Aims. Several different designs of hemiarthroplasty are used to treat intracapsular fractures of the proximal femur, with large variations in costs. No clinical benefit of modular over monoblock designs has been reported in the literature. Long-term data are lacking. The aim of this study was to report the ten-year implant survival of commonly used designs of hemiarthroplasty. Methods. Patients recorded by the Australian Orthopaedic Association National Joint Replacement Registry (AOANJRR) between 1 September 1999 and 31 December 2020 who underwent hemiarthroplasty for the treatment of a hip fracture with the following implants were included: a cemented monoblock Exeter Trauma Stem (ETS), cemented Exeter V40 with a bipolar head, a monoblock Thompsons prosthesis (Cobalt/Chromium or Titanium), and an Exeter V40 with a Unitrax head. Overall and age-defined cumulative revision rates were compared over the ten years following surgery. Results. A total of 41,949 hemiarthroplasties were included. Exeter V40 with a Unitrax head was the most commonly used (n = 20,707, 49.4%). The overall rate of revision was small. A total of 28,201 patients (67.2%) were aged > 80 years. There were no significant differences in revision rates across all designs of hemiarthroplasty in patients of this age at any time. The revision rates for all designs were < 3.5%, three years postoperatively. At subsequent times the
Hip hemiarthroplasty is a standard treatment for intracapsular
proximal femoral fractures in the frail elderly. In this study we
have explored the implications of early return to theatre, within
30 days, on patient outcome following hip hemiarthroplasty. We retrospectively reviewed the hospital records of all hip hemiarthroplasties
performed in our unit between January 2010 and January 2015. Demographic
details, medical backround, details of the primary procedure, complications,
subsequent procedures requiring return to theatre, re-admissions,
discharge destination and death were collected.Aims
Patients and Methods
A total of 56 male patients with a displaced
intracapsular fracture of the hip and a mean age of 81 years (62
to 94), were randomised to be treated with either a cemented hemiarthroplasty
(the Exeter Trauma Stem) or reduction and internal fixation using
the Targon Femoral Plate. All surviving patients were reviewed one
year after the injury, at which time restoration of function and
pain in the hip was assessed. There was no statistically significant
difference in mortality between the two groups (7/26; 26.9% for
hemiarthroplasty These results indicate that cemented hemiarthroplasty gives better
results than internal fixation in elderly men with a displaced intracapsular
fracture of the hip. Cite this article:
National Institute of Clinical Excellence guidelines
state that cemented stems with an Orthopaedic Data Evaluation Panel
(ODEP) rating of >
3B should be used for hemiarthroplasty when treating
an intracapsular fracture of the femoral neck. These recommendations
are based on studies in which most, if not all stems, did not hold
such a rating. This case-control study compared the outcome of hemiarthroplasty
using a cemented (Exeter) or uncemented (Corail) femoral stem. These
are the two prostheses most commonly used in hip arthroplasty in
the UK. Data were obtained from two centres; most patients had undergone
hemiarthroplasty using a cemented Exeter stem (n = 292/412). Patients
were matched for all factors that have been shown to influence mortality
after an intracapsular fracture of the neck of the femur. Outcome
measures included: complications, re-operations and mortality rates
at two, seven, 30 and 365 days post-operatively. Comparable outcomes
for the two stems were seen. There were more intra-operative complications in the uncemented
group (13% This study therefore supports the use of both cemented and uncemented
stems of proven design, with an ODEP rating of 10A, in patients
with an intracapsular fracture of the neck of the femur. Cite this article:
Recent guidance recommends the use of a well-proven
cemented femoral stem for hemiarthroplasty in the management of
fractures of the femoral neck, and the Exeter Trauma Stem (ETS)
has been suggested as an example of such an implant. The design
of this stem was based on the well-proven Exeter Total Hip Replacement
stem (ETHRS). This study assessed the surface finish of the ETS
in comparison with the ETHRS. Two ETSs and two ETHRSs were examined
using a profilometer with a precision of 1 nm and compared with
an explanted Exeter Matt stem. The mean roughness average (RA) of
the ETSs was approximately ten times higher than that of the ETHRSs (0.235 μm
(0.095 to 0.452) versus 0.025 μm (0.011 to 0.059);
p <
0.001). The historical Exeter Matt stem roughness measured
a mean RA of 0.973 μm (0.658 to 1.159). The change of the polished
Exeter stem to a matt surface finish in 1976 resulted in a high
stem failure rate. We do not yet know whether the surface differences
between
We evaluated the cost and consequences of proximal femoral fractures requiring further surgery because of complications. The data were collected prospectively in a standard manner from all patients with a proximal femoral fracture presenting to the trauma unit at the John Radcliffe Hospital over a five-year period. The total cost of treatment for each patient was calculated by separating it into its various components. The risk factors for the complications that arose, the location of their discharge and the mortality rates for these patients were compared to those of a matched control group. There were 2360 proximal femoral fractures in 2257 patients, of which 144 (6.1%) required further surgery. The mean cost of treatment in patients with complications was £18 709 (£2606.30 to £60.827.10), compared with £8610 (£918.54 to £45 601.30) for uncomplicated cases (p <
0.01), with a mean length of stay of 62.8 (44.5 to 79.3) and 32.7 (23.8 to 35.0) days, respectively. The probability of mortality after one month in these cases was significantly higher than in the control group, with a mean survival of 209 days, compared with 496 days for the controls. Patients with complications were statistically less likely to return to their own home (p <
0.01). Greater awareness and understanding are required to minimise the complications of proximal femoral fractures and consequently their cost.