The survival of humeral hemiarthroplasties in patients with relatively intact glenoid cartilage could theoretically be extended by minimizing the associated postoperative glenoid erosion. Ceramic has gained attention as an alternative to metal as a material for hemiarthroplasties because of its superior tribological properties. The aim of this study was to assess the in vitro wear performance of ceramic and metal humeral hemiarthroplasties on natural glenoids. Intact right cadaveric shoulders from donors aged between 50 and 65 years were assigned to a ceramic group (n = 8, four male cadavers) and a metal group (n = 9, four male cadavers). A dedicated shoulder wear simulator was used to simulate daily activity by replicating the relevant joint motion and loading profiles. During testing, the joint was kept lubricated with diluted calf serum at room temperature. Each test of wear was performed for 500,000 cycles at 1.2 Hz. At intervals of 125,000 cycles, micro-CT scans of each glenoid were taken to characterize and quantify glenoid wear by calculating the change in the thickness of its articular cartilage.Aims
Methods
Metal and ceramic humeral head bearing surfaces are available choices in anatomical shoulder arthroplasties. Wear studies have shown superior performance of ceramic heads, however comparison of clinical outcomes according to bearing surface in total shoulder arthroplasty (TSA) and hemiarthroplasty (HA) is limited. This study aimed to compare the rates of revision and reoperation following metal and ceramic humeral head TSA and HA using data from the National Joint Registry (NJR), which collects data from England, Wales, Northern Ireland, Isle of Man and the States of Guernsey. NJR shoulder arthroplasty records were linked to Hospital Episode Statistics and the National Mortality Register. TSA and HA performed for osteoarthritis (OA) in patients with an intact rotator cuff were included. Metal and ceramic humeral head prostheses were matched within separate TSA and HA groups using propensity scores based on 12 and 11 characteristics, respectively. The primary outcome was time to first revision and the secondary outcome was non-revision reoperation.Aims
Methods
This study reports the ten-year wear rates, incidence of osteolysis, clinical outcomes, and complications of a multicentre randomized controlled trial comparing oxidized zirconium (OxZr) versus cobalt-chrome (CoCr) femoral heads with ultra-high molecular weight polyethylene (UHMWPE) and highly cross-linked polyethylene (XLPE) liners in total hip arthroplasty (THA). Patients undergoing primary THA were recruited from four institutions and prospectively allocated to the following treatment groups: Group A, CoCr femoral head with XLPE liner; Group B, OxZr femoral head with XLPE liner; and Group C, OxZr femoral head with UHMWPE liner. All study patients and assessors recording outcomes were blinded to the treatment groups. The outcomes of 262 study patients were analyzed at ten years’ follow-up.Aims
Methods
To clarify the effectiveness of the induced membrane technique (IMT) using beta-tricalcium phosphate (β-TCP) for reconstruction of segmental bone defects by evaluating clinical and radiological outcomes, and the effect of defect size and operated site on surgical outcomes. A review of the medical records was conducted of consecutive 35 lower limbs (30 males and five females; median age 46 years (interquartile range (IQR) 40 to 61)) treated with IMT using β-TCP between 2014 and 2018. Lower Extremity Functional Score (LEFS) was examined preoperatively and at final follow-up to clarify patient-centered outcomes. Bone healing was assessed radiologically, and time from the second stage to bone healing was also evaluated. Patients were divided into ≥ 50 mm and < 50 mm defect groups and into femoral reconstruction, tibial reconstruction, and ankle arthrodesis groups.Aims
Methods
Periprosthetic joint infection (PJI) is one of the most dreaded complications after arthroplasty surgery; thus numerous approaches have been undertaken to equip metal surfaces with antibacterial properties. Due to its antimicrobial effects, silver is a promising coating for metallic surfaces, and several types of silver-coated arthroplasty implants are in clinical use today. However, silver can also exert toxic effects on eukaryotic cells both in the immediate vicinity of the coated implants and systemically. In most clinically-used implants, silver coatings are applied on bulk components that are not in direct contact with bone, such as in partial or total long bone arthroplasties used in tumour or complex revision surgery. These implants differ considerably in the coating method, total silver content, and silver release rates. Safety issues, such as the occurrence of argyria, have been a cause for concern, and the efficacy of silver coatings in terms of preventing PJI is also controversial. The application of silver coatings is uncommon on parts of implants intended for cementless fixation in host bone, but this option might be highly desirable since the modification of implant surfaces in order to improve osteoconductivity can also increase bacterial adhesion. Therefore, an optimal silver content that inhibits bacterial colonization while maintaining osteoconductivity is crucial if silver were to be applied as a coating on parts intended for bone contact. This review summarizes the different methods used to apply silver coatings to arthroplasty components, with a focus on the amount and duration of silver release from the different coatings; the available experience with silver-coated implants that are in clinical use today; and future strategies to balance the effects of silver on bacteria and eukaryotic cells, and to develop silver-coated titanium components suitable for bone ingrowth. Cite this article:
Antibiotic resistance represents a threat to human health. It has been suggested that by 2050, antibiotic-resistant infections could cause ten million deaths each year. In orthopaedics, many patients undergoing surgery suffer from complications resulting from implant-associated infection. In these circumstances secondary surgery is usually required and chronic and/or relapsing disease may ensue. The development of effective treatments for antibiotic-resistant infections is needed. Recent evidence shows that bacteriophage (phages; viruses that infect bacteria) therapy may represent a viable and successful solution. In this review, a brief description of bone and joint infection and the nature of bacteriophages is presented, as well as a summary of our current knowledge on the use of bacteriophages in the treatment of bacterial infections. We present contemporary published in vitro and in vivo data as well as data from clinical trials, as they relate to bone and joint infections. We discuss the potential use of bacteriophage therapy in orthopaedic infections. This area of research is beginning to reveal successful results, but mostly in nonorthopaedic fields. We believe that bacteriophage therapy has potential therapeutic value for implant-associated infections in orthopaedics. Cite this article:
Ceramic bearings have several desirable properties, such as resistance to wear, hardness, and biocompatibility, that favour it as an articulating surface in hip arthroplasty. However, ceramic fracture remains a concern. We have reviewed the contemporary literature, addressing the factors that can influence the incidence of ceramic bearing surface fracture. Cite this article:
Aims. Early evidence has emerged suggesting that ceramic-on-ceramic
articulations induce a different tissue reaction to ceramic-on-polyethylene
and metal-on-metal bearings. Therefore, the aim of this study was
to investigate the tissue reaction and cellular response to ceramic
total hip arthroplasty (THA) materials in vitro,
as well as the tissue reaction in capsular tissue after revision
surgery of ceramic-on-ceramic THAs. Patients and Methods. We investigated tissue collected at revision surgery from nine
ceramic-on-ceramic articulations. we compared our findings with
tissue obtained from five metal-on-metal THA revisions, four ceramic-on-polyethylene
THAs, and four primary osteoarthritis synovial membranes. The latter
were analyzed to assess the amount of tissue fibrosis that might
have been present at the time of implantation to enable evaluation,
in relation to implantation time, of any subsequent response in
the tissues. Results. There was a significant increase in tissue fibrosis with implantation
time for all implant types tested. Interestingly, the tissue fibrosis
in ceramic-on-ceramic THAs was significantly increased compared
with metal-on-metal and ceramic-on-polyethylene. Additionally, we
found ceramic wear particles in the periprosthetic tissue of ceramic implants.
Fibroblasts responded with expression of cytokines when cultured
on alumina-toughened zirconia (ATZ) and zirconia-toughened alumina
(ZTA) ceramic surfaces. This response was more pronounced on ATZ
ceramics compared with ZTA
The development and pre-clinical evaluation of
nano-texturised, biomimetic, surfaces of titanium (Ti) implants treated
with titanium dioxide (TiO2) nanotube arrays is reviewed. Cite this article:
The Corail stem has good long-term results. After four years
of using this stem, we have detected a small group of patients who
have presented with symptomatic metaphyseal debonding. The aim of
this study was to quantify the incidence of this complication, to
delineate the characteristics of patients presenting with this complication
and to compare these patients with asymptomatic controls to determine
any important predisposing factors. Of 855 Corail collarless cementless stems implanted for osteoarthritis,
18 presented with symptomatic metaphyseal debonding. A control group
of 74 randomly selected patients was assembled. Clinical and radiological
parameters were measured and a logistic regression model was created
to evaluate factors associated with metaphyseal debonding.Aims
Patients and Methods
Ceramic-on-ceramic (CoC) bearings in total hip arthroplasty (THA)
are commonly used, but concerns exist regarding ceramic fracture.
This study aims to report the risk of revision for fracture of modern
CoC bearings and identify factors that might influence this risk,
using data from the National Joint Registry (NJR) for England, Wales, Northern
Ireland and the Isle of Man. We analysed data on 223 362 bearings from 111 681 primary CoC
THAs and 182 linked revisions for bearing fracture recorded in the
NJR. We used implant codes to identify ceramic bearing composition
and generated Kaplan-Meier estimates for implant survivorship. Logistic
regression analyses were performed for implant size and patient specific
variables to determine any associated risks for revision.Aims
Patients and Methods
To determine the effect of a change in design of a cementless
ceramic acetabular component in fixation and clinical outcome after
total hip arthroplasty We compared 342 hips (302 patients) operated between 1999 and
2005 with a relatively smooth hydroxyapatite coated acetabular component
(group 1), and 337 hips (310 patients) operated between 2006 and
2011 using a similar acetabular component with a macrotexture on
the entire outer surface of the component (group 2). The mean age of
the patients was 53.5 (14 to 70) in group 1 and 53.0 (15 to 70)
in group 2. The mean follow-up was 12.7 years (10 to 17) for group
1 and 7.2 years (4 to 10) for group 2.Aims
Patients and Methods
We conducted a prospective study of a delta ceramic total hip
arthroplasty (THA) to determine the rate of ceramic fracture, to
characterise post-operative noise, and to evaluate the mid-term
results and survivorship. Between March 2009 and March 2011, 274 patients (310 hips) underwent
cementless THA using a delta ceramic femoral head and liner. At
each follow-up, clinical and radiological outcomes were recorded.
A Kaplan-Meier analysis was undertaken to estimate survival.Aims
Patients and Methods
This review summarises the technique of impaction
grafting with mesh augmentation for the treatment of uncontained
acetabular defects in revision hip arthroplasty. The ideal acetabular revision should restore bone stock, use
a small socket in the near-anatomic position, and provide durable
fixation. Impaction bone grafting, which has been in use for over
40 years, offers the ability to achieve these goals in uncontained
defects. The precepts of modern, revision impaction grafting are
that the segmental or cavitary defects must be supported with a
mesh; the contained cavity is filled with vigorously impacted morselised
fresh-frozen allograft; and finally, acrylic cement is used to stabilise
the graft and provide rigid, long-lasting fixation of the revised
acetabular component. Favourable results have been published with this technique. While
having its limitations, it is a viable option to address large acetabular
defects in revision arthroplasty. Cite this article:
To compare the structural durability of hydroxyapatite-tricalcium
phosphate (HATCP) to autologous iliac crest bone graft in calcaneal
lengthening osteotomy (CLO) for pes planovalgus in childhood. We present the interim results of ten patients (HATCP, n = 6
and autograft, n = 5) with a mean age of 11.5 years (8.2 to 14.2)
from a randomised controlled non-inferiority trial with six months
follow-up. The primary outcome was the stability of the osteotomy
as measured by radiostereometric analysis. A non-inferiority margin
of ≤ 2 mm osteotomy compression was set.Aims
Patients and Methods
The use of large-diameter metal-on-metal (MoM)
components in total hip arthroplasty (THA) is associated with an increased
risk of early failure due to adverse local tissue reaction to metal
debris (ARMD) in response to the release of metal ions from the
bearing couple and/or head-neck taper corrosion. The aim of this
paper was to present a review of the incidence and natural history
of ARMD and the forms of treatment, with a focus on the need for
and extent of resection or debulking of the pseudotumour. An illustrative
case report is presented of a patient with an intra-pelvic pseudotumour
associated with a large diameter MoM THA, which was treated successfully
with revision of the bearing surface to a dual mobility couple and
retention of the well-fixed acetabular and femoral components. The
pseudotumour was left Cite this article:
The leading indication for revision total hip
arthroplasty (THA) remains aseptic loosening owing to wear. The younger,
more active patients currently undergoing THA present unprecedented
demands on the bearings. Ceramic-on-ceramic (CoC) bearings have
consistently shown the lowest rates of wear. The recent advances,
especially involving alumina/zirconia composite ceramic, have led
to substantial improvements and good results in vitro. Alumina/zirconia composite
The number of arthroplasties being undertaken
is expected to grow year on year, and periprosthetic joint infections will
be an increasing socioeconomic burden. The challenge to prevent
and eradicate these infections has resulted in the emergence of
several new strategies, which are discussed in this review. Cite this article: