The Morse taper was adapted into orthopaedics as a connecting junction in total hip replacements. The benefit of modular systems includes the adjustment of leg length, offset and the ability to remove the head for acetabular exposure during revision surgery. The design of the Morse taper facilitates the intimate contact of the conical trunnion of the femoral stem (male component), with the conical bore of the femoral head (female component). Design consideration for tapers involve physical parameters (angle, length, diameter and contact points), and manufacturing finish (surface characteristics). Orthopaedic trunnion tapers are not standardized and vary in length, taper angle and base dimension. Variations in the design and surface characteristics of the trunnion, which directly reflects on the interface at the taper junction can influence the likelihood of subsequent wear, corrosion and ultimately effect longevity of the implant. We studied 11 clean un-used commercial stems from five different companies (DePuy, Stryker, Biomet, Wright Medical, Smith and Nephew). Trunnion surface was scanned with an optical profilometer (Bruker ContourGT-I 3D Optical Microscope (Karlsruhe, Germany), an interference microscopes with the capability to analyse 3 dimensional topographical features of materials. Three measurements of each trunnion were taken on the anterior surface at the proximal, mid-point, and distal segments. Each region scan was conducted under a 20x objective with the scan length of 1000 um and a back scan of 500 um.Introduction
Methods and Materials