The Birmingham Hip Resurfacing (BHR) was introduced in 1997 to address the needs of young active patients using a historically proven large-diameter metal-on-metal (MoM) bearing. A single designer surgeon’s consecutive series of 130 patients (144 hips) was previously reported at five and ten years, reporting three and ten failures, respectively. The aim of this study was to extend the follow-up of this original cohort at 25 years. The study extends the reporting on the first consecutive 144 resurfacing procedures in 130 patients for all indications. All operations were undertaken between August 1997 and May 1998. The mean age at operation was 52.1 years (SD 9.93; 17 to 76), and included 37 female patients (28.5%). Failure was defined as revision of either component for any reason. Kaplan-Meier survival analysis was performed. Routine follow-up with serum metal ion levels, radiographs, and Oxford Hip Scores (OHSs) was undertaken.Aims
Methods
This study aims to assess the relationship between history of pseudotumour formation secondary to metal-on-metal (MoM) implants and periprosthetic joint infection (PJI) rate, as well as establish ESR and CRP thresholds that are suggestive of infection in these patients. We hypothesized that patients with a pseudotumour were at increased risk of infection. A total of 1,171 total hip arthroplasty (THA) patients with MoM articulations from August 2000 to March 2014 were retrospectively identified. Of those, 328 patients underwent metal artefact reduction sequence MRI and had minimum two years’ clinical follow-up, and met our inclusion criteria. Data collected included demographic details, surgical indication, laterality, implants used, history of pseudotumour, and their corresponding preoperative ESR (mm/hr) and CRP (mg/dl) levels. Multivariate logistic regression modelling was used to evaluate PJI and history of pseudotumour, and receiver operating characteristic curves were created to assess the diagnostic capabilities of ESR and CRP to determine the presence of infection in patients undergoing revision surgery.Aims
Methods
The aims of this study were to determine if an increasing serum cobalt (Co) and/or chromium (Cr) concentration is correlated with a decreasing Harris Hip Score (HHS) and Hip disability and Osteoarthritis Outcome Score (HOOS) in patients who received the Articular Surface Replacement (ASR) hip resurfacing arthroplasty (HRA), and to evaluate the ten-year revision rate and show if sex, inclination angle, and Co level influenced the revision rate. A total of 62 patients with an ASR-HRA were included and monitored yearly postoperatively. At follow-up, serum Co and Cr levels were measured and the HHS and the HOOS were scored. In addition, preoperative patient and implant variables and the need for revision surgery were recorded. We used a linear mixed model to relate the serum Co and Cr levels to different patient-reported outcome measures (PROMs). For the survival analyses we used the Kaplan-Meier and Cox regression model.Aims
Methods
Knee arthroplasty surgery is a highly effective treatment for arthritis and disorders of the knee. There are a wide variety of implant brands and types of knee arthroplasty available to surgeons. As a result of a number of highly publicized failures, arthroplasty surgery is highly regulated in the UK and many other countries through national registries, introduced to monitor implant performance, surgeons, and hospitals. With time, the options available within many brand portfolios have grown, with alternative tibial or femoral components, tibial insert materials, or shapes and patella resurfacings. In this study we have investigated the effect of the expansion of implant brand portfolios and where there may be a lack of transparency around a brand name. We also aimed to establish the potential numbers of compatible implant construct combinations. Hypothetical implant brand portfolios were proposed, and the number of compatible implant construct combinations was calculated.Aims
Methods
To find out if there is an inverse association between estimated glomerular filtration rate (eGFR) and whole blood cobalt (Co) and chromium (Cr) levels in patients with metal-on-metal (MoM) hip arthroplasties and renal insufficiency, suggesting that renal insufficiency could cause accumulation of Co and Cr in blood. Out of 2,520 patients with 3,013 MoM hip arthroplasties, we identified 1,244 patients with whole blood Co, Cr, and creatinine measured within no more than a one-year interval. We analyzed the correlation of blood metal ion levels and eGFR to identify a potential trend of accumulating Co or Cr with decreasing eGFR.Aims
Methods
Adverse local tissue reactions associated with abnormal wear considerably slowed down the general use of metal-on-metal (MoM) hip resurfacing arthroplasty (HRA), now limited to a few specialized centres. In this study, we provide the clinical results of 400 consecutive MoM HRAs implanted more than 20 years ago in one such centre. A total of 355 patients (400 hips) were treated with Conserve Plus HRA between November 1996 and November 2000. There were 96 female (27%) and 259 male patients (73%). Their mean age was 48.2 years (SD 10.9). The University of California, Los Angeles (UCLA) hip scores and 12-item Short Form Survey (SF-12) quality of life scores were reported. Survivorship was assessed using Kaplan-Meier analyses.Aims
Methods
For this retrospective cohort study, patients aged ≤ 30 years
(very young) who underwent total hip arthroplasty (THA) were compared
with patients aged ≥ 60 years (elderly) to evaluate the rate of
revision arthroplasty, implant survival, the indications for revision,
the complications, and the patient-reported outcomes. We retrospectively reviewed all patients who underwent primary
THA between January 2000 and May 2015 from our institutional database.
A total of 145 very young and 1359 elderly patients were reviewed.
The mean follow-up was 5.3 years (1 to 18). Logistic generalized
estimating equations were used to compare characteristics and the revision
rate. Survival was evaluated using Kaplan–Meier curves and hazard
rates were created using Cox regression.Aims
Patients and Methods
The aim of this study was to determine whether the rates of revision
for metal-on-metal (MoM) total hip arthroplasties (THAs) with Pinnacle
components varied according to the year of the initial operation,
and compare these with the rates of revision for other designs of
MoM THA. Data from the National Joint Registry for England and Wales included
36 mm MoM THAs with Pinnacle acetabular components which were undertaken
between 2003 and 2012 with follow-up for at least five years (n
= 10 776) and a control group of other MoM THAs (n = 13 817). The
effect of the year of the primary operation on all-cause rates of revision
was assessed using Cox regression and interrupted time-series analysis.Aims
Patients and Methods
The aim of this study was to determine whether patients with
metal-on-metal (MoM) arthroplasties of the hip have an increased
risk of cardiac failure compared with those with alternative types
of arthroplasties (non-MoM). A linkage study between the National Joint Registry, Hospital
Episodes Statistics and records of the Office for National Statistics
on deaths was undertaken. Patients who underwent elective total
hip arthroplasty between January 2003 and December 2014 with no
past history of cardiac failure were included and stratified as
having either a MoM (n = 53 529) or a non-MoM (n = 482 247) arthroplasty.
The primary outcome measure was the time to an admission to hospital
for cardiac failure or death. Analysis was carried out using data
from all patients and from those matched by propensity score.Aims
Patients and Methods
We sought to determine whether cobalt-chromium alloy (CoCr) femoral
stem tapers (trunnions) wear more than titanium (Ti) alloy stem
tapers (trunnions) when used in a large diameter (LD) metal-on-metal
(MoM) hip arthroplasty system. We performed explant analysis using validated methodology to
determine the volumetric material loss at the taper surfaces of
explanted LD CoCr MoM hip arthroplasties used with either a Ti alloy
(n = 28) or CoCr femoral stem (n = 21). Only 12/14 taper constructs
with a rough male taper surface and a nominal included angle close
to 5.666° were included. Multiple regression modelling was undertaken
using taper angle, taper roughness, bearing diameter (horizontal
lever arm) as independent variables. Material loss was mapped using
a coordinate measuring machine, profilometry and scanning electron
microscopy.Aims
Patients and Methods
A contact patch to rim (CPR) distance of <
10 mm has been
associated with edge-loading and excessive wear. However, not all
arthroplasties with a low CPR distance show problems with wear.
Therefore, CPR distance may not be the only variable affecting the
post-operative metal ion concentrations. We used multiple logistic regression to determine what variables
differed between the patients who had high and low cobalt (CoS)
and chromium (CrS) serum ion concentrations within a cohort of patients
with low (<
10 mm) CPR distances. A total of 56 patients treated
with unilateral hip resurfacing arthroplasty (HRA) had CoS and CrS
ion studies performed more than one year after surgery. The mean
age of the patients at the time of surgery was 51.7 years (29 to
70), with 38 women (68%) and 18 men (32%).Aims
Patients and Methods
To determine ten-year failure rates following 36 mm metal-on-metal
(MoM) Pinnacle total hip arthroplasty (THA), and identify predictors
of failure. We retrospectively assessed a single-centre cohort of 569 primary
36 mm MoM Pinnacle THAs (all Corail stems) followed up since 2012
according to Medicines and Healthcare Products Regulation Agency
recommendations. All-cause failure rates (all-cause revision, and
non-revised cross-sectional imaging failures) were calculated, with predictors
for failure identified using multivariable Cox regression.Aims
Patients and Methods
Our first aim was to determine whether there are significant
changes in the level of metal ions in the blood at mid-term follow-up,
in patients with an Articular Surface Replacement (ASR) arthroplasty.
Secondly, we sought to identify risk factors for any increases. The study involved 435 patients who underwent unilateral, metal-on-metal
(MoM) hip resurfacing (HRA) or total hip arthroplasty (THA). These
patients all had one measurement of the level of metal ions in the
blood before seven years had passed post-operatively (early evaluation)
and one after seven years had passed post-operatively (mid-term evaluation).
Changes in ion levels were tested using a Wilcoxon signed-rank test.
We identified subgroups at the highest risk of increase using a
multivariable linear logistic regression model.Aims
Patients and Methods
We wished to investigate the influence of metal debris exposure
on the subsequent immune response and resulting soft-tissue injury
following metal-on-metal (MoM) hip arthroplasty. Some reports have
suggested that debris generated from the head-neck taper junction
is more destructive than equivalent doses from metal bearing surfaces. We investigated the influence of the source and volume of metal
debris on chromium (Cr) and cobalt (Co) concentrations in corresponding
blood and hip synovial fluid samples and the observed agglomerated
particle sizes in excised tissues using multiple regression analysis
of prospectively collected data. A total of 199 explanted MoM hips
(177 patients; 132 hips female) were analysed to determine rates
of volumetric wear at the bearing surfaces and taper junctions. Aims
Patients and Methods
Hip resurfacing arthroplasty (HRA) is an alternative to conventional
total hip arthroplasty for patients with osteonecrosis (ON) of the
femoral head. Our aim was to report the long-term outcome of HRA,
which is not currently known. Long-term survivorship, clinical scores and radiographic results
for 82 patients (99 hips) treated with HRA for ON over a period
of 18 years were reviewed retrospectively. The mean age of the 67
men and 15 women at the time of surgery was 40.8 years (14 to 64).
Patients were resurfaced regardless of the size of the osteonecrotic
lesion.Aims
Patients and Methods
Joint replacement of the hip and knee remain
very satisfactory operations. They are, however, expensive. The
actual manufacturing of the implant represents only 30% of the final
cost, while sales and marketing represent 40%. Recently, the patents
on many well established and successful implants have expired. Companies
have started producing and distributing implants that purport to
replicate existing implants with good long-term results. The aims of this paper are to assess the legality, the monitoring
and cost saving implications of such generic implants. We also assess
how this might affect the traditional orthopaedic implant companies. Cite this article:
Surgeons have commonly used modular femoral heads and stems from
different manufacturers, although this is not recommended by orthopaedic
companies due to the different manufacturing processes. We compared the rate of corrosion and rate of wear at the trunnion/head
taper junction in two groups of retrieved hips; those with mixed
manufacturers (MM) and those from the same manufacturer (SM). We identified 151 retrieved hips with large-diameter cobalt-chromium
heads; 51 of two designs that had been paired with stems from different
manufacturers (MM) and 100 of seven designs paired with stems from
the same manufacturer (SM). We determined the severity of corrosion
with the Goldberg corrosion score and the volume of material loss
at the head/stem junction. We used multivariable statistical analysis
to determine if there was a significant difference between the two
groups. Aims
Materials and Methods
There are many guidelines that help direct the management of
patients with metal-on-metal (MOM) hip arthroplasties. We have undertaken
a study to compare the management of patients with MOM hip arthroplasties in
different countries. Six international tertiary referral orthopaedic centres were
invited to participate by organising a multi-disciplinary team (MDT)
meeting, consisting of two or more revision hip arthroplasty surgeons
and a musculoskeletal radiologist. A full clinical dataset including
history, blood tests and imaging for ten patients was sent to each
unit, for discussion and treatment planning. Differences in the
interpretation of findings, management decisions and rationale for
decisions were compared using quantitative and qualitative methods.Aims
Methods