the survivors was 22
(73%; 16 to 23). The implant survival was 95% (95% CI 69.5 to 99.3)
at five years. The use of a THER in the treatment of malignant tumours of bone
is oncologically safe; it gives consistent and predictable results
with low rates of complication.
We report the effectiveness of revision of total
elbow replacement by re-cementing. Between 1982 and 2004, 53 elbows
in 52 patients were treated with re-cementing of a total elbow replacement
into part or all of the existing cement mantle or into the debrided
host-bone interface, without the use of structural bone augmentation
or a custom prosthesis. The original implant revision was still
In light of the growing number of elderly osteopenic
patients with distal humeral fractures, we discuss the history of
their management and current trends. Under most circumstances operative
fixation and early mobilisation is the treatment of choice, as it
gives the best results. The relative indications for and results
of total elbow replacement
We studied, ten patients (11 elbows) who had
undergone 14 allograft-prosthesis composite reconstructions following
failure of a previous total elbow replacement with massive structural
bone loss. There were nine women and one man with a mean age of
64 years (40 to 84), who were reviewed at a mean of 75 months (24
to 213). One patient developed a deep infection after 26 months
and had the allograft-prosthesis composite removed, and two patients
had mild pain. The median flexion-extension arc was 100° (95% confidence
interval (CI) 76° to 124°). With the exception of the patient who
had the infected failure, all the patients could use their elbows
comfortably without splints or braces for activities of daily living.
The mean Mayo Elbow Performance Index improved from 9.5 (95% CI
4.4 to 14.7) pre-operatively to 74 (95% CI 62.4 to 84.9) at final
review. Radiologically, the rate of partial resorption was similar in
the humeral and ulnar allografts (three of six and four of eight,
respectively; p >
0.999). The patterns of resorption, however, were
different. Union at the host-bone-allograft junction was also different
between the humeral and ulnar allografts (one of six and seven of
eight showing union, respectively; p = 0.03). At medium-term follow-up, allograft-prosthesis composite reconstruction
appears to be a useful salvage technique for failed elbow replacements
with massive bone loss. The effects of allograft resorption and
host-bone-allograft junctional union on the longevity of allograft-prosthesis
composite reconstruction, however, remain unknown, and it is our
view that these patients should remain under long-term regular review.
Between September 1993 and September 1996, we performed 34 Kudo 5 total elbow replacements in 31 rheumatoid patients. All 22 surviving patients were reviewed at a mean of 11.9 years (10 to 14). Their mean age was 56 years (37 to 78) at the time of operation. All had Larsen grade IV or V rheumatoid changes on X-ray. Nine (three bilateral replacements and six unilateral) had died from unrelated causes. One who had died before ten years underwent revision for dislocation. Of the 22 total elbow replacements reviewed six had required revision, four for aseptic loosening (one humeral and three ulnar) and two for infection. Post-operatively, one patient had neuropraxia of the ulnar nerve and one of the radial nerve. Two patients had valgus tilting of the