Advertisement for orthosearch.org.uk
Results 1 - 20 of 31
Results per page:
The Bone & Joint Journal
Vol. 99-B, Issue 11 | Pages 1435 - 1441
1 Nov 2017
Buttaro MA Oñativia JI Slullitel PA Andreoli M Comba F Zanotti G Piccaluga F

Aims

The Corail stem has good long-term results. After four years of using this stem, we have detected a small group of patients who have presented with symptomatic metaphyseal debonding. The aim of this study was to quantify the incidence of this complication, to delineate the characteristics of patients presenting with this complication and to compare these patients with asymptomatic controls to determine any important predisposing factors.

Patients and Methods

Of 855 Corail collarless cementless stems implanted for osteoarthritis, 18 presented with symptomatic metaphyseal debonding. A control group of 74 randomly selected patients was assembled. Clinical and radiological parameters were measured and a logistic regression model was created to evaluate factors associated with metaphyseal debonding.


The Bone & Joint Journal
Vol. 99-B, Issue 5 | Pages 607 - 613
1 May 2017
Mäkinen TJ Abolghasemian M Watts E Fichman SG Kuzyk P Safir OA Gross AE

Aims

It may not be possible to undertake revision total hip arthroplasty (THA) in the presence of massive loss of acetabular bone stock using standard cementless hemispherical acetabular components and metal augments, as satisfactory stability cannot always be achieved. We aimed to study the outcome using a reconstruction cage and a porous metal augment in these patients.

Patients and Methods

A total of 22 acetabular revisions in 19 patients were performed using a combination of a reconstruction cage and porous metal augments. The augments were used in place of structural allografts. The mean age of the patients at the time of surgery was 70 years (27 to 85) and the mean follow-up was 39 months (27 to 58). The mean number of previous THAs was 1.9 (1 to 3). All patients had segmental defects involving more than 50% of the acetabulum and seven hips had an associated pelvic discontinuity.


The Bone & Joint Journal
Vol. 97-B, Issue 5 | Pages 582 - 589
1 May 2015
Brennan SA Ní Fhoghlú C Devitt BM O’Mahony FJ Brabazon D Walsh A

Implant-associated infection is a major source of morbidity in orthopaedic surgery. There has been extensive research into the development of materials that prevent biofilm formation, and hence, reduce the risk of infection. Silver nanoparticle technology is receiving much interest in the field of orthopaedics for its antimicrobial properties, and the results of studies to date are encouraging. Antimicrobial effects have been seen when silver nanoparticles are used in trauma implants, tumour prostheses, bone cement, and also when combined with hydroxyapatite coatings. Although there are promising results with in vitro and in vivo studies, the number of clinical studies remains small. Future studies will be required to explore further the possible side effects associated with silver nanoparticles, to ensure their use in an effective and biocompatible manner. Here we present a review of the current literature relating to the production of nanosilver for medical use, and its orthopaedic applications.

Cite this article: Bone Joint J 2015; 97-B:582–9.


The Bone & Joint Journal
Vol. 97-B, Issue 5 | Pages 595 - 602
1 May 2015
McCalden RW Korczak A Somerville L Yuan X Naudie DD

This was a randomised controlled trial studying the safety of a new short metaphyseal fixation (SMF) stem. We hypothesised that it would have similar early clinical results and micromovement to those of a standard-length tapered Synergy metaphyseal fixation stem. Using radiostereometric analysis (RSA) we compared the two stems in 43 patients. A short metaphyseal fixation stem was used in 22 patients and a Synergy stem in 21 patients. No difference was found in the clinical outcomes pre- or post-operatively between groups. RSA showed no significant differences two years post-operatively in mean micromovement between the two stems (except for varus/valgus tilt at p = 0.05) (subsidence 0.94 mm (sd 1.71) vs 0.32 mm (sd 0.45), p = 0.66; rotation 0.96° (sd 1.49) vs 1.41° (sd 2.95), p = 0.88; and total migration 1.09 mm (sd 1.74) vs 0.73 mm (sd 0.72), p = 0.51). A few stems (four SMF and three Synergy) had initial migration > 1.0 mm but stabilised by three to six months, with the exception of one SMF stem which required revision three years post-operatively. For most stems, total micromovement was very low at two years (subsidence < 0.5 mm, rotation < 1.0°, total migration < 0.5 mm), which was consistent with osseous ingrowth. The small sample makes it difficult to confirm the universal applicability of or elucidate the potential contraindications to the use of this particular new design of stem.

Cite this article: Bone Joint J 2015; 97-B:595–602.


The Bone & Joint Journal
Vol. 96-B, Issue 7 | Pages 889 - 895
1 Jul 2014
Fink B Urbansky K Schuster P

We report our experience of revision total hip replacement (THR) using the Revitan curved modular titanium fluted revision stem in patients with a full spectrum of proximal femoral defects. A total of 112 patients (116 revisions) with a mean age of 73.4 years (39 to 90) were included in the study. The mean follow-up was 7.5 years (5.3 to 9.1). A total of 12 patients (12 hips) died but their data were included in the survival analysis, and four patients (4 hips) were lost to follow-up. The clinical outcome, proximal bone regeneration and subsidence were assessed for 101 hips.

The mean Harris Hip Score was 88.2 (45.8 to 100) after five years and there was an increase of the mean Barnett and Nordin-Score, a measure of the proximal bone regeneration, of 20.8 (-3.1 to 52.7). Five stems had to be revised (4.3%), three (2.9%) showed subsidence, five (4.3%) a dislocation and two of 85 aseptic revisions (2.3%) a periprosthetic infection.

At the latest follow-up, the survival with revision of the stem as the endpoint was 95.7% (95% confidence interval 91.9% to 99.4%) and with aseptic loosening as the endpoint, was 100%. Peri-prosthetic fractures were not observed.

We report excellent results with respect to subsidence, the risk of fracture, and loosening after femoral revision using a modular curved revision stem with distal cone-in-cone fixation. A successful outcome depends on careful pre-operative planning and the use of a transfemoral approach when the anatomy is distorted or a fracture is imminent, or residual cement or a partially-secured existing stem cannot be removed. The shortest appropriate stem should, in our opinion, be used and secured with > 3 cm fixation at the femoral isthmus, and distal interlocking screws should be used for additional stability when this goal cannot be realised.

Cite this article: Bone Joint J 2014;96-B:889–95.


The Bone & Joint Journal
Vol. 96-B, Issue 5 | Pages 569 - 573
1 May 2014
Sullivan MP McHale KJ Parvizi J Mehta S

Nanotechnology is the study, production and controlled manipulation of materials with a grain size < 100 nm. At this level, the laws of classical mechanics fall away and those of quantum mechanics take over, resulting in unique behaviour of matter in terms of melting point, conductivity and reactivity. Additionally, and likely more significant, as grain size decreases, the ratio of surface area to volume drastically increases, allowing for greater interaction between implants and the surrounding cellular environment. This favourable increase in surface area plays an important role in mesenchymal cell differentiation and ultimately bone–implant interactions.

Basic science and translational research have revealed important potential applications for nanotechnology in orthopaedic surgery, particularly with regard to improving the interaction between implants and host bone. Nanophase materials more closely match the architecture of native trabecular bone, thereby greatly improving the osseo-integration of orthopaedic implants. Nanophase-coated prostheses can also reduce bacterial adhesion more than conventionally surfaced prostheses. Nanophase selenium has shown great promise when used for tumour reconstructions, as has nanophase silver in the management of traumatic wounds. Nanophase silver may significantly improve healing of peripheral nerve injuries, and nanophase gold has powerful anti-inflammatory effects on tendon inflammation.

Considerable advances must be made in our understanding of the potential health risks of production, implantation and wear patterns of nanophase devices before they are approved for clinical use. Their potential, however, is considerable, and is likely to benefit us all in the future.

Cite this article: Bone Joint J 2014; 96-B: 569–73.


The Bone & Joint Journal
Vol. 96-B, Issue 3 | Pages 299 - 305
1 Mar 2014
Bell KR Clement ND Jenkins PJ Keating JF

We performed a case–control study to compare the rates of further surgery, revision and complications, operating time and survival in patients who were treated with either an uncemented hydroxyapatite-coated Corail bipolar femoral stem or a cemented Exeter stem for a displaced intracapsular fracture of the hip. The mean age of the patients in the uncemented group was 82.5 years (53 to 97) and in the cemented group was 82.7 years (51 to 99) We used propensity score matching, adjusting for age, gender and the presence or absence of dementia and comorbidities, to produce a matched cohort receiving an Exeter stem (n = 69) with which to compare the outcome of patients receiving a Corail stem (n = 69). The Corail had a significantly lower all-cause rate of further surgery (p = 0.016; odds ratio (OR) 0.18, 95% CI 0.04 to 0.84) and number of hips undergoing major further surgery (p = 0.029; OR 0.13, 95% CI 0.01 to 1.09). The mean operating time was significantly less for the Corail group than for the cemented Exeter group (59 min [12 to 136] vs 70 min [40 to 175], p = 0.001). The Corail group also had a lower risk of a peri-prosthetic fracture (p = 0.042; OR 0.19, 95% CI 0.01 to 1.42) . There was no difference in the mortality rate between the groups. There were significantly fewer complications in the uncemented group, suggesting that the use of this stem would result in a decreased rate of morbidity in these frail patients. Whether this relates to an improved functional outcome remains unknown.

Cite this article: Bone Joint J 2014;96-B:299–305.


The Bone & Joint Journal
Vol. 95-B, Issue 12 | Pages 1610 - 1616
1 Dec 2013
Epinette J Asencio G Essig J Llagonne B Nourissat C

We report a multicentre prospective consecutive study assessing the long-term outcome of the proximally hydroxyapatite (HA)-coated ABG II monobloc femoral component in a series of 1148 hips in 1053 patients with a mean age at surgery of 64.77 years (22 to 80) at a mean follow-up of 10.84 years (10 to 15.25). At latest follow-up, the mean total Harris hip score was 94.7 points (sd; 6.87; 49 to 100), and the mean Merle d’Aubigné–Postel score was 17.6 points (sd 1.12; 7 to 18). The mean total Engh radiological score score was 21.54 (sd 5.77; 3.5 to 27), with 95.81% of ‘confirmed ingrowth’, according to Engh’s classification. With aseptic loosening or pain as endpoints, three AGB II stems (0.26%) failed, giving a 99.7% survival rate (se 0.002; 95% confidence interval (CI) 0.994 to 1) at 14 years’ follow-up. The survival of patients ≤ 50 years of age (99.0% (se 0.011; 95% CI 0.969 to 1)) did not differ significantly from those of patients aged > 50 years (99.8% (se 0.002; 95% CI 0.994 to 1)). This study confirmed the excellent long-term results currently achieved with the ABG II proximally HA-coated monobloc stem.

Cite this article: Bone Joint J 2013;95-B:1610–16.


The Bone & Joint Journal
Vol. 95-B, Issue 11 | Pages 1538 - 1543
1 Nov 2013
Kendrick BJL Wilson HA Lippett JE McAndrew AR Andrade AJMD

The National Institute for Health and Clinical Excellence (NICE) guidelines from 2011 recommend the use of cemented hemi-arthroplasty for appropriate patients with an intracapsular hip fracture. In our institution all patients who were admitted with an intracapsular hip fracture and were suitable for a hemi-arthroplasty between April 2010 and July 2012 received an uncemented prosthesis according to our established departmental routine practice. A retrospective analysis of outcome was performed to establish whether the continued use of an uncemented stem was justified. Patient, surgical and outcome data were collected on the National Hip Fracture database. A total of 306 patients received a Cathcart modular head on a Corail uncemented stem as a hemi-arthroplasty. The mean age of the patients was 83.3 years (sd 7.56; 46.6 to 94) and 216 (70.6%) were women. The mortality rate at 30 days was 5.8%. A total of 46.5% of patients returned to their own home by 30 days, which increased to 73.2% by 120 days. The implant used as a hemi-arthroplasty for intracapsular hip fracture provided satisfactory results, with a good rate of return to pre-injury place of residence and an acceptable mortality rate. Surgery should be performed by those who are familiar with the design of the stem and understand what is required for successful implantation.

Cite this article: Bone Joint J 2013;95-B:1538–43.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 11 | Pages 1540 - 1545
1 Nov 2012
Reigstad O Lütken T Grimsgaard C Bolstad B Thorkildsen R Røkkum M

The Motec cementless modular metal-on-metal ball-and-socket wrist arthroplasty was implanted in 16 wrists with scaphoid nonunion advanced collapse (SNAC; grades 3 or 4) and 14 wrists with scapholunate advanced collapse (SLAC) in 30 patients (20 men) with severe (grades 3 or 4) post-traumatic osteoarthritis of the wrist. The mean age of the patients was 52 years (31 to 71). All prostheses integrated well radiologically. At a mean follow-up of 3.2 years (1.1 to 6.1) no luxation or implant breakage occurred. Two wrists were converted to an arthrodesis for persistent pain. Loosening occurred in one further wrist at five years post-operatively. The remainder demonstrated close bone–implant contact. The clinical results were good, with markedly decreased Disabilities of the Arm Shoulder and Hand (DASH) and pain scores, and increased movement and grip strength. No patient used analgesics and most had returned to work.

Good short-term function was achieved using this wrist arthroplasty in a high-demand group of patients with post-traumatic osteoarthritis.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 6 | Pages 774 - 781
1 Jun 2012
Kim Y Oh J

We compared the clinical and radiological outcomes of two cementless femoral stems in the treatment of patients with a Garden III or IV fracture of the femoral neck. A total of 70 patients (70 hips) in each group were enrolled into a prospective randomised study. One group received a short anatomical cementless stem and the other received a conventional cementless stem. Their mean age was 74.9 years (50 to 94) and 76.0 years (55 to 96), respectively (p = 0.328). The mean follow-up was 4.1 years (2 to 5) and 4.8 years (2 to 6), respectively. Perfusion lung scans and high resolution chest CTs were performed to detect pulmonary microemboli.

At final follow-up there were no statistically significant differences between the short anatomical and the conventional stems with regard to the mean Harris hip score (85.7 (66 to 100) versus 86.5 (55 to 100); p = 0.791), the mean Western Ontario and McMaster Universities Osteoarthritis Index (17 (6 to 34) versus 16 (5 to 35); p = 0.13) or the mean University of California, Los Angeles activity score (5 (3 to 6) versus 4 (3 to 6); p = 0.032). No patient with a short stem had thigh pain, but 11 patients (16%) with a conventional stem had thigh pain. No patients with a short stem had symptomatic pulmonary microemboli, but 11 patients with a conventional stem had pulmonary microemboli (symptomatic in three patients and asymptomatic in eight patients). One hip (1.4%) in the short stem group and eight (11.4%) in the conventional group had an intra-operative undisplaced fracture of the calcar. No component was revised for aseptic loosening in either group. One acetabular component in the short stem group and two acetabular components in the conventional stem group were revised for recurrent dislocation.

Our study demonstrated that despite the poor bone quality in these elderly patients with a fracture of the femoral neck, osseo-integration was obtained in all hips in both groups. However, the incidence of thigh pain, pulmonary microemboli and peri-prosthetic fracture was significantly higher in the conventional stem group than in the short stem group.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 4 | Pages 564 - 569
1 Apr 2012
Pendegrass CJ El-Husseiny M Blunn GW

The success of long-term transcutaneous implants depends on dermal attachment to prevent downgrowth of the epithelium and infection. Hydroxyapatite (HA) coatings and fibronectin (Fn) have independently been shown to regulate fibroblast activity and improve attachment. In an attempt to enhance this phenomenon we adsorbed Fn onto HA-coated substrates. Our study was designed to test the hypothesis that adsorption of Fn onto HA produces a surface that will increase the attachment of dermal fibroblasts better than HA alone or titanium alloy controls.

Iodinated Fn was used to investigate the durability of the protein coating and a bioassay using human dermal fibroblasts was performed to assess the effects of the coating on cell attachment. Cell attachment data were compared with those for HA alone and titanium alloy controls at one, four and 24 hours. Protein attachment peaked within one hour of incubation and the maximum binding efficiency was achieved with an initial droplet of 1000 ng. We showed that after 24 hours one-fifth of the initial Fn coating remained on the substrates, and this resulted in a significant, three-, four-, and sevenfold increase in dermal fibroblast attachment strength compared to uncoated controls at one, four and 24 hours, respectively.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 3 | Pages 425 - 430
1 Mar 2012
Picardo NE Blunn GW Shekkeris AS Meswania J Aston WJ Pollock RC Skinner JA Cannon SR Briggs TW

In skeletally immature patients, resection of bone tumours and reconstruction of the lower limb often results in leg-length discrepancy. The Stanmore non-invasive extendible endoprosthesis, which uses electromagnetic induction, allows post-operative lengthening without anaesthesia. Between 2002 and 2009, 55 children with a mean age of 11.4 years (5 to 16) underwent reconstruction with this prosthesis; ten patients (18.2%) died of disseminated disease and one child underwent amputation due to infection. We reviewed 44 patients after a mean follow-up of 41.2 months (22 to 104). The mean Musculoskeletal Tumor Society score was 24.7 (8 to 30) and the Toronto Extremity Salvage score was 92.3% (55.2% to 99.0%). There was no local recurrence of tumour. Complications developed in 16 patients (29.1%) and ten (18.2%) underwent revision.

The mean length gained per patient was 38.6 mm (3.5 to 161.5), requiring a mean of 11.3 extensions (1 to 40), and ten component exchanges were performed in nine patients (16.4%) after attaining the maximum lengthening capacity of the implant. There were 11 patients (20%) who were skeletally mature at follow-up, ten of whom had equal leg lengths and nine had a full range of movement of the hip and knee.

This is the largest reported series using non-invasive extendible endoprostheses after excision of primary bone tumours in skeletally immature patients. The technique produces a good functional outcome, with prevention of limb-length discrepancy at skeletal maturity.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 2 | Pages 163 - 166
1 Feb 2012
Zahos K Mehendale S Ward AJ Smith EJ Nichols M

We report the use of a 15° face-changing cementless acetabular component in patients undergoing total hip replacement for osteoarthritis secondary to developmental dysplasia of the hip. The rationale behind its design and the surgical technique used for its implantation are described. It is distinctly different from a standard cementless hemispherical component as it is designed to position the bearing surface at the optimal angle of inclination, that is, < 45°, while maximising the cover of the component by host bone.


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 12 | Pages 1592 - 1596
1 Dec 2011
Babis GC Sakellariou VI Chatziantoniou AN Soucacos PN Megas P

We report the results of 62 hips in 62 patients (17 males, 45 females) with mean age of 62.4 years (37 to 81), who underwent revision of the acetabular component of a total hip replacement due to aseptic loosening between May 2003 and November 2007. All hips had a Paprosky type IIIa acetabular defect. Acetabular revision was undertaken using a Procotyl E cementless oblong implant with modular side plates and a hook combined with impaction allografting.

At a mean follow-up of 60.5 months (36 to 94) with no patients lost to follow-up and one died due to unrelated illness, the complication rate was 38.7%. Complications included aseptic loosening (19 hips), deep infection (3 hips), broken hook and side plate (one hip) and a femoral nerve palsy (one hip). Further revision of the acetabular component was required in 18 hips (29.0%) and a further four hips (6.4%) are currently loose and awaiting revision.

We observed unacceptably high rates of complication and failure in our group of patients and cannot recommend this implant or technique.


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 10 | Pages 1320 - 1326
1 Oct 2011
Regis D Sandri A Bonetti I Braggion M Bartolozzi P

Revision after failed femoral components may be technically demanding due to loss of peri-prosthetic bone. This retrospective study evaluated the long-term results of femoral revision using the cementless Wagner Self-Locking stem. Between 1992 and 1998, 68 consecutive hips in 66 patients underwent femoral revision using this implant. A total of 25 patients died from unrelated causes without further revision; the remaining 41 hips in 41 patients (12 men and 29 women) with a mean age of 61 years (29 to 80) were reviewed at a mean follow-up of 13.9 years (10.4 to 15.8). A transfemoral approach was used in 32 hips. A total of five stems required further revision because of infection in two, progressive subsidence in two and recurrent dislocation in one. Four hips had dislocated and eight stems had subsided ≥ 10 mm. The mean Harris hip score improved from 33 points pre-operatively to 75 points at final follow-up (p < 0.001). In all, 33 stems (91.7%) showed radiological signs of stable bone fixation. The cumulative survival rates at 15.8 years with femoral revision for any reason and for stem failure as the endpoints were 92.0% (95% confidence interval (CI) 86.0% to 98.4%) and 96.6% (95% CI 92.2% to 100%), respectively. The survivorship with revision and ≥ 10 mm migration of the stem as the endpoint was 83.6% (95% CI 76.6% to 91.4%).

This study shows quite good survival and moderate clinical outcome when using a monoblock tapered titanium stem for supporting the regeneration of bone in complex revision hip surgery.


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 8 | Pages 1098 - 1103
1 Aug 2011
Ruggieri P Mavrogenis AF Guerra G Mercuri M

We retrospectively studied 14 patients with proximal and diaphyseal tumours and disappearing bone (Gorham’s) disease of the humerus treated with wide resection and reconstruction using an allograft-resurfacing composite (ARC). There were ten women and four men, with a mean age of 35 years (8 to 69). At a mean follow-up of 25 months (10 to 89), two patients had a fracture of the allograft. In one of these it was revised with a similar ARC and in the other with an intercalary prosthesis. A further patient had an infection and a fracture of the allograft that was revised with a megaprosthesis. In all patients with an ARC, healing of the ARC-host bone interface was observed. One patient had failure of the locking mechanism of the total elbow replacement. The mean post-operative Musculoskeletal Tumor Society score for the upper extremity was 77% (46.7% to 86.7%), which represents good and excellent results; one patient had a poor result (46.7%).

In the short term ARC effectively relieves pain and restores shoulder function in patients with wide resection of the proximal humerus. Fracture and infection remain significant complications.


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 8 | Pages 1111 - 1117
1 Aug 2011
Sewell MD Hanna SA McGrath A Aston WJS Blunn GW Pollock RC Skinner JA Cannon SR Briggs TWR

The best method of reconstruction after resection of malignant tumours of the tibial diaphysis is unknown. In the absence of any long-term studies analysing the results of intercalary endoprosthetic replacement, we present a retrospective review of 18 patients who underwent limb salvage using a tibial diaphyseal endoprosthetic replacement following excision of a malignant bone tumour. There were ten men and eight women with a mean age of 42.5 years (16 to 76). Mean follow-up was 58.5 months (20 to 141) for all patients and 69.3 months (20 to 141) for the 12 patients still alive. Cumulative patient survival was 59% (95% confidence interval (CI) 32 to 84) at five years. Implant survival was 63% (95% CI 35 to 90) at ten years. Four patients required revision to a proximal tibial replacement at a mean follow-up of 29 months (10 to 54). Complications included metastases in five patients, aseptic loosening in four, peri-prosthetic fracture in two, infection in one and local recurrence in one. The mean Musculoskeletal Tumor Society score and the mean Toronto Extremity Salvage Score were 23 (17 to 28) and 74% (53 to 91), respectively.

Although rates of complication and revision were high, custom-made tibial diaphyseal replacement following resection of malignant bone tumours enables early return to function and provides an attractive alternative to other surgical options, without apparent compromise of patient survival.


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 6 | Pages 760 - 768
1 Jun 2011
ten Broeke RHM Alves A Baumann A Arts JJC Geesink RGT

Four uncemented Symax hip stems were extracted at three weeks and nine, 13 and 32 months, respectively, for reasons other than loosening. The reasons for implant removal were infection in two cases, recurrent dislocation in one and acetabular fracture in one. They were analysed to assess the effect and behaviour of an electrochemically deposited, completely resorbable biomimetic BONIT-hydroxyapatite (HA) coating (proximal part) and a DOTIZE surface treatment (distal part) using qualitative histology, quantitative histomorphometry and scanning electron microscopy (SEM). Early and direct bone-implant bonding with signs of active remodelling of bone and the HA coating were demonstrated by histology and SEM. No loose BONIT-HA particles or delamination of the coating were observed, and there was no inflammation or fibrous interposition at the interface. Histomorphometry showed bone-implant contact varying between 26.5% at three weeks and 83.5% at 13 months at the HA-coated implant surface. The bone density in the area of investigation was between 24.6% at three weeks and 41.1% at 32 months. The DOTIZE surface treatment of the distal part of the stem completely prevented tissue and bone apposition in all cases, thereby optimising proximal stress transfer. The overall features of this implant, in terms of geometry and surface texture, suggest a mechanically stable design with a highly active biomimetic coating, resulting in rapid and extensive osseo-integration, exclusively in the metaphyseal part of the stem. Early remodelling of the HA coating does not seem to have a detrimental effect on short-term bone-implant coupling. There were no adverse effects identified from either the BONIT-HA coating or the DOTIZE surface treatment


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 3 | Pages 399 - 403
1 Mar 2011
Griffiths D Gikas PD Jowett C Bayliss L Aston W Skinner J Cannon S Blunn G Briggs TWR Pollock R

Between 1997 and 2007, 68 consecutive patients underwent replacement of the proximal humerus for tumour using a fixed-fulcrum massive endoprosthesis. Their mean age was 46 years (7 to 87). Ten patients were lost to follow-up and 16 patients died. The 42 surviving patients were assessed using the Musculoskeletal Tumor Society (MSTS) Score and the Toronto Extremity Salvage Score (TESS) at a mean follow-up of five years and 11 months (one year to ten years and nine months). The mean MSTS score was 72.3% (53.3% to 100%) and the mean TESS was 77.2% (58.6% to 100%).

Four of 42 patients received a new constrained humeral liner to reduce the risk of dislocation. This subgroup had a mean MSTS score of 77.7% and a mean TESS of 80.0%. The dislocation rate for the original prosthesis was 25.9; none of the patients with the new liner had a dislocation at a mean of 14.5 months (12 to 18).

Endoprosthetic replacement for tumours of the proximal humerus using this prosthesis is a reliable operation yielding good results without the documented problems of unconstrained prostheses. The performance of this prosthesis is expected to improve further with a new constrained humeral liner, which reduces the risk of dislocation.