Variations in component positioning of total hip replacements can lead to edge loading of the liner, and potentially affect device longevity. These effects are evaluated using ISO 14242:4 edge loading test results in a dynamic system. Mediolateral translation of one of the components during testing is caused by a compressed spring, and therefore the kinematics will depend on the spring stiffness and damping coefficient, and the mass of the translating component and fixture. This study aims to describe the sensitivity of the liner plastic strain to these variables, to better understand how tests using different simulator designs might produce different amounts of liner rim deformation. A dynamic explicit deformable finite element model with 36mm Pinnacle metal-on-polyethylene bearing geometry (DePuy Synthes, Leeds, UK) was used with material properties for conventional UHMWPE. Setup was 65° clinical inclination, 4mm mismatch, 70N swing phase load, and 100N/mm spring. Fixture mass was varied from 0.5-5kg, spring damping coefficient was varied from 0-2Ns/mm. They were changed independently, and in combination. Maximum separation values were relatively insensitive to changes in the mass, damping coefficient, or both. The sensitivity of peak plastic strain, to this range of inputs, was similar to changing the swing phase load from 70N to approximately 150N – 200N. Increasing the fixture mass and/or damping coefficient increased the peak plastic strain, with values from 0.15-0.19. Liner plastic deformation was sensitive to the spring damping and fixture mass, which may explain some of the differences in fatigue and deformation results in UHMWPE liners tested on different machines or with modified fixtures. These values should be described when reporting the results of ISO14242:4 testing. Acknowledgements Funded by EPSRC grant EP/N02480X/1; CAD supplied by DePuy Synthes.
The importance of cup position on the performance of total hip replacements (THR) has been demonstrated in Pelvic movement data for walking for 39 unilateral THR patients was acquired (Leeds Biomedical Research Centre). Patient's elected walking speed was used to group patients into high- and low-functioning (mean speed, 1.36(SD 0.09)ms−1 and 0.85(SD 0.08)ms−1 respectively). A computational algorithm (Python3.7) was developed to calculate cup version during gait cycle. Inputs were pelvic angles and initial cup orientation (assumed to be 45° inclination and 7° version, anterior pelvic plane was parallel to radiological frontal plane). Outputs were cup version angles during a gait cycle (101 measurements/cycle). Minimum, maximum and average cup version during gait cycle were measured for each patient. Two-sample t-test (p=0.05) was used to compare groups.Abstract
Objectives
Methods
Mismatch of bearing component centres and tension of soft tissues surrounding the hip joint can lead to component separation during gait cycle and cause the femoral head to contact the rim of an acetabular liner, which could increase wear and shorten lifespan of an implant. This study aims to investigate the contact and wear mechanics of a metal-on-polyethylene hip joint under dynamic separation by using Finite Element Analysis (FEA). A Pinnacle® cup with a Marathon neutral liner 36×56mm with a 45° inclination was constrained by a spring element in the medial-lateral axis. The spring was pre-compressed by 4mm to represent the corresponding translational mismatch of a simulator testing. Archard's law was used to predict wear over one ISO 14242-1 gait cycle. Contact pressure is proportional to the load input during the stance phase, associated with concentric contact condition; it increases threefold just before the swing phase (time C), reaching 46.2MPa, where edge loading occurs. Consequently, separation climbs to 3.54mm, which is comparable to the mathematical prediction (3.34mm) and dynamic FEA (3.2mm). The predicted volumetric wear after this gait cycle is 1.22 × 10–5 mm3. Dynamic separation between femoral head and acetabular liner can result in edge loading, consequently high contact pressure on the edge of a liner. In combination with cyclic loading, fatigue damage could take place and may be worth investigating in the future.
Increased Dkk-1 signaling is associated with OA occurrence and joint microenvironment damage. Interruption of Dkk1 action is beneficial to improve OA knees. Osteoarthritis (OA) is a leading cause of disability and healthcare financial burden for total knee arthroplasty, rehabilitation, and disability. Inappropriate mechanical stress, immunological, or biochemical regulation reportedly disturbs homeostasis among cartilage, synovium and subchondral bone microstructure that contributes to OA pathogenesis. Control of joint-deleterious factor action is an emerging strategy to ameliorate OA-induced joint deterioration. Dickkopf-1 (Dkk-1) is a potent inhibitor for Wnt/β-catenin signaling regulation of tissue development and remodeling in physiological or pathological contexts. Dkk-1 also acts as a master deleterious factor that represses osteoblast differentiation capacity and bone repair. Associations among Dkk-1 expression, chondrocyte fate, synovial fibroblast behavior or OA incidence are merit of characterization.Summary Statement
Introduction