header advert
Results 1 - 1 of 1
Results per page:
Orthopaedic Proceedings
Vol. 105-B, Issue SUPP_13 | Pages 66 - 66
7 Aug 2023
Holthof S Amis A Van Arkel R Rock M
Full Access

Abstract

Introduction

Mid-flexion instability may cause poor outcomes following TKA. Surgical technique, patient-specific factors, and implant design could all contribute to it, with modelling and fluoroscopy data suggesting the latter may be the root cause. However, current implants all pass the preclinical stability testing standards, making it difficult to understand the effects of implant design on instability. We hypothesized that a more physiological test, analysing functional stability across the range of knee flexion-extension, could delineate the effects of design, independent of surgical technique and patient-specific factors.

Methods

Using a SIMvitro-controlled six-degree-of-freedom robot, a dynamic stability test was developed, including continuous flexion and reporting data in a trans-epicondylar axis system. 3 femoral geometries were tested: gradually reducing radius, multi-radius and single-radius, with their respective tibial inserts. 710N of compression force (body weight) was applied to the implants as they were flexed from 0–140° with three levels of anterior/posterior (AP) tibial force applied (−90N,0N,90N).